Datasheet

For more information, please visit the product page.
date
09/11/2012
page
1 of 4
SERIES: VSKM-S10 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
U
up to 10 W continuous power
compact board mount design
universal input (85~264 Vac / 120~370 Vdc)
single output from 3.3~24 V
over current and short circuit protections
full medical approvals
efficiency up to 80%
RoHS
VSKM-S10-3R3U
output
voltage
output
current
(Vdc)
max
(A)
3.3
VSKM-S10-5U
5
VSKM-S10-9U
9
12
VSKM-S10-15U
15
VSKM-S10-24U
efficiency
max
(W)
typ
(mVp-p)
typ
(%)
2
6.6
50
70
2
10
50
74
1.1
10
50
76
0.9
10
50
76
0.7
10
50
78
24
0.45
10
50
80
1. Ripple and noise measured at 20 MHz bandwidth
O
Notes:
ripple
and noise1
N
VSKM-S10-12U
output
power
TI
MODEL
N
•
•
•
•
•
•
•
ED
FEATURES
VSKM-S10 - XXU
Base Number
Output Voltage
D
IS
C
PART NUMBER KEY
cui.com
For more information, please visit the product page.
CUI Inc │ SERIES: VSKM-S10 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/11/2012 │ page 2 of 4
INPUT
conditions/description
min
voltage
frequency
current
at 110 Vac
at 230 Vac
inrush current
at 110 Vac
at 230 Vac
input fuse
slow blow, 250 V
leakage current
230 V ac, 50 Hz
parameter
conditions/description
at 230 Vac
voltage set point accuracy
3.3 V output
all other models
switching frequency
PROTECTIONS
parameter
conditions/description
short circuit protection
safety approvals
safety class
primary to secondary (for 1 minute)
230
150
mA
mA
10
20
A
A
2
A
0.1
mA
typ
max
units
±0.5
%
±1
%
0.02
%/°C
50
ms
±3
±2
%
%
60
kHz
typ
max
units
110
%
min
typ
max
units
4,000
Vac
IEC 60601-1, EN 60601-1, UL 60601-1
class II
EN 55011 (level B), IEC/EN 61000-4-2 level 4 (8kV/15kV), IEC/EN 61000-4-3, IEC/EN 61000-4-4 level
4 (4kV), IEC/EN 61000-4-5 level 4 (2kV/4kV)
IS
C
EMI/EMC1
conditions/description
O
isolation voltage
Hz
auto recovery with no damage from a short on any output
SAFETY & COMPLIANCE
parameter
63
min
N
over current protection
47
TI
temperature coefficient
hold-up time
Vac
Vdc
N
10 ~ 100%
units
264
370
min
line regulation
load regulation
typ
U
OUTPUT
max
85
120
ED
parameter
RoHS compliant
yes
MTBF
25°C
300,000
hrs
Notes: 1. external EMC application circuit is required
ENVIRONMENTAL
parameter
conditions/description
min
typ
max
units
-25
70
°C
storage temperature
-40
105
°C
95
°C
95
%
D
operating temperature
case temperature
operating humidity
non-condensing
cui.com
For more information, please visit the product page.
CUI Inc │ SERIES: VSKM-S10 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/11/2012 │ page 3 of 4
DERATING CURVES
2. output power vs. input voltage
MECHANICAL
conditions/description
min
dimensions
2.1 x 1.12 x 0.75 (53.5 x 28.5 x 19.0 mm)
case material
UL94V-0
typ
N
parameter
U
ED
1. output power vs. ambient temperature
47
MECHANICAL DRAWING
inch
50
53
g
O
N
units: mm [inches]
tolerance: ±0.5 [±0.02]
pin section tolerance: ±0.10 mm [±0.004]
units
TI
weight
max
1.20[0.047]
IS
C
53.50 2.106
1
4
19.00 0.748
2
1.00 0.039
(TYP)
6.00 0.236 (TYP)
recommended PCB footprint (top view)
grid: 2.54 mm (0.10 inch)
53.50 2.106
D
10.16 0.400
PIN CONNECTIONS
2
4
10.16 0.400
1
3
3
10.16 0.400
28.5 1.122
14.25 0.561
45.72 1.800
cui.com
PIN
FUNCTION
1
AC(N)
2
AC(L)
3
+Vo
4
-Vo
For more information, please visit the product page.
CUI Inc │ SERIES: VSKM-S10 │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 09/11/2012 │ page 4 of 4
TYPICAL APPLICATION CIRCUIT
L
+Vo
C1
C2 TVS
C7
+
NF
C6
*MOV
C8
-Vo
N
N
ED
*Fuse *NTC
L
RL
U
EXTERNAL CAPACITORS TYPICAL VALUE (Unit: μF)
MODEL
C1
C2
TVS
VSKM-S10-3R3U
220 μF/10 V
0.1 μF/50 V
P6KE6.8A
220 μF/10 V
0.1 μF/50 V
P6KE6.8A
120 μF/25 V
0.1 μF/50 V
P6KE12A
VSKM-S10-12U
120 μF/25 V
0.1 μF/50 V
P6KE20A
VSKM-S10-15U
120 μF/25 V
0.1 μF/50 V
P6KE20A
VSKM-S10-24U
68 μF/35 V
0.1 μF/50 V
P6KE30A
TI
Notes:
N
VSKM-S10-5U
VSKM-S10-9U
1. Output filtering capacitor C1 is an electrolytic capacitor. It is recommended to use high frequency
and low impedance electrolytic capacitors. For capacitance and current of capacitor please refer
to manufacture’s datasheet. Voltage derating of capacitor should be 80% or above. C2 is ceramic
capacitor, it is used to filter high frequency noise. TVS is a recommended component to protect
post-circuits (if converter fails).
2. MOV is required to protect the device under surge.
N
2. It is recommended to use a 2A/250V slow blow FUSE. External input NTC is recommended to
use 5D-9.
O
3. If EMC performance is required, it is recommended to add “EMC filter” at the input end (see EMC
Application Figure).
C6: X capacitor, recommended parameter 0.1μF/275V;
C7,C8: Y capacitor, recommended parameter 2200pF/400V;
NF: common model choke, recommended inductance is about 10mH-30mH.
IS
C
REVISION HISTORY
rev.
description
date
1.0
initial release
01/19/2012
1.01
V-Infinity branding removed
09/11/2012
D
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
[email protected]
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to
affect its safety or effectiveness.