ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Products contained in this shipment may be subject to ITAR regulations. Warning: The export of these commodity(ies), technology, or software are subject either to the U.S. Commerce Department Export Administration Regulations (E.A.R.), or to the U.S. State Department International Traffic In Arms Regulations (I.T.A.R.). Diversion, the shipment to unauthorized locations or entities, or the disclosure of related technical data or software to unauthorized foreign nationals is contrary to U.S. law and is prohibited. If export is authorized to a specific country or end-users, compliance with the U.S. export laws is required prior to transfer, transshipment on a non-continuous voyage, or disposal in any other country, or to any other end-user of these commodities, either in their original form or after being incorporated into other end-items. An ISO 9001:2008 and DLA Certified Company 1 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Enhanced Low Dose Rate Sensitivity (ELDRS) Radiation Testing of the RH119W High Performance Dual Comparator for Linear Technology Customer: Linear Technology, PO# 73783L RAD Job Number: 15-0663 Part Type Tested: RH119W High Performance Dual Comparator, Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C Traceability Information: Fab Lot Number: F0258.1, Lot Number: 601387.1, Wafer Number: 13, Date Code: 1049A. See photograph of unit under test in Appendix A. Quantity of Units: 11 units received, 5 units for biased irradiation, 5 units for unbiased irradiation and 1 unit for control. Serial numbers 1802, 1803, 1804, 1805 and 1806 were biased during irradiation, serial numbers 1807, 1808, 1809, 1841 and 1842 were unbiased during irradiation and serial number 1850 was used as control. See Appendix B for the radiation bias connection table. Radiation and Electrical Test Increments: 10mrad(Si)/s ionizing radiation with electrical test increments: pre-irradiation, 10krad(Si), 20krad(Si), 30krad(Si), 50krad(Si) and 100krad(Si). Pre-Irradiation Burn-In: Burn-In performed by Linear Technology prior to receipt by RAD Overtest and Post-Irradiation Anneal: No overtest. 24-hour room temperature anneal followed by a 168-hour 100°C anneal. Both anneals were performed in the same electrical bias condition as the irradiations. Electrical measurements were made following each anneal increment. Radiation Test Standard: MIL-STD-750 TM1019 and/or MIL-STD-883 TM1019 Condition D and Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C. Test Hardware and Software: LTS2020 Automated Tester, Entity ID TS04, Calibration Date: 5/28/2015, Calibration Due: 5/28/2016. LTS2101 Family Board, Entity ID FB10. LTS0608 Test Fixture, Entity ID TF08. RH119 DUT Board. Test Program: RH119AX.SRC Facility and Radiation Source: Aeroflex RAD's, Colorado Springs, CO. Gamma rays provided by Co60 (GB-150) low dose rate source. Dosimetry performed by Air Ionization Chamber (AIC) traceable to NIST. Aeroflex RAD's dosimetry has been audited by DLA and Aeroflex RAD has been awarded Laboratory Suitability for MIL-STD-750 and MIL-STD-883 TM 1019. Irradiation and Test Temperature: Room temperature controlled to 24°C±6°C per MIL-STD-883 and MIL-STD-750. Low Dose Rate Test Result: PASSED the enhanced low dose rate sensitivity test to the maximum tested dose level of 100krad(Si) with all parameters remaining within their datasheet specifications. Further the units do not exhibit ELDRS as defined in the current test method. An ISO 9001:2008 and DLA Certified Company 2 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 1.0. Overview and Background It is well known that total dose ionizing radiation can cause parametric degradation and ultimately functional failure in electronic devices. The damage occurs via electron-hole pair production, transport and trapping in the dielectric regions. In advanced CMOS technology nodes (0.6µm and smaller) the bulk of the damage is manifested in the thicker isolation regions, such as shallow trench or local oxidation of silicon (LOCOS) oxides (also known as "birds-beak" oxides). However, many linear and mixed signal devices that utilize bipolar minority carrier elements exhibit an enhanced low dose rate sensitivity (ELDRS). At this time there is no known or accepted a priori method for predicting susceptibility to ELDRS or simulating the low dose rate sensitivity with a "conventional" room temperature 50-300rad(Si)/s irradiation (Condition A in MIL-STD-883H TM 1019). Over the past 10 years a number of accelerating techniques have been examined, including an elevated temperature anneal, such as that used for MOS devices (see ASTM-F-1892 for more technical details) and irradiating at various temperatures. However, none of these techniques have proven useful across the wide variety of linear and/or mixed signal devices used in spaceborne applications. The latest requirement incorporated in MIL-STD-883H TM 1019 requires that devices that could potentially exhibit ELDRS "shall be tested either at the intended application dose rate, at a prescribed low dose rate to an overtest radiation level, or with an accelerated test such as an elevated temperature irradiation test that includes a parameter delta design margin". While the recently released MIL-STD883H TM 1019 allows for accelerated testing, the requirements for this are to essentially perform a low dose rate ELDRS test to verify the suitability of the acceleration method on the component of interest before the acceleration technique can be instituted. Based on the limitations of accelerated testing and to meet the requirements of MIL-STD-883H TM1019 Condition D, we have performed a low dose rate test at 10mrad(Si)/s. 2.0. Radiation Test Apparatus The low dose rate testing described in this final report was performed using the facilities at Aeroflex RAD's Longmire Laboratories in Colorado Springs, CO. The low dose rate source is a GB-150 irradiator modified to provide a panoramic exposure. The Co-60 rods are held in the base of the irradiator heavily shielded by lead. During the irradiation exposures the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for this irradiator in this configuration ranges from approximately 1mrad(Si)/s to a maximum of approximately 50rad(Si)/s, determined by the distance from the source. For low dose rate testing described in this report, the devices are placed approximately 2-meters from the Co-60 rods. The irradiator calibration is maintained by Aeroflex RAD's Longmire Laboratories using air ionization chamber (AIC) dosimetry traceable to the National Institute of Standards and Technology (NIST). Figure 2.1 shows a photograph of the GB-150 Co-60 irradiator at Aeroflex RAD's Longmire Laboratory facility. An ISO 9001:2008 and DLA Certified Company 3 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 2.1. Aeroflex RAD's Co-60 irradiator. The dose rate is obtained by positioning the device-under-test at a fixed distance from the gamma cell. The dose rate for this irradiator varies from approximately 50rad(Si)/s close to the rods down to <1mrad(Si)/s at a distance of approximately 4-meters. An ISO 9001:2008 and DLA Certified Company 4 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 3.0. Radiation Test Conditions The RH119W High Performance Dual Comparator described in this final report were irradiated using a single-sided supply potential of +/-15V and with all pins tied to ground, that is biased and unbiased. See Appendix B for details on the biasing conditions during radiation exposure. In our opinion, this bias circuit satisfies the requirements of MIL-STD-883H TM1019 Section 3.9.3 Bias and Loading Conditions which states "The bias applied to the test devices shall be selected to produce the greatest radiation induced damage or the worst-case damage for the intended application, if known. While maximum voltage is often worst case some bipolar linear device parameters (e.g. input bias current or maximum output load current) exhibit more degradation with 0 V bias." The devices were irradiated to a maximum total ionizing dose level of 100krad(Si) with incremental readings at 10krad(Si), 20krad(Si), 30krad(Si) and 50krad(Si). Electrical testing occurred within one hour following the end of each irradiation segment. For intermediate irradiations, the units were tested and returned to total dose exposure within two hours from the end of the previous radiation increment. The radiation exposure bias board was positioned in the Co-60 cell to provide the targeted dose rate of10mrad(Si)/s and was located inside a lead-aluminum enclosure. The lead-aluminum enclosure is required under MIL-STD-883H TM1019 Section 3.4 that reads as follows: "Lead/Aluminum (Pb/Al) container. Test specimens shall be enclosed in a Pb/Al container to minimize dose enhancement effects caused by low-energy, scattered radiation. A minimum of 1.5 mm Pb, surrounding an inner shield of at least 0.7 mm Al, is required. This Pb/Al container produces an approximate charged particle equilibrium for Si and for TLDs such as CaF2. The radiation field intensity shall be measured inside the Pb/Al container (1) initially, (2) when the source is changed, or (3) when the orientation or configuration of the source, container, or test-fixture is changed. This measurement shall be performed by placing a dosimeter (e.g., a TLD) in the device-irradiation container at the approximate test-device position. If it can be demonstrated that low energy scattered radiation is small enough that it will not cause dosimetry errors due to dose enhancement, the Pb/Al container may be omitted". The final dose rate within the lead-aluminum box was determined based on air ionization chamber (AIC) dosimetry measurements just prior to the beginning of the total dose irradiations. The final dose rate for this work was 10mrad(Si)/s with a precision of ±5%. An ISO 9001:2008 and DLA Certified Company 5 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 4.0. Tested Parameters During the enhanced low dose rate sensitivity testing the following electrical parameters were measured pre- and post-irradiation: 1. Positive Supply Current (A) @ VS=+/-15V, VCM=0V 2. Negative Supply Current (A) @ VS=+/-15V, VCM=0V 3. Offset Voltage (V) @ VS=+/-15V, VCM=0V 4. Input Offset Current (A) @ VS=+/-15V, VCM=0V 5. Positive Input Bias Current (A) @ VS=+/-15V, VCM=0V 6. Negative Input Bias Current (A) @ VS=+/-15V, VCM=0V 7. Large Signal Voltage Gain (V/mV) @ VS=+/-15V, VCM=0V 8. Common Mode Rejection Ratio (dB) @ VS=+/-15V, VCM=+/-10V 9. Saturation Voltage1 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA 10. Saturation Voltage2 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA 11. Output Leakage Current (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V Appendix C details the measured parameters, test conditions, pre-irradiation specification and measurement resolution for each of the measurements. The parametric data was obtained as "read and record" and all the raw data plus an attributes summary are contained in this report as well as in a separate Excel file. The attributes data contains the average, standard deviation and the average with the KTL values applied. The KTL value used in this work is 2.742 per MIL-HDBK-814 using one sided tolerance limits of 90/90 and a 5-piece sample size. The 90/90 KTL values were selected to match the statistical levels specified in the MIL-PRF-38535 sampling plan for the qualification of a radiation hardness assured (RHA) component. Note that the following criteria must be met for a device to pass the low dose rate test: following the radiation exposure each of the 5 pieces irradiated under electrical bias shall pass the specification value. The units irradiated without electrical bias and the KTL statistics are included in this report for reference only. If any of the 5 pieces irradiated under electrical bias exceed the device post radiation data sheet specification limits, then the lot could be logged as a failure. Further, MIL-STD-883H, TM 1019 Section 3.13.1.1 Characterization test to determine if a part exhibits ELDRS' states the following: Select a minimum random sample of 21 devices from a population representative of recent production runs. Smaller sample sizes may be used if agreed upon between the parties to the test. All of the selected devices shall have undergone appropriate elevated temperature reliability screens, e.g. burn-in and high temperature storage life. Divide the samples into four groups of 5 each and use the remaining part for a control. Perform pre-irradiation electrical characterization on all parts assuring that they meet the Group A electrical tests. Irradiate 5 samples under a 0 volt bias and another 5 under the irradiation bias given in the acquisition specification at 50-300 rad(Si)/s and room temperature. Irradiate 5 samples under a 0 volt bias and another 5 under irradiation bias given in the acquisition specification at < 10mrad(Si)/s and room temperature. Irradiate all samples to the same dose An ISO 9001:2008 and DLA Certified Company 6 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 levels, including 0.5 and 1.0 times the anticipated specification dose, and repeat the electrical characterization on each part at each dose level. Post irradiation electrical measurements shall be performed per paragraph 3.10 where the low dose rate test is considered Condition D. Calculate the radiation induced change in each electrical parameter (Δpara) for each sample at each radiation level. Calculate the ratio of the median Δpara at low dose rate to the median Δpara at high dose rate for each irradiation bias group at each total dose level. If this ratio exceeds 1.5 for any of the most sensitive parameters then the part is considered to be ELDRS susceptible. This test does not apply to parameters which exhibit changes that are within experimental error or whose values are below the pre-irradiation electrical specification limits at low dose rate at the specification dose. Therefore, the data in this report can be analyzed along with the high dose rate report titled "Total Ionizing Dose (TID) Radiation Testing of the RH119W High Performance Dual Comparator for Linear Technology" to demonstrate that these parts do not exhibit ELDRS as defined in the current test method. An ISO 9001:2008 and DLA Certified Company 7 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 5.0. ELDRS Test Results Based on this criterion the RH119W High Performance Dual Comparator (from the lot traceability information provided on the first page of this test report) PASSED the enhanced low dose rate sensitivity test to the maximum tested dose level of 100krad(Si) with all parameters remaining within their datasheet specifications. Figures 5.1 through 5.20 show plots of all the measured parameters versus total ionizing dose while Tables 5.1 - 5.20 show the corresponding raw data for each of these parameters. In the data plots the solid diamonds are the average of the measured data points for the sample irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the units irradiated with all pins tied to ground. The black lines (solid or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the sample irradiated in the biased condition while the shaded lines (solid or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the sample irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. The control units, as expected, show no significant changes to any of the parameters. Therefore we can conclude that the electrical testing remained in control throughout the duration of the tests and the observed degradation was due to the radiation exposure. Appendix D lists the figures used in this section to facilitate the location of a particular parameter. An ISO 9001:2008 and DLA Certified Company 8 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.1. Plot of Positive Supply Current (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 9 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.1. Raw data for Positive Supply Current (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Positive Supply Current (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 100 7.55E-03 7.21E-03 7.29E-03 7.71E-03 7.47E-03 7.62E-03 7.50E-03 7.44E-03 7.32E-03 7.36E-03 7.38E-03 24-hr Anneal 110 7.56E-03 7.21E-03 7.28E-03 7.73E-03 7.48E-03 7.62E-03 7.50E-03 7.46E-03 7.31E-03 7.36E-03 7.36E-03 168-hr Anneal 120 7.54E-03 7.24E-03 7.31E-03 7.73E-03 7.48E-03 7.64E-03 7.51E-03 7.46E-03 7.31E-03 7.37E-03 7.37E-03 0 7.58E-03 7.26E-03 7.32E-03 7.76E-03 7.50E-03 7.63E-03 7.52E-03 7.47E-03 7.35E-03 7.39E-03 7.38E-03 10 7.56E-03 7.23E-03 7.30E-03 7.74E-03 7.49E-03 7.63E-03 7.51E-03 7.47E-03 7.34E-03 7.39E-03 7.38E-03 Total Dose (krad(Si)) 20 30 7.53E-03 7.51E-03 7.19E-03 7.19E-03 7.27E-03 7.26E-03 7.71E-03 7.70E-03 7.46E-03 7.45E-03 7.61E-03 7.62E-03 7.50E-03 7.50E-03 7.46E-03 7.45E-03 7.34E-03 7.33E-03 7.38E-03 7.38E-03 7.37E-03 7.38E-03 7.48E-03 2.01E-04 8.04E-03 6.93E-03 7.46E-03 2.05E-04 8.03E-03 6.90E-03 7.43E-03 2.07E-04 8.00E-03 6.86E-03 7.42E-03 2.04E-04 7.98E-03 6.86E-03 7.43E-03 2.02E-04 7.98E-03 6.87E-03 7.45E-03 2.00E-04 7.99E-03 6.90E-03 7.45E-03 2.09E-04 8.02E-03 6.88E-03 7.46E-03 1.95E-04 8.00E-03 6.92E-03 7.47E-03 1.13E-04 7.78E-03 7.16E-03 1.15E-02 PASS 7.47E-03 1.15E-04 7.78E-03 7.15E-03 1.15E-02 PASS 7.46E-03 1.07E-04 7.75E-03 7.17E-03 1.15E-02 PASS 7.45E-03 1.12E-04 7.76E-03 7.15E-03 1.15E-02 PASS 7.45E-03 1.16E-04 7.77E-03 7.13E-03 1.15E-02 PASS 7.45E-03 1.17E-04 7.77E-03 7.13E-03 1.15E-02 PASS 7.45E-03 1.20E-04 7.78E-03 7.12E-03 1.15E-02 PASS 7.46E-03 1.27E-04 7.81E-03 7.11E-03 1.15E-02 PASS 50 7.53E-03 7.20E-03 7.27E-03 7.71E-03 7.45E-03 7.61E-03 7.50E-03 7.45E-03 7.32E-03 7.37E-03 7.37E-03 An ISO 9001:2008 and DLA Certified Company 10 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.2. Plot of Negative Supply Current (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 11 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.2. Raw data for Negative Supply Current (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Negative Supply Current (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status 100 -3.27E-03 -3.14E-03 -3.16E-03 -3.36E-03 -3.25E-03 -3.35E-03 -3.27E-03 -3.25E-03 -3.20E-03 -3.21E-03 -3.21E-03 24-hr Anneal 110 -3.27E-03 -3.14E-03 -3.16E-03 -3.36E-03 -3.25E-03 -3.35E-03 -3.27E-03 -3.26E-03 -3.20E-03 -3.21E-03 -3.21E-03 168-hr Anneal 120 -3.28E-03 -3.14E-03 -3.17E-03 -3.36E-03 -3.25E-03 -3.35E-03 -3.27E-03 -3.25E-03 -3.19E-03 -3.21E-03 -3.21E-03 0 -3.29E-03 -3.16E-03 -3.18E-03 -3.37E-03 -3.26E-03 -3.36E-03 -3.27E-03 -3.26E-03 -3.20E-03 -3.21E-03 -3.21E-03 10 -3.28E-03 -3.15E-03 -3.18E-03 -3.37E-03 -3.26E-03 -3.35E-03 -3.27E-03 -3.25E-03 -3.20E-03 -3.21E-03 -3.21E-03 Total Dose (krad(Si)) 20 30 -3.28E-03 -3.28E-03 -3.14E-03 -3.14E-03 -3.17E-03 -3.17E-03 -3.36E-03 -3.36E-03 -3.25E-03 -3.25E-03 -3.35E-03 -3.35E-03 -3.27E-03 -3.27E-03 -3.26E-03 -3.25E-03 -3.20E-03 -3.20E-03 -3.21E-03 -3.21E-03 -3.21E-03 -3.21E-03 -3.25E-03 8.70E-05 -3.01E-03 -3.49E-03 -3.25E-03 8.89E-05 -3.00E-03 -3.49E-03 -3.24E-03 8.86E-05 -3.00E-03 -3.48E-03 -3.24E-03 8.96E-05 -2.99E-03 -3.49E-03 -3.24E-03 8.84E-05 -3.00E-03 -3.48E-03 -3.24E-03 8.81E-05 -2.99E-03 -3.48E-03 -3.24E-03 8.87E-05 -2.99E-03 -3.48E-03 -3.24E-03 8.75E-05 -3.00E-03 -3.48E-03 -3.26E-03 6.30E-05 -3.08E-03 -3.43E-03 -4.50E-03 PASS -3.26E-03 6.19E-05 -3.09E-03 -3.43E-03 -4.50E-03 PASS -3.26E-03 6.16E-05 -3.09E-03 -3.43E-03 -4.50E-03 PASS -3.26E-03 6.20E-05 -3.09E-03 -3.43E-03 -4.50E-03 PASS -3.26E-03 6.23E-05 -3.08E-03 -3.43E-03 -4.50E-03 PASS -3.26E-03 6.13E-05 -3.09E-03 -3.42E-03 -4.50E-03 PASS -3.26E-03 6.21E-05 -3.08E-03 -3.43E-03 -4.50E-03 PASS -3.25E-03 6.30E-05 -3.08E-03 -3.43E-03 -4.50E-03 PASS 50 -3.28E-03 -3.14E-03 -3.17E-03 -3.36E-03 -3.25E-03 -3.35E-03 -3.27E-03 -3.25E-03 -3.20E-03 -3.21E-03 -3.21E-03 An ISO 9001:2008 and DLA Certified Company 12 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.3. Plot of Offset Voltage_1 (V) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 13 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.3. Raw data for Offset Voltage_1 (V) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Offset Voltage_1 (V) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 0 2.74E-04 1.92E-04 2.11E-04 2.03E-04 6.30E-05 -1.50E-05 -7.00E-05 -1.12E-04 3.39E-04 1.98E-04 3.67E-04 Biased Statistics Average Biased 1.89E-04 Std Dev Biased 7.71E-05 Ps90%/90% (+KTL) Biased 4.00E-04 Ps90%/90% (-KTL) Biased -2.29E-05 Un-Biased Statistics Average Un-Biased 6.80E-05 Std Dev Un-Biased 1.93E-04 Ps90%/90% (+KTL) Un-Biased 5.97E-04 Ps90%/90% (-KTL) Un-Biased -4.61E-04 Specification MIN -4.00E-03 Status PASS Specification MAX 4.00E-03 Status PASS 100 3.03E-04 2.03E-04 2.37E-04 2.12E-04 9.20E-05 -5.30E-05 -1.02E-04 -1.50E-04 3.05E-04 1.67E-04 3.66E-04 24-hr Anneal 110 3.03E-04 2.03E-04 2.40E-04 2.11E-04 9.20E-05 -5.40E-05 -1.04E-04 -1.51E-04 3.00E-04 1.65E-04 3.67E-04 168-hr Anneal 120 2.85E-04 1.81E-04 2.12E-04 1.88E-04 7.70E-05 -4.40E-05 -9.90E-05 -1.53E-04 3.04E-04 1.74E-04 3.68E-04 2.09E-04 7.70E-05 4.20E-04 -2.27E-06 2.09E-04 7.64E-05 4.19E-04 -1.13E-07 2.10E-04 7.67E-05 4.20E-04 -4.72E-07 1.89E-04 7.48E-05 3.94E-04 -1.64E-05 4.72E-05 1.91E-04 5.72E-04 -4.78E-04 -4.00E-03 PASS 4.00E-03 PASS 3.34E-05 1.94E-04 5.66E-04 -4.99E-04 -4.00E-03 PASS 4.00E-03 PASS 3.12E-05 1.93E-04 5.60E-04 -4.98E-04 -4.00E-03 PASS 4.00E-03 PASS 3.64E-05 1.94E-04 5.70E-04 -4.97E-04 -4.00E-03 PASS 4.00E-03 PASS 10 2.92E-04 1.98E-04 2.28E-04 2.09E-04 8.00E-05 -1.90E-05 -7.00E-05 -1.17E-04 3.33E-04 1.92E-04 3.65E-04 Total Dose (krad(Si)) 20 30 3.03E-04 3.04E-04 2.02E-04 2.06E-04 2.34E-04 2.36E-04 2.14E-04 2.15E-04 8.50E-05 8.50E-05 -2.50E-05 -3.00E-05 -7.60E-05 -8.10E-05 -1.22E-04 -1.27E-04 3.27E-04 3.22E-04 1.88E-04 1.84E-04 3.67E-04 3.66E-04 50 3.01E-04 2.04E-04 2.38E-04 2.13E-04 8.90E-05 -3.40E-05 -8.70E-05 -1.35E-04 3.16E-04 1.76E-04 3.64E-04 2.01E-04 7.70E-05 4.13E-04 -9.77E-06 2.08E-04 7.89E-05 4.24E-04 -8.75E-06 2.09E-04 7.93E-05 4.27E-04 -8.33E-06 6.38E-05 1.91E-04 5.88E-04 -4.61E-04 -4.00E-03 PASS 4.00E-03 PASS 5.84E-05 1.91E-04 5.83E-04 -4.66E-04 -4.00E-03 PASS 4.00E-03 PASS 5.36E-05 1.92E-04 5.79E-04 -4.72E-04 -4.00E-03 PASS 4.00E-03 PASS An ISO 9001:2008 and DLA Certified Company 14 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.4. Plot of Offset Voltage_2 (V) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 15 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.4. Raw data for Offset Voltage_2 (V) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Offset Voltage_2 (V) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 0 -1.67E-04 1.05E-04 3.98E-04 1.39E-04 5.67E-04 -2.80E-05 -1.60E-05 4.40E-05 3.51E-04 -1.67E-04 1.59E-04 Biased Statistics Average Biased 2.08E-04 Std Dev Biased 2.83E-04 Ps90%/90% (+KTL) Biased 9.85E-04 Ps90%/90% (-KTL) Biased -5.68E-04 Un-Biased Statistics Average Un-Biased 3.68E-05 Std Dev Un-Biased 1.92E-04 Ps90%/90% (+KTL) Un-Biased 5.63E-04 Ps90%/90% (-KTL) Un-Biased -4.89E-04 Specification MIN -4.00E-03 Status PASS Specification MAX 4.00E-03 Status PASS 100 -1.07E-04 1.47E-04 4.61E-04 1.83E-04 6.19E-04 -6.00E-05 -5.10E-05 1.10E-05 3.20E-04 -1.94E-04 1.59E-04 24-hr Anneal 110 -1.07E-04 1.46E-04 4.58E-04 1.80E-04 6.19E-04 -6.40E-05 -5.40E-05 6.00E-06 3.20E-04 -1.95E-04 1.58E-04 168-hr Anneal 120 -1.31E-04 1.23E-04 4.25E-04 1.69E-04 6.01E-04 -5.50E-05 -4.50E-05 1.80E-05 3.20E-04 -1.94E-04 1.56E-04 2.52E-04 2.88E-04 1.04E-03 -5.38E-04 2.61E-04 2.84E-04 1.04E-03 -5.18E-04 2.59E-04 2.84E-04 1.04E-03 -5.19E-04 2.37E-04 2.83E-04 1.01E-03 -5.39E-04 1.82E-05 1.92E-04 5.46E-04 -5.09E-04 -4.00E-03 PASS 4.00E-03 PASS 5.20E-06 1.91E-04 5.29E-04 -5.19E-04 -4.00E-03 PASS 4.00E-03 PASS 2.60E-06 1.92E-04 5.29E-04 -5.24E-04 -4.00E-03 PASS 4.00E-03 PASS 8.80E-06 1.90E-04 5.31E-04 -5.13E-04 -4.00E-03 PASS 4.00E-03 PASS 10 -1.40E-04 1.27E-04 4.29E-04 1.59E-04 5.93E-04 -3.20E-05 -2.10E-05 4.40E-05 3.50E-04 -1.68E-04 1.58E-04 Total Dose (krad(Si)) 20 30 -1.32E-04 -1.26E-04 1.35E-04 1.37E-04 4.44E-04 4.48E-04 1.69E-04 1.70E-04 6.05E-04 6.09E-04 -3.70E-05 -4.10E-05 -2.50E-05 -3.10E-05 3.80E-05 3.30E-05 3.45E-04 3.40E-04 -1.72E-04 -1.75E-04 1.57E-04 1.58E-04 50 -1.24E-04 1.40E-04 4.53E-04 1.76E-04 6.15E-04 -4.60E-05 -3.60E-05 2.40E-05 3.34E-04 -1.85E-04 1.58E-04 2.34E-04 2.85E-04 1.01E-03 -5.47E-04 2.44E-04 2.87E-04 1.03E-03 -5.42E-04 2.48E-04 2.87E-04 1.03E-03 -5.38E-04 3.46E-05 1.92E-04 5.62E-04 -4.93E-04 -4.00E-03 PASS 4.00E-03 PASS 2.98E-05 1.92E-04 5.57E-04 -4.97E-04 -4.00E-03 PASS 4.00E-03 PASS 2.52E-05 1.92E-04 5.50E-04 -5.00E-04 -4.00E-03 PASS 4.00E-03 PASS An ISO 9001:2008 and DLA Certified Company 16 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.5. Plot of Input Offset Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 17 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.5. Raw data for Input Offset Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Input Offset Current_1 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -3.51E-08 -3.44E-08 -3.64E-08 -3.23E-08 -2.73E-08 -3.53E-09 1.34E-08 7.27E-09 -8.30E-10 -1.57E-09 -8.58E-09 24-hr Anneal 110 -3.51E-08 -3.48E-08 -3.60E-08 -3.24E-08 -2.73E-08 -1.20E-09 1.49E-08 9.20E-09 1.52E-09 -3.00E-10 -8.44E-09 168-hr Anneal 120 -2.10E-08 -1.70E-08 -1.99E-08 -1.81E-08 -1.48E-08 -1.00E-11 8.03E-09 5.83E-09 -1.85E-09 -1.21E-09 -8.70E-09 0 -7.11E-09 -2.34E-09 -5.08E-09 -2.69E-09 -2.55E-09 -3.14E-09 1.04E-09 -1.63E-09 -7.32E-09 -3.79E-09 -8.58E-09 10 -1.44E-08 -1.26E-08 -1.27E-08 -1.19E-08 -1.07E-08 -1.45E-09 3.07E-09 -3.10E-10 -4.80E-09 -2.91E-09 -8.59E-09 Total Dose (krad(Si)) 20 30 -1.86E-08 -2.17E-08 -1.47E-08 -1.81E-08 -1.76E-08 -2.02E-08 -1.66E-08 -1.90E-08 -1.49E-08 -1.65E-08 -1.53E-09 -2.24E-09 4.78E-09 5.48E-09 1.48E-09 2.56E-09 -3.11E-09 -2.92E-09 -1.99E-09 -2.29E-09 -8.51E-09 -8.80E-09 -3.95E-09 2.09E-09 1.77E-09 -9.67E-09 -1.25E-08 1.38E-09 -8.71E-09 -1.63E-08 -1.65E-08 1.69E-09 -1.19E-08 -2.11E-08 -1.91E-08 2.00E-09 -1.36E-08 -2.46E-08 -2.29E-08 1.81E-09 -1.79E-08 -2.78E-08 -3.31E-08 3.57E-09 -2.33E-08 -4.29E-08 -3.31E-08 3.49E-09 -2.35E-08 -4.27E-08 -1.81E-08 2.45E-09 -1.14E-08 -2.49E-08 -2.97E-09 3.06E-09 5.43E-09 -1.14E-08 -7.50E-08 PASS 7.50E-08 PASS -1.28E-09 2.96E-09 6.82E-09 -9.38E-09 -7.50E-08 PASS 7.50E-08 PASS -7.40E-11 3.20E-09 8.70E-09 -8.85E-09 -1.00E-07 PASS 1.00E-07 PASS 1.18E-10 3.72E-09 1.03E-08 -1.01E-08 -1.00E-07 PASS 1.00E-07 PASS 6.74E-10 4.36E-09 1.26E-08 -1.13E-08 -1.50E-07 PASS 1.50E-07 PASS 2.96E-09 7.17E-09 2.26E-08 -1.67E-08 -3.00E-07 PASS 3.00E-07 PASS 4.82E-09 6.96E-09 2.39E-08 -1.43E-08 -3.00E-07 PASS 3.00E-07 PASS 2.16E-09 4.47E-09 1.44E-08 -1.01E-08 -3.00E-07 PASS 3.00E-07 PASS 50 -2.33E-08 -2.19E-08 -2.51E-08 -2.37E-08 -2.03E-08 -1.57E-09 7.70E-09 2.10E-09 -1.97E-09 -2.89E-09 -8.66E-09 An ISO 9001:2008 and DLA Certified Company 18 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.6. Plot of Input Offset Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 19 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.6. Raw data for Input Offset Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Input Offset Current_2 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -3.40E-08 -5.38E-08 -4.84E-08 -3.38E-08 -4.27E-08 -6.94E-09 -1.44E-08 -1.45E-08 -1.60E-08 -9.21E-09 -3.56E-09 24-hr Anneal 110 -3.38E-08 -5.40E-08 -4.85E-08 -3.37E-08 -4.23E-08 -3.42E-09 -1.01E-08 -1.21E-08 -1.43E-08 -7.18E-09 -3.62E-09 168-hr Anneal 120 -1.46E-08 -2.48E-08 -2.32E-08 -1.58E-08 -2.44E-08 3.42E-09 6.50E-10 -1.16E-09 -3.79E-09 2.41E-09 -3.61E-09 0 -2.34E-09 -1.17E-08 -8.05E-09 -3.72E-09 -8.81E-09 5.10E-10 4.60E-10 -1.49E-09 -4.96E-09 -5.40E-10 -2.62E-09 10 -1.26E-08 -2.25E-08 -1.82E-08 -1.34E-08 -2.11E-08 1.27E-09 -2.30E-10 -1.25E-09 -3.89E-09 1.06E-09 -2.33E-09 Total Dose (krad(Si)) 20 30 -1.76E-08 -1.96E-08 -2.94E-08 -3.41E-08 -2.46E-08 -2.82E-08 -1.91E-08 -2.12E-08 -2.76E-08 -2.93E-08 3.30E-10 -4.30E-10 -1.13E-09 -1.70E-09 -3.55E-09 -4.50E-09 -5.99E-09 -5.80E-09 1.20E-10 -1.00E-09 -3.48E-09 -3.60E-09 -6.92E-09 3.84E-09 3.60E-09 -1.74E-08 -1.76E-08 4.46E-09 -5.32E-09 -2.98E-08 -2.37E-08 5.15E-09 -9.55E-09 -3.78E-08 -2.64E-08 6.01E-09 -9.95E-09 -4.29E-08 -3.13E-08 7.22E-09 -1.15E-08 -5.10E-08 -4.25E-08 8.83E-09 -1.83E-08 -6.67E-08 -4.25E-08 8.96E-09 -1.79E-08 -6.70E-08 -2.05E-08 4.94E-09 -7.01E-09 -3.41E-08 -1.20E-09 2.25E-09 4.98E-09 -7.39E-09 -7.50E-08 PASS 7.50E-08 PASS -6.08E-10 2.10E-09 5.15E-09 -6.37E-09 -7.50E-08 PASS 7.50E-08 PASS -2.04E-09 2.69E-09 5.34E-09 -9.43E-09 -1.00E-07 PASS 1.00E-07 PASS -2.69E-09 2.34E-09 3.73E-09 -9.10E-09 -1.00E-07 PASS 1.00E-07 PASS -6.29E-09 3.74E-09 3.95E-09 -1.65E-08 -1.50E-07 PASS 1.50E-07 PASS -1.22E-08 3.91E-09 -1.48E-09 -2.29E-08 -3.00E-07 PASS 3.00E-07 PASS -9.44E-09 4.27E-09 2.26E-09 -2.11E-08 -3.00E-07 PASS 3.00E-07 PASS 3.06E-10 2.88E-09 8.20E-09 -7.59E-09 -3.00E-07 PASS 3.00E-07 PASS 50 -2.21E-08 -3.90E-08 -3.48E-08 -2.52E-08 -3.51E-08 -1.24E-09 -4.08E-09 -9.33E-09 -1.03E-08 -6.51E-09 -3.63E-09 An ISO 9001:2008 and DLA Certified Company 20 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.7. Plot of Positive Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 21 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.7. Raw data for Positive Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Positive Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -4.43E-07 -4.78E-07 -4.66E-07 -4.58E-07 -4.28E-07 -5.37E-07 -5.41E-07 -5.59E-07 -5.19E-07 -4.76E-07 -3.21E-07 24-hr Anneal 110 -4.43E-07 -4.78E-07 -4.64E-07 -4.59E-07 -4.28E-07 -5.31E-07 -5.35E-07 -5.55E-07 -5.13E-07 -4.72E-07 -3.20E-07 168-hr Anneal 120 -3.82E-07 -4.07E-07 -3.99E-07 -4.03E-07 -3.77E-07 -4.23E-07 -4.33E-07 -4.40E-07 -4.10E-07 -3.88E-07 -3.21E-07 0 -3.12E-07 -3.33E-07 -3.29E-07 -3.40E-07 -3.15E-07 -3.40E-07 -3.50E-07 -3.55E-07 -3.29E-07 -3.19E-07 -3.22E-07 10 -3.44E-07 -3.62E-07 -3.60E-07 -3.67E-07 -3.47E-07 -3.65E-07 -3.75E-07 -3.81E-07 -3.54E-07 -3.44E-07 -3.21E-07 Total Dose (krad(Si)) 20 30 -3.67E-07 -3.79E-07 -3.88E-07 -4.05E-07 -3.83E-07 -3.96E-07 -3.87E-07 -3.97E-07 -3.71E-07 -3.78E-07 -3.87E-07 -4.09E-07 -3.95E-07 -4.15E-07 -4.02E-07 -4.22E-07 -3.76E-07 -3.94E-07 -3.66E-07 -3.79E-07 -3.21E-07 -3.22E-07 -3.26E-07 1.20E-08 -2.93E-07 -3.59E-07 -3.56E-07 1.01E-08 -3.28E-07 -3.84E-07 -3.79E-07 9.73E-09 -3.52E-07 -4.06E-07 -3.91E-07 1.19E-08 -3.58E-07 -4.24E-07 -4.10E-07 1.48E-08 -3.69E-07 -4.51E-07 -4.54E-07 1.94E-08 -4.01E-07 -5.08E-07 -4.54E-07 1.93E-08 -4.01E-07 -5.07E-07 -3.93E-07 1.34E-08 -3.57E-07 -4.30E-07 -3.39E-07 1.50E-08 -2.97E-07 -3.80E-07 -5.00E-07 PASS 5.00E-07 PASS -3.64E-07 1.48E-08 -3.23E-07 -4.04E-07 -5.00E-07 PASS 5.00E-07 PASS -3.85E-07 1.46E-08 -3.45E-07 -4.25E-07 -7.50E-07 PASS 7.50E-07 PASS -4.04E-07 1.73E-08 -3.56E-07 -4.51E-07 -7.50E-07 PASS 7.50E-07 PASS -4.41E-07 1.68E-08 -3.95E-07 -4.87E-07 -1.00E-06 PASS 1.00E-06 PASS -5.26E-07 3.16E-08 -4.40E-07 -6.13E-07 -1.50E-06 PASS 1.50E-06 PASS -5.21E-07 3.12E-08 -4.36E-07 -6.07E-07 -1.50E-06 PASS 1.50E-06 PASS -4.19E-07 2.06E-08 -3.62E-07 -4.75E-07 -1.50E-06 PASS 1.50E-06 PASS 50 -3.88E-07 -4.23E-07 -4.19E-07 -4.18E-07 -4.02E-07 -4.28E-07 -4.45E-07 -4.68E-07 -4.38E-07 -4.27E-07 -3.20E-07 An ISO 9001:2008 and DLA Certified Company 22 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.8. Plot of Positive Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 23 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.8. Raw data for Positive Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Positive Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -4.60E-07 -5.06E-07 -4.91E-07 -4.75E-07 -4.52E-07 -5.70E-07 -5.78E-07 -5.97E-07 -5.42E-07 -4.97E-07 -3.18E-07 24-hr Anneal 110 -4.61E-07 -5.06E-07 -4.90E-07 -4.76E-07 -4.52E-07 -5.62E-07 -5.70E-07 -5.91E-07 -5.36E-07 -4.93E-07 -3.17E-07 168-hr Anneal 120 -3.88E-07 -4.21E-07 -4.11E-07 -4.12E-07 -3.92E-07 -4.39E-07 -4.47E-07 -4.58E-07 -4.17E-07 -3.96E-07 -3.18E-07 0 -3.10E-07 -3.32E-07 -3.29E-07 -3.36E-07 -3.19E-07 -3.38E-07 -3.50E-07 -3.56E-07 -3.28E-07 -3.16E-07 -3.19E-07 10 -3.48E-07 -3.71E-07 -3.67E-07 -3.71E-07 -3.58E-07 -3.70E-07 -3.79E-07 -3.86E-07 -3.56E-07 -3.45E-07 -3.18E-07 Total Dose (krad(Si)) 20 30 -3.74E-07 -3.88E-07 -4.03E-07 -4.22E-07 -3.95E-07 -4.10E-07 -3.95E-07 -4.07E-07 -3.87E-07 -3.95E-07 -3.95E-07 -4.19E-07 -4.02E-07 -4.26E-07 -4.12E-07 -4.34E-07 -3.83E-07 -4.01E-07 -3.70E-07 -3.84E-07 -3.18E-07 -3.19E-07 -3.25E-07 1.07E-08 -2.96E-07 -3.55E-07 -3.63E-07 9.94E-09 -3.36E-07 -3.90E-07 -3.91E-07 1.08E-08 -3.61E-07 -4.20E-07 -4.04E-07 1.33E-08 -3.68E-07 -4.41E-07 -4.27E-07 1.79E-08 -3.77E-07 -4.76E-07 -4.77E-07 2.20E-08 -4.16E-07 -5.37E-07 -4.77E-07 2.16E-08 -4.18E-07 -5.36E-07 -4.05E-07 1.41E-08 -3.66E-07 -4.43E-07 -3.38E-07 1.63E-08 -2.93E-07 -3.82E-07 -5.00E-07 PASS 5.00E-07 PASS -3.67E-07 1.66E-08 -3.22E-07 -4.13E-07 -5.00E-07 PASS 5.00E-07 PASS -3.92E-07 1.65E-08 -3.47E-07 -4.38E-07 -7.50E-07 PASS 7.50E-07 PASS -4.13E-07 2.01E-08 -3.58E-07 -4.68E-07 -7.50E-07 PASS 7.50E-07 PASS -4.57E-07 1.93E-08 -4.04E-07 -5.10E-07 -1.00E-06 PASS 1.00E-06 PASS -5.57E-07 3.87E-08 -4.51E-07 -6.63E-07 -1.50E-06 PASS 1.50E-06 PASS -5.50E-07 3.74E-08 -4.48E-07 -6.53E-07 -1.50E-06 PASS 1.50E-06 PASS -4.32E-07 2.48E-08 -3.64E-07 -5.00E-07 -1.50E-06 PASS 1.50E-06 PASS 50 -3.98E-07 -4.44E-07 -4.37E-07 -4.31E-07 -4.23E-07 -4.42E-07 -4.63E-07 -4.88E-07 -4.52E-07 -4.40E-07 -3.17E-07 An ISO 9001:2008 and DLA Certified Company 24 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.9. Plot of Negative Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 25 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.9. Raw data for Negative Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Negative Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -4.09E-07 -4.44E-07 -4.30E-07 -4.26E-07 -4.02E-07 -5.37E-07 -5.58E-07 -5.70E-07 -5.21E-07 -4.77E-07 -3.13E-07 24-hr Anneal 110 -4.09E-07 -4.44E-07 -4.29E-07 -4.27E-07 -4.02E-07 -5.32E-07 -5.52E-07 -5.67E-07 -5.16E-07 -4.73E-07 -3.12E-07 168-hr Anneal 120 -3.61E-07 -3.91E-07 -3.79E-07 -3.85E-07 -3.62E-07 -4.24E-07 -4.41E-07 -4.47E-07 -4.09E-07 -3.88E-07 -3.12E-07 0 -3.05E-07 -3.31E-07 -3.24E-07 -3.38E-07 -3.12E-07 -3.37E-07 -3.52E-07 -3.54E-07 -3.22E-07 -3.15E-07 -3.14E-07 10 -3.31E-07 -3.51E-07 -3.48E-07 -3.56E-07 -3.37E-07 -3.65E-07 -3.79E-07 -3.81E-07 -3.50E-07 -3.42E-07 -3.13E-07 Total Dose (krad(Si)) 20 30 -3.48E-07 -3.58E-07 -3.74E-07 -3.88E-07 -3.66E-07 -3.76E-07 -3.70E-07 -3.79E-07 -3.57E-07 -3.62E-07 -3.86E-07 -4.08E-07 -4.02E-07 -4.22E-07 -4.06E-07 -4.25E-07 -3.74E-07 -3.92E-07 -3.65E-07 -3.77E-07 -3.13E-07 -3.13E-07 -3.22E-07 1.33E-08 -2.86E-07 -3.58E-07 -3.44E-07 1.03E-08 -3.16E-07 -3.73E-07 -3.63E-07 1.04E-08 -3.34E-07 -3.91E-07 -3.72E-07 1.25E-08 -3.38E-07 -4.07E-07 -3.88E-07 1.47E-08 -3.48E-07 -4.29E-07 -4.22E-07 1.69E-08 -3.76E-07 -4.69E-07 -4.22E-07 1.67E-08 -3.76E-07 -4.68E-07 -3.75E-07 1.34E-08 -3.39E-07 -4.12E-07 -3.36E-07 1.75E-08 -2.88E-07 -3.84E-07 -5.00E-07 PASS 5.00E-07 PASS -3.63E-07 1.72E-08 -3.16E-07 -4.11E-07 -5.00E-07 PASS 5.00E-07 PASS -3.87E-07 1.73E-08 -3.39E-07 -4.34E-07 -7.50E-07 PASS 7.50E-07 PASS -4.05E-07 2.03E-08 -3.49E-07 -4.61E-07 -7.50E-07 PASS 7.50E-07 PASS -4.44E-07 1.97E-08 -3.90E-07 -4.98E-07 -1.00E-06 PASS 1.00E-06 PASS -5.32E-07 3.64E-08 -4.33E-07 -6.32E-07 -1.50E-06 PASS 1.50E-06 PASS -5.28E-07 3.62E-08 -4.29E-07 -6.27E-07 -1.50E-06 PASS 1.50E-06 PASS -4.22E-07 2.41E-08 -3.56E-07 -4.88E-07 -1.50E-06 PASS 1.50E-06 PASS 50 -3.65E-07 -4.03E-07 -3.95E-07 -3.96E-07 -3.83E-07 -4.28E-07 -4.55E-07 -4.72E-07 -4.38E-07 -4.26E-07 -3.12E-07 An ISO 9001:2008 and DLA Certified Company 26 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.10. Plot of Negative Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 27 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.10. Raw data for Negative Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Negative Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status Specification MAX Status 100 -4.27E-07 -4.53E-07 -4.43E-07 -4.42E-07 -4.10E-07 -5.67E-07 -5.68E-07 -5.85E-07 -5.29E-07 -4.92E-07 -3.15E-07 24-hr Anneal 110 -4.27E-07 -4.53E-07 -4.42E-07 -4.43E-07 -4.10E-07 -5.61E-07 -5.62E-07 -5.82E-07 -5.23E-07 -4.88E-07 -3.14E-07 168-hr Anneal 120 -3.74E-07 -3.97E-07 -3.88E-07 -3.97E-07 -3.68E-07 -4.43E-07 -4.49E-07 -4.57E-07 -4.15E-07 -3.99E-07 -3.14E-07 0 -3.08E-07 -3.20E-07 -3.22E-07 -3.33E-07 -3.10E-07 -3.39E-07 -3.51E-07 -3.54E-07 -3.22E-07 -3.15E-07 -3.16E-07 10 -3.35E-07 -3.48E-07 -3.50E-07 -3.57E-07 -3.37E-07 -3.71E-07 -3.80E-07 -3.85E-07 -3.53E-07 -3.47E-07 -3.15E-07 Total Dose (krad(Si)) 20 30 -3.57E-07 -3.69E-07 -3.74E-07 -3.89E-07 -3.71E-07 -3.83E-07 -3.76E-07 -3.87E-07 -3.60E-07 -3.66E-07 -3.96E-07 -4.21E-07 -4.03E-07 -4.25E-07 -4.09E-07 -4.31E-07 -3.77E-07 -3.97E-07 -3.71E-07 -3.85E-07 -3.15E-07 -3.15E-07 -3.18E-07 1.01E-08 -2.91E-07 -3.46E-07 -3.45E-07 9.01E-09 -3.21E-07 -3.70E-07 -3.67E-07 8.65E-09 -3.44E-07 -3.91E-07 -3.79E-07 1.04E-08 -3.50E-07 -4.07E-07 -3.96E-07 1.33E-08 -3.59E-07 -4.33E-07 -4.35E-07 1.66E-08 -3.89E-07 -4.80E-07 -4.35E-07 1.66E-08 -3.90E-07 -4.80E-07 -3.85E-07 1.33E-08 -3.48E-07 -4.21E-07 -3.36E-07 1.71E-08 -2.89E-07 -3.83E-07 -5.00E-07 PASS 5.00E-07 PASS -3.67E-07 1.65E-08 -3.22E-07 -4.12E-07 -5.00E-07 PASS 5.00E-07 PASS -3.91E-07 1.63E-08 -3.47E-07 -4.36E-07 -7.50E-07 PASS 7.50E-07 PASS -4.12E-07 1.99E-08 -3.57E-07 -4.66E-07 -7.50E-07 PASS 7.50E-07 PASS -4.52E-07 1.82E-08 -4.03E-07 -5.02E-07 -1.00E-06 PASS 1.00E-06 PASS -5.48E-07 3.78E-08 -4.45E-07 -6.52E-07 -1.50E-06 PASS 1.50E-06 PASS -5.43E-07 3.73E-08 -4.41E-07 -6.46E-07 -1.50E-06 PASS 1.50E-06 PASS -4.33E-07 2.46E-08 -3.65E-07 -5.00E-07 -1.50E-06 PASS 1.50E-06 PASS 50 -3.76E-07 -4.05E-07 -4.03E-07 -4.07E-07 -3.88E-07 -4.42E-07 -4.60E-07 -4.81E-07 -4.43E-07 -4.36E-07 -3.14E-07 An ISO 9001:2008 and DLA Certified Company 28 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.11. Plot of Large Signal Voltage Gain_1 (V/mV) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 29 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.11. Raw data for Large Signal Voltage Gain_1 (V/mV) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Large Signal Voltage Gain_1 (V/mV) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status 100 4.70E+01 4.84E+01 4.80E+01 4.68E+01 4.85E+01 4.85E+01 4.86E+01 4.90E+01 4.91E+01 4.86E+01 4.94E+01 24-hr Anneal 110 4.72E+01 4.84E+01 4.79E+01 4.70E+01 4.85E+01 4.86E+01 4.85E+01 4.92E+01 4.90E+01 4.86E+01 4.94E+01 168-hr Anneal 120 4.72E+01 4.86E+01 4.81E+01 4.70E+01 4.85E+01 4.90E+01 4.87E+01 4.92E+01 4.92E+01 4.88E+01 4.93E+01 0 4.80E+01 4.97E+01 4.89E+01 4.84E+01 4.95E+01 5.00E+01 4.93E+01 4.99E+01 5.01E+01 4.93E+01 4.93E+01 10 4.84E+01 4.95E+01 4.90E+01 4.81E+01 4.94E+01 4.98E+01 4.94E+01 5.01E+01 5.00E+01 4.96E+01 4.93E+01 Total Dose (krad(Si)) 20 30 4.78E+01 4.78E+01 4.91E+01 4.88E+01 4.86E+01 4.98E+01 4.78E+01 4.77E+01 4.91E+01 4.90E+01 4.97E+01 4.95E+01 4.92E+01 4.92E+01 5.00E+01 4.99E+01 5.01E+01 4.98E+01 4.94E+01 4.93E+01 4.95E+01 4.95E+01 4.89E+01 7.18E-01 5.09E+01 4.69E+01 4.89E+01 6.14E-01 5.06E+01 4.72E+01 4.85E+01 6.53E-01 5.03E+01 4.67E+01 4.86E+01 8.79E-01 5.10E+01 4.62E+01 4.82E+01 6.27E-01 5.00E+01 4.65E+01 4.77E+01 7.92E-01 4.99E+01 4.56E+01 4.78E+01 6.82E-01 4.97E+01 4.59E+01 4.79E+01 7.40E-01 4.99E+01 4.59E+01 4.97E+01 3.90E-01 5.08E+01 4.87E+01 1.00E+01 PASS 4.98E+01 2.86E-01 5.06E+01 4.90E+01 1.00E+01 PASS 4.97E+01 3.83E-01 5.07E+01 4.86E+01 1.00E+01 PASS 4.95E+01 3.05E-01 5.04E+01 4.87E+01 1.00E+01 PASS 4.93E+01 2.51E-01 4.99E+01 4.86E+01 1.00E+01 PASS 4.88E+01 2.70E-01 4.95E+01 4.80E+01 1.00E+01 PASS 4.88E+01 3.03E-01 4.96E+01 4.79E+01 1.00E+01 PASS 4.90E+01 2.28E-01 4.96E+01 4.84E+01 1.00E+01 PASS 50 4.77E+01 4.90E+01 4.84E+01 4.75E+01 4.86E+01 4.93E+01 4.90E+01 4.95E+01 4.95E+01 4.90E+01 4.92E+01 An ISO 9001:2008 and DLA Certified Company 30 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.12. Plot of Large Signal Voltage Gain_2 (V/mV) @ VS=+/-15V, VCM=0V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 31 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.12. Raw data for Large Signal Voltage Gain_2 (V/mV) @ VS=+/-15V, VCM=0V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Large Signal Voltage Gain_2 (V/mV) @ VS=+/-15V, VCM=0V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status 100 5.08E+01 5.18E+01 4.92E+01 4.89E+01 4.94E+01 5.27E+01 5.22E+01 5.03E+01 5.03E+01 5.17E+01 5.17E+01 24-hr Anneal 110 5.05E+01 5.18E+01 4.90E+01 4.89E+01 4.93E+01 5.27E+01 5.18E+01 5.04E+01 5.01E+01 5.17E+01 5.16E+01 168-hr Anneal 120 5.05E+01 5.22E+01 4.94E+01 4.92E+01 4.94E+01 5.34E+01 5.23E+01 5.08E+01 5.03E+01 5.22E+01 5.17E+01 0 5.17E+01 5.37E+01 5.07E+01 5.03E+01 5.07E+01 5.46E+01 5.31E+01 5.15E+01 5.16E+01 5.31E+01 5.17E+01 10 5.16E+01 5.31E+01 5.04E+01 5.03E+01 5.03E+01 5.43E+01 5.30E+01 5.14E+01 5.14E+01 5.29E+01 5.19E+01 Total Dose (krad(Si)) 20 30 5.13E+01 5.09E+01 5.26E+01 5.29E+01 4.99E+01 4.99E+01 4.98E+01 4.96E+01 5.00E+01 4.98E+01 5.39E+01 5.37E+01 5.30E+01 5.27E+01 5.14E+01 5.11E+01 5.13E+01 5.11E+01 5.28E+01 5.26E+01 5.16E+01 5.17E+01 5.14E+01 1.38E+00 5.52E+01 4.76E+01 5.11E+01 1.23E+00 5.45E+01 4.78E+01 5.07E+01 1.22E+00 5.41E+01 4.74E+01 5.06E+01 1.37E+00 5.44E+01 4.69E+01 5.04E+01 1.32E+00 5.40E+01 4.67E+01 5.00E+01 1.23E+00 5.34E+01 4.66E+01 4.99E+01 1.24E+00 5.33E+01 4.65E+01 5.01E+01 1.26E+00 5.36E+01 4.67E+01 5.28E+01 1.28E+00 5.63E+01 4.93E+01 1.00E+01 PASS 5.26E+01 1.23E+00 5.60E+01 4.92E+01 1.00E+01 PASS 5.25E+01 1.11E+00 5.55E+01 4.94E+01 1.00E+01 PASS 5.22E+01 1.13E+00 5.53E+01 4.92E+01 1.00E+01 PASS 5.20E+01 1.16E+00 5.52E+01 4.89E+01 1.00E+01 PASS 5.14E+01 1.10E+00 5.45E+01 4.84E+01 1.00E+01 PASS 5.13E+01 1.07E+00 5.43E+01 4.84E+01 1.00E+01 PASS 5.18E+01 1.25E+00 5.52E+01 4.84E+01 1.00E+01 PASS 50 5.10E+01 5.24E+01 4.96E+01 4.93E+01 4.95E+01 5.36E+01 5.26E+01 5.10E+01 5.08E+01 5.22E+01 5.18E+01 An ISO 9001:2008 and DLA Certified Company 32 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.13. Plot of Common Mode Rejection Ratio_1 (dB) @ VS=+/-15V, VCM=+/-10V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 33 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.13. Raw data for Common Mode Rejection Ratio_1 (dB) @ VS=+/-15V, VCM=+/-10V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Common Mode Rejection Ratio_1 (dB) @ VS=+/-15V, VCM=+/-10V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status 100 1.13E+02 1.53E+02 1.22E+02 1.43E+02 1.23E+02 1.22E+02 1.16E+02 1.26E+02 1.31E+02 1.14E+02 1.58E+02 24-hr Anneal 110 1.13E+02 1.59E+02 1.22E+02 1.42E+02 1.22E+02 1.22E+02 1.16E+02 1.26E+02 1.31E+02 1.14E+02 1.64E+02 168-hr Anneal 120 1.13E+02 1.50E+02 1.22E+02 1.43E+02 1.23E+02 1.22E+02 1.16E+02 1.25E+02 1.33E+02 1.14E+02 1.63E+02 0 1.13E+02 1.42E+02 1.22E+02 1.40E+02 1.22E+02 1.22E+02 1.16E+02 1.25E+02 1.32E+02 1.14E+02 1.55E+02 10 1.13E+02 1.48E+02 1.22E+02 1.42E+02 1.22E+02 1.22E+02 1.16E+02 1.25E+02 1.32E+02 1.14E+02 1.54E+02 Total Dose (krad(Si)) 20 30 1.13E+02 1.13E+02 1.48E+02 1.58E+02 1.22E+02 1.22E+02 1.41E+02 1.41E+02 1.22E+02 1.22E+02 1.21E+02 1.22E+02 1.16E+02 1.16E+02 1.25E+02 1.25E+02 1.32E+02 1.31E+02 1.14E+02 1.14E+02 1.67E+02 1.57E+02 1.28E+02 1.25E+01 1.62E+02 9.36E+01 1.29E+02 1.50E+01 1.70E+02 8.84E+01 1.29E+02 1.48E+01 1.70E+02 8.84E+01 1.31E+02 1.80E+01 1.80E+02 8.16E+01 1.30E+02 1.59E+01 1.73E+02 8.60E+01 1.30E+02 1.66E+01 1.76E+02 8.51E+01 1.32E+02 1.88E+01 1.83E+02 8.01E+01 1.30E+02 1.57E+01 1.73E+02 8.69E+01 1.22E+02 7.44E+00 1.42E+02 1.01E+02 9.00E+01 PASS 1.22E+02 7.22E+00 1.42E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.18E+00 1.41E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.06E+00 1.41E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.23E+00 1.42E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.05E+00 1.41E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.11E+00 1.41E+02 1.02E+02 9.00E+01 PASS 1.22E+02 7.58E+00 1.43E+02 1.01E+02 9.00E+01 PASS 50 1.13E+02 1.52E+02 1.22E+02 1.40E+02 1.22E+02 1.22E+02 1.16E+02 1.25E+02 1.32E+02 1.14E+02 1.52E+02 An ISO 9001:2008 and DLA Certified Company 34 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.14. Plot of Common Mode Rejection Ratio_2 (dB) @ VS=+/-15V, VCM=+/-10V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 35 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.14. Raw data for Common Mode Rejection Ratio_2 (dB) @ VS=+/-15V, VCM=+/-10V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Common Mode Rejection Ratio_2 (dB) @ VS=+/-15V, VCM=+/-10V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MIN Status 100 1.44E+02 1.18E+02 1.20E+02 1.11E+02 1.30E+02 1.14E+02 1.15E+02 1.12E+02 1.17E+02 1.29E+02 1.12E+02 24-hr Anneal 110 1.42E+02 1.18E+02 1.20E+02 1.11E+02 1.29E+02 1.14E+02 1.15E+02 1.12E+02 1.17E+02 1.29E+02 1.12E+02 168-hr Anneal 120 1.40E+02 1.18E+02 1.20E+02 1.11E+02 1.29E+02 1.14E+02 1.15E+02 1.12E+02 1.16E+02 1.28E+02 1.12E+02 0 1.40E+02 1.18E+02 1.20E+02 1.11E+02 1.29E+02 1.14E+02 1.15E+02 1.12E+02 1.17E+02 1.28E+02 1.12E+02 10 1.39E+02 1.18E+02 1.20E+02 1.11E+02 1.29E+02 1.14E+02 1.15E+02 1.12E+02 1.17E+02 1.28E+02 1.12E+02 Total Dose (krad(Si)) 20 30 1.41E+02 1.40E+02 1.18E+02 1.18E+02 1.20E+02 1.20E+02 1.11E+02 1.11E+02 1.29E+02 1.28E+02 1.14E+02 1.14E+02 1.15E+02 1.15E+02 1.12E+02 1.12E+02 1.17E+02 1.17E+02 1.28E+02 1.28E+02 1.12E+02 1.12E+02 1.24E+02 1.11E+01 1.54E+02 9.30E+01 1.23E+02 1.09E+01 1.53E+02 9.34E+01 1.24E+02 1.14E+01 1.55E+02 9.22E+01 1.23E+02 1.13E+01 1.54E+02 9.24E+01 1.23E+02 1.10E+01 1.53E+02 9.32E+01 1.25E+02 1.27E+01 1.60E+02 8.97E+01 1.24E+02 1.18E+01 1.56E+02 9.16E+01 1.24E+02 1.13E+01 1.55E+02 9.29E+01 1.17E+02 6.20E+00 1.34E+02 1.00E+02 9.00E+01 PASS 1.17E+02 6.21E+00 1.34E+02 1.00E+02 9.00E+01 PASS 1.17E+02 6.36E+00 1.35E+02 9.98E+01 9.00E+01 PASS 1.17E+02 6.37E+00 1.35E+02 9.97E+01 9.00E+01 PASS 1.17E+02 6.36E+00 1.35E+02 9.97E+01 9.00E+01 PASS 1.17E+02 6.56E+00 1.35E+02 9.93E+01 9.00E+01 PASS 1.17E+02 6.63E+00 1.36E+02 9.92E+01 9.00E+01 PASS 1.17E+02 6.31E+00 1.35E+02 9.99E+01 9.00E+01 PASS 50 1.40E+02 1.18E+02 1.20E+02 1.11E+02 1.29E+02 1.14E+02 1.15E+02 1.12E+02 1.17E+02 1.28E+02 1.12E+02 An ISO 9001:2008 and DLA Certified Company 36 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.15. Plot of Saturation Voltage1_1 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 37 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.15. Raw data for Saturation Voltage1_1 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Saturation Voltage1_1 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 100 3.04E-01 3.07E-01 3.10E-01 3.09E-01 3.08E-01 3.13E-01 3.11E-01 3.11E-01 3.13E-01 3.07E-01 3.07E-01 24-hr Anneal 110 3.04E-01 3.07E-01 3.10E-01 3.09E-01 3.08E-01 3.12E-01 3.11E-01 3.10E-01 3.13E-01 3.07E-01 3.06E-01 168-hr Anneal 120 3.06E-01 3.09E-01 3.11E-01 3.10E-01 3.10E-01 3.14E-01 3.13E-01 3.12E-01 3.14E-01 3.09E-01 3.09E-01 0 2.99E-01 3.00E-01 3.04E-01 3.03E-01 3.03E-01 3.08E-01 3.09E-01 3.07E-01 3.09E-01 3.05E-01 3.07E-01 10 2.98E-01 3.01E-01 3.04E-01 3.03E-01 3.03E-01 3.07E-01 3.06E-01 3.06E-01 3.08E-01 3.03E-01 3.07E-01 Total Dose (krad(Si)) 20 30 2.99E-01 3.00E-01 3.02E-01 3.02E-01 3.04E-01 3.05E-01 3.03E-01 3.05E-01 3.04E-01 3.04E-01 3.07E-01 3.08E-01 3.08E-01 3.08E-01 3.06E-01 3.07E-01 3.08E-01 3.09E-01 3.04E-01 3.05E-01 3.06E-01 3.07E-01 3.02E-01 2.17E-03 3.08E-01 2.96E-01 3.02E-01 2.39E-03 3.08E-01 2.95E-01 3.02E-01 2.07E-03 3.08E-01 2.97E-01 3.03E-01 2.17E-03 3.09E-01 2.97E-01 3.04E-01 2.68E-03 3.12E-01 2.97E-01 3.08E-01 2.30E-03 3.14E-01 3.01E-01 3.08E-01 2.30E-03 3.14E-01 3.01E-01 3.09E-01 1.92E-03 3.14E-01 3.04E-01 3.08E-01 1.67E-03 3.12E-01 3.03E-01 4.00E-01 PASS 3.06E-01 1.87E-03 3.11E-01 3.01E-01 4.00E-01 PASS 3.07E-01 1.67E-03 3.11E-01 3.02E-01 4.00E-01 PASS 3.07E-01 1.52E-03 3.12E-01 3.03E-01 4.00E-01 PASS 3.08E-01 1.92E-03 3.13E-01 3.03E-01 4.00E-01 PASS 3.11E-01 2.45E-03 3.18E-01 3.04E-01 4.00E-01 PASS 3.11E-01 2.30E-03 3.17E-01 3.04E-01 4.00E-01 PASS 3.12E-01 2.07E-03 3.18E-01 3.07E-01 4.00E-01 PASS 50 3.00E-01 3.03E-01 3.06E-01 3.06E-01 3.06E-01 3.09E-01 3.09E-01 3.08E-01 3.10E-01 3.05E-01 3.06E-01 An ISO 9001:2008 and DLA Certified Company 38 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.16. Plot of Saturation Voltage1_2 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 39 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.16. Raw data for Saturation Voltage1_2 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Saturation Voltage1_2 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 100 3.06E-01 3.07E-01 3.09E-01 3.11E-01 3.05E-01 3.10E-01 3.11E-01 3.11E-01 3.11E-01 3.05E-01 3.07E-01 24-hr Anneal 110 3.06E-01 3.07E-01 3.09E-01 3.11E-01 3.05E-01 3.10E-01 3.11E-01 3.11E-01 3.11E-01 3.05E-01 3.06E-01 168-hr Anneal 120 3.08E-01 3.10E-01 3.10E-01 3.13E-01 3.07E-01 3.11E-01 3.12E-01 3.12E-01 3.12E-01 3.06E-01 3.08E-01 0 3.02E-01 3.01E-01 3.04E-01 3.07E-01 3.01E-01 3.06E-01 3.08E-01 3.07E-01 3.08E-01 3.03E-01 3.06E-01 10 3.01E-01 3.01E-01 3.03E-01 3.05E-01 3.00E-01 3.05E-01 3.06E-01 3.05E-01 3.06E-01 3.01E-01 3.06E-01 Total Dose (krad(Si)) 20 30 3.02E-01 3.02E-01 3.02E-01 3.03E-01 3.04E-01 3.05E-01 3.06E-01 3.06E-01 3.01E-01 3.01E-01 3.05E-01 3.06E-01 3.06E-01 3.06E-01 3.05E-01 3.06E-01 3.06E-01 3.07E-01 3.01E-01 3.02E-01 3.06E-01 3.06E-01 3.03E-01 2.55E-03 3.10E-01 2.96E-01 3.02E-01 2.00E-03 3.07E-01 2.97E-01 3.03E-01 2.00E-03 3.08E-01 2.98E-01 3.03E-01 2.07E-03 3.09E-01 2.98E-01 3.04E-01 2.39E-03 3.11E-01 2.98E-01 3.08E-01 2.41E-03 3.14E-01 3.01E-01 3.08E-01 2.41E-03 3.14E-01 3.01E-01 3.10E-01 2.30E-03 3.16E-01 3.03E-01 3.06E-01 2.07E-03 3.12E-01 3.01E-01 4.00E-01 PASS 3.05E-01 2.07E-03 3.10E-01 2.99E-01 4.00E-01 PASS 3.05E-01 2.07E-03 3.10E-01 2.99E-01 4.00E-01 PASS 3.05E-01 1.95E-03 3.11E-01 3.00E-01 4.00E-01 PASS 3.06E-01 1.92E-03 3.11E-01 3.01E-01 4.00E-01 PASS 3.10E-01 2.61E-03 3.17E-01 3.02E-01 4.00E-01 PASS 3.10E-01 2.61E-03 3.17E-01 3.02E-01 4.00E-01 PASS 3.11E-01 2.61E-03 3.18E-01 3.03E-01 4.00E-01 PASS 50 3.02E-01 3.03E-01 3.05E-01 3.08E-01 3.03E-01 3.06E-01 3.07E-01 3.07E-01 3.08E-01 3.03E-01 3.06E-01 An ISO 9001:2008 and DLA Certified Company 40 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.17. Plot of Saturation Voltage2_1 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 41 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.17. Raw data for Saturation Voltage2_1 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Saturation Voltage2_1 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 100 6.18E-01 6.28E-01 6.29E-01 6.33E-01 6.23E-01 6.35E-01 6.32E-01 6.39E-01 6.39E-01 6.29E-01 6.21E-01 24-hr Anneal 110 6.18E-01 6.26E-01 6.31E-01 6.31E-01 6.21E-01 6.35E-01 6.32E-01 6.36E-01 6.37E-01 6.28E-01 6.22E-01 168-hr Anneal 120 6.30E-01 6.33E-01 6.37E-01 6.40E-01 6.31E-01 6.41E-01 6.40E-01 6.45E-01 6.46E-01 6.34E-01 6.30E-01 0 6.14E-01 6.18E-01 6.24E-01 6.25E-01 6.19E-01 6.29E-01 6.30E-01 6.33E-01 6.31E-01 6.23E-01 6.23E-01 10 6.14E-01 6.22E-01 6.24E-01 6.25E-01 6.18E-01 6.28E-01 6.26E-01 6.30E-01 6.30E-01 6.21E-01 6.21E-01 Total Dose (krad(Si)) 20 30 6.19E-01 6.18E-01 6.23E-01 6.24E-01 6.28E-01 6.29E-01 6.28E-01 6.29E-01 6.20E-01 6.22E-01 6.31E-01 6.31E-01 6.28E-01 6.29E-01 6.31E-01 6.33E-01 6.29E-01 6.32E-01 6.23E-01 6.23E-01 6.20E-01 6.21E-01 6.20E-01 4.53E-03 6.32E-01 6.08E-01 6.21E-01 4.56E-03 6.33E-01 6.08E-01 6.24E-01 4.28E-03 6.35E-01 6.12E-01 6.24E-01 4.72E-03 6.37E-01 6.11E-01 6.25E-01 4.32E-03 6.37E-01 6.13E-01 6.26E-01 5.81E-03 6.42E-01 6.10E-01 6.25E-01 5.86E-03 6.41E-01 6.09E-01 6.34E-01 4.21E-03 6.46E-01 6.23E-01 6.29E-01 3.77E-03 6.40E-01 6.19E-01 1.50E+00 PASS 6.27E-01 3.74E-03 6.37E-01 6.17E-01 1.50E+00 PASS 6.28E-01 3.29E-03 6.37E-01 6.19E-01 1.50E+00 PASS 6.30E-01 3.97E-03 6.40E-01 6.19E-01 1.50E+00 PASS 6.32E-01 3.94E-03 6.43E-01 6.21E-01 1.50E+00 PASS 6.35E-01 4.38E-03 6.47E-01 6.23E-01 1.50E+00 PASS 6.34E-01 3.65E-03 6.44E-01 6.24E-01 1.50E+00 PASS 6.41E-01 4.76E-03 6.54E-01 6.28E-01 1.50E+00 PASS 50 6.20E-01 6.25E-01 6.29E-01 6.30E-01 6.22E-01 6.32E-01 6.31E-01 6.35E-01 6.36E-01 6.26E-01 6.22E-01 An ISO 9001:2008 and DLA Certified Company 42 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.18. Plot of Saturation Voltage2_2 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 43 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.18. Raw data for Saturation Voltage2_2 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Saturation Voltage2_2 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 100 6.13E-01 6.17E-01 6.23E-01 6.29E-01 6.12E-01 6.26E-01 6.26E-01 6.27E-01 6.28E-01 6.18E-01 6.19E-01 24-hr Anneal 110 6.12E-01 6.17E-01 6.23E-01 6.28E-01 6.11E-01 6.25E-01 6.27E-01 6.26E-01 6.29E-01 6.18E-01 6.20E-01 168-hr Anneal 120 6.23E-01 6.24E-01 6.32E-01 6.38E-01 6.21E-01 6.33E-01 6.34E-01 6.34E-01 6.38E-01 6.27E-01 6.28E-01 0 6.11E-01 6.12E-01 6.19E-01 6.24E-01 6.10E-01 6.21E-01 6.24E-01 6.22E-01 6.24E-01 6.15E-01 6.21E-01 10 6.08E-01 6.11E-01 6.18E-01 6.22E-01 6.07E-01 6.19E-01 6.20E-01 6.18E-01 6.21E-01 6.12E-01 6.19E-01 Total Dose (krad(Si)) 20 30 6.12E-01 6.13E-01 6.14E-01 6.16E-01 6.21E-01 6.22E-01 6.26E-01 6.26E-01 6.09E-01 6.12E-01 6.21E-01 6.22E-01 6.21E-01 6.23E-01 6.20E-01 6.21E-01 6.21E-01 6.23E-01 6.13E-01 6.14E-01 6.20E-01 6.20E-01 6.15E-01 6.06E-03 6.32E-01 5.99E-01 6.13E-01 6.53E-03 6.31E-01 5.95E-01 6.16E-01 6.95E-03 6.35E-01 5.97E-01 6.18E-01 6.02E-03 6.34E-01 6.01E-01 6.18E-01 6.42E-03 6.36E-01 6.01E-01 6.19E-01 7.16E-03 6.38E-01 5.99E-01 6.18E-01 7.26E-03 6.38E-01 5.98E-01 6.28E-01 7.16E-03 6.47E-01 6.08E-01 6.21E-01 3.70E-03 6.31E-01 6.11E-01 1.50E+00 PASS 6.18E-01 3.54E-03 6.28E-01 6.08E-01 1.50E+00 PASS 6.19E-01 3.49E-03 6.29E-01 6.10E-01 1.50E+00 PASS 6.21E-01 3.78E-03 6.31E-01 6.10E-01 1.50E+00 PASS 6.23E-01 3.58E-03 6.32E-01 6.13E-01 1.50E+00 PASS 6.25E-01 4.00E-03 6.36E-01 6.14E-01 1.50E+00 PASS 6.25E-01 4.18E-03 6.36E-01 6.14E-01 1.50E+00 PASS 6.33E-01 3.96E-03 6.44E-01 6.22E-01 1.50E+00 PASS 50 6.13E-01 6.15E-01 6.23E-01 6.27E-01 6.13E-01 6.23E-01 6.23E-01 6.23E-01 6.27E-01 6.17E-01 6.21E-01 An ISO 9001:2008 and DLA Certified Company 44 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.19. Plot of Output Leakage Current_1 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 45 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.19. Raw data for Output Leakage Current_1 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Output Leakage Current_1 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 50 2.00E-08 1.70E-08 1.40E-08 1.70E-08 1.10E-08 1.20E-08 1.00E-08 9.00E-09 1.00E-08 9.00E-09 8.00E-09 100 2.30E-08 2.30E-08 1.70E-08 2.10E-08 1.50E-08 1.20E-08 1.00E-08 1.00E-08 9.00E-09 1.00E-08 8.00E-09 24-hr Anneal 110 2.00E-08 1.80E-08 1.40E-08 1.30E-08 9.00E-09 8.00E-09 8.00E-09 7.00E-09 8.00E-09 8.00E-09 6.00E-09 1.36E-08 2.97E-09 2.17E-08 5.47E-09 1.58E-08 3.42E-09 2.52E-08 6.42E-09 1.98E-08 3.63E-09 2.98E-08 9.84E-09 1.48E-08 4.32E-09 2.67E-08 2.94E-09 1.00E-08 1.87E-09 1.51E-08 4.87E-09 7.40E-09 8.94E-10 9.85E-09 4.95E-09 2.00E-06 PASS 1.00E-08 1.22E-09 1.34E-08 6.64E-09 2.00E-06 PASS 1.02E-08 1.10E-09 1.32E-08 7.20E-09 2.00E-06 PASS 7.80E-09 4.47E-10 9.03E-09 6.57E-09 2.00E-06 PASS 7.00E-09 7.07E-10 8.94E-09 5.06E-09 2.00E-06 PASS Total Dose (krad(Si)) 20 30 1.90E-08 1.80E-08 1.20E-08 1.40E-08 1.20E-08 1.20E-08 1.30E-08 1.40E-08 6.00E-09 1.00E-08 8.00E-09 8.00E-09 1.00E-08 8.00E-09 8.00E-09 8.00E-09 9.00E-09 7.00E-09 7.00E-09 6.00E-09 7.00E-09 6.00E-09 0 1.40E-08 1.10E-08 9.00E-09 9.00E-09 7.00E-09 9.00E-09 9.00E-09 1.00E-08 8.00E-09 7.00E-09 6.00E-09 10 2.00E-08 1.50E-08 1.40E-08 1.40E-08 9.00E-09 9.00E-09 1.00E-08 9.00E-09 9.00E-09 7.00E-09 7.00E-09 1.00E-08 2.65E-09 1.73E-08 2.75E-09 1.44E-08 3.91E-09 2.51E-08 3.67E-09 1.24E-08 4.62E-09 2.51E-08 -2.55E-10 8.60E-09 1.14E-09 1.17E-08 5.47E-09 2.00E-06 PASS 8.80E-09 1.10E-09 1.18E-08 5.80E-09 2.00E-06 PASS 8.40E-09 1.14E-09 1.15E-08 5.27E-09 2.00E-06 PASS An ISO 9001:2008 and DLA Certified Company 46 168-hr Anneal 120 1.20E-08 1.20E-08 9.00E-09 9.00E-09 8.00E-09 8.00E-09 7.00E-09 7.00E-09 7.00E-09 6.00E-09 5.00E-09 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Figure 5.20. Plot of Output Leakage Current_2 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V versus total dose. The solid diamonds are the average of the measured data points for the samples irradiated under electrical bias while the shaded diamonds are the average of the measured data points for the samples irradiated with all pins tied to ground. The black lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated under electrical bias while the gray lines (solid and/or dashed) are the upper and/or lower confidence limits, as determined by KTL statistics, on the samples irradiated in the unbiased condition. The red dotted line(s) are the pre- and/or post-irradiation minimum and/or maximum specification value as defined in the datasheet and/or test plan. An ISO 9001:2008 and DLA Certified Company 47 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table 5.20. Raw data for Output Leakage Current_2 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V versus total dose, including the statistical analysis, specification and the status of the testing (pass/fail). Output Leakage Current_2 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V Device 1802 1803 1804 1805 1806 1807 1808 1809 1841 1842 1850 Biased Statistics Average Biased Std Dev Biased Ps90%/90% (+KTL) Biased Ps90%/90% (-KTL) Biased Un-Biased Statistics Average Un-Biased Std Dev Un-Biased Ps90%/90% (+KTL) Un-Biased Ps90%/90% (-KTL) Un-Biased Specification MAX Status 50 6.00E-09 5.00E-09 5.00E-09 4.00E-09 2.00E-09 2.00E-09 0.00E+00 0.00E+00 1.00E-09 0.00E+00 1.00E-09 100 8.00E-09 8.00E-09 4.00E-09 8.00E-09 3.00E-09 0.00E+00 0.00E+00 0.00E+00 0.00E+00 1.00E-09 2.00E-09 24-hr Anneal 110 6.00E-09 4.00E-09 2.00E-09 3.00E-09 2.00E-09 2.00E-09 2.00E-09 2.00E-09 2.00E-09 3.00E-09 3.00E-09 2.40E-09 1.67E-09 6.99E-09 -2.19E-09 4.40E-09 1.52E-09 8.56E-09 2.42E-10 6.20E-09 2.49E-09 1.30E-08 -6.28E-10 3.40E-09 1.67E-09 7.99E-09 -1.19E-09 4.00E-10 5.48E-10 1.90E-09 -1.10E-09 2.60E-09 1.52E-09 6.76E-09 -1.56E-09 2.00E-06 PASS 6.00E-10 8.94E-10 3.05E-09 -1.85E-09 2.00E-06 PASS 2.00E-10 4.47E-10 1.43E-09 -1.03E-09 2.00E-06 PASS 2.20E-09 4.47E-10 3.43E-09 9.74E-10 2.00E-06 PASS 3.60E-09 5.48E-10 5.10E-09 2.10E-09 2.00E-06 PASS 0 2.00E-09 0.00E+00 1.00E-09 0.00E+00 1.00E-09 2.00E-09 0.00E+00 0.00E+00 0.00E+00 0.00E+00 0.00E+00 10 6.00E-09 2.00E-09 2.00E-09 2.00E-09 0.00E+00 1.00E-09 2.00E-09 2.00E-09 1.00E-09 2.00E-09 3.00E-09 Total Dose (krad(Si)) 20 30 4.00E-09 4.00E-09 2.00E-09 4.00E-09 1.00E-09 2.00E-09 2.00E-09 2.00E-09 0.00E+00 0.00E+00 2.00E-09 0.00E+00 1.00E-09 3.00E-09 2.00E-09 3.00E-09 3.00E-09 3.00E-09 2.00E-09 4.00E-09 4.00E-09 5.00E-09 8.00E-10 8.37E-10 3.09E-09 -1.49E-09 2.40E-09 2.19E-09 8.41E-09 -3.61E-09 1.80E-09 1.48E-09 5.87E-09 -2.27E-09 4.00E-10 8.94E-10 2.85E-09 -2.05E-09 2.00E-06 PASS 1.60E-09 5.48E-10 3.10E-09 9.81E-11 2.00E-06 PASS 2.00E-09 7.07E-10 3.94E-09 6.11E-11 2.00E-06 PASS An ISO 9001:2008 and DLA Certified Company 48 168-hr Anneal 120 1.00E-09 0.00E+00 0.00E+00 0.00E+00 1.00E-09 4.00E-09 4.00E-09 3.00E-09 3.00E-09 4.00E-09 4.00E-09 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 6.0. Summary / Conclusions The low dose rate testing described in this final report was performed using the facilities at Aeroflex RAD's Longmire Laboratories in Colorado Springs, CO. The low dose rate source is a GB-150 irradiator modified to provide a panoramic exposure. The Co-60 rods are held in the base of the irradiator heavily shielded by lead. During the irradiation exposures the rod is raised by an electronic timer/controller and the exposure is performed in air. The dose rate for this irradiator in this configuration ranges from approximately 1mrad(Si)/s to a maximum of approximately 50rad(Si)/s, determined by the distance from the source. The parametric data was obtained as "read and record" and all the raw data plus an attributes summary are contained in this report as well as in a separate Excel file. The attributes data contains the average, standard deviation and the average with the KTL values applied. The KTL value used in this work is 2.742 per MIL-HDBK-814 using one sided tolerance limits of 90/90 and a 5-piece sample size. The 90/90 KTL values were selected to match the statistical levels specified in the MIL-PRF-38535 sampling plan for the qualification of a radiation hardness assured (RHA) component. Note that the following criteria must be met for a device to pass the low dose rate test: following the radiation exposure each of the 5 pieces irradiated under electrical bias shall pass the specification value. The units irradiated without electrical bias and the KTL statistics are included in this report for reference only. If any of the 5 pieces irradiated under electrical bias exceed the device post radiation data sheet specification limits, then the lot could be logged as a failure. Based on this criterion the RH119W High Performance Dual Comparator (from the lot date code identified on the first page of this test report) PASSED the enhanced low dose rate sensitivity test to the maximum tested dose level of 100krad(Si) with all parameters remaining within their datasheet specifications. Further, the data in this report can be analyzed along with the high dose rate report titled "Total Ionizing Dose (TID) Radiation Testing of the RH119W High Performance Dual Comparator for Linear Technology" to demonstrate that these parts do not exhibit ELDRS as defined in the current test method. An ISO 9001:2008 and DLA Certified Company 49 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Appendix A: Photograph of Packing Label and a Sample Unit-Under-Test to Show Part Traceability An ISO 9001:2008 and DLA Certified Company 50 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Appendix B: Radiation Bias Connections and Absolute Maximum Ratings ELDRS Radiation Biased Conditions: Extracted from Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C. Pin 1 2 3 4 5 6 7 8 9 10 Function Connection / Bias OUT 1 To +15V via 10kΩ Resistor GND 1 To GND +IN 1 To Pin 8 To Pin 9, -IN 1 To +15V via 10kΩ Resistor, To Pin 8 via 100Ω Resistor VTo -15V OUT 2 To +15V via 10kΩ Resistor GND 2 To GND To Pin 3, +IN 2 To -15V via 10kΩ Resistor, To Pin 4 via 100Ω Resistor -IN 2 To Pin 4 V+ To +15V Figure B.1. Irradiation bias circuit. This figure was extracted from Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C. An ISO 9001:2008 and DLA Certified Company 51 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 ELDRS Radiation Unbiased Conditions: All pins grounded. Pin 1 2 3 4 5 6 7 8 9 10 Function OUT 1 GND 1 +IN 1 -IN 1 VOUT 2 GND 2 +IN 2 -IN 2 V+ Connection / Bias GND GND GND GND GND GND GND GND GND GND Figure B.2. W Package drawing (for reference only). This figure was extracted from Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C. Absolute Maximum Ratings: Parameter Supply Voltage Output to Negative Supply Voltage Ground to Negative Supply Voltage Ground to Positive Supply Voltage Differential Input Voltage Differential Input Current Output Short-Circuit Duration Max Rating 36V 36V 25V 18V ±5V ±5mA 10 sec An ISO 9001:2008 and DLA Certified Company 52 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Appendix C: Electrical Test Parameters and Conditions The expected ranges of values as well as the measurement conditions are taken from Linear Technology RH119 Datasheet I.D No. 66-10-0176 Revision C. All electrical tests for this device are performed on one of Aeroflex RAD's LTS2020 Test Systems. The LTS2020 Test System is a programmable parametric tester that provides parameter measurements for a variety of digital, analog and mixed signal products including voltage regulators, voltage comparators, D to A and A to D converters. The LTS2020 Test System achieves accuracy and sensitivity through the use of software self-calibration and an internal relay matrix with separate family boards and custom personality adapter boards. The tester uses this relay matrix to connect the required test circuits, select the appropriate voltage / current sources and establish the needed measurement loops for all the tests performed. The measured parameters and test conditions are shown in Table C.1. A listing of the measurement precision/resolution for each parameter is shown in Table C.2. The precision/resolution values were obtained from test data or from the DAC resolution of the LTS-2020 for the particular test shown, whichever is greater. To generate the precision/resolution shown in Table C.2, one of the units-under-test was tested repetitively (a total of 10-times with re-insertion between tests) to obtain the average test value and standard deviation. Using this test data MIL-HDBK-814 90/90 KTL statistics were applied to the measured standard deviation to generate the final measurement range. This value encompasses the precision/resolution of all aspects of the test system, including the LTS2020 mainframe, family board, socket assembly and DUT board as well as insertion error. In some cases, the measurement resolution is limited by the internal DACs, which results in a measured standard deviation of zero. In these instances the precision/resolution will be reported back as the LSB of the DAC. Note that the testing and statistics used in this document are based on an “analysis of variables” technique, which relies on small sample sizes to qualify much larger lot sizes (see MIL-HDBK-814, p. 91 for a discussion of statistical treatments). Not all measured parameters are well suited to this approach due to inherent large variations. If necessary, larger samples sizes could be used to qualify these parameters using an “attributes” approach. An ISO 9001:2008 and DLA Certified Company 53 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table C.1. Measured parameters and test conditions for the RH119W High Performance Dual Comparator. Parameter Symbol Test Conditions Positive Supply Current (A) +IS VS=+/-15V, VCM=0V Negative Supply Current (A) -IS VS=+/-15V, VCM=0V Offset Voltage (V) VOS VS=+/-15V, VCM=0V Input Offset Current (A) IOS VS=+/-15V, VCM=0V Positive Input Bias Current (A) +IB VS=+/-15V, VCM=0V Negative Input Bias Current (A) -IB VS=+/-15V, VCM=0V Large Signal Voltage Gain (V/mV) AVOL VS=+/-15V, VCM=0V Common Mode Rejection Ratio (dB) CMRR Saturation Voltage1 (V) VSAT1 Saturation Voltage2 (V) VSAT2 VS=+/-15V, VCM=+/-10V V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA Output Leakage Current (A) ICEX VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V An ISO 9001:2008 and DLA Certified Company 54 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Table C.2. Measured parameters, pre-irradiation specifications and measurement precision for the RH119W High Performance Dual Comparator. Pre-Irradiation Specification Parameter MIN MAX Positive Supply Current (A) Measurement Precision/Resolution 1.15E-02 Negative Supply Current (A) -4.50E-03 ±4.92E-05 ±3.24E-06 Offset Voltage (V) 4.00E-03 ±7.43E-06 7.50E-08 ±4.93E-10 Positive Input Bias Current (A) 5.00E-07 ±3.57E-09 Negative Input Bias Current (A) 5.00E-07 ±3.56E-09 Input Offset Current (A) -7.50E-08 Large Signal Voltage Gain (V/mV) 1.00E+01 ±9.10E-01% Common Mode Rejection Ratio (dB) 9.00E+01 ±1.17E+00 Saturation Voltage1 (V) 4.00E-01 ±9.97E-04 Saturation Voltage2 (V) 1.50E+00 ±4.94E-03 Output Leakage Current (A) 2.00E-06 ±1.75E-09 An ISO 9001:2008 and DLA Certified Company 55 ELDRS Report 15-0663 03/17/16 R1.0 Aeroflex RAD 5030 Centennial Blvd. Colorado Springs, CO 80919 (719) 531-0800 Appendix D: List of Figures Used in the Results Section (Section 5) 5.1. 5.2. 5.3. 5.4. 5.5. 5.6. 5.7. 5.8. 5.9. 5.10. 5.11. 5.12. 5.13. 5.14. 5.15. 5.16. 5.17. 5.18. 5.19. 5.20. Positive Supply Current (A) @ VS=+/-15V, VCM=0V Negative Supply Current (A) @ VS=+/-15V, VCM=0V Offset Voltage_1 (V) @ VS=+/-15V, VCM=0V Offset Voltage_2 (V) @ VS=+/-15V, VCM=0V Input Offset Current_1 (A) @ VS=+/-15V, VCM=0V Input Offset Current_2 (A) @ VS=+/-15V, VCM=0V Positive Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V Positive Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V Negative Input Bias Current_1 (A) @ VS=+/-15V, VCM=0V Negative Input Bias Current_2 (A) @ VS=+/-15V, VCM=0V Large Signal Voltage Gain_1 (V/mV) @ VS=+/-15V, VCM=0V Large Signal Voltage Gain_2 (V/mV) @ VS=+/-15V, VCM=0V Common Mode Rejection Ratio_1 (dB) @ VS=+/-15V, VCM=+/-10V Common Mode Rejection Ratio_2 (dB) @ VS=+/-15V, VCM=+/-10V Saturation Voltage1_1 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA Saturation Voltage1_2 (V) @ V+=+4.5V, V-=0V, VCM=0V, VIN=-6mV, ISINK=3.2mA Saturation Voltage2_1 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA Saturation Voltage2_2 (V) @ VS=+/-15V, VCM=0V, VIN=-5mV, IO=25mA Output Leakage Current_1 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V Output Leakage Current_2 (A) @ VS=+/-15V, VCM=0V, VIN=5mV, VO to V-=35V An ISO 9001:2008 and DLA Certified Company 56