Installation Note

Installation Note
Agilent PSA Series Spectrum Analyzers
Secure Memory Erase, Option 117
Part Number E4440-90323
Printed in USA July 2006
Notice.
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material, including
but not limited to, the implied warranties of merchantability and fitness for a particular
purpose. Agilent Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing, performance, or use
of this material.
© Copyright 2006 Agilent Technologies, Inc.
Secure Memory Erase, Option 117
Secure Memory Erase, Option 117
Products Affected:
PSA E4440A
PSA E4443A
PSA E4445A
PSA E4446A
PSA E4447A
PSA E4448A
Serial Numbers:
US0000 / US9999
MY0000 / MY9999
SG0000 / SG9999
To Be Performed By:
(X) Agilent Service Center
(X) Personnel Qualified by Agilent
( ) Customer
Estimated Hardware Installation Time:
Estimated Firmware Download and Verification
Time:
30 minutes
1 Hour
Introduction
This installation note explains how to install the hardware and firmware that allows
enhanced security features on the PSA series spectrum analyzers.
The new hardware consists of a memory board where only files that can be modified by
the user are stored. These “user” files consist of screen captures, instrument state and
trace files, limit lines, amplitude correction files, user preset files, mode cache files, and
power on last state files. The instrument main memory is mounted read only, and this
main memory contains the instrument core firmware, the license keywords that enable
the options, the measuring personality files, and the instrument network settings (IP
address, gateway settings, GPIB address, etc). Therefore, to secure the instrument, the
contents of the memory on the optional memory board can be erased without loosing
operational data required to run the PSA.
Once the hardware is installed, it is mandatory to update the PSA firmware so the PSA
can recognize the new hardware.
NOTE
1. The security feature is NOT to be turned on by the retrofit kit installer. As mentioned
above, once the secure erase feature is enabled, all things stored in main memory
cannot be changed by the end user. Therefore, the end user must be given a chance to
complete the instrument configuration such as adding optional measurement
personalities and configuring their I/O settings. When the end user is satisfied with
the instrument configuration, only then will they enable the security feature.
2. The Memory board has a USB port, but this port is only active if Option 111, USB
Installation Note E4440-90323
3
Secure Memory Erase, Option 117
Device Side I/O, is installed.
3. This option is licensed for one instrument model/serial number combination. The
license key will only install on the designated instrument.
Contents
Description
Quantity
Agilent Part
Number
1
Installation Note
This note
1
Option Upgrade Entitlement Certificate
5964-5141
1
User's and Programmers Guide Vol. 1
1
Memory Board
E4440-60252
1
Cover Plate, rear panel
E4440-00061
1
Label (used to cover USB port on cover plate)
E4440-80099
9
Screw, M3 X 0.5 8 mm LG TORX
--
0515-0372
Tools Required
❏ T-20 Torx driver
❏ T-10 Torx driver
❏ PSA Series Spectrum Analyzer Service Guide. This manual is available as part of the
E444xAU Option OBW kits.
❏ Calibration software. Latest software information and downloads available at
http://www.agilent.com/find/calibrationsoftware
❏ Test equipment supported by the calibration software.
❏ Microsoft Windows based personnel computer with LAN card and CD-ROM drive.
❏ Windows 2000 or Windows XP
❏ Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable
with cross pinning, Agilent p/n 8121-0545.
❏ Firmware A.0 9.00 or later. Download the latest revision from
http://www.agilent.com/find/psa_firmware, or order the Firmware Update kit.
Depending on your model number, the firmware update kit is ordered as: Order
E444xAU (Qty.1) Option UE2 (Qty. 1), where the “x” is the last digit of the model
number. For example: E4440AU option UE2.
4
Installation Note E4440-90323
Installation Procedure
Installation Procedure
Preliminary Hardware Verification
1. Check for the presence of Option 111 or Option 115 by pressing System, More, Show
System, and look at the option list.
NOTE
Option 111, Option 115, and Option 117 share the same PC assembly.
Therefore, if the latest version of the PC assembly is currently
installed, you can skip the hardware installation portion of this
procedure. Instruments with serial prefix US/MY/SG 4618 or greater
were factory configured with Option 115.
2. If Option 111 or Option 115 is present, press System, More, Show Hardware. Note the
part number of the USB/Flash Driver.
3. If Option 115 is present it must be uninstalled before installing Option 117.*
• Write down the complete option list from the instrument “Show System” screen by
pressing System, More, Show System. A hard copy of this list is needed because the
list of measurement personalities will disappear from the instrument “Show
System” screen once the Option 115 license is deleted. Your option list may look
something like: BAF, BAC, 215, 219.
Later in the firmware update procedure you will need to assure these options are
chosen for reinstallation.
• Delete the Option 115 license by pressing System, More, More, Licensing, Option.
Use the front panel numerical key pad to enter the option designator 115 and press
the Enter key. Press Delete Option.
• Check that Option 115 has been removed by pressing System, More, Show System,
and looking at the option list.
* Option 115, 512 MB User Memory uses a different memory map and is not
compatible with Option 117. Changing from Option 115 to Option 117 causes the
non-user memory to be reduced from 512 MB to 64 MB, limiting the number of
measurement personalities that can be stored in the instrument.
4. Determine the part number of the PC assembly in the kit. The assembly in the kit
should be the latest part number.
• If Option 111 and/or Option 115 are present and the part numbers match, skip to
the “Install the Option Designator and License Keyword” portion of this
installation note.
• If neither Option 111 nor Option 115 is present, or if the part numbers do not
match, follow the procedure below to install the latest PC assembly.
Installation Note E4440-90323
5
Installation Procedure
Remove the Outer Case
CAUTION
If the instrument is placed on its face during any of the following
procedures, be sure to use a soft surface or soft cloth to avoid damage to
the front panel, keys, or input connector.
1. Disconnect the instrument from ac power.
2. Refer to Figure 1. Remove the two handles on the sides of the instrument as shown.
Use the T-20 driver to loosen the screws that attach each handle (1). Remove the
handles.
3. Remove the four bottom feet (2). Lift up on the tabs on the feet, and slide the feet in
the direction indicated by the arrows.
4. Remove the four screws (3) that hold the rear feet (4) in place.
5. Pull the instrument cover (5) off toward the rear of the instrument.
Figure 1
6
Instrument Outer Case Removal
Installation Note E4440-90323
Installation Procedure
Remove the Top Brace
1. Refer to Figure 2. Use the T-10 driver to remove the top screws (3) (one screw is
under the security label), and the side screws (2) attaching the top brace (1) to the
deck.
2. Remove the top brace from the deck.
Figure 2
Top Brace Removal
Installation Note E4440-90323
7
Installation Procedure
Installing the Retrofit Kit
1. Locate the A6 (SCSI Interface board). When facing the rear panel of the instrument,
the A6 board is in the right side of the instrument.
2. Remove the 6 screws from the rear plate of the A6 SCSI interface board. The plate
will stay attached to the board. See the picture below.
8
Installation Note E4440-90323
Installation Procedure
3. Remove the A6 SCSI interface board. To do this, gently lift the board vertically from
its socket on the PSA motherboard. Once the A6 SCSI interface board is unseated
from the motherboard, it can be removed through the rear panel.
Installation Note E4440-90323
9
Installation Procedure
4. Insert the Memory board as shown below. Make sure to firmly insert the connectors
into the PSA motherboard sockets, but do not exert so much force as to break the
connector.
After installation, the memory board should look like this:
10
Installation Note E4440-90323
Installation Procedure
5. Put the rear plate in place on the PSA rear panel, and screw it into place as shown
below. New screws are included in the kit.
6. Apply the label, included in the kit, over the USB port.
Replace the Top Brace
1. Refer to Figure 2.
2. Carefully position the top brace on the deck. There is an alignment pin on the mid
web/fan assembly that should match up with the alignment hole on the top brace.
Make sure that no coaxial cables will get pinched underneath the brace.
3. Using the T-10 driver, replace the top screws first, and then the side screws after the
top screws are tightened. Torque to 9 inch pounds.
Replace the Outer Case
1. Refer to Figure 1.
2. Slide the instrument cover back onto the deck from the rear. The seam on the cover
should be on the bottom. Be sure the cover seats into the gasket groove in the front
frame.
3. Replace the four rear feet onto the rear of the instrument. Torque to 236 Ncm (21
in-lb).
4. Use the T-20 driver to replace the handles. Torque to 236 Ncm (21 in-lb).
5. Replace the four bottom feet by pressing them into the holes in the case and sliding
them in the opposite direction of the arrows until they click into place. Note that the
feet at the front have the tilt stands.
Installation Note E4440-90323
11
Installation Procedure
Install the Option Designator and License Keyword
NOTE
The option designator 117 and the license keyword must be entered into
instrument memory in addition to the correct firmware before the
hardware will function.
1. Locate the Option Upgrade Entitlement Certificate in the kit and follow the
directions to redeem it. You will receive an email that has the license keyword
required for step 4.
2. Plug in instrument and power up.
3. On the instrument front panel press: System, More, until the Licensing softkey is
visible. Press Licensing and Option. This will activate the alpha editor menu. Use the
front panel numerical keypad to enter the option designator 117. Press the Enter key.
Note that 117 now appears on the Option key.
4. Press License Key. The license key number is a hexadecimal number that will require
the entry of both letters and numbers. Use the alpha editor and the front panel
numerical keypad to enter the license key number. Your entry will appear in the
active function area of the display. If you make a typing error, use the backspace key
to correct the error. Check the license key number you entered. Press Enter, Activate
License.
5. Cycle instrument power and allow instrument to perform the auto align routine.
Install New Instrument Firmware
Download the PSA Update Program and the PSA Firmware Procedure from
http://www.agilent.com/find/psa_firmware
Alternate method:
Install the Firmware Upgrade Kit E4440AU Option UE2.
A complete firmware installation is required even if the instrument currently contains
the latest firmware. If the instrument does contain the latest firmware, do the following:
• At the Choose Components to load screen, check the box for Operating System
Version X.XX for Firmware Version A.XX.XX.
This will allow the instrument memory to be reconfigured.
If Option 115 was previously installed on the instrument then refer to the list of options
that were written down during the Preliminary Hardware Verification step 3 and check
the appropriate boxes on the “Chose Components to Load” screen of the Firmware
Update Program. If a large number of measurement personalities were previously
installed they may no longer all fit.
DO NOT enable this option until the end user has fully configured the instrument, setup
all System, Config I/O settings, added or removed options and licenses, and configured the
desired power-up state. After the security mode is enabled, the main Flash memory
12
Installation Note E4440-90323
Installation Procedure
becomes read-only, rather than read/write memory. This makes it impossible to add or
remove personalities or their license keys, or otherwise change the instrument's
configuration. Refer to “Managing Security” in the PSA User's and Programmer's
Reference, Volume 1 for more information.
Adjustments and Performance Tests Required
None. Installing this memory assembly does not affect instrument calibration.
Verify the Option
Once the license keyword is installed and the firmware updated, the instrument will
automatically update the memory following a power cycle.
Check for the presence of Option 117 by pressing System, Show System and looking for
117 in the Options section of the display.
End of Installation
For assistance, contact your nearest Agilent Technologies Sales and Service Office. To
find your local Agilent office access the following URL:
http://www.agilent.com/find/assist
Installation Note E4440-90323
13
Installation Procedure
14
Installation Note E4440-90323