MMK150S series

M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
MMK150S
series
88-264VAC Auto Ranging Input
Compact Footprint
High Efficiency
Regulated Outputs
5656VDC Isolation
Built in EMI Filter
3.5”W x 6”L x 1.75”H
Model Number
Output Voltage
Output Amps (max) Ripple/Noise pp,typ
MMK150S-5
MMK150S-12
MMK150S-15
MMK150S-24
5 VDC
12 VDC
15 VDC
24 VDC
30
12.5
10
6.25
WWW.ASTRODYNE.COM
100mv
100mv
100mv
100mv
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
INPUT SPECIFICATIONS
Input Voltage Range
Frequency Range
Inrush Current, typ:
Leakage Current
series
ENVIRONMENTAL SPECIFICATIONS
88-132/176-264 VAC
Auto Voltage Select
47-63 Hz
20A @ 100VAC *
40A @ 200VAC *
<500UA @ 264VAC, 50Hz
OUTPUT SPECIFICATIONS
Voltage and Current
Line/Load Regulation (NL-FL)
Voltage Tolerance
DC Voltage Adjust (typ)
Temperature Coefficient
Ripple/Noise
Over Voltage Protection
Over Current Protection
Short Circuit Protection
Hold Up Time
Turn On Time
Rise Time
Over/Under Shoot
Leakage Current
MMK150S
Oper. Temperature
Storage Temperature
Relative Humidity
EMI (Designed to Meet)
MTBF
0 to +60°C
(See Derate Curve)
-25 to +85°C *
10% to 90%, non-cond *
EN55011, Class B
100,000 Hrs
Mil Std 217, 25°C
PHYSICAL SPECIFICATIONS
See Selection Chart
± 0.5%, typ
± 1%, typ
± 6%, typ
± 0.03%/°C
100mV, typ
130%, Latching, Re-power *
160%, Auto-recovery
Auto-recovery
20mS, typ (Nom I/P, FL)
≤ 1.0 sec, typ
≤ 20 msec, typ
< 10% of FS
≤ 0.3 mA @ 240VAC, 50 Hz
Size
Construction
Weight
6” x 3.5” x 1.75”
Enclosed
1.37 lb, (630g)
GENERAL SPECIFICATIONS
Isolation
Efficiency
Switching Frequency
Safety UL/cUL:
UL-EU:
CB:
CE:
I/P-O/P: 5656VDC
I/P-Ground: 2121VDC
O/P-Ground: 707VDC
77~80%, typ
100Khz, (fixed, typical)
ANSI/AAMI ES60601-1 3rd ed.
CSA C22.2 No.60601-1 3rd ed.
EN60601-1 3rd ed.
IEC 60601-1 3rd ed.
EN60601-1-2
All specifications are typical at nominal input, full load, and 25°C unless otherwise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
WWW.ASTRODYNE.COM
Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
MMK150S
M e d i c a l G r a d e 1 50 Watt Enclosed Switching Power Supplies
MECHANICAL DIMENSIONS
(See Note Below)
(See Note Below)
Pin #
Outputs
1
2
3
4
5
6
7
ACL
ACN
GND
+ Output
+ Output
- Output
- Output
series
(See Note Below)
AAASSSTTTRRRO
O
D
N
O
OD
DYYYN
NEEE PPPAAACCCIIIFFFIIICCC CCCO
ORRRPPP
Note: For our M3 x 0.5 aluminum threaded holes:
Recommended
2.5 in-lbs, 3Curve
in-lbs maximum (2.9 kgf-cm/3.5 kgf-cm max)
7.torque:
Derating
PLEASE NOTE:
To improve specifications and safety isolation, connection
designations
EEE
LLLEEECCCTTTRRRIIICCCAAALLLmay
SSSPPPEEhave
CCCAAATTTIIIO
O
N
ECCCIIIFFFIIIchanged
ON
Nfrom prior revisions. Please
verify connections per product labels on input and output
connections prior to applying power to the unit
OUTPUT DERATING CURVE
120%
100%
Load(%)
80%
Line 2
60%
Line 1
Line 3
40%
20%
0%
℃
10℃
20℃
30℃
40℃
50℃
60℃
70℃
1. Line 1 : Force air cooling.
2. Line 2 : 12V,15V & 24V convection cooling.
3. Line 3 : 5V convection cooling.
WWW.ASTRODYNE.COM
1413
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458