EXTECH 480826

User Guide
Three Axis Electromagnetic Field (EMF) Meter
Model 480826
Introduction
Congratulations on your purchase of Extech’s Model 480826 Electromagnetic Field
(EMF) Meter. The Model 480826 is a battery powered meter that measures and displays
EMF in Gauss and Tesla units with a frequency bandwidth of 30 to 300Hz. The 3 axis
sensor allows for three component (xyz) measurement coverage. The Model 480826 is
specifically designed to determine the magnitude of electromagnetic fields generated by
power lines, computers electric appliances, televisions and many other similar devices.
This meter is shipped fully tested and calibrated and, with proper use, will provide years
of reliable service.
Meter Operation
1. Press the POWER button to turn the meter ON.
2. Press the UNIT button to select either µTesla or mGauss units.
3. If the approximate range of the measurement is known, select the suitable meter range
using the RANGE button. For unknown measurements, start with the highest range
and work down through the ranges until the optimum range is reached.
4. Hold the probe by its handle and move it slowly toward the object under test. If the LCD
display is completely blank or if the low battery symbol appears on the LCD, check the
9V battery.
5. Notice that the field intensity reading increases as you move closer to a field.
6. Use the XYZ button to read the EMF measurement in the X, Y, or Z axis.
7. If the meter’s display indicates a “1” on the left side of the LCD, an overload condition
exists. This indicates that the measured radiation is higher than the capability of the
currently selected range. Find the appropriate range using the RANGE button as
described above.
Measurement Notes
Due to environmental electromagnetic interference the display may show small EMF
values before testing. This is normal and due to the high sensitivity of the meter. Once a
signal is detected by the sensor, the meter will display accurately.
If the object under test is turned off in the middle of testing, the meter reading should fall
close to zero unless a field from another source is detected.
Data Hold feature
To freeze a displayed reading, press the HOLD button. The DH display icon will switch
on. To unlock the display and return to normal operation, press the HOLD button again.
The DH indicator will switch off.
Battery Replacement
When the low battery icon appears on the left corner of the LCD, the 9V battery has fallen
to a critically low voltage level and should be replaced as soon as possible. The battery
compartment cover is located at the bottom rear of the meter. Remove the Phillips head
screw that secures the battery compartment and slide off the battery compartment cover.
Replace the battery and secure the compartment cover before use.
2
480826 V3.0 3/06
Meter Description
1.
Sensor plug shown inserted in meter’s sensor jack
2.
LCD Display
3.
XYZ axis select button
4.
Manual Range button
5.
Power button
6.
Data Hold button
7.
Unit select button
8.
Sensor
10
9.
Sensor grip handle
11
8
1
9
2
5
3
6
10. Tripod mount
12
11. Pull-out tilt stand
12. Battery compartment access
screw
4
7
13
OPEN
13. Battery compartment cover
Specifications
Display
3-1/2 digit (2000 count) LCD
Measurement rate
Approx. 0.4 seconds
Ranges and resolution
20µTesla (0.01) and 200mGauss (0.1)
200µTesla (0.1) and 2000mGauss (1)
2000µTesla (1) and 20,000mGauss (10)
NOTE: 1 µTesla = 10 mGauss
Accuracy
± (4%FS + 3 digits) for 20µTesla and 200mGauss ranges
(stated for 50/60Hz)
± (5%FS + 3 digits) for 200µTesla and 2000mGauss
ranges
Frequency bandwidth
30 to 300Hz
Over-range indication
“1___” is displayed
Operating
Temperature/Humidity
Temperature: 32 to 122 F (0 to 50°C)
Power source
9V Battery
Power consumption
Approx. 2.7mA DC
Dimensions
Meter: 7.6 x 2.6 x 1.2” (195 x 68 x 30mm)
± (10%FS + 5 digits) 2000µTesla & 20,000mGauss ranges
o
RH: 90% max. from 32 to 95°F (0 to 35°C);
80% max. from 95 to 122°F (35 to 50°C)
Probe: 2.8 x 2.3 x 8.7” (70 x 58 x 220mm)
Sensor Cable Length
3 ft (1m) approx.
Weight
16.2 oz. (460g) including probe and battery
3
480826 V3.0 3/06
EMF Exposure
The effect of EMF exposure is a modern day concern. At the time of this writing, to the best
of our knowledge, no standards or recommendations exist regarding limits of EMF
exposure. Exposure limits of 1 to 3mG have been suggested by several international
bodies. Until evidence suggests that there is not a health risk associated with EMF
exposure, common sense would dictate that a practice of minimal exposure be exercised.
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set
forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Service Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2006 Extech Instruments Corporation.
All rights reserved including the right of reproduction in whole or in part in any form.
4
480826 V3.0 3/06