SLPOWER MINT1180

Features

3” x 5” x 1.3” Package

180W Convection Cooled Power

Ideal for 1U Applications

Class I and Class II versions

Universal Input 90-264 VAC

Approved to CSA/EN/IEC/UL60601-1, 3rd
Edition, 2 x MOPP Isolation

Approved to CSA/EN/IEC/UL60950-1 2nd
Edition

Efficiency 90% typical

3 Year Warranty

RoHS Compliant
Description
The MINT1180 product family is the most cost-effective medically approved supply in the market with 180 watts
of convection cooled output power in 3”X5” footprint. The series comes in eight different models ranging from
12 to 48V and has been approved to IEC60601-1 3rd Edition, rated for 2 MOPP (Means of Patient Protection)
isolation. This new single output power supply family has both Class I and II input versions and accepts wide
input range of 90 to 264Vac. The typical 90% efficiency and convection cooled make this supply an excellent
solution for HomeCare applications where the use of fans is not desired.
Model Selection
Model Number
Volts
Output
Current*
Minimum Load
Ripple &
Noise**
Total
Regulation
OVP Threshold
MINT1180A1275K01
12V
15.0A
0A
120mV pk-pk
3%
14.0  1.1V
MINT1180A1575K01
15V
12.0A
0A
150mV pk-pk
3%
18.5 ± 1.2V
MINT1180A1875K01
18V
10.0A
0A
180mV pk-pk
3%
21.5 ± 2.0V
MINT1180A2475K01
24V
7.50A
0A
240mV pk-pk
3%
29.0 ± 2.5V
MINT1180A2875K01
28V
6.40A
0A
280mV pk-pk
3%
33.5 ± 2.5V
MINT1180A3275K01
32V
5.62A
0A
320mV pk-pk
3%
36.0 ± 3.0V
MINT1180A4875K01
48V
3.75A
0A
480mV pk-pk
3%
56.0 ± 3.0V
Notes:
* Total convection power is 180 Watts.
** Measured with noise probe directly across output terminals, and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
General Specifications
AC Input
Input Current
100-240Vac, ±10%, 47-63Hz, 1
120–370Vdc
Turn On Time
Less than 3 sec. @115Vac & Full Load
115Vac: 1.8A, 230Vac: 0.9A
Hold-up Time
>16mS at 180W, 120Vac
MINT1180
180W Single Output
Medical Series
General Specifications (continued)
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
Output Power
Transient
Response
Signals
264Vac, cold start: will not exceed 55A
Overload
Protection
F1, F2: 3.15A, 250VAC fuses provided on all
models
Short Circuit
Protection
<275μA@264Vac, 60Hz, NC; <400µA SFC
Overvoltage
Protection
88% typical
Switching
Frequency
180W convection cooled
500μS typ. for return to within 0.5% of
nominal, 50% load step. i/t <0.2A/µS.
Max Volt Deviation = 3%
Isolation
N/A
120%-150% of rating, Hiccup Mode
Provided - no damage will occur if the output is
shorted, auto recovery
OVP latch at 110%-130% of rated output
voltage.
PFC: Fixed, 65kHz
Main Converter: Variable 35-200kHz, 65-70kHz
at full load.
Input-Output: 4000Vac, 2 x MOPP
Input-Ground: 1800Vac, 1 x MOPP
Output-Ground: 1500Vac
Ripple and Noise
See chart
Operating
Temperature
-10°C to +70°C
Start Up at -40°C, full load
Output Voltage
See chart
Temperature
Derating
Derate output power linearly above 50°C to 50%
at 70°C
+/-5% from nominal
Storage
Temperature
-40°C to +85°C
Voltage
Adjustability
Minimum Load
Altitude
Not required
Total Regulation
+/- 3% combined line, load and initial setting.
Relative Humidity
Vibration
Operating: 0.003g²/Hz, 1.5grms overall, 3
axes, 10 min/axis
Non-Operating: 0.026g²/Hz, 5.0grms overall,
3 axes, 1 hr/axis
Shock
Dimensions
Weight
Safety Standards
W: 3.0” x L: 5.0” x H: 1.3”
MTBF
325g
Operating: -500 to 10,000 ft.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: Half-sine, 20gpk, 10ms, 3 axes, 6
shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3
axes, 6 shocks total
EN/CSA/UL/IEC 60601-1, 3rd Edition
EN/CSA/UL/IEC 60950-1, 2nd Edition
405,000 hours, 25°C, 110Vac
Output vs. Temperature Derating Curve
180W convection cooled, derate output power to 50% at 70C.
Output Power (W)
200
150
100
50
0
-10
0
Convection
www.slpower.com
10
20
30
40
50
60
70
Temperature (°C)
MINT1180 REV 1.4, 11-FEB-13
2