SL Power ME60A0551N01 60w single output external power Datasheet

ME60
60W Single Output External Power
Medical Series
Features

Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*

Approved to EN/IEC/UL60601-1, 3rd edition
with isolation levels which satisfy the 2 MOPP
requirements

Meets DoE Efficiency Level VI Requirements


No load input power
Average Efficiency

Up to 60W of AC-DC Power

Universal Input 90-264Vac Input Range

Desktop Style Package

Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin

E-cap life of >7 years

3 Year Warranty

IP22 Rated Enclosure
Description
A high performance AC to DC external power supply family designed for medical applications. The ME60A Medical Series low
power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition and EN/IEC/UL60601-1-11:2010
for Home Healthcare (non-hospital use) applications with and isolation levels which satisfy the 2 MOPP requirements and
*
designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME60A Series models will operate at universal input range of 90 to
264Vac over the wide temperature range of -20°C to +70°C, delivering full rated output power up to +40°C and applicable
output power derating at 70°C. These models are available in desktop versions, include an IP22 rating per IEC60529 for the
enclosure, and the output cable can be terminated at a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME60A0551F01
Volts
5.0V
Output
Current
7.00A
Output
Power
35W
Ripple &
Noise¹
75mV pk-pk
Line
Regulation
±1%
Load
Regulation
±5%
ME60A1203F01
12.0V
5.00A
60W
120mV pk-pk
±1%
±5%
ME60A1503F01
15.0V
4.00A
60W
150mV pk-pk
±1%
±5%
ME60A1803F01
18.0V
3.30A
60W
180mV pk-pk
±1%
±5%
ME60A2403F01
24.0V
2.70A
60W
240mV pk-pk
±1%
±5%
ME60A4803F01
48.0V
1.35A
60W
480mV pk-pk
±1%
±5%
ME60A0551N01
5.0V
7.00A
35W
75mV pk-pk
±1%
±5%
ME60A1203N01
12.0V
5.00A
60W
120mV pk-pk
±1%
±5%
ME60A1503N01
15.0V
4.00A
60W
150mV pk-pk
±1%
±5%
ME60A1803N01
18.0V
3.30A
60W
180mV pk-pk
±1%
±5%
ME60A2403N01
24.0V
2.70A
60W
240mV pk-pk
±1%
±5%
ME60A4803N01
48.0V
1.35A
60W
480mV pk-pk
±1%
±5%
ME60A0551Q01
5.0V
7.00A
35W
75mV pk-pk
±1%
±5%
ME60A1203Q01
12.0V
5.00A
60W
120mV pk-pk
±1%
±5%
ME60A1503Q01
15.0V
4.00A
60W
150mV pk-pk
±1%
±5%
ME60A1803Q01
18.0V
3.30A
60W
180mV pk-pk
±1%
±5%
ME60A2403Q01
24.0V
2.70A
60W
240mV pk-pk
±1%
±5%
Output
Connector
Output
Cable
6 pin Molex Type2
1150mm, #18AWG
2.5 x 5.5 x 9.5mm
Straight Barrel
Type, center
positive
9V:1150mm
18AWG
All others:
1500mm, #18AWG
6 pin Molex Type3
1150mm, #18AWG
2.5 x 5.5 x 9.5mm
Straight Barrel
Type, center
positive
9V:1150mm
18AWG
All others:
1500mm, #18AWG
6 pin Molex Type3
1150mm, #18AWG
2.5 x 5.5 x 9.5mm
Straight Barrel
Type, center
positive
9V:1150mm
18AWG
All others:
1500mm, #18AWG
Input
Configuration
Class I Desktop,
IEC60320 C14
Receptacle
Class II Desktop,
IEC60320 C8
Receptacle
Class II Desktop,
IEC60320 C18
Receptacle
ME60A4803Q01
48.0V
1.35A
60W
480mV pk-pk
±1%
±5%
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V
models, values listed are typical, 100mV pk-pk maximum.
2. Molex p/n 39-01-2060 or equivalent. See outline drawing for pinout information.
3. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE60B1203F01).
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
www.slpower.com
ME60 Rev 2.7
2-AUG-16
1
ME60
60W Single Output External Power
Medical Series
General Specifications
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Less than 1 sec @115Vac, full load
100Vac: 1.5A, 240Vac: 0.7A
Hold-up Time
20mS min., at full Load, 100Vac input
264Vac, cold start: will not exceed 40A
F1, F2: 2A, 250Vac fuses (line & neutral lines)
provided on all models
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
Overtemperature
Protection
Overload Protection
Short Circuit
Protection
Overvoltage
Protection
>88%, typical
Output Power
60W continuous – See models chart for
specific voltage model ratings.
No Load Input
Power
<0.210W (meets DoE Efficiency Level VI
Requirements)
Isolation
Safety Standards
Will shutdown upon an overtemperature
condition, auto-recovery.
130 to 180% of rating, Hiccup Mode
Hiccup Mode, auto recovery.
130 to 150% of output voltage (max. 60V on
48V model), hiccup mode
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1 MOPP
EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
Ripple and Noise
See models chart on pg 1.
Operating
Temperature
-20°C to +70°C.
Output Voltage
See models chart on pg 1.
Temperature
Derating
See derating curve.
Transient
Response
Regulation
Drop Test
Vibration
Dimensions
Weight
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Storage Temperature
Altitude
See models chart on pg 1.
1.4m from table top to wooden platform, 4
faces.
Relative Humidity
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
W: 2.67” x L: 4.25” x H: 1.29”
W: 67.9mm x L: 108mm x H: 32.7mm
MTBF
>250,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6.
E-Cap Life
400g
>7 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day. (80% load on 5V, 12V model)
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
www.slpower.com
ME60 Rev 2.7
2-AUG-16
2
ME60
60W Single Output External Power
Medical Series
EMI/EMC Compliance
Conducted Emissions:
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4.4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
EN55024/IECEN61000-4-11: --100% dip for 10 mS, at 0, 45, 90, 135, 180, 225,
270 and 315 degrees, 100% dip for 20mS, 0 deg., Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted. Consult factory for information regarding testing for or usage under special
environments.
Mechanical Drawing
See
Note 3:
9V through 48V Models: 2.5 x 5.5 x 9.5mm
5V Models: Output Connector: 6 pin Molex 39-01-2060
or equiv. Pins 1, 4 = (+), pins 3, 6 = (-), pins 2, 5 = NC²
Barrel Connector, center positive²
Notes:
1) All dimensions in mm.
2) Other options are available.
3) Cable length on 12V through 48V models is 1500mm, nominal.
4) The unit should not be covered or enclosed to protect against excessive case temperature rise.
www.slpower.com
ME60 Rev 2.7
2-AUG-16
3
ME60
60W Single Output External Power
Medical Series
Derating Chart:
70
Output Power- Watts
60
50
40
12V
30
18V
20
24V
48V
10
0
-20
-10
0
10
20
30
40
50
60
70
Ambient Temperature- ͦC
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive (6-pin Molex type - #51 – on 5V models).
Other standard options are listed below. The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME60 Rev 2.7
2-AUG-16
4
Similar pages