PM60 Series

MEDICAL
PM60 Series
3 7 - 6 4 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T
FEATURES
•
•
•
•
•
•
•
Medical and ITE approvals
Compact size 2” x4” x1.18”
Single, dual and triple outputs
Wide-range input 90-264 VAC
Low earth leakage current
Level B emissions
RoHS compliant
SPECIFICATIONS
INPUT
OPERATING
Voltage range
90-264 VAC
Frequency
Inrush current
47-63 Hz
30 A @ 115 VAC, or 60 A @ 230 VAC,
at25°C cold start
1.3 A (rms) for 100 VAC
0.7 A (rms) for 240 VAC
150 µA max. @ 264 VAC, 63 Hz
Input current
Earth leakage current
Isolation
Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
OUTPUT
Efficiency
Switching frequency
400,000 hours at full load at 25°C
ambient, calculated per MIL-HDBK-217F
80-88% typical except PM60-31-3A and
PM60-31-5 A at 75% typical
62 KHz tpical
ENVIRONMENT
Operating temperature:
-10°C to +70°C
Storage temperature:
-40°C to +85°C
Relative humidity:
Derating:
5% to 95% non-condensing
Derate from 100% to +50°C linearly to
50% at +70°C
Output voltage
See table
Output current
Ripple and noise
Safety
UL 60601-1, CSA C22.2 No. 601.1, TÜV EN
60601-1,
UL 60950-1, CSA C22.2 No.
60950-1, TÜV EN 60950-1
Line regulation
See table
100 mV peak to peak on 3.3 V & 5.0 V
models, 1% peak to peak on other models
±0.5% maximum at full load
C-Tick
AS/NZS CISPR11 Group1, Class A
Load regulation
See table
EMC standards
Minimum load current
See table
Transient response
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
IEC60601-1-2, EN55011 / EN55022 and
FCC, VCCI: Class B conducted . class B
radiated
Immunity standards
Overvoltage protection
Temperature coefficient
Provided on output #1 only; set at
112–132% of its nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /°C maximum
EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8,
EN61000-4-11
Hold-up time
12 ms minimum at 110 VAC
Overcurrent protection
394
MTBF
STANDARDS AND APPROVALS
MECHANICAL
Dimensions
101x50.8x30mm
Weight
205g
Doc. No. PBC9 Rev.C 28-4-11
PM60 Series
SELECTION TABLE
MODEL(1)
OUTPUT #1
CLASS I
PM60-10A
PM60-12A
PM60-13A
PM60-14A
PM60-18A
PM60-23A
PM60-25A
PM60-31A
PM60-31-3A
PM60-31-5A
PM60-32A
PM60-39A
OUTPUT #2
MAX.
MIN.
MAX. CURRENT
V1 CURREN CURREN AT 5 CFM
T
T
(2)
5V
12 V
15 V
24 V
48 V
+5 V
+5 V
+5 V
+3.3 V
+5 V
+5 V
+5 V
0A
0A
0A
0A
0A
0.5 A
0.5 A
0.5 A
0.8 A
0.5 A
0.5 A
0.5 A
11.0 A
5.0 A
4.3 A
2.7 A
1.35 A
6.0 A
6.0 A
6.0 A
6.0 A
6.0 A
6.0 A
6.0 A
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
8A
8A
8A
8A
8A
8A
8A
TOL.
±2%
±2%
±2%
±2%
±2%
±3%
±3%
±3%
±3%
±3%
±3%
±3%
V2
+12 V
+24 V
+12 V
+5.2 V
+3.3 V
+15 V
+24 V
OUTPUT #3
MIN.
MAX.
CURREN CURREN
T
T
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0.1 A
3.0 A
0.1 A
1.5 A
0.1 A
3.0 A
0.1 A
3.0 A
0A
1.5 A
0.1 A
2.4 A
0.1 A
1.5 A
TOL.
V3
MIN.
MAX.
CURREN CURREN
T
T
±5%
±5% (N/A)
±5% -12 V
±5% +12 V
±5% +12 V
±5% -15 V
±5% -12 V
TOL.
MAX.
OUTPUT
POWER
±5%
±4%
±4%
±4%
±4%
55 W
60 W
64 W
64 W
64 W
55 W
55 W
55 W
37.5 W
37.5 W/47.5 W
55 W
55 W
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A
0A
0A
0A
0A
0A
0.5 A
0.3 A
0.3 A
0.5 A
0.5 A
MEDICAL
3 7 - 6 4 W AT T S - A C / D C S I N G L E & M U LT I P L E O U T P U T
NOTES:
1. Safety approvals are for PCB form only. To order unit with cover fitted, change suffix “A” to “C”.
2. Maximum current of output #1 of multi-output models can be 8 A at 5 CFM forced air provided by user. 3. 47.5 W with 5 CFM forced air provided
by user
4. The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is out of stated limits. All
models may be operated at no-load without damage.
5. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load
ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output.
MECHANICAL SPECIFICATIONS
OUTPUT POWER DERATING CURVE
NOTES:
1. Dimensions shown in inches [mm]
2. Tolerance 0.02 [0.5] maximum
3. Connector CN1: Molex header 09-65-2038 or equivalent, mating with
Molex housing 09-50-1031 or equivalent.
4. Connector CN2: Molex header 09-65-2068 or equivalent, mating with
Molex housing 09-50-1061 or equivalent.
5. Ground tab is 0.25 [6.35] x 0.032 [0.8] fast-on connector.
6. To ensure compliance with level B emissions, connect the two “*“ marked mounting holes with metallic standoffs to chassis.
7. Weight: 205 grams (0.45 lbs.) approx.
PIN CHART
MODEL
PIN
PM60-10A
PM60-14A
PM60-23A
PM60-12A
PM60-18A
PM60-25A
PM60-13A
PM60-31A
PM60-32A
PM60-39A
PM60-31-3A PM60-31-5A
Doc. No. PBC9 Rev.C 28-4-11
1
2
3
4
5
6
+V1
+V1
V1 Return
V1 Return
N.C.
N.C.
V1
V1
Common Return
N.C
V2
V1
V1
Common Return
V3
V2
V1
V1
Common Return
V3
V2
Powering Progress – www.powerbox.com.au
395