XP Power ALM65US15 Ac-dc power supply Datasheet

ALM65 Series
AC-DC Power Supplies
65 Watts
• Medical & IT Safety Approvals
• Energy Efficiency Level VI
• 4th Edition Medical EMC
• IP32 Environmental Rating
• Class I and Class II Versions
• <0.21 W Standby Power
• 0 °C to 60 °C Operation
• Low Earth Leakage Current
• 3 Year Warranty
Dimensions:
The ALM65 series of medical external power supplies is fully approved to international medical
safety standards. It has been designed with very high efficiency and low standby power, enabling it
to meet the latest environmental legislation. The unit has a fully sealed enclosure complying with
IP32 and a smooth surface finish making it easier to wipe down in a clinical setting.
ALM65:
4.94 x 2.19 x 1.32” (125.5 x 55.5 x 33.5 mm)
Models & Ratings
Output Power
Output Voltage
Output Current
12.0 V
15.0 V
19.0 V
24.0 V
48.0 V
65 W
5.40
4.30
3.40
2.70
1.35
A
A
A
A
A
Total Regulation
Efficiency(1)
±5%
88%
88%
88%
88%
88%
Model Number(2,3,4)
ALM65US12
ALM65US15
ALM65US19
ALM65US24
ALM65US48
Notes
1. Typical average of efficiencies measured at 25%, 50%, 75% and 100% load and
115VAC input.
2. For class II versions, add suffix 'C2-8' to the end of the part number e.g.
ALM65US24C2-8.
3. For optional input connector retention clip add suffix ‘-A’ to the model number,
e.g. ALM65US24-A (not available for C2 versions)
4. 5.5 x 2.1 mm output connector add suffix B1 to the part number,
e.g. ALM65US24-B1, ALM65US24C2-8B1, ALM65US24-AB1.
Mechanical Details
Class I
IEC 320-C14
4.941 (125.5)
1.319 (33.5)
Class II
IEC 320-C8
1.512
(38.4)
2.185 (55.5)
0.177 (4.50)
35.433 ±0.8 (900±20)
ø 0.685
(17.4)
1.969 ±0.394 (50±10)
0.374 ±0.012 (9.50±0.30)
1.319 (33.5)
0.039 (1.0)
RETURN ( - )
OUTPUT ( + )
4.488 (114.0)
1.732 (44.0)
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
output cable length.
2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
for C8 versions.
3. Weight: 0.68 lbs (308 g) approx.
4. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
1
www.xppower.com
ALM65 Series
AC-DC Power Supplies
Input
Units
Notes & Conditions
264
VAC
Derate linearly from 100% load at 90 VAC to 80% load
at 80 VAC
63
60/120
Hz
A
A
250
0.21
µA
W
Max.
48
±2
Units
VDC
%
2
40
±0.5
±5
s
ms
ms
%
%
4
%
Ripple and Noise
1.5
% pk-pk
Overshoot
Overload Protection
Overvoltage Protection
Short Circuit Protection
Thermal Protection
Temperature Coefficient
Patient Leakage Current
10
175
170
%
%
%
100
%/°C
µA
264 VAC, 60 Hz
Maximum
+60
Units
°C
Notes & Conditions
Derate from 100% load at 40 °C to 50% load at 60 °C
Characteristic
Minimum
Input Voltage
80
Input Frequency
Input Current
Inrush Current
Power Factor
Earth Leakage Current
No Load Input Power
Input Protection
47
Typical
Maximum
1.2/0.6
Measured at 115/230 VAC
115/230 VAC, cold start at 25 °C
EN61000-3-2 Class A
264 VAC, 60 Hz
T3.15A/250 VAC internal fuse in both line
Output
Characteristic
Output Voltage
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
Line Regulation
Total Regulation
Min.
12
Typ.
20
Transient Response
115
145
Trip and restart (hiccup), auto resetting
Measured internally, auto resetting
0.05
Notes & Conditions
See Models and Ratings table
At 50% load
No minimum load required
Full load and 230 VAC
Maximum deviation, recovering to less than 1% within
500 µs for 25% step load
20 MHz bandwidth, measured with 20 MHz Bandwidth and
10 µF electrolytic in parallel with 0.1 µF ceramic capacitor.
At turn on / turn off
Recycle mains to reset
Environmental
Characteristic
Operating Temperature
Cooling
Operating Humidity
Storage Temperature
Operating Altitude
Shock
Vibration
Minimum
0
Natural convection
10
-40
Typical
90
%RH
+85
°C
5000
m
IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
IEC68-2-6, 10-500 Hz, 2 g 10 mins/sweep, 60 mins for each of 3 axes
Non-condensing
Derating Curve
Output Power (%)
100%
80%
60%
40%
20%
0
10
20
Ambient Temperature (ºC)
www.xppower.com
40
60
2
ALM65 Series
AC-DC Power Supplies
General
Characteristic
Minimum
Efficiency
Typical
Maximum
Notes & Conditions
Units
%
88
Switching Frequency
Power Density
Mean Time Between Failure
Weight
See Models and Ratings table and curves.
4000
VAC
2 x MOPP
1500
VAC
1 x MOPP (Class I versions only)
500
VAC
Class I versions only
65 KHz
kHz
PWM
4.55
W/in3
Isolation: Input to Output
Input to Ground
Output to Ground
>300
kHrs
0.68 (308)
lb (g)
MIL-HDBK-217F at 25 °C GB
Efficiency Curves
ALM65US12
100%
90%
80%
Efficiency
70%
60%
50%
115V
40%
230V
30%
20%
10%
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Load
ALM65US24
100%
90%
80%
Efficiency
70%
60%
50%
115V
40%
230V
30%
20%
10%
0%
10%
20%
30%
40%
50%
60%
70%
80%
90%
100%
Load
EMC: Emissions
Phenomenon
Emissions
Harmonic Current
Voltage Flicker
Standard
EN55022
EN61000-3-2
EN61000-3-3
Test Level
Level B
Class A
www.xppower.com
Notes & Conditions
Conducted & Radiated
3
ALM65 Series
AC-DC Power Supplies
EMC: Immunity
Phenomenon
ESD
Radiated
EFT/Burst
Surge
Conducted
Magnetic Fields
Test Level
3
10 V/m
3
Installation Class 3
10 V
4
Dip: 30% 500 ms
Dip: 60% 200 ms
Dip: 80% 5000 ms
Dip: 100% 5000 ms
Dip: 30% 25 AC Cycles
Dip: 60% 5 AC Cycles
Dip: 100% 0.5 AC Cycles
Int.: >95% 5000 ms
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Dips and Interruptions
EN60601-1-2
Criteria
A
A
A
A
A
A
A/B
A/B
B
B
A
A
A
B
Notes & Conditions
±15 kV air/±8 kV contact
80-2700 MHz. IEC60601-1-2 Ed.4 at other frequencies
High Line/Low Line
High Line/Low Line
230 VAC 100% load, 100 VAC 75%
230 VAC 100% load, 100 VAC 20%
At 8 angles
Safety Approvals
Safety Agency
Safety Standard
UL
Notes & Conditions
UL60950-1 2nd Edition
TUV
EN60950-1
CB
IEC60950-1
CE
LVD
UL
ANSI/AAMI ES 60601-1
CSA
CSA C22.2 No. 60601
TUV
EN60601-1
CB
IEC60601-1
Information Technology
Medical
Mechanical Details
Class I
IEC 320-C14
4.941 (125.5)
1.319 (33.5)
Class II
IEC 320-C8
35.433 ±0.8 (900±20)
1.512
(38.4)
2.185 (55.5)
0.177 (4.50)
ø 0.685
(17.4)
Optional Input Connector Retention (-A)
1.40 (35.7)
1.969 ±0.394 (50±10)
0.374 ±0.012 (9.50±0.30)
1.319 (33.5)
0.039 (1.0)
RETURN ( - )
OUTPUT ( + )
4.488 (114.0)
1.732 (44.0)
Notes
1. All dimensions shown in inches (mm). Tolerance is 0.02 (0.5) maximum, except
5. For UK mains lead order part UK-MAINS-IEC for C14 versions, or UK-MAINS-8
for C8 versions.
output cable length.
2. Output connector: 5.5 outer diameter, 2.5 mm inner diameter with centre positive. 6. For US mains lead order part US-MAINS-IEC for C14 versions, or US-MAINS-8
for C8 versions.
3. Weight: 0.68 lbs (308 g) approx.
4. For European mains lead order part EU-MAINS-IEC for C14 versions,
or EU-MAINS-8 for C8 versions.
www.xppower.com
10 Mar 16
Similar pages