CUI ETMA1601250U

date
08/13/2012
page
1 of 5
SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
FEATURES
•
•
•
•
•
•
•
•
up to 200 W power
universal input (90~264 Vac)
single regulated output from 12~48 V
over voltage, overload and short circuit protections
full medical safety approvals
active PFC
level V efficiency
custom designs available
RoHS
®
IEC
60601-1
EN
60601-1
output
voltage
output
current
output
power
ripple1
(Vdc)
max
(A)
max
(W)
max
(mVp-p)
ETMA1201666U
12
16.66
200
240
V
ETMA1601250U
16
12.5
200
300
V
ETMA1901053U
19
10.53
200
300
V
MODEL
efficiency
level
ETMA2001000U
20
10
200
300
V
ETMA240833U
24
8.33
200
300
V
ETMA360555U
36
5.55
200
300
V
ETMA480416U
48
4.16
200
300
V
Notes:
1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with 10 μF aluminum electrolytic and 0.1 μF ceramic capacitors.
PART NUMBER KEY
ETMA1201666U C - P51 - WP - CXX
Base Number
example of 12 Vdc, 16.66 A
Reserved for Custom
Configurations
DC Plug Type
Input Cable:
C = included
blank = no cable
Factory
Designation
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 2 of 5
INPUT
parameter
conditions/description
min
voltage
frequency
current
at 90 Vac, 60 Hz
inrush current
at 230 Vac, cold start
max
units
90
typ
264
Vac
47
63
Hz
3
A
no load power consumption
150
A
0.5
W
max
units
OUTPUT
parameter
conditions/description
min
line regulation
±1
load regulation
hold-up time
typ
%
±5
at 115 Vac
%
16
temperature coefficient
ms
±0.05
%/°C
PROTECTIONS
parameter
conditions/description
over voltage protection
12 V output
16, 20 V outputs
19 V output
24, 36 V outputs
48 V output
110
110
111
110
110
133
130
137
125
120
%
%
%
%
%
overload protection
shutdown and auto restart
110
160
%
short circuit protection
continuous, auto restart
max
units
5,656
2,121
Vdc
Vdc
SAFETY & COMPLIANCE
parameter
conditions/description
isolation voltage
input to output for 2 seconds
input to frame ground for 2 seconds
min
typ
insulation resistance
input to output, at 500 Vdc
input to frame ground, at 500 Vdc
safety approvals
UL/cUL (UL 60601-1), EN 60601-1, IEC 60601-1
EMI/EMC
FCC 47 CFR Part 18, CE, EN 61000-3-(2, 3), EN 55011, IEC 61000-4-(2, 3, 4, 5, 6, 8, 11)
leakage current
earth, at 264 Vac
enclosure, at 264 Vac
MTBF
at 25°C, max. load
RoHS compliant
yes
20
20
MΩ
MΩ
190
10
300
100
140,000
μA
μA
hours
ENVIRONMENTAL
parameter
conditions/description
min
operating temperature
storage temperature
humidity
non-condensing
cui.com
max
units
0
typ
50
°C
-10
70
°C
20
90
%
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 3 of 5
MECHANICAL
parameter
conditions/description
dimensions
7.42 x 4.11 x 2.36 (188.5 x 104.5 x 60 mm)
min
input plug
IEC320 / C14
weight
typ
units: mm
104.5 0.5
60.0 0.5
188.5 0.5
cui.com
units
inch
1.06
MECHANICAL DRAWING
max
kg
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 4 of 5
DC CORD
PIN 2
Blue (GND)
PIN 4
Black (GND)
[2464#16AWG 4C+1(16AWG)]
PIN 3
Yellow (+Vo)
D.W. (GND)
SECTIONAL DRAWING
Core
14.2*8.0*28.5mm
TERMINAL
ø3.2*2
PIN 1
Red (+Vo)
4P POWER DIN
FRONT VIEW
UL Heat Shrinkable Tube
Black+Blue+D.W.
Yellow+Red
4PIN POWER DIN
COMPONENT LIST
1. Cable: color black, UL/CSA 2464 80/90°C 300V
16 AWG X 4C+Aluminum Shield (16AWG)
2. PVC Wire: color: blue/black/yellow/red
3. Connector: 4PIN power DIN or EQ
4. 4PIN power DIN (female) MDP-002-4PIN (P530600004F0001)
AC CORD
(IEC 320 C13)
(NEMA 5-15P Hospital Grade)
183±5cm
SPECIFICATION:
1)
Length: 1.83m
2)
Color: black
3)
Plug: SL-22 (A)10 (V)125 Approval: UL/cUL
4)
Connector: SL-3(A)10 (V)125 Approval: UL/cUL
5)
Cable: SJT 18AWG/3C Approval: UL/CSA
20±2cm
cui.com
CUI Inc │ SERIES: ETMA 200W │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
date 08/13/2012 │ page 5 of 5
REVISION HISTORY
rev.
description
date
1.0
initial release
03/05/2012
1.01
V-Infinity branding removed, safety and EMI/EMC data updated
08/13/2012
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
[email protected]
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to