XP Power JHL0324S15 Dc-dc converter Datasheet

JHL03 Series
DC-DC Converter
3 Watts
xxx Series
• International Medical Approvals
• 4000 VAC Reinforced Insulation
• Meets IEC60601-1, 3rd Edition
• 2 MOPP Isolation at 250 VAC
• 2 µA Patient Leakage Current
• DIP24 Package
• EN55011 Level A With No External Components
• 3 Year Warranty
Dimensions:
JHL03:
1.25 x 0.80 x 0.40” (31.15 x 20.32 x 10.20 mm)
Models & Ratings
Input Voltage
Output Voltage
10-17 V
20-30 V
Input Current
No Load(1)
Full Load(2)
Output Current
Maximum
Capacitive Load(3)
Efficiency(4)
Model Number
5.0 V
600 mA
56 mA
325 mA
720 µF
76%
JHL0312S05
12.0 V
250 mA
72 mA
316 mA
300 µF
78%
JHL0312S12
15.0 V
200 mA
67 mA
315 mA
240 µF
78%
JHL0312S15
±12.0 V
±125 mA
43 mA
304 mA
±140 µF
81%
JHL0312D12
±15.0 V
±100 mA
56 mA
303 mA
±120 µF
80%
JHL0312D15
5.0 V
600 mA
38 mA
167 mA
720 µF
74%
JHL0324S05
12.0 V
250 mA
37 mA
165 mA
300 µF
78%
JHL0324S12
15.0 V
200 mA
23 mA
146 mA
240 µF
82%
JHL0324S15
±12.0 V
±125 mA
29 mA
150 mA
±140 µF
80%
JHL0324D12
±15.0 V
±100 mA
42 mA
166 mA
±120 µF
80%
JHL0324D15
Notes
1. Input current measured at nominal input voltage.
2. Input current measured at lowest input voltage.
3. Maximum capacitive load is per output.
4. Typical values.
Mechanical Details
1.25
(31.75)
14
0.80
(20.32)
Pin Connections
0.10
(2.54)
0.20
(5.08)
16
22
23
0.60
(15.24)
Bottom View
11 10
0.18
(4.57)
0.40 (10.20)
9
3
2
0.90
(22.86)
Notes:
1. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
2. Pin pitch tolerance: ±0.014 (±0.35)
3. Case Tolerance: ±0.02 (±0.5)
0.12 (3.05) Min.
1
www.xppower.com
Pin
Single
Dual
2
-Vin
-Vin
3
-Vin
-Vin
9
No Pin
Common
10
Trim
Trim
11
No Pin
-Vout
14
+Vout
+Vout
16
-Vout
Common
22
+Vin
+Vin
23
+Vin
+Vin
Notes
1. All dimensions are in inches (mm)
2. Weight: 0.04 lbs (20 g) approx.
3. Pin diameter: 0.02 ±0.002 (0.5 ±0.05)
4. Pin pitch tolerance: ±0.014 (±0.35)
5. Case tolerance: ±0.02 (±0.5)
JHL03 Series
DC-DC Converter
Input
Characteristic
Input Voltage Range
Minimum
Typical
Units
17
VDC
12 V nominal
20
30
VDC
24 V nominal
25
A
2
µA
Input Current
See Models and Ratings table
Inrush Current
Input Filter
At 30 VDC input
Pi type
Patient Leakage Current
Undervoltage Lockout
Notes & Conditions
Maximum
10
On at >8.8 V. Off <8.3 V
12 V models
On at >17.5 V. Off <17.0 V
24 V models
Input Surge
25
VDC
12 V models for 3 s
50
VDC
24 V models for 3 s
Maximum
Units
Notes & Conditions
30
V
See Models and Ratings table
±10
%
Via external resistors, see Application Notes
±1
%
on V1
±2
%
on V2 of dual output models
A
No minimum load required
Output
Characteristic
Minimum
Output Voltage
5
Typical
Output Voltage Trim
Initial Set Accuracy
Minimum Load
0
Start Up Delay
5
Start Up Rise Time
2
ms
ms
Line Regulation
±0.3
%
Load Regulation
±1
%
0 - 100% load
On dual output models with one output set to 50% load and
the other varied from 10% to 100% load (D05 20% to 100%)
Cross Regulation
±4
%
Transient Response
4
% deviation
Ripple & Noise
1
% pk-pk
Short Circuit Protection
Recovery to within 1% in <500 µs for a 50% load change
at 0.25 A/µs rate
20 MHz bandwidth
Trip & Restart (hiccup mode), auto recovery
Overload Protection
120
200
%
Overvoltage Protection
115
140
%
0.03
%/ ˚C
Maximum
Units
Temperature Coefficient
Trip & Restart (hiccup mode)
General
Characteristic
Minimum
Efficiency
Isolation
Typical
80
%
Patient Leakage Current
2
Input to Output Capacitance
20
Switching Frequency
pF
kHz
Power Density
7.5
Weight
For 1 min. Double/reinforced with a working voltage of 250
VAC. Meets 2 x MOPP per 3rd edition of IEC60601-1 5000
VAC for 10 ms in accordance with IEC60664-1
µA
250
Mean Time Between Failure
See Models and Ratings table
VAC
4000
Notes & Conditions
W/in3
>1
MHrs
0.04 (20.0)
lb (g)
MIL-HDBK-217F, +25 °C GB
Environmental
Characteristic
Minimum
Operating Temperature
-20
Storage Temperature
-40
Case Temperature
Humidity
5
Typical
Maximum
Units
Notes & Conditions
+80
°C
See derating curve
+100
°C
+100
°C
90
%RH
Cooling
Non-condensing
Natural convection
Shock
±3 shocks in each plane, total 18 shocks of 30 g : 11 ms halfsine. Conforms to EN60068-2-27 & EN60068-2-47
Vibration
10-500 Hz at 2 g sweep and endurance at resonance in all 3 planes. Conforms to EN60068-2-6
www.xppower.com
2
JHL03 Series
DC-DC Converter
EMC: Emissions
Phenomenon
Standard
Test Level
Conducted
EN55011
Level A
Radiated
EN55011
Level A
Notes & Conditions
EMC: Immunity
Standard
Test Level
Criteria
Immunity
Phenomenon
IEC60601-1-2
Ed 4.0: 2014
As Below
ESD Immunity
EN61000-4-2
±8 kv Contact, ±15 kv Air
A
Radiated Immunity
EN61000-4-3
10 V/m
A
EFT/Burst
EN61000-4-4
2
A
Surges
EN61000-4-5
1
A
Conducted Immunity
EN61000-4-6
3 Vm
A
Magnetic Fields
EN61000-4-8
30 A/m
A
Safety Approvals
Notes & Conditions
80 MHz - 2.7 GHz plus discrete communication
proximity field frequencies
ANSI/AMMI ES60601-1 3rd Edition, CSA-22.2 No.60601-1:2008, IEC60601-1 3rd Edition
Safety Approvals
Safety Agency
Safety Standard
Notes & Conditions
CB Report
IEC60601-1 Including Risk Management
Medical
UL
ANSI/AAMI ES60601-1 3rd Ed. & CSA C22.2, No.60601-1:2008
Medical
EN
EN60601-1
Medical
Application Notes
Derating Curve
100
1
2
3
Load (%)
80
60
40
20
0
-20
-10
0
10
20
30
40
50
60
70
80
90
Ambient Temperature (°C)
External Output Trim
Trim
+ Vout
Trim Up
-Vout
3
Trim Down
Trim
For 5 V output:
Trim +10%, R = 3.4 k typical
Trim -10%, R = 1.1 k typical
For ±12 V output:
Trim +10%, R = 12.8 k typical
Trim -10%, R = 9.5 k typical
For 12 V output:
Trim +10%, R = 5.9 k typical
Trim -10%, R = 11.3 k typical
For ±15 V output:
Trim +10%, R = 18 k typical
Trim -10%, R = 14.8 k typical
For 15 V output:
Trim +10%, R = 8.4 k typical
Trim -10%, R = 10.4 k typical