Technical data PBMH Hygienic

PBMH
Fully welded pressure transmitter for hygienic
applications
Main benefits
„„
High temperature resistance
for SIP and CIP processes
„„
Surface roughness of flush
process connection Ra ≤ 0.8
for highest hygienic requirements
„„
Fully welded and compact
design for washdowns without
residuals
„„Excellent active temperature
compensation for increased
process stability
„„External programming of zero
point and span with FlexProgrammer 9701
Applications
„„Food
„„Beverage
„„Biotechnology
„„Pharmaceutical
TYPE EL - CLASS I
DECEMBER 2012
Measuring principle Piezoresistive silicon sensor
Measuring ranges
-1 ... 0 bar up to 0 ... 40 bar
Type of pressure
Relative / Absolute
Accuracy (20 °C)
≤ ± 0.25% FS
± 0.1% FS
(includes linearity, hysteresis,
≤
repeatability, error of span
and zero point according limit
point adjustment)
Turn down
5:1
Performance after
Turn down
[Turn-Down] * [Accuracy] FS
Zero thermal drift
≤ ± 0.03% FS/10 K
Span thermal drift
≤ ± 0.03% FS/10 K
Long term stability
≤ ± 0.1% FS / Year
Response time
(10 … 90%)
≤ 5 ms
Process connections See page 3
Bump
IEC60068-2-27
100 g/2 ms
4000 x Imp. per axis and direction
Random
IEC60068-2-64
0.1 g2/Hz (20 Hz – 1 KHz)
30 min per axis (>10 g RMS)
Electrical specification
Output signal /
Power supply
Load impedance
Current ouput
Voltage output
Insulation resistance
4 … 20 mA / 8 … 30 VDC
0…10 V / 13 … 30 VDC
RΩ = (Usupply-8 V) / 20 mA
> 5 KΩ
>100 MΩ at 500 VDC
Electrical connections See page 3
Material
Process connection SS 1.4404 AISI 316L, SS 1.4435 AISI 316L or
Hastelloy-C
SS 1.4404 AISI 316L
Diaphragm
SS 1.4435 AISI 316L or Hastelloy-C
Temperature
Storage
-40 ... + 85°C
Compensated range -40 ... + 85°C
-40 ... +125°C
Medium
Sealing
EPDM O-rings are conform to 3-A
Sanitary
Standard 18-03 Class II
EPDM gaskets are conform to 3-A
Sanitary
Standard 18-03 Class I
Cable
PUR
(without cooling neck)
Medium
(with cooling neck)
-40 ... +200°C
Ambient
-40 ... + 85°C
SIP/CIP compatibility Medium temperature up to 150 °C
(< 60 minutes) without cooling neck
Medium temperature up to 200 °C
(permanent) with cooling neck
Protection rating
IP65 (EN 60529) up to IP67 depending on
electrical connection
Vibration
IEC60068-2-6
1.5 mm p-p (10 – 57 Hz), 10 g (58 Hz – 2 KHz)
10 cycles within 2.5 h per axis
Shock
IEC60068-2-27
50 g/11 ms 100 g/6 ms
10 x Imp. per axis and direction
www.baumer.com
2014-10-03
Housing
Environment
Design and specifications subject to change without notice
Technical specifications
Data sheet A11.34
Page 1 / 4
PBMH
Fully welded pressure transmitter for hygienic
applications
ATEX
ATEX II 1G
Ex ia IIC T4/T6 Ga
ATEX II 1/2G
Ex ia IIC T4/T6
Ga/Gb
Surface roughness (in contact with medium)
Process connection
All versions without cooling neck,
DIN connector and with output
signal code A1
All versions without cooling neck,
with DIN connector and output
signal code A1
ATEX II 1G
All versions with cooling neck,
Ex ia IIC T3/T4/T6 Ga output signal code A1 and without
DIN connector
All versions with cooling neck,
ATEX II 1/2G
Ex ia IIC T3/T4/T6
DIN connector and output signal
Ga/Gb
code A1
ATEX II 1D
Ex ia IIIC T107°C
IP6X Da
All versions with output signal
code A1
Barrier data
Ui ≤ 30 V
Ii ≤ 100 mA
Pi ≤ 750 mW
Capacity
Ci ≤ 31 nF
CCable ≤ 0.12 nF/m
Inductivity
Li ≤ 3 μH
LCable ≤ 1.1 μH/m
ISO 2852 / TriClamp / DIN 32676 / DIN 11864-3
Ra ≤ 0.4 μm
DN38 Hygienic Connection
Ra ≤ 0.8 μm
DN76 Hygienic Connection
Ra ≤ 0.8 μm
GEA Tuchenhagen Varivent® ball housing
Ra ≤ 0.8 μm
Weld joint
Ra ≤ 0.8 μm
Diaphragm
Ra ≤ 0.4 μm
Approvals
CE conformity
EMC directive 2004/108/CE in accordance with
EN 61000-6-2 & EN 61000-6-3
Hygienic
3-A 74-06
EHEDG, EL Class I
Temperature class T1 ... T3: -40 <Tamb < 45/70/75/85 °C
(ambient temperature) T1 ... T4: -40 <T < 85 °C
amb
T1 ... T6: -40 <Tamb < 70 °C
Temperature class T1 ... T3: -40 <Tmed < 130/150/160/170/200 °C
(medium temperature) T1 ... T4: -40 <T < 115/130 °C
med
T1 ... T6: -40 <Tmed < 75/80 °C
For the application in Ex zone you have to respect the conditions
mentioned in the ATEX Type Examination Certificate (SEV 11 ATEX
0129).
You find the certificates and manuals under http://www.baumer.com/
Measuring ranges and overpressure safety
0 ... 0.1
0 ... 0.16
0 ... 0.25
-0.1 ... 0.1
-0.2 ... 0.2
0 ... 0.4
0 ... 0.6
0 ... 1
-1 ... 0
-1 ... 0.6
0 ... 1.6
0 ... 2
0 ... 2.5
-1 ... 1.5
0 ... 4
-1 ... 3
-1 ... 5
0 ... 6
0 ... 10
-1 ... 9
0 ... 16
-1 ... 15
0 ... 20
0 ... 25
-1 ... 24
0 ... 40
-1 ... 39
1
3
15
60
70
135
Burst pressure
2
6
30
120
140
270
2014-10-03
Over pressure
Design and specifications subject to change without notice
Pressure in bar
Pressure range
www.baumer.com
Data sheet A11.34
Page 2 / 4
PBMH
Fully welded pressure transmitter for hygienic
applications
Electrical connections
Signal
4...20 at
mA
/ Signal at 0...10 V
Signal at 4...20
mA /atSignal
0...10V
Connector
DIN 43650
M12
4-pins
4
1
+VS / +VS
3
IOUT / GND
n.c. / +VOUT
3
2
n.c. / +VOUT
IOUT / GND
1
2
+VS / +VS
n.c. / +VOUT
earth
earth
Cable
output
Field housing
(Red) / +VS / +VS
Not included in delivery
clamping range: 8...10 mm
n.c. / +VOUT
(Blue) / IOUT / GND
4
3
2
1
IOUT / GND
(White) / n.c. / +VOUT
+VS / +VS
(Shield) / earth
(Shield) / earth
Dimensions (mm)
EN 175301-803A
(DIN 43650 A)
4 pins, IP65
Field Housing,
Cable gland stainless
steel, IP67
Shielded cable
(1,5 m),
IP67
Ø 8...10
Ø 54,5
Ø 6,3
51,6
64
94,5
min. 91,8
M12
4 pins, IP67
Ø 22
Ø 22
Ø 22
Code 44
Code 14
Ø 22
Code 54
Code 53
optional with cooling neck
Code 57
Code 51
Code 54
Code 53
DN3/4" TriClamp /
DN3/4" DIN 32676
Series C
Code 50
Ø 34
Ø 19
Membrane
DN21.3 ISO 2852 /
DN20 DIN 32676
Series A
Ø 50,5
Ø 25
Membrane
DN33.7 ISO 2852 /
DN38 ISO 2852 /
DN25 DIN 32676 Series A /
DN32 DIN 32676 Series A /
DN40 DIN 32676 Series A /
DN33.7 DIN 32676 Series B /
DN1 1/2" DIN 32676 Series C
Ø 50,5
Ø 19
Membrane
DN25 ISO 2852 /
DN26.9 DIN 32676 Series B /
DN1" DIN 32676 Series C /
Code 62
Ø 25
Membrane
DN40 ISO 2852 /
DN51 ISO 2852 /
DN50 DIN 32676 Series A /
DN42.4 DIN 32676 Series B /
DN48.3 DIN 32676 Series B /
DN2" DIN 32676 Series C
Code 61
Code 58
51
27
Code 56
Ø 64
Ø 50
Ø 25
Membrane
Ø 66
Ø 25
Ø 19
Membrane
Ø 72
Ø 25
Membrane
Ø 91
DN76 Hygienic
Connection 3-A
Ø 25
Membrane
Ø 84
Varivent® type-N
Ø 50,5
Ø 25
Membrane
DN25 AsepticClamp
DIN 11864-3
BKS form A
2014-10-03
Ø 50,5
DN38 Hygienic
Connection 3-A
Varivent® type-F
Ø 68
Design and specifications subject to change without notice
Ø 16
Membrane
17,3
Ø 24,9
50.8 - 52.8
17,3
Code 52
www.baumer.com
Data sheet A11.34
Page 3 / 4
PBMH
Fully welded pressure transmitter for hygienic
applications
Ordering details PBMH hygienic
PBMH
Model
Pressure transmitter
.
x
.
xxx
.
x
.
xx
.
2
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
Only pressure type relative
xx
.
x
.
x
.
x
.
.
B2H
B4G
B59
B72
B74
B76
B77
B79
B81
B82
B1L
B08
B09
B10
B11
B12
B15
B16
B17
B18
B19
B20
B22
B24
B25
B26
B27
.
R
A
Output signal
4...20mA
0...10V
.
A1
A2
Output connection
M12, 4 pins
DIN 43650, 4 pins
Shielded cable (1.5 m)
Fiel housing (only with output signal A1)
.
14
44
53
54
Process connection
DN 38 Hygienic Connection 3-A
DN 33.7 - DN 38 ISO2852 / TriClamp 1 1/2" / DN 25 - DN 40 DIN32676
DN 3/4" TriClamp / DN 3/4" DIN32676 (without 3-A)
DN 25 ISO2852 / DN 26.9 DIN32676
DN 21.3 ISO 2852 / DN 20 DIN32676
DN 40 - DN51 ISO2852 / DN 42.4 - DN 48.3 DIN32676
DN76 Hygienic Connection 3-A
DN 25 Aseptic Clamp DIN 11864-3 BKS, Series A, Form A
Varivent® type N
Varivent® type F
DN 33.7 - DN 38 ISO2852 / TriClamp 1 1/2" / DN 25 - DN 40 DIN32676 with cooling neck
DN 3/4" TriClamp / DN 3/4" DIN32676 with cooling neck (without 3-A)
DN 25 ISO2852 / DN 26.9 DIN32676 with cooling neck
DN 40 - DN51 ISO2852 / DN 42.4 - DN 48.3 DIN32676 with cooling neck
DN 21.3 ISO 2852 / DN 20 DIN32676 with cooling neck
DN 25 Aseptic clamp DIN 11864-3 BKS, Series A, Form A with cooling neck
Process connection material
Stainless steel 1.4404 AISI 316L
Stainless steel 1.4435 AISI 316L
Stainless steel 1.4435 AISI 316L electropolished
Hastelloy-C
In combination with process connection Code 51/52/54/57/81/82/84/87
In combination with process connection Code 51/52/54/57/81/82/84/87
Sealing
Without
EPDM (FDA / 3-A)
In combination with process connection Code 51/ 81
EPDM (EHEDG)
* Size of sealing according bold description at process connection
T ≥ -10°C
.
50
51
52
53
57
54
56
58
61
62
81
82
83
84
87
88
.
2
5
F
H
.
0
2
7
.
1
2
.
0
0
1
In combination with output signal code A1
0
S0x
2014-10-03
Display
Without
ATEX
Without
ATEX according to SEV 11 ATEX 0129
Approval
None
Downscale
Without downscale
With downscale
.
.
4
5
P > 250 mbar
Type of pressure
Relative
Absolute
Oil filling
Silicon oil
FDA approved white oil
xx
Design and specifications subject to change without notice
Pressure range and unit in bar
-0.1...0.1
-0.2...0.2
-1...0
-1...0.6
-1...1.5
-1...3
-1...5
-1...9
-1...15
-1...24
-1...39
0...0.1
0...0.16
0...0.25
0...0.4
0...0.6
0...1
0...1.6
0...2
0...2.5
0...4
0...6
0...10
0...16
0...20
0...25
0...40
2
–
Housing material
Stainless steel 1.4404 AISI 316L
Accuracy
0.25% FS
0.10% FS
–
PBMH
www.baumer.com
Data sheet A11.34
Page 4 / 4