FUJI EDS3-143

INFRARED GAS ANALYZER
DATA SHEET
ZPA
This gas analyzer (ZPA) is capable of measuring the concentration of NO,SO 2 ,CO 2 ,CO,CH 4 and O 2 components in
sample gas. NO, SO 2 , CO 2 , CO and CH 4 are measured by
non-dispersion infrared method (NDIR), while O 2 is measured by fuel cell or paramagnetic and zirconia method.
Optimum use as an analyzer unit for measurement of
combustion exhaust gas from refuse incinerator, boiler
exhaust or gas from various industrial furnaces.
FEATURES
1.High sensitivity
Equipped with the newest upgraded mass flow sensor.
2.Easy maintenance
Maintenance is easy due to simple measurement unit of
single-beam system adapted.
3.Small and light
The size is small 133x483x382mm (HxWxD). (This is
for 19inch rack mounting type) and light (9kg). Unit is
capable max. 5 component measurement in one analyzer.
4.Extensive functions
Various optional functions are available such as auto
calibration control, atmospheric pressure correction, high
and low concentration alarm, remote range switch, and
range identification signal,etc.
SPECIFICATIONS
Standard Specifications
Principle of measurement:
NO, SO 2 , CO 2 , CO, CH 4 ;
Non-dispersion infrared-ray absorption
method
Single light source and single beams
(single beam system)
O 2 ;Fuel cell O 2 sensor (build-in) or paramagnetic O 2 sensor (build-in) or zirconia O 2
sensor (externally installed TYPE: ZFK7)
Measurable gas components and measuring range:
NO
SO2
CO2
CO
CH4
O2
Built in
fuel cell
O2
Built-in
Paramagnetic
O2
External
Zirconia
Minimum range
0 - 200ppm
0 - 200ppm
0 - 100ppm
0 - 200ppm
0 - 500ppm
Maximum range
0 - 5000ppm
0 - 10vol%
0 - 100vol%
0 - 100vol%
0 - 100vol%
0 - 10vol%
0 - 25vol%
0 - 5vol%
0 - 100vol%
None
100 - 95vol%
0 - 5vol%
0 - 25vol%
•Max. 5 components measurement including O 2 .
For reverse range O 2 measurement, infrared gas measurement is not available
(single range O 2 only).
•Measuring range ratio max. 1:10 (except
O2 )
•Measuring ranges are changeable between the specified minimum and maximum range
Settable one range or two ranges
* In measurement range low range is
called first range, high range is called
second range.
•For possible combinations of components
and ranges, refer to Table1.
Measured value indication:
Digital indication in 4 digits
(LCD panel with LED back light)
•Instantaneous value of each component
•Instantaneous value after O 2 correction
(only in NO, SO 2 , CO measurement with
O2 )
•Average value after O 2 correction
(only in NO, SO 2 , CO measurement with
O2 )
•O 2 average value
EDS3-143
Date
Nov. 2, 2011
ZPA
Analog output signals:
4 to 20mA DC or 0 to 1V DC, isolated
internally from circuit and ground. Output
lines are non-isolated each other.; 12
outputs max.
Allowable load 550Ω for 4 to 20mA DC
Allowable load 100KΩ for 0 to 1V DC
* Refer to Table2 for the channel No. of
displayed values and analog output
signals.
Analog input signal:
For signal input from externally installed
O 2 sensor.
Signal requirement;
(1) Signal from Fuji’s Zirconia O 2 sensor
(TYPE: ZFK7)
(2) 0 to 1V DC from an O 2 sensor
Input section is not isolated. This
feature is effective when an O 2 sensor
is not built in.
* Externally installed O 2 sensor should
be purchased separately.
Digital output: (Option)
1c contact (24V DC/1A, resistive load)
max.15 outputs
Instrument error, calibration error,
range identification, auto calibration
status, solenoid valve drive for auto
calibration, High/Low limit alarm contact output.
* All relay contacts are isolated mutually
and from the internal circuit.
Digital input: (Option)
Voltage contact (supply 12-24VDC (15mA
Max.)) Max. 9 inputs
Remote range change over, auto calibration remote start, remote hold,
average value reset, Isolated from
the internal circuit with photocoupler.
Power supply: Voltage rating
;100V to 240V AC
Allowable range
;85V to 264V AC
Frequency
;50Hz/60Hz
Power consumption ;100VA max.
Operation conditions:
Ambient temperature ;
- 5˚C to 45˚C
(40˚C max. when 2 optical system at 200V AC power source)
Ambient humidity ;90% RH max.,
non-condensing
Storage conditions:
Ambient temperature ; -20˚C to 60˚C
Ambient humidity ;100% RH max.,
non-condensing
Dimensions (H × W × D):
133 x 483 x 382mm
Mass:
Approx. 9 kg
Finish color: Front panel; Cool gray (PANTON 1C-F)
Enclosure: Steel casing, for indoor use
Material of gas-contacting parts:
Gas inlet/outlet; SUS304
Sample cell; SUS304,chloroprene rubber
Infrared-ray transmitting window; CaF2
Internal piping;Toaron,Teflon, Polypropylene
Paramagnetic O 2 sensor cell : SUS316
Fuel cell O 2 sensor cell : ABS resin
2
Gas inlet/outlet: Rc1/4 or NPT1/4 internal thread
Purge gas flow rate: 1L/min ( when required)
Life time of fuel cell O 2 sensor: 2 years
Standard Functions
Output signal holding:
Output signals are held unchanged during
manual and auto calibrations by activation
of holding (turning “ON” its setting).
The values held are those just before start
calibration mode or setting value.
It is selectable.
Indication of instantaneous values will
not be held.
Switch ranges: The switch ranges function is available in
manual, auto, and remote modes. Only
preset switch method is effective.
Manual: Allows range to switch by key operation.
Auto:
Automatically switched from first range to
second range when the measured value
exceeds to 90%FS of first range.
Automatically switched from second range
to first range when the measured value
drops to 80% or less first range.
Remote: Voltage contact input (for measurable
(Option) components)
Allows range to switch via an external
signal when remote range switch input
is received.
When the contact input terminals for
each component are input voltage, the
first range is selected, and it is switched
to the second range when the terminals
are open.
* These range value are settable between original
first range and second range.
Optional Functions
Remote output holding:
Output signal is held at the last value or
preset value by voltage input the remote
output holding input terminals.
Holding is maintained while the voltage
input the terminals. Indication of instantaneous values are not held.
Range identification signal:
The present measuring range is identified
by a contact position.
The contact output terminals for each
component close when the first range
is selected, and when the second range
is selected, the terminals are open.
Auto calibration:
Auto calibration is carried out periodically
at the preset cycle.
When a standard gas cylinder for calibration and a solenoid valve for opening /
closing the gas flow line are prepared
externally by the customer, calibration
will be carried out with the solenoid
valve drive contacts for zero calibration
and each span calibration turned on/off
sequentially at the set auto calibration
timing.
Auto calibration cycle setting:
Auto calibration cycle is set.
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day).
Gas flow time setting:
The time for flowing each calibration gas
in auto calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
Auto calibration remote start:
Auto calibration starts by opening the auto
calibration remote start input terminal
after short circuiting for 1.5 sec or longer.
Auto calibration starts when contacts
open.
Auto zero calibration:
Auto zero calibration is carried out periodically at the preset cycle.
This cycle is independent on “Auto calibration” cycle.
When zero calibration gas and solenoid
valve for opening/closing the calibration
gas flow line are prepared externally by
the customer, zero calibration will be
carried out with the solenoid valve drive
contact for zero calibration turned on/off
at the set auto zero calibration timing.
Auto zero calibration cycle setting:
Auto zero calibration cycle is set.
Setting is variable within 1 to 99 hours
(in increments of 1 hour) or 1 to 40 days
(in increments of 1 day)
Gas flow time setting:
The timing for flowing zero gas in auto
zero calibration is set.
Settable within 60 to 900 seconds (in
increments of 1 second)
High/low limit alarm:
Alarm contact output turns on when
measurement value reaches the preset
high or low limit alarm value.
Contacts close when the instantaneous
value of each channel exceeds the high
alarm limit value or falls below the low
alarm limit value.
Instrument error contact output:
Contacts turn on at occurrence of analyzer
error No. 1, 2, 3 or 10.
Calibration error contact output:
Contacts turn on at occurrence of manual
or auto calibration error (any of errors No.
4 to 9).
Auto calibration status contact outputs:
Contacts turn on during auto calibration.
O 2 correction:
Correction of measured NO, SO 2 and CO
gas concentrations into values at reference O 2 concentration
Correction formula:
C=
21-On
21-Os
x Cs
C :Sample gas concentration after O 2
correction
C S :Measured concentration of sample
gas
O S :Measured O 2 concentration (Limit
setting: 1 to 20% O 2 )
O n :Reference O 2 concentration
(value changeable by setting.0 to
19% O 2 )
Average value after O 2 correction and O 2 average value
calculation:
The result of O 2 correction or instantaneous O 2 value can be output as an
average value over the preset period of
time.
Used for averaging is the moving average
method in which sampling is carried out
at intervals of 30 seconds.
(Output is updated every 30 seconds. It
is the average value in the determined
period of time just before the latest updating.)
Averaging time is settable within 1 to 59
minutes (in increments of 1 minute) or 1
to 4 hours (in increments of 1 hour).
Average value resetting:
The above-mentioned output of average
value is started from the initial state by
opening the average value resetting input
terminals after short circuiting for 1.5 sec
or longer.
Output is reset by input voltage and
restarted by opening
Communication function:
RS-485 (9pins D-sub connector)
Half-duplex bit serial
Start-stop synchronization
ModbusTM protocol
Contents :Read/ Write parameters
Read measurement concentration and instrument status
Remark : W hen connecting via RS 232C interface, an RS-232C
↔ RS-485 converter should
be used.
Atmospheric pressure correction:
Measure atmospheric pressure and calculate compensation (for use, be sure to
relieve the exhaust gas from analyzer to
the atmosphere)
After atmospheric pressure correction;
Zero point:No influenced
Span point:The change is 0.5%%measured value or less relating
to the change of the atmospheric pressure 1%.
Correction range: 700hPa-1050hPa
3
ZPA
Performance
±0.5% of full scale
1% of full scale
prior to atmospheric pressure correction
(option)
Zero drift:
±2% of full scale/week
In the case of Auto zero calibration use
for 500ppm or less range.
Span drift:
±2% of full scale/week
Response time (for 90% FS response) :
1 to 15 sec electrical response. Within
10-30 seconds including replacement time
of sampling gas.
Gas replacement time depends on the
number of measuring components,and
measuring range.
Interference from other gases:
Repeatability:
Linearity:
Interference
component
NO
CO2
CO
CH4
SO2
analyzer analyzer analyzer analyzer analyzer
CO
1000ppm
≤1.0%FS
CO2 15%
H2O saturation
at 20°C
H2O saturation
at 2°C
CH4
1000ppm
––
≤1.0%FS ≤1.0%FS ≤1.0%FS
––
*1)
*2)
≤1.0%FS ≤1.0%FS ≤1.0%FS ≤1.0%FS
*2)
≤1.0%FS ≤1.0%FS ≤1.0%FS
––
≤2.0%FS
≤1.0%FS ≤1.0%FS
––
––
––
≤2.0%FS ≤2.0%FS
––
≤20ppm
––
Corrosive component: 1 ppm or less
Standard gas for calibration:
1))Infrared-ray measurable component, standard O 2
Zero gas ;Dry N2
Span gas ; Each sample gas having concentration 90 to 100% of its
measuring range (recommended).
In case a zirconia O2 analyzer is installed
externally and calibration is carried out on
the same calibration gas line:
Zero gas ;Dry air or atmospheric air (Do
not use with CO2 measurement)
Span gas ;For other than O2 measurement, each sample gas having concentration 90 to 100%
of its measuring range
For O2 measurement, O2 gas
of 1 to 2 vol%/remains N2 gas
2))Reverse range O2 measurement
Zero gas ;100% O2
Span gas ;For O2 measurement, O2 gas
of 95.0%~95.5%/remains N2
gas
*For reverse range O 2 measurement,
infrared measurable component is not
measurable.
*1) 0-200ppm range ≤ 2.0%FS
*2) 0-500ppm range or less ≤ 2.0%FS
Installation Requirements
EC Directive Compliance
The product conforms to the requirements of the Low Voltage Directive 2006/95/EC and EMC directive 89/336/EEC
(as amended by Directive 92/31/EEC), both as amended by
Directive 93/68/EEC.
It conforms to following standards for product safety and
electromagnetic compatibility ;
EN61010-1 : 2010Safety requirements for electrical equipment for measurement, control and laboratory
use.
“Installation Category II”
“Pollution Degree 2”
EN61326-1 : 2006
Electrical equipment for measurement, control
and laboratory use — EMC requirements.
Standard Requirements for Sample Gas
Flow rate:
Temperature:
Pressure:
Dust:
Mist:
Moisture:
4
0.5L / min ±0.2L / min
0 to 50˚C
10 kPa or less (Gas outlet side should be
open to the atmospheric air.)
100 µg/Nm3 or less in particle size of 0.3
µm or less
Unallowable
Below a level where saturation occurs at
room temperature (condensation unallowable).
Below the level where saturation occurs
at 2˚C for CO measurement in 0 to 200
ppm range, NO measurement, and SO 2
measurement.
•Indoor use (Select a place where the equipment does
not receive direct sunlight, draft/rain or radiation from hot
substances. If such a place cannot be found, a roof or
cover should be prepared for protection.)
•Avoid a place where unit receives heavy vibration
•Select a place where atmospheric air is clean
Principle diagram of NDIR type measurement (For NO, SO2, CO2, CO, CH4)
Front expansion chamber
Gas inlet
Infrared-ray
light source
Gas outlet
Rear
expansion
chamber
Detector
Motor
Mass flow
sensor
Measuring cell
Chopper
Preamplifier
Signal
processing
and calculation
block
Indication
Output
Principle diagram of fuel cell type measurment (For O2)
Resistor
Thermistor
Electrolyte
Negative
electrode
Diaphragm
Positive
electrode
Principle diagram of paramagnetic type measurment (For O2)
Permanent magnet
Mirror
Dumb-bell
Gas inlet
Measuring cell
Magnetic field
Gas outlet
Preamplifier
Permanent
magnet
Photo diode
Light emitting diode
Indication
Signal processing and
calculation block
Output
5
ZPA
Examples of sampling system configuration including gas analyzer
To measure low moisture content (saturated at room temperature or lower) sample gas (CO, CO2, CH4)
9
Mist filter
Sample
gas inlet
1
Flow meter
10
Pump
3
Two-way
Solenoid
valve
7
7
7
7
Gas
analyzer
(ZPA)
Solenoid
valve
Drain pot
5
11
Membrane
filter
Demister
O2 analyzer
6
Ball valve
6 Ball valve
(In the case of zirconia type)
Exhaust
(atmospheric pressure)
Drain
Reference gas
for zero
calibration
8
Reference gas for
span calibration
To measure high moisture content sample gas, NO, SO2, or CO (0 to 200 ppm range)
13
1
Mist filter
3
Pump
Sample
gas inlet
3
Solenoid
valve
9
10
7
Pump
Atmospheric
(Air)
*1)
2
4
Safety
drain trap
Flow meter
Two-way
Solenoid
valve
12
7
7
7
Electronic
cooler
5
6
NOX/NO converter
(for NOX measurement)
Solenoid
valve
Drain
pot
Ball
valve
Drain
Membrane
filter
Gas
analyzer
(ZPA)
O2 analyzer
Reference gas
for zero
calibration
8
Reference gas for
span calibration
(In the case of
zirconia type)
Exhaust
(atmospheric
pressure)
*1) Be sure to use a dehumidifier such as electronic cooler for NO, SO2, and CO
analyzers of 0 to 200 ppm range (=2ºC saturation or lower).
List of sampling devices (example)
No.
Device name
Fuji’s type
1
Mist filter
ZBBK1V03-0
2
Safety drain trap
ZBH51603
3
Pump
ZBG80
4
Electoric cooler
ZBC91004
5
Drain pot
ZBH13003 (Length 255mm)
6
Ball valve
ZBFB1
7
Two-way solenoid valve
8
Standard gas for calibration
ZBM
9
Flow meter
ZBD42203
10
Membrane filter
ZBBM2V03-0
11
Demister
ZBH35003
12
NO2/NO converter
ZDL02001
13
Y04-0
(Codes in
to be selected depending on application)
Three-way solenoid valve
Note) The above is a typical configuration example. As configuration may differ depending on measuring objects,
please consult us.
6
CODE SYMBOLS
standard multi components
Description
Digit
4
<Specification/Structure>
Horizontal type (Terminal block for power supply)
Horizontal type (Power inlet, with lock)
5
<Mounting>
19inch rack mounting type EIA conformity
6
<Measurable component (NDIR)>
1st component 2nd component 3rd component 4th component
None
NO
SO2
CO2
CO
CH4
NO
SO2
NO
CO
CO2
CO
CH4
CO
CO2
CH4
NO
SO2
CO
CO
CH4
CO2
NO
SO2
CO2
CO
Others
7
<Measurable component (O2)>
None
External O2 analyzer
External zirconia O2 sensor (ZFK7)
Built-in fuel cell O2 sensor
Built-in paramagnetic O2 sensor
8
<Revision code>
9
<Measuring range (NDIR)>1st component, 1st range
10 <Measuring range (NDIR)>1st component, 2nd range
11 <Measuring range (NDIR)>2nd component, 1st range
12 <Measuring range (NDIR)>2nd component, 2nd range
13 <Measuring range (NDIR)>3rd component, 1st range
14 <Measuring range (NDIR)>3rd component, 2nd range
15 <Measuring range (NDIR)>4th component, 1st range
16 <Measuring range (NDIR)>4th component, 2nd range
17 <Measuring range (O2)>
None
0-5/10%
0-5/25%
0-10/25%
0-5%
0-10%
0-25%
0-50%
0-100%
100-95%
Others
18 <Gas connection>
Rc1/4
NPT1/4
19 <Output>
DC0-1V
DC4-20mA
DC0-1V+Communication function
DC4-20mA+Communication function
20 <Indication/power supply cable>
Japanese (PSE)
English (UL)
English (CEE)
Chinese (CCC)
21 <O2 correction and O2 correction average output>
None
O2 correction
O2 correction average
O2 correction and O2 correction average
22 <Optional function (DIO)>
FAULT A. Cal. H/L Alarm RangeID/Remote range
None
1 2 3 4 5 6 7 8
note
note1
ZPA
B
9 10 11 12 13
14 15 16 17 18 19 20
21 22 23 24 25
1
Digit
A
D
B
note2
note3
note4
note4
note4
note4
note4
note4
note4
note4
note5
Y
P
A
D
B
E
F
G
J
K
L
N
T
V
Z
Y
1
2
3
4
1
Y
A
B
C
L
M
V
P
R
S
Z
1
2
A
B
C
D
note6
note6
note6
note6
note7
J
E
U
C
Y
A
B
C
note8
Y
A
B
C
D
E
F
G
H
7
ZPA
1 2 3 4 5 6 7 8
9 10 11 12 13
14 15 16 17 18 19 20
21 22 23 24 25
1
Digit
Y
1
A
B
note9
A
C
D
Z
note10
RANGE CODE
Range
None
0~100ppm
0~200ppm
0~250ppm
0~300ppm
0~500ppm
0~1000ppm
0~2000ppm
0~2500ppm
0~3000ppm
0~5000ppm
ZRE
note
Description
Digit
23 <Pressure compensation>
None
Pressure compensation
24 <Un>
ppm, Vol%
mg/m3, g/m3
25 <Adjustment>
For standard
For heat treatment furnace
For converter
Others
O2 measurement range
Code
Y
B
C
D
S
E
F
G
U
T
H
Range
0~1%
0~2%
0~3%
0~5%
0~10%
0~20%
0~25%
0~40%
0~50%
0~70%
0~100%
Others
Code
J
K
Q
L
M
N
V
W
P
X
R
Z
Measurement Range Fuel cell O2 Paramagnet Zirconia O2
range
code sensor
ic O2 sensor sensor
(external)
(built - in) (built - in)
A
0~5/10 vol%
0~5/25 vol%
B
0~10/25 vol%
C
0~5 vol%
L
0~10 vol%
M
0~25 vol%
V
0~50 vol%
P
0~100 vol%
R
100~95 vol%
S
note1)When “D” is specified at 4th digit, Power cable is supplied in the scope of supply. Cable specification should be
specified at the 20th digit.
note2)When only O2 measurement is necessary, "Y" should be specified at 6th digit.
note3)When "1" is specified at 7th digit, O 2 pt sensor signal has to be set as 0-1V DC linear corresponding to full scale.
External zirconia O2 sensor and external O2 analyzer are not included in the scope of supply, and has to be separately
ordered.
note4)Refer to Table 1 for possible combination of measuring components and ranges in the data sheet. When "Y" is specified
at 6th digit, "Y" should be specified at 9th to 16th digit. For fuel cell O 2 sensor, range is 0-10% or more.
note5)Select the type of voltage rating, plug type and applicable standard of the power cable by 20th digit. Select a power
cable for using at the location of end-user. When “A” is specified at 4th digit, Power cable will not be included in scope
of delivery. When English is specified for display language, "E" should be selected at 20th digit.
note6)O2 correction is calculated only for NO, SO 2 and CO.
note7)When 5 components measurement is specified, "H" must not be specified at 22nd digit.
When 4 components measurement is specified and “H” is specified at 22nd digit, 3 points is maximum for alarm
output function.
note8)When "B" is specified at 24th digit, measuring range should be specified by ppm range code. In this case NO, SO 2 and
CO measuring range are corresponding range in mg/m 3. Please refer to the table shown below for the corresponding
range code based on "mg/m 3".
note9)When A to D is specified at 25th digit, the analyzer will be adjusted and delivered with the following gasses. Standard
"A": balance gas N2. For heat treatment furnace "C": balance gas 30% H 2/remamins N2.
For converter "D": balance gas CO, CO 2.
When other adjustment is required, please specify "Z".
When "Z" is specified, please attach a list of gas composition contained in the measuring gas.
Corresponding mg/m3
Range code
C
D
S
E
F
G
U
T
H
8
Unit : ppm
0-200ppm
0-250ppm
0-300ppm
0-500ppm
0-1,000ppm
0-2,000ppm
0-2,500ppm
0-3,000ppm
0-5,000ppm
Corresponding range in mg/m3
NO
SO2
CO
0-260mg/m3
0-250mg/m3
0-570mg/m3
0-325mg/m3
0-700mg/m3
0-300mg/m3
0-400mg/m3
0-850mg/m3
0-375mg/m3
0-650mg/m3
0-1,400mg/m3
0-600mg/m3
0-1,300mg/m3
0-2,800mg/m3
0-1,250mg/m3
0-2,600mg/m3
0-5,600mg/m3
0-2,500mg/m3
0-3,300mg/m3
0-7,100mg/m3
0-3,000mg/m3
0-4,000mg/m3
0-8,500mg/m3
0-3,750mg/m3
0-6,600mg/m3
0-14.00g/m3
0-6,250mg/m3
The conversion formula "ppm" unit
into "mg/m3" unit.
NO (mg/m3) = 1.34 × NO (ppm)
SO2 (mg/m3) = 2.86 × SO2 (ppm)
CO (mg/m3) = 1.25 × CO (ppm)
Table 1 Measurable component and range - availability check table Procedure of range selection
On one component analyzer:
First determine 1st range, then select 2nd range from the right of your determination range in following tables.
More than two components:
The procedure is same as one component. Difference is 2nd range in the tables.
The 2nd range in the tables for two and more components is maximum available range.
2nd range is selectable from 1st range to 2nd range(max) on the table.
1-component analyzer : CO
1st range
0 - 200ppm
0 - 250ppm
0 - 300ppm
0 - 500ppm
0 - 1000ppm
0 - 2000ppm
0 - 2500ppm
0 - 3000ppm
0 - 5000ppm
0 - 1%
0 - 2%
0 - 3%
0 - 5%
0 - 10%
0 - 20%
0 - 25%
0 - 40%
0 - 50%
0 - 70%
0 - 100%
1-component analyzer : NO
2nd range
None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
None, 0 - 3000ppm,5000ppm,1%,2%
None, 0 - 5000ppm,1%,2%
None, 0 - 1%,2%,3%,5%
None, 0 - 2%,3%,5%,10%
None, 0 - 3%,5%,10%,20%
None, 0 - 5%,10%,20%,25%
None, 0 - 10%,20%,25%,40%,50%
None, 0 - 20%,25%,40%,50%,70%,100%
None, 0 - 25%,40%,50%,70%,100%
None, 0 - 40%,50%,70%,100%
None, 0 - 50%,70%,100%
None, 0 - 70%,100%
None, 0 - 100%
None
1-component analyzer : CO2
1st range
0 - 100ppm
0 - 200ppm
0 - 250ppm
0 - 300ppm
0 - 500ppm
0 - 1000ppm
0 - 2000ppm
0 - 2500ppm
0 - 3000ppm
0 - 5000ppm
0 - 1%
0 - 2%
0 - 3%
0 - 5%
0 - 10%
0 - 20%
0 - 25%
0 - 40%
0 - 50%
0 - 70%
0 - 100%
2nd range
None, 0 - 200ppm,250ppm,300ppm,500ppm,1000ppm
None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
None, 0 - 3000ppm,5000ppm,1%,2%
None, 0 - 5000ppm,1%,2%
None, 0 - 1%,2%,3%,5%
None, 0 - 2%,3%,5%,10%
None, 0 - 3%,5%,10%,20%
None, 0 - 5%,10%,20%,25%
None, 0 - 10%,20%,25%,40%,50%
None, 0 - 20%,25%,40%,50%,70%,100%
None, 0 - 25%,40%,50%,70%,100%
None, 0 - 40%,50%,70%,100%
None, 0 - 50%,70%,100%
None, 0 - 70%,100%
None, 0 - 100%
None
2-component analyzer : NO/SO2
1-component : NO
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
1st range
0 - 200ppm
0 - 250ppm
0 - 300ppm
0 - 500ppm
0 - 1000ppm
0 - 2000ppm
0 - 2500ppm
0 - 3000ppm
0 - 5000ppm
2nd range
None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2500ppm,3000ppm,5000ppm
None, 0 - 3000ppm,5000ppm
None, 0 - 5000ppm
None
1-component analyzer : SO2
1st range
0 - 200ppm
0 - 250ppm
0 - 300ppm
0 - 500ppm
0 - 1000ppm
0 - 2000ppm
0 - 2500ppm
0 - 3000ppm
0 - 5000ppm
0 - 1%
0 - 2%
0 - 3%
0 - 5%
0 - 10%
2nd range
None, 0 - 250ppm,300ppm,500ppm,1000ppm,2000ppm
None, 0 - 300ppm,500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 500ppm,1000ppm,2000ppm,2500ppm
None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
None, 0 - 3000ppm,5000ppm,1%,2%
None, 0 - 5000ppm,1%,2%
None, 0 - 1%,2%,3%,5%
None, 0 - 2%,3%,5%,10%
None, 0 - 3%,5%,10%
None, 0 - 10%
None, 0 - 10%
None
1-component analyzer : CH4
1st range
0 - 500ppm
0 - 1000ppm
0 - 2000ppm
0 - 2500ppm
0 - 3000ppm
0 - 5000ppm
0 - 1%
0 - 2%
0 - 3%
0 - 5%
0 - 10%
0 - 20%
0 - 25%
0 - 40%
0 - 50%
0 - 70%
0 - 100%
2nd range
None, 0 - 1000ppm,2000ppm,2500ppm,3000ppm,5000ppm
None, 0 - 2000ppm,2500ppm,3000ppm,5000ppm,1%
None, 0 - 2500ppm,3000ppm,5000ppm,1%,2%
None, 0 - 3000ppm,5000ppm,1%,2%
None, 0 - 5000ppm,1%,2%
None, 0 - 1%,2%,3%,5%
None, 0 - 2%,3%,5%,10%
None, 0 - 3%,5%,10%,20%
None, 0 - 5%,10%,20%,25%
None, 0 - 10%,20%,25%,40%,50%
None, 0 - 20%,25%,40%,50%,70%,100%
None, 0 - 25%,40%,50%,70%,100%
None, 0 - 40%,50%,70%,100%
None, 0 - 50%,70%,100%
None, 0 - 70%,100%
None, 0 - 100%
None
2-component analyzer : NO/CO
2-component : SO2
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
• The 2nd component should
be selected as shown in the
right table.
1-component : NO
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
2-component : CO
1st range 2nd range (max.)
0 - 200ppm 0 - 2000ppm
0 - 250ppm 0 - 2500ppm
0 - 300ppm 0 - 2500ppm
0 - 500ppm 0 - 5000ppm
0 - 1000ppm 0 - 5000ppm
0 - 2000ppm 0 - 5000ppm
0 - 2500ppm 0 - 5000ppm
0 - 3000ppm 0 - 5000ppm
0 - 5000ppm None
• The 2nd component should
be selected as shown in the
right table.
9
ZPA
2-component analyzer: CO2/CO
2-component: CO
1-component: CO2
1st range 2nd range (max.) 1st range/2nd range (max.)
0-100ppm
0-1000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm
0-200ppm
0-2000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%,
0-250ppm
0-2500ppm
0-3%
0-300ppm
0-500ppm
0-500ppm
0-5000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/3%, 0-3%
0-1000ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-2000ppm
0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-1000ppm
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-2000ppm
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-2000ppm
0-2%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-2500ppm
0-1%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-2500ppm
0-2%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%,
0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-3000ppm
0-1%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%,
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-3000ppm
0-2%
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/25%,
0-10/100%,0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5000ppm
0-3%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-1%
0-5%
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-5%
0-5000ppm
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-3%
0-25%
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5%
0-50%
0-10%
0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,
0-20%
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-25%
0-40%
0-50%
0-70%
0-100%
None
2-component analyzer: CH4/CO
2-component: CO
1-component: CH4
1st range 2nd range (max.) 1st range/2nd range (max.)
0-500ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm
0-1000ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm
0-1000ppm
0-1%
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm
0-2000ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%,
0-3/20%, 0-5/20%, 0-10/20%
0-2500ppm
0-5000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% 0-3000ppm/1%,
0-3000ppm
0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-2000ppm
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%
0-2500ppm
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-3000ppm
0-2000ppm
0-2%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%
0-2500ppm
0-2%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/25%
0-3000ppm
0-5000ppm
0-1%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%,
0-25/50%, 0-40/50%, 0-50%
0-5000ppm
0-3%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-5000ppm
0-5%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5% 0-2/5%, 0-3/25%, 0-5/25%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%
0-1%
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-1%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1% ,0-3000ppm/1%, 0-5000ppm/1%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-3%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-3%
0-25%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-5%
0-25%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-5%
0-50%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/5%, 0-2/5%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-10%
0-50%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-10%
0-100%
0-5000ppm/5%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-20%
0-50%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%,
0-25%
0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-40%
0-20%
0-100%
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-25%
0-40%
0-50%
0-70%
0-100%
None
10
2-component analyzer: CO2/CH4
2-component: CH4
1-component: CO2
1st range 2nd range (max.) 1st range/2nd range (max.)
0-100ppm
0-1000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm
0-200ppm
0-2000ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/1%, 0-5000ppm/1%, 0-1%
0-250ppm
0-2500ppm
0-300ppm
0-2500ppm
0-500ppm
0-2500ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%
0-500ppm
0-5000ppm
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/2%, 0-2/10%, 0-3/10%, 0-5/10%, 0-10%
0-1000ppm
0-2500ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-1000ppm
0-5000ppm
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3% 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-1000ppm
0-1%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/3%, 0-1/3%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2000ppm
0-2500ppm
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2000ppm
0-5000ppm
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2000ppm
0-2%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/20%, 0-20%
0-2500ppm
0-5000ppm
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-2500ppm
0-2%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5%, 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-3000ppm
0-2%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/5% 0-2/20%, 0-3/20%, 0-5/25%, 0-10/25%, 0-20/25%, 0-25%
0-5000ppm
0-3%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-5000ppm
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/20%, 0-5/20%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-1%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%,
0-40/50%, 0-50%
0-2%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-10%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-3%
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-3%
0-25%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-5%
0-50%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%,
0-50/100%, 0-70/100%, 0-100%
0-10%
0-20%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%,
0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-10%
0-50%
0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%,
0-20%
0-50/100%, 0-70/100%, 0-100%
0-25%
0-40%
0-10%
0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/10%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-70/100%, 0-100%
0-20%
0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/10%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-70/100%, 0-100%
0-25%
0-100%
0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/2%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%,
0-40%
0-70/100%, 0-100%
0-50%
0-70%
0-100%
None
3-component analyzer: NO/SO2/CO >>> Combination of 1st component NO and 2nd component SO2 / 3rd component CO
2-component: SO2
3-component: CO
1-component: NO
1st range 2nd range (max.)
1st range 2nd range (max.) 1st range/2nd range (max.)
0-200ppm
0-2000ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm
0-200ppm
0-2000ppm
0-250ppm
0-2500ppm
0-250ppm
0-2500ppm
0-300ppm
0-300ppm
0-2500ppm
0-500ppm
0-2500ppm
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-5000ppm
0-500ppm
0-5000ppm
0-1000ppm
0-1000ppm
0-5000ppm
0-2500ppm
None
0-2000ppm
0-5000ppm
0-1000ppm
0-5000ppm
0-500/5000ppm, 0-1000/5000ppm, 0-2000/5000ppm, 0-2500/5000ppm, 0-3000/5000ppm, 0-5000ppm
0-2500ppm
0-5000ppm
0-2000ppm
0-3000ppm
0-5000ppm
0-2500ppm
0-5000ppm
None
0-3000ppm
0-5000ppm
None
3-component analyzer: CO2/CO/CH4 >>> Combination of 1st component CO2 / 2nd component CO and 3rd component CH4
1-component: CO2
2-component: CO
3-component: CH4
1st range 2nd range (max.) 1st range/2nd range (max.)
1st range 2nd range (max.)
0-5000ppm
0-5%
0-5000ppm
0-3%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-1%
0-5%
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-5%
0-5000ppm
0-5%
0-500/5000ppm, 0-1000ppm/1%, 0-2000ppm/2%, 0-2500ppm/2%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-1%
0-10%
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-2%
0-20%
0-1%
0-10%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3/25%, 0-5/25%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-3%
0-25%
0-2%
0-20%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-5%
0-10%
0-3/25%, 0-5/50%, 0-10/50%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-10%
0-20%
0-2%
0-10%
0-500/5000ppm , 0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%, 0-5000ppm/5%, 0-1/10%, 0-2/20%,
0-3%
0-25%
0-3/25%, 0-5/50%, 0-10/100%, 0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-20%
0-25%
0-5%
0-50%
0-25%
0-40%
0-10%
0-100%
0-1000ppm/1%, 0-2000ppm/1%, 0-2500ppm/1%, 0-3000ppm/2%,
0-40%
0-50%
0-20%
0-5000ppm/5%, 0-1/10%, 0-2/20%, 0-3/25%, 0-5/50%, 0-10/100%,
0-25%
0-20/100%, 0-25/100%, 0-40/100%, 0-50/100%, 0-70/100%, 0-100%
0-50%
0-70%
0-40%
0-50%
0-70%
0-100%
None
0-70%
0-100%
0-100%
None
Availability of product
Product available only
when CO analyzer max.
measuring range is
50% or less
Product available
Product available only
when CO analyzer
measuring range is 0 to
1000ppm or more.
Product available only
when CO analyzer
measuring range is 0 to
5000ppm or more.
Product available only
when CO analyzer
measuring range is more
than 5000ppm or CO2
analyzer range is more
than 2%.
11
ZPA
4-component analyzer: NO/SO2/CO2/CO >>> Combination of 1st component NO /4th component CO and component 2nd component SO2/3rd component CO2
1-component: NO
1st range 2nd range (max.)
0-200ppm
0-2000ppm
0-250ppm
0-2500ppm
0-300ppm
0-2500ppm
0-500ppm
0-2000ppm
0-1000ppm
0-2000ppm
0-2000ppm
None
0-500ppm
0-5000ppm
0-1000ppm
0-5000ppm
0-2000ppm
0-5000ppm
0-2500ppm
0-5000ppm
0-3000ppm
0-5000ppm
0-5000ppm
None
4-component: CO
1st range/2nd range (max.)
0-200/2000ppm, 0-250/2500ppm, 0-300/2500ppm, 0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None
0-500/2500ppm, 0-1000/2500ppm, 0-2000/2500ppm, 0-2500ppm, None
2-component analyzer: SO2 3-component analyzer: CO2
1st range 2nd range (max.) 1st range/2nd range (max.)
0-200ppm
0-2000ppm
0-250ppm
0-2500ppm
0-300ppm
0-2500ppm
0-500ppm
0-5000ppm
0-1/10%, 0-2/20%, 0-3/20%, 0-5/50%, 0-10/50%, 0-20/50%, 0-25/50%, 0-40/50%, 0-50%/None
0-1000ppm
0-5000ppm
0-2000ppm
0-5000ppm
0-2500ppm
0-5000ppm
0-3000ppm
0-5000ppm
0-5000ppm
None
Table 2 Channel (Ch) No. and display/output contents comparison table
Code symbol
6th digit 7th digit 21st digit
Y
1 to 3
Y
P
Y
Y
A
Y
Y
D
Y
Y
B
Y
Y
E
Y
Y
F
Y
Y
G
Y
Y
J
Y
Y
K
Y
Y
L
Y
Y
N
Y
Y
T
Y
Y
V
Y
Y
P
1 to 3
Y
A
1 to 3
Y
D
1 to 3
Y
B
1 to 3
Y
E
1 to 3
Y
F
1 to 3
Y
G
1 to 3
Y
J
1 to 3
Y
K
1 to 3
Y
L
1 to 3
Y
N
1 to 3
Y
T
1 to 3
Y
V
1 to 3
Y
P
1 to 3
A*
A
1 to 3
A*
B
1 to 3
A*
F
1 to 3
A*
G
1 to 3
A*
J
1 to 3
A*
N
1 to 3
A*
V
1 to 3
A*
P
1 to 3
C *
A
1 to 3
C *
B
1 to 3
C *
F
1 to 3
C *
G
1 to 3
C *
J
N
1 to 3
1 to 3
C *
C *
V
1 to 3
C *
Display/output contents
Ch1:O2
Ch1:NO
Ch1:SO2
Ch1:CO2
Ch1:CO
Ch1:CH4
Ch1:NO, Ch2:SO2
Ch1:NO, Ch2:CO
Ch1:CO2, Ch2:CO
Ch1:CH4, Ch2:CO
Ch1:CO2, Ch2:CH4
Ch1:NO, Ch2:SO2, Ch3:CO
Ch1:CO2, Ch2:CO, Ch3:CH4
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO
Ch1:NO, Ch2:O2
Ch1:SO2, Ch2:O2
Ch1:CO2, Ch2:O2
Ch1:CO, Ch2:O2
Ch1:CH4, Ch2:O2
Ch1:NO, Ch2:SO2, Ch3:O2
Ch1:NO, Ch2:CO, Ch3:O2
Ch1:CO2, Ch2:CO, Ch3:O2
Ch1:CH4, Ch2:CO, Ch3:O2
Ch1:CO2, Ch2:CH4, Ch3:O2
Ch1:NO, Ch2:SO2, Ch3:CO, Ch4:O2
Ch1:CO2, Ch2:CO, Ch3:CH4, Ch4:O2
Ch1:NO, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2
Ch1:NOx, Ch2:O2, Ch3:corrected NOx
Ch1:SO2, Ch2:O2, Ch3:corrected SO2
Ch1:CO, Ch2:O2, Ch3:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO
Ch1:NOx, Ch2:O2, Ch3:corrected NOx, Ch4:corrected NOx average
Ch1:SO2, Ch2:O2, Ch3:corrected SO2, Ch4:corrected SO2 average
Ch1:CO, Ch2:O2, Ch3:corrected CO, Ch4corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:O2, Ch4:corrected NOx, Ch5:corrected SO2, Ch6:corrected NOx average,
Ch7:corrected SO2 average
Ch1:NOx, Ch2:CO, Ch3:O2, Ch4:corrected NOx, Ch5:corrected CO, Ch6:corrected NOx average,
Ch7:corrected CO average
Ch1:CO2, Ch2:CO, Ch3:O2, Ch4:corrected CO, Ch5:corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:CO, Ch4:O2, Ch5:corrected NOx, Ch6:corrected SO2, Ch7:corrected CO,
Ch8:corrected NOx average, Ch9:corrected SO2 average, Ch10:corrected CO average
Ch1:NOx, Ch2:SO2, Ch3:CO2, Ch4:CO, Ch5:O2, Ch6:corrected NOx, Ch7:corrected SO2, Ch8:corrected CO,
Ch9:corrected NOx average, Ch10:corrected SO2 average2, Ch11:corrected CO average
* When the 21st digit code is A or C, the component of the NO analyzer is displayed as NOx.
12
OUTLINE DIAGRAMS (Unit : mm)
Refer to appendix. ((TC drawing)
429
57.2
<FRONT VIEW>
132.5
7
<TOP VIEW>
415
463
483
Power Switch
Analog Output Connctor (A/O)
Digital Input/output Connector (DIO1 to 3)
Purge Gas Inlet Rc1/4 or NPT1/4
Communication Connector (RS485)
Sample Gas Inlet Rc1/4 or NPT1/4
Analog Input Connector (A/I)
Sample Gas Outlet Rc1/4 or NPT1/4
Fuse
<REAR VIEW>
(M4 Terminal Block Type)
124.5
Power Inlet(100V to 240V AC 50/60Hz)
<SIDE VIEW>
24
2
25
380
396
16
(M4 Terminal Block Type)
(Power Inlet Type(100V to 240V AC 50/60Hz))
13
ZPA
Mounting method
The analyzer weight should be supported at the bottom of the case.
19-inch rack mounting type
57.2
<Rack dimensions>
450 or more
465
Mounting diagram
Guide rails
Guide rails
EXTERNAL CONNECTION
<Analog output> A/O connector
13
1
<RS485 communication signal>
1
14
2
25
14
15
3
D-sub 25pins female
16
4
* In standard, displayed Channel No.
and Analog Output No. are same.
17
5
18
6
<Screw terminal (M4)>
P
N
L
19
7
20
8
100 to 240V AC
21
9
22
10
<Power inlet>
23
11
24
12
POWER INLET
100-240V AC
25
13
14
AO1+
AO1AO2+
AO2AO3+
AO3AO4+
AO4AO5+
AO5AO6+
AO6AO7+
AO7AO8+
AO8AO9+
AO9AO10+
AO10AO11+
AO11AO12+
AO12NC
5
(GND)
1
1
6
RT×D+
2
9
6
7
RT×D-
3
D-sub 9pins female
8
4
9
5
<Analog input> A/I connector (O2 signal input)
AI-
AI+
<Digital I/O> DIO 1 to 3 connector (option)
13
DIO1
DIO2
DIO3
connector connector connector
1
1
14
25
2
14
15
3
D-sub 25pins female
16
4
* DIO 1 to 3 are all as same connector.
17
5
18
6
19
7
20
Contents of digital input
signal
DI1
Remote hold
DI2
Average value reset
DI3
A. cal. start
DI4
A. zero. cal. start
DI5
Remote range Ch1
DI6
Remote range Ch2
DI7
Remote range Ch3
DI8
Remote range Ch4
DI9
Remote range Ch5
8
21
9
22
10
23
11
NC
com
NO
NC
com
NO
NC
com
NO
NC
com
NO
NC
com
NO
DI1+
DI1DI2+
DI2DI3+
DI3-
DI4+
DI4DI5+
DI5DI6+
DI6-
DI7+
DI7DI8+
DI8DI9+
DI9-
DO1
DO6
DO11
DO2
DO7
DO12
DO3
DO8
DO13
DO4
DO9
DO14
DO5
DO10
DO15
Digital input
OFF : 0V
ON : 12 to 24V DC
Digital output
max. contact load
rating 24V DC/1A
24
12
25
13
Contents of digital output signal
Independent on the
number of component
1-component analyzer
22th digit
A, C
B, E
D, F, G, H
B, D, E, F, G, H
B, D, E, F, G, H
DO1
Instrument error
Instrument error
Instrument error
Instrument error
Instrument error
DO2
Calibration error
Calibration error
Calibration error
Calibration error
Calibration error
DO3
A.cal.status
(A.cal.status)
(A.cal.status)
(A.cal.status)
DO4
For zero gas
(For zero gas)
(For zero gas)
(For zero gas)
For span gas Ch1
(For span gas Ch1)
(For span gas Ch1)
(For span gas Ch1)
(For span gas Ch2)
(For span gas Ch2)
DO5
2-component analyzer 3-component analyzer
DO6
(Alarm1)
(Alarm1)
DO7
(Alarm2)
(Alarm2)
(For span gas Ch3)
DO8
(Alarm3)
(Alarm3)
(Range identification Ch1)
DO9
(Alarm4)
(Alarm4)
DO10
(Alarm5)
(Alarm5)
(Range identification Ch1)
(Range identification Ch2)
Range identification Ch1
(Range identification Ch2)
(Range identification Ch3)
DO11
(Alarm1)
(Alarm1)
(Alarm1)
DO12
(Alarm2)
(Alarm2)
(Alarm2)
DO13
(Alarm3)
(Alarm3)
(Alarm3)
DO14
(Alarm4)
(Alarm4)
(Alarm4)
DO15
(Alarm5)
(Alarm5)
(Alarm5)
4-component analyzer
The items in the parentheses
may not be available depending on the selected type on
22th digit.
The normal open side (NO) of
digital output is close when
the function is active without
range ID.
In case of range ID, normal
open (NO) side is close with
First range.
The normal close (NC) side is
close with Second range.
5-component analyzer
22th digit
B, E
D, F
G
H
B, E
D, F
G
DO1
Instrument error
Instrument error
Instrument error
Instrument error
Instrument error
Instrument error
Instrument error
DO2
Calibration error
Calibration error
Calibration error
Calibration error
Calibration error
Calibration error
Calibration error
DO3
A.cal.status
A.cal.status
A.cal.status
A.cal.status
A.cal.status
DO4
For zero gas
For zero gas
For zero gas
For zero gas
For zero gas
DO5
For span gas Ch1
For span gas Ch1
For span gas Ch1
For span gas Ch1
For span gas Ch1
DO6
For span gas Ch2
For span gas Ch2
For span gas Ch2
For span gas Ch2 Range identification Ch1
For span gas Ch2
DO7
For span gas Ch3 Range identification Ch1
For span gas Ch3
For span gas Ch3
For span gas Ch3 Range identification Ch2
For span gas Ch3
DO8
For span gas Ch4 Range identification Ch2
For span gas Ch4
For span gas Ch4
For span gas Ch4 Range identification Ch3
For span gas Ch4
DO9
Range identification Ch3
Range identification Ch1
For span gas Ch5 Range identification Ch4
For span gas Ch5
DO10
Range identification Ch4
Range identification Ch2
Range identification Ch5
DO11
(Alarm1)
(Alarm1)
(Alarm1)
(Alarm1)
(Alarm1)
Range identification Ch1
DO12
(Alarm2)
(Alarm2)
Range identification Ch1
(Alarm2)
(Alarm2)
(Alarm2)
Range identification Ch2
DO13
(Alarm3)
(Alarm3)
Range identification Ch2
(Alarm3)
(Alarm3)
(Alarm3)
Range identification Ch3
DO14
(Alarm4)
(Alarm4)
Range identification Ch3
Range identification Ch3
(Alarm4)
(Alarm4)
Range identification Ch4
DO15
(Alarm5)
(Alarm5)
Range identification Ch4
Range identification Ch4
(Alarm5)
(Alarm5)
Range identification Ch5
SCOPE OF DELIVERY
ORDERING INFORMATION
•Gas analyzer … 1 unit
•Replacement fuse (250V, 2A AC, delay type) … 2 pcs
•Instruction manual … 1 copy
•Connector for I/O connection … 1 set
•Power cable(standard inlet type 2m) … 1 pc
1. Code symbols
2.Application and composition of sample gas
15
ZPA
Gas inlet/outlet size:
Rc1/4 or NPT1/4
Power supply: Rated voltage
Exclusive Zirconia O2 Sensor
(to be purchased separately)
For O2 correction, the gas analyzer ZPA can accept linearized
0 to 1V DC signal coming from analyzer calibrated 0 to 25%
O2 full scale. If the analyzer is not available, Fuji can supply
exclusive Zirconia O2 sensor Model ZFK.
Measuring method:
Zirconia system
Measurable component and measuring range:
Measurable component Range
O 2
Oxygen
0 to 25vol%
Repeatability: Within ± 0.5% of full scale
Linearity:
Within ± 1% of full scale
Zero drift:
Within ± 1% of full scale/week
Span drift:
Within ± 2% of full scale/week
Response time: Approx. 20 seconds (for 90% response)
Measured gas flow rate:
0.5 ± 0.25L / min
Remark:The Zirconia system, due to its principle,
may produce a measuring error due to relative concentration versus the combustible
O2 gas concentration. Also, a corrosive
gas (SO2 of 250 ppm or more, etc.) may
affect the life of the sensor.
OUTLINE DIAGRAM
;100 to 115V AC or
200 to 240V AC
Rated frequency ;50Hz/60Hz
Max. rated power;215VA (during power
ON)
65VA (during steadystate operation)
Enclosure:
Steel casing, for indoor application
Indication:
Temperature indication (LED)
Temperature alarm output:
Contact output 1a contact,
Contact capacity 220V, 1A AC (resistive
load)
Outer dimensions (H x W x D):
141 x 170 x 190mm
Mass {weight}: Approx. 3kg
Finish color:
Munsell 5Y 7/1
CODE SYMBOLS
1 2 3 4 5 6 7 8
Z F K 7 Y Y
9 10 11 12 13
4 -
Measuring method
Zirconia method
7YY
Power supply
100 to 115V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(Standard)
200 to 240V AC 50/60Hz(CE mark)
9
B
C
(Unit:mm)
Description
Y 0 Y Y
Gas inlet/outlet size
Rc 1/4
NPT 1/4
1
8
170
4
152
OXYGEN ANALYZER
US
1
AC250V
T3.15A
2
3
4
5
6
7
E
F
110
131
141
TEMP.
L
N
AC
INPUT
OUT
ALM
9TH DIGIT
1: RC1/4
8: NPT1/4
ø5
EXTERNAL CONNECTION DIAGRAM
1
2
3
4
+
AC power supply
E
5
6
INLET
OUTLET
7
Output
Temperature
to analyzer alarm output
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fujielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan