4205 Series Instruction Manual

IM-612
IM-612
Model 4205
Operation & Installation Manual
Model 4205 Series
MMIC Digital Attenuator
This documentation may not be reproduced in any form, for any
purpose unless authorized in writing by API Weinschel, Inc.
© API Weinschel, Inc.
Frederick, Maryland
2013, 2015
API
Weinschel
Manual
Rev. B (11-24-2015)
1
SAFETY SUMMARY
SAFETY SUMMARY
• To minimize shock hazard, the instrument chassis
DEFINITIONS.
The following definitions apply to WARNINGS,
CAUTIONS, and NOTES found throughout this manual.
WARNING
must be connected to an electrical ground. Using
the supplied three-conductor power cable ensures
that the instrument can be firmly connected to the
ac power source and electrical ground at a
grounded power outlet. If using a 3-2 wire
adapter be sure to connect the ground lead to earth
ground.
• Use the buddy system any time work involving
An operating or maintenance procedure, practice,
statement, condition, etc., which, if not strictly
observed, could result in injury and/or death of
personnel. Do not proceed beyond a WARNING
symbol until all the indicated conditions have been
fully understood and/or met.
active high voltage components is required. Turn
OFF the power before making/breaking any
electrical connection. Regard any exposed
connector, terminal board, or circuit board as a
possible shock hazard. DO NOT replace any
component or module with power applied.
• If test conditions to live equipment are required,
ground the test equipment before probing the
voltage or signal to be tested.
CAUTION
• Personnel working with or near high voltage
should be familiar with modern methods of
resuscitation.
An operating or maintenance procedure, practice,
statement, condition, etc., which, if not strictly
observed, could result in damage or destruction of the
equipment or long-term health hazards to personnel.
Do not proceed beyond a CAUTION symbol until all
the indicated conditions have been fully understood
and/or met.
NOTE
An essential operating or maintenance procedure,
condition, or statement that must be highlighted.
• DO NOT wear jewelry (rings, bracelets, metal
watches, and/or neck chains) while working on
exposed equipment. Be very cautious about
using hand tools near exposed backplanes, bus
bars, and/or power supply terminals. Use
properly insulated tools. When making test
connections to the power supply terminals and
bus bars, use only insulated probe tips.
• Verify that the instrument is set to match the
available line voltage and the correct fuse is
installed.
• DO NOT install substitute parts or perform any
GENERAL PRECAUTIONS.
The following are general precautions that are not related
to any specific procedure and, therefore, do not appear
elsewhere in this publication. These are precautions that
personnel must understand and apply during various phases of
instrument operation or service.
WARNING
unauthorized modification to this instrument.
Contact API Weinschel to acquire any information on replacement parts or returning the
instrument for repair. Unauthorized modification can cause injury to personnel and/or
destruction of the instrument.
• Operating personnel must not remove instrument
covers. Component replacement or adjustments
MUST BE performed by qualified service
personnel.
• DO NOT operate the instrument near or in the
presence of flammable gases or fumes.
• Potentially lethal voltages are present in this
instrument. Serious shock hazards from voltages
above 70 volts may exist in any connector,
chassis, or circuit board. Observe the following
precautions:
DETAILED PRECAUTIONS.
The following WARNINGS, CAUTIONS and NOTES
appear throughout the text of this manual and are repeated here
for emphasis.
i
SAFETY SUMMARY
NOTE
CAUTION
• All procedures and/or steps identified as
must be followed exactly as written and
according to industry accepted ESDS device
handling procedures. Failure to comply WILL
RESULT in ESDS damage.
• DO NOT use a nylon bristle brush in any solvent
as the bristles may dissolve and cause damage to
the circuit card or component.
DO NOT return any instrument or component to
Weinschel Corporation without receiving prior
factory authorization.
SAFETY SYMBOLS.
The following symbols are used to identify safety hazards
found throughout this publication and/or located on the
instrument.
• DO NOT use ultrasonic cleaning on parts or
assemblies containing electrical or electronic
components.
• DO NOT bend pins of electrical connectors
when using fiber-bristle brush.
CAUTION
HIGH VOLTAGE
• Compressed air used for cleaning and/or drying
can create airborne particles that may enter the
eye. Goggles/faceshields should be worn. DO
NOT direct air stream towards self or other
personnel. Pressure should be restricted to a
maximum of 15 psi to avoid personal injury.
• Under no circumstances should a wire brush, steel
wool, or abrasive compound be used on any
surface. Using these items will cause extensive
damage to the instrument's surface.
ii
WARNING
HIGH
VOLTAGE
Model 4205
IM-612
TABLE OF CONTENTS
1. GENERAL INFORMATION ................................................................................................................................ 3
1-1.
1-2.
1-3.
1-4.
1-5.
1-6.
1-7.
1-8.
1-9.
1-10.
PURPOSE .......................................................................................................................................3
SCOPE ...........................................................................................................................................3
EQUIPMENT DESCRIPTION ..............................................................................................................3
UNPACKING AND INSPECTION .........................................................................................................3
RESHIPMENT INSTRUCTIONS ..........................................................................................................3
STORAGE INSTRUCTIONS ...............................................................................................................4
RELATED MANUALS.........................................................................................................................4
ELECTROSTATIC DISCHARGE SENSITIVE (ESD) ................................................................................4
ABBREVIATIONS & ACRONYMS ........................................................................................................4
SAFETY CONSIDERATIONS .............................................................................................................4
2. SPECIFICATIONS .............................................................................................................................................. 5
2-1. GENERAL SPECIFICATIONS ...............................................................................................................5
2-2. CONFIGURATION MATRIX ...............................................................................................................5
3. INSTALLATION & OPERATION ................................................................................................................... 6-21
3-1. MOUNTING.....................................................................................................................................6
3-2. RF CONNECTORS & CABLE INSTALLATION .......................................................................................6
3-3. CONTROL CONNECTORS .............................................................................................................6-7
3-3.1. PARALLEL TTL CONNECTOR (J4) ...........................................................................................6
3-3.2. USB CONNECTOR (J3) ..........................................................................................................7
3-4. USB DRIVER INSTALLATION ...................................................................................................... 7-12
3-5. COMMAND OPERATION .......................................................................................................... 12-13
3-5.1 COMMAND REFERENCE ...................................................................................................... 13
3-5.2 APPLICATION SPECIFIC COMMANDS ............................................................................... 13-15
3-5.3. 488.2 COMMON COMMANDS ......................................................................................... 16-17
3-5.4. SETUP AND CONFIGURATION COMMANDS ......................................................................... 18
3-6. AW4205_CDC.INF INSTALLATION FILE ...................................................................................... 19-21
4. CONTACTING API WEINSCHEL, INC. . ........................................................................................................ 22
5. API WEINSCHEL WARRANTY ...................................................................................................................... 22
6. SPECIFICATION/ICD DRAWINGS
ICD, ATTENUATOR PROGRAMMABLE, MODEL 4205-31.5 ........................................................089-4451B
ICD, ATTENUATOR PROGRAMMABLE, MODEL 4205-63.5 ........................................................089-4460A
ICD, ATTENUATOR PROGRAMMABLE, MODEL 4205-95.5 ........................................................089-4465F
API Weinschel
2
Model 4205
IM-612
1. GENERAL INFORMATION:
1-1 PURPOSE: This manual contains setup and operation information for the
API Weinschel Model 4205 Series, MMIC Digital Attenuators.
1-2 SCOPE: This manual is to be used in conjunction with the operation and
installation of the Model 4205 Series. The manual also provides a description of
the assembly; block diagrams: and general maintenance procedures to maintain
the assembly.
1-3 EQUIPMENT DESCRIPTION: API Weinschel Model 4205 series of MMIC
Digital Attenuators operates over the 0.2 to 6 GHz frequency range and is in a
variety of attenuation ranges up to 95.5 dB in 0.5 dB steps. These units can be
controlled using either standard TTL or USB interfaces.
1-4. UNPACKING AND INSPECTION: Upon unpacking the equipment, retain the shipping container and packing
material for future shipment for recalibration. Perform the following initial inspection:
a. Carefully look at the outside of the shipping container for discoloration, stains, charring, or other signs of
exposure to excessive heat, moisture, or liquid chemicals. Check for any physical damage to the shipping
container such as dents, snags, rips, crushed sections or areas, or similar signs of excessive shock or
careless handling.
b. With the equipment and any accessory package removed from the shipping container, check each item
against the packing list or Items Supplied List. If any items are missing, contact the API Weinschel
Customer Service Department.
c.
Carefully inspect the equipment looking for dents, deep scratches, damaged or loose connector, or any
other signs of physical abuse or careless handling. If damage is found, forward an immediate request to
the delivering carrier to perform an inspection and prepare a concealed-damage report. DO NOT destroy
any packing material until it has been examined by an agent of the carrier. Concurrently, report the nature
and extent of damage to API Weinschel, giving equipment model and serial numbers, so that necessary
action can be taken. Under U.S. shipping regulations, damage claims must be collected by the
consignee; DO NOT return the equipment to API Weinschel until a claim for damages has been
established.
1-5. RESHIPMENT: Use the best packaging materials available to protect the unit during storage or reshipment.
When possible, use the original packing container and cushioning material. If the original packing materials are not
available, use the following procedure:
a. Wrap the storage cases in sturdy paper or plastic;
b. Place the wrapped storage cases in a strong shipping container and place a layer of shock-absorbing material
(3/4 inch minimum thickness) around all sides of the unit to provide a firm cushion and to prevent movement
inside the container.
c.
If shipping the unit for service, attach a tag to indicate:
1.
2.
3.
4.
5.
6.
model and serial numbers
service required
description of malfunction
return address
authorization to conduct repairs
return authorization number
d. Thoroughly seal the shipping container and mark it FRAGILE. Ship to:
API Weinschel, Inc.
Attn: Customer Service Department
5305 Spectrum Drive
Frederick, MD 21703-7362
or to an authorized sales representative.
API Weinschel
3
Model 4205
IM-612
1-6. STORAGE: Storage of the Model 4205 is possible for extended periods without incurring damage to internal
circuitry if the Model 4205 is packaged according to the instructions above. The safe limits for storage environment
are as follows:
Temperature:
Humidity:
-67° to +167 °F (-55° to +75°C)
5% to 85% (non-condensing)
1-7. RELATED MANUALS: The following manuals contain information that may be used in conjunction with this
manual to operate, service, or calibrate this instrument.
Title
Manual
IM-611
Aeroflex USB Control Center Software, P/N 189-164
1-8. ELECTROSTATIC DISCHARGE SENSITIVE: The equipment documented in this manual contains certain
Electrostatic Discharge Sensitive (ESDS) components or parts. Therefore, certain procedures/steps are identified by
the use of the symbol .
. This symbol is used in two ways:
CAUTION
All procedures and/or steps identified as must be followed exactly as written and according to
accepted ESDS device handling procedures. Failure to comply WILL RESULT in ESDS damage.
a. When the ESDS symbol is placed between a paragraph number and title
, all of that
paragraph, including all subparagraphs, is considered ESDS device handling procedure.
b. When the ESDS symbol is placed between a procedure/step number and the text
that procedure is considered an ESDS device handling procedure.
, all of
1-9. ABBREVIATIONS AND ACRONYMS: The following list contains abbreviations used throughout this manual.
Abbreviations and acronyms that are not listed conform to MIL-STD-12D.
ESDS
TBD
Electrostatic Discharge Sensitive
To Be Determined
1-10. SAFETY CONSIDERATIONS: The Model 4205 Series and all related documentation must be reviewed for
familiarization with safety markings and procedures before any operation and/or service. Refer to the SAFETY
SUMMARY located at the beginning of this manual for a summary of safety information and procedures. Following
these simple safety precautions will ensure safe operation and service of the unit.
API Weinschel
4
Model 4205
IM-612
2. SPECIFICATIONS:
2-1. GENERAL SPECIFICATIONS:
Power Requirements
5 VDC @ 50 mA typical
Environmental
Operating Temperature
Storage Temperature:
Humidity:
Altitude:
0° to +70°C
-67° to +167 °F (-55° to +75°C)
5% to 85% (non-condensing)
40,000' (12,192M)
Parallel TTL
Connector: 10 pin .025 square post header on .1 center. Mates with Amp connector
746285-1 or equivalent (one mating connector included with each unit).
USB 2.0
Connector:
RF Characteristics &
Physical Dimensions
Refer to API Weinschel ICD/Specification Drawings:
USB, MINI B
Model
4205-31.5
4205-63.5
4205-95.5
Specification/ICD #
089-4451
089-4460
089-4465
2-2. CONFIGURATION MATRIX:
Model Number
Attenuation Cells (dB)
Attenuation
Range / Step Size
4305-31.5
0.5, 1, 2, 4, 8 & 16
31.5 in 0.5 dB steps
4305-63.5
0.5, 1, 2, 4, 8, 16 & 32
63.5 in 0.5 dB steps
4305-95.5
0.5, 1, 2, 4, 8, 16, 32 & 32
95.5 in 0.5 dB steps
API Weinschel
5
Model 4205
IM-612
3. INSTALLATION & OPERATION:
3-1. MOUNTING: Each Digital Attenuator is supplied with 12 mounting holes (4-40 UNC-2B x 4.1 [0.16 Deep])
located on the front, back, top and bottom end of the attenuator. Refer to the appropriate API Weinschel
Specification/ICD drawing for mounting hole location.
Figure 1. Connector Location
CAUTION

When applying an RF signal to the RF INPUT connector, DO NOT exceed the maximum allowable
power level specifications of the unit.

Do not over torque the SMA connectors more than 7 inch pounds. Damage may occur.
3-2. RF CONNECTORS & CABLE INSTALLATION: The Model 4205 series contains two SMA female connectors
labeled J1 and J2 that mate nondestructively with SMA male connectors per MIL-STD-39012. Care should be taken
to prevent strain on the interconnecting cables, since damage here may not always be apparent. Check the attaching
RF cables and the SMA connectors for signs of cracked insulation and/or bent or worn pins prior to connection. Tests
show that connectors must be clean for accuracy and stability. This requires an inspection and cleaning of each
connector immediately before use. When cleaning precautions are observed regularly, connectors can maintain
their stability for over several thousand connection cycles. API Weinschel recommends a torque value of 7 to 10
inch pounds when connecting any RF cable to the attenuator's RF connectors.
3-3 CONTROL CONNECTORS: Model 4205 attenuators can be controlled in one of two modes: parallel TTL or
USB. The TTL control connector is a standard 10-pin ribbon cable connector that mates with an AMP 746285-1 style
connector. The USB connector is a standard 5-pin USB mini B series. NOTE: Only one interface should be connected
at a time. DO NOT supply a connection to both control ports simultaneously.
3-3.1. Parallel TTL Connector (J4): In parallel TTL mode up to eight TTL control signals (model dependent) are
used to activate or deactivate a particular attenuation cell. A TTL high will enable the attenuation state for the cell,
whereas a TTL low will maintain the cell in its zero attenuation state. In addition, there are two DC supply voltage pins
which are used to supply power to the unit.
API Weinschel
6
Model 4205
IM-612
Pin
4205-31.5
Signal
4205-63.5
Signal
4205-95.5
Signal
1
0.5 dB
0.5 dB
0.5 dB
2
1 dB
1 dB
1 dB
3
2 dB
2 dB
2 dB
4
4 dB
4 dB
4 dB
5
8 dB
8 dB
8 dB
6
16 dB
16 dB
16 dB
7
NC
32 dB
32 dB
8
NC
NC
32 dB
9
+5VDC
+5VDC
+5VDC
10
GND
GND
GND
Note: NC pins are Do Not Connect
3-3.2. USB Connector (J3): In USB mode the attenuator is controlled and powered via a standard USB 2.0
connection to a USB host. The 4205 operates as a USB CDC device, so it may be controlled via any software that
can communicate to a standard COM port. Programming is done via simple ASCII strings to control the device (see
the Command section paragraph 3-4).
For ease of use, the 4205 has two modes of operation: console and raw mode. Console mode provides a simple
command-line based interface that can be used in conjunction with any standard terminal emulator program. Console
mode sends command prompts ('>'), echoes received characters, issues error messages, and supports the
backspace key for simple editing, while raw mode is more suitable for programming. By default, the unit is shipped
with Console mode enabled, but this operation can be change by the user (see the CONSOLE command for more
details).
3-4. USB DRIVER INSTALLATION: When you connect the 4205 to the USB for the first time, you should be
presented with the New Hardware Wizard.
Follow the steps shown below to install the USB CDC .inf file. (NOTE: A copy of the INF information is included in
paragraph 3-6 of this document. If you do not have an electronic copy, you can create one using Notepad. Copy and
paste the information into Notepad and save it as a plain text file with the name AW4205_CDC.INF)
API Weinschel
7
Model 4205
API Weinschel
IM-612
8
Model 4205
IM-612
Navigate to the drive/folder containing the AW4205_CDC.INF file, and select 'Next'
API Weinschel
9
Model 4205
API Weinschel
IM-612
10
Model 4205
IM-612
To verify that the driver has installed properly, view the Ports section in Device Manager. You should see the 4205
listed as a USB COM port. Note the assigned COM port number.
Using a terminal emulator, open a connection to the COM port shown above. The default COM port settings should be
acceptable, as these are unused by the 4205. If the 4205 is in Console mode (the default), you should see the signon
message.
API Weinschel
11
Model 4205
IM-612
The COM port numbers are assigned by Windows based on the device USB VID, PID, and Serial Number. The 4205
uses different USB PID codes for each of the three models as shown below:
Model
USB PID
4205-31.5
106F
4205-63.5
106E
4205-95.5
106D
By default, the attenuators are all shipped with the USB serial number set to '000000'. If you are using a single
attenuator of a given model, you can ignore this setting. If you wish to use multiple attenuators of the same type, you
can change the serial number to a unique setting so that each attenuator is assigned a different COM port. This can
be done through the Console using the IDN command, which allows the user to change the serial number/ID to any 6character setting. As the units are serialized at the factory on the product label, you may wish to use this number for
easy identification.
To assign a new serial number to the attenuator, connect the unit as normal. It is recommended to do this with only a
single attenuator connected to the USB to avoid any confusion. Note the serial number located on the product label
(ie 100), and send the command 'IDN serialno', replacing serialno with the actual desired string. Unplug and reattach
the attenuator's USB connection. You will have to reinstall the driver (this time you can let Windows install it
automatically), after which it should be re-enumerated with a different COM port number.
3-5. COMMAND OPERATION: Commands are comprised of text-based ASCII strings. The command parser is
case-insensitive, so either upper or lower case characters are acceptable. Command parameters may be separated
with either an ASCII SPACE char (0x20) or an ASCII COMMA char (0x2E), but the separator character used must be
the same within an individual command string. Additional SPACE characters are ignored. Input program messages
may be terminated using either an ASCII CR character (0x0D) or an ASCII LF character (0x0A). Command message
strings are limited to 64 characters total, including the terminator. Multiple commands can be included in one
message by separating the individual commands with an ASCII SEMICOLON character ';' (0x3B), up to the 64
character message limit. Typically, Response messages sent from the device are terminated using both a CR (0x0D)
and LF (0x0A) to terminate the message. The output terminator sequence may be changed using the RMT command.
A list of supported commands can be seen by typing 'HELP' at the Console prompt.
The command structure/operation is similar to that used in IEEE 488.2, and includes some of the 488.2 Common
Commands such as *IDN?, *RST, *CLS, and *OPC?, in addition to device specific commands. In 488.2, programming
commands take one of two forms: a Program message or a Query message. Program messages are used to send
commands to the device, while Query messages are used to elicit a response. Query commands are those that
contain a '?' character. In general, the device does not generate any response to a program message unless the
message contains a valid Query command. (Note that this does not apply when operating in Console mode, or when
using some commands such as HELP which are designed to provide the user general information). You can use this
feature to provide a method to synchronize command execution with the controller by appending a Query to the
desired command, and waiting for the response. For example, sending "*CLS;*OPC?" will place a "1" in the output
queue when the *CLS command has been executed. Query commands that return multiple values will have the
values separated by an ASCII COMMA character (0x2E). If multiple Query commands are included in the same
message, the individual query responses will be separated with an ASCII SEMICOLON character (0x3B).
An Error Queue is provided that logs the results of command/execution errors in a FIFO fashion. The queue entries
can be read using the ERR? command, which returns both an error code and a descriptive text message, such as
101, "invalid command"
When the queue is empty, ERR? returns the message 0, "no error". The queue can be emptied by repeatedly
API Weinschel
12
Model 4205
IM-612
sending ERR? until all entries are read from the queue, or via sending the *CLS message.
Unless otherwise specified, commands revert to their default setting at system reset/poweron, with the exception of
the system setup and configuration commands which store their setting in non-volatile memory (NVM).
3-5.1. Command Reference: In the command descriptions that follow, argument types are described using the
following additional conventions to indicate the relative size of the parameter:
byte
word
int8
int16
int32
string
-
used to indicate an 8-bit unsigned integer
used to indicate a 16-bit unsigned integer
8-bit integer
16-bit integer
32-bit integer
character data, including the max number of characters allowable. (ie string8 has a max of 8 chars)
Numeric arguments default to decimal (base 10) notation, but may optionally be provided in hex if appropriate by
using a "0x" prefix (ie 0x0a = 10)
Required command keywords are shown in CAPITAL letters, and arguments are shown in italics. Square brackets '[]'
may be used to indicate an optional parameter, for example [select]. Optional parameters, if not supplied by the user,
assume the default setting specified in the text.
3-5.2 Application Specific Commands: Attenuation value settings are specified in dB, with up to 1 digit of
precision after the decimal point for attenuators that support step sizes of < 1dB. When specifying integral dB values,
usage of the decimal point format is strictly option (ie '10' is the same as '10.0'). The attenuation setting must be a
multiple of the attenuators intrinsic step size or the command will generate an error. For example, an attenuator that
has a stepsize of 0.5 dB will accept settings of 0.5, but will generate an error if set to 0.3. For responses, attenuation
values will be formatted to the base precision of the attenuator.
Attenuation switching speed requirements are built into the command execution time and are always
enforced such that if you execute a command sequence such as "ATTN 10;*OPC?" the response will not be sent until
the attenuator has been programmed and has changed state. The attenuator can be switched at the maximum rate
that commands are executed, which is typically in the 1-2ms range.
ATTN
Function:
set attenuator
Syntax:
ATTN setting
attenuator setting, in dB. setting=0-max attenuation value, or MAX
Argument(s): setting
Remarks:
This command sets the RF attenuator to the dB value provided by setting. If setting is MAX, then the
maximum dB value for that attenuator will be used.
Return Value: none
Example(s):
ATTN 10
// sets attn to 10 dB
ATTN MAX
// sets attn to max (95.5dB)
ATTN 15.5
// sets attn to 15.5 dB
ATTN?
Function:
read attenuator setting
Syntax:
ATTN?
Argument(s): none
Remarks:
This command returns the current setting of the attenuator
Return Value: attenuator setting, in dB
Example(s):
ATTN 10
// sets attn to 10 dB
ATTN?
// read attn setting
10.0
// returns attn setting (10 dB)
API Weinschel
13
Model 4205
IM-612
STEPSIZE
Function:
set attenuator stepsize
Syntax:
STEPSIZE setting
attenuator stepsize, in dB. Setting=0-max attenuation value
Argument(s): setting
Remarks:
This command sets the attenuation stepsize. This command can be used with the INCR, DECR, and
FADE commands to change the step value. Specifying a setting of 0 sets the stepsize to the intrinsic step value for
the attenuator, in effect removing any previous STEPSIZE command.
Return Value: none
Example(s):
STEPSIZE 10
// sets attn stepsize to 10dB
STEPSIZE 0
// sets the stepsize back to native value (ie 0.5dB)
STEPSIZE?
Function:
read attenuator stepsize setting
Syntax:
STEPSIZE?
Argument(s): none
Remarks:
This command returns the current stepsize
Return Value: attenuator setting, in dB
Example(s):
STEPSIZE 10
// sets attn stepsize to 10dB
STEPSIZE?
// read attn stepsize setting
10.0
// returns attn stepsize (10dB)
INCR
Function:
increment attenuator setting
Syntax:
INCR
Argument(s): none
Remarks:
This command increments the current setting of the attenuator by the STEPSIZE setting.
Return Value: none
Example(s):
ATTN 5
// sets attn to 5dB
STEPSIZE 10
// sets attn stepsize to 10dB
INCR
// increments attn by current stepsize (5dB + 10dB = 15dB)
DECR
Function:
decrement attenuator setting
Syntax:
DECR
Argument(s): none
Remarks:
This command decrements the current setting of the attenuator by the STEPSIZE setting.
Return Value: none
Example(s):
ATTN 15
// sets attn to 15dB
STEPSIZE 10
// sets attn stepsize to 10dB
DECR
// decrements attn by current stepsize (15dB - 10dB = 5dB)
API Weinschel
14
Model 4205
IM-612
FADE
FADE?
Function:
Syntax:
fade attenuation value over time
FADE start_db end_db interval
FADE? start_db end_db interval
initial attenuation setting
Argument(s): start_db
end_db
final attenuation setting
dwell time per increment, in msec (max=10000)
interval
Remarks:
This command ramps the attenuation value at a user specified rate. The start_db and end_db values
can be any settings valid for the attenuator, and the attenuation will ramp up or down as required. The attenuation
value will step according to the current STEPSIZE setting. Attenuation values are changed at the rate specified by
interval, which is in units of msec. Note that no other commands can be executed until the fade operation is
completed. The query form of the command (FADE?) will return the current attenuation values as the steps are
executed.
Return Value: none
Example(s):
STEPSIZE 1.0
// set stepsize to 1 dB
FADE 10.0 85.0 500
// fade from 10dB to 85dB in STEPSIZE increments at a rate of 500ms
STEPSIZE 0
// set stepsize to default (0.5dB)
FADE? 90 0 1000
// fade from 90dB to 0dB in 0.5dB steps at a rate of 1000ms (1sec)
90.0
// values are sent as the attenuation changes
89.5
….
0.0
RFCONFIG?
Function:
displays current RF configuration
Syntax:
RFCONFIG?
Argument(s): none
Remarks:
This command displays the current attenuator configuration, including the model, max attn, default
stepsize, and frequency range
Example(s):
RFCONFIG?
4205-95.5, 95.5, 0.5, 0.2-6GHz
API Weinschel
15
Model 4205
IM-612
3-5.3. 488.2 COMMON COMMANDS:
*CLS
Function:
Syntax:
Argument(s):
Remarks:
Return Value:
Example(s):
*CLS
clears the error status
*CLS
none
This function is used to clear the Error Queue
none
*IDN?
Function:
Reads the system identification information
Syntax:
*IDN?
Argument(s): none
Remarks:
This function is used to read the system identification info, which is a string consisting of the following
data: manufacturer, model, serial number, and firmware version.
Return Value: idstr
string
id info
Example(s):
*IDN?
Aeroflex Weinschel, 4205, 002, V1.00
*OPC?
Function:
Operation complete query
Syntax:
*OPC?
Argument(s): none
Remarks:
This function loads a '1' into the output queue when the Program Message Unit is executed. It’s
primary use is to provide an indication of command completion by including the command as the last one in a series
of commands. It can be useful to synchronize operation and to prevent input buffer overflow.
Return Value: 1
integer constant
command completed
Example(s):
CMD1 1; CMD2 2; *OPC?
1
// sends a '1' response when the three commands have been parsed and executed
*ESR?
Function:
Syntax:
Argument(s):
Remarks:
Return Value:
Example(s):
*ESR?
32
*RST
Function:
Syntax:
Argument(s):
Remarks:
Return Value:
Example(s):
*RST
API Weinschel
Event Status Register query
*ESR?
none
This function reads the 488.2 Event Status Register. Reading the register also clears it.
int8
integer
status register
// indicates a Command Error
Performs a device application level reset.
*RST
none
This function is used to reset the device application settings.
none
16
Model 4205
IM-612
*TST?
Function:
Self-test query
Syntax:
*TST?
Argument(s): none
Remarks:
This function performs an internal self-test. Upon completion, the results of the test are loaded into
the output queue.
Return Value: testresults
integer
'0' indicates test passed. Non-zero indicates test failed.
Example(s):
*TST?
0
// returns a '0' when the test completes successfully.
ERR?
Function:
Read the Error Queue
Syntax:
ERR?
Argument(s): none
Remarks:
This function returns the last entry in the error status queue, and a string description of the error code.
Repeating the command will return the next entry, until the error queue is empty and returns a zero. The error queue
may be cleared via the *CLS command. Note that when using the command-line interface the Error Queue contents
are automatically displayed after each command prior to issuing the CLI prompt.
Return Value: error number, "error description"
Example(s):
ERR?
101, "invalid command"
ERR?
0, "no error"
API Weinschel
17
Model 4205
IM-612
3-5.4. Setup and Configuration Commands
RMT
Function:
sets the Response Message Terminator end of string character(s)
Syntax:
RMT val
Argument(s): val
word, eos characters
Remarks:
This function sets the output Response Message Terminator (RMT) sequences. The val parameter
specifies the character sequence used, and can specify up to two characters, typically as a hex word high byte-low
byte pair. Common definitions for the terminators include the ASCII CR (0x0D) and LF (0x0A) characters. A single
character may be specified either by using 0 for the high byte, such as 0x000D, or by only specifying a single
character (ie 0x0D). On output the characters are sent low byte then high byte, unless it is specified as 0. Note that
the CONSOLE mode will always use a fixed CRLF (0x0A0D) sequence. This value is stored in non-volatile memory
(NVM), and does not take effect until the next system restart event.
Return Value: none
Example(s):
RMT 0x0A0D
// set output sequence as CR-LF
RMT 0x0D
// set output sequence as a single CR character
CONSOLE
Function:
Console mode enable
Syntax:
CONSOLE mode
Argument(s): mode
OFF, ON, ENABLE, or DISABLE
OFF
turns console mode off (and updates NVM setting)
ON
turns console mode on (and updates NVM setting)
ENABLE
enables console mode
DISABLE
disables console mode
Remarks:
This function enables/disables the console mode command-line interface and optionally updates the
nvm setting. Setting mode parameter OFF or ON updates the NVM default setting, while modes ENABLE and
DISABLE do not.
Return Value: none
Example(s):
CONSOLE ON
// turns on the console and updates nvm setting
CONSOLE ENABLE
// turns on console for this session only
CONSOLE DISABLE // turns off console for this session only
IDN
Function:
Identification serial number setup
Syntax:
IDN serno
Argument(s): serno
6-character string
Remarks:
This command sets the serial number parameter returned by the *IDN? Command, and also used as
the USB serial number. Alpha-numeric values are allowed, up to a max of 6 characters. The value is saved in NVM.
Return Value: none
Example(s):
IDN 001
API Weinschel
18
Model 4205
IM-612
3-6. AW4205_CDC.INF INSTALLATION FILE
; Windows USB CDC ACM Setup File
; Based on INF template which was:
; Copyright (c) 2000 Microsoft Corporation
; Copyright (C) 2004 Microchip Technology Inc.
[Version]
Signature="$Windows NT$"
Class=Ports
ClassGuid={4D36E978-E325-11CE-BFC1-08002BE10318}
Provider=%MFG%
LayoutFile=layout.inf
DriverVer=01/07/2007,5.1.2800.0
[Manufacturer]
%MFG%=DeviceList, NTx86, NTia64, NTamd64
[DestinationDirs]
DefaultDestDir=12
;-----------------------------------------------------------------------------; Vendor and Product ID Definitions
;-----------------------------------------------------------------------------; Note: One INF file can be used for multiple devices with different
;
VID and PIDs. For each supported device, append
;
",USB\VID_xxxx&PID_yyyy" to the end of the line.
;-----------------------------------------------------------------------------[SourceDisksFiles]
[SourceDisksNames]
[DeviceList.NTx86]
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106D
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106E
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106F
; 4205-95.5
; 4205-63.5
; 4205-31.5
[DeviceList.NTia64]
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106D
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106E
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106F
; 4205-95.5
; 4205-63.5
; 4205-31.5
[DeviceList.NTamd64]
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106D
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106E
%DESCRIPTION%=DriverInstall, USB\VID_25EA&PID_106F
; 4205-95.5
; 4205-63.5
; 4205-31.5
;-----------------------------------------------------------------------------; 32 bit section for Windows 2000/2003/XP/Vista
;-----------------------------------------------------------------------------[DriverInstall.NTx86]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
API Weinschel
19
Model 4205
IM-612
AddReg=DriverInstall.NTx86.AddReg
[DriverInstall.NTx86]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
AddReg=DriverInstall.NTx86.AddReg
[DriverCopyFiles]
usbser.sys,,,0x20
[DriverInstall.NTx86.AddReg]
HKR,,DevLoader,,*ntkern
HKR,,NTMPDriver,,usbser.sys
HKR,,EnumPropPages32,,"MsPorts.dll,SerialPortPropPageProvider"
[DriverInstall.NTx86.Services]
AddService=usbser, 0x00000002, DriverService
;-----------------------------------------------------------------------------; 64 bit section for Intel Itanium based systems
;-----------------------------------------------------------------------------[DriverInstall.NTia64]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
AddReg=DriverInstall.NTia64.AddReg
[DriverInstall.NTia64]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
AddReg=DriverInstall.NTia64.AddReg
[DriverCopyFiles]
usbser.sys,,,0x20
[DriverInstall.NTia64.AddReg]
HKR,,DevLoader,,*ntkern
HKR,,NTMPDriver,,usbser.sys
HKR,,EnumPropPages32,,"MsPorts.dll,SerialPortPropPageProvider"
[DriverInstall.NTia64.Services]
AddService=usbser, 0x00000002, DriverService
;-----------------------------------------------------------------------------; 64 bit section for AMD64 and Intel EM64T based systems
;-----------------------------------------------------------------------------[DriverInstall.NTamd64]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
AddReg=DriverInstall.NTamd64.AddReg
[DriverInstall.NTamd64]
include=mdmcpq.inf
CopyFiles=DriverCopyFiles
API Weinschel
20
Model 4205
IM-612
AddReg=DriverInstall.NTamd64.AddReg
[DriverCopyFiles]
usbser.sys,,,0x20
[DriverInstall.NTamd64.AddReg]
HKR,,DevLoader,,*ntkern
HKR,,NTMPDriver,,usbser.sys
HKR,,EnumPropPages32,,"MsPorts.dll,SerialPortPropPageProvider"
[DriverInstall.NTamd64.Services]
AddService=usbser, 0x00000002, DriverService
;-----------------------------------------------------------------------------;
;-----------------------------------------------------------------------------[DriverService]
DisplayName=%SERVICE%
ServiceType=1
StartType=3
ErrorControl=1
ServiceBinary=%12%\usbser.sys
;-----------------------------------------------------------------------------; String Definitions
;-----------------------------------------------------------------------------[Strings]
MFG="AeroflexWeinschel"
DESCRIPTION="Weinschel 4205 USB COM Port"
SERVICE="USB RS232 Emulation Driver"
API Weinschel
21
Model 4205
IM-612
4. CONTACTING API WEINSCHEL, INC.:
In the event of a malfunction, contact API Weinschel. An apparent malfunction of an instrument or
component may be diagnosed over the phone by first contacting the Customer Service Department at API Weinschel.
DO NOT send the instrument or component back to the factory without prior authorization. When it is necessary to
return an item, state the symptoms, catalog and type number of the instrument or component, and date of original
purchase. Also write the Company name and your name and phone number on a card and tape the card to the item
returned. Page provides further information regarding preparation of a unit for reshipment. Contact API Weinschel
Customer Service Department as follows:
Via mail:
API Weinschel, Inc.
5305 Spectrum Drive
Frederick, MD 21703-7362
U.S.A.
Via Telefax:
301-846-9116
Via Phone:
Call TOLL FREE 800-638-2048
Toll call # 301-846-9222
Via Website: www.aeroflex.com/weinschel
Via e-mail:
[email protected]
5. API WEINSCHEL INC. WARRANTY:
PRODUCTS - API Weinschel, Inc. warrants each product it manufactures to be free from defects in material and
workmanship under normal use and service anywhere in the world. API Weinschel only obligation under this Warranty
is to repair or replace, at its plant, any product or part thereof that is returned with transportation charges prepaid to
API Weinschel by the original purchaser within ONE YEAR from the date of shipment.
The foregoing Warranty does not apply in API Weinschel sole opinion to products that have been subject to improper
or inadequate maintenance, unauthorized modifications, misuse, or operation outside the environmental specifications
for the product.
SOFTWARE PRODUCTS- API Weinschel, Inc. software products are supplied without representation or Warranty of
any kind. API Weinschel, therefore, assume no responsibility and will not accept liability (consequential or otherwise)
arising from the use of program materials, disk, or tape.
The Warranty period is controlled by the Warranty document furnished with each product and begins on the date of
shipment. All Warranty returns must be authorized by API Weinschel prior to their return.
API Weinschel’s Quality System Certified to:
© 2013-2015, API Weinschel, Inc.
API Weinschel
22
4
3
NOTES:
2
PIN 1
PIN 5
1. ALL DIMENSIONS ARE GIVEN IN mm [in.]
D
2. ALL MATERIALS AND PROCESSES ARE TO BE IN COMPLIANCE WITH
THE EUROPEAN DIRECTIVE RESTRICTION OF HAZARDOUS
SUBSTANCES (RoHS) (REF: A/W 080-638)
9.5
[.38]
8.4
[.33]
10.2
[.40]
9.0
[.36]
DETAIL A
SCALE 4 : 1
63.5
[2.50]
43.18
[1.700]
9.5
[.38]
5.6
[.22]
SEE
DETAIL
A
47.5
[1.87]
6.6
[.26]
9.0
[.36]
4X 2-56 UNC-2B
X 4.1[.16 DEEP]
2X 4-40 UNC-2B
X 4.1[.16 DEEP]
2.2 10.9
[.09] [.43]
SPECIFICATIONS:
Frequency:
Impedance:
Attenuation:
Attenuation Cells :
Switching Time:
SWR:
A
17.1
[.68]
200-6000 MHz
50 ohms nominal
0 to 31.5 dB in 0.5 dB step size
0.5,1,2,4,8,16 dB
< 1 microsecond
0.2 - 5GHz: 1.5:1 max
5 - 6GHz: 1.7:1 max
Insertion Loss:
0.2 - 3GHz: 3dB max
3 - 6GHz: 4dB max
Accuracy:
0.5 to 7.5 dB;
+/- 0.5 dB
8 to 11.5 dB; +/- 1.0 dB
12 to 31.5 dB ; +/- 1.25 dB
Supply Voltage :
5 Volts @ 50 mA typical
Control Logic :
TTL or USB
RF Power Rating: +23 dBm
Temp Range:
Connectors:
Monotonicity:
29.21
[1.150]
0°C - 70°C
SMA Female
200-5000 MHz (min of 0.5 dB step)
5000-6000 MHz (min of 1.0 dB step)
4
D
43.18
[1.700]
C
USB CONTROL CONNECTOR
J3 PIN LOCATIONS:
PIN No. (J3)
FUNCTION
V BUS (+5V)
1
2
D3
D+
ID (NC)
4
5
GROUND
TTL CONTROL CONNECTOR
J4 PIN LOCATIONS:
PIN No. (J4) 4205-31.5 (CELL)
0.5
1
2
1
3
2
4
4
5
8
6
16
7
NC
8
NC
+5V
9
10
GND
6.6
[.26]
SPECIFICATIONS CONTINUE:
CONTROL CONNECTORS:
1) AMP latch 10 pin ribbon
cable connector mates
with AMP P/N 746285-1
(supplied with each unit)
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
.XX
.01
1/2
1/64
.XXX
.003
MATERIAL
PROPRIETARY NOTICE
3
B
INTERFACE CONTROL DRAWING
2) USB - Mini-B
THIS DOCUMENT IS COMMERCIAL
CONFIDENTIAL AND THE INFORMATION
CONTAINED THEREIN MAY NOT BE DISSEMNATED, NOR IS REPRODUCTION
PERMITTED EXCEPT BY WRITTEN
AUTHORIZATION.
10.2
[.40]
2.2
[.09]
13.8
[.54]
B
17.1
[.68]
59.18
[2.330]
44.3
[1.75]
10.9
[.43]
29.21
[1.150]
APPROVED
PIN 2
59.18
[2.330]
63.5
[2.50]
2X 4-40 UNC-2B
X 4.1[.16 DEEP]
4X 2-56 UNC-2B
X 4.1[.16 DEEP]
PIN 1
CAGE: 93459
MODEL: 4205-31.5
SERNO: XXXX
0.2-6GHz 0-31.5dB / 0.5dB
C
1
REVISION HISTORY
ZONE REV
DESCRIPTION
DATE
A
ERN 13-022
FEB 13
B ECN 13-188
RS MAR 2013
101-2463
NEXT ASSY
4205-31.5
USED ON
APPLICATION
N/A
CONTRACT NO.
APPROVALS
1/18/2013
CHECKED
1/21/2013
A. HOPKINS
FINISH
N/A
ENGR.
Z. DARMON
DO NOT SCALE DRAWING
2
DATE
DRAWN
R. SHOCK
ISSUED
A
Aeroflex / Weinschel
1/21/2013
ICD, ATTENUATOR PROGRAMMABLE
MODEL 4205-31.5
SIZE FSCM NO.
C
SCALE
93459
2:1
DWG NO
REV
B
089-4451
SHEET
1
1
OF
1
4
3
2
NOTES:
PIN 1
PIN 5
1. ALL DIMENSIONS ARE GIVEN IN mm [in.]
D
2X 4-40 UNC-2B
2. ALL MATERIALS AND PROCESSES ARE TO BE IN COMPLIANCE WITH
THE EUROPEAN DIRECTIVE RESTRICTION OF HAZARDOUS
SUBSTANCES (RoHS) (REF: A/W 080-638)
10.2
[.40]
9.5
[.38]
5.6
[.22]
CAGE: 93459
MODEL: 4205-63.5
SERNO: XXXX
0.2-6GHz 0-63.5dB / 0.5dB
47.5
[1.87]
9.0
[.36]
6.6
[.26]
4X 2-56 UNC-2B
2X 4-40 UNC-2B
4.1[.16]
10.9
[.43]
4.1[.16]
2.2
[.09]
Frequency:
Impedance:
Attenuation:
Attenuation Cells :
Switching Time:
SWR:
17.1
[.68]
200-6000 MHz
50 ohms nominal
0 to 63.5 dB in 0.5 dB step size
0.5,1,2,4,8,16,32 dB
< 1 microsecond
0.2 - 0.8GHz: 1.8:1 max
0.8 - 6GHz: 1.5:1 max
Insertion Loss:
0.2 - 3GHz: 4.5dB max
3 - 6GHz: 6dB max
Accuracy:
0.5 to 7.5 dB;
+/- 0.5 dB
8 to 11.5 dB; +/- 1.0 dB
12 to 63.5 dB ; +/- 1.25 dB
Supply Voltage :
5 Volts @ 50 mA typical
Control Logic :
TTL or USB
RF Power Rating: +23 dBm
29.21
[1.150]
0°C - 70°C
SMA Female
200-5000 MHz (min of 0.5 dB step)
5000-6000 MHz (min of 1.0 dB step)
Temp Range:
Connectors:
Monotonicity:
4
C
USB CONTROL CONNECTOR
J3 PIN LOCATIONS:
PIN No. (J3)
FUNCTION
V BUS (+5V)
1
2
D3
D+
ID (NC)
4
5
GROUND
TTL CONTROL CONNECTOR
J4 PIN LOCATIONS:
PIN No. (J4) 4205-63.5 (CELL)
0.5
1
2
1
3
2
4
4
5
8
6
16
7
32
8
NC
+5V
9
10
GND
6.6
[.26]
SPECIFICATIONS CONTINUE:
CONTROL CONNECTORS:
1) AMP latch 10 pin ribbon
cable connector mates
with AMP P/N 746285-1
(supplied with each unit)
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
.XX
.01
1/2
1/64
.XXX
.003
MATERIAL
PROPRIETARY NOTICE
3
B
INTERFACE CONTROL DRAWING
2) USB - Mini-B
THIS DOCUMENT IS COMMERCIAL
CONFIDENTIAL AND THE INFORMATION
CONTAINED THEREIN MAY NOT BE DISSEMNATED, NOR IS REPRODUCTION
PERMITTED EXCEPT BY WRITTEN
AUTHORIZATION.
10.2
[.40]
2.2
[.09]
13.8
[.54]
SPECIFICATIONS:
43.18
[1.700]
59.18
[2.330]
44.3
[1.75]
10.9
[.43]
17.1
[.68]
PIN 2
59.18
[2.330]
63.5
[2.50]
B
A
4X 2-56 UNC-2B
PIN 1
APPROVED
D
4.1[.16]
SEE
VIEW
A
C
29.21
[1.150]
4.1[.16]
9.0
[.36]
VIEW A
SCALE 4 : 1
63.5
[2.50]
43.18
[1.700]
9.5
[.38]
8.4
[.33]
1
REVISION HISTORY
ZONE REV
DESCRIPTION
DATE
A
ERN 13-042
RS
MAR 13
101-2464
NEXT ASSY
4205-63.5
USED ON
APPLICATION
N/A
CONTRACT NO.
APPROVALS
DRAWN
R. SHOCK
DATE
ICD, ATTENUATOR PROGRAMMABLE
MODEL 4205-63.5
3/26/2013
CHECKED
A. HOPKINS
FINISH
N/A
ENGR.
Z. DARMON
DO NOT SCALE DRAWING
2
ISSUED
A
Aeroflex / Weinschel
SIZE FSCM NO.
C
SCALE
93459
2:1
DWG NO
REV
A
089-4460
SHEET
1
1
OF
1
4
3
NOTES:
2
PIN 1
PIN 5
1. ALL DIMENSIONS ARE GIVEN IN mm [in.]
D
2. ALL MATERIALS AND PROCESSES ARE TO BE IN COMPLIANCE WITH
THE EUROPEAN DIRECTIVE RESTRICTION OF HAZARDOUS
SUBSTANCES (RoHS) (REF: A/W 080-638)
9.5
[.38]
8.4
[.33]
2X 4-40 UNC-2B
9.0
[.36]
DETAIL A
SCALE 4 : 1
63.5
[2.50]
43.18
[1.700]
10.2
[.40]
9.5
[.38]
CAGE: 93459
MODEL: 4205-95.5
SERNO: XXXX
0.2-6GHz 0-95.5dB / 0.5dB
47.5
[1.87]
PIN 1
4X 2-56 UNC-2B
2X 4-40 UNC-2B
4.1[.16]
6.6
[.26]
59.18
[2.330]
4.1[.16]
2.2 10.9
[.09] [.43]
B
SPECIFICATIONS:
17.1
[.68]
200-6000 MHz
50 ohms nominal
0 to 95.5 dB in 0.5 dB step size
0.5;1;2;4;8;16;32 & 32 dB
< 1 microsecond
0.2-0.8 GHz: 2:1 max
0.8-5 GHz:
1.6:1 max
5-6 GHz:
1.9:1 max
Insertion Loss:
0.2-3 GHz:
6.5 dB max
3-6 GHz:
8 dB max
Accuracy:
0.5 to 7.5 dB; +/- 0.5 dB
8 to 11.5 dB; +/- 1.0 dB
12 to 85.5 dB ; +/- 1.25 dB or 4%
86 to 95.5 dB ; 5 %
A Supply Voltage :
5 Volts @ 50 mA typical
Control Logic :
TTL or USB
RF Power Rating: +23 dBm
Temperature Range: 0°C - 70°C
Connectors:
SMA Female
Monotonicity:
200-5000 MHz (min of 0.5 dB step)
5000-6000 MHz (min of 1.0 dB step)
4
C
4X 2-56 UNC-2B
29.21
[1.150]
4.1[.16]
2.2
[.09]
43.18
[1.700]
TTL CONTROL CONNECTOR
J4 PIN LOCATIONS:
PIN No. (J4) 4205-95.5 dB (CELL)
0.5
1
2
1
3
2
4
4
5
8
6
16
7
32
8
32
+5V
9
10
GND
13.8
[.54]
Frequency:
Impedance:
Attenuation:
Attenuation Cells :
Switching Time:
SWR:
D
44.3
[1.75]
10.9
[.43]
9.0
[.36]
17.1
[.68]
PIN 2
59.18
[2.330]
63.5
[2.50]
29.21
[1.150]
4.1[.16]
APPROVED
Z. DARMON
R.SINNO
5.6
[.22]
SEE
DETAIL
A
C
ZONE REV
D
E
F
1
REVISION HISTORY
DESCRIPTION
DATE
ECN 12-230
VB.
10/12
ECN 12-310
VB.
12/12
ECN 13-188
RS
MAR 2013
6.6
[.26]
SPECIFICATIONS CONTINUE:
CONTROL CONNECTORS:
B
USB CONTROL CONNECTOR
J3 PIN LOCATIONS:
PIN No. (J3)
1
2
3
4
5
1) AMP latch 10 pin ribbon
cable connector mates
with AMP P/N 746285-1
(supplied with each unit)
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE IN INCHES
TOLERANCES ARE:
FRACTIONS DECIMALS ANGLES
.XX
.01
1/2
1/64
.XXX
.003
MATERIAL
PROPRIETARY NOTICE
3
FUNCTION
V BUS (+5V)
DD+
ID (NC)
GROUND
INTERFACE CONTROL DRAWING
2) USB - Mini-B
THIS DOCUMENT IS COMMERCIAL
CONFIDENTIAL AND THE INFORMATION
CONTAINED THEREIN MAY NOT BE DISSEMNATED, NOR IS REPRODUCTION
PERMITTED EXCEPT BY WRITTEN
AUTHORIZATION.
10.16
[.400]
101-2424
NEXT ASSY
4205-95.5
USED ON
APPLICATION
N/A
CONTRACT NO.
APPROVALS
5/15/2012
CHECKED
7/16/2012
A. HOPKINS
FINISH
N/A
ENGR.
Z. DARMON
DO NOT SCALE DRAWING
2
DATE
DRAWN
V. BETSIS
ISSUED
A
Aeroflex / Weinschel
7/17/2012
ICD, ATTENUATOR PROGRAMMABLE
MODEL 4205-95.5
SIZE FSCM NO.
C
SCALE
93459
2:1
DWG NO
REV
F
089-4365
SHEET
1
1
OF
1
APPENDIX A
APPENDIX A
CARE AND HANDLING OF MICROWAVE COAXIAL
CABLE ASSEMBLIES
A-1 CARE AND HANDLING OF ASSEMBLIES.
A-3 MAKING INITIAL CONNECTIONS.
To ensure accurate measurements and optimal
performance of Weinschel products, the microwave coaxial
cable assemblies used in system and test setups must be
properly used and maintained. Proper connections, routine
inspection of all cables, and cleaning of the connectors are
extremely important procedures which can prolong the
longevity and accuracy of equipment.
Exercise caution when mating cables. Poor connections
lead to poor system performance. They can also damage not
only the cable assembly, but more significantly, front or rear
panel connectors on the equipment itself which may be more
difficult to repair.
A-2 CABLE INSPECTION.
Routinely check external cables for signs of cracked
insulation, dents, twists, flattening, signs of jacket abrasion, or
other signs of abuse. Wrinkles in the jacket indicate that the
minimum bend radius has been exceeded. Most often, this
occurs near the marker tubes and connectors.
Also inspect the connector interfaces for the following:
•
•
•
•
•
•
•
Bent pins (male).
A-3.2 MATING CONNECTORS. Gent ly mate the
connectors by hand, taking care not to force the coupling nut
at the slightest resistance. It is often possible to feel whether
or not the pins are mated. If the coupling nut is difficult to
turn, either the pins are not mated, the coupling nut is
cross-threaded, or one of the connectors has been damaged by
excess torque.
Bent or missing tines (female).
Worn or chipped plating.
Damaged or displaced dielectric inserts.
Thread damage.
Folded or mushroomed outer interface rims.
Mushroomed pin shoulders (male) or tine ends
(female).
• Score lines on pins and outer interface rims
visible to the unaided eye.
• Recessed or protruding pins.
It is advisable to clean the connectors prior to inspection to
make subtle damage more apparent. If any of the above is
noted, replace the assembly before its further use results in
equipment damage. Also inspect the mating connectors for
similar damage.
Inspect the connector interface for signs of debris. Debris
may be in the form of:
•
•
•
•
A-3.1 ALIGNING CONNECTORS. Align the center
lines of two connectors before actual mating. Male retaining
nuts contain a small amount of necessary play which may make
it possible to mate the threads without the pins being properly
aligned. Pin misalignment can damage pins and dielectric
inserts.
Plating chips or other metal particles.
Dust or dirt.
Oily films.
Other miscellaneous foreign particles.
Never hold a male connector coupling nut
stationary while screwing a female connector into it. This
rotation can erode the plating and damage both the outer
interface rim as well as the pin. If the pins become locked,
serious damage can result to both the equipment and the cable
assembly.
A-4 ENSURING PROPER CONNECTOR TORQUE.
A-4.1 OVERTORQUING. Once connectors have been
properly mated, apply only the proper amount of torque.
Overtorquing damages both connectors involved. Also, a
connector which has been damaged by overtorquing, in turn,
damages every connector to which it is subsequently mated.
It usually leads to poor system performance as well.
Overtorque can cause:
•
•
•
•
•
•
•
•
Bent pins.
Recessed or protruding pins.
Recessed or protruding dielectrics.
Chipped plating.
Damaged coupling threads.
Coupling nut retaining ring damage.
Mushroomed outer interface shells.
Mushroomed pin shoulders.
If signs of debris are present, clean the connector interface
as directed in Paragraph A-6.
A-i
CARE & HANDLING OF COAXIAL CABLES
A-4.2 HEX-NUT TYPES. To mate a connector of the
hex-nut type, always use a torque wrench set to the correct
torque value. Tighten the connector slowly until the wrench
snaps. Tightening too quickly can cause the wrench to exceed
its set limit. Do not snap the wrench more than once as this
also causes overtorque.
A-4.3 KNURLED NUTS. T ighten connectors with
knurled nuts by hand. If this does not provide sufficient
tightness use a hex-nut connector and torque wrench instead.
Never use pliers to tighten a connector. Table A-1
recommends torque specifications for the various types of
connectors.
Table A-1. Recommended Torque Values
Connector
Recommended
Torque
GPC-7 (7mm) w/hex nut
14 in/lbs ± 1 in/lbs
Type N w/hex nut
14 in/lbs ± 1 in/lbs
SMA, 2.92mm, 3.5mm
2.4mm, WPM, WPM-3
WPM-4
7.5 in/lbs ± 0.5 in/lbs
Type N & TNC (knurled)
Hand-tight
BNC (knurled)
Hand-tight
A-5 PROPER CABLE HANDLING.
Never exceed the minimum bend radius specified for a
cable. Guard against tight bends at the end of connector strain
relief tubing, or at the ends of marker tubing where they may
be less noticeable. Although cable bend may seem slight, the
actual radius of the bend at the point of angular departure may
be far smaller than the acceptable radius.
Never pinch, crush or drop objects on cable assemblies.
Also, do not drag a cable over sharp edges as this will pinch it
and cause it to exceed the minimum bend radius.
Never use a cable assembly to pull a piece of equipment.
Cables and connectors are not designed to support or move
equipment.
A-ii
A-5.1 SECURING CABLES. Use toothed, rubber-lined
"P-clamps" to hold cables in place. If it is necessary to use
tie-wraps, use the widest possible wrap and the lowest setting
on the gun to ensure the minimum pressure on the cable.
A-5.2 STORING CABLES.W he n st oring cables,
minimize cable "set" by coiling them in large diameters (1 or
2 feet). Unroll the cable properly when it is ready to be used;
do not pull the loops out hastily. Similarly, re-roll them when
storing them away again.
A-6 CLEANING CONNECTOR INTERFACES.
Use the following guidelines in cleaning connector
interfaces:
a. Do not use chlorinated solvents including common tap
water. These solvents are extremely penetrating and
sometimes ruin otherwise good devices and assemblies.
b. Moisten a cotton swab with isopropyl alcohol. Roll
the swab on a paper towel to remove excess.
c. Use the moistened cotton swab to wipe away debris.
Do not try to dissolve the debris by overwetting the swab.
d. Repeat the cleaning process using additional swabs as
necessary. If metallic particles are embedded in the
dielectric, use an eyeglass and a sharp pick in an attempt
to dislodge them. Swab again.
e. When satisfied that the interfaces are clean, blow them
dry with dry compressed air, or preferably dry nitrogen
(pressurized spray cans work well). Do not use breath.
f. Clean the mating connectors. These may be the source
of the debris.
Model 4205
IM-612
Model 4205 Series (IM-612) Revision Record
DATE
A
3-28-2013
ERN 13-043: Initial Issue includes API Weinschel
ECNs 12-230, 12-310, 13-022 & 13-188.
All Units
B
11-24-2015
ECN 15-319: Company Name and Logo change to
API Weinschel.
All unit built after
11-24-15
API Weinschel
DESCRIPTION
APPLICABLE
SERIAL NUMBERS
REVISION
23