SL Power MINT1065X1875C01 65w single output sery Datasheet

MINT1065
65W Single Output Series
Medical & ITE – Universal Input Power Supply
3 Year Warranty
•65W Convection Cooled AC-DC Power Supply
•Universal Input 90-264 VAC
•Less than 0.5W No Load Power Dissipation
•Rugged EMC Compliant Design
•2” x 4” x 1.2” Package
nd
Edition
nd
Edition
•Approved to EN/IEC/UL60601-1, 2
•Approved to EN/IEC/UL60950-1, 2
•Meets Efficiency Level IV Efficiency
•
Compliant (LVD, RoHS)
•Optional Chassis/Cover
Specifications
All Specifications are typical at nominal input, full load at 25C unless otherwise stated.
AC Input
100-240VAC +/- 10%, 47-63 Hz single phase.
Input Current
115VAC: 0.7A, 230VAC: 1.5A.
Inrush Current
The inrush at 264VAC, at a cold start, will not exceed 40A.
Input Fuses
F1, F2: 2.5A, 250VAC fuses provided on all models.
Earth Leakage Current
<300μA @ 264Vac, 60Hz, normal; <500μA SFC.
Efficiency
Up to 90% typical.
Output Power
65W continuous.
Ripple and Noise
1% pk-pk max., 20MHz BW, differential mode. Measured with noise
probe directly across output terminals.
Output Voltage
See chart.
Voltage Adjustability
Voltage pre-set by fixed resistance.
Minimum Load
Not required.
Total Regulation
+/- 2%. See chart.
Transient Response
500μs max., 50% load step, typical.
Turn On Time
Less than 1 second under all rated load conditions.
Hold-up Time
<17 ms at 65W, nominal input (120VAC).
Overvoltage Protection
Built-in
Overload Protection
Hiccup Mode
Short Circuit Protection
Hiccup Mode
Switching Frequency
Variable 50-75kHz typical.
Medical Safety Standards
EN/IEC/UL 60601-1, 2nd Edition, EN/IEC/UL60950-1, 2nd Edition
Isolation
Input-Output: 4000VAC, Input-Ground: 1800VAC, Output-Ground:
500VAC
Operating Temperature
0 to 70C (derate from full rated power above 40C).
Cooling
Convection
Output Power Derating
See derating chart
Storage Temperature
-40 to +85C.
Relative Humidity
5% to 95%, non-condensing
______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp  6050 King Drive Ventura, CA 93003  Phone 805.486.4565  Fax 858.712.2040  Email: [email protected]  www.slpower.com
MINT1065
65W Single Output Series
Medical & ITE – Universal Input Power Supply
3 Year Warranty
Output Parameters
Model
Number
Volts (V)
Output Current
Open Frame
w/cover
Ripple
& Noise
Total
Regulation
OVP
Threshold
MINT1065X1275C01
12V
5.25A
4.00A
120mV
±2%
15.0 ± 1.5V
MINT1065X1375C01
13.2V
4.80A
3.63A
140mV
±2%
16.0 ± 1.5V
MINT1065X1575C01
15V
4.33A
3.20A
150mV
±2%
18.0 ± 1.5V
MINT1065X1875C01
18V
3.50A
2.66A
180mV
±2%
21.0 ± 2.0V
MINT1065X2075C01
20V
3.25A
2.40A
200mV
±2%
23.0 ± 2.0V
MINT1065X2475C01
24V
2.70A
2.00A
240mV
±2%
27.0 ± 2.0V
MINT1065X4875C01
48V
1.35A
1.00A
480mV
±2%
55.0 ± 4.0V
Specifications (continued)
Vibration
Operating:
0.003 g2/Hz, 1.5 grms overall, 3 axes, 10 min. / axis
Non-Operating:
0.026 g2/Hz, 5.0 grms overall, 3 axes, 1 hr. / axis
Dimensions
W 2.0” x L 4.0” x H 1.17”. See outline drawing.
Weight
150 grams
Shock
Operating:
Half-sine, 20 gpk, 10 ms, 3 axes, 6 shocks total
Non-Operating: Half-sine, 40 gpk, 10 ms, 3 axes, 6 shocks total
Operating Altitude
-500 to 10,000 ft.
Non-operating Altitude
-500 to 40,000 ft.
EMI/EMC Compliance
Conducted Emissions
EN55011 Class B, FCC Part 15, Class B.
Radiated Emissions
EN55011 Class A, FCC Part 15, Class A.
Static Discharge Immunity
EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity
EN61000-4-3, 3V/m.
EFT/Burst Immunity
EN61000-4-4, 2kV/5kHz..
Line Surge Immunity
EN61000-4-5, 1kV differential, 2kV common-mode
Conducted RF Immunity
EN61000-4-6, 3Vrms
Power Frequency Magnetic Field Immunity
EN61000-4-8, 3A/m
Voltage Dip Immunity
EN61000-4-11 240VAC, 0%/0.5 cycle, 40%/5cycles, 70%/25
cycles
Model Number Key
MINT 1 065 X 12 75 C 01
01 = Standard Model, 02 and higher indicates a modified model.
Input Connector - "C" = 2 pin header
"75" = Output Connector - 4 pin header
Output Voltage: "12" = 12V output, "24" = 24V output, etc.
Model Configuration: "A" = Class I (grounded); "B" = Class II (ungrounded); "C" =
Chassis/cover provided (class I only)
Output Power (Watts) - "065" = 65W
# of Outputs
Product Family: "M" = Medical, "I" = Internal, "NT" = New Technology
______________________________________________________________________________________________________________________________ 2 of 3
SL Power Electronics Corp  6050 King Drive Ventura, CA 93003  Phone 805.486.4565  Fax 858.712.2040  Email: [email protected]  www.slpower.com
MINT1065
65W Single Output Series
Medical & ITE – Universal Input Power Supply
3 Year Warranty
Mechanical Drawings and Connector Information
MINT1065 Series Connectors
INPUT:
J1
AMP P/N 641937-1
PIN 1) AC LINE
PIN 2) AC NEUTRAL
OUTPUT:
J2
AMP P/N 640445-4
PIN 1) +Vout
PIN 2) +Vout
PIN 3) -Vout
PIN 4) -Vout
MATING CONNECTORS:
P1
P2
Pins:
Molex 09-50-3031
Molex 09-50-3041
Molex 08-50-0105
GND CONNECTION (Class I version)
0.187" FASTON TAB
Output Power Derating Curve
Data Sheet © 2013 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication.
Rev. 2-1-2013
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
______________________________________________________________________________________________________________________________ 3 of 3
SL Power Electronics Corp  6050 King Drive Ventura, CA 93003  Phone 805.486.4565  Fax 858.712.2040  Email: [email protected]  www.slpower.com