MPS45 series

MPS45
4 5 W S i n g l e O u t put for Medical Type
series
Universal AC input
Low leakage current ≤ 0.3mA
Protections: Short circuit / Overload / Over voltage / Over
temperature
Fixed switching frequency at 45KHz
Cooling by free air convection
Model Number
Output Volts Output Amps Ripple & Noise
Line Reg Load Reg Tolerance Min Load
Efficiency
SINGLE OUTPUT
MPS45-3.3
MPS45-5
MPS45-7.5
MPS45-12
MPS45-13.5
MPS45-15
MPS45-24
MPS45-27
MPS45-48
3.3 Volts(DC)
5 Volts(DC)
7.5 Volts(DC)
12 Volts(DC)
13.5 Volts(DC)
15 Volts(DC)
24 Volts(DC)
27 Volts(DC)
48 Volts(DC)
WWW.ASTRODYNE.COM
8.0 Amps
8.0 Amps
5.4 Amps
3.7 Amps
3.3 Amps
3.0 Amps
1.9 Amps
1.7 Amps
1.0 Amps
80mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
100mV pk-pk
±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±1.0% ±3.0%
±3.0%
±3.0%
±2.0%
±2.0%
±2.0%
±2.0%
±2.0%
±2.0%
±3.0%
±3.0%
±3.0%
±2.0%
±2.0%
±2.0%
±2.0%
±2.0%
±2.0%
0~10.7 Amps
0~10.5 Amps
0~7 Amps
0~4.4 Amps
0~3.9 Amps
0~3.5 Amps
0~2.2 Amps
0~1.95 Amps
0~1.1 Amps
65%
72%
75%
76%
76%
77%
78%
78%
78%
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
MPS45
4 5 W S i n g l e O u t put for Medical Type
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current, typ: (cold start)
Input Current
Leakage current
90-264VAC / 127-370 Volts(DC)
47-440 Hz
15 Amps @ 115VAC Input
30 Amps @ 230VAC
1.2 Amps max @ 115VAC
0.7 Amps max @ 230VAC
< 0.3mA / 230VAC
series
Efficiency
See Selection Chart
Isolation
4000VAC Input - Output
1500VAC Input - Ground
500VAC Output - Ground
EMS
Compliance to
EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, medical level, criteria A
OUTPUT SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Voltage and Current
Line Regulation
Load Regulation
Voltage Tolerance (Note 2)
Ripple/Noise (Note 1)
Hold Up Time @ FL
Setup, Rise Time @ FL
Over Temperature Protection
Over Voltage Protection
Over Current Protection
DC Voltage Adjust
PHYSICAL SPECIFICATIONS
See Selection Chart
See Selection Chart See Selection Chart
See Selection Chart
See Selection Chart
50mS @ 230VAC
16mS @ 115VAC
800mS, 30mS/230VAC
1200mS. 30mS/115VAC
Tj 135°C typ. (U1) detect on main conrol IC, auto recover
115~135%
Hiccup mode, auto recover
53~75W rated output power
36~55W (3.3Volts DC)
Hiccup mode, auto recover
±10%
GENERAL SPECIFICATIONS
Safety
Insulation Resistance
EMI
Harmonic Current
UL2601-1
TUV EN60601-1
IEC60601-1 approved
≥ 100MΩ / 500 Volts(DC)
Compliance to EN55011 (CISPR11) Class B
Compliance to EN61000-3-2,-3
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
WWW.ASTRODYNE.COM
Oper. Temperature
Storage Temperature
Relative Humidity
Temperature Coefficient
MTBF
Vibration
-10°C to +60°C (See Derate Curve)
-20°C to +85°C, 10~95% RH
20 to +90% RH non cond
0.04% / °C (0-50°C)
366.7KHrs min, MIL-HDBK-217F(25°C)
10~500Hz, 2G10min./1cycle, period for 60min. each along X, Y, Z axes
Size
Millimeters
127 x 76 x 28
Inches
5” x 2.99” x 1.10”
Weight
6.35 oz (180g)
Note
1. Ripple and Noise are measured at 20MHz of bandwidth by using a
12” twisted pair-wire terminated with a 0.1uf & 47 uf parallel capacitor.
2. Tolerance: includes set up tolerance, line regulation and load regulation.
3. Mounting holes M1 and M2 should be grounded for EMI purposes.
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
4 5 W S i n g l e O u t put for Medical Type
WWW.ASTRODYNE.COM
MPS45
series
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
4 5 W S i n g l e O u t put for Medical Type
WWW.ASTRODYNE.COM
MPS45
series
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458