Liebert Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C; Medical Match-Up; 50/60Hz (R 7/00) (SL-16548)

Keeping Business in Business
•
•
•
•
•
•
•
•
Process Fluid Chiller Designed
For Varian CLINAC 4/100, 6/100 & 600C
Dedicated, equipment-matched cooling supplies the proper temperature and water flow year round.
Factory-tested, piped, wired and charged for hassle-free startup and operation.
Clean, non-ferrous cooling loop, required for sensitive medical equipment.
Local service and parts for quick and knowledgeable maintenance and service.
Optional switch-over to alternate water source provides redundant, continuous operation.
Optional remote monitoring systems keep you informed of chiller operation.
UL/CSA listing on standard models.
ISO 9001 certified company.
CLINAC 4/100, 6/100 & 600C Specifications
Fluid Specified
Water
Heat Dissipation—BTUH (kW)
23,900 (7)
Fluid Temperature Range—°F (°C)
Fluid Flow Rate—GPM (l/m)
50-70 (10-25)
1 (3.8)
Fluid Pressure Drop—PSI (kPa)
25-40 (172-276)
Note: Performance is based on 95°F condenser ambient.
Process Fluid Chiller Specifications
Model Number
PS024A
PS036A
PS036A
Fluid Circulated
Water
Ethylene Glycol
Propylene Glycol
Net Cooling Capacity—BTUH (kW)
24,300 (7.1)
24,400 (7.2)
23,900 (7.0)
Leaving Fluid Temperature—°F (°C)
61 (16.1)
47 (8.3)
50 (10)
Fluid Flow Rate—GPM (l/m)
1.3 (4.9)
1.5 (5.7)
1.6 (6.1)
* (*)
* (*)
* (*)
Chiller Pressure Drop—PSI (kPa)
6.3 (43.4)
2.3 (15.9)
2.8 (19.3)
Pump Total Head—PSI (kPa)
41 (282.7)
41 (282.7)
41 (282.7)
3/4 (.56)
3/4 (.56)
3/4 (.56)
CLINAC Pressure Drop—PSI (kPa)
Pump HP (kW)
*Data not available
Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C
1
A
AIRFLOW
IN
B
D
AIRFLOW
OUT
C*
SHADED AREAS
INDICATE A
RECOMMENDED CLEARANCE
FOR COMPONENT REMOVAL,
ACCESS, AND AIR FLOW.
OPTIONAL REFRIGERANT
PRESSURE GAUGES
SHADED AREAS
INDICATE A
RECOMMENDED CLEARANCE
FOR COMPONENT REMOVAL,
ACCESS, AND AIR FLOW.
2"
18"
(457mm) (51mm)
36"
(914mm)
36"
(914mm)
I
I
H
1" MAX.
(25mm)
36"
(914mm)
H
UNIT DIMENSIONAL DATA
1" MAX.
(25mm)
UNIT DIMENSIONAL DATA
1" NPT
Model No.
A-Depth
in. (mm)
B-Width
in. (mm)
C-Height
in. (mm)
Weight
lbs. (kg)
Voltage
(60 Hz)
Voltage
(50 Hz)
PS018A
36 (914)
40 (1016)
25.25 (641))
394 (178)
208/230, 1-phase
200/230, 1-phase
PS024A
36 (914)
40 (1016)
25.25 (641)
404 (183)
208/230, 1-phase
200/230, 1-phase
PS036A
36 (914)
48 (1219)
32.5 (825)
515 (233)
460, 3-phase
208/230, 1- or 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS048A
36 (914)
53 (1346)
38.5 (978)
592 (268)
460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
PS060A
36 (914)
53 (1346)
38.5 (978)
642 (291)
460, 3-phase
208/230, 3-phase
400, 3-phase
200/230, 1- or 3-phase
Standard Features
Accessories
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Indoor/outdoor cabinet
Hermetic compressor
Refrigerant dehydrator
Expansion valve
Refrigerant receiver
Stainless steel fluid reservoir expansion tank
Pressure relief valve
Hot gas bypass valve capacity control
Optional Features
• Refrigerant gauges and reservoir heater
• Stainless steel pump
1050 Dearborn Drive
P.O. Box 29186
Columbus, OH 43229
Phone: 800.877.9222
Fax: 614.841.6022
Via Leonardo Da Vinci 8
Zona Industriale Tognana
35028 Piove Di Sacco
Italy
Phone: 39.049.9719.111
Fax: 39.049.5841257
Liebert Web Site—http://www.liebert.com
Emergency water bypass
Enviro-source emergency bypass
Water level control and alarm system
Fluid pressure gauge
Check valve
Dial thermometer
Relief valve
Circuit setter
Ball valve
High temperature thermostat
Low flow switch
Liebert RCM8 monitor
23/F Allied Kajima Bldg.
138 Gloucester Road
Wanchai
Hong Kong
Phone: 852.2.572.2201
Fax: 852.2.572.1182
©2000 Liebert Corporation. All rights
reserved throughout the world.
Specifications subject to change without
notice.
®Liebert and the Liebert logo are
registered trademarks of Liebert
Corporation. All names referred to are
trademarks or regisered trademarks of
their respective owners.
SL-16548 (07/00)
Process Fluid Chiller Designed For Varian CLINAC 4/100, 6/100 & 600C
2