CLAMP ON SENSOR 9660

CLAMP ON SENSOR 9660 < Clamp Sensors < Products
Clamp Sensors (Current sensor for Models 3169 and
3196)
CLAMP ON SENSOR 9660
CAT III 300 V
POWER QUALITY ANALYZER 3196
OVERVIEW
Clamp-on Sensors for the POWER QUALITY ANALYZER 3196 and CLAMP ON POWER
HiTESTER 3169
For input up to AC 500A, select our Model 9661, and for smaller currents of up to AC 100A, select Model 9660.
Both sensors are designed with a safety barrier to protect the user from being in direct contact with live
conductors and completely meet all CE marking requirements.
Basic specifications
Rated input current
100 A AC
Rated output current
0.1 V AC
Accuracy (at 45 to 66 Hz)
Amplitude: ±0.5 % rdg. ±0.2 % f.s. , Full scale (f.s.) is the 3196 range value,
Phase: ±1 ° or less (at 90 A or less), ±1.3 ° or less (at 90 A to 100 A)
Frequency characteristics
66 Hz to 5 kHz (±1 % or less from accuracy)
Max. allowable input
continuous 130 A rms (45 to 66 Hz)
Max. circuit voltage
300 V AC rms (insulated wire)
Core jaw dia.
Ø15 mm(0.59 in)
Dimensions, mass
46 mm(1.81 in)W × 135 mm(5.31 in)H × 21 mm(0.83 in)D, 230 g (8.1 oz), cord length: 3
m(9.84 ft)
OVERVIEW