VUFM-S400 Datasheet - MEDICAL AC

For more information, please visit the product page.
date
01/21/2014
page
1 of 5
SERIES: VUFM-S400 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
•
•
•
•
up to 400 W continuous power
compact 1U size
5.56 W/inch3 power density
universal input (90~264 Vac)
single output from 3.3~48V
short circuit, over voltage and over temperature protections
full medical approvals
built-in active PFC function
built-in remote sense function
efficiency up to 83%
®
IEC
60601-1
EN
60601-1
output
voltage
output
current
output
power
ripple and
noise4
efficiency
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
typ
(%)
VUFM-S400-03R
3.3
60
2001
33
70
VUFM-S400-05R
5
60
300
50
75
VUFM-S400-12R
12
33.34
4003
120
80
VUFM-S400-18R
18
25
3
400
180
83
VUFM-S400-24R
24
18.19
4003
240
83
VUFM-S400-36R
36
12.9
3
400
360
83
VUFM-S400-48R
48
9.53
4003
480
83
MODEL
Notes:
1.
2.
3.
4.
total continuous output power will not exceed
total continuous output power will not exceed
total continuous output power will not exceed
1% minimum load is required to maintain the
2
200 W forced air (23 CFM), 150 W without fan
300 W forced air (23 CFM), 225 W without fan
400 W forced air (23 CFM), 250 W without fan
ripple and regulation
PART NUMBER KEY
VUFM-S400 - XX R X
Base Number
Preset Output Voltage
cui.com
Current Sharing:
“blank” = N/A
I = available
For more information, please visit the product page.
CUI Inc │ SERIES: VUFM-S400 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 01/21/2014 │ page 2 of 5
INPUT
parameter
conditions/description
min
voltage
typ
90
frequency
current
47
at 90 Vac, full load
max
units
264
Vac
63
Hz
6.35
A
35
A
inrush current
at 230 Vac, full load, cold start
input fuse
Built-in ac fuse. A blown fuse usually indicates permanent damage to the power supply serviceable by
factory only.
power factor correction
meets EN 61000-3-2 Class D
OUTPUT
parameter
conditions/description
min
total regulation
transient response
typ
max
±1
units
%
Output voltage returns to within 1% in less than 2.5 ms for a 50% load change. Peak transient does
not exceed 5%.
start-up time
at 120 Vac
hold-up time
at 120 Vac, 80% load
adjustability
output user adjustable
1
20
±5
switching frequency
s
ms
30
%
kHz
overshoot
Turn-on and turn-off overshoot shall not exceed 5% over nominal voltage.
remote sense1
Designated as RS+ and RS- on CN3. Total voltage compensation for cable losses with respect to the
main output.
remote on/off
Defined RSW on CN3, requiring a low signal to inhibit output.
LED display (LED 1)
Green - the power supply is operating normally.
Orange - when any protection occurs or RSW is low.
power good
Designated as PG on CN3. This signal goes high 100~500 ms after the output reaches regulation. It
goes low at least 1 ms before loss of regulation.
current sharing
Designated as CSH on CN3, optional single wired for forced surrent sharing function and parallel up to
4 units within 10% accuracy at full load.
current monitor
Designated as CMN on CN3 for for current sense for 0.5~3 Vdc to represent 0~100% output current.
AC fail (optional)
Designated as ACF on CN3 to monitor the input voltage when input goes under 80 ±5 Vac the signal
will go low (0 V) and then go high (+5 V) once it reappears over 86 Vac.
Notes:
1. Not available for current sharing models
PROTECTIONS
parameter
conditions/description
input under voltage protection
Power supply shuts down when ac input is under 80 ±5 Vac. When ac line reappears over 86 ±5 Vac,
the power supply restarts automatically.
min
over voltage protection
shutdown and latches, ac input reset required to restart
over current protection
auto recovery
short circuit protection
auto recovery upon removal of short
110
over temperature protection
shutdown
85
cui.com
typ
max
units
130
%
140
%Io
°C
For more information, please visit the product page.
CUI Inc │ SERIES: VUFM-S400 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 01/21/2014 │ page 3 of 5
SAFETY & COMPLIANCE
parameter
conditions/description
isolation voltage
primary to secondary at 2 mA for 3 seconds
primary to transformer core at 2 mA for 3 seconds
primary to earth ground at 2 mA for 3 seconds
min
typ
isolation resistance
allowable resistance measured when 40 A current
is applied from the ground pin of the three prong
plug to the farthest earthed connection point.
safety approvals
UL 60601-1, CSA C22.2 No. 601.1-M90, TUV EN 60601-1, CE Mark (LVD)
EN 61204-3/60601-1-2/61000-3-(2,3) & IEC 61000-4 Series Regulations, CB
EMI/EMC
FCC Part 15, CISPR22 Class B, conducted
yes
MTBF
according to MIL-HBK-217F at 30°C
units
Vac
Vac
Vac
leakage current
RoHS compliant
max
4,000
1,500
1,500
0.1
Ω
300
μA
100,000
hours
ENVIRONMENTAL
parameter
conditions/description
operating temperature
derating linearly at 2.5% from 50~70°C
min
storage temperature
max
units
0
typ
70
°C
-20
85
°C
non-condensing
5
90
%RH
storage humidity
non-condensing
5
95
%RH
vibration
5 ~ 50 Hz, per axis
operating humidity
±7.35
DERATING CURVE
Load (%)
100
80
60
40
20
0
10
20
30
40 50 60 70 80
Ambient Temperature (°C)
cui.com
90
100
m/s2
For more information, please visit the product page.
CUI Inc │ SERIES: VUFM-S400 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 01/21/2014 │ page 4 of 5
MECHANICAL
parameter
conditions/description
dimensions
203.2 x 127 x 40.64 (8 x 5 x 1.6 inch)
min
typ
max
units
mm
weight
1.3
kg
MECHANICAL DRAWING
units: mm[inches]
tolerance: ±0.5[0.02]
VO-
VO-
VO+
VO+
B
CN1
B
B
C
C
INPUT CONNECTOR (CN1)
Howder terminal block
HD-121-3P
(option 1)
OUTPUT CONNECTOR (CN2)
Howder terminal block
HD-121-6P
(option 1)
Molex 09-50-3161
(16 pins)
(option 2)
PIN
FUNCTION
PIN
FUNCTION
1~3
+Vo
1~8
+Vo
4~6
-Vo
9~16
-Vo
Note:
FAN
LOGIC CONNECTOR (CN3)
Molex 09-50-3071
(7 pins, 5 used)
(option 2)
---
Suggested mating connector
--- or equivalent
Contact: SXH-002T-P0.6
PIN
FUNCTION
PIN
FUNCTION
1
MAG - margin
6
PG - power good signal
2
CMN - current monitoring
7
RSW - remore on/off
3
CSH - current sharing
8
RS- - remote sense (-)
4
RTN - return
9
RS+ - remote sense (+)
5
ACF - AC fail
1. Bottom mounting holes have a maximum screw depth of 3.8mm. Side mounting holes have a maximum screw depth of 6.3mm.
cui.com
Suggested mating
connector
--- or equivalent,
Contact: SXH-002TP0.6
For more information, please visit the product page.
CUI Inc │ SERIES: VUFM-S400 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 01/21/2014 │ page 5 of 5
REVISION HISTORY
rev.
description
date
1.0
initial release
07/06/2006
1.01
updated features, added preset voltage data
12/27/2006
1.02
updated connector data and drawing
12/12/2007
1.03
new template applied
05/11/2012
1.04
V-Infinity branding removed
08/28/2012
1.05
added derating curve
10/30/2012
1.06
updated spec, Molex connectors changed
01/21/2014
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
[email protected]
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical component is any component of a
life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to affect its safety or effectiveness.