Si91-Maxum-Methanator-Purge_Rev1.pdf

SIEMENS
Automation & Drives
Technical Support Information No. 91
September 2003
Process Instrumentation and Analytics - A&D PI 2 - Karlsruhe
Page 1 of 1
Information zu / to
Maxum II Methanator Spülung und Oxydation
Zusammenfassung
Der Maxum II Methanator wird mittels der Gönnheimer APU gespült um die Temperatur-Betriebssicherheit zu
erfüllen.
Wie auf dem Aufkleber angebracht am Maxum II aufgeführt, darf der GC erst geöffnet werden, nachdem
sichergestellt ist, dass die Umgebung frei von feuergefährlichem Material bzw. erst nachdem 30 Minuten zuvor alle
elektronischen Bauelemente abgeschaltet wurden. Die hohe Spülrate sollte mindestens 5 Minuten und bei
mindestens 10 psig Ofenluft verwendet werden.
Maxum II Methanator Purge and Oxidation
Purging the Maxum II Methanator
The Maxum II Methanator is purged, with the EC purge system, to meet the safety temperature rating of the
analyzer. As stated on the tag on the Maxum II with a Methanator, the analyzer shall not be opened unless the area
is known to be free of flammable material or all devices in the analyzer have been turned off for at least 30 minutes.
This allows the heater in the methanator to cool down to a safe level. To ensure the Methanator is purged
of possible flammable material, fast purge needs to run for at least 5 minutes.
The following text is on a tag which is attached to all Maxum II analyzers with a Methanator:
When operating this unit in a Div 2 area when equipped with Methanator Model 2021601: This enclosure shall not
be opened unless the area is known to be free of flammable material or unless all devices within have been deenergized for at least 30 minutes. Power shall not be restored after enclosure has been opened until the "FAST
PURGE" has been on for 5 minutes with an "OVEN AIR" of at least 10 psig.
Background: The Optichrom Methanator was not purged because its housing is an explosion proof enclosure.
Purging the Maxum II Methanator keeps explosive gasses from reaching the hot surfaces inside the Methanator.
Attached is a pdf which includes; the assembly drawing for the Maxum II Methanator, a section from the manual
and the installation drawing for the Methanator installation kit.
Refer to TSI No. 82 for additional informations about Methanator catalyst in oxidized state!
Spare Parts
2021601-001 - Maxum II Methanator, Temperature Controlled 115/230 VAC (complete assembly)
1621004-012 - Heater, Cartridge
1622001-020 - RTD, Push-In Style
©
SIEMENS AG Karlsruhe A&D PI 2 AS 2 Niko Benas Tel. +49 721 595 7216, Fax 595 6564
ITEM
PART
NO.
NUMBER
DESCRIPTION
REFERENCE/ZONE
ASSEMBLY/QUANTITY
DESIGNATOR
2021601-001
ASSEMBLY, METHANATOR, 120/240
1
2021601-AD
ASSEMBLY DRAWING
2
2021606-001
3
2021604-001
4
UNIT
-001
MEAS
X
REF
EA
COVER, METHANATOR ASSEMBLY
1
EA
1
EA
1531000-026
HEATER, BLOCK, METHANATOR
O-RING,SILICONE RUBBER,76.0MM OD
2.5MM DIAMETER
1
EA
5
2021623-001
ASSEMBLY, METHANATOR BASE
1
EA
6
1312530-007
WASHER, M4, HELICAL SPRING LOCK, SST
4
EA
7
1312420-346
SCREW, M4X6, SOCKET HEAD CAP, SST
5
EA
8
2021603-001
INSULATION, METHANATOR, LOWER
1
EA
9
2021603-002
INSULATION, METHANATOR, CENTER
1
EA
10
2021603-003
INSULATION, METHANATOR, UPPER
1
EA
REV
ACTION
DRAWN
CHECKED
APPROVED
DATE
PRODUCT LINE
ANALYTICAL
Siemens Applied Automation
LEVEL
3
DESIGN
DATE
AA
L9667
SIMONS
SMITH
KESSELH
02-3
PFALZER
02-03
DRAWN
DATE
AB
D9785
SMITH
SMITH
KESSELH
02-4
SIMONS
02-03
AC
D10371
WIDENER
WIDENR
COLLINS
02-10
SMITH
PARTS LIST FOR
ASSEMBLY, METHANATOR, 120/240
MAXUM II
CHECKED
02-3
APPROVED
KESSELHUTH
02-3
SCALE
SIZE
TYPE
NONE
A
PL
DRAWING NO.
2021601
REV
AC
SHEET
1
OF
2
ITEM
PART
NO.
NUMBER
DESCRIPTION
REFERENCE/ZONE
DESIGNATOR
ASSEMBLY/QUANTITY
UNIT
-001
MEAS
11
2021262-002
PLATE, METHANATOR MOUNTING
1
EA
12
2021604-002
COVER, HEATER, BLOCK, METHANATOR
1
EA
13
1312413-036
SCREW,M4X10,FLAT HEAD SOCKET,SST
3
EA
14
2021172-005
COVER, DETECTOR, METHANATOR
1
EA
15
2021636-001
PIPE, MODIFIED, METHANATOR
1
EA
16
1317001-026
GROMMET, ID 13/16", GROOVE DIA. 15/16"
1
EA
17
2021607-001
CABLE, GROUND, METHANATOR, LONG
1
EA
18
2021607-002
CABLE, GROUND, METHANATOR, SHORT
1
EA
19
1684021-002
CABLE TIE,4",TEFZEL,HI TEMP, UV RESIST
2
EA
20
1621004-012
HEATER, 1” CARTRIDGE, 120V, FIREROD
2
EA
21
N80005
NUT,TRU-SEAL,1/2,STL
1
EA
22
F10900
CONDUIT,NUT,DRIVE 1/2
1
EA
23
1622001-020
RTD, PUSH-IN STYLE, 1” LONG, .188”OD
1
EA
24
T08800
TAPE,1/2" GLASS CLOTH
AR
MM
25
2021696-001
TAG, METHANATOR, CENELEC
1
EA
26
1371000-001
GREASE, FLUORINATED, KRYTOX, 240AC
AR
EA
27
2020184-001
TAG, METHANATOR, CATALYST WARNING
1
EA
SCALE
NONE
SIZE
A
TYPE
PL
DRAWING NO.
2021601
REV
AC
SHEET
2
OF
2
Methanator
Description
The methantor is used with an FID when detection of CO and CO2 is
required.
Methanator Operation
In the methanator CO and CO2 are catalytically reduced to methane
under excess hydrogen, which can be detected in the FID:
CO + 3 H2
CO2 + 4 H2
Ni
→
Ni
→
CH4 + H2O
CH4 + 2 H2O
Other areas of application are the reduction of formic acid and
formaldehyde to CH4. This permits quantitative detection of small
concentrations in the ppm range.
HCOOH + 3 H2
HCOH + 2 H2
Ni
→
Ni
→
CH4 + 2 H2O
CH4 + H2O
If possible, the samples should not contain any halogenated or sulfurous
components.
Operational Note
Hydrocarbons above C8 are partially decomposed through the catalyst.
They appear with tailing and too small. The simultaneous detection of
higher boiling components and CO/CO2 is therefore not possible.
Hydration Reactor
(Methanator)
The hydration reactor consists of a stainless steel U pipe that is filled
with a catalyst. A heating plate is plugged on the U pipe. Two heating
cartridges and a Pt 100 temperature sensor regulate heating to 400 °C.
The U pipe is installed in a base plate airtight. The gas inlets and gas
outlets are configured as flame barriers. A protective cover is installed
airtight over the base plate. The cover is insulated to prevent the surface
temperature from exceeding 180 °C.
In the ex-version the U pipe is encapsulated and pressure proof. The
base plate contains a gas connector and exhaust nozzles for purge gas.
The gas connector has a pipe, which extends into the purged shaft
behind the installation chamber. This pipe is used to hold the connection
lines for heating cartridges and temperature sensor. Purge gas flows
from the purged shaft into the methanator housing through this pipe and
prevents explosive gas from reaching the heater connections.
ITEM
PART
NO.
NUMBER
2021564-001
DESCRIPTION
REFERENCE/ZONE
2021564-005
1
2021564 -A D
ASSWMBLY DRAWING
2
2021171-001
3
2021564-002
2021564-003
2021564-004
-001
DESIGNATOR
METHANATOR, INSTALLATION PACKAGE, NONE
METHANATOR, INSTALLATION PACKAGE, AIR
BATH OVEN, RIGHT POSITION
METHANATOR, INSTALLATION PACKAGE, AIR
BATH OVEN, LEFT POSITION
METHANATOR, INSTALLATION PACKAGE, AIR
BATH OVEN, BOTH POSITIONS
METHANATOR, INSTALLATION PACKAGE,
AIRLESS OVEN, RIGHT POSITION
ASSEMBLY/QUANTITY
-002
-003
UNIT
-004
-005
MEAS
X
X
X
X
X
REF
REF
REF
REF
REF
EA
GASKET, DETECTOR
2
2
2
2
2
EA
2021172-001
COVER, DETECTOR W/O HOLES
2
1
1
-
1
EA
4
1312601-004
NUTS, M4 WITH LOCKWASHER
8
8
8
8
8
EA
5
2021215-001
PLATE, OVEN COVER, DETECTOR
2
1
1
-
1
EA
6
1312421-069
SCREW, M4 X 8 SOCKETM BUTTON HEAD CAP
8
8
8
8
8
EA
7
1312530-007
WASHER, M4, HELICAL SPRING LOCK, SST
8
8
8
8
8
EA
8
2021601-001
ASSEMBLY, METHANATOR, 120/240
-
1
1
2
1
EA
9
1212005-004
WIRE NUT, 22-24 GAUGE
-
1
1
2
1
EA
10
T08500
TAPE,3/4” ELECTRICAL
-
AR
AR
AR
AR
AR
REV
ACTION
AA
L9877
DRAWN
SIMONS
CHECKED
SMITH
APPROVED
KESSELH
DATE
02-5
PRODUCT LINE
ANALYTICAL
DESIGN
LEVEL
SIMONS
02-04
DRAWN
DATE
SIMONS
02-04
CHECKED
DATE
SMITH
02-5
APPROVED
DATE
KESSELHUTH
02-5
Siemens Applied Automation
3
DATE
PARTS LIST FOR
METHANATOR, INSTALLATION PACKAGE
MAXUM II
SCALE
NONE
SIZE
A
TYPE
PL
DRAWING NO.
2021564
REV
AA
SHEET
1
OF
4
ITEM
PART
NO.
NUMBER
11
2021700-001
DESCRIPTION
REFERENCE/ZONE
DESIGNATOR
ASSEMBLY/QUANTITY
-001
-002
-003
-004
-005
MEAS
-
1
1
1
1
EA
TAG, METHANATOR WARNING
SCALE
NONE
SIZE
A
UNIT
TYPE
PL
DRAWING NO.
2021564
REV
AA
SHEET
2
OF
4
ITEM
PART
NO.
NUMBER
DESCRIPTION
REFERENCE/ZONE
2021564-007
1
2021564 -A D
ASSWMBLY DRAWING
2
2021171-001
3
2021564-006
-006
DESIGNATOR
METHANATOR, INSTALLATION PACKAGE,
AIRLESS OVEN, LEFT POSITION
METHANATOR, INSTALLATION PACKAGE,
AIRLESS OVEN, BOTH POSITIONS
ASSEMBLY/QUANTITY
UNIT
-007
MEAS
X
X
REF
REF
EA
GASKET, DETECTOR
2
2
EA
2021172-001
COVER, DETECTOR W/O HOLES
1
-
EA
4
1312601-004
NUTS, M4 WITH LOCKWASHER
8
8
EA
5
2021215-001
PLATE, OVEN COVER, DETECTOR
1
0
EA
6
1312421-069
SCREW, M4 X 8 SOCKETM BUTTON HEAD CAP
8
8
EA
7
1312530-007
WASHER, M4, HELICAL SPRING LOCK, SST
8
8
EA
8
2021601-001
ASSEMBLY, METHANATOR, 120/240
1
2
EA
9
1212005-004
WIRE NUT, 22-24 GAUGE
1
2
EA
AR
AR
AR
10
T08500
REV
ACTION
TAPE,3/4” ELECTRICAL
DRAWN
CHECKED
APPROVED
DATE
PRODUCT LINE
Siemens Applied Automation
LEVEL
DESIGN
DATE
DRAWN
DATE
CHECKED
DATE
APPROVED
DATE
PARTS LIST FOR
SCALE
NONE
SIZE
A
TYPE
PL
DRAWING NO.
2021564
REV
AA
SHEET
3
OF
4
ITEM
PART
NO.
NUMBER
11
2021700-001
DESCRIPTION
REFERENCE/ZONE
DESIGNATOR
ASSEMBLY/QUANTITY
-006
-007
MEAS
1
1
EA
TAG, METHANATOR WARNING
SCALE
NONE
SIZE
A
UNIT
TYPE
PL
DRAWING NO.
2021564
REV
AA
SHEET
4
OF
4