RLAT 200K Report_RH119W_F0258.1 W18

RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Radiation Lot Acceptance Test (RLAT) of the RH119W High Performance
Dual Comparator for Linear Technology
Customer: Linear Technology, PO 50963L
RAD Job Number: 08-279
Part Type Tested: Linear Technology RH119W High Performance Dual Comparator
Commercial Part Number: RH119W
Traceability Information: FAB #F0258.1, Wafer 18, Lot #472013.1
Quantity of Units: 11 units total, 5 units for biased irradiation, 5 units for unbiased irradiation and 1 control unit.
External Traveler: None required
Pre-Irradiation Burn-In: Burn-In performed by Linear Technology prior to receipt by RAD, Inc.
TID Dose Rate and Maximum Total Dose: 50 to 300rad(Si)/s to 200krad(Si) total ionizing dose
TID Test Increments: Pre-Irradiation, 20krad(Si), 50krad(Si) 100krad(Si) and 200krad(Si)
TID Overtest and Post-Irradiation Anneal: No overtest. No anneal.
TID Test Standard: MIL-STD-883G, Method 1019.7, Condition A
TID Electrical Test Conditions: Pre-irradiation, and within one hour following each radiation exposure.
Test Hardware and Software: LTS2020 tester, TS-04 Calibration Date: 11-03-08, Calibration Due: 05-03-09.
Test program: RH119JH.SRC
TID Bias Conditions: Serial numbers 694, 695, 698, 699 and 700, were biased during irradiation, serial numbers
701, 702, 704, 705 and 706 were unbiased during irradiation and serial number 1280 was used as the control.
Facility and Radiation Source: Radiation Assured Devices Longmire Laboratories, Colorado Springs, CO using
their JLSA 81-24 Co60 cell. Dosimetry performed using CaF TLDs traceable to NIST. RAD’s dosimetry has been
audited by DSCC and RAD has been awarded Laboratory Suitability for MIL-STD-750 TM 1019.5
Irradiation and Test Temperature: Ambient, room temperature
RLAT Result: Units passed with only minor degradation to the input offset and
input bias parameters and no significant degradation to any other measured
parameter.
An ISO 9001:2000 Certified Company
1
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
1.0. Overview and Background
It is well known that total dose ionizing radiation can cause parametric degradation and ultimately
functional failure in electronic devices. The damage occurs via electron-hole pair production, transport
and trapping in the dielectric and interface regions. In discrete devices the bulk of the damage is
frequently manifested as a reduction in the gain and/or breakdown voltage of the device. The damage
will usually anneal with time following the end of the radiation exposure. Due to this annealing, and to
ensure a worst-case test condition MIL-STD-883 TM1019.7 calls out a dose rate of 50 to 300rad(Si)/s as
Condition A and further specifies that the time from the end of an incremental radiation exposure and
electrical testing shall be 1-hour or less and the total time from the end of one incremental irradiation to
the beginning of the next incremental radiation step should be 2-hours or less. The work described in
this report was performed to meet MIL-STD-883 TM1019.7 Condition A.
2.0. Radiation Test Apparatus
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices’ Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 84-21 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead, during the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source. For high-dose rate experiments the bias
boards are placed in a radial fashion equidistant from the raised Co-60 rods with the distance adjusted to
provide the required dose rate. The irradiator calibration is maintained by Radiation Assured Devices
Longmire Laboratories using thermoluminescent dosimeters (TLDs)) traceable to the National Institute
of Standards and Technology (NIST). Figure 2.1 shows a photograph of the JLSA 81-24 Co-60
irradiator at RAD’s Longmire Laboratory facility.
RAD is currently certified by the Defense Supply Center Columbus (DSCC) for Laboratory Suitability
under MIL STD 750. Additional details regarding Radiation Assured Devices dosimetry for TM1019
Condition A testing are available in RAD’s report to DSCC entitled: “Dose Rate Mapping of the J.L.
Shepherd and Associates Model 81 Irradiator Installed by Radiation Assured Devices”
An ISO 9001:2000 Certified Company
2
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure 2.1. Radiation Assured Devices’ high dose rate Co-60 irradiator. The dose rate is obtained by
positioning the device-under-test at a fixed distance from the gamma cell. The dose rate for this
irradiator varies from approximately 120rad(Si)/s close to the rods down to 1rad(Si)/s at a distance of
approximately 2-feet.
An ISO 9001:2000 Certified Company
3
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
3.0. Radiation Test Conditions
The RH119W dual comparators described in this final report were irradiated using a split 15V supply
and with all pins tied to ground, that is biased and unbiased. See the TID Bias Table in Appendix A for
the full bias circuits and electrical connections during the radiation exposure. These bias circuits satisfy
the requirements of MIL-STD-883G TM1019.7 Section 3.9.3 Bias and Loading Conditions which states
“The bias applied to the test devices shall be selected to produce the greatest radiation induced damage
or the worst-case damage for the intended application, if known. While maximum voltage is often worst
case some bipolar linear device parameters (e.g. input bias current or maximum output load current)
exhibit more degradation with 0 V bias.”
The devices were irradiated to a maximum total ionizing dose level of 200krad(Si) with incremental
readings at 20, 50, 100 and 200krad(Si). Electrical testing occurred within one hour following the end
of each irradiation segment. For intermediate irradiations, the parts were tested and returned to total
dose exposure within two hours from the end of the previous radiation increment.
The TID bias board was positioned in the Co-60 cell to provide the required minimum of 50rad(Si)/s
and was located inside a lead-aluminum enclosure. The lead-aluminum enclosure is required under
MIL-STD-883G TM1019.7 Section 3.4 that reads as follows: “Lead/Aluminum (Pb/Al) container. Test
specimens shall be enclosed in a Pb/Al container to minimize dose enhancement effects caused by lowenergy, scattered radiation. A minimum of 1.5 mm Pb, surrounding an inner shield of at least 0.7 mm
Al, is required. This Pb/Al container produces an approximate charged particle equilibrium for Si and
for TLDs such as CaF2. The radiation field intensity shall be measured inside the Pb/Al container (1)
initially, (2) when the source is changed, or (3) when the orientation or configuration of the source,
container, or test-fixture is changed. This measurement shall be performed by placing a dosimeter (e.g.,
a TLD) in the device-irradiation container at the approximate test-device position. If it can be
demonstrated that low energy scattered radiation is small enough that it will not cause dosimetry errors
due to dose enhancement, the Pb/Al container may be omitted”.
The final dose rate within the lead-aluminum enclosure was determined based on TLD dosimetry
measurements (see previous section). The final dose rate for this work was 64rad(Si)/s with a precision
of ±5%.
An ISO 9001:2000 Certified Company
4
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
4.0. Tested Parameters
During the radiation lot acceptance testing the pre- and post-irradiation electrical parameters measured
were:
1. Input Offset Voltage, VOS1 and VOS2
2. Common Mode Rejection Ratio, CMRR1 and CMRR2
3. Input Offset Current, IOS1 and IOS2
4. Input Bias Current, Non-Inverting Input, +IB1 and +IB2
5. Input Bias Current, Inverting Input, -IB1and -IB2
6. Large Signal Voltage Gain, AVOL1, AVOL2
7. Saturation Voltage, VSAT1 and VSAT2
8. Saturation Voltage, VSAT3 and VSAT4
9. Output Leakage Current, ICEX1 and ICEX2
10. Positive Supply Current, +IS
11. Negative Supply Current, -IS
The parametric data was obtained as “read and record” and all the raw data plus an attributes summary
are contained in this report as well as in a separate Excel file. The attributes data contains the average,
standard deviation and the average with the KTL values applied. The KTL values used is 2.742 per
MIL HDBK 814 using one sided tolerance limits of 90/90 and a 5-piece sample size. This survival
probability/level of confidence is consistent with a 22-piece sample size and zero failures analyzed using
a lot tolerance percent defective (LTPD) approach. Note that the following criteria must be met for a
device to pass the TID testing: following the radiation exposure the units shall pass the specification
value and the average value for the each device must pass the specification value when the KTL limits
are applied. If either of these conditions is not satisfied following the radiation exposure, then the lot
could be logged as an RLAT failure.
An ISO 9001:2000 Certified Company
5
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
5.0. Total Ionizing Dose Test Results
The units passed the radiation lot acceptance test with only minor degradation to the input offset and
input bias parameters and no significant degradation to any other measured parameter.
Figures 5.1 and 5.22 show plots of all the measured parameters versus total ionizing dose while Tables
5.1 – 5.22 show the corresponding raw data for each of these parameters. In the data plots the solid
diamonds are the average of the measured data points for the samples irradiated under electrical bias
while the shaded diamonds are the average of the measured data points for the units irradiated with all
pins tied to ground. The black lines (solid or dashed) are the average of the data points after application
of the KTL statistics on the sample irradiated in the biased condition while the shaded lines (solid or
dashed) are the average of the data points after application of the KTL statistics on the sample irradiated
in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or
maximum specification value as defined in the datasheet and/or test plan.
As seen clearly in these figures and tables, the pre-irradiation data is well within the specification and
the control sample remains stable during the course of the measurements. Therefore, we can conclude
that any of the observed degradation was due to the radiation exposure.
An ISO 9001:2000 Certified Company
6
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Positive Supply Current (A); +/-15V
1.40E-02
1.20E-02
1.00E-02
8.00E-03
6.00E-03
4.00E-03
2.00E-03
0.00E+00
0
50
100
150
Total Dose (krad(Si))
Figure 5.1. Plot of positive supply current versus total dose. The data show no significant degradation with total
dose. The solid diamonds are the average of the measured data points for the sample irradiated under electrical
bias while the shaded diamonds are the average of the measured data points for the units irradiated with all pins
tied to ground. The black lines (solid or dashed) are the average of the data points after application of the KTL
statistics on the sample irradiated in the biased condition while the shaded lines (solid or dashed) are the average
of the data points after application of the KTL statistics on the sample irradiated in the unbiased condition. The
red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in
the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
7
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.1. Raw data of the positive supply current versus total dose, including the statistical analysis, the
specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Positive Supply Current (A); +/-15V
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
7.28E-03
7.28E-03
7.58E-03
7.45E-03
7.31E-03
7.36E-03
7.37E-03
7.20E-03
7.21E-03
7.28E-03
7.25E-03
20
7.28E-03
7.27E-03
7.58E-03
7.44E-03
7.30E-03
7.32E-03
7.34E-03
7.19E-03
7.19E-03
7.27E-03
7.24E-03
50
7.28E-03
7.25E-03
7.57E-03
7.42E-03
7.28E-03
7.36E-03
7.36E-03
7.19E-03
7.21E-03
7.28E-03
7.24E-03
100
7.25E-03
7.24E-03
7.54E-03
7.41E-03
7.25E-03
7.36E-03
7.36E-03
7.19E-03
7.20E-03
7.25E-03
7.24E-03
200
7.26E-03
7.23E-03
7.53E-03
7.39E-03
7.25E-03
7.36E-03
7.37E-03
7.18E-03
7.21E-03
7.28E-03
7.26E-03
7.38E-03
1.32E-04
7.74E-03
7.02E-03
7.37E-03
1.34E-04
7.74E-03
7.01E-03
7.36E-03
1.35E-04
7.73E-03
6.99E-03
7.34E-03
1.33E-04
7.70E-03
6.97E-03
7.33E-03
1.27E-04
7.68E-03
6.98E-03
7.28E-03 7.26E-03 7.28E-03 7.27E-03 7.28E-03
8.02E-05 7.05E-05 8.03E-05 8.35E-05 8.57E-05
7.50E-03 7.46E-03 7.50E-03 7.50E-03 7.52E-03
7.06E-03 7.07E-03 7.06E-03 7.04E-03 7.04E-03
1.15E-02 1.15E-02 1.15E-02 1.15E-02 1.15E-02
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
8
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Negative Supply Current (A); +/- 15V
0.00E+00
-5.00E-04
-1.00E-03
-1.50E-03
-2.00E-03
-2.50E-03
-3.00E-03
-3.50E-03
-4.00E-03
-4.50E-03
-5.00E-03
0
50
100
150
Total Dose (krad(Si))
Figure 5.2. Plot of negative supply current versus total dose. The data show no significant degradation with total
dose. The solid diamonds are the average of the measured data points for the sample irradiated under electrical
bias while the shaded diamonds are the average of the measured data points for the units irradiated with all pins
tied to ground. The black lines (solid or dashed) are the average of the data points after application of the KTL
statistics on the sample irradiated in the biased condition while the shaded lines (solid or dashed) are the average
of the data points after application of the KTL statistics on the sample irradiated in the unbiased condition. The
red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in
the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
9
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.2. Raw data for the negative supply current versus total dose, including the statistical analysis, the
specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Negative Supply Current (A); +/- 15V
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
-3.17E-03
-3.15E-03
-3.28E-03
-3.23E-03
-3.17E-03
-3.17E-03
-3.19E-03
-3.11E-03
-3.12E-03
-3.16E-03
-3.15E-03
20
-3.16E-03
-3.14E-03
-3.27E-03
-3.22E-03
-3.16E-03
-3.17E-03
-3.19E-03
-3.10E-03
-3.11E-03
-3.16E-03
-3.15E-03
50
-3.16E-03
-3.14E-03
-3.27E-03
-3.22E-03
-3.15E-03
-3.17E-03
-3.19E-03
-3.11E-03
-3.12E-03
-3.15E-03
-3.16E-03
100
-3.15E-03
-3.14E-03
-3.25E-03
-3.21E-03
-3.15E-03
-3.17E-03
-3.19E-03
-3.11E-03
-3.12E-03
-3.16E-03
-3.15E-03
200
-3.15E-03
-3.14E-03
-3.25E-03
-3.21E-03
-3.15E-03
-3.17E-03
-3.19E-03
-3.10E-03
-3.12E-03
-3.16E-03
-3.16E-03
-3.20E-03 -3.19E-03 -3.19E-03 -3.18E-03 -3.18E-03
5.39E-05 5.39E-05 5.54E-05 4.80E-05 4.80E-05
-3.05E-03 -3.04E-03 -3.04E-03 -3.05E-03 -3.05E-03
-3.35E-03 -3.34E-03 -3.34E-03 -3.31E-03 -3.31E-03
-3.15E-03
3.39E-05
-3.06E-03
-3.24E-03
-4.50E-03
PASS
-3.15E-03
3.91E-05
-3.04E-03
-3.25E-03
-4.50E-03
PASS
-3.15E-03
3.35E-05
-3.06E-03
-3.24E-03
-4.50E-03
PASS
An ISO 9001:2000 Certified Company
10
-3.15E-03
3.39E-05
-3.06E-03
-3.24E-03
-4.50E-03
PASS
-3.15E-03
3.70E-05
-3.05E-03
-3.25E-03
-4.50E-03
PASS
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.00E-02
Input Offset Voltage 1 (V)
8.00E-03
6.00E-03
4.00E-03
2.00E-03
0.00E+00
-2.00E-03
-4.00E-03
-6.00E-03
-8.00E-03
-1.00E-02
0
50
100
150
Total Dose (krad(Si))
Figure 5.3. Plot of input offset voltage (comparator 1) versus total dose. The data show no significant
degradation with total dose. The solid diamonds are the average of the measured data points for the sample
irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the
units irradiated with all pins tied to ground. The black lines (solid or dashed) are the average of the data points
after application of the KTL statistics on the sample irradiated in the biased condition while the shaded lines
(solid or dashed) are the average of the data points after application of the KTL statistics on the sample irradiated
in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum
specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
11
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.3. Raw data of input offset voltage (comparator 1) versus total dose, including the statistical
analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Input Offset Voltage 1 (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
4.10E-04
3.00E-04
5.20E-04
5.10E-04
5.00E-04
3.40E-04
5.00E-04
3.90E-04
5.50E-04
7.80E-04
6.90E-04
20
4.50E-04
3.20E-04
5.40E-04
5.40E-04
5.20E-04
3.30E-04
4.60E-04
3.60E-04
5.50E-04
7.80E-04
7.10E-04
50
4.60E-04
3.20E-04
5.30E-04
5.50E-04
4.50E-04
3.30E-04
4.70E-04
3.40E-04
5.40E-04
7.70E-04
6.90E-04
100
4.40E-04
3.30E-04
5.20E-04
4.90E-04
5.00E-04
2.80E-04
4.40E-04
3.10E-04
5.20E-04
7.60E-04
7.00E-04
200
4.50E-04
3.10E-04
5.40E-04
5.30E-04
5.20E-04
2.50E-04
4.40E-04
3.00E-04
5.00E-04
7.20E-04
7.00E-04
4.48E-04
9.36E-05
7.05E-04
1.91E-04
4.74E-04
9.37E-05
7.31E-04
2.17E-04
4.62E-04
9.04E-05
7.10E-04
2.14E-04
4.56E-04
7.64E-05
6.65E-04
2.47E-04
4.70E-04
9.62E-05
7.34E-04
2.06E-04
5.12E-04
1.72E-04
9.83E-04
4.13E-05
-4.00E-03
PASS
4.00E-03
PASS
4.96E-04
1.81E-04
9.92E-04
-6.71E-08
-4.00E-03
PASS
4.00E-03
PASS
4.90E-04
1.80E-04
9.83E-04
-3.18E-06
-4.00E-03
PASS
4.00E-03
PASS
4.62E-04
1.93E-04
9.91E-04
-6.70E-05
-4.00E-03
PASS
4.00E-03
PASS
4.42E-04
1.86E-04
9.51E-04
-6.67E-05
-8.00E-03
PASS
8.00E-03
PASS
An ISO 9001:2000 Certified Company
12
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
1.00E-02
Iniput Offset Voltage 2 (V)
8.00E-03
6.00E-03
4.00E-03
2.00E-03
0.00E+00
-2.00E-03
-4.00E-03
-6.00E-03
-8.00E-03
-1.00E-02
0
50
100
150
Total Dose (krad(Si))
Figure 5.4. Plot of input offset voltage (comparator 2) versus total dose. The data show no significant
degradation with total dose. The solid diamonds are the average of the measured data points for the sample
irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the
units irradiated with all pins tied to ground. The black lines (solid or dashed) are the average of the data points
after application of the KTL statistics on the sample irradiated in the biased condition while the shaded lines
(solid or dashed) are the average of the data points after application of the KTL statistics on the sample irradiated
in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum
specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
13
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.4. Raw data of input offset voltage (comparator 2) versus total dose, including the statistical
analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Iniput Offset Voltage 2 (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
4.40E-04
9.10E-04
4.30E-04
2.90E-04
8.50E-04
6.10E-04
3.80E-04
6.00E-04
3.60E-04
6.80E-04
5.80E-04
20
4.90E-04
9.60E-04
4.30E-04
3.30E-04
8.80E-04
6.00E-04
3.80E-04
6.00E-04
3.60E-04
6.90E-04
5.90E-04
50
5.00E-04
9.70E-04
4.30E-04
3.50E-04
9.00E-04
5.90E-04
3.80E-04
5.90E-04
3.50E-04
6.80E-04
6.00E-04
100
5.10E-04
9.90E-04
4.60E-04
3.30E-04
9.10E-04
5.90E-04
3.60E-04
5.80E-04
3.40E-04
6.80E-04
5.90E-04
200
5.10E-04
9.70E-04
4.70E-04
3.60E-04
9.10E-04
5.70E-04
3.50E-04
5.70E-04
3.30E-04
6.60E-04
5.90E-04
5.84E-04 6.18E-04 6.30E-04 6.40E-04 6.44E-04
2.77E-04 2.83E-04 2.85E-04 2.92E-04 2.77E-04
1.34E-03 1.39E-03 1.41E-03 1.44E-03 1.40E-03
-1.77E-04 -1.58E-04 -1.50E-04 -1.60E-04 -1.14E-04
5.26E-04 5.26E-04 5.18E-04 5.10E-04 4.96E-04
1.46E-04 1.47E-04 1.45E-04 1.51E-04 1.47E-04
9.26E-04 9.30E-04 9.15E-04 9.25E-04 9.00E-04
1.26E-04 1.22E-04 1.21E-04 9.51E-05 9.23E-05
-4.00E-03 -4.00E-03 -4.00E-03 -4.00E-03 -8.00E-03
PASS
PASS
PASS
PASS
PASS
4.00E-03 4.00E-03 4.00E-03 4.00E-03 8.00E-03
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
14
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
6.00E-07
Input Offset Current 1 (A)
4.00E-07
2.00E-07
0.00E+00
-2.00E-07
-4.00E-07
-6.00E-07
0
50
100
150
Total Dose (krad(Si))
Figure 5.5. Plot of input offset current (comparator 1) versus total dose. The data show a modest
degradation with total dose, however it stays within specification to the maximum total dose tested. The
solid diamonds are the average of the measured data points for the sample irradiated under electrical bias
while the shaded diamonds are the average of the measured data points for the units irradiated with all pins
tied to ground. The black lines (solid or dashed) are the average of the data points after application of the
KTL statistics on the sample irradiated in the biased condition while the shaded lines (solid or dashed) are
the average of the data points after application of the KTL statistics on the sample irradiated in the
unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum
specification value as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
15
200
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table 5.5. Raw data of input offset current (comparator 1) versus total dose, including the statistical
analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Input Offset Current 1 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
20
50
100
200
1.92E-08 1.46E-08 2.43E-08 1.97E-08 -1.39E-08
1.76E-08 8.75E-09 3.91E-09 -2.00E-11 3.79E-09
2.20E-08 2.39E-08 4.06E-08 6.08E-09 -1.74E-08
2.66E-08 2.33E-08 2.18E-08 4.75E-08 -1.42E-08
2.16E-08 1.09E-08 9.16E-08 2.39E-08 -5.85E-09
2.78E-08 2.51E-08 3.81E-08 4.72E-08 6.35E-08
2.85E-08 3.21E-08 4.97E-08 6.48E-08 8.55E-08
2.39E-08 2.57E-08 3.79E-08 4.82E-08 6.80E-08
1.31E-08 2.12E-08 2.84E-08 3.70E-08 4.78E-08
1.85E-08 2.37E-08 3.28E-08 5.00E-08 7.48E-08
-6.61E-09 -3.80E-09 -7.14E-09 -7.82E-09 -7.22E-09
2.14E-08 1.63E-08 3.64E-08 1.94E-08 -9.51E-09
3.43E-09 6.99E-09 3.35E-08 1.85E-08 8.57E-09
3.08E-08 3.55E-08 1.28E-07 7.01E-08 1.40E-08
1.20E-08 -2.89E-09 -5.53E-08 -3.12E-08 -3.30E-08
2.24E-08 2.56E-08 3.74E-08 4.94E-08 6.79E-08
6.54E-09 4.04E-09 7.97E-09 9.96E-09 1.40E-08
4.03E-08 3.66E-08 5.92E-08 7.67E-08 1.06E-07
4.42E-09 1.45E-08 1.55E-08 2.21E-08 2.96E-08
-7.50E-08 -1.00E-07 -1.50E-07 -3.00E-07 -5.00E-07
PASS
PASS
PASS
PASS
PASS
7.50E-08 1.00E-07 1.50E-07 3.00E-07 5.00E-07
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
16
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
6.00E-07
Input Offset Current 2 (A)
4.00E-07
2.00E-07
0.00E+00
-2.00E-07
-4.00E-07
-6.00E-07
0
50
100
150
Total Dose (krad(Si))
Figure 5.6. Plot of input offset current (comparator 2) versus total dose. The data show a modest degradation
with total dose, however it stays within specification to the maximum total dose tested. The solid diamonds are
the average of the measured data points for the sample irradiated under electrical bias while the shaded
diamonds are the average of the measured data points for the units irradiated with all pins tied to ground. The
black lines (solid or dashed) are the average of the data points after application of the KTL statistics on the
sample irradiated in the biased condition while the shaded lines (solid or dashed) are the average of the data
points after application of the KTL statistics on the sample irradiated in the unbiased condition. The red dotted
line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the
datasheet and/or test plan.
An ISO 9001:2000 Certified Company
17
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.6. Raw data of input offset current (comparator 2) versus total dose, including the statistical
analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Input Offset Current 2 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
-2.93E-08
-3.18E-08
-3.57E-08
-3.27E-08
-3.31E-08
-4.01E-08
-3.37E-08
-3.81E-08
-3.61E-08
-3.81E-08
-1.08E-08
20
-4.84E-08
-5.29E-08
-5.22E-08
-4.81E-08
-5.54E-08
-3.92E-08
-3.76E-08
-3.96E-08
-3.66E-08
-4.02E-08
-1.23E-08
50
-5.83E-08
-7.04E-08
-8.66E-08
-7.31E-08
-7.56E-08
-4.75E-08
-4.89E-08
-5.95E-08
-4.27E-08
-5.40E-08
-1.12E-08
100
-7.85E-08
-9.93E-08
-9.75E-08
-1.12E-07
-1.06E-07
-6.14E-08
-6.26E-08
-6.78E-08
-5.75E-08
-6.84E-08
-1.19E-08
200
-1.39E-07
-1.09E-07
-1.36E-07
-1.52E-07
-1.55E-07
-7.68E-08
-7.56E-08
-8.11E-08
-7.26E-08
-8.97E-08
-1.18E-08
-3.25E-08 -5.14E-08 -7.28E-08 -9.88E-08 -1.38E-07
2.31E-09 3.12E-09 1.02E-08 1.27E-08 1.86E-08
-2.62E-08 -4.28E-08 -4.50E-08 -6.38E-08 -8.71E-08
-3.89E-08 -5.99E-08 -1.01E-07 -1.34E-07 -1.89E-07
-3.72E-08
2.44E-09
-3.05E-08
-4.39E-08
-7.50E-08
PASS
7.50E-08
PASS
-3.86E-08
1.47E-09
-3.46E-08
-4.27E-08
-1.00E-07
PASS
1.00E-07
PASS
-5.05E-08
6.44E-09
-3.28E-08
-6.82E-08
-1.50E-07
PASS
1.50E-07
PASS
An ISO 9001:2000 Certified Company
18
-6.35E-08
4.55E-09
-5.10E-08
-7.60E-08
-3.00E-07
PASS
3.00E-07
PASS
-7.92E-08
6.63E-09
-6.10E-08
-9.73E-08
-5.00E-07
PASS
5.00E-07
PASS
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-06
Positive Bias Current 1 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.7. Plot of input bias current, non-inverting input (comparator 1) versus total dose. The data show a
modest degradation with total dose, however it stays within specification to the maximum total dose tested.
The solid diamonds are the average of measured data points from the biased sample (devices irradiated with an
electrical bias) while the shaded diamonds are the average from the un-biased sample. The black lines show
the effects on the data after application of the biased KTL statistics (solid and dashed lines) while the gray lines
show the effects on the data after application of the unbiased KTL statistics (solid and dashed lines). The red
dashed lines are the minimum and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
19
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.5. Raw data of input bias current, non-inverting input (comparator 1) versus total dose, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Positive Bias Current 1 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
PASS
PASS
0
3.18E-07
2.99E-07
3.27E-07
3.36E-07
3.27E-07
3.14E-07
3.11E-07
3.11E-07
3.21E-07
3.09E-07
3.27E-07
20
4.01E-07
3.84E-07
4.23E-07
4.16E-07
4.15E-07
3.84E-07
3.80E-07
3.83E-07
3.89E-07
3.73E-07
3.29E-07
50
4.52E-07
4.66E-07
5.32E-07
5.16E-07
4.53E-07
4.65E-07
4.51E-07
4.64E-07
4.43E-07
4.21E-07
3.32E-07
100
5.75E-07
5.64E-07
6.05E-07
7.04E-07
5.90E-07
5.67E-07
5.63E-07
5.70E-07
5.80E-07
5.47E-07
3.30E-07
200
7.26E-07
6.67E-07
7.44E-07
7.58E-07
7.32E-07
7.09E-07
7.06E-07
7.05E-07
7.18E-07
7.31E-07
3.32E-07
3.21E-07
1.41E-08
3.60E-07
2.83E-07
4.08E-07
1.53E-08
4.50E-07
3.66E-07
4.84E-07
3.77E-08
5.87E-07
3.80E-07
6.08E-07
5.58E-08
7.60E-07
4.55E-07
7.26E-07
3.48E-08
8.21E-07
6.30E-07
3.13E-07 3.82E-07 4.49E-07 5.65E-07 7.14E-07
4.76E-09 5.84E-09 1.82E-08 1.19E-08 1.08E-08
3.26E-07 3.98E-07 4.99E-07 5.98E-07 7.43E-07
3.00E-07 3.66E-07 3.99E-07 5.33E-07 6.84E-07
-5.00E-07 -7.50E-07 -1.00E-06 -1.50E-06 -2.00E-06
PASS
PASS
PASS
PASS
5.00E-07 7.50E-07 1.00E-06 1.50E-06 2.00E-06
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
20
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-06
Positive Bias Current 2 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.8. Plot of input bias current, non-inverting input (comparator 2) versus total dose. The data show a
modest degradation with total dose, however it stays within specification to the maximum total dose tested. The
solid diamonds are the average of measured data points from the biased sample (devices irradiated with an
electrical bias) while the shaded diamonds are the average from the un-biased sample. The black lines show the
effects on the data after application of the biased KTL statistics (solid and dashed lines) while the gray lines
show the effects on the data after application of the unbiased KTL statistics (solid and dashed lines). The red
dashed lines are the minimum and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
21
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.6. Raw data of input bias current, non-inverting input (comparator 2) versus total dose, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Positive Bias Current 2 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
3.47E-07
3.24E-07
3.57E-07
3.62E-07
3.56E-07
3.47E-07
3.43E-07
3.45E-07
3.44E-07
3.38E-07
3.36E-07
20
4.42E-07
4.02E-07
4.55E-07
4.58E-07
4.09E-07
4.21E-07
4.16E-07
4.22E-07
4.18E-07
4.11E-07
3.36E-07
50
5.40E-07
5.20E-07
6.18E-07
5.73E-07
5.62E-07
5.20E-07
5.19E-07
5.43E-07
4.63E-07
5.11E-07
3.35E-07
100
6.51E-07
6.34E-07
6.82E-07
8.04E-07
6.81E-07
6.34E-07
6.44E-07
6.51E-07
6.35E-07
6.28E-07
3.36E-07
200
8.29E-07
7.53E-07
8.29E-07
8.61E-07
8.46E-07
7.96E-07
8.13E-07
8.15E-07
7.93E-07
8.47E-07
3.36E-07
3.49E-07
1.53E-08
3.91E-07
3.07E-07
4.33E-07
2.60E-08
5.04E-07
3.62E-07
5.63E-07
3.69E-08
6.64E-07
4.61E-07
6.90E-07
6.67E-08
8.73E-07
5.08E-07
8.24E-07
4.15E-08
9.37E-07
7.10E-07
3.43E-07 4.18E-07 5.11E-07 6.39E-07 8.13E-07
3.24E-09 4.46E-09 2.93E-08 9.18E-09 2.14E-08
3.52E-07 4.30E-07 5.91E-07 6.64E-07 8.72E-07
3.34E-07 4.05E-07 4.31E-07 6.13E-07 7.54E-07
-5.00E-07 -7.50E-07 -1.00E-06 -1.50E-06 -2.00E-06
PASS
PASS
PASS
PASS
PASS
5.00E-07 7.50E-07 1.00E-06 1.50E-06 2.00E-06
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
22
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-06
Negative Bias Current 1 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.9. Plot of input bias current, inverting input (comparator 1) versus total dose. The data show a modest
degradation with total dose, however it stays within specification to the maximum total dose tested. The solid
diamonds are the average of measured data points from the biased sample (devices irradiated with an electrical
bias) while the shaded diamonds are the average from the un-biased sample. The black lines show the effects on
the data after application of the biased KTL statistics (solid and dashed lines) while the gray lines show the
effects on the data after application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines
are the minimum and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
23
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.7. Raw data for the input bias current, inverting input (comparator 1) versus total dose, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Negative Bias Current 1 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
3.37E-07
3.15E-07
3.46E-07
3.61E-07
3.44E-07
3.42E-07
3.40E-07
3.38E-07
3.35E-07
3.23E-07
3.21E-07
20
4.15E-07
3.90E-07
4.50E-07
4.43E-07
4.24E-07
4.10E-07
4.12E-07
4.09E-07
4.05E-07
3.96E-07
3.22E-07
50
5.07E-07
4.75E-07
5.51E-07
5.35E-07
5.43E-07
5.06E-07
5.08E-07
5.03E-07
5.01E-07
4.88E-07
3.20E-07
100
5.41E-07
5.31E-07
6.14E-07
7.52E-07
6.15E-07
6.15E-07
6.27E-07
6.17E-07
6.13E-07
5.73E-07
3.20E-07
200
7.13E-07
6.75E-07
7.22E-07
7.47E-07
7.28E-07
7.70E-07
7.94E-07
7.77E-07
7.64E-07
8.04E-07
3.22E-07
3.40E-07
1.68E-08
3.86E-07
2.94E-07
4.25E-07
2.38E-08
4.90E-07
3.59E-07
5.22E-07
3.13E-08
6.08E-07
4.36E-07
6.11E-07
8.82E-08
8.52E-07
3.69E-07
7.17E-07
2.66E-08
7.90E-07
6.44E-07
3.36E-07 4.06E-07 5.01E-07 6.09E-07 7.82E-07
7.51E-09 6.55E-09 7.88E-09 2.08E-08 1.68E-08
3.56E-07 4.24E-07 5.23E-07 6.66E-07 8.28E-07
3.15E-07 3.88E-07 4.80E-07 5.52E-07 7.36E-07
-5.00E-07 -7.50E-07 -1.00E-06 -1.50E-06 -2.00E-06
PASS
PASS
PASS
PASS
PASS
5.00E-07 7.50E-07 1.00E-06 1.50E-06 2.00E-06
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
24
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MIN
Specification MAX
2.50E-06
Negative Bias Current 2 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.10. Plot of input bias current, inverting input (comparator 2) versus total dose. The data show a
modest degradation with total dose, however it stays within specification to the maximum total dose tested.
The solid diamonds are the average of measured data points from the biased sample (devices irradiated with
an electrical bias) while the shaded diamonds are the average from the un-biased sample. The black lines
show the effects on the data after application of the biased KTL statistics (solid and dashed lines) while the
gray lines show the effects on the data after application of the unbiased KTL statistics (solid and dashed
lines). The red dashed lines are the minimum and maximum specification values as defined in the datasheet
and/or test plan.
An ISO 9001:2000 Certified Company
25
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.10. Raw data for the input bias current, inverting input (comparator 2) versus total dose, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Negative Bias Current 2 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
Specification MAX
Status
0
3.18E-07
2.90E-07
3.22E-07
3.27E-07
3.22E-07
3.07E-07
3.09E-07
3.04E-07
3.05E-07
3.00E-07
3.22E-07
20
3.94E-07
3.66E-07
4.21E-07
4.09E-07
4.02E-07
3.79E-07
3.78E-07
3.81E-07
3.79E-07
3.71E-07
3.21E-07
50
4.80E-07
4.49E-07
5.30E-07
4.99E-07
4.85E-07
4.69E-07
4.70E-07
4.81E-07
4.68E-07
4.56E-07
3.22E-07
100
5.49E-07
5.34E-07
5.44E-07
6.90E-07
5.74E-07
5.57E-07
5.34E-07
5.55E-07
5.31E-07
5.60E-07
3.21E-07
200
6.90E-07
5.95E-07
6.94E-07
7.05E-07
6.90E-07
7.17E-07
7.37E-07
7.33E-07
7.20E-07
7.53E-07
3.23E-07
3.16E-07
1.49E-08
3.57E-07
2.75E-07
3.98E-07
2.07E-08
4.55E-07
3.42E-07
4.89E-07
2.93E-08
5.69E-07
4.08E-07
5.78E-07
6.44E-08
7.55E-07
4.02E-07
6.75E-07
4.51E-08
7.98E-07
5.51E-07
3.05E-07 3.77E-07 4.68E-07 5.48E-07 7.32E-07
3.51E-09 3.96E-09 8.84E-09 1.38E-08 1.42E-08
3.15E-07 3.88E-07 4.93E-07 5.85E-07 7.71E-07
2.95E-07 3.66E-07 4.44E-07 5.10E-07 6.93E-07
-5.00E-07 -7.50E-07 -1.00E-06 -1.50E-06 -2.00E-06
PASS
PASS
PASS
PASS
PASS
5.00E-07 7.50E-07 1.00E-06 1.50E-06 2.00E-06
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
26
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
9.00E+01
Open Loop Gain 1 (V/mV)
8.00E+01
7.00E+01
6.00E+01
5.00E+01
4.00E+01
3.00E+01
2.00E+01
1.00E+01
0.00E+00
0
50
100
150
Total Dose (krad(Si))
Figure 5.11. Plot of open loop gain (comparator 1) versus total dose using the datasheet units of V/mV. The data
show no significant degradation with total dose, however the KTL statistics are erratic due to the sensitivity of the
measurement to input conditions and the limitations of the test equipment. The solid diamonds are the average of
measured data points from the biased sample (devices irradiated with an electrical bias) while the shaded
diamonds are the average from the un-biased sample. The black lines show the effects on the data after
application of the biased KTL statistics (solid and dashed lines) while the gray lines show the effects on the data
after application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum
and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
27
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.11. Raw data for the open loop gain (comparator 1) versus total dose using the datasheet units of
V/mV, including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Open Loop Gain 1 (V/mV)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
7.86E+01
8.78E+01
5.58E+01
6.69E+01
7.03E+01
7.55E+01
5.87E+01
7.71E+01
7.31E+01
7.15E+01
9.97E+01
20
6.70E+01
9.16E+01
7.08E+01
7.85E+01
8.67E+01
7.57E+01
8.25E+01
6.86E+01
9.54E+01
9.14E+01
7.16E+01
50
7.63E+01
7.49E+01
8.05E+01
8.23E+01
7.57E+01
7.80E+01
7.86E+01
6.68E+01
7.00E+01
1.16E+02
7.74E+01
100
8.93E+01
7.60E+01
7.93E+01
7.33E+01
8.62E+01
9.08E+01
6.86E+01
6.66E+01
8.64E+01
7.93E+01
5.53E+01
200
7.69E+01
8.78E+01
8.55E+01
7.77E+01
8.54E+01
8.07E+01
7.09E+01
9.12E+01
6.99E+01
6.93E+01
5.40E+01
7.19E+01
1.21E+01
1.05E+02
3.87E+01
7.89E+01
1.04E+01
1.07E+02
5.05E+01
7.79E+01
3.26E+00
8.69E+01
6.90E+01
8.08E+01
6.76E+00
9.94E+01
6.23E+01
8.27E+01
4.99E+00
9.64E+01
6.90E+01
7.12E+01
7.30E+00
9.12E+01
5.12E+01
1.00E+01
PASS
8.27E+01
1.10E+01
1.13E+02
5.25E+01
1.00E+01
PASS
8.18E+01
1.95E+01
1.35E+02
2.82E+01
1.00E+01
PASS
7.83E+01
1.07E+01
1.08E+02
4.91E+01
1.00E+01
PASS
7.64E+01
9.49E+00
1.02E+02
5.04E+01
5.00E+00
PASS
An ISO 9001:2000 Certified Company
28
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
1.05E+02
1.00E+02
Open Loop Gain 1 (dB)
9.50E+01
9.00E+01
8.50E+01
8.00E+01
7.50E+01
7.00E+01
6.50E+01
6.00E+01
0
50
100
150
Total Dose (krad(Si))
Figure 5.12. Plot of open loop gain (comparator 1) versus total dose using units of dB. The data show no
significant degradation with total dose. Note that using units of dB to measure the gain (a commonly used
unit for this measurement) removes the erratic behavior of the KTL statistics observed in Figure 5.11. The
solid diamonds are the average of measured data points from the biased sample (devices irradiated with an
electrical bias) while the shaded diamonds are the average from the un-biased sample. The black lines show
the effects on the data after application of the biased KTL statistics (solid and dashed lines) while the gray
lines show the effects on the data after application of the unbiased KTL statistics (solid and dashed lines).
The red dashed lines are the minimum and maximum specification values as defined in the datasheet and/or
test plan.
An ISO 9001:2000 Certified Company
29
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.12. Raw data for the open loop gain (comparator 1) versus total dose using units of dB, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Open Loop Gain 1 (dB)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
97.91
98.87
94.93
96.51
96.94
97.56
95.37
97.74
97.28
97.09
99.97
9.70E+01
1.49E+00
1.01E+02
9.30E+01
9.70E+01
9.48E-01
9.96E+01
9.44E+01
8.00E+01
PASS
Total Dose (krad(Si))
20
50
100
96.52
97.65
99.02
99.24
97.49
97.62
97.00
98.12
97.99
97.90
98.31
97.30
98.76
97.58
98.71
97.58
97.84
99.16
98.33
97.91
96.73
96.73
96.50
96.47
99.59
96.90
98.73
99.22
101.25
97.99
97.10
97.77
94.85
9.79E+01 9.78E+01 9.81E+01 9.83E+01
1.14E+00 3.61E-01 7.23E-01 5.30E-01
1.01E+02 9.88E+01 1.00E+02 9.98E+01
9.47E+01 9.68E+01 9.61E+01 9.69E+01
9.83E+01
1.17E+00
1.02E+02
9.51E+01
8.00E+01
PASS
9.81E+01
1.87E+00
1.03E+02
9.29E+01
8.00E+01
PASS
An ISO 9001:2000 Certified Company
30
200
97.72
98.87
98.64
97.81
98.63
98.14
97.01
99.20
96.89
96.81
94.65
9.78E+01
1.19E+00
1.01E+02
9.45E+01
8.00E+01
PASS
9.76E+01
1.04E+00
1.00E+02
9.48E+01
7.40E+01
PASS
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
1.00E+02
9.00E+01
Open Loop Gain 2 (V/mV)
8.00E+01
7.00E+01
6.00E+01
5.00E+01
4.00E+01
3.00E+01
2.00E+01
1.00E+01
0.00E+00
0
50
100
150
Total Dose (krad(Si))
Figure 5.13. Plot of open loop gain (comparator 2) versus total dose using the datasheet units of V/mV. The
data show no significant degradation with total dose, however the KTL statistics are erratic due to the
sensitivity of the measurement to input conditions and the limitations of the test equipment. The solid
diamonds are the average of measured data points from the biased sample (devices irradiated with an
electrical bias) while the shaded diamonds are the average from the un-biased sample. The black lines show
the effects on the data after application of the biased KTL statistics (solid and dashed lines) while the gray
lines show the effects on the data after application of the unbiased KTL statistics (solid and dashed lines).
The red dashed lines are the minimum and maximum specification values as defined in the datasheet and/or
test plan.
An ISO 9001:2000 Certified Company
31
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.13. Raw data for the open loop gain (comparator 2) versus total dose using the datasheet units of
V/mV, including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Open Loop Gain 2 (V/mV)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
9.15E+01
7.74E+01
7.24E+01
7.92E+01
7.31E+01
7.84E+01
9.33E+01
8.55E+01
9.00E+01
7.40E+01
9.40E+01
20
1.08E+02
7.74E+01
6.56E+01
8.97E+01
1.01E+02
7.00E+01
9.28E+01
9.14E+01
8.07E+01
7.65E+01
8.55E+01
50
9.02E+01
7.65E+01
7.25E+01
9.04E+01
7.57E+01
8.44E+01
6.95E+01
8.03E+01
8.26E+01
8.03E+01
9.40E+01
100
9.82E+01
7.65E+01
8.29E+01
7.86E+01
8.13E+01
9.54E+01
6.88E+01
8.13E+01
9.04E+01
6.79E+01
1.06E+02
200
7.31E+01
8.90E+01
9.51E+01
6.77E+01
8.44E+01
8.90E+01
7.95E+01
9.02E+01
6.79E+01
6.65E+01
8.78E+01
7.87E+01
7.69E+00
9.98E+01
5.76E+01
8.84E+01
1.72E+01
1.36E+02
4.11E+01
8.11E+01
8.57E+00
1.05E+02
5.76E+01
8.35E+01
8.58E+00
1.07E+02
6.00E+01
8.19E+01
1.13E+01
1.13E+02
5.09E+01
8.42E+01
8.00E+00
1.06E+02
6.23E+01
1.00E+01
PASS
8.23E+01
9.75E+00
1.09E+02
5.55E+01
1.00E+01
PASS
7.94E+01
5.81E+00
9.53E+01
6.35E+01
1.00E+01
PASS
8.08E+01
1.24E+01
1.15E+02
4.67E+01
1.00E+01
PASS
7.86E+01
1.12E+01
1.09E+02
4.78E+01
5.00E+00
PASS
An ISO 9001:2000 Certified Company
32
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
1.05E+02
1.00E+02
Open Loop Gain 2 (dB)
9.50E+01
9.00E+01
8.50E+01
8.00E+01
7.50E+01
7.00E+01
6.50E+01
6.00E+01
0
50
100
150
Total Dose (krad(Si))
Figure 5.14. Plot of open loop gain (comparator 2) versus total dose using units of dB. The data show no
significant degradation with total dose. Note that using units of dB to measure the gain (a commonly used unit
for this measurement) removes the erratic behavior of the KTL statistics observed in Figure 5.13. The solid
diamonds are the average of measured data points from the biased sample (devices irradiated with an electrical
bias) while the shaded diamonds are the average from the un-biased sample. The black lines show the effects
on the data after application of the biased KTL statistics (solid and dashed lines) while the gray lines show the
effects on the data after application of the unbiased KTL statistics (solid and dashed lines). The red dashed
lines are the minimum and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
33
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.14. Raw data for the open loop gain (comparator 2) versus total dose using units of dB, including
the statistical analysis, the specification and the status of the testing (pass/fail).
Open Loop Gain 2 (dB)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
99.23
97.77
97.19
97.97
97.28
97.89
99.40
98.64
99.08
97.38
99.46
9.79E+01
8.17E-01
1.00E+02
9.57E+01
9.85E+01
8.34E-01
1.01E+02
9.62E+01
8.00E+01
PASS
Total Dose (krad(Si))
20
50
100
100.66
99.10
99.84
97.77
97.67
97.67
96.34
97.21
98.37
99.06
99.12
97.91
100.11
97.58
98.20
96.90
98.53
99.59
99.35
96.84
96.75
99.22
98.09
98.20
98.14
98.34
99.12
97.67
98.09
96.64
98.64
99.46
100.52
9.88E+01 9.81E+01 9.84E+01 9.82E+01
1.76E+00 9.08E-01 8.50E-01 1.22E+00
1.04E+02 1.01E+02 1.01E+02 1.02E+02
9.40E+01 9.56E+01 9.61E+01 9.48E+01
9.83E+01
1.04E+00
1.01E+02
9.54E+01
8.00E+01
PASS
9.80E+01
6.62E-01
9.98E+01
9.62E+01
8.00E+01
PASS
An ISO 9001:2000 Certified Company
34
200
97.28
98.99
99.56
96.61
98.53
98.99
98.01
99.10
96.64
96.46
98.87
9.81E+01
1.34E+00
1.02E+02
9.44E+01
8.00E+01
PASS
9.78E+01
1.26E+00
1.01E+02
9.44E+01
7.40E+01
PASS
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Common Mode Rejection Ratio 1 (dB)
1.50E+02
1.40E+02
1.30E+02
1.20E+02
1.10E+02
1.00E+02
9.00E+01
8.00E+01
7.00E+01
6.00E+01
5.00E+01
0
50
100
150
Total Dose (krad(Si))
Figure 5.15. Plot of common mode rejection ratio (comparator 1) versus total dose. The data show no
significant degradation with total dose. The solid diamonds are the average of measured data points from
the biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the average
from the un-biased sample. The black lines show the effects on the data after application of the biased KTL
statistics (solid and dashed lines) while the gray lines show the effects on the data after application of the
unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum and maximum
specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
35
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.15. Raw data for the common mode rejection ratio (comparator 1), including the statistical
analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Common Mode Rejection Ratio 1 (dB)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
1.36E+02
1.28E+02
1.18E+02
1.22E+02
1.30E+02
1.28E+02
1.15E+02
1.23E+02
1.24E+02
1.23E+02
1.15E+02
20
1.34E+02
1.39E+02
1.17E+02
1.48E+02
1.34E+02
1.21E+02
1.14E+02
1.21E+02
1.18E+02
1.21E+02
1.13E+02
50
1.24E+02
1.20E+02
1.28E+02
1.26E+02
1.31E+02
1.23E+02
1.16E+02
1.17E+02
1.13E+02
1.23E+02
1.19E+02
100
1.26E+02
1.19E+02
1.27E+02
1.46E+02
1.28E+02
1.21E+02
1.14E+02
1.20E+02
1.15E+02
1.29E+02
1.17E+02
200
1.40E+02
1.58E+02
1.39E+02
1.42E+02
1.28E+02
1.21E+02
1.16E+02
1.20E+02
1.20E+02
1.34E+02
1.21E+02
1.27E+02
7.25E+00
1.47E+02
1.07E+02
1.35E+02
1.14E+01
1.66E+02
1.03E+02
1.26E+02
4.22E+00
1.37E+02
1.14E+02
1.29E+02
1.01E+01
1.57E+02
1.02E+02
1.42E+02
1.10E+01
1.72E+02
1.11E+02
1.23E+02
4.95E+00
1.36E+02
1.09E+02
9.00E+01
PASS
1.19E+02
3.13E+00
1.28E+02
1.11E+02
9.00E+01
PASS
1.18E+02
4.30E+00
1.30E+02
1.07E+02
9.00E+01
PASS
1.20E+02
6.00E+00
1.37E+02
1.04E+02
9.00E+01
PASS
1.22E+02
6.97E+00
1.41E+02
1.03E+02
9.00E+01
PASS
An ISO 9001:2000 Certified Company
36
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (-KTL) Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Common Mode Rejection Ratio 2 (dB)
1.30E+02
1.20E+02
1.10E+02
1.00E+02
9.00E+01
8.00E+01
7.00E+01
6.00E+01
5.00E+01
0
50
100
150
Total Dose (krad(Si))
Figure 5.16. Plot of common mode rejection ratio (comparator 2) versus total dose. The data show no
significant degradation with total dose. The solid diamonds are the average of measured data points from the
biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the average from the
un-biased sample. The black lines show the effects on the data after application of the biased KTL statistics
(solid and dashed lines) while the gray lines show the effects on the data after application of the unbiased KTL
statistics (solid and dashed lines). The red dashed lines are the minimum and maximum specification values as
defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
37
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.16. Raw data for the common mode rejection ratio (comparator 2) versus total dose, including the
statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Common Mode Rejection Ratio 2 (dB)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MIN
Status
0
1.12E+02
1.30E+02
1.19E+02
1.15E+02
1.14E+02
1.32E+02
1.13E+02
1.19E+02
1.24E+02
1.14E+02
1.34E+02
20
1.14E+02
1.19E+02
1.18E+02
1.14E+02
1.15E+02
1.39E+02
1.13E+02
1.18E+02
1.19E+02
1.18E+02
1.33E+02
50
1.12E+02
1.15E+02
1.19E+02
1.12E+02
1.16E+02
1.32E+02
1.13E+02
1.20E+02
1.19E+02
1.16E+02
1.39E+02
100
1.13E+02
1.25E+02
1.14E+02
1.11E+02
1.15E+02
1.32E+02
1.12E+02
1.20E+02
1.23E+02
1.15E+02
1.30E+02
200
1.15E+02
1.28E+02
1.16E+02
1.11E+02
1.15E+02
1.32E+02
1.13E+02
1.19E+02
1.21E+02
1.14E+02
1.42E+02
1.18E+02
7.29E+00
1.38E+02
9.81E+01
1.16E+02
2.36E+00
1.22E+02
1.09E+02
1.15E+02
3.15E+00
1.23E+02
1.06E+02
1.16E+02
5.51E+00
1.31E+02
1.00E+02
1.17E+02
6.59E+00
1.35E+02
9.91E+01
1.20E+02
7.58E+00
1.41E+02
9.94E+01
9.00E+01
PASS
1.21E+02
1.01E+01
1.49E+02
9.37E+01
9.00E+01
PASS
1.20E+02
7.01E+00
1.39E+02
1.01E+02
9.00E+01
PASS
1.20E+02
7.87E+00
1.42E+02
9.89E+01
9.00E+01
PASS
1.20E+02
7.77E+00
1.41E+02
9.84E+01
9.00E+01
PASS
An ISO 9001:2000 Certified Company
38
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Voltage Output Low 1 @ 3.2mA (V)
4.50E-01
4.00E-01
3.50E-01
3.00E-01
2.50E-01
2.00E-01
0
50
100
150
Total Dose (krad(Si))
Figure 5.17. Plot of the saturation voltage (VSAT) at 3.2mA for comparator 1 versus total dose. The data
show no significant degradation with total dose. The solid diamonds are the average of measured data points
from the biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the
average from the un-biased sample. The black lines show the effects on the data after application of the
biased KTL statistics (solid and dashed lines) while the gray lines show the effects on the data after
application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum
and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
39
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.17. Raw data for the saturation voltage (VSAT) at 3.2mA for comparator 1 versus total dose,
including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Voltage Output Low 1 @ 3.2mA (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
2.48E-01
2.44E-01
2.44E-01
2.48E-01
2.48E-01
2.46E-01
2.42E-01
2.41E-01
2.40E-01
2.41E-01
2.51E-01
20
2.47E-01
2.44E-01
2.41E-01
2.49E-01
2.48E-01
2.45E-01
2.41E-01
2.41E-01
2.40E-01
2.40E-01
2.52E-01
50
2.48E-01
2.44E-01
2.43E-01
2.51E-01
2.50E-01
2.48E-01
2.43E-01
2.43E-01
2.42E-01
2.42E-01
2.52E-01
100
2.51E-01
2.48E-01
2.49E-01
2.51E-01
2.52E-01
2.51E-01
2.47E-01
2.46E-01
2.46E-01
2.44E-01
2.52E-01
200
2.56E-01
2.53E-01
2.53E-01
2.57E-01
2.57E-01
2.56E-01
2.52E-01
2.51E-01
2.50E-01
2.51E-01
2.52E-01
2.46E-01
2.19E-03
2.52E-01
2.40E-01
2.46E-01
3.27E-03
2.55E-01
2.37E-01
2.47E-01
3.56E-03
2.57E-01
2.37E-01
2.50E-01
1.64E-03
2.55E-01
2.46E-01
2.55E-01
2.05E-03
2.61E-01
2.50E-01
2.42E-01 2.41E-01 2.44E-01 2.47E-01 2.52E-01
2.35E-03 2.07E-03 2.51E-03 2.59E-03 2.35E-03
2.48E-01 2.47E-01 2.50E-01 2.54E-01 2.58E-01
2.36E-01 2.36E-01 2.37E-01 2.40E-01 2.46E-01
4.00E-01 4.00E-01 4.00E-01 4.00E-01 4.00E-01
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
40
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Voltage Output Low 2 @ 3.2mA (V)
4.50E-01
4.00E-01
3.50E-01
3.00E-01
2.50E-01
2.00E-01
0
50
100
150
Total Dose (krad(Si))
Figure 5.18. Plot of the saturation voltage (VSAT) at 3.2mA for comparator 2 versus total dose. The data
show no significant degradation with total dose. The solid diamonds are the average of measured data points
from the biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the
average from the un-biased sample. The black lines show the effects on the data after application of the
biased KTL statistics (solid and dashed lines) while the gray lines show the effects on the data after
application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum
and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
41
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.18. Raw data for the saturation voltage (VSAT) at 3.2mA for comparator 2 versus total dose,
including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Voltage Output Low 2 @ 3.2mA (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
2.45E-01
2.38E-01
2.41E-01
2.48E-01
2.45E-01
2.46E-01
2.43E-01
2.38E-01
2.40E-01
2.38E-01
2.51E-01
20
2.43E-01
2.38E-01
2.39E-01
2.48E-01
2.46E-01
2.43E-01
2.41E-01
2.36E-01
2.38E-01
2.36E-01
2.51E-01
50
2.46E-01
2.39E-01
2.41E-01
2.50E-01
2.47E-01
2.46E-01
2.44E-01
2.37E-01
2.41E-01
2.39E-01
2.52E-01
100
2.48E-01
2.41E-01
2.45E-01
2.50E-01
2.49E-01
2.49E-01
2.46E-01
2.41E-01
2.43E-01
2.40E-01
2.52E-01
200
2.54E-01
2.47E-01
2.49E-01
2.57E-01
2.54E-01
2.55E-01
2.52E-01
2.45E-01
2.49E-01
2.48E-01
2.52E-01
2.43E-01
3.91E-03
2.54E-01
2.33E-01
2.43E-01
4.32E-03
2.55E-01
2.31E-01
2.45E-01
4.51E-03
2.57E-01
2.32E-01
2.47E-01
3.65E-03
2.57E-01
2.37E-01
2.52E-01
4.09E-03
2.63E-01
2.41E-01
2.41E-01 2.39E-01 2.41E-01 2.44E-01 2.50E-01
3.46E-03 3.11E-03 3.65E-03 3.70E-03 3.83E-03
2.50E-01 2.47E-01 2.51E-01 2.54E-01 2.60E-01
2.32E-01 2.30E-01 2.31E-01 2.34E-01 2.39E-01
4.00E-01 4.00E-01 4.00E-01 4.00E-01 4.00E-01
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
42
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Voltage Output Low 1 @ 25mA (V)
1.60E+00
1.40E+00
1.20E+00
1.00E+00
8.00E-01
6.00E-01
4.00E-01
2.00E-01
0
50
100
150
Total Dose (krad(Si))
Figure 5.19. Plot of the saturation voltage (VSAT) at 25mA for comparator 1 versus total dose. The data
show no significant degradation with total dose. The solid diamonds are the average of measured data points
from the biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the
average from the un-biased sample. The black lines show the effects on the data after application of the
biased KTL statistics (solid and dashed lines) while the gray lines show the effects on the data after
application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum
and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
43
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.19. Raw data for the saturation voltage (VSAT) at 25mA for comparator 1 versus total dose versus
total dose, including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Voltage Output Low 1 @ 25mA (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
6.33E-01
6.27E-01
6.22E-01
6.29E-01
6.34E-01
6.29E-01
6.17E-01
6.21E-01
6.24E-01
6.19E-01
6.43E-01
20
6.26E-01
6.22E-01
6.12E-01
6.24E-01
6.29E-01
6.28E-01
6.15E-01
6.17E-01
6.23E-01
6.15E-01
6.42E-01
50
6.28E-01
6.25E-01
6.17E-01
6.27E-01
6.32E-01
6.27E-01
6.15E-01
6.18E-01
6.26E-01
6.17E-01
6.42E-01
100
6.33E-01
6.30E-01
6.25E-01
6.31E-01
6.37E-01
6.31E-01
6.18E-01
6.22E-01
6.29E-01
6.22E-01
6.41E-01
200
6.39E-01
6.36E-01
6.31E-01
6.40E-01
6.45E-01
6.37E-01
6.26E-01
6.30E-01
6.35E-01
6.26E-01
6.40E-01
6.29E-01
4.85E-03
6.42E-01
6.16E-01
6.23E-01
6.47E-03
6.40E-01
6.05E-01
6.26E-01
5.54E-03
6.41E-01
6.11E-01
6.31E-01
4.38E-03
6.43E-01
6.19E-01
6.38E-01
5.17E-03
6.52E-01
6.24E-01
6.22E-01 6.20E-01 6.21E-01 6.24E-01 6.31E-01
4.69E-03 5.73E-03 5.50E-03 5.41E-03 5.07E-03
6.35E-01 6.35E-01 6.36E-01 6.39E-01 6.45E-01
6.09E-01 6.04E-01 6.06E-01 6.10E-01 6.17E-01
1.50E+00 1.50E+00 1.50E+00 1.50E+00 1.50E+00
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
44
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
Voltage Output Low 2 @ 25mA (V)
1.60E+00
1.40E+00
1.20E+00
1.00E+00
8.00E-01
6.00E-01
4.00E-01
2.00E-01
0
50
100
150
Total Dose (krad(Si))
Figure 5.20. Plot of the saturation voltage (VSAT) at 25mA for comparator 2 versus total dose. The data
show no significant degradation with total dose. The solid diamonds are the average of measured data points
from the biased sample (devices irradiated with an electrical bias) while the shaded diamonds are the
average from the un-biased sample. The black lines show the effects on the data after application of the
biased KTL statistics (solid and dashed lines) while the gray lines show the effects on the data after
application of the unbiased KTL statistics (solid and dashed lines). The red dashed lines are the minimum
and maximum specification values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
45
200
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table 5.20. Raw data of the saturation voltage (VSAT) at 25mA for comparator 2 versus total dose,
including the statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Voltage Output Low 2 @ 25mA (V)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
6.24E-01
6.13E-01
6.17E-01
6.30E-01
6.19E-01
6.26E-01
6.13E-01
6.12E-01
6.20E-01
6.09E-01
6.34E-01
20
6.18E-01
6.08E-01
6.10E-01
6.25E-01
6.15E-01
6.25E-01
6.09E-01
6.09E-01
6.17E-01
6.05E-01
6.32E-01
50
6.20E-01
6.12E-01
6.12E-01
6.28E-01
6.17E-01
6.23E-01
6.10E-01
6.10E-01
6.18E-01
6.06E-01
6.32E-01
100
6.26E-01
6.16E-01
6.20E-01
6.31E-01
6.22E-01
6.27E-01
6.14E-01
6.14E-01
6.22E-01
6.12E-01
6.32E-01
200
6.32E-01
6.22E-01
6.27E-01
6.40E-01
6.31E-01
6.32E-01
6.21E-01
6.21E-01
6.28E-01
6.15E-01
6.31E-01
6.21E-01
6.58E-03
6.39E-01
6.03E-01
6.15E-01
6.76E-03
6.34E-01
5.97E-01
6.18E-01
6.65E-03
6.36E-01
6.00E-01
6.23E-01
5.74E-03
6.39E-01
6.07E-01
6.30E-01
6.66E-03
6.49E-01
6.12E-01
6.16E-01 6.13E-01 6.13E-01 6.18E-01 6.23E-01
6.89E-03 8.00E-03 6.91E-03 6.42E-03 6.66E-03
6.35E-01 6.35E-01 6.32E-01 6.35E-01 6.42E-01
5.97E-01 5.91E-01 5.94E-01 6.00E-01 6.05E-01
1.50E+00 1.50E+00 1.50E+00 1.50E+00 1.50E+00
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
46
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
2.50E-06
Output Leakage Current 1 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.21. Plot of output leakage current (comparator 1) versus total dose. The data show no significant
degradation with total dose. The solid diamonds are the average of measured data points from the biased
sample (devices irradiated with an electrical bias) while the shaded diamonds are the average from the unbiased sample. The black lines show the effects on the data after application of the biased KTL statistics
(solid and dashed lines) while the gray lines show the effects on the data after application of the unbiased
KTL statistics (solid and dashed lines). The red dashed lines are the minimum and maximum specification
values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
47
200
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table 5.21. Raw data of the output leakage current (comparator 1) versus total dose, including the
statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Output Leakage Current 1 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
20
50
100
200
4.00E-09 2.00E-09 2.00E-09 2.00E-09 4.00E-09
3.00E-09 2.00E-09 3.00E-09 3.00E-09 2.00E-09
0.00E+00 1.00E-09 2.00E-09 4.00E-09 1.00E-09
2.00E-09 0.00E+00 0.00E+00 0.00E+00 4.00E-09
2.00E-09 0.00E+00 0.00E+00 2.00E-09 0.00E+00
0.00E+00 1.00E-09 2.00E-09 4.00E-09 3.00E-09
3.00E-09 1.00E-09 1.00E-09 3.00E-09 3.00E-09
0.00E+00 1.00E-09 2.00E-09 3.00E-09 4.00E-09
2.00E-09 3.00E-09 2.00E-09 0.00E+00 4.00E-09
2.00E-09 5.00E-09 2.00E-09 3.00E-09 2.00E-09
1.00E-09 0.00E+00 1.00E-09 1.00E-09 1.00E-09
2.20E-09 1.00E-09 1.40E-09 2.20E-09 2.20E-09
1.48E-09 1.00E-09 1.34E-09 1.48E-09 1.79E-09
6.27E-09 3.74E-09 5.08E-09 6.27E-09 7.11E-09
-1.87E-09 -1.74E-09 -2.28E-09 -1.87E-09 -2.71E-09
1.40E-09 2.20E-09 1.80E-09 2.60E-09 3.20E-09
1.34E-09 1.79E-09 4.47E-10 1.52E-09 8.37E-10
5.08E-09 7.11E-09 3.03E-09 6.76E-09 5.49E-09
-2.28E-09 -2.71E-09 5.74E-10 -1.56E-09 9.06E-10
2.00E-06 2.00E-06 2.00E-06 2.00E-06 2.00E-06
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
48
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Average Biased
Average Un-Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (+KTL) Un-Biased
Specification MAX
2.50E-06
Output Leakage Current 2 (A)
2.00E-06
1.50E-06
1.00E-06
5.00E-07
0.00E+00
-5.00E-07
-1.00E-06
-1.50E-06
-2.00E-06
-2.50E-06
0
50
100
150
Total Dose (krad(Si))
Figure 5.22. Plot of output leakage current (comparator 2) versus total dose. The data show no significant
degradation with total dose. The solid diamonds are the average of measured data points from the biased
sample (devices irradiated with an electrical bias) while the shaded diamonds are the average from the unbiased sample. The black lines show the effects on the data after application of the biased KTL statistics
(solid and dashed lines) while the gray lines show the effects on the data after application of the unbiased
KTL statistics (solid and dashed lines). The red dashed lines are the minimum and maximum specification
values as defined in the datasheet and/or test plan.
An ISO 9001:2000 Certified Company
49
200
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table 5.22. Raw data for the output leakage current (comparator 2) versus total dose, including the
statistical analysis, the specification and the status of the testing (pass/fail).
Total Dose (krad(Si))
Output Leakage Current 2 (A)
Device
694
695
698
699
700
701
702
704
705
706
1280
Biased Statistics
Average Biased
Std Dev Biased
Ps90%/90% (+KTL) Biased
Ps90%/90% (-KTL) Biased
Un-Biased Statistics
Average Un-Biased
Std Dev Un-Biased
Ps90%/90% (+KTL) Un-Biased
Ps90%/90% (-KTL) Un-Biased
Specification MAX
Status
0
20
50
100
200
0.00E+00 3.00E-09 2.00E-09 2.00E-09 0.00E+00
1.00E-09 2.00E-09 3.00E-09 2.00E-09 2.00E-09
1.00E-09 0.00E+00 0.00E+00 0.00E+00 2.00E-09
0.00E+00 2.00E-09 2.00E-09 0.00E+00 2.00E-09
1.00E-09 0.00E+00 3.00E-09 2.00E-09 2.00E-09
1.00E-09 2.00E-09 2.00E-09 0.00E+00 1.00E-09
3.00E-09 1.00E-09 1.00E-09 3.00E-09 3.00E-09
1.00E-09 2.00E-09 1.00E-09 4.00E-09 4.00E-09
3.00E-09 2.00E-09 1.00E-09 2.00E-09 2.00E-09
0.00E+00 1.00E-09 0.00E+00 1.00E-09 1.00E-09
0.00E+00 0.00E+00 2.00E-09 2.00E-09 0.00E+00
6.00E-10 1.40E-09 2.00E-09 1.20E-09 1.60E-09
5.48E-10 1.34E-09 1.22E-09 1.10E-09 8.94E-10
2.10E-09 5.08E-09 5.36E-09 4.20E-09 4.05E-09
-9.02E-10 -2.28E-09 -1.36E-09 -1.80E-09 -8.53E-10
1.60E-09 1.60E-09 1.00E-09 2.00E-09 2.20E-09
1.34E-09 5.48E-10 7.07E-10 1.58E-09 1.30E-09
5.28E-09 3.10E-09 2.94E-09 6.34E-09 5.78E-09
-2.08E-09 9.81E-11 -9.39E-10 -2.34E-09 -1.38E-09
2.00E-06 2.00E-06 2.00E-06 2.00E-06 2.00E-06
PASS
PASS
PASS
PASS
PASS
An ISO 9001:2000 Certified Company
50
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
6.0. Summary / Conclusions
The total ionizing dose testing described in this final report was performed using the facilities at
Radiation Assured Devices’ Longmire Laboratories in Colorado Springs, CO. The high dose rate total
ionizing dose (TID) source is a JLSA 84-21 irradiator modified to provide a panoramic exposure. The
Co-60 rods are held in the base of the irradiator heavily shielded by lead, during the radiation exposures
the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for
this irradiator in this configuration ranges from <1rad(Si)/s to a maximum of approximately
120rad(Si)/s, determined by the distance from the source.
The parametric data was obtained as “read and record” and all the raw data plus an attributes summary
were presented in this report. The attributes data contains the average, standard deviation and the
average with the KTL values applied. The KTL value used was 2.742 per MIL HDBK 814 using onesided tolerance limits of 90/90 and a 5-piece sample size. Note that the following criteria was used to
determine the outcome of the testing: following the radiation exposure each parameter had to pass the
specification value and the average value for the five-piece sample must pass the specification value
when the KTL limits are applied. If these conditions were not both satisfied following the radiation
exposure, then the lot could be logged as an RLAT failure.
The units passed the radiation lot acceptance test with only minor degradation to the input offset and
input bias current parameters and no significant degradation to any other measured parameter.
An ISO 9001:2000 Certified Company
51
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix A: TID Bias Connections
Biased Samples:
Pin
Function
Connection / Bias
1
OUT1
+15V Via 10kΩ
2
GND1
GND
3
+IN1
To Pin 8 and –15V via 10kΩ
4
-IN1
To Pin 9 and +15V via 10kΩ and Pin 8 via 100Ω
5
V-
-15V
6
OUT2
+15V Via 10kΩ
7
GND2
GND
8
+IN2
To Pin 3 and –15V via 10kΩ and Pin 4 via 100Ω
9
-IN2
To Pin 4 and +15V via 10kΩ
10
V+
+15V
Unbiased Samples:
Pin
Function
Connection / Bias
1
OUT1
2
GND1
GND
GND
3
+IN1
GND
4
-IN1
GND
5
V-
GND
6
OUT2
GND
7
GND2
GND
8
+IN2
GND
9
-IN2
GND
10
V+
GND
An ISO 9001:2000 Certified Company
52
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Figure A.1. Irradiation bias drawing for the units to be irradiated under electrical bias. This figure was extracted
from LINEAR TECHNOLOGY CORPORATION, RH119 Datasheet.
Figure A.2. Package drawing (for reference only). This figure was extracted from LINEAR TECHNOLOGY
CORPORATION, RH119 Datasheet.
An ISO 9001:2000 Certified Company
53
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix B: Photograph of device-under-test to show part markings
An ISO 9001:2000 Certified Company
54
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Appendix C: Electrical Test Parameters and Conditions (All Subgroup 1 Parameters)
All electrical tests for this device are performed on one of Radiation Assured Device’s LTS2020 Test
Systems. The LTS2020 Test System is a programmable parametric tester that provides parameter
measurements for a variety of digital, analog and mixed signal products including dual comparators,
voltage comparators, D to A and A to D converters. The LTS2020 Test System achieves accuracy and
sensitivity through the use of software self-calibration and an internal relay matrix with separate family
boards and custom personality adapter boards. The tester uses this relay matrix to connect the required
test circuits, select the appropriate voltage / current sources and establish the needed measurement loops
for all the tests performed. The measured parameters and test conditions are shown in Table C.1
A listing of the measurement precision/resolution for each parameter is shown in Table C.2. The
precision/resolution values were obtained either from test data or from the DAC resolution of the LTS2020 for the particular test shown, whichever is greater. To generate the precision/resolution shown in
Table C.2, one of the units-under-test was tested repetitively (a total of 10-times with re-insertion
between tests) to obtain the average test value and standard deviation. Using this test data MIL-HDBK814 90/90 KTL statistics were applied to the measured standard deviation to generate the final
measurement range. This value encompasses the precision/resolution of all aspects of the test system,
including the LTS2020 mainframe, family board, socket assembly and DUT board as well as insertion
error. In some cases, the measurement resolution is limited by the internal DACs, which results in a
measured standard deviation of zero. In these instances the precision/resolution will be reported back as
the LSB of the DAC.
An ISO 9001:2000 Certified Company
55
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
RLAT Report
R1.1 081220
Table C.1. Measured parameters and test conditions for the RH119W.
TEST
DESCRIPTION
TEST
CONDITIONS
Input Offset Voltage, VOS1, VOS2
VS = ±15V, VCM = 0V
Common Mode Rejection Ratio, CMRR1, CMRR2
VS = ±15V
Input Offset Current, IOS1, IOS2
VS = ±15V, VCM = 0V
Input Bias Current, Non-Inverting Input, +IB1, +IB2
VS = ±15V, VCM = 0V
Input Bias Current, Inverting Input, -IB1, -IB2
VS = ±15V, VCM = 0V
Large Signal Voltage Gain, AVOL1, AVOL2
VS = ±15V, VCM = 0V
Saturation Voltage, VSAT1, VSAT2
+VS = +4.5V, -VS=0V, ISINK = 3.2mA
Saturation Voltage, VSAT3, VSAT4
VS = ±15V, IOUT = 25mA
Output Leakage Current, ICEX1, ICEX2
VIN=5mV, VOUT to V– = 35V
Positive Supply Current, +IS
VS = ±15V
Negative Supply Current, -IS
VS = ±15V
An ISO 9001:2000 Certified Company
56
RLAT Report
R1.1 081220
Radiation Assured Devices
5017 N. 30th Street
Colorado Springs, CO 80919
(719) 531-0800
Table C.2. Measured parameters, pre-irradiation specifications and measurement resolution
for the RH119W. The resolution is based either on test data or the LSB of the DAC for a
given measurement.
Measured Parameter
Pre-Irradiation
Specification
Measurement
Resolution/Precision
4mV MAX
±3.38E-05V
90dB MIN
±9.70E+00dB
75nA MAX
±9.24E-09A
Input Offset Voltage, VOS1, VOS2
Common Mode Rejection Ratio, CMRR1,
CMRR2
Input Offset Current, IOS1, IOS2
Input Bias Current, Non-Inverting Input,
+IB1, +IB2
Input Bias Current, Inverting Input, -IB1, -IB2
500nA MAX
±1.00E-08A
500nA MAX
±1.00E-08A
Large Signal Voltage Gain, AVOL1, AVOL2
10V/mV MIN
±2.06E+01V/mV
Large Signal Voltage Gain, AVOL1, AVOL2
80dB MIN
±2.29E+00dB
Saturation Voltage, VSAT1, VSAT2
0.4V MAX
±1.00E-03V
Saturation Voltage, VSAT3, VSAT4
1.5V MAX
±5.00E-03V
Output Leakage Current, ICEX1, ICEX2
2µA MAX
±3.24E-09A
Positive Supply Current, +IS
11.5mA MAX
±6.54E-05 A
Negative Supply Current, -IS
4.5mA MAX
±9.97E-06A
An ISO 9001:2000 Certified Company
57