MTC30 - ARCH Electronics Corp.

VER : A_1
update : 2015.09.11
AC -DC Medical Power Module
MTC30 SERIES
30 Watts
KEY FEATURES
 Switching Power Module for PCB Mountable
 Fully Encapsulated Plastic Case
 Universal Input Range 90-264VAC
 Regulated Output and Low Ripple and Noise
 <0.15W No Load Input Power
 Isolation Class II
 CE, CB, UL, cUL Approval
 3-Year Product Warranty
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
( Single Output )
Max Output Wattage (W)
Voltage
MTC30-5S
25W
(Note 1)
Frequency (Hz)
Input
Output
Protection
Isolation
Environment
Physical
MTC30-12S
30W
MTC30-15S
30W
MTC30-24S
30W
90-264 VAC or 120-370 VDC
47-440 Hz
Current (Full load)
650 mA max. (115 VAC) / 400 mA max. (230 VAC)
Inrush current (<2ms,Cold Start)
30 A max. (115 VAC) / 60 A max. (230 VAC)
Leakage Current
< 0.1mA / 264 VAC (Touch Current)
No Load Input Power (<240 VAC)
<0.15W
External Fuse (recommend)
3.15 A slow blow type
Voltage (V.DC.)
5V
Voltage Accuracy
±2%
Current (mA) max
5000
Line Regulation (LL-HL) (typ.)
±0.5%
Load Regulation (10-100%) (typ.)
±1%
12V
15V
24V
2500
2000
1250
Maximum Capacitive Load (at 230 VAC)
6800uF
1600uF
1200uF
470uF
Ripple & Noise (max.)
100mVp-p
150mVp-p
150mVp-p
240mVp-p
Efficiency (at 230 VAC)
84%
89%
86%
86%
Hold-up Time
10 ms min.
Over Power Protection
Hiccup technique, auto-recovery
Over Voltage Protection
Zener diode clamp
(Note 2)
Short Circuit Protection
Hiccup mode, indefinite (automatic recovery)
Input-Output (V.AC)
4000V
Operating Temperature
-40°C…+80°C
Storage Temperature
-40°C…+90°C
Temperature Coefficient
±0.05%/°C
Altitude During Operation
5000m
Humidity
up to 95% RH
MTBF
>250,000 h @ 25°C (MIL-HDBK-217F)
Atmospheric Pressure
70 kPa to 106 kPa
Dimension (L x W x H)
2.52 x 1.77 x 0.92 Inches
Case Material
Plastic resin (flammability to UL 94V-0)
Weight
130 g
Cooling Method
Free air convection
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
( 64.0 x 45.0 x 23.5 mm ) Tolerance 0.5 mm
-1-
VER : A_1
update : 2015.09.11
AC -DC Medical Power Module
MTC30 SERIES
30 Watts
ELECTRICAL SPECIFICATIONS
All specifications valid at normal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Model No.
Safety
( Single Output )
Approval
Conducted and radiated EMI
EMC
MTC30-5S
MTC30-12S
MTC30-15S
MTC30-24S
cUL / UL Standard:
UL 60950-1, CAN/CSA C22.2 No. 60950-1-07
ANSI/AAMI ES60601-1 (2005 + C1:09 + A2:10), CAN/CSA-C22.2 No. 60601-1
(2008), 2 x MOPP
CB Standard:
IEC 60950-1:2005 (2nd Edition) + Am 1:2009 + Am 2:2013
IEC 60601-1:2005 (3rd Edition) + CORR. 1 (2006) + CORR. 2 (2007) + AM1
(2012) or IEC 60601-1 (2012 reprint), 2 x MOPP
EN55011 class B
ESD
EN61000-4-2 air ± 8kV , Contact ± 4kV
Radiated Immunity
EN61000-4-3 10V/m
Fast Transient
EN61000-4-4 ± 2kV
Surge
EN61000-4-5 ±1kV
Conducted Immunity
EN61000-4-6 10Vrms
PFMF
EN61000-4-8 30A/m
Dips
EN61000-4-11 30% 10ms
Interruption
EN61000-4-11 >95% 5000ms
NOTE
1. This product is not designed for use in critical life support systems, equipment used in hazardous environment,
nuclear control systems or other such applications which necessitate specific safety and regulatory standards
other the ones listed in this datasheet.
2. Ripple & Noise are measured at 20MHz of bandwidth with 0.1uF & 47uF parallel capacitor.
3. Safety approvals cover frequency 47-63 Hz.
4. That “natural convection” is about 20LFM but is not equal to still air (0 LFM).
5. It's recommended to add Varistor 14S471K at L / N input side in parallel.
6. Please refer to our PDF file "AC-DC Application" on our website: www.archcorp.com.tw
MECHANICAL DIMENSION
( Top View )
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
PIN#
Single
1
AC IN (N)
2
AC IN (L)
3
-DC OUT
4
NO PIN
5
+DC OUT
-2-
VER : A_1
update : 2015.09.11
AC -DC Medical Power Module
MTC30 SERIES
30 Watts
DERATING
BLOCK DIAGRAM
http://www.archcorp.com.tw
TEL: +886-2-26989508 FAX: +886-2-26981319
-3-