DACU 820

Charge Amplifier DACU 820
Features
• Multi-range charge amplifier for industrial
application
• 4 selectable ranges channel 1
3 fixed ranges 100'000 pC - 500'000 pC
1 variable range 100'000 pC - 500'000 pC
• 4 fixed ranges channel 2
4 fixed ranges 2'000 pC - 20'000 pC
• Adjustable limit value with switching output
• Peak value and test function
• Serial RS 232 interface
Electrical Data
Order Code
Voltage supply10...40 VDC
DACU 820-2.0-500BS
Capacity draw< 1,5 W (1)
Measuring range channel 1±100'000...500'000 pC
Measuring range channel 2±2'000...20'000 pC
Output signal±10 V
Included
Characteristic curve
deviation
• Mounting screws 2 pcs. M4 x 16
< 1% FS
Linearity< 0,02%FS
Output offset< ±5 mV
Noise voltage< 5 mVpp (0,1 Hz...100 kHz) (2)
Dimensions (mm)
Output impedance10 Ω
Reset operate offset< ± 10 mV
Drift< 0,03 pC/s at 23 °C (3)
Frequency range (-3 dB)0...20 kHz (4)
Control input± 5 V...± 45 V, galv. separated
Switching outputmax. 45 V, max. 100 mA
galv. separated
(1)
< 55 mA at 24 V
(2)
< 20 mVpp in the 2000 pC range
(3)
DACU at least 30 min attached to operational voltage
(4)
@ 2'000 pC...100'000 pC; > 2 kHz @ 500'000 pC
All specifications at ambient temperature (23°C ±2°C)
Mechanical Data
Control connection25 pin D-Sub
Sensor connectionBNC male
Enclosure materialAluminum die cast
Environmental Conditions
Operating temp. range-5...+60 °C
Storage temperature-20...+80 °C
Protection classIP 40
EMC
EN 61000-6-2 immunity
EN 61000-6-4 emission
9.6
www.baumer.com
DACU 820
Electrical Connections D-Sub 25
Measuring Range Selection
Pin
Function
Range CH1
1
Signal out CH1
A1
2
Peak out
0 0
500'000
3
Level 1 (In or Out)
0 1
200'000
4
Level 0 (In or Out)
1 0
100'000
5
Range B1 (CH2)
1 1
100'000...500'000
6
Range B0 (CH2)
Range CH2
7
Range A1 (CH1)
B1
8
Range A0 (CH1)
0 0
20'000
9
Supply GND
Measuring Range
A0
pC/10V
B0
0 1
10'000
10+Supply
1 0
5'000
11
Code 2
1 1
2'000
12
Code 0
13
Code Supply +
14
Signal GND
15
Alarm 1 (CH1)
16
Alarm 0 (CH1)
Accessories
Connecting cable with open end
17RX
18TX
19
Com Logic Input
20Operate
80% Test
22
Supply GND
23
Signal out CH2
24
Code 1
25
Com Logic Output (Alarm)
Control Elements
50
L
21
Length (L) Order code
5 m
DZCS 05/DACU 8
9
Range
Adjustment of variable range
L0
Adjustment of value 1. The limit switch voltage
can be set with the potentiometer and has to
be measured at pin 4. Alternatively a voltage can be supplied as well.
L1
Adjustment limit value 2. ditto, pin 3 only