Manual - Extech Instruments

 USER GUIDE Video Particle Counter with built‐in Camera Model VPC300 Introduction Thank you for selecting the Extech Instruments Model VPC300 Particle Counter with Camera. The VPC300 has a Color TFT LCD display, a 74MB internal data storage bank, a micro‐SD memory card slot for capturing images and video for viewing on a PC, and a built‐in 320x240 pixel camera for capturing stills/video of test locations. The VPC300 offers quick and accurate readings for particle count, air temperature, most surface temperatures, and relative humidity. The VPC300 also offers a Dew Point and Wet Bulb temperature calculation geared for energy efficiency and environmental protection applications. Supplied PC software allows the user to generate reports based on records data, images, and video. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.extech.com) to check for the latest version of this User Guide, Product Updates, and Customer Support. Features 
















2.8” TFT Color LCD display Convenient keystroke menu structure for measurements, setup, and other programming 320 x 240 pixel built‐in camera for still and video imaging JPEG images and 3GP video Meter’s internal memory 74MB. Meter can store 5000 records (date, time, counts, humidity, temperature, sample volumes, and location label) and 20 minutes of video to internal memory. Micro‐SD memory card slot (8GB maximum); card not included. Selectable Test sample time, count data, and programmable delay Simultaneously measure and display 6 channels of particle sizes (down to 0.3μm) Air temperature and relative humidity Dew point and Wet bulb temperature calculations MIN/MAX/AVG/DIF record with date/time stamp Auto power off Language selection for English, French, German, and Spanish Meter is shipped NIST calibrated with calibration certificate Tripod mounting option Includes universal AC adaptor/charger with a variety of plug options, USB cable, PC software, tripod, Zero count filter, 7.4V Lithium‐ion polymer rechargeable battery, and protective case 2 VPC300‐en‐GB_v1.5 5/15 Meter Description 1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Isokinetic probe Temperature and RH Sensor (note its protective cover that slides up and down) Color TFT LCD Compartment for USB interface port and AC Adaptor/Charger Function F1, F2, F3 buttons Control buttons: Page Up/Down, Enter, Run/Stop, and ESC ON/OFF button Battery and Micro‐SD card compartment (SD card slot underneath battery) Tripod mount Camera lens 2
1
10
3
4
9
5
6
8
7
3 VPC300‐en‐GB_v1.5 5/15 Operation Power ON‐OFF Press and hold the power button to switch the instrument ON or OFF. Important Notes Note: Operating the particle counter with the isokinetic probe cap in place will damage the pump and other internal components. Note: Electrical noise, sensor leakage, or other interference may cause the VPC300 to show incorrect particle count data.
Getting Started When the meter powers ON, the Particle Counter icon is displayed at the center of the LCD and the date, time, and battery status can be observed on the top row of the LCD. Three bottom row options are also displayed and can be selected using the Function keys (F1, F2, and F3) as shown below. These are explained in greater detail later in this user guide:
Memory (where data, images, and video are located) System Setup Parameters Help File Symbol Glossary Cumulative mode Concentration mode Differential mode Test is running Video mode Test is paused CAM (Camera) mode Video Record start Color coded particle scale (green is low, yellow is moderate, and red is high particle count) Alarm ON Battery Status Particle Count Setup 4 VPC300‐en‐GB_v1.5 5/15 Sequence of Operation To ensure that you measure the amount of particles accurately, perform this sequence of operation. 1. Zero the particle count sensor 2. Setup your counter for proper operation. Sample time, mode, cycle, and interval. 3. Run your test cycles and evaluate the results. Basic Operation Press and hold the ON/OFF button for 3 to 5 seconds until the display illuminates to power on the meter. Press the enter button ( ) to get to setup mode. Press F3 to enter settings mode. Set each parameter to your preferred setting by pressing ENTER to enter each parameter. Then press ENTER to enable editing of each parameter. Press ESC to exit parameter setup. Sample Time: Length of time each test runs (3 to 60 seconds) Start Delay: Duration of time from RUN to actual start of test. (3 to 100 seconds) Channel Display: Particle sizes displayed during test. Ambient Temp/RH: Allows the display of the air temp and %RH values Sample Cycle: Number of test cycles to run (1 to 100) Sample Mode: Cumulative, differential, concentration Interval: Test cycle delay measured in seconds. Level Indication: Particle level selection (highlighted on display) Press ESC to return to testing screen Press RUN/STOP to start the automated test sequence. 5 VPC300‐en‐GB_v1.5 5/15 Taking Measurements After powering the meter, press the Enter ( ) button to access the Particle Measurement screen. Then open the temperature/RH sensor by sliding its protective cover downward. When measuring particle counts, these correspond to the function keys (F1,F2,F3). icons
F1 is to enter Camera and video mode, F2 is to view your saved data, and F3 is for Setup mode. The display will show the particle count channels with readings, air temperature (AT), Dew point temperature (DP), Wet bulb temperature (WB), and Relative Humidity (%RH) as shown in accompanying diagram. Press the RUN/STOP button to start the air pump and the particle count measurement process. When the programmed sampling time has elapsed, the particle measurement cycle will automatically end and the measurement data will be automatically stored. To set the sampling time, refer to the Particle Count SETUP MODE section. If desired, press the RUN/STOP button and stop the session manually. When measuring particle counts, these icons correspond to the function keys (F1,F2,F3). F1 is to enter Camera and video mode, F2 is to view your saved data, and F3 is for Setup mode. Capturing Images While measuring particle counts, press F1 to enter the camera mode. Press F1 to capture a Still image. Then press F1 to Save the image or F3 to delete it. Press ESC to get back to exit this mode. Capturing Video While measuring particle counts, press F1 to enter the camera mode. Press F3 to enter the video mode. The LCD will show . Now press F2 to capure video and press F2 again to stop the video capture. The video file will be automatically stored. The System Setup mode section allows the user to store video to the meter’s internal memory or to an installed Micro‐SD card. Press ESC to exit this mode. Particle Counter Setup Mode When measuring, these display icons are visible. Press F3 to enter the Setup mode. Use the up and down arrow keys to select an option to edit and press ENTER ( ) to access the selection. SAMPLE TIME (Test Cycle Time) Select Sample Time. Press the Enter ( ) button to enter the option and Press Enter again to activate the setting for editing. Use the arrow buttons to set the sampling time/gas volume (3 to 60 seconds). Use the Esc button to save and return to the menu. 6 VPC300‐en‐GB_v1.5 5/15 START DELAY The test start delay time is adjustable (3 to 100 seconds). From the Particle setup menu, use the Up and Down arrows to select Start Delay. Press the Enter ( ) button to enter the option and Press Enter again to activate the setting for editing. Use the arrow keys to select the desire delay time. The test will not start until the programmed delay time has elapsed. Use the Esc button to save the setting and to return to the menu list. CHANNEL DISPLAY Select channels that are to be displayed. From the Particle setup menu, use the Up and Down arrows to select Channel Display. Press the Enter ( ) button to enter the option. Use the arrow keys to select the channel and press the Enter key to select or de‐select the channel. Use the Esc key to save the setting and to return to the menu. AMBIENT TEMPERAURE / %RH Enable or Disable the ambient temperature and relative humidity displays. From the Particle setup menu, use the Up and Down arrows to select Ambient Temp/RH. Press the Enter ( )
button to enter the option. Use the arrow keys to select Enable or Disable and use the ESC button to save and return to the menu. SAMPLE CYCLE Set the desired number of test cycles to run. From the Particle setup menu, use the Up and Down arrows to select Sample Cycle. Press the Enter ( ) button to enter the option and Press Enter again to activate the setting for editing. Use the arrow keys to select the number of cycles to run the test (1 to 100). Press ESC to save setting and to return to the menu list. 7 VPC300‐en‐GB_v1.5 5/15 SAMPLE MODE Concentration). From the Particle Select desired mode ( Cumulative, Differential, or setup menu, use the Up and Down arrows to select Sample Mode. Press the Enter ( ) button to enter the option. Use the arrow buttons to select desired mode and then use ESC to save and to return to the menu list. Cumulative Mode – measures all particles that are greater than or equal to the particle size selected in the Sample Volume field. Differential Mode – measures all particles that are greater than or equal to the particle size selected in the Sample Volume field, but less than the next largest particle size. Concentration Mode – used to take a quick sample of airborne particulates, performed especially in areas where the particulate levels are unknown and may exceed the operating levels of this counter. INTERVAL (TIME BETWEEN REPEATED TEST CYCLES) Set the time between tests for tests with more than one cycle (1 to 100 seconds). From the Particle setup menu, use the Up and Down arrows to select Interval Time. Press the Enter ( )
button to enter the option and Press Enter again to activate the setting for editing. Use the arrow buttons to set the desired interval time. Use the ESC button to save setting and to return to the menu. Note: The INTERVAL setting is only for programming the time between test cycles, not between individual measurements. 8 VPC300‐en‐GB_v1.5 5/15 LEVEL (ALARM LIMIT) INDICATION Select the Alarm Limit (Level) of the corresponding particle size. When the selected particle size is exceeded, the instrument alerts the user. From the Particle setup menu, use the Up and Down arrows to select Level Indication. Press the Enter ( ) button to enter the option. Use the arrow buttons to select the desired alarm limit and then press ESC to save setting and return to the menu. Memory Storage Browser When the meter is switched ON, these LCD icons are visible
. Press F1 to enter the data memory section. There are three options available in the memory mode, Picture, Video, and Particle Logs. Use the arrow buttons to select one and then press ENTER ( ) to access the selection. When browsing images, data, and video use the arrow buttons to navigate, use the Enter button to select, and use the ESC or F3 button to return to the menu. Press ESC again to exit to the power‐on screen. 9 VPC300‐en‐GB_v1.5 5/15 System Setup Parameters When the meter is switched ON, these LCD icons are visible
. Click on the F2 key and the menu list will appear. Detailed explanations are provided below for each parameter. Use the Arrow keys to choose the desired parameter. Set the Date and Time. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to change a value, and use the Enter button to go to the next item. Press the ESC button to save the setting return to the Setup menu. FONT COLOR Select the color for the Display font. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select a color and use the ESC button to save the setting and return to the Setup menu. DATE/TIME LANGUAGE Select the desired language for menu text. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select a language and use the ESC button to save the setting and return to the Setup menu. 10 VPC300‐en‐GB_v1.5 5/15 BRIGHTNESS AUTO POWER OFF Adjust the display to the desired brightness level. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select a level and use the ESC button to save the setting and return to the Setup menu. Select elapsed time value before meter switches OFF. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select the setting and use the ESC button to save the setting and return to the Setup menu. DISPLAY TIME‐OUT Select display auto‐off delay. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select a setting and use the ESC button to save the setting and return to the Setup menu. 11 VPC300‐en‐GB_v1.5 5/15 MEMORY STATUS Display the memory and the micro‐SD card status. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select ‘Device’ or ‘SD Card’ and then view the status for the selected item. Use the ESC button to return to the Setup menu. FACTORY DEFAULT UNITS Restore factory default settings. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select YES or NO and then use the ESC button to save the setting and return to the Setup menu. Select the desired unit of measure for temperature C or F. Press the Enter ( ) button to open the selection for editing. Use the arrow keys to select the Unit and use the ESC button to save the setting and return to the Setup menu. 12 VPC300‐en‐GB_v1.5 5/15 Zero the Particle count Sensor Particles must be purged (removed) from the sensor before and after each use in a high sample count environment. Zero the sensor: 1. Unscrew and remove the Isokinetic probe and attach the Zero count filter. 2. Turn on the meter and in the Setup mode, a. Set the Sample mode to Cumulative. b. Set the Channel Display to ensure all channels are selected. c. Set Sample time to 60 seconds d. Set Sample Cycle to 10 3. Start the meter 4. Allow the meter to run until all of the particle counts read zero. You may need to run the meter multiple times to ensure that all channels read zero. 5. Turn off the meter. 6. Remove the Zero count filter and attach the Isokinetic probe. 13 VPC300‐en‐GB_v1.5 5/15 Particle Counter Considerations Particle pollution is a mixture of various components including smoke, dust, chemicals, mold, and gases. Particle concentrations can affect our health and comfort levels. Evaluating the quality of the air by determining the concentration and size of these particles can help to determine if a problem exists. Locating the source of these various particles can help to determine effective methods of reducing these and improving the Indoor Air Quality (IAQ). One way to determine the source of air particulates in a room is to test the air outside of the room under investigation. Determine the concentration and size of particulates at the air intake to that room. Purge or Zero the particle counter and then measure the concentration and size of particles in multiple areas of the room to be tested, to determine the effectiveness of the incoming air filtration. The air temperature, relative humidity, and CO2 levels are also an important consideration for IAQ as high humidity and lower temperatures can bring about mold creation. If a room is classified as an ISO clean room or clean zone, the counts of particle sizes for that class cannot be exceeded when tested. Refer to the ISO Clean Room Class chart (ISO 14644‐1:1999) for values. The VPC300 is equipped with a still/video camera to directly see the areas under test. Data, images, and video can be stored on up to an 8GB micro‐SD card (not included) or in the meter’s internal 74MB memory. The life of the meter can be maximized by testing only when necessary and securely storing the instrument when it is not in use. Continuous use is not recommended and can shorten the life of the instrument especially in dusty environments. Setting up a common sense maintenance schedule and testing to the schedule will go a long way in lengthening the life of the instrument sensor. PC Interface and Supplied PC software This meter has the capability to connect to and communicate with a PC. A USB cable is supplied and connects to the jack on the meter’s left side compartment. To install and use the software, please refer to the instructions provided on the supplied CD‐
ROM and/or the instructions provided in the HELP Utility within the software program. Check the software download page of the website www.extech.com for the latest version of the PC software and its operating system compatibility. 14 VPC300‐en‐GB_v1.5 5/15 Battery Charging and Replacement When the battery icon appears drained on the LCD or if the meter will not switch ON, the battery must be recharged using the supplied AC adaptor/charger. The charger plug connects to the jack located on the meter’s left side compartment. The battery compartment is located on the rear of the instrument. The battery is a 7.4V Lithium‐ion polymer rechargeable battery. To access the meter’s battery for replacement: 1. On the rear of the meter, remove the Phillips head screw that secures the battery compartment. 2. Open the compartment by carefully lifting the compartment cover. 3. Replace the battery with one of the same type (note that the Micro‐SD card slot is located underneath the battery as explained below). 4. Close the compartment and secure the compartment cover with the Phillips head screw Battery Safety Reminders o
o
o
Please dispose of batteries responsibly; observe local, state, and national regulations. Never dispose of batteries in fire; batteries may explode or leak. Always install a new battery of the same type. Micro‐SD Card Slot (optional) The micro‐SD card slot is located inside the battery compartment underneath the battery. Micro‐SD cards up to 8GB can be inserted in the slot. To access the SD memory card slot: 1. Turn off power to the meter and remove the AC adapter cable if connected. 2. Open the battery compartment as described in the battery replacement section. 3. Remove the battery. 4. Insert a micro‐SD card (8GB max.) in the card slot underneath the battery by lifting up on the metal card holder and placing the Micro SD card in the connector. The pins facing towards the top of the meter. Close the metal holder. 5. Insert the battery and secure the compartment cover before powering on the meter. 15 VPC300‐en‐GB_v1.5 5/15 Specifications Particle Counter Channels Six channels: 0.3, 0.5, 1.0, 2.5, 5.0, 10μm Flow Rate 3
0.1ft /min (2.83L/min) Counting Efficiency 50% at 0.3μm; 100% for particles > 0.45μm Coincidence Loss 3
5% at 2,000,000 particles per ft Particle Count Modes Cumulative, Differential, and Concentration modes Air Temperature and Relative Humidity Measurements Air Temperature Range 32 to 122oF (0 to 50oC) Dew Point/Wet Bulb Temperature Range 32 to 122oF (0 to 50oC) Relative Humidity Range 0 to 100%RH Air Temperature Accuracy o
o
o
o
o
±0.9 F (0.5 C) 50 to 104 F (10 C to 40 C) o
o
±1.8 F (1.0 C) all other ranges Dew Point/Wet Bulb Temperature Accuracy ±0.9oF (0.5oC) 50 to 104oF (10oC to 40oC) o
o
1.8 F (1.0 C) all other ranges Relative Humidity Accuracy ±3%RH from 40 to 60%RH ±3.5%RH from 20 to 40%RH and from 60 to 80%RH ±5.0%RH from 0 to 20%RH and from 80 to 100%RH General Specifications Display 2.8” 320 x 240 pixel TFT Color LCD with backlight Low battery indication Battery symbol appears on the LCD Micro‐SD Card Slot Accommodates up an 8GB card Meter internal memory 5000 sample records; 74MB Power supply Rechargeable 7.4V Lithium‐ion polymer battery Battery life Approx. 4 hours of continuous use 2 hours approx. with AC adaptor/charger Battery charge time Operating Temperature o
o
32 to 122 F (0 to 50 C) Operating/Storage Humidity 10 to 90% RH non‐condensing Storage Temperature o
o
14 to 140 F (‐10 to 60 C) Dimensions 9.4 x 3.0 x 2.2” (240 x 75 x 57mm) Weight 1.26 lbs. (570g) Copyright © 2014‐2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form ISO‐9001 Certified www.extech.com 16 VPC300‐en‐GB_v1.5 5/15