AGILENT 4285A

Agilent 4285A
Precision LCR Meter
Data Sheet
Specifications
The complete Agilent Technologies 4285A specifications are listed below. These specifications are
the performance standards or limits against which
the instrument is tested. When shipped from the
factory, the 4285A meets the specifications listed
in this section. The specification test procedures
are covered in Agilent 4285A Maintenance Manual
(Agilent Part Number 04285-90030).
Measurement Functions
Mathematical functions
The deviation and the percent of deviation of
measurement values from a programmable reference value.
Equivalent measurement circuit
Parallel and series
Ranging
Auto and manual (hold/up/down)
Trigger
Internal, external, BUS (GPIB), and manual
Measurement parameters
|Z| = Absolute value of impedance
|Y| = Absolute value of admittance
L = Inductance
C = Capacitance
R = Resistance
G = Conductance
D = Dissipation factor
Q = Quality factor
Rs = Equivalent series resistance
Rp = Parallel resistance
X = Reactance
B = Suceptance
θ = Phase angle
Delay time
Programmable delay from the trigger command to
the start of the measurement, 0 to 60.000 s in 1 ms
steps.
Measurement terminals
Four-terminal pair
Test cable length
0 m, 1 m, and 2 m selectable
Combinations of measurement parameters
|Z|, |Y|
L, C
R
G
θ (deg), θ (rad)
D, Q, Rs, Rp, G
X
B
Integration time
Short, medium, and long selectable (refer to Supplemental Performance Characteristics, page 14,
for the measurement time)
Averaging
1 to 256, programmable
Frequency
75 kHz to 30 MHz, 100 Hz solution
Output impedance
The following output impedance is specified for the
test cable length of 0 m:
Frequency accuracy
±0.01%
(25 + 0.5 fm) Ω ± 10 + 2 fm %
3
Signal modes
Normal—Program selected voltage or current at the
measurement terminals when they are opened or
shorted, respectively
where:
Test Signal
(
Constant—Maintains selected voltage or current at
the device under test (DUT) independent of
changes in the device’s impedance.
Signal level
The following test signal level accuracy is specified
for an ambient temperature range of 23 °C ± 5 °C
and the test cable length is 0 m.
Mode
Voltage Normal
Constant1
Current Normal
Constant1
Range
Setting accuracy
5 m Vrms to 2 Vrms
±{(8 + 0.4 fm)% + 1 m Vrms}
10 mVrms to 1 Vrms
±{(6 + 0.2 fm)% + 1 m Vrms}
200 µArms to 20 mArms
±{(8 + 1 fm)% + 40 µArms}
100 µArms to 20 mArms
±{(6 + 0.2 fm)% + 40 µArms}
1. When the ALC function is set to ON
For the temperature range of 0 °C to 55 °C, multiply the temperature induced setting error listed in
Figure 1-5 to the test signal setting accuracy. When
test cable length is 1 m or 2 m, add the following
error due to test cable length.
fm
)
= Test frequency [MHz]
Test signal level monitor
The following test signal level monitor accuracy is
specified for an ambient temperature range of
23 °C ± 5 °C and the test cable length is 0 m.
Mode
Range
Voltage
0.01 mVrms – 2.000 Vrms
Current
0.001 µArms – 20.00 mArms
Monitor accuracy
{
± Smon
4 + 0.2 fm
Sset
+
100
500
For the temperature range of 0 °C to 55 °C, multiply the temperature induced setting error listed in
Figure 1-5. When test cable length is 1 m or 2 m,
add the following error due to test cable length.
Smon 0.2 fm L [V]
where:
fm =
L =
Smon =
Sset =
Test frequency [MHz]
Test cable length [m]
Readout value of test signal level
Setting value of test signal level
For example,
0.2 fm L [%]
where:
fm
L
2
= Test frequency [MHz]
= Test cable length [m]
} [V ]
Test frequency:
Test signal level:
Monitor readout value:
Cable length:
Ambient temperature:
1 MHz
1 Vrms
500 mVrms
1m
25 °C
Then, voltage level monitor accuracy Vma is
ΔVma = 0.5 4 + 0.2 1 1 0.2
0.2
+ 0.5 11
+
100
100
100
= 0.024 [V]
Vma =
0.024
0.5
Measurement Accuracy
The measurement accuracy includes stability, temperature coefficient, linearity, repeatability, and
calibration interpolation error. The measurement
accuracy is specified when all of the following conditions are satisfied:
• Warm-up time: ≥ 30 minutes
≈ 4.8 [%]
• Test cable length: 0 m, 1 m (Agilent 16048A), or
2 m (Agilent 16048D). For the 1 m or 2 m cable
length operation (with Agilent 16048A/D),
CABLE CORRECTION has been performed
Display Range
Parameter
Range
|Z|, R, X
0.00001 Ω to 99.9999 MΩ
|Y|, G, B
0.00001 µS to 99.9999 S
C
0.00001 pF to 999.999 µF
L
0.001 nH to 99.9999 H
D
0.000001 to 9.99999
Q
0.01 to 99999.9
θ
-180.000° to 180.000°
Δ
-999.999% to 999.999%
• OPEN and SHORT corrections have been performed.
• The optimum measurement range is selected by
matching the DUT’s impedance to the effective
measuring range shown in Figure 1-1 and Figure
1-2. (For example, if the DUT’s impedance is
3 kΩ and oscillator level is less than or equal to
1 V, the optimum range is the 500 Ω range.)
• Measurement accuracy is specified at the following reference planes:
• Test frequency ≤ 1 MHz
At the UNKNOWN terminals on the Agilent
4285A front panel or at the end of the standard test leads (Agilent 16048A/D).
• Test frequency ≥ 1.001 MHz
At the 1-port terminal of the Agilent 16085B
Terminal Adapter, which should be connected
to the UNKNOWN terminals of the Agilent
4285A or to the end of the standard test leads
(Agilent 16048A/D).
3
Figure 1-1. Effective measurement range (oscillator level ≤ 1 Vrms)
4
Figure 1-2. Effective measurement range (oscillator level > 1 Vrms)
5
|Z|, |Y|, L, C, R, X, G, and B accuracy
|Z|, |Y|, L, C, R, X, G, and B accuracy Ae is given as
Q accuracy
Q accuracy Qe is given as
Ae = ±(An + Ac) Kt [%]
Qe = ±
Q2x De
1 Qx De
where:
where:
An = Basic accuracy equation given from the A1
to A16 shown in Table 1-1. The applicable
frequency range and impedance range of
equations A1 to A16 are shown in Figure 1-3
and Figure 1-4. (Refer to Basic Accuracy
Equations on page 7.)
Ac = Cable length factor (Refer to Cable Length
Factor on page 10.)
Kt = Temperature factor (Refer to Temperature
Factor on page 10.)
L, C, X, and B accuracies apply when Dx (measured
D value) ≤ 0.1.
When Dx > 0.1, multiply Ae by `b
1 + D2x for L, C, X,
and B accuracies.
R and G accuracies apply when Qx (measured Q
value) ≤ 0.1.
Qx = Measured Q value
De = D accuracy
Q accuracy applies when Qx De <1.
θ Accuracy
θ accuracy θe is given as
θe = ±
180 Ae
[deg]
π 100
where:
Ae = |Z|, |Y|, L, C, R, X, G, and B accuracy
G Accuracy
G accuracy Ge is given as
Ge = ±Bx De
When Qx > 0.1, multiply Ae by `b
1 + Q x for R and G
accuracies. G accuracy given by the equation above
applies to the G-B combination only.
(
D accuracy
D accuracy De is given as
where:
2
Bx = 2 πfCx =
=
=
=
=
=
[S]
1
2 πfLx
)
where:
Bx
Cx
Lx
De
f
Ae = |Z|, |Y|, L, C, R, X, G, and B accuracy
G accuracy applies when Dx (measured D value) ≤ 0.1.
De = ± Ae
100
D accuracy applies when Dx (measured D value) ≤ 0.1.
When Dx > 0.1, multiply De by (1 + Dx).
6
Measured B value [S]
Measured C value [F]
Measured L value [H]
D accuracy
Test frequency [Hz]
G accuracy given by the equation above applies to
the Cp-G and Lp-G combinations only.
Rp accuracy
Rp accuracy Rpe is given as
Rpe = ±
Rpx De
[Ω]
Dx De
Table 1-1. An equations
where:
Rs accuracy
Rs accuracy Rse is given as
Rse = ± Xx De [Ω]
fm 2 .
m|
0.1% + |Z500
5
)
f
[0.02% + ( 30m ) . 0.05%] . Ki . Kosc
f
1
2 πfCx
A5 = 0.18% + |Z5mk| . 0.02% . Ki . Kosc
f
f
A6 = 0.18% + ( 30m )2 . 3% + |Z5mk| [0.02% + ( 10m ) . 0.03%] . Ki . Kosc
f
f
A7 = 0.5% + ( 30m )2 . 3% + |Z5mk| . ( 30m ) . 0.2% . Ki . Kosc
m| .
. .
A8 = 0.18% + |Z
50 k 0.03% Ki Kosc
f 2
f
A9 = N2% + ( 30m ) . 3% + |Z100
[0.02% + ( 30m ) . 0.1%] . Ki
m|
f
)
where:
=
=
=
=
=
f
f
Rp accuracy applies when Dx (measured D value) ≤ 0.1.
Xx
Cx
Lx
De
f
f
m|
A4 = 0.3% + ( 30m )2 . 3% + |Z500
[0.05% + ( 30m ) . 0.1%] . Ki . Kosc
Rpx = Measured Rp value [Ω]
Dx = Measured D value
De = D accuracy
Xx = 2 πfLx =
f
A2 = N1% + ( 30m )2 . 3% + |Z50m| [0.02% + ( 30m ) . 0.05%] . Ki . Kosc
A3 = N1% + (
(
f
A1 = N1% + ( 30m )2 . 3% + |Z50m| [0.02% + ( 30m ) . 0.1%] . Ki . Kosc
Measured X value [Ω]
Measured C value [F]
Measured L value [H]
D accuracy
Test frequency [Hz]
Rs accuracy applies when Dx (measured D value) ≤ 0.1.
Basic accuracy equations
The basic accuracy An is calculated from the
following procedure:
1. Determine An equation from Figure 1-3 or
Figure 1-4. In both charts, boundary line belongs
to the better accuracy area.
When the oscillator level is ≤ 1 Vrms, determine An
to be applied, value of Ki and value of Kosc from the
Figure 1-3. If the determined Kosc ≤ 1, then round
up Kosc to 1.
When the oscillator level is > 1 Vrms, determine An
to be applied and value of Ki from the Figure 1-4.
f
m|
A10 = N2% + ( 30m )2 . 3% + |Z100
[0.02% + ( 30m ) . 0.05%] . Ki
f
f
A11 = 0.18% + ( 5m )2 . 0.1% + |Z1mk| [0.02% + ( 30m ) . 0.05%] . Ki
f
f
A12 = 0.3% + ( 30m )2 . 3% + |Z1mk| [0.05% + ( 30m ) . 0.1%] . Ki
m| .
.
A13 = 0.18% + |Z
10 k 0.02% Ki
f
f
m .
m|
.
A14 = 0.18% + ( 30m )2 . 3% + |Z
10 k [0.02% + ( 10 ) 0.03%] Ki
f
m| .
A15 = 0.5% + ( 30m )2 . 3% + |Z
10 k (
fm
30
) . 0.2%] . Ki
m| .
.
A16 = 0.18% + |Z
100 k 0.03% Ki
|Zm|: Absolute value of measured impedance in [Ω]
fm: Test frequency in [MHz]
N accuracy factors
N1 and N2 in the An equations have the following
values:
Table 1-2. N accuracy factors
Frequency(f)
N1
N2
75 kHz ≤ f ≤ 200 kHz
0.15
0.15
200 kHz < f ≤ 3 MHz
0.08
0.15
3 MHz < f ≤ 5 MHz
.15
0.38
5 MHz < f ≤ 30 MHz
0.30
0.38
2. Calculate An from the formula to be applied. The
n accuracy factor included in the An equation is
shown in Table 1-2. Use Ki and Kosc factors determined in previous step.
7
Figure 1-3. Accuracy equations, Ki factor, and Kosc factor (test signal level ≤ 1 Vrms)
8
Figure 1-4. Accuracy equations and Ki factor (test signal level > 1 Vrms )
9
Cable length factor
Add the following cable length factor Ac to the
measurement accuracy when the cable length is set
to 1 m (for 16048A) or 2 m (for 16048D) in CABLE
field, after performing the cable correction and the
OPEN/SHORT correction. When the cable length is
0 m, Ac is 0 percent.
Aco is the additional error when the impedance
range is above 5 kΩ.
. .
Aco = |Zm| fm Kt [%]
1000
Zm = 2 πfmLx
≈ 35 [Ω]
where:
fm
Lx
where:
fm = Test frequency in [MHz]
Zm = Absolute value of measured impedance in [kΩ]
Kl = Test cable length in [m]
Temperature factor
Multiply the sum of the basic accuracy and the
cable length factor by the following temperature
induced error Kt, when the temperature range is
0 °C to 55 °C. The boundary belongs to the smaller
multiplier.
= Test frequency [Hz]
= Measured inductance value of the DUT [H]
3. Choose an accuracy chart from Figure 1-3 and
Figure 1-4. The oscillator level is 1 Vrms, then
Figure 1-3 is chosen for this measurement.
4. Find the frequency point of fm (25.2 MHz) along
the X axis in Figure 1-3. Both axes are in log format. Interpolation may be required.
5. Find the impedance point of Zm (35 Ω) along the
Y axis in Figure 1-3 determined in step 2. Both
axes are in log format. Interpolation may be
required.
6. Mark the intersection of above two steps and
determine the basic accuracy equation An, integration factor Ki, and oscillator level factor Kosc.
Figure 1-5. Temperature factor Kt
Measurement Accuracy Calculation Example
Example of Ls-Q accuracy calculation
Measurement conditions
10
Ae = ±(An + Ac) Kt
2. First of all, to determine the measurement
accuracy Ae, calculate the impedance value from
the DUT’s inductance value. So the measurement
impedance Zm is:
Ac = fm + Aco [%]
15
Measured inductance Lx of DUT:
Measured Q value of DUT:
Test signal level:
Test frequency fm:
Integration time:
Cable length:
Operating temperature:
Determine inductance measurement accuracy Ae
1. From |Z|, |Y|, L, C, R, X, G, and B Accuracy
(see page 6), measurement accuracy Ae is determined as below:
220 nH
30
1 Vrms
25.2 MHz
LONG
0m
28 °C
From:
Test frequency fm:
DUT’s impedance Zm:
Integration time:
Test signal level:
25.2 MHz
35 Ω
LONG
1 Vrms
Then, An = A1, Ki = 1, and Kosc = 1 (rounded from
0.02).
7. From Table 1-1, the actual accuracy equation to
be applied is determined as A1.
Then,
f
f 2
A1 = N1% + ( 30 ) . 3% + |Z50m| [0.02% + ( 30 ) . 0.1%] . Ki . Kosc
8. Determine N1 from Table 1-2.
From frequency = 25.2 MHz, then, N1 = 0.3.
9. Then,
An = 0.3% + (
≈ 2.6 [%]
25.2 2 .
50
3% + |35|
30
)
[0.02% + (
25.2
30
Correction Functions
Zero open
Eliminates measurement errors due to parasitic
stray admittance (C, G) of the test fixture.
Zero short
Eliminates measurement errors due to parasitic
residual impedances (L, R) of the test fixture.
Load
Improves the measurement accuracy by using a
device whose value is accurately known (a working
standard) as a reference.
) . 0.1%] . 1 . 1
10. Cable length is 0 m, then Ac = 0.
11. Operating temperature is 28 °C, then Kt = 1
(from Figure 1-5).
List Sweep
A maximum of ten frequencies or test signal levels
can be programmed. Single or sequential tests can
be performed. When Option 4285A-001 is installed,
DC bias voltages can also be programmed.
Comparator Function
12. Therefore, inductance measurement accuracy
Ae is:
Ae = ±(An + Ac) Kt
= ±(2.6 + 0) 1
= ±2.6 [%]
Determine Q measurement accuracy Qe
1. From Q Accuracy (see page 6), Q measurement
accuracy Qe is determined as below:
Ten-bin sorting for the primary measurement
parameter, and IN/OUT decision output for the
secondary measurement parameter.
Sorting modes
Sequential mode. Sorting into unnested bins with
absolute upper and lower limits.
Tolerance mode. Sorting into nested bins with
absolute or percent limits.
2
Qe = ± Qx De
1 Qx De
Bin count
0 to 999999
2. Determine D accuracy De for calculating Qe.
From the previous step Determine inductance
measurement accuracy Ae (see page 10), Ae is
2.6 [%], then:
List sweep comparator
HIGH/IN/LOW decision output for each point in
the list sweep table.
De = ± Ae
100
= ± 0.026
3. Therefore Qe is:
Qe = ±
Qx2 De
1 Qx De
2
Qe = ± 30 0.026
1 30 0.026
≈ +106/–13
11
Other Functions
Store/Load
Ten instrument control settings, including comparator limits and list sweep programs, can be
stored and loaded into and from the internal nonvolatile memory. Ten additional settings can also
be stored and loaded from each memory card.
GPIB
All control settings, measured values, comparator
limits, and list sweep program can be controlled or
monitored. Uses ASCII and 64-bit binary data format.
GPIB buffer memory can store measured values for
a maximum of 128 measurements and output packed
data over the GPIB bus. Complies with IEEE-488.1
and 488.2. The programming language is Test and
Measurement Systems Language (TMSL).
DC bias monitor terminal
DC bias voltage or current can be monitored at the
rear panel BNC connector.
The following monitor accuracies are applied when
the digital volt meter whose input impedance is ≥
10 MΩ is used.
• DC bias voltage monitor
DC bias voltage across the DUT 1
Output impedance: 11 kΩ
Monitor accuracy: ±(0.2% of reading + 2 + 0.8
Idut) mV
where:
Idut is current flowing through the DUT in [mA].
GPIB interface functions
SH1, AH1, T5, L4, SR1, RL1, DC1, DT1, C0, E1
Self test
Softkey controllable. Provides a means to confirm
proper operation.
Option 4285A-001 (Internal DC Bias)
Adds the variable DC bias voltage function.
DC bias level
The following DC bias level accuracy is specified
for an ambient temperature range of 23 °C ± 5 °C.
Multiply the temperature induced setting error, Kt
listed in Figure 1-5 for the temperature range of
0 °C to 55 °C.
Voltage range
Resolution
Setting accuracy
±(0.000 to 4.000) V
1 mV
±(0.1% of setting +1 mV)
±(4.002 to 8.000) V
2 mV
±(0.1% of setting +2 mV)
±(8.005 to 20.000) V
5 mV
±(0.1% of setting +5 mV)
±(20.01 to 40.00) V
10 mV
±(0.1% of setting +10 mV)
A maximum DC bias current of 100 mA can be
applied to the DUT.
12
• DC bias current monitor
DC bias current through the DUT 10 Ω (1 V at
100 mA)
Output impedance: 10 kΩ
Monitor accuracy: ±(1% of reading + 0.3) mA
Other Options
Option 4285A-700: No DC bias
Option 4285A-002: Accessory Control Interface
Allows the 4285A to control the Agilent 42841A
bias current source or the Agilent 42851A precision Q adapter.
The voltage ratio meaurement accuracy, when the
4285A is used with the 42851A precision Q
adapter, is described in the 42851A’s operation
manual.
Option 4285A-201: Handler interface
Option 4285A-202: Handler interface
Option 4285A-301: Scanner interface
Option 4285A-710: Blank panel
Option 4285A-907: Front handle kit
Option 4285A-908: Rack mount kit
Option 4285A-909: Rack flange and handle kit
Option 4285A-915: Add service manual
Option 4285A-ABA: Add English manual
Option 4285A-ABD: Add German manual
Option 4285A-ABJ: Add Japanese manual
Furnished Accessories
Power Requirements
Power cord
Depends on the country
where the 4285A is being
used
Line voltage
100, 120, 220 Vac ±10%, 240 Vac +5% –10%
100 Ω resistor box
Agilent P/N 04285-61001
BNC female-female
Adapter
Agilent P/N 1250-0080 (4 ea.)
Line frequency
47 to 66 Hz
Fuse
Only for Option 4285A-201
Agilent P/N 2110-0046
(2 ea.)
Accessories Available
Test fixture/test leads
16034E
Test fixture for SMD or chip type
DUT, f ≤ 40 MHz
16044A
Four-terminal test fixture for
SMD or chip type DUT, f ≤ 10 MHz
16047A
Test fixture for axial or radial
DUT, f ≤ 13 MHz
16047D
Test fixture for axial or radial
DUT, f ≤ 40 MHz
16048A
Test leads, length 1 m (BNC
connector)
16048D
Test leads, length 2 m (BNC
connector)
16048G
Test fixture for SMD or chip type
DUT, f ≤ 110 MHz
16048H
Test fixture for array-type SMD or
chip type DUT, f ≤ 110 MHz
16065A
External voltage bias fixture
16334A
Tweezer-type test fixture for SMD
or chip type DUT, f ≤ 15 MHz
16451B
Dielectric test fixture
42842C
Bias current test fixture
42851-61100
SMD test fixture
(Option 42842C-201)
DC bias source
42841A
Bias current source
Memory card
04278-89001
Memory card, 1 ea.
Power consumption
200 VA max.
Operating Environment
Temperature
0 °C to 55 °C
Humidity
≤ 95% R.H. at 40 °C
Dimensions
426 (W) by 177 (H) by 498 (D) (mm)
Weight
Approximately 16 kg (35.3 lb., standard)
Display
LCD dot-matrix display
Capable of displaying
Measured values
Control settings
Comparator limits and decisions
List sweep tables
Self test message and annunciations
Number of display digits
6-digits, maximum display count 999999
GPIB interconnection cables
10833A
GPIB cable, 1 m
10833B
GPIB cable, 2 m
10833C
GPIB cable, 4 m
10833D
GPIB cable, 0.5 m
13
Supplemental Performance Characteristics
The Agilent 4285A supplemental performance
characterisics are listed below. These supplemental
performance characteristics are not specifications,
but are typical characteristics included as supplemental information for the operator.
Stability
When the following conditions are satisfied,
Integration time:
LONG
Operating temperature: Constant operating temperature of 23 °C ± 5 °C
Parameter
≤ 1 MHz
30 MHz
|Z|, |Y|, L, C, R
< 0.01%/day
< 0.05%/day
D
< 0.0001/day
< 0.0005/day
Temperature coefficient
When the following conditions are satisfied,
Integration time:
LONG
Test signal voltage:
≥ 20 mVrms
Operating temperature: 23 °C ± 5 °C
Parameter
≤ 1 MHz
30 MHz
|Z|, |Y|, L, C, R
< 0.004%/°C
< 0.05%/°C
D
< 0.00004/°C
< 0.0005/°C
Input protection
Internal circuit protection, when a charged capacitor is connected to the UNKNOWN terminals.
The maximum capacitor voltage is:
`b
Vmax = 1 [V]
C
where:
Vmax = ≤ 200 V
C = Capacitance value in Farads
Measurement time
Typical measurement times from the trigger to the
output of EOM at the Handler Interface. (EOM:
End of Measurement)
Integration time
Measurement time
SHORT
30 ms
MEDIUM
65 ms
LONG
200 ms
In the following condition an additional measurement time, approx. 300 ms, is added to the measurement time.
Test signal voltage:
Measurement range:
Test signal current:
0.51 V – 2 Vrms
0 Ω Range
≥ 22 mA
Settling time
Display time
Frequency (fm)
Display time for each display format is given as:
< 50 msec.
< 100 msec.
MEAS DISPLAY page
Approx. 8 ms
BIN No. DISPLAY page
Approx. 5 ms
BIN COUNT DISPLAY page Approx. 0.5 ms
Measurement range
GPIB data output time
< 50 msec./range shift
Internal GPIB data, processing time from EOM output to measurement data output on GPIB lines
(excluding display time):
Test signal level
• Approx. 10 ms
14
Option 4285A-001 (internal DC bias)
Maximum DC bias current when the normal measurement can be performed is 100 mA.
DC bias settling time
When DC bias is set to ON, add 5 ms to the measurement time. This settling time does not include
the DUT charge time.
Sum of DC bias settling time plus DUT (capacitor)
charge time is shown in the following figure.
Figure 1-6. Sum of the DC bias settling time and DUT
(capacitor) charge time
15
Web Resources
www.agilent.com/find/lcrmeters
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the
products and applications you select.
Remove all doubt
www.agilent.com
Our repair and calibration services will get
your equipment back to you, performing
like new, when promised. You will get
full value out of your Agilent equipment
throughout its lifetime. Your equipment
will be serviced by Agilent-trained technicians using the latest factory calibration
procedures, automated repair diagnostics
and genuine parts. You will always have the
utmost confidence in your measurements.
For more information on Agilent
Technologies’ products, applications
or services, please contact your local
Agilent office. The complete list is
available at:
www.agilent.com/find/contactus
Americas
Canada
Latin America
United States
(877) 894-4414
305 269 7500
(800) 829-4444
Agilent offers a wide range of additional
expert test and measurement services for
your equipment, including initial start-up
assistance onsite education and training,
as well as design, system integration, and
project management.
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Thailand
1 800 629 485
800 810 0189
800 938 693
1 800 112 929
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
1 800 226 008
For more information on repair and
calibration services, go to
www.agilent.com/find/removealldoubt
Agilent Direct
www.agilent.com/find/agilentdirect
Quickly choose and use your test
equipment solutions with confidence.
Agilent
Open
www.agilent.com/find/open
Agilent Open simplifies the process of
connecting and programming test systems
to help engineers design, validate and
manufacture electronic products. Agilent
offers open connectivity for a broad range
of system-ready instruments, open industry
software, PC-standard I/O and global
support, which are combined to more
easily integrate test system development.
www.lxistandard.org
LXI is the LAN-based successor to GPIB,
providing faster, more efficient connectivity.
Agilent is a founding member of the LXI
consortium.
Europe & Middle East
Austria
01 36027 71571
Belgium
32 (0)2 404 93 40
Denmark
45 70 13 15 15
Finland
358 (0)10 855 2100
France
0825 010 700*
*0.125 ¤/minute
Germany
07031 464 6333
Ireland
1890 924 204
Israel
972-3-9288-504/544
Italy
39 02 92 60 8484
Netherlands
31 (0)20 547 2111
Spain
34 (91)631 3300
Sweden
0200-88 22 55
Switzerland
0800 80 53 53
United Kingdom 44 (0)118 9276201
Other European Countries:
www.agilent.com/find/contactus
Revised: October 6, 2008
Product specifications and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 1995, 2000.
2003, 2008
Printed in USA, November 12, 2008
5963-5395E