MPD MD212MD-15EHI Miniature, 2w sip medical approved dc/dc converter Datasheet

MD200MEHI Series
Miniature, 2W SIP
Medical Approved
DC/DC Converters
Key Features:
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
• EN 60601 3RD Ed. Approved Input
Parameter
• 2W Output Power
• 4.2 kVAC Isolation
• Reinforced Insulation
Conditions
Min.
5 VDC Input
4.5
12 VDC Input 10.8
24 VDC Input 21.6
Capacitor Filter
Input Voltage Range
Input Filter
• 1 x MOPP & 2 x MOOP per Reflected Ripple Current
EN 60601-1 3RD Edition & Output
ANSI/AAMI ES 60601-1
• 2 µA Leakage Current Max
• Compact SIP Case
• Single & Dual Outputs
• 3.5 MH MTBF
Typ.
5.0
12.0
24.0
Max.
5.5
13.2
26.4
Units
VDC
0.2
Parameter
Output Voltage Accuracy
Line Regulation
Load Regulation, See Note 1
Efficiency
Ripple & Noise (20 MHz), See Note 2
Temperature Coefficient
Output Short Circuit
Conditions
Min.
A
Typ.
±5.0
Max.
VIN Change of 1%
5.0 VOUT
All Other Models
See Model Selection Guide
5.0 VOUT
80
All Other Models
100
±0.02
See Note 3
±1.5
20
15
Units
%
%
%
150
150
mV P - P
%/°C
S
3.0
General
Parameter
Conditions
Isolation Voltage
Leakage Current, See Note 4
Isolation Resistance
Isolation Capacitance
Switching Frequency
EMI Characteristics, See Note 5
Parameter
Radiated Emissions
Conducted Emissions
ESD
Environmental , See Note 6
Parameter
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
Physical
Case Size
Case Material
Weight
Min.
60 Seconds 4,200
6,000
250 VAC, 50/60 Hz
500 VDC
1.0
100 kHz, 0.1V
Standard
EN 55022
EN 55022
EN 61000-4-2
Typ.
Max.
Units
VAC
VDC
µA
G
pF
kHz
2.0
5
100
Criteria
Level
Level B
Level B
±8 kV Contact
B
Conditions
Min.
Typ.
-40
-55
Free Air Convection
RH, Non-condensing
Ambient
Max.
+85
+125
Units
°C
°C
95
%
See Mechanical Diagram (Page 3)
Non-Conductive Black Plastic (UL94-V0)
0.14 Oz (4.2g)
Reliability Specifications
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
T: (781) 344-8226
F: (781) 344-8481
E: [email protected]
W: www.micropowerdirect.com
Parameter
MTBF
Transformer Creepage
Transformer Clearance
PCB Creepage & Clearance
Safety Approvals, See Note 7
Conditions
MIL HDBK 217F, 25°C, Gnd Benign
Min.
Typ.
Max.
Units
3.5
MHours
5.0
mm
5.0
mm
5.5
mm
IEC/EN 60601-1 3RD Edition, 1XMOPP & 2xMOOP
ANSI/AAMI ES 60601-1 1xMOPP & 2xMOOP Recognition (UL Certificate)
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1
Absolute Maximum Ratings
Parameter
Conditions
Input Voltage Surge (1 Sec)
Lead Temperature
Min.
Typ.
5 VDC Input
12 VDC Input
24 VDC Input
1.5 mm From Case For 10 Sec
Max.
9.0
18.0
30.0
300
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com
Units
VDC
°C
www.micropowerdirect.com
Model Selection Guide
Model
Number
MD205MS-05EHI
MD205MS-12EHI
MD205MS-15EHI
MD205MD-05EHI
MD205MD-09EHI
MD205MD-12EHI
MD205MD-15EHI
MD212MS-05EHI
MD212MS-12EHI
MD212MS-15EHI
MD212MD-05EHI
MD212MD-09EHI
MD212MD-12EHI
MD212MD-15EHI
MD224MS-05EHI
MD224MS-12EHI
MD224MS-15EHI
MD224MD-05EHI
MD224MD-09EHI
MD224MD-12EHI
MD224MD-15EHI
Input
Output
Voltage (VDC)
Current (mA)
Voltage
Nominal
Range
Full-Load No-Load (VDC)
5
4.5 - 5.5
519
35
5.0
5
4.5 - 5.5
506
35
12.0
5
4.5 - 5.5
506
35
15.0
5
4.5 - 5.5
513
35
±5.0
5
4.5 - 5.5
513
35
±9.0
5
4.5 - 5.5
513
35
±12.0
5
4.5 - 5.5
500
35
±15.0
12
10.8 - 13.2
216
15
5.0
12
10.8 - 13.2
208
15
12.0
12
10.8 - 13.2
203
15
15.0
12
10.8 - 13.2
214
15
±5.0
12
10.8 - 13.2
203
15
±9.0
12
10.8 - 13.2
203
15
±12.0
12
10.8 - 13.2
208
15
±15.0
24
21.6 - 26.4
105
10
5.0
24
21.6 - 26.4
102
10
12.0
24
21.6 - 26.4
99
10
15.0
24
21.6 - 26.4
105
10
±5.0
24
21.6 - 26.4
103
10
±9.0
24
21.6 - 26.4
102
10
±12.0
24
21.6 - 26.4
103
10
±15.0
Notes:
1. Output load regulation is specified for a load change of 10% to 100%.
2. When measuring output ripple, it is recommended that an external 1.0 µF ceramic capacitor and a
10 µF electrolytic capacitor be connected in parallel from the +VOUT pin to the -VOUT pin for single output
models, or from each output to common for dual output models.
3. In the event of an output short circuit that lasts less than 3 seconds, the unit must be restarted.
4. Patient leakage current and reinforced insulation is based on a 250 VAC, 50/60 Hz system
input voltage.
5. These converters operate without external components. However, to meet specific EMC standards,
some external components may be required. Typical connections are shown below.
Typical Connection, Single Output Models
Current
Current
Capacitive
Load
(mA, Max)
(mA, Min)
(µF, Max)
400.0
167.0
133.0
±200.0
±111.0
±83.0
±67.0
400.0
167.0
133.0
±200.0
±111.0
±83.0
±67.0
400.0
167.0
133.0
±200.0
±111.0
±83.0
±67.0
40.0
17.0
133.0
±20.0
±12.0
±9.0
±7.0
40.0
17.0
133.0
±20.0
±12.0
±9.0
±7.0
40.0
17.0
133.0
±20.0
±12.0
±9.0
±7.0
Efficiency
(%, Typ)
1,000
470
470
470
470
220
220
1,000
470
470
470
470
220
220
1,000
470
470
470
470
220
220
Fuse Rating
Slow-Blow
(mA)
77
79
79
78
78
78
80
77
80
82
78
82
82
80
79
82
84
79
81
82
81
1,000
1,000
1,000
1,000
1,000
1,000
1,000
400
400
400
400
400
400
400
200
200
200
200
200
200
200
6. All models operate to 85ºC ambient without derating. They are not specified for operation over
85ºC. Contact the factory for more information.
7. The safety certification (EN 60601) and 1xMOPP/2xMOOP rating is for a system input
voltage within 250 VAC, 50/60 Hz.
8. Operation at no-load will not damage these units. However, they may not meet all
specifications.
9. It is recommended that a fuse be used on the input of a power supply for protection.
See the table above for the correct rating.
Typical Connection, Dual Output Models
These converters are specified for operation without external components. However, in some applications the addition of
input/output capacitors will enhance stability and reduce output ripple. For applications that require this, the diagrams above
illustrates a typical connection of the MD200Mx-xxEHI series. Suggested component values are given in the table at right.
Capacitors C2 and C3 are not required to meet specifications, but may be used if a lower level of output ripple is required.
VIN
C1
VOUT
C2
C3
5 VDC
12 VDC
24 VDC
10 µF/50V
4.7 µF/50V
2.2 µF/50V
5.0 VDC
12 VDC
15 VDC
±5.0 VDC
±9.0 VDC
±12 VDC
±15 VDC
10 µF
2.2 µF
1.0 µF
4.7 µF
2.2 µF
1.0 µF
1.0 µF
------4.7 µF
2.2 µF
1.0 µF
1.0 µF
Typical Connection, EMI Input Filter
For applications that require meeting EMC standards, the diagram at left
illustrates a typical connection of the MD200Mx-xxEHI series. Suggested
component values are given in the table at below.
The value for capacitor C2 is given above.
Page 2
We Power Your Success - For Less
VIN
C4
L1
C5
5 VDC
12 VDC
24 VDC
4.7 µF/50V
4.7 µF/50V
4.7 µF/50V
6.8 µH
6.8 µH
15 µH
4.7 µF/50V
4.7 µF/50V
4.7 µF/50V
www.micropowerdirect.com
Typical Efficiency vs Output, Nom Input
Typical Efficiency vs Output, Nom Input
Typical Efficiency vs Input, Full Load
Typical Efficiency vs Input, Full Load
Mechanical Dimensions
Pin Connections
Pin
Single
Dual
1
+VIN
+VIN
2
-VIN
-VIN
5
-VOUT
-VOUT
6
No Pin
Common
7
+VOUT
+VOUT
Mechanical Notes:
• All dimensions are typical in inches (mm)
• Tolerance x.xx = ±0.02 (±0.50)
MicroPower Direct
We Power Your Success - For Less!
292 Page Street Ste D Stoughton, MA 02072
•
TEL: (781) 344-8226
•
FAX: (781) 344-8481
•
E-Mail: [email protected]
Similar pages