MAGNETROL L35-2N56-JHF

Series 3
Liquid Level Switches
Installation and Operating Manual
Liquid
Level
Switches
ASME
B31.3
Construction
Read this Manual Before Installing
This manual provides information on Series 3 Liquid
Level Switches. It is important that all instructions are read
carefully and followed in sequence. Detailed instructions
are included in the Installation section of this manual.
Conventions Used in this Manual
Certain conventions are used in this manual to convey
specific types of information. General technical material,
support data, and safety information are presented in
narrative form. The following styles are used for notes,
cautions, and warnings.
NOTES
Notes contain information that augments or clarifies
an operating step. Notes do not normally contain
actions. They follow the procedural steps to which
they refer.
Cautions
Cautions alert the technician to special conditions that
could injure personnel, damage equipment, or reduce
a component’s mechanical integrity. Cautions are also
used to alert the technician to unsafe practices or the
need for special protective equipment or specific
materials. In this manual, a caution box indicates a
potentially hazardous situation which, if not avoided,
may result in minor or moderate injury.
WARNINGS
Warnings identify potentially dangerous situations or
serious hazards. In this manual, a warning indicates an
imminently hazardous situation which, if not avoided,
could result in serious injury or death.
Safety Messages
Follow all standard industry procedures for servicing electrical equipment when working with or around high
voltage. Always shut off the power supply before touching
any components.
Low Voltage Directive
For use in Installation Category II, Pollution Degree 2. If
equipment is used in a manner not specified by the
manufacturer, protection provided by the equipment may
be impaired.
Notice of Trademark, Copyright, and Limitations
Magnetrol & Magnetrol logotype are registered trademarks of Magnetrol International.
Copyright © 2010 Magnetrol International,
Incorporated. All rights reserved.
Performance specifications are effective with date of issue
and are subject to change without notice. Magnetrol
reserves the right to make changes to the product
described in this manual at any time without notice.
Magnetrol makes no warranty with respect to
the accuracy of the information in this manual.
Warranty
All Magnetrol mechanical level and flow controls are warranted free of defects in materials or workmanship for five
full years from the date of original factory shipment.
Repair parts are warranted free of defects in materials and
workmanship for one year from the date of shipment.
Materials, specifications, and contents are subject to
change without prior written notice.
If returned within the warranty period; and, upon factory
inspection of the control, the cause of the claim is
determined to be covered under the warranty; then,
Magnetrol will repair or replace the control at no cost to
the purchaser (or owner) other than transportation.
Magnetrol shall not be liable for misapplication, labor
claims, direct or consequential damage or expense arising
from the installation or use of equipment.
There are no other warranties expressed or implied,
except special written warranties covering some
Magnetrol products.
Quality Assurance
The quality assurance system in place at Magnetrol guarantees the highest level of quality throughout
the company. Magnetrol is committed to providing full
customer satisfaction both in quality products and quality
service.
Magnetrol’s quality assurance system
is registered to ISO 9001 affirming its
commitment to known international
quality standards providing the
strongest assurance of product/service
quality available.
Series 3
Liquid Level Switches
ASME B31.3 Construction
Table of Contents
1.0 Introduction
1.1 Principle of Operation ..............................................4
1.2 Operating Cycle ........................................................4
2.0 Installation
2.1 Unpacking ................................................................4
2.2 Critical Alarm Function ............................................5
2.3 Piping .......................................................................5
2.4 Mounting..................................................................5
2.5 Wiring ......................................................................6
3.0 Switch Differential Adjustment
3.1 Low Level Controls ...................................................8
3.2 High Level Controls..................................................9
4.0 Preventive Maintenance
4.1 What To Do............................................................10
4.1.1 Keep Control Clean .....................................10
4.1.2 Inspect Switch Mechanisms, Terminals, and
Connections Monthly ..................................10
4.1.3 Inspect Entire Unit Periodically....................11
4.1.4 Control Head Removal and Installation .......12
4.1.5 Functional Test of Unit ................................13
4.2 Troubleshooting ......................................................14
4.2.1 Check Switch Mechanism ............................14
4.2.2 Check Sensing Unit......................................15
4.3 What To Avoid........................................................16
5.0 Reference Information
5.1 Agency Approvals....................................................17
5.2 Specifications ..........................................................18
5.2.1 Dimensional Specifications — Sealed
Cage Models ................................................18
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.2.2 Dimensional Specifications — Flanged
Cage Models ................................................20
5.2.3 Actuating Levels, Steam Service Ratings
and Specific Gravities ...................................23
5.3 Replacement Parts...................................................25
5.3.1 Sealed Cage Float Models B35, C35,
G35, K35, L35, V35 and Z35
Parts Identification ......................................25
5.3.2 Sealed Cage Float Models B35, C35, G35
K35, L35, V35 and Z35 Part Numbers ......25
5.3.3 Sealed Cage Displacer Model B34
Parts Identification ......................................26
5.3.4 Sealed Cage Displacer Model B34
Part Numbers..............................................26
5.3.5 Flanged Cage Displacer Model C34
Parts Identification ......................................27
5.3.6 Flanged Cage Displacer Model C34
Part Numbers..............................................27
5.3.7 Flanged Cage Float Models B3F, G3F
K3F and Z3F w/Material Code 1
Parts Identification ......................................28
5.3.8 Flanged Cage Float Models B3F, G3F, K3F
and Z3F w/Material Code 1 Part Numbers ..28
5.3.9 Flanged Cage Float Models B3F, G3F, K3F
and Z3F w/Material Code 2 & N
Parts Identification ......................................29
5.3.10 Flanged Cage Float Models B3F, G3F, K3F
and Z3F w/Material Code 2 & N
Part Numbers..............................................29
5.4 Model Numbers
5.4.1 Sealed Cage Models ....................................30
5.4.2 Flanged Cage Models ..................................32
1.0
Introduction
Magnetrol’s Series 3 level switches are float operated units
suitable for use on clean liquid applications for level alarm,
pump control and safety shutdown functions. Units are
designed, fabricated and certified to compliance with
ASME B31.3 specifications.
Return spring
Enclosing tube
(non-magnetic)
Magnet
Attraction
sleeve
Swing-in
position
1.1
The design of Magnetrol float operated level switches is
based upon the principle that a magnetic field will penetrate non-magnetic materials such as 316 stainless steel.
In this case, the float moves a magnetic attraction sleeve
within a non-magnetic enclosing tube and actuates a
switch mechanism. The enclosing tube provides a pressure
seal to the chamber and therefore to the process.
HL
Figure 1
1.2
Switch Tripped
Figure 2
Switch Released
Operating Cycle
As the liquid level rises in the chamber (refer to Figure 1),
the float moves the magnetic attraction sleeve up within
the enclosing tube and into the field of the switch mechanism magnet. As a result, the magnet is drawn in tightly to
the enclosing tube causing the switch to trip, “making” or
“breaking” an electrical circuit. As the liquid level falls, the
float drops and moves the attraction sleeve out of the magnetic field, releasing the switch at a predetermined “low
level” (refer to Figure 2). The tension spring ensures the
return of the switch in a snap action.
Swing-out
position
LL
Principle of Operation
2.0
Installation
2.1
Unpacking
Unpack the instrument carefully. Inspect all units for
damage. Report any concealed damage to carrier within
24 hours. Check the contents of the packing slip and
purchase order. Check and record the serial numbers and
model numbers for future reference when ordering parts.
Serial # _____________________________________
Model # _____________________________________
4
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
2.2
Critical Alarm Function
It is recommended that for critical alarm functions, an
additional level switch be installed as a high–high or
low–low level alarm for maximum protection.
Conduit
outlet
Pressure
vessel
2.3
Piping
shows a typical piping installation of a Magnetrol
Series 3 control to a pressure vessel. Level decals on
control identify the actuation levels for the bottom switch
mechanism of a unit with three switches at minimum
specific gravity. Charts defining actuation levels for a unit
with one switch at different specific gravities can be found
on pages 23–24, Actuating Levels and Specific Gravities,
Section 5.2.3.
Figure 3
Switch
actuating
level
reference
marks
Shutoff
valve
Max. 12"
Use pipe of sufficient strength to support the control. If
necessary, provide a stand or hanger to help support its
weight. All piping should be straight and free of “low
spots” or “pockets” so that lower liquid line will drain
towards the vessel and upper vapor line will drain toward
the control. Shut-off valves are recommended for installation between the vessel and the control. If control is to be
used with a low temperature liquid (one which will “boil”
in the float chamber if outside heat is absorbed), the
chamber and piping should be insulated. Such boiling
in the chamber will cause false level indications.
DO NOT INSULATE SWITCH MECHANISM
HOUSING.
Drain
valve
Figure 3
Typical Piping Arrangement
On controls equipped with pneumatic switch assemblies,
consult bulletin on mechanism furnished for air (or gas)
piping instructions. Refer to the Switch Instruction
Reference Bulletins chart on page 7 to find bulletin
numbers for pneumatic switches.
2.4
Mounting
Caution: If equipment is used in a manner not specified by the
manufacturer, protection provided by the equipment may
be impaired.
Caution: This instrument is intended for use in Installation
Category II, Pollution Degree 2.
Adjust piping as required to bring control to a vertical
position. Magnetrol controls must be mounted within 3°
of vertical. A 3° slant is noticeable by eye, but installation
should be checked with a spirit level on top and/or sides
of float chamber.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5
Controls should be mounted as close to the vessel as possible.
This will result in a more responsive and accurate level
change in the control. Liquid in a long line may be cooler
and more dense than liquid in the vessel causing lower level
indication in the control than actual level in the vessel.
Caution: Operation of all buoyancy type level devices should be
done in such a way as to minimize the action of dynamic
forces on the float or displacer sensing element. Good
practice for reducing the likelihood of damage to the control is to equalize pressure across the device slowly.
2.5
Wiring
NOTE: A switch or circuit breaker shall be installed in close proximity
to equipment and within easy reach of operator. It shall be
marked as the disconnecting device for the equipment.
Screw
Set
Screw
NOTE: For supply connections in installations with ambient temperature up to +158° F (+70° C), use wire with a minimum rating of
+167° F (+75° C) as required by the process conditions.
Installations with ambient temperatures up to +176° F (+80° C)
require wire with a minimum rating of +185° F (+85° C) as
required by the process conditions. Use a minimum of
14 AWG wire for power and ground field wires.
Screw
Figure 4
NEMA 4X/7/9
Caution: Level controls are shipped from the factory with the
enclosing tube tightened and the middle set screw, on the
housing base, locked to the enclosing tube. Failure to
loosen the set screw prior to repositioning the conduit
connection may cause the enclosing tube to loosen,
resulting in the possible leakage of the process liquid
or vapor.
Internal Circuit
(Left) Switch
Close on high level
4
1
5
2
6
3
Internal Circuit
(Right) Switch
Load
Load
1. Loosen set screw(s) at base of switch housing. Refer to
Figure 4.
Close on high level
Common
Common
Close on low level
Close on low level
Load
Load
Line
Series 3 controls are shipped with the conduit entry of
the switch housing placed 180° opposite to the tank configurations to simplify installation in most cases. If this
configuration is appropriate to the installation, proceed to
step 4 to begin wiring the unit. If another configuration is
desired, the switch housing can be easily rotated by first
following steps 1, 2, and 3.
2. Switch housing may be rotated 360° to allow correct positioning of conduit outlet.
Line
3. Tighten set screw(s) at base of switch housing.
Figure 5
Wiring diagram for DPDT
Series B, C, D, F, R, 8 & 9 switches
6
4. Unscrew and remove switch housing cover. The threads
have been lubricated to facilitate removal.
NOTE: On high temperature applications (above +250° F [+121° C] in
float chamber), high temperature wire should be used
between control and first junction box located in a cooler area.
On non-hazardous applications, flexible conduit may be used
between the control and the first junction box.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5. The switch terminals are located next to the conduit outlet
to facilitate wiring. Bring supply wires through conduit
outlet. Route extra wire around enclosing tube under the
baffle plate, and connect them to the proper terminals.
Refer to the wiring diagram in your switch bulletin for this
information. See Switch Instruction Reference Bulletins
chart below for switch instruction bulletin numbers.
Switch Instruction Reference Bulletins
Switch Series
Letter
B, C, D
Description
Bulletin
No.
Dry Contact Switch
42-683
High Temperature Hermetically Sealed
Snap Switch
42-683
HS, H1
Hermetically Sealed Snap Switch
42-694
R, 8, 9
High Temperature Snap Switch
42-799
J
Bleed Type Pneumatic Switch
42-685
K
Non-Bleed Type Pneumatic Switch
42-686
F
NOTE: For models with a Series HS switch with high temperature
lead wire, the leads are routed out through the conduit opening by the factory. A suitable conduit box should be provided
for the connection of the leads to the control wiring.
6. Dress wiring to ensure no interference or contact with the
switch housing cover.
OBSERVE ALL APPLICABLE ELECTRICAL CODES
AND PROPER WIRING PROCEDURES.
Prevent moisture seepage into the enclosure by installing
approved seal-drain fittings in the conduit run leading into
the unit.
Caution: In hazardous areas, do not power the unit until the
conduit is sealed and the enclosure cover is screwed
down securely.
7. Replace housing cover.
8. If control has been furnished with an explosion proof or
moisture proof (gasketed) switch housing, it must be
sealed at the conduit outlet with a suitable compound or
non-hardening sealant to prevent entrance of air.
9. Test switch action by varying liquid level in float chamber.
NOTE: If switch mechanism fails to function properly, check vertical
alignment of control housing and consult installation bulletin
on switch mechanism furnished.
10. Check cover to base fit to be certain gasketed joint is tight.
A positive seal is necessary to prevent infiltration of moisture laden air or corrosive gases into switch housings.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
7
3.0
Switch Differential Adjustment
The standard differential of Series 3 float models with
one switch mechanism may be field-adjusted. Adjustment
may be necessary if a wider differential needs to be set to
overcome switch chatter caused by the process.
The differential, or the amount of level travel between
“switch-on” and “switch-off ,” may be adjusted by
repositioning the lower jam nuts on the float stem.
This adjustment is different for high level and low level
controls. Refer to the appropriate section below for
adjustment instructions.
Caution: Maximum differential adjustment is 1 inch (25 mm).
Differential adjustment may be done only on models with
a single switch mechanism.
3.1
Low Level Controls
On low level controls the switch trips on the lower actuation
point and resets on the higher actuation point. Widening
the differential will allow the switch to trip on the original
actuation point and reset at a later, or higher, point.
Switch housing
cover
Enclosing tube
Refer to Figures 7 and 8
Conduit
connection
Housing base
Enclosing tube nut
NOTE: To widen the differential one inch, the lower jam nuts must be
set proportionately lower on the stem (i.e., in this example
1 inch).
Chamber
Float
Figure 6
8
The differential on low level controls may be adjusted by
repositioning the lower jam nuts on the float stem. The
standard factory setting is for a minimum amount of play
(gap) between the top jam nuts and the attraction sleeve as
shown in Figure 7 on page 9.
1. Determine what change in differential is necessary.
2. Make sure power source is turned off.
3. NEMA 4X/7/9—Unscrew and remove switch housing cover.
4. Disconnect power supply wires from switch mechanism.
Pull wires out of conduit connection opening in housing
base. Refer to Figure 6.
5a. Perform system shut-down procedures as required to
relieve pressure from float chamber of control. Allow unit
to cool.
5b. Close shut-off valves (if so equipped) to isolate control
from tank. Drain off liquid in float chamber. Refer to
Figure 3 on page 5.
5c. On installations without shut-off valves, relieve pressure
from the tank. Drain liquid in tank to a level below the
connections of the float chamber.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
NOTE: Level control, connections and pipe lines need not be
removed from the tank.
6. Loosen enclosing tube nut with a 15⁄16" wrench. Unscrew
Slight
enclosing tube counterclockwise (switch and housing base
play (gap)
will rotate also) until it is free. Refer to Figure 6 on page 8.
Must be
Replace in same position
allowed
7. Lift enclosing tube, switch, and base off float chamber.
D
(0.03"
typical)
Jam nuts and attraction sleeve are now accessible.
Maximum gap
1.00
setting (applies to
(25 mm)
8. Measure the distance “D” from the top edge of the upper
models having a
Position of
single switch
jam nuts to the top of the float stem. Refer to Figure 8.
bottom
Drop bottom
mechanism with
jam nuts
jam
nuts
to
Record this measurement.
a single magnet
(normal
increase gap
actuator only)
factory
setting (refer 9. Loosen and remove upper jam nuts, guide washer, and
setting)
to above
attraction sleeve.
instructions)
10. Loosen and adjust lower jam nuts to the desired position.
Tighten lower jam nuts securely. Refer to Figure 8.
Figure 8
Figure 7
11. Replace attraction sleeve on stem.
12. Replace upper jam nuts and guide washer on the stem in
the position previously noted. Tighten upper jam nuts
securely. Refer to Figure 8.
NOTE: Use a new enclosing tube gasket when reassembling enclosing tube to the chamber. Make certain that all gasket surfaces
are thoroughly cleaned to allow proper gasket seating. Coat
enclosing tube threads with “anti-seizing” compound.
13. Replace enclosing tube, switch, and base on chamber.
Screw tube clockwise until tightened to 200–225 footpounds of torque.
14. Rotate switch housing to correct position and tighten set
screw at base of switch housing. Refer to Figure 4 on
page 6.
15. Bring supply wires through conduit outlet. Follow steps
5 through 10 in Wiring, Section 2.5 on pages 6 and 7.
16. Test switch action by varying liquid level in float chamber.
NOTE: If switch mechanism fails to function properly, check vertical
alignment of control housing and consult installation bulletin on
switch mechanism. If the unit still fails to function properly, consult the factory.
3.2
High Level Controls
On high level controls the switch trips on the higher actuation point and resets on the lower actuation point.
Caution: On high level controls, widening the differential requires
raising the trip point a proportional amount. The reset
point will remain the same.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
9
To widen the differential by raising the trip point, follow
steps 1 through 16 in Low Level Controls, Section 3.1,
pages 8–9).
Caution: After increasing gap setting, be certain to check for
proper operation of switch mechanism by raising and lowering float assembly. Magnet must “snap” cleanly with
additional float movement available after magnet snaps.
4.0
Preventive Maintenance
Periodic inspections are a necessary means to keep your
level control in good working order. This control is a safety
device to protect the valuable equipment it serves. A systematic program of “preventive maintenance” must be
implemented when the control is placed into service. If the
following sections on “what to do” and “what to avoid” are
observed, your control will provide reliable protection of
your equipment for many years.
4.1
What To Do
4.1.1 Keep Control Clean
Be sure the switch housing cover is always in place on the
control. This cover is designed to keep dust and dirt from
interfering with switch mechanism operation. In addition,
it protects against damaging moisture and acts as a safety
feature by keeping bare wires and terminals from being
exposed. Should the housing cover or any seals become
damaged or misplaced, obtain a replacement immediately.
4.1.2 Inspect Switch Mechanisms, Terminals, and
Connections Monthly
1. Dry contact switches should be inspected for excessive
wear on actuating lever or misalignment of adjustment
screw at point of contact between screw and lever. Such
wear can cause false switch actuating levels. See switch
mechanism bulletin supplied with control should switch
adjustment or replacement be necessary.
2. DO NOT operate your control with defective or maladjusted switch mechanisms. Refer to the Switch
Instruction Reference Bulletins chart on page 7 to find
appropriate bulletins for service instructions.
3. Level controls may sometimes be exposed to excessive
heat or moisture. Under such conditions, insulation on
electrical wiring may become brittle, eventually breaking
or peeling away. The resulting “bare” wires can cause
short circuits.
NOTE: Check wiring carefully and replace at the first sign of brittle
insulation.
10
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
4. Vibration may sometimes cause terminal screws to loosen.
Check all terminal connections to be certain that screws
are tight.
5. On units with pneumatic switches, air (or gas) lines
subjected to vibration, may eventually crack or loosen at
connections causing leakage. Check lines and connections
carefully and repair or replace, if necessary.
NOTE: Spare switches should be kept on hand at all times.
4.1.3 Inspect Entire Unit Periodically
Isolate control from vessel. Raise and lower liquid level to
check for switch contact and reset.
Figure 9
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
11
4.1.4 Control Head Removal and Installation
Inspection of the interior of the chamber is possible on
flanged cage models. To do this, the control head must be
removed to provide proper access to the chamber. This
procedure will assure that the head assembly is removed
and reinstalled properly.
1. Allow vessel to get to safe temperature and pressure before
opening the pressure boundary.
2. Power down all wiring to the unit.
3. Remove all wiring and conduit from the unit.
4. Remove flange bolts.
5. Prepare an area where the head assembly can be placed
such that it rests on the flange in its normal orientation
(two 2×4 boards placed across an open drum work well).
Refer to Figure 9 on page 11.
6. Carefully lift the head assembly by holding the sides of the
flange. Maintain the head assembly as much as possible in
the vertical position.
NOTE: Care must be taken not to place side force on the float which
could bend the stem.
7. With the unit resting in its temporary fixture in an upright
position, it can now be inspected or repaired.
Top jam
nuts
Bottom jam
nuts
Attraction
sleeve
Enclosing tube
(ref.)
Dim. A
0.03"
Minimum
gap setting
Head
flange
(ref.)
Stop tube
Sleeve
stop strap
Retaining
screws
Float and stem
assembly
Caution: Do not place the unit on its side — this could result in
damage to the stem.
8. Conduct a check for float stem straightness as follows:
8.1 Remove the stem assembly including attraction sleeve
from the head assembly by removing the two screws
holding the stop strap to the bottom of the flange.
See Figure 10 for assembly schematic.
8.2 Check the stem straightness by aligning the stem
against a straight edge. Stems that are bent in excess of
1
⁄8 inch over any 6 inch length should not be used
in this condition. Replace any bent stems. See
Replacement Parts, Section 5.3 beginning on page 25.
8.3 Ensure that the components in the stem assembly are
in good condition and surfaces are clear of potential
obstructions. The stem assembly is reinstalled by
securing the stop strap to the flange. Ensure that the
attraction sleeve moves freely through the enclosing
tube and that the stem assembly surfaces are smooth.
NOTE: Care must be taken during the straightness check to be certain that the attraction sleeve is not removed from the stem
and jam nuts settings are not changed.
Figure 10
12
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
9. In the container that the unit is set on, raise the liquid
level sufficiently to lift the float and trip the switch mechanism. Inspect inside of chamber for any obstructions or
scale buildup. If excessive buildup is present, the interval
between inspections should be shortened. Remove old gasket such that mating flange surfaces are completely clean.
Position new gasket on chamber flange before replacing
the head assembly.
NOTE: The float is restricted from moving laterally by the chamber, in
the test configuration the float and stem can move further off
center than possible under normal operation. Avoid excess
movement in this lateral direction during the test. If switch
trips, the unit may be reassembled.
10. Replace the assembly onto the float cage keeping it as
vertical as possible. Carefully lower the assembly into the
chamber making sure that the stem does not become bent
during installation. Refer to Figure 9 on page 11.
11. Tighten flange bolts per torque values as indicated in table
below.
Model
Flange Bolting
B3F – 150#
100 – 120 ft-lbs
B3F – 300#, G3F – 150#, G3F – 300#
180 – 200 ft-lbs
B3F – 600#
300 – 320 ft-lbs
G3F – 600#, Z3F – 600#
470 – 490 ft-lbs
K3F – 600#
675 – 710 ft-lbs
Z3F – 900#
775 – 850 ft-lbs
Enclosing Tube
200–225 ft-lbs
12. Attach wiring and connect power supply as described in
Wiring, Section 2.5 on page 6.
4.1.5 Functional Test of Unit
This procedure will assure that the complete unit is
functioning properly. This test should be conducted after
switch and wiring inspections have been completed.
1. Adjust piping as required to bring control to a vertical
position. Control must be mounted within 3° of vertical.
Installation should be checked with an indicating level on
top or sides of float chamber.
2. Check that the chamber is mounted within 3° of vertical
after vessel installation is complete.
3. Raise level in the vessel sufficient to raise the float and
trip the switch. Lower level in the vessel sufficient to
lower float and reset the switch. Confirm switch trips and
resets at expected levels (contact Magnetrol for specific
information on trip points; have the serial number of the
unit readily available).
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
13
4.2
Troubleshooting
Usually the first indication of improper operation is failure
of the controlled equipment to function, i.e., pump will
not start (or stop), signal lamps fail to light, etc. When
these symptoms occur, whether at time of installation or
during routine service thereafter, check the following
potential external causes first:
• Fuses may be blown
• Reset button(s) may need resetting
• Power switch may be open
• Controlled equipment may be faulty
• Wiring leading to control may be defective
If a thorough inspection of these possible conditions fails
to locate the trouble, proceed next to a check of the
control's switch mechanism.
4.2.1 Check Switch Mechanism
1.
2.
3.
4.
5.
6.
7.
8.
Pull switch or otherwise disconnect power to the control.
Remove switch housing cover.
Disconnect power wiring from switch assembly.
Swing magnet assembly in and out by hand to check carefully for any sign of binding. Assembly should require
minimal force to move it through its full swing.
If binding exists, magnet may be rubbing enclosing tube.
If magnet is rubbing, loosen magnet clamp screw and shift
magnet position. Retighten magnet clamp screw.
If switch magnet assembly swings freely and mechanism
still fails to actuate, check installation of control to be
certain it is within the specified 3° of vertical.
Examine the switch closely as described in Inspect Switch
Mechanisms, Terminals, and Connections Monthly, Section
4.1.2, step 1 on page 10. Check continuity of switch with
ohmmeter. If switch does not function properly, replace it
immediately.
If switch mechanism is operating satisfactorily, proceed to
check sensing unit.
NOTE: Spare switches should be kept on hand at all times.
14
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
4.2.2 Check Sensing Unit
Shutoff valve
(if used)
Conduit
outlet
1. Reconnect power supply. Carefully actuate the switch
mechanism manually (use a non-conductive tool on electrical switch mechanisms) to determine whether controlled
equipment will operate.
Caution: With electrical power on, avoid contact with switch leads
and connections at terminal block.
Switch
actuating
level reference
mark
Optional liquid leg
connection to vessel
Shutoff valve
(if used)
Optional blow-down
or drain valve
Figure 11
2. If controlled equipment responds to manual switch actuation test, trouble may be located in the level sensing portion of the control, float(s), stem(s), and magnetic attraction sleeve(s).
3. Check to be certain liquid is entering float chamber.
A valve may be closed or piping plugged.
Caution: Be certain to pull disconnect switch or otherwise ensure
that electrical circuit(s) through control is deactivated.
Close operating medium supply valve on controls
equipped with pneumatic switch mechanisms.
4. Disconnect wiring from supply side of switch mechanism(s)
and remove electrical conduit or operating medium line
connections to switch housing.
5. Perform system shutdown to relieve pressure from float
chamber of control and allow unit to cool.
6. Close shutoff valves (if equipped) to isolate control from
vessel. Drain off liquid in chamber if necessary
7. On installations without shutoff valves, relieve pressure
from vessel and drain off liquid head above control
mounting level.
NOTE: Control chamber, connections, and pipe lines need not be
removed from vessel or boiler.
8. Remove switch housing assembly by loosening enclosing
tube hex nut, which is located immediately below housing
base. Refer to Figure 6 on page 8.
9. With switch housing assembly removed, inspect attraction
sleeve(s) and inside of enclosing tube for excessive corrosion
or solids buildup, which could restrict movement, preventing sleeve(s) from reaching field of switch magnet(s).
10. If differential has been changed in the field by repositioning the lower jam nuts on the float stem, check tightness
and position of the jam nuts.
11. Check float to be certain it is buoyant in the liquid (float
chamber or vessel must have adequate liquid level). If float
is determined to be filled with liquid, or it is collapsed, it
must be replaced immediately.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
15
Caution: Do not attempt to repair a float. See Replacement Parts,
Section 5.3 beginning on page 25.
If all components in the control are in operating condition,
the trouble is likely located external to the control. Repeat
inspection of external conditions previously described.
NOTE: If difficulties are encountered which cannot be identified,
consult the factory or your local representative for assistance.
A complete description of the trouble should be provided
along with information concerning your piping and mounting
arrangement, plus a description of your operating sequence.
Sketches or photographs showing the installation are also
beneficial.
When communicating about your control, be certain to
always specify the complete Model and Serial numbers.
4.3
What To Avoid
1. Never leave switch housing cover off the control longer
than necessary to make routine inspections.
2. Never place a jumper wire across terminals to “cut-out”
the control. If a “jumper” is necessary for test purposes, be
certain it is removed before placing control into service.
3. Never attempt to make adjustments or replace switches
without reading instructions carefully. Certain adjustments
provided for in level controls should not be attempted in
the field. When in doubt, consult the factory or your local
representative.
4. Never use lubricants on pivots of switch mechanisms.
A sufficient amount of lubricant has been applied at the
factory to ensure a lifetime of service. Further oiling is
unnecessary and will only tend to attract dust and dirt
which can interfere with mechanism operation.
Caution: Operation of all buoyancy type level devices should be
done in such a way as to minimize the action of dynamic
forces on the float or displacer sensing element. Good
practice for reducing the likelihood of damage to the
control is to equalize pressure across the device very
slowly.
16
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.0
Reference Information
5.1
Agency Approvals
AGENCY
APPROVED MODEL
FM
All with an electric switch mechanism and a housing
listed as Type 4X/7/9
All with an electric switch mechanism and a housing
listed as Type 4X/7/9 Class I, Div 1, Group B
CSA
All with a Series HS, H1, F, 8 & 9 electric switch mechanism
and a housing listed as CSA Type 4X
All with an electric switch mechanism and a housing
listed as Type 4X/7/9
ATEX / IEC Ex CE
All with an electric switch mechanism and a housing
listed as Type 4X/7/9 Class I, Div 1, Group B
All with an electric switch mechanism and an
ATEX housing Low Voltage Directives 2006/95/EC
Per Harmonized Standard:
EN 61010-1/1993 & Amendment No. 1
APPROVAL CLASSES
Class I, Div 1, Groups C & D
Class II, Div 1, Groups E, F & G
Class I, Div 1, Groups B, C & D
Class II, Div 1, Groups E, F & G
Class I, Div 2, Groups B, C & D
Class I, Div 1, Groups C & D
Class II, Div 1, Groups E, F & G
Class I, Div 1, Groups B, C & D
Class II, Div 1, Groups E, F & G
ATEX II 2 G EEx d IIC T6
94/9/EC
IEC Ex Ex d IIC T6
IP 66
Installation Category II
Pollution Degree 2
Controls with two “HS” or “H1” switches are not ATEX approved.
IEC Installation Instructions:
The cable entry and closing devices shall be Ex d certified suitable for the conditions of use
and correctly installed.
For ambient temperatures above +55° C or for process temperatures above +150° C, suitable
heat resistant cables shall be used.
Heat extensions (between process connection and housing) shall never be insulated.
Special conditions for safe use:
When the equipment is installed in process temperatures higher than +85° C the temperature
classification must be reduced according to the following table as per IEC60079-0.
Maximum Process
Temperature
Temperature
Classification
< 85° C
T6
< 100° C
T5
< 135° C
T4
< 200° C
T3
< 300° C
T2
< 450° C
T1
These units are in conformity with IECEx KEM 05.0020X
Classification Ex d IIC T6
Tambient -40° C to +70° C
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
17
5.2
Specifications
5.2.1 Dimensional Specifications — Sealed Cage Models
L
M
L
NEMA 1: 8.50 (216)
NEMA 4X/7/9: 10.12 (257)
Group B: 10.12 (257)
Plugged
K
M
K
K
Plugged
K
HL
HL
(Ref) C
LL
(Ref) F
LL
A
D
B
E
Figure 12
Threaded & Socket Weld
Upper Side/Bottom
Figure 13
Flanged Upper Side/Bottom
L
M
K
K
Plugged
Conduit Connections K
HL
(Ref) J
LL
G
Electrical Switches
NEMA 4X/7/9: 1" NPT
Group B:
1" NPT
Pneumatic Switches
1
NEMA 1:
⁄4" NPT
Actuation levels
Outline
Dimensions
NEMA 4X/7/9
NEMA 4X/7/9,
Group B
H
Figure 14
Flanged Side/Side
NEMA 1
L
M
3.87
(98)
5.93
(151)
5.00
(127)
4.62
(117)
Allow 8" (203 mm) over head
clearance for cover removal.
All housings rotatable 360°.
18
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.2
Specifications
5.2.1 Dimensional Specifications – Sealed Cage Models (cont.)
Inches (mm)
CHAMBERS WITH 1- INCH CONNECTIONS — 150 LB. AND 300 LB. ANSI CLASS
1" NPT Threaded & Socket Weld
Model
Code
B34
B35
C35
G35
L35
V35
Z35
K35
A
B
C
8.76
(223)
8.92
(227)
8.92
(227)
9.80
(249)
9.80
(249)
8.92
(227)
9.80
(249)
3.63
(92)
4.15
(105)
4.15
(105)
4.69
(119)
4.69
(119)
4.15
(105)
4.69
(119)
13.84
(352)
15.12
(384)
15.12
(384)
16.17
(410)
16.17
(410)
15.12
(384)
16.17
(410)
A
B
C
12.04
(306)
5.68
(144)
18.92
(481)
1" Flanged Upper Side/Bottom
D
F
E
Std.
14"
Std.
14"
12.87
6.38
17.95
—
—
(327)
(162)
(459)
11.56
14.00
6.69
17.76
20.20
(294)
(356)
(170)
(451)
(513)
11.56
14.00
6.69
17.76
20.20
(294)
(356)
(170)
(451)
(513)
12.31
14.00
7.19
18.68
20.36
(313)
(356)
(183)
(474)
(517)
12.31
14.00
7.19
18.68
20.36
(313)
(356)
(183)
(474)
(517)
11.56
14.00
6.68
17.76
20.20
(294)
(356)
(170)
(451)
(513)
12.31
14.00
7.19
18.68
20.36
(313)
(356)
(183)
(474)
(517)
D
F
E
Std.
16"
Std.
16"
14.56
16.00
8.19
21.45
22.89
(370)
(406)
(208)
(545)
(581)
1" Flanged Side/Side
G
Std.
13.62
(346)
12.21
(310)
12.21
(310)
12.96
(329)
12.96
(329)
12.21
(310)
12.96
(329)
14"
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
G
Std.
15.21
(386)
16"
16.00
(406)
H
6.38
(162)
6.69
(170)
6.69
(170)
7.19
(183)
7.19
(183)
6.68
(170)
7.19
(183)
H
8.19
(208)
J
Std.
18.70
(478)
18.41
(468)
18.41
(468)
19.33
(490)
19.33
(490)
18.41
(468)
19.33
(490)
14"
—
20.20
(513)
20.20
(513)
20.36
(517)
20.36
(517)
20.20
(513)
20.36
(517)
J
Std.
22.10
(561)
16"
22.89
(581)
CHAMBERS WITH 11⁄2 - INCH CONNECTIONS — 150 LB. AND 300 LB. ANSI CLASS
11⁄2" NPT Threaded & Socket Weld
Model
Code*
B35
C35
G35
L35
V35
Z35
K35
A
B
C
8.92
(227)
8.92
(227)
9.80
(249)
9.80
(249)
8.92
(227)
9.80
(249)
4.22
(107)
4.22
(107)
4.75
(121)
4.75
(121)
4.22
(107)
4.75
(121)
15.12
(384)
15.12
(384)
16.17
(410)
16.17
(410)
15.12
(384)
16.17
(410)
A
B
C
12.04
(306)
5.75
(146)
18.92
(481)
11⁄2" Flanged Upper Side/Bottom
D
F
E
Std.
14"
Std.
14"
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
D
F
E
Std.
16"
Std.
16"
15.56
16.00
9.19
22.45
22.89
(395)
(406)
(233)
(570)
(581)
11⁄2" Flanged Side/Side
G
Std.
13.21
(336)
13.21
(336)
13.96
(355)
13.96
(355)
13.21
(336)
13.96
(355)
14"
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
G
Std.
16.21
(412)
16"
16.00
(406)
H
7.69
(195)
7.69
(195)
8.19
(208)
8.19
(208)
7.69
(195)
8.19
(208)
H
9.19
(233)
J
Std.
19.41
(493)
19.41
(493)
20.33
(516)
20.33
(516)
19.41
(493)
20.33
(516)
14"
20.20
(513)
20.20
(513)
20.36
(517)
20.36
(517)
20.20
(513)
20.36
(517)
J
Std.
23.10
(587)
16"
22.89
(581)
CHAMBERS WITH 2 - INCH CONNECTIONS — 150 LB. AND 300 LB. ANSI CLASS
2" NPT Threaded & Socket Weld
Model
Code*
B35
C35
G35
L35
V35
Z35
K35
A
B
C
8.92
(227)
8.92
(227)
9.80
(249)
9.80
(249)
8.92
(227)
9.80
(249)
4.34
(110)
4.34
(110)
4.88
(124)
4.88
(124)
4.34
(110)
4.88
(124)
15.12
(384)
15.12
(384)
16.17
(410)
16.17
(410)
15.12
(384)
16.17
(410)
A
B
C
12.04
(306)
5.88
(149)
18.92
(481)
2" Flanged Upper Side/Bottom
D
F
E
Std.
14"
Std.
14"
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
12.56
14.00
7.69
18.76
20.20
(319)
(356)
(195)
(477)
(513)
13.31
14.00
8.19
19.68
20.36
(338)
(356)
(208)
(500)
(517)
D
F
E
Std.
16"
Std.
16"
15.56
16.00
9.19
22.45
22.89
(395)
(406)
(233)
(570)
(581)
2" Flanged Side/Side
G
Std.
13.21
(336)
13.21
(336)
13.96
(355)
13.96
(355)
13.21
(336)
13.96
(355)
14"
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
G
Std.
16.21
(412)
16"
16.00
(406)
H
7.69
(195)
7.69
(195)
8.19
(208)
8.19
(208)
7.69
(195)
8.19
(208)
H
9.19
(233)
J
Std.
19.41
(493)
19.41
(493)
20.33
(516)
20.33
(516)
19.41
(493)
20.33
(516)
14"
20.20
(513)
20.20
(513)
20.36
(517)
20.36
(517)
20.20
(513)
20.36
(517)
J
Std.
23.10
(587)
16"
22.89
(581)
* Model B34 is not available for chambers with 11⁄2" and 2" connections
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
19
5.2
Specifications
5.2.2 Dimensional Specifications — Flanged Cage Models
L
M
M
L
K
NEMA 1: 8.50 (216)
NEMA 4X/7/9: 10.12 (257)
Group B: 10.12 (257)
K
Plugged
Plugged
K
K
HL
(Ref) J
HL
LL
(Ref) C
LL
G
A
B
H
Figure 15
Threaded & Socket Weld
Upper Side/Bottom
L
Figure 16
Flanged Side/Side
M
K
K
Plugged
Conduit Connections K
Electrical Switches
NEMA 4X/7/9: 1" NPT
Group B:
1" NPT
Pneumatic Switches
1
NEMA 1:
⁄4" NPT
HL
(Ref) F
LL
Actuation levels
D
Outline
Dimensions
NEMA 4X/7/9
NEMA 4X/7/9,
Group B
NEMA 1
E
Figure 17
Flanged Upper Side/Bottom
20
L
M
3.87
(98)
5.93
(151)
5.00
(127)
4.62
(117)
Allow 8" (203 mm) over head
clearance for cover removal.
All housings rotatable 360°.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.2
Specifications
5.2.2 Dimensional Specifications – Flanged Cage Models (cont.)
Inches (mm)
CHAMBERS WITH 1-INCH CONNECTIONS
Head
Model
Flg.
Code Rating
(lbs.)
300
C34
600
150
B3F
300
600
150
G3F
300
600
600
Z3F
900
150
K3F
300
600
1" NPT Threaded
& Socket Weld
1" Flanged
Upper Side/Bottom
A
B
C
15.63
(397)
15.63
(397)
9.12
(232)
9.12
(232)
9.12
(232)
10.12
(257)
10.12
(257)
10.12
(257)
10.11
(257)
10.11
(257)
3.13
(80)
3.13
(80)
3.63
(92)
3.63
(92)
3.63
(92)
4.69
(119)
4.69
(119)
4.69
(119)
4.68
(119)
4.68
(119)
22.59
(574)
22.72
(577)
16.64
(423)
17.32
(440)
18.45
(469)
18.30
(465)
19.12
(486)
20.56
(522)
20.19
(513)
21.13
(537)
A
B
C
11.62
(295)
11.62
(295)
11.62
(295)
5.69
(145)
5.69
(145)
5.69
(145)
20.38
(518)
21.25
(540)
22.93
(582)
D
Std.
18.33
(466)
18.33
(466)
12.06
(306)
12.06
(306)
12.06
(306)
13.06
(332)
13.06
(332)
13.06
(332)
13.06
(332)
13.06
(332)
14"
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
D
Std.
14.56
(370)
14.56
(370)
14.56
(370)
16"
16.00
(406)
16.00
(406)
16.00
(406)
E
5.45
(138)
5.45
(138)
6.13
(156)
6.13
(156)
6.13
(156)
7.19
(183)
7.19
(183)
7.19
(183)
7.19
(183)
7.19
(183)
E
8.19
(208)
8.19
(208)
8.19
(208)
1" Flanged
Side/Side
F
Std.
25.29
(642)
25.42
(646)
19.58
(497)
20.26
(515)
21.39
(543)
21.25
(540)
22.06
(560)
23.50
(597)
23.14
(588)
23.14
(588)
G
14"
21.51
(546)
22.19
(564)
23.32
(592)
22.19
(564)
23.00
(584)
24.44
(621)
24.08
(612)
24.08
(612)
Std.
18.97
(482)
18.97
(482)
12.71
(323)
12.71
(323)
12.71
(323)
13.71
(348)
13.71
(348)
13.71
(348)
13.71
(348)
13.71
(348)
16"
24.77
(629)
25.64
(651)
27.32
(694)
Std.
15.21
(386)
15.21
(386)
15.21
(386)
—
—
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
F
Std.
23.33
(593)
24.20
(615)
25.88
(657)
H
14"
G
16"
16.00
(406)
16.00
(406)
16.00
(406)
5.45
(138)
5.45
(138)
6.13
(156)
6.13
(156)
6.13
(156)
7.19
(183)
7.19
(183)
7.19
(183)
7.18
(182)
7.18
(182)
H
8.19
(208)
8.19
(208)
8.19
(208)
J
Std.
25.93
(659)
26.06
(662)
20.23
(514)
20.91
(531)
22.04
(560)
21.89
(556)
22.71
(577)
24.15
(613)
23.79
(604)
23.79
(604)
14"
—
—
21.51
(546)
22.19
(564)
23.32
(592)
22.17
(563)
23.00
(584)
24.43
(621)
24.08
(612)
24.08
(612)
J
Std.
23.98
(609)
24.85
(631)
26.53
(674)
16"
24.77
(629)
25.64
(651)
27.32
(694)
CHAMBERS WITH 11⁄2-INCH CONNECTIONS
Head
Model
Flg.
Code Rating
(lbs.)
300
C34
600
150
B3F
300
600
150
G3F
300
600
600
Z3F
900
1" NPT Threaded
& Socket Weld
A
B
C
15.63
(397)
15.63
(397)
9.12
(232)
9.12
(232)
9.12
(232)
10.12
(257)
10.12
(257)
10.12
(257)
10.11
(257)
10.11
(257)
3.19
(81)
3.19
(81)
3.69
(94)
3.69
(94)
3.69
(94)
4.75
(121)
4.75
(121)
4.75
(121)
4.75
(121)
4.75
(121)
22.59
(574)
22.72
(577)
16.64
(423)
17.32
(440)
18.45
(469)
18.30
(465)
19.12
(486)
20.56
(522)
20.19
(513)
21.13
(537)
A
B
C
1" Flanged
Upper Side/Bottom
D
Std.
18.33
(466)
18.33
(466)
13.06
(332)
13.06
(332)
13.06
(332)
14.06
(357)
14.06
(357)
14.06
(357)
14.06
(357)
14.06
(357)
Std.
11.62
5.75
20.38
15.56
150
(295)
(146)
(518)
(395)
G3F with 600# flanges
centerlines
only.
11.62 available
5.75 in 16"21.25
15.56
K3F
300
(295)
(146)
(540)
(395)
11.62
5.75
22.93
15.56
600
(295)
(146)
(582)
(395)
14"
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
D
16"
16.00
(406)
16.00
(406)
16.00
(406)
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
E
6.19
(157)
6.19
(157)
7.13
(181)
7.13
(181)
7.13
(181)
8.19
(208)
8.19
(208)
8.19
(208)
8.18
(208)
8.18
(208)
E
9.19
(233)
9.19
(233)
9.19
(233)
1" Flanged
Side/Side
F
Std.
25.29
(642)
25.42
(646)
20.58
(523)
21.26
(540)
22.39
(568)
22.25
(565)
23.06
(586)
24.50
(622)
24.14
(613)
24.14
(613)
G
14"
21.51
(546)
22.19
(564)
23.32
(592)
22.19
(564)
23.00
(584)
24.44
(621)
24.08
(612)
24.08
(612)
Std.
18.97
(482)
18.97
(482)
13.71
(348)
13.71
(348)
13.71
(348)
14.71
(374)
14.71
(374)
14.71
(374)
14.71
(374)
14.71
(374)
16"
24.77
(629)
25.64
(651)
27.32
(694)
Std.
16.21
(412)
16.21
(412)
16.21
(412)
—
—
F
Std.
24.33
(618)
25.50
(648)
26.88
(683)
14"
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
G
16"
16.00
(406)
16.00
(406)
16.00
(406)
H
6.19
(157)
6.19
(157)
7.13
(181)
7.13
(181)
7.13
(181)
8.19
(208)
8.19
(208)
8.19
(208)
8.18
(208)
8.18
(208)
H
9.19
(233)
9.19
(233)
9.19
(233)
J
Std.
25.93
(659)
26.06
(662)
21.23
(539)
21.91
(556)
23.04
(585)
22.89
(581)
23.71
(602)
25.15
(639)
24.79
(630)
24.79
(630)
14"
—
—
21.51
(546)
22.19
(564)
23.32
(592)
22.19
(564)
23.00
(584)
24.44
(621)
24.08
(612)
24.08
(612)
J
Std.
24.98
(634)
25.85
(657)
27.53
(699)
16"
24.77
(629)
25.64
(651)
27.32
(694)
21
5.2
Specifications
5.2.2 Dimensional Specifications – Flanged Cage Models (cont.)
Inches (mm)
CHAMBERS WITH 2-INCH CONNECTIONS
Head
Model
Flg.
Code Rating
(lbs.)
300
C34
600
150
B3F
300
600
150
G3F
300
600
600
Z3F
900
150
K3F
300
600
22
2" NPT Threaded
& Socket Weld
A
B
C
15.63
(397)
15.63
(397)
9.12
(232)
9.12
(232)
9.12
(232)
10.12
(257)
10.12
(257)
10.12
(257)
10.11
(257)
10.11
(257)
3.31
(84)
3.31
(84)
3.81
(97)
3.81
(97)
3.81
(97)
4.88
(124)
4.88
(124)
4.88
(124)
4.87
(124)
4.87
(124)
22.59
(574)
22.72
(577)
16.64
(423)
17.32
(440)
18.45
(469)
18.30
(465)
19.12
(486)
20.56
(522)
20.19
(513)
21.13
(537)
A
B
C
11.62
(295)
11.62
(295)
11.62
(295)
5.88
(149)
5.88
(149)
5.88
(149)
20.38
(518)
21.25
(540)
22.93
(582)
2" Flanged
Upper Side/Bottom
D
Std.
18.33
(466)
18.33
(466)
13.06
(332)
13.06
(332)
13.06
(332)
14.06
(357)
14.06
(357)
14.06
(357)
14.06
(357)
14.06
(357)
14"
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
D
Std.
15.56
(395)
15.56
(395)
15.56
(395)
16"
16.00
(406)
16.00
(406)
16.00
(406)
E
6.25
(159)
6.25
(159)
7.13
(181)
7.13
(181)
7.13
(181)
8.19
(208)
8.19
(208)
8.19
(208)
8.18
(208)
8.18
(208)
E
9.19
(233)
9.19
(233)
9.19
(233)
2" Flanged
Side/Side
F
Std.
25.29
(642)
25.42
(646)
20.58
(523)
21.26
(540)
22.39
(568)
22.25
(565)
23.06
(586)
24.50
(622)
24.14
(613)
24.14
(613)
G
14"
21.51
(546)
22.19
(564)
23.32
(592)
22.19
(564)
23.00
(584)
24.44
(621)
24.08
(612)
24.08
(612)
Std.
18.97
(482)
18.97
(482)
13.71
(348)
13.71
(348)
13.71
(348)
14.71
(374)
14.71
(374)
14.71
(374)
14.71
(374)
14.71
(374)
16"
24.77
(629)
25.64
(651)
27.32
(694)
Std.
16.21
(412)
16.21
(412)
16.21
(412)
—
—
F
Std.
24.33
(618)
25.20
(648)
26.88
(683)
14"
—
—
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
14.00
(356)
G
16"
16.00
(406)
16.00
(406)
16.00
(406)
H
6.25
(159)
6.25
(159)
7.13
(181)
7.13
(181)
7.13
(181)
8.19
(208)
8.19
(208)
8.19
(208)
8.18
(208)
8.18
(208)
H
9.19
(233)
9.19
(233)
9.19
(233)
J
Std.
25.93
(659)
26.06
(662)
21.23
(539)
21.91
(556)
23.04
(585)
22.89
(581)
23.71
(602)
25.15
(639)
24.79
(630)
24.79
(630)
14"
—
—
21.51
(546)
22.19
(564)
23.32
(592)
22.19
(564)
23.00
(584)
24.44
(621)
24.08
(612)
24.08
(612)
J
Std.
24.98
(634)
25.85
(657)
27.53
(699)
16"
24.77
(629)
25.64
(651)
27.32
(694)
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.2
Specifications
5.2.3 Actuating Levels* and Specific Gravities
For float operated units, minimum specific gravities and actuating levels vary depending upon the
material of construction code used with the unit.
NOTE: The minimum specific gravities and actuating levels shown are for single switch units with 1" process
connections only. Minimum specific gravities and levels will change for multistage units and levels
will change for units with 11⁄2 or 2" process connections. Consult factory for these changes.
*The level decals on the control identify the actuation levels for units with three switches at the corresponding minimum specific gravity of the three switch configuration.
Inches (mm)
SEALED CAGE FLOAT MODELS WITH 1-INCH PROCESS CONNECTIONS
Models w/Materials of Construction Code 1
Model
Code
Minimum
Specific Gravity
Actuating Level
@ Minimum Specific Gravity
HL
LL
Models w/Materials of Construction Code 2
Minimum
Specific Gravity
Actuating Level
@ Minimum Specific Gravity
HL
LL
B35
0.69
3.06 (78)
3.62 (92)
0.72
3.06 (78)
3.79 (96)
C35
0.57
3.15 (80)
3.86 (98)
0.59
3.09 (78)
3.96 (101)
G35
0.55
3.25 (82)
3.83 (97)
0.56
3.20 (81)
3.90 (99)
K35
0.40
3.56 (90)
4.06 (103)
0.40
3.40 (86)
4.02 (102)
L35
0.41
3.31 (84)
3.98 (101)
0.43
3.34 (85)
4.14 (105)
V35
0.71
3.06 (78)
3.81 (97)
0.75
3.00 (76)
3.64 (92)
Z35
0.66
2.94 (75)
3.63 (92)
0.68
2.90 (74)
3.47 (88)
Levels ± 0.25" (6 mm)
FLANGED CAGE FLOAT MODELS WITH 1-INCH PROCESS CONNECTIONS
Model
Code
B3F
G3F
K3F
Head
Flange
(Lbs.)
Models w/Materials of Construction Code 1
Minimum
Specific
Gravity
Actuating Level
@ Minimum Specific Gravity
HL
LL
Models w/Materials of Construction Code 2 or N
Minimum
Specific
Gravity
Actuating Level
@ Minimum Specific Gravity
HL
LL
150
0.78
3.00 (76)
3.87 (98)
0.83
2.97 (75)
4.05 (103)
300
0.66
2.17 (55)
2.97 (75)
0.66
2.10 (53)
3.10 (79)
600
0.66
2.06 (52)
3.12 (79)
0.71
2.19 (56)
2.94 (75)
150
0.41
3.58 (91)
4.26 (108)
0.43
3.61 (92)
4.42 (112)
300
0.55
3.52 (89)
4.10 (104)
0.56
3.47 (88)
4.18 (106)
600
0.70
4.00 (102)
4.81 (122)
0.76
4.13 (105)
4.75 (121)
150
0.33
3.08 (78)
3.64 (92)
0.33
2.97 (75)
3.67 (112)
300
0.40
3.05 (77)
3.56 (90)
0.40
2.96 (75)
3.57 (93)
600
0.60
3.30 (84)
3.85 (98)
0.62
3.44 (87)
3.88 (99)
600
0.67
3.10 (79)
3.80 (97)
0.70
3.12 (79)
3.68 (93)
900
0.67
3.10 (79)
3.80 (97)
0.70
3.12 (79)
3.68 (93)
Z3F
Levels ± 0.25" (6 mm)
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
23
5.2
Specifications
5.2.3 Actuating Levels*, Steam Service Ratings and Specific Gravities (cont.)
Inches (mm)
DISPLACER MODELS WITH 1-INCH PROCESS CONNECTIONS
Liquid Specific Gravity
Model
Code
B34
C34
Process
Temp. ° F
(° C)
HL
LL
HL
LL
HL
LL
HL
LL
HL
LL
HL
LL
HL
LL
100
(38)
1.30
(33)
4.77
(121)
2.17
(55)
5.00
(127)
2.73
(69)
5.16
(131)
3.13
(80)
5.28
(134)
3.44
(87)
5.37
(136)
3.67
(93)
5.45
(138)
3.86
(98)
5.51
(140)
200
(93)
1.24
(31)
4.70
(119)
2.13
(54)
4.95
(126)
2.70
(69)
5.11
(130)
3.11
(79)
5.24
(133)
3.42
(87)
5.34
(136)
3.66
(93)
5.41
(137)
3.85
(98)
5.48
(139)
300
(149)
1.18
(30)
4.62
(117)
2.09
(53)
4.88
(124)
2.67
(68)
5.06
(129)
3.09
(78)
5.19
(132)
3.40
(86)
5.30
(135)
3.64
(92)
5.38
(137)
3.84
(98)
5.45
(138)
100
(38)
4.56
(116)
8.44
(214)
5.75
(146)
8.94
(227)
6.56
(167)
9.25
(235)
7.13
(181)
9.50
(241)
7.50
(191)
9.69
(246)
7.88
(200)
9.81
(249)
8.13
(207)
9.94
(252)
200
(93)
4.19
(106)
8.06
(205)
5.50
(140)
8.63
(219)
6.31
(160)
9.00
(229)
6.94
(176)
9.25
(235)
7.38
(187)
9.50
(241)
7.69
(195)
9.63
(245)
8.00
(203)
9.75
(248)
300
(149)
3.81
(97)
7.63
(194)
5.19
(132)
8.25
(210)
6.06
(154)
8.69
(221)
6.69
(170)
9.00
(229)
7.19
(183)
9.25
(235)
7.50
(191)
9.44
(240)
7.81
(198)
9.56
(243)
400
(205)
n/a
n/a
4.81
(122)
7.88
(200)
5.75
(146)
8.38
(213)
6.44
(164)
8.75
(222)
6.94
(176)
9.00
(229)
7.31
(186)
9.25
(235)
7.63
(194)
9.38
(238)
500
(260)
n/a
n/a
4.44
(113)
7.44
(189)
5.44
(138)
8.00
(203)
6.19
(157)
8.44
(214)
6.69
(170)
8.75
(222)
7.13
(181)
9.25
(229)
7.44
(189)
9.19
(233)
0.40
0.50
0.60
0.70
0.80
0.90
1.00
Levels ± 0.25" (6 mm)
24
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.3
Replacement Parts
5.3.1 Sealed Cage Float Models B35, C35, G35, K35, L35, V35 and Z35 Parts Identification
1
4
5
2
6
8
3
7
Figure 18
Housing and Switch Mechanism
9
Item
1
2
3
4
5
6
7
8
9
10
10
Description
Housing cover
Housing base
Switch mechanism
Jam nuts (qty. 4)
Washer
Attraction sleeve
Stop tube
Enclosing tube E-tube gasket
Chamber assembly
Figure 19
Sealed Cage Models B35, C35, G35, K35, L35
5.3.2 Sealed Cage Float Models B35, C35, G35, K35, L35, V35 and Z35 Part Numbers
B35, C35, G35, K35 & L35
Material Code
1
Housing kit
Material Code
2
Material Code
1
Material Code
2
Refer to bulletin on switch mechanism and housing base
(includes items 1 and 2)
Switch mechanism
V35 & Z35
assembly furnished (see page 7)
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
Sleeve kit
89-3426-001
89-3426-002
89-3426-005
89-3426-006
Z32-6346-002
Z32-6346-202
Z32-6346-003
Z32-6346-203
(includes items 4 through 7)
Enclosing tube E-tube gasket
Chamber assembly*
12-1204-001
Available as complete sensing units only
An enclosing tube order must also include an E-tube gasket order.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
* Specify model and serial number of control when ordering.
25
5.3
Replacement Parts
5.3.3 Sealed Cage Displacer Models B34 Parts Identification
4
5
Item
1
2
3
4
5
6
7
8
9
10
6
7
8
9
Description
Housing cover
Housing base
Switch mechanism
Jam nuts (qty. 5)
Washer
Attraction sleeve
Stop tube
Enclosing tube E-tube gasket
Chamber assembly
Figure 20
Sealed Cage Displacer Model B34
5.3.4 Sealed Cage Displacer Models B34 Part Numbers
Housing kit
(includes items 1 and 2)
Switch mechanism
Sleeve kit
(includes items 4 through 6)
Enclosing tube E-tube gasket
Chamber assembly*
All models w/Material Code 1
All models w/Material Code 2
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
89-3426-003
Z32-6346-001
12-1204-001
Available as complete sensing units only
An enclosing tube order must also include an E-tube gasket order.
26
89-3426-004
Z32-6346-201
* Specify model and serial number of control when ordering.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.3
Replacement Parts
5.3.5 Flanged Cage Displacer Model C34 Parts Identification
5
4
7
Item
6
1
Housing cover
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
16
15
10
8 9
14
Description
13
11
12
17
Housing base
Switch mechanism
Jam nuts (qty. 5)
Enclosing tube E-tube gasket
Attraction sleeve
Washer (qty. 3)
Screws (qty. 3)
Spring & stem assembly
Displacer assembly
Retainer
Washer
Cotter pin (qty. 2)
Head flange gasket
Head flange Chamber assembly
Figure 21
Flanged Cage Displacer Model C34
5.3.6 Flanged Cage Displacer Model C34 Part Numbers
Housing kit
(includes items 1 and 2)
Switch mechanism
Jam nuts
Enclosing tube E-tube gasket
Attraction sleeve
Displacer and stem kits
(includes items
8 through 15)
Head flange 300# rating
600# rating
300# rating
600# rating
300#
600#
300#
600#
rating
rating
rating
rating
All models
All models
All models
w/Material Code 1
w/Material Code 2
w/Material Code N
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
10-2106-004
10-2107-001
10-2107-001
Z32-6346-201
Z32-6346-201
Z32-6339-002
Z32-6346-204
Z32-6346-204
Z32-6339-001
12-1204-001
32-4205-002
32-4511-202
32-4203-004
32-4511-201
89-3258-001
89-3258-002
Chamber assembly*
89-3258-003
89-3258-004
Z04-8639-001
Z04-8639-002
Consult factory
An enclosing tube order must also include an E-tube gasket order.
A head flange order must also include a head flange gasket order. Consult factory for part number.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
Consult factory
Consult factory
* Specify model and serial number
of control when ordering.
27
5.3
Replacement Parts
5.3.7 Flanged Cage Float Models B3F, G3F, K3F and Z3F w/Material Code 1 Parts Identification
6
7
8
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
4
9
5
14
13
10
11
12
15
Figure 22
Description
Housing cover
Housing base
Switch mechanism
Enclosing tube E-tube gasket
Jam nuts (qty. 4)
Washer (qty. 2)
Attraction sleeve
Stop tube
Stop strap
Screws (qty. 2)
Float & stem assembly
Head flange gasket
Head flange Chamber assembly
Flanged Cage Models B3F, G3F, K3F and Z3F
5.3.8 Flanged Cage Float Models B3F, G3F, K3F and Z3F w/Material Code 1 Part Numbers
150# Head Flange
Housing kit
Enclosing tube (Code N)
Refer to bulletin on switch mechanism and housing base assembly furnished (see page 7)
Z32-6346-002
Z32-6346-002
Z32-6339-004
Z32-6339-004
Z32-6339-005
12-1204-001
E-tube gasket
(includes items 6
through 13)
Head Flange
Z32-6346-003
Z32-6346-003
n/a
B3F
G3F
12-1204-024
12-1204-012
12-1204-002
n/a
12-1204-025
12-1204-013
12-1204-028
n/a
K3F
Z3F
B3F
G3F
12-1204-026
12-1204-027
12-1204-029
n/a
n/a
12-1204-028
12-1204-047
89-3258-005
89-3258-008
89-3258-011
n/a
89-3258-006
89-3258-009
89-3258-012
n/a
89-3258-007
89-3258-010
89-3258-013
n/a
K3F
Z3F
B3F
G3F
K3F
Z3F
n/a
n/a
n/a
89-3258-023
89-3258-024
Z04-8637-001
Z04-8637-002
Z04-8637-003
n/a
Z04-8637-004
Z04-8637-005
Z04-8637-006
n/a
Z04-8637-007
Z04-8637-008
Z04-8637-009
n/a
n/a
n/a
Z04-8637-006
Z04-8637-013
Chamber assembly*
An enclosing tube order must also include an E-tube gasket order.
A head flange order must also include a head flange gasket order.
Consult factory for part number.
28
900# Head Flange
assembly furnished (see page 7)
Enclosing tube (Codes 1, 2)
Float & Stem Kits
600# Head Flange
Refer to bulletin on switch mechanism and housing base
(includes items 1 and 2)
Switch mechanism
Head Flange Gasket
300# Head Flange
Consult factory
* Specify model and serial number of control when ordering.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.3
Replacement Parts
5.3.9 Flanged Cage Float Models B3F, G3F, K3F and Z3F w/Material Codes 2 & N Parts Identification
6
7
8
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
4
9
5
14
13
10
11
12
15
Description
Housing cover
Housing base
Switch mechanism
Enclosing tube E-tube gasket
Jam nuts (qty. 4)
Washer (qty. 2)
Attraction sleeve
Stop tube
Stop strap
Screws (qty. 2)
Float & stem assembly
Head flange gasket
Head flange Chamber assembly
Figure 23
Flanged Cage Models B3F, G3F, K3F and Z3F
5.3.10 Flanged Cage Float Models B3F, G3F, K3F and Z3F w/Material Codes 2 & N Part Numbers
150# Head Flange
Material Code 2 & N
Housing kit
Enclosing tube (Codes 1, 2)
Enclosing tube (Code N)
(includes items 6
through 13)
Head Flange
900# Head Flange
Material Code 2
Z32-6346-202
Z32-6346-202
Z32-6346-203
Z32-6346-203
Z32-6339-004
Z32-6339-004
Z32-6339-005
n/a
E-tube gasket
Float & Stem Kits
600# Head Flange
Material Code 2 & N
Refer to bulletin on switch mechanism and housing base
assembly furnished (see page 7)
Refer to bulletin on switch mechanism and housing base assembly furnished (see page 7)
(includes items 1 and 2)
Switch mechanism
Head Flange Gasket
300# Head Flange
Material Code 2 & N
B3F
G3F
K3F
Z3F
B3F
G3F
K3F
Z3F
B3F
G3F
K3F
Z3F
12-1204-024
12-1204-001
12-1204-012
12-1204-002
n/a
12-1204-025
12-1204-013
12-1204-028
n/a
12-1204-026
12-1204-027
12-1204-029
n/a
n/a
n/a
12-1204-028
12-1204-047
89-3258-014
89-3258-017
89-3258-020
n/a
89-3258-015
89-3258-018
89-3258-021
n/a
89-3258-016
89-3258-019
89-3258-022
n/a
n/a
n/a
89-3258-025
89-3258-026
Z04-8637-001
Z04-8637-002
Z04-8637-003
n/a
Z04-8637-004
Z04-8637-005
Z04-8637-006
n/a
Z04-8637-007
Z04-8637-008
Z04-8637-009
n/a
n/a
n/a
Z04-8637-006
Z04-8637-013
Chamber assembly*
An enclosing tube order must also include an E-tube gasket order.
A head flange order must also include a head flange gasket order.
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
Consult factory
* Specify model and serial number of control when ordering.
29
5.4
Model Numbers
5.4.1 Sealed Cage Models
MODEL NUMBER CODE
Model
Code
Pressure Rating Min. S.G. for Models
with Material of
Construction Code 1
psig @ ° F
bar @ ° C
2
100
550
750
800
900
1000
38
288
399
427
482
538
—
—
—
—
—
138
68.9
60.0
49.6
41.0
34.5
30.0
27.6
27.2
134
51.7
45.0
41.4
40.9
121
41.4
36.0
33.1
32.7
20.7
154.4
18.0
131.9
16.5
100.3
16.3
83.4
—
—
—
—
—
9.5
138
50.2
9.2
n/a
151.2
131.9
98.3
81.7
49.2
n/a
38
93
149
89.6
84.4
78.9
FLOAT MODELS
B35
0.69
0.72
1000
870
716
594
C35
0.57
0.59
500
435
400
395
G35
0.54
0.56
750
653
600
593
0.40
0.40
600
522
480
474
0.41
0.71
0.43
0.81
300
2240
261
1913
240
1455
237
1210
728
134
n/a
0.66
0.71
2193
1913
1425
1185
713
n/a
100
200
300
1300
1225
1145
K35 L35
V35 Z35 DISPLACER MODEL
B34 0.40
0.40
n/a
9.5
9.2
8.3
n/a
MATERIALS OF CONSTRUCTION
1
2
Carbon steel chamber, 316L stainless steel float, 400 stainless steel sleeve, ASME B31.3 Construction
Carbon steel chamber, 316L stainless steel float, 316 stainless steel sleeve, ASME B31.3 Construction
TANK CONNECTION TYPE AND SIZE
Type
Threaded
Socket Weld
Size
11⁄2" C/C
1"
Std.
Std.
B20
B30
2" C20
C30
D20
D30
Cage Mounting ANSI Flange Rating (lbs.)
Flanged
Upper Side/
Bottom
Flanged
Side/Side
Std.
14"
16"
Std.
14"
16"
150
300
600
900
150
300
600
900
150
300
600
900
N30
N34
N36
S30
S34
S36
N40
N44
N46
S40
S44
S46
N50
N54
N56
S50
S54
S56
N60
N64
N66
S60
S64
S66
P30
P34
P36
T30
T34
T36
P40
P44
P46
T40
T44
T46
P50
P54
P56
T50
T54
T56
P60
P64
P66
S60
S64
S66
Q30
Q34
Q36
V30
V34
V36
Q40
Q44
Q46
V40
V44
V46
Q50
Q54
Q56
V50
V54
V56
Q60
Q64
Q66
V60
V64
V66
PNEUMATIC SWITCH MECHANISM AND ENCLOSURE
Switch
Description
Series J
Bleed Type
Series K
Non-Bleed
Maximum
Supply
Pressure
Maximum
Process
Temperature
Bleed
Orifice
Diameter
Models B35, C35, Models B34, V35,
G35, K35 & L35 Z35 & All Models
with Material of with Material of
Constr. Code 1 Constr. Code 2
psig
bar
°F
°C
inches
mm
NEMA 1
NEMA 1
100
7
400
204
.063
1.6
JGH
JGF
60
4
400
204
.094
2.3
JHH
JHF
60
4
700
371
.055
1.3
JJH
JJF
100
7
400
204
—
—
—
KOF
40
3
400
204
—
—
KOH
—
Minimum SG ratings apply to single stage units. Consult factory for two and
three stage units.
Models are limited to maximum temperature rating of selected switch
mechanism.
Consult factory for 14" center-to-center dimensions on models K35 & B34.
Models V35 and Z35 contain 17–7 ph stainless steel floats.
Consult factory for Model B34 above +300° F (+149° C). Available with
single switch only.
Codes 11⁄2" and 2" NPT tank connections are not applicable to the B34
displacer model.
Codes for a 16" center-to-center dimension are applicable to the K35 model
only.
30
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.4
Model Numbers
5.4.1 Sealed Cage Models (cont.)
ELECTRIC SWITCH MECHANISM AND ENCLOSURE
Switch
Description
Series B
Snap Switch
Series C
Snap Switch
Series D DC Current
Snap Switch
Series F
Hermetically Sealed
Snap Switch
Series HS Hermetically Sealed
5-amp Snap Switch
with Wiring Leads
Series HS Hermetically Sealed
5-amp Snap Switch
with Terminal Block
Series H1 Hermetically Sealed
1-amp Snap Switch
with Wiring Leads
Series R
High Temperature
Snap Switch
Series 8
Hermetically Sealed
Snap Switch
Series 9
High Temperature
Hermetically Sealed
Snap Switch
Switch
Description
Series R
High Temperature
Snap Switch
Series 9
High Temperature
Hermetically Sealed
Snap Switch
Models B35, C35, K35 & L35
Models B34, V35, Z35 & all models
Process with Material of Construction Code 1
with Material of Construction Code 2
Temperature
Set
TYPE 4X/7/9 Aluminum Enclosure
Contacts
Range
Points
ATEX
ATEX
Class I, Div 1 Class I, Div 1
Class I, Div 1 Class I, Div 1
Ex II 2 G EEx
Ex II 2 G EEx
°F (°C)
Groups C&D
Group B
Groups C&D
Group B
d IIC T6
d IIC T6
1
BKA
BKJ
BCC
BKB
BKK
BC9
SPDT
2
BLA
BLJ
BDC
BLB
BLK
BD9
-40 to +250
3
BMA
BMJ
BEC
BMB
BMK
BE9
(-40 to +121)
1
BNA
BNJ
BFC
BNB
BNK
BF9
DPDT
2
BOA
BOJ
BGC
BOB
BOK
BG9
1
CKA
CKJ
CCC
CKB
CKK
CC9
SPDT
2
CLA
CLJ
CDC
CLB
CLK
CD9
-40 to +450
3
CMA
CMJ
CEC
CMB
CMK
CE9
(-40 to +232)
1
CNA
CNJ
CFC
CNB
CNK
CF9
DPDT
2
COA
COJ
CGC
COB
COK
CG9
1
DKB
DKK
DC9
DKB
DKK
DC9
SPDT
2
DLB
DLK
DD9
DLB
DLK
DD9
-40 to +250
3
DMB
DMK
DE9
DMB
DMK
DE9
(-40 to +121)
1
DNB
DNK
DF9
DNB
DNK
DF9
DPDT
2
DOB
DOK
DG9
DOB
DOK
DG9
1
FKA
FKJ
FCC
FKB
FKK
FC9
SPDT
2
FLA
FLJ
FDC
FLB
FLK
FD9
-50 to +750
(-46 to +399)
1
FNA
FNJ
FFC
FNB
FNK
FF9
DPDT
2
FOA
FOJ
FGC
FOB
FOK
FG9
1
HMJ
HMK
HMJ
HMK
SPDT
2
HMN
HMP
HMN
HMP
-50 to +550 N/A
N/A
(-46 to +288)
1
HMS
HMT
HMS
HMT
DPDT
2
HMY
HMZ
HMY
HMZ
-50 to +550 (-46 to +288)
-50 to +750
(-46 to +399)
SPDT
1
HM3
HM4
HA9
HM3
HM4
HA9
DPDT
1
HM7
HM8
HB9
HM7
HM8
HB9
SPDT
1
HKJ
HKK
HKJ
HKK
HKN
HKP
N/A
DPDT
2
HKN
1
2
1
2
1
2
3
1
2
1
2
3
1
2
RKB
RLB
RNB
ROB
8KA
8LA
8MA
8NA
8OA
9KA
9LA
9MA
9NA
9OA
HKP
N/A
RKK
RC9
RKB
RKK
RC9
RLK
RD9
RLB
RLK
RD9
RNK
RF9
RNB
RNK
RF9
DPDT
ROK
RG9
ROB
ROK
RG9
8KJ
8CC
8KB
8KK
8C9
SPDT
8LJ
8DC
8LB
8LK
8D9
-50 to +750
8MJ
8EC
8MB
8MK
8E9
(-46 to +399)
8NJ
8FC
8NB
8NK
8F9
DPDT
8OJ
8GC
8OB
8OK
8G9
9KJ
9CC
9KB
9KK
9C9
SPDT
9LJ
9DC
9LB
9LK
9D9
-50 to +750
9MJ
9EC
9MB
9MK
9E9
(-46 to +399)
9NJ
9FC
9NB
9NK
9F9
DPDT
9OJ
9GC
9OB
9OK
9G9
Cast Iron
Cast Iron
CS/Aluminum
CS/Aluminum
Process Set
Temp. Range Contacts
Class I, Div 1 Class I, Div 1
Class I, Div 1 Class I, Div 1
NEMA 4X
Points NEMA 4X
Groups C&D
Group B
Groups C&D
Group B
°F (°C)
1
R1M
RKM
RKW
R1M
RKM
RKW
SPDT
2
R3M
RLM
RLW
R3M
RLM
RLW
-40 to +1000
(-40 to +538)
1
RDM
RNM
RNW
RDM
RNM
RNW
DPDT
2
REM
ROM
ROW
REM
ROM
ROW
1
9AD
9KD
9KV
9AM
9KM
9KW
SPDT
2
9BD
9LD
9LV
9BM
9LM
9LW
-50 to +1000
3
9CD
9MD
9MV
9CM
9MM
9MW
(-46 to +538)
1
9DD
9ND
9NV
9DM
9NM
9NW
DPDT
2
9ED
9OD
9OV
9EM
9OM
9OW
-40 to +750
(-40 to +399)
SPDT
Process temperature based on +100° F (+38° C) ambient.
On condensing applications, temperature down-rated to +400° F (+204° C)
process at +100° F (+38° C) ambient.
HS & H1 switches can be used with only materials of construction code 1
on models V35 & Z35
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
31
5.4
Model Numbers
5.4.2 Flanged Cage Models
MODEL NUMBER CODE
Model
Code
Min. S.G. for Models with Material of
Construction Code
Head
Flange
ANSI
Class
FLOAT MODELS
150#
B3F
300#
600# 150#
G3F
300#
600# 150#
K3F 300#
600# 600# Z3F
900# DISPLACER MODELS
300#
C34 600# Minimum SG
Pressure Rating psig @ ° F
bar @ ° C
1
2 or N
100
550
750
800
900
1000
38
288
399
427
482
538
0.78
0.66
0.66
0.41
0.55
0.70
0.33
0.40
0.60
0.67
0.67
0.83
0.68
0.71
0.43
0.56
0.76
0.33
0.40
0.62
0.74
0.73
285
740
1000
285
740
1000
285
600
1480
1480
2193
100
740
1480
0.40
155
588
870
155
588
870
155
522
1147
1147
1718
300
655
1310
0.40
95
505
800
95
505
800
95
480
1010
1010
1425
500
600
1310
0.44
80
410
790
80
410
790
80
474
825
825
1185
—
—
—
—
—
—
—
—
20
50
105
20
50
105
20
50
n/a
n/a
n/a
19.6
51.0
68.9
19.6
51.0
68.9
19.6
41.4
102.0
102.0
151.2
100
51.0
102.0
0.40
10.7
39.6
60.0
10.7
45.0
60.0
10.7
36.0
79.1
79.1
118.5
300
45.2
90.7
0.40
6.5
34.8
55.6
6.5
41.1
55.6
6.5
33.1
69.6
69.6
98.3
500
41.1
82.7
0.44
5.5
28.3
54.5
5.5
28.3
54.5
5.5
32.7
56.9
56.9
81.7
—
—
—
—
—
—
—
—
1.4
3.4
7.2
1.4
3.4
7.2
1.4
3.4
n/a
n/a
n/a
n/a
345
345
515
n/a
92.7
92.7
35.5
n/a
MATERIALS OF CONSTRUCTION
1
2
N
Carbon steel chamber, 316L stainless steel float, 400 stainless steel sleeve, ASME B31.3 Construction
Carbon steel chamber, 316L stainless steel float, 316 stainless steel sleeve, ASME B31.3 Construction
Carbon steel chamber, 316L stainless steel float, 316 stainless steel sleeve, NACE and ASME B31.3 Construction
NACE construction: float cage is thermally stress relieved after welding, then hardness tested to assure compliance.
TANK CONNECTION TYPE AND SIZE
Connection
Type
C/C
Size 11⁄2"
11"
2"
ANSI Head Flange Rating (lbs.)
Threaded s/b
Socket Weld s/b
Flanged
Upper Side/
Bottom
Flanged
Side/Side
Std.
Std.
Std.
14" 16" Std.
14" 16" 150
B20
B30
N30
N34
N36
S30
S34
S36
300
B60
B70
N40
N44
N46
S40
S44
S46
600
B80
B90
N50
N54
N56
S50
S54
S56
900
B40
B50
N60
N64
N66
S60
S64
S66
150
C20
C30
P30
P34
P36
T30
T34
T36
300
C60
C70
P40
P44
P46
T40
T44
T46
600
C80
C90
P50
P54
P56
T50
T54
T56
900
C40
C50
P60
P64
P66
T60
T64
T66
150
D20
D30
Q30
Q34
Q36
V30
V34
V36
300
D60
D70
Q40
Q44
Q46
V40
V44
V46
600
D80
D90
Q50
Q54
Q56
V50
V54
V56
900
D40
D50
Q60
Q64
Q66
V60
V64
V66
PNEUMATIC SWITCH MECHANISM AND ENCLOSURE
Switch
Description
Series J
Bleed Type
Series K
Non-Bleed
Maximum
Supply
Pressure
psig
100
50
60
100
40
bar
7
4
4
7
3
Maximum
Process
Temperature
°F
400
400
700
400
400
°C
204
204
371
204
204
Bleed
Orifice
Diameter
inches
.063
.094
.055
—
—
mm
1.6
2.3
1.3
—
—
All Models with Material
of Construction Code 1
except B3F, C34, K34,
G3F & Z3F with 600# or
900# ANSI rating
Model C34 and All Models
with Material of Construction Code 2 or N and B3F,
G3F, K3F & Z3F with
600# or 900# ANSI rating
NEMA 1
JGH
JHH
JJH
—
KOH
NEMA 1
JGF
JHF
JJF
KOF
—
Minimum SG ratings apply to single stage units. Consult factory for two and three stage units.
Models are limited to maximum temperature rating of selected switch mechanism.
600# and 900# flanged cage models must use switch codes listed under heading
“Model C34 & All Models w/Materials of Construction Codes 2 or N and B3F, G3F, K3F &
Z3F with 600# or 900# ANSI rating”.
HS switches can be used with Materials of Construction Code 1 only.
Consult factory for 14" center-to-center dimensions on models K3F and C34.
Construction Code N is not applicable to this unit due to 17–7 ph stainless steel float material.
Consult factory for Model C34 up to +750° F (+399° C).
Example: B3F-1B80-HSG.
32
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
5.4
Model Numbers
5.4.2 Flanged Cage Models (cont.)
ELECTRIC SWITCH MECHANISM AND ENCLOSURE
Switch
Description
Series B
Snap Switch
Series C
Snap Switch
Series D DC Current
Snap Switch
Series F
Hermetically Sealed
Snap Switch
Series HS
Hermetically Sealed
5-amp Snap Switch
with Wiring Leads
Series HS
Hermetically Sealed
5-amp Snap Switch
with Terminal Block
Series H1
Hermetically Sealed
1-amp Snap Switch
with Wiring Leads
Series R
High Temperature
Snap Switch
Series 8
Hermetically Sealed
Snap Switch
Series 9
High Temperature
Hermetically Sealed
Snap Switch
Switch
Description
Series R
High Temperature
Snap Switch
Series 9
High Temperature
Hermetically Sealed
Snap Switch
All models C34 and all models
All models C34 and all models with
with Material of Construction Code 1
Material of Construction Codes 2 & N
except B3F, G3F, K3F & Z3F
and B3F, G3F, K3F & Z3F
Process with 600# or 900# ANSI ratings
with 600# or 900# ANSI ratings
Temperature
Set
Contacts
Range
TYPE 4X/7/9 Aluminum Enclosure
Points
ATEX
ATEX
°F (°C)
Class I, Div 1 Class I, Div 1
Class I, Div 1 Class I, Div 1
Ex II 2 G EEx
Ex II 2 G EEx
Groups C&D
Group B
Groups C&D
Group B
d IIC T6
d IIC T6
1
BKA
BKJ
BCC
BKB
BKK
BC9
SPDT
2
BLA
BLJ
BDC
BLB
BLK
BD9
-40 to +250
3
BMA
BMJ
BEC
BMB
BMK
BE9
(-40 to +121)
1
BNA
BNJ
BFC
BNB
BNK
BF9
DPDT
2
BOA
BOJ
BGC
BOB
BOK
BG9
1
CKA
CKJ
CCC
CKB
CKK
CC9
SPDT
2
CLA
CLJ
CDC
CLB
CLK
CD9
-40 to +450
3
CMA
CMJ
CEC
CMB
CMK
CE9
(-40 to +232)
1
CNA
CNJ
CFC
CNB
CNK
CF9
DPDT
2
COA
COJ
CGC
COB
COK
CG9
1
DKB
DKK
DC9
DKB
DKK
DC9
SPDT
2
DLB
DLK
DD9
DLB
DLK
DD9
-40 to +250
3
DMB
DMK
DE9
DMB
DMK
DE9
(-40 to +121)
1
DNB
DNK
DF9
DNB
DNK
DF9
DPDT
2
DOB
DOK
DG9
DOB
DOK
DG9
1
FKA
FKJ
FCC
FKB
FKK
FC9
SPDT
2
FLA
FLJ
FDC
FLB
FLK
FD9
-50 to +750
(-46 to +399)
1
FNA
FNJ
FFC
FNB
FNK
FF9
DPDT
2
FOA
FOJ
FGC
FOB
FOK
FG9
1
HMJ
HMK
HMJ
HMK
SPDT
2
HMN
HMP
HMN
HMP
-50 to +550 11
N/A
N/A
(-46 to +288)
1
HMS
HMT
HMS
HMT
DPDT
2
HMY
HMZ
HMY
HMZ
-50 to +550 11
(-46 to +288)
-50 to +750
(-46 to +399)
SPDT
1
HM3
HM4
HA9
HM3
HM4
HA9
DPDT
1
HM7
HM8
HB9
HM7
HM8
HB9
SPDT
1
HKJ
HKK
HKJ
HKK
HKN
HKP
N/A
DPDT
2
HKN
1
2
1
2
1
2
3
1
2
1
2
3
1
2
RKB
RLB
RNB
ROB
8KA
8LA
8MA
8NA
8OA
9KA
9LA
9MA
9NA
9OA
HKP
N/A
RKK
RC9
RKB
RKK
RC9
RLK
RD9
RLB
RLK
RD9
RNK
RF9
RNB
RNK
RF9
DPDT
ROK
RG9
ROB
ROK
RG9
8KJ
8CC
8KB
8KK
8C9
SPDT
8LJ
8DC
8LB
8LK
8D9
-50 to +750
8MJ
8EC
8MB
8MK
8E9
(-46 to +399)
8NJ
8FC
8NB
8NK
8F9
DPDT
8OJ
8GC
8OB
8OK
8G9
9KJ
9CC
9KB
9KK
9C9
SPDT
9LJ
9DC
9LB
9LK
9D9
-50 to +750
9MJ
9EC
9MB
9MK
9E9
(-46 to +399)
9NJ
9FC
9NB
9NK
9F9
DPDT
9OJ
9GC
9OB
9OK
9G9
Cast Iron
Cast Iron
CS/Aluminum
CS/Aluminum
Process Set
Temp. Range Contacts
Class I, Div 1 Class I, Div 1
Class I, Div 1 Class I, Div 1
NEMA 4X
Points NEMA 4X
°F (°C)
Groups C&D
Group B
Groups C&D
Group B
1
R1M
RKM
RKW
R1M
RKM
RKW
SPDT
2
R3M
RLM
RLW
R3M
RLM
RLW
-40 to +1000
(-40 to +538)
1
RDM
RNM
RNW
RDM
RNM
RNW
DPDT
2
REM
ROM
ROW
REM
ROM
ROW
1
9AD
9KD
9KV
9AM
9KM
9KW
SPDT
2
9BD
9LD
9LV
9BM
9LM
9LW
-50 to +1000
3
9CD
9MD
9MV
9CM
9MM
9MW
(-46 to +538)
1
9DD
9ND
9NV
9DM
9NM
9NW
DPDT
2
9ED
9OD
9OV
9EM
9OM
9OW
-40 to +750
(-40 to +399)
SPDT
Codes for 150 lb. flanges are not applicable to the C34 displacer model.
Codes for a 16" center-to-center dimension are applicable to the K3F model only.
Process temperature based on +100° F (+38° C) ambient.
11 On condensing applications, temperature down-rated to +400° F (+204° C) process at
+100° F (+38° C) ambient.
12 On condensing applications, temperature down-rated to +400° F (+204° C) process at
+100° F (+38° C) ambient
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
33
NOTES:
34
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
NOTES:
46-622 Series 3 Liquid Level Switches ASME B31.3 Construction
35
ASSURED QUALITY & SERVICE COST LESS
Service Policy
Return Material Procedure
Owners of Magnetrol may request the return of a control
or any part of a control for complete rebuilding or replacement. They will be rebuilt or replaced promptly. Controls
returned under our service policy must be returned by
Prepaid transportation. Magnetrol will repair or replace
the control at no cost to the purchaser (or owner) other
than transportation if:
So that we may efficiently process any materials that are
returned, it is essential that a “Return Material
Authorization” (RMA) number be obtained from the
factory, prior to the material's return. This is available
through Magnetrol’s local representative or
by contacting the factory. Please supply the following
information:
1. Returned within the warranty period; and
2. The factory inspection finds the cause of the claim
to be covered under the warranty.
If the trouble is the result of conditions beyond our control; or, is NOT covered by the warranty, there will be
charges for labor and the parts required to rebuild or
replace the equipment.
In some cases it may be expedient to ship replacement
parts; or, in extreme cases a complete new control, to
replace the original equipment before it is returned. If this
is desired, notify the factory of both the model and serial
numbers of the control to be replaced. In such cases, credit for the materials returned will be determined on the
basis of the applicability of our warranty.
1.
2.
3.
4.
5.
Company Name
Description of Material
Serial Number
Reason for Return
Application
Any unit that was used in a process must be properly
cleaned in accordance with OSHA standards, before it is
returned to the factory.
A Material Safety Data Sheet (MSDS) must accompany
material that was used in any media.
All shipments returned to the factory must be by prepaid
transportation.
All replacements will be shipped F.O.B. factory.
No claims for misapplication, labor, direct or consequential damage will be allowed.
5300 Belmont Road • Downers Grove, Illinois 60515-4499 • 630-969-4000 • Fax 630-969-9489 • www.magnetrol.com
145 Jardin Drive, Units 1 & 2 • Concord, Ontario Canada L4K 1X7 • 905-738-9600 • Fax 905-738-1306
Heikensstraat 6 • B 9240 Zele, Belgium • 052 45.11.11 • Fax 052 45.09.93
Regent Business Ctr., Jubilee Rd. • Burgess Hill, Sussex RH15 9TL U.K. • 01444-871313 • Fax 01444-871317
Copyright © 2011 Magnetrol International, Incorporated. All rights reserved. Printed in the USA.
CSA logotype is a registered trademark of Canadian Standards Association
BULLETIN: 46-622.10
EFFECTIVE: August 2010
SUPERSEDES: June 2009