Download Data Sheet

WMIHPU15 SERIES
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies
Rev A
Wall Industries, Inc.
FEATURES
 Class II System
 Single Outputs Ranging from 3VDC to 55VDC
 RoHS Compliant
 -10°C to +70°C Operating Temperature Range
 Up to 15 Watts Output Power
 Meets FCC Part-18 Class B & CISPR-11 EN55011
 86% High Efficiency
Class B Emission Limits
 Energy Star 2.0, Efficiency Level V
 UL/cUL 3rd Edition Medical Approvals
 80-275VAC Input Voltage Range
 Interchangeable Plug Options: United States,
 Over Voltage & Over Load Protection
 Input to Output: 2MOPP
Europe, Australia, & United Kingdom Types
 Optional Output Connectors Available
SAFETY APPROVALS
Size:
2.36 x 1.71 x 1.58 inches
60.0 x 43.5 x 40.2 mm
Weight:
5.82oz (165g)
DESCRIPTION
AC Plug Types:
• Unites States (“U” Suffix)
• Europe (“E” Suffix)
• Australia (“A” Suffix”)
• United Kingdom (“K” Suffix)
Applications:
• Home Healthcare Equipment
• Patient Monitors
• Blood Pressure Systems
• Portable Medical Devices
• ECG Machines
The WMIHPU15 series of Class II medical AC/DC wall mount power supplies offers up to 15
watts of output power in a 2.36” x 1.71” x 1.58” package. This series consists of single output
models ranging from 3VDC to 55VDC with an 80~275VAC input voltage range and a -10°C
to +70°C operating temperature. This series meets FCC Part-18 Class B and CISPR-11
EN55011 Class B Emission Limits and has ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd edition)
and CE safety approvals. All units are RoHS and Energy Star Level V compliant. Plugs come
in United States (“U” suffix), Europe (“E” suffix), Australia (“A” suffix), and United Kingdom
(“K” suffix) types. Plugs are sold separately so please contact factory for ordering details.
MODEL SELECTION TABLE
Model Number (1)
Input Voltage
Adjustable Voltage Range Adjustable Current Range
Maximum
Ripple & Noise
Output Power
Min
Max
Min
Max
2.40 A
7.2W
66mVp-p
Efficiency
No Load
Consumption
±5%
65%
0.2W
WMIHPU15-101x
3.0 VDC
4.99 VDC
WMIHPU15-102x
5.0 VDC
5.99 VDC
2.00 A
2.40 A
12W
50mVp-p
±5%
80%
0.3W
WMIHPU15-103x
6.5 VDC
8.0 VDC
1.50 A
1.84 A
12W
70mVp-p
±5%
82%
0.3W
WMIHPU15-104x
8.0 VDC
11.0 VDC
1.22 A
1.68 A
13.5W
90mVp-p
±5%
83%
0.3W
WMIHPU15-105x
11.0 VDC
13.0 VDC
1.15 A
1.36 A
15W
100mVp-p
±5%
84%
0.3W
WMIHPU15-106x
13.0 VDC
16.0 VDC
0.94 A
1.15 A
15W
100mVp-p
±5%
85%
0.3W
WMIHPU15-107x
16.0 VDC
21.0 VDC
0.72 A
0.94 A
15W
100mVp-p
±5%
85%
0.3W
WMIHPU15-108x
21.0 VDC
27.0 VDC
0.55 A
0.72 A
15W
100mVp-p
±5%
86%
0.3W
WMIHPU15-109x
27.0 VDC
33.0 VDC
0.45 A
0.55 A
15W
100mVp-p
±3%
86%
0.3W
WMIHPU15-110x
33.0 VDC
40.0 VDC
0.37 A
0.45 A
15W
100mVp-p
±3%
86%
0.3W
WMIHPU15-111x
40.0 VDC
48.0 VDC
0.32 A
0.37 A
15W
100mVp-p
±3%
86%
0.3W
WMIHPU15-112x
48.0 VDC
55.0 VDC
0.28 A
0.32 A
15W
100mVp-p
±3%
86%
0.3W
80 ~ 275VAC
1.44 A
Total
Regulation
NOTES
1. The “x” in the model number can be “U” for United States type plug; “E” for Europe type plug, “A” for Australia type plug, or “K” for United Kingdom type
plug. Plugs are sold separately so please contact factory for ordering details.
2. Output can provide up to peak load when power supply starts up. Continuous staying in more than rated load is not allowed.
3. At factory, in 60% rated load condition, each output is checked to be within voltage accuracy.
4. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
5. Load regulation is defined by changing ±40% of measured output load from 60% rated load.
6. Ripple & noise is measured by using 20MHz limited bandwidth and with 0.47F capacitor in parallel across the output at nominal line and rated load.
7. Hold-up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
8. Optional output connectors are available for this series. Please call factory for ordering details.
08/04/2014
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 1 of 4
Wall Industries, Inc.
Rev A
WMIHPU15 SERIES
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies
TECHNICAL SPECIFICATIONS: WMIHPU15 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Input Voltage
Input Frequency
Input Current
Inrush Current
No Load Power Consumption
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation
Total Regulation
Output Power
Output Current
Ripple & Noise (peak to peak)
Hold-up Time
Start-up Time
Transient Response Time
Temperature Coefficient
PROTECTION
Short Circuit Protection
Over Load Protection
GENERAL SPECIFICATIONS
Efficiency
Dielectric Withstanding Voltage
Protection Classes
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Cooling
Flammability Rating
Vibration
Operating Altitude (Elevation)
MTBF
PHYSICAL SPECIFICATIONS
Weight
Dimensions (L x W x H)
Safety Approvals Input Voltage Range
Derating linearly from 100% load at 90VAC to 80% load at 80VAC
Sine wave
100VAC, full load
240VAC, full load
100VAC, full load, 25°C, cold start
240VAC, full load, 25°C, cold start
230VAC, no load
See Note 6
See Note 7
100~240VAC, full load
110VAC, Full load
Compliance
CEC & Energy Star
08/04/2014
12
0.3
Recovers automatically after fault condition is removed
110
Nominal line and rated load
Primary to Secondary, limit current <10mA
5656
Hz
A
A
1
%
2
4
+0.04
ms
s
ms
%/°C
Auto-recovery
150
%
See Table
VDC
Double insulated, Class II
Derating linearly from 100% Load at 40°C to 50% load at 70°C
10~95% RH
Non-condensing
10~500Hz, 10min./1cycle, 60 min each along X, Y, Z axes
-10
-40
0
0
+70
+85
95
95
Free air convection
UL94V-1
5
3000
MIL-HDBK-217F, 25°C
Safety Approvals
Compliance to EN55011 (CISPR11), EN61000-3-2,-3
Electro Static Discharge
VAC
See Table
See Table
See Table
See Table
-0.04
Output Connector
SAFETY, EMC, & COMPLIANCE
EMC Emission
Surge Voltage
240
275
63
0.6
0.4
35
85
See Table
See Table
See Note 4
See Note 5
“U” suffix
“K” suffix
“E” suffix
“A” suffix
AC Plug
100
80
47
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
100,000
°C
°C
%RH
%RH
G
m
hours
5.82oz (165g)
2.36 x 1.71 x 1.58 inches (60.0 x 43.5 x 40.2 mm)
United States
United Kingdom
Europe
Australia
Several options available
ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd ed.)
IEC60601-1 3rd edition
CE
B
Class
2
kV
8
kV
6
RoHS Compliant
CEC and Energy Star 2.0, Efficiency Level V
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
WMIHPU15 SERIES
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies
Rev A
Wall Industries, Inc.
DERATING
Input Voltage Derating Curve
Temperature Derating Curve
110
110
100
100
90
80
Load (%)
Load (%)
90
80
70
70
60
50
40
30
60
20
50
10
0
-10
80
100
120
140
160
180
200
Input Voltage (VAC)
220
240
260
280
0
10
20
30
40
Temperature (°C)
50
60
70
80
MECHANICAL DRAWING
1.58±0.04 [40.2±1.0]
OPTIONAL
2.36±0.04 [60.0±1.0]
1.71±0.04 [43.5±1.0]
United States Type
(Suffix “U”)
United Kingdom Type
(Suffix “K”)
Europe Type
(Suffix “E”)
NOTES:
1. Unit: inches [mm]
2. Tolerance: ±0.04 [±1.0]
3. Weight: 5.82oz (165g)
4. Models WMIHPU15-101~104 need to use AWG#18/4FT output cable in order to meet the total regulation specified.
Models WMIHPU15-105~109 need to use AWG#20/4FT output cable in order to meet the total regulation specified.
Models WMIHPU15-110~112 need to use AWG#22/4FT output cable in order to meet the total regulation specified.
The technical specifications will change if a different output cable is used.
5. Plugs are sold separately. Please call factory for ordering details.
6. Optional output connectors available. Please call factory for ordering details.
7. All dimensions are for reference only.
08/04/2014
AustraliaType
(Suffix “A”)
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 4
Rev A
Wall Industries, Inc.
WMIHPU15 SERIES
Up to 15 Watts, Medical AC/DC Wall Mount Power Supplies
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will
provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one
example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its
customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing
support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
08/04/2014
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 4