Model 7541 - Honeywell Test and Measurement Sensors

Model 7541
AC Carrier or DC Strain Gage
Conditioner Indicator
DESCRIPTION
These advanced instruments provide engineering unit display
The alphanumeric readout can display measured and computed
of a strain gage (mV/V) input and a frequency input. They also
data, units of measure, and test status. During setup, it guides
compute power and perform 21 functions including limit checks,
you with English language prompts. There are no manual adjust-
tare, hold and max./min. capture. You needn’t write code or add
ments. To calibrate, enter the full scale value in engineering units
hardware to be up and running a productive test.
and auto-cal takes over. Provides 0.01 % (max.) resolution and
±5 V and/or ±10 V analog outputs at full scale. The keyboard accesses measured data, held data, max./min. data, data spread,
limit status, and/or I/O status without disruption. Password protection may be used, if needed.
FEATURES
Order code
•
May be calibrated in two directions (clockwise and counter-
7541-111
ac excitation, single display
clockwise or tension and compression)
7541-112
ac excitation, dual display (two inputs)
Fast, rock-solid readings with high noise immunity - 2000
7541-115
ac excitation, torque, RPM, hp (displays 2 of the 3)
samples/sec for torque, head, or drawbar force input - 1 ms
7541-101
dc excitation, single display
response for speed
7541-102
dc excitation, dual display (two inputs)
•
One or two channel input
7541-105
dc excitation, torque, RPM, hp (displays 2 of the 3)
•
Six-digit engineering unit with legends and 0.01 % resolu-
•
tion max.
•
RS-232, RS-422, RS-485 serial communication
•
Auto-scaled ±5 V and/or ±10 V analog outputs
•
No pots, batteries, fans, maintenance, or external power
supplies
•
Excitation: ac carrier models, 3.28 kHz, 3 V rms; dc models,
5 V or 10 V dc, user selectable
•
CE approved
Model 7541
General SPECIFICATIONS
Characteristic
Strain gage input
ENVIRONMENTAL SPECIFICATIONS
Measure
Characteristic
Measure
Any 80 ohm to 2K ohm transducer, directly wired or transformer coupled. 4, 6,
or 7 wire circuits are accommodated
Temperature effects, zero
±0.001 % of full scale/°F max.
Temperature effect, span
±0.001 % of full scale/°F max.
Frequency input
Any unidirectional or bidirectional
(quadrature) source including selfgenerating and zero velocity magnetic
pickups, optical encoders, flowmeters,
etc. When used with bidirectional sensors, the conditioner outputs both direction and magnitude
Input impedance and
configuration
Differential or single ended inputs. 100
kohm differential, 50 kohm single ended
Input threshold
(keypad selectable)
10, 20, 50, 100 or 200 mV pk-pk (between inputs) or TTL
Maximum voltage
±130 Vdc or 130 Vrms
Input signal bandwidth
0.001 kHz to 200 kHz (10 mV to 200 mV
pk-pk threshold) 0.001 kHz to 400 kHz
(TTL threshold)
Display ranges and
resolution
Rangeless (use any full scale engineering unit value) with 50 % overrange.
Resolution is 0.01 % of full scale
Low pass filter
(keypad selectable)
20 kHz (-3 dB) or none (this filter is not
available for TTL inputs)
Response time
Greater of: 1 ms, typ (2 ms worst case)
or the input pulse length
Common mode rejection
80 dB (60 Hz), 55 dB (0 kHz to 10 kHz)
Low pass filtering of
sampled data
Unfiltered or four-pole Bessel filter. Cutoff frequencies from 0.1 Hz to 100 Hz in
1-2-5 steps
Overall accuracy
0.01 % of full scale @ 25 °C [77 °F],
0.015 % of full scale @ 5 °C to 50 °C [41
°F to 122 °F]
Excitation supplies
12 V @ 125 mA2 or 5 V @ 250 mA2.
Short circuit (current limit and overvoltage (fuses) protected
Maximum transducer
cable length
500 ft, except 200 ft for 100 ohm or
lower strain gage transducers
Transducer excitation,
ac models
3 Vrms, 3.28 kHz ±0.01 % sine wave.
Regulated and short circuit protected
Transducer excitation,
dc models
5 V or 10 V dc, user selectable, regulated
Sensitivity
1.0 mV/V to 4.5 mV/V with 50 % overrange; automatically scaled
Input resistance
100 mohm in parallel with 33pF
Automatic null
In Phase: ±10 % of full scale (with 50 %
overrange), ±60 % of full scale (with 0 %
overrange). Quadrature: ±1 mV/V
Auto calibration
Dual polarity shunt calibration with provision for CAL resistor feedback
Spurious signal rejection
60 Hz: 120 dB common mode, 100 dB
normal mode. Carrier quadrature: 60 dB
Anti-alias filter
200 Hz, 7-pole Bessel response filter
Low-pass filtering
Four-pole Bessel response digital filter
with 11 cutoff frequencies from 0.1 Hz to
200 Hz in 1-2-5 steps
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Signal-to-noise ratio1
With 1/10/100/200 Hz filters 86/76/66/62
dB @ 1 mV/V full scale and 86/80/72/66
dB @ 5 mV/V FS
Resolution
0.01 % of full scale
Overall accuracy @
25 °C [77 ºF]
0.02 % of full scale, worst case.
DISPLAY SPECIFICATIONS
Characteristic
Measure
System display
Two line by 16 alphanumeric characters, each 0.2 in wide by 0.3 in high.
Backlit LCD with adjustable contrast
Views
Select either two channels, one Channel
with limit status or one channel with I/O
status
Data displayed
Select from current, max., min., spread,
held data and tare value
Data format
Engineering units with six digits (1-2-5
format) and five character, upper or
lower case, user-entered legend/descriptor
2 Honeywell • Sensing and Control
System response (per channel)
SPECIFICATIONS
Characteristic
Measure
Data samp. & max./min.
update rates
2000 Hz (hardware channels), 50 Hz
(CH3 calculation)
Limit checking rate
1000 Hz (hardware channels), 50 Hz
(CH3 calculation)
Logic I/O response time
1 ms (hardware channels), 20 ms (CH3
calculation)
Update rate for each
analog output
1000 Hz
AC Carrier Strain Gage Conditioner Indicator
System SPECIFICATIONS
Logic inputs SPECIFICATIONS
Characteristic
Measure
Characteristic
Measure
System control
All I/O functions can be ordered in any
combination. The pattern function adds
ANDing capabilities
Four Logic inputs
Each with programmable destination,
protected to ±130 Vdc or 130 Vrms
Four Logic type
Input actions/channel
Logic inputs, outputs and internal Matrix
signals control the following actions:
tare, clear tare, hold, clear hold, reset
max./min., clear latched limits, check
limits, do max./mins., apply +CAL
TTL compatible, Scmitt Trigger, low-true
with 47 kohm pull-up. Input current is
-100 µA @ 0 V
Six Logic outputs
Each with programmable source, short
circuit (current and thermal limits) and
overvoltage (fuse) protected.
The following events drive logic outputs
and internal Matrix signals: HI limit, NOT
HI limit, IN limit, NOT IN limit, LO limit,
NOT LO limit, at max., NOT at max., at
min., NOT at min.
Six Logic type
Open collector, low-true. Operating @
24 V (max.) and 0.3 A max sink current
External +5 Vdc power
(on I/O conn.)
250 mA, short circuit (current limit) and
overvoltage (fuse) protected.
Three user-defined patterns
Patterns of logic inputs, outputs and
Matrix signals drive Logic outputs and
internal Matrix signals
Serial communication port
(selectable as RS232, RS422 or RS485)
Supports 32 devices on RS485 port and
1 device on RS232/422
Limit Checking
Each channel has a HI and LO limit
which may be latched or unlatched,
absolute or signed, and with or without
hysteresis. Select either current, max.,
min. spread or held data for limit
checking. Limit violations on any or all
channels can be set to trigger backlight flashing in any of the display view
modes
Baud rate
300 to 38400. Maximum cable length:
4000 ft (RS422/RS485), 50ft (RS232)
120 ohm terminal resistors
(RS485)
User selectable for RXD and TXD
RS422/485 transceivers
Slew-rate limited, short circuit protected
(current and thermal limits)
RS232 drivers
Short circuit protected (current limit)
Serial I/Os
Use a nine-pin D connector. ±15 kV
ESD protected and float (100 kohm)
with respect to earth ground
Commands
Control of all modes, settings and measurements
Non-volatile memory
storage settings
EEPROM, batteries are not used
Dual analog outputs
Each assignable to any of the three
channels are short circuit (current limit)
and overvoltage (fuse) protected
Output impedance/min.
load resis.
< 1 ohm/10 kohm
Full scale
±5 V or ±10 V (user selectable). Resolution is 2 mV @ ±5 V full scale or 4 mV @
±10 V full scale
Overrange
±8.2 V @ ±5 V full scale or ±13.5 V @
±10 V full scale
Nonlinearity
±2 mV @ ±5 V full scale or ±4 mV @
±10 V full scale
Output events/channel
Mechanical SPECIFICATIONS
Characteristic
Measure
Size (W x H x D)
165,1 mm x 73,66 mm x 220,98 mm [6.5
in x 2.9 in x 8.7 in]
Weight
3 lb
Operating temperature
5 °C to 50 °C [41 °F to 122 °F]
Input power
90 Vac to 250 Vac, 50/60 Hz @ 25 VA,
max. Two 2 A/250 V fuses, line filter and
rear power switch
Overall error
(worst case, incl. temp.
effects)
Filter
±5 mV @ ±5 V full scale or ±10 mV @
±10 V full scale
100 Hz, 5 pole Bessel response low
pass filter
Honeywell • Sensing and Control 3
Model 7541
AC Carrier Strain Gage Conditioner Indicator
Notes
1. The ratio expressed in decibels (dB), of full scale (FS) to noise spread.
Measurements are made for a one-minute interval using a 350 ohm
bridge.
2. Both excitation voltages can be used simultaneously with the following
restrictions: 4.8 x (12 V current) + (5 V current) < 700 mA; 12 V current < 125 mA; 5 V current < 250 mA.
3. Applies to strain gage channel only. Frequency measurement is
absolute with enhanced accuracy and only requires user engineering
unit scale.
4. Specification is subject to change without notice.
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.