Bel MBC120-1030L 3 x 2 footprint Datasheet

The MBC120 Series of open frame medical
power supplies feature a wide universal AC
input range of 85 V – 264 VAC, offering 120 W
of output power in a compact footprint, with a
variety of isolated single output voltages.

3 x 2 Footprint

120 Watts with Forced Air Cooling

Approval to EN60601 3rd Edition

Efficiencies up to 93%

-40 To 70 Degree Operating Temperature

Dual Fusing

Thermal Shut-Down Feature

>3.00 Million Hours, Telcordia -SR332-Issue 3

Standby Power < 0.3W

Class II Option Available

RoHS Compliant

CE Marked
The MBC series was designed to 3rd edition
medical approvals and provides 2x MOPP
(Means of Patient Protection) isolation for
Class I and Class II installations.
These power supplies are ideal for medical,
telecom, Datacom, industrial equipment and
other applications.






Diagnostic
Drug Pump
Dialysis
Home Health Care
Monitoring
Portable Equipment
North America
+1-866.513.2839
Asia-Pacific
+86.755.29885888
Europe, Middle East
+353 61 225 977
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
MBC120 Series
MODEL NUMBER2
MBC120-1T12L
MBC120-1012L
MBC120-1T15L
MBC120-1015L
MBC120-1T24L
MBC120-1024L
MBC120-1T30L
MBC120-1030L
MBC120-1T48L
MBC120-1048L
MBC120-1T58L
MBC120-1058L
COVER-120-XBC
1
2
DESCRIPTION
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
Screw Terminal
Molex Header
metal cover kit accessory
VOLTAGE
MAX. LOAD
(CONVECTION)
MAX. LOAD
(200 LFM)
MIN. LOAD
RIPPLE &
NOISE 1
12 V
8.33 A
10.0 A
0.0 A
1%
15 V
6.66 A
8.0 A
0.0 A
1%
24 V
4.16 A
5.0 A
0.0 A
1%
30 V
3.33 A
4.0 A
0.0 A
1%
48 V
2.08 A
2.5 A
0.0 A
1%
58 V
1.72 A
2.07 A
0.0 A
1%
Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line
voltage and load ranges.
For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC120-1012L-2).
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
MBC120 Series
Specifications are for nominal input voltage, 25°C unless otherwise stated.
PARAMETER
DESCRIPTION / CONDITION
SPECIFICATION
Input Voltage
Universal (see derating under output power)
85-264 VAC / 390 VDC3
Power Factor
115 VAC:
230 VAC:
Typical
115 VAC:
230 VAC:
264 VAC:
Typical (N.A. For Class II Option)
Touch current
@ Full Load
47-63 Hz
1.2 A max.
0.65 A max.
< 0.3 W
25 A
45 A
75 A
300 uA
<100 uA
> 0.95
Switching Frequency
Typical
60 KHz
DESCRIPTION / CONDITION
With 200 LFM:
Convection:
48 V, 58 V:
24 V, 30 V:
12 V, 15 V:
Typical
SPECIFICATION
120 W
Up to 100 W
93%
91%
90%
>10 ms
Input Frequency
Input Current
No Load Power
Inrush Current
Leakage Current
3
Functional, not approved.
PARAMETER
Output Power
Efficiency
Hold-up Time
Line Regulation
+/-0.5%
Load Regulation
Transient Response
+/-1%
25% step load change, at 0.1A/uS slew rate,
50% duty cycle, 50 Hz = 4%
Voltage Adjustment
Rise Time
Typical
recovery time < 5 ms
+/-3%
55 ms
Set Point Tolerance
+/-1%
Over Current Protection
> 110%
Over Voltage Protection
Latch type (AC recycling required)
Short Circuit Protection
Hiccup mode
PARAMETER
DESCRIPTION / CONDITION
Operating Temperature
110 to 140%
Startup guaranteed, derate linearly above 50°C to 70°C as per Fig 1.
Storage Temperature
Cooling
Relative Humidity
Altitude
Reliability
Forced: with 300 LFM (Refer to Mech. Drawing)
Convection: for input 100-264 VAC (derate linearly to 80W @ 85VAC)
Noncondensing
Operating:
Nonoperating:
MTBF according to Telcordia -SR332-Issue 3
SPECIFICATION
-40 to +70°C
-40 to 0°C
-40 to +85°C
120 W
100 W
5% to 95%
16,000 ft
40,000 ft.
3.00 million hours
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
MBC120 Series
PARAMETER
DESCRIPTION / CONDITION
Conducted Emissions
EN55022-B, CISPR22-B, FCC PART15 - B
Static Discharge
EN61000-4-2:
Level-3
RF Field Susceptibility
EN61000-4-3:
Level-3
Fast Transients/Bursts
Level-3
Level A
Level B
Surge Susceptibility
EN61000-4-4:
Radiated:
Radiated with external core:
(King core K5B RC 25x12x15-M in input cable with 5 Turns)
EN61000-4-5:
Harmonic Current
EN61000-3-2:
Class D
PARAMETER
SPECIFICATION
4000 VAC
1500 VAC
1500 VAC
500 VAC
Agency Approvals
DESCRIPTION / CONDITION
Input to Output: (For medical applications)
Input to GND: (Not Applicable For Class II Option)
Output to GND: for type BF
for type B (N/A For Class II Option)
Approved to the latest edition of the following standards:
CSA/UL60601-1, EN60601-1 and IEC60601-1.
Nemko, UL, C-UL
CE mark
Complies with LVD Directive
Radiated Emissions
Isolation Voltage
Safety Standard(s)
SPECIFICATION
Level-3
Figure 1 – Derating Curve
CONNECTOR
AC Input Connector
DC Output Connector
PIN
J1
J2
DESCRIPTION / CONDITION
Pin 1
Pin 2
Pin 3
AC Line
Not Fitted
AC Neutral
Pin 1, 2
Pin 3, 4
V1 +VE
V1 - VE
Screw Terminal (Option 1)
Molex Header (Option 2)
Screw Terminal (Option 1)
Molex Header (Option 2)
MANUFACTURER / PN
Molex: 39357-0003
Tyco-2-1776112-3
Molex: 1722861103
(Mating conn: Molex 1722561003)
Molex: 39357-0004
Tyco-2-1776112-4
Molex: 1722861104
(Mating conn: Molex 1722561004)
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
MBC120 Series
PARAMETER
DESCRIPTION / CONDITION
Weight
150 g
Dimensions
76.2 x 50.8 x 30.1 mm (3 x 2 x 1.18 inch)
Figure 2 - Mechanical Drawing - Screw Terminal (Option 1)
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
MBC120 Series
Figure 3 - Mechanical Drawing - Molex Header (Option 2)
NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support
systems, equipment used in hazardous environments, or nuclear control systems.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending
on the date manufactured. Specifications are subject to change without notice.
866.513.2839
[email protected]
© 2015 Bel Power Solutions, Inc.
BCD.00667_001
Similar pages