Keysight N1082A Wide-bandwidth oscilloscope mainframe and module Datasheet

Keysight Technologies
Infiniium DCA-X 86100D
Wide-Bandwidth Oscilloscope
Mainframe and Modules
Data Sheet
02 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Introduction
See the TRUE performance of your designs.
The 86100D DCA-X performs precision measurements on high-speed digital designs from 50 Mb/s to more than 80 Gb/s on up to 16
channels simultaneously. Applications include:
Optical
–– Transceiver design and manufacturing
Electrical
–– ASIC/FPGA/IC design and characterization
TDR/TDT/S-Parameter
–– Serial bus designs, cables, and PCB characterization
Table of Contents
Introduction................................................................................................................................................. 02
Overview of the Infiniium DCA-X................................................................................................................. 03
Specifications............................................................................................................................................... 04
Modules Selection Table.............................................................................................................................. 09
Modules Specifications................................................................................................................................ 11
Typical System Configurations.................................................................................................................... 23
Software Applications................................................................................................................................. 24
Measurements............................................................................................................................................. 25
Ordering Information................................................................................................................................... 29
Keysight Oscilloscopes................................................................................................................................ 33
03 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Overview of the Infiniium DCA-X
The 86100D DCA-X can be viewed as four powerful instruments in one:
Scope mode
Jitter mode
High-fidelity waveform characterization (Yellow: raw trace, Blue:
de-embedded waveform)
Precision jitter, amplitude, and frequency analysis capability
Eye/Mask mode
TDR/TDT mode
Fast transmitter characterization using eye diagram analysis and
automated mask margin measurements
Accurate time domain reflectometry/transmission and S-Parameter
measurements
These modes are further complimented by the following features
that provide additional insight and analysis capability:
–– De-embedding, embedding, equalizer capability
–– Phase noise/Jitter spectrum analysis
–– Phase locked loop (PLL) analysis
–– And more…
Modular
Precision measurements, more margin, and more
insight
The 86100D DCA-X oscilloscope combines high analog
bandwidth, low jitter, and low noise performance to accurately
characterize optical and electrical designs from 50 Mb/s to over
80 Gb/s. The mainframe provides the foundation for powerful
insight and measurement capability, such as de-embedding of
cables and fixtures, that improve margins and allow engineers to
see the true performance of their designs.
The modular system means that the instrument can grow to
meet your needs, when you need it. There’s no need to purchase
capability that you don’t need now. The DCA-X supports a wide
range of modules for testing optical and electrical designs. Select
modules to get the specific bandwidth, filtering, and sensitivity
you need. The DCA-X supports all modules in the DCA family and
is 100% backwards compatible with the 86100C mainframe.
High channel count
The DCA-X can measure from one to 16 channels simultaneously
with no significant loss in throughput. Create an affordable
high-capacity test system by measuring multiple signals in
parallel.
Software
The DCA-X provides powerful analysis capability that is enabled
through licensed software options.
Examples include 86100D-200 for fast and accurate jitter
analysis, 86100D-SIM for de-embedding and/or embedding of
fixtures and cables, and the N1012A/N1014A/N108XA Series test
applications.
04 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications
General and mainframe
Comparing specifications
Specifications describe warranted performance over the
temperature range of +10 to +40 °C (unless otherwise noted).
The specifications are applicable for the temperature after the
instrument is turned on for one hour, and while self-calibration
is valid. Many performance parameters are enhanced through
frequent, simple user calibrations. Characteristics provide useful,
non-warranted information about the functions and performance
of the instrument. Characteristics are printed in italic typeface.
Product specifications and descriptions in this document are
subject to change without notice.
When comparing performance attributes between instruments,
ensure you compare the same type of parameter. For example,
compare warranted specifications from both instruments, or
compare characteristics or typical performance. Warranted
specifications include measurement uncertainties and are
conservative compared to other types of unwarranted attributes.
Factory calibration cycle - For optimum performance, the
instrument should have a complete verification of specifications
once every twelve months.
General specifications
Temperature
Operating
Non-operating
Altitude
Operating
Power (max)
100/120 Vac 50/60/400 Hz
220/240 Vac 50/60 Hz, 700 W maximum
Weight
Mainframe without modules
Typical module
Mainframe dimensions (excluding handle) Without front connectors and rear feet
Max relative humidity
Supply voltage variation
10 to +40 °C (50 to +104 °F)
–40 to +65 °C (–40 to +158 °F)
Up to 4,600 meters (15,000 feet)
With typical modules: 150 to 230 VA at 25 °C
20.5 kg (43 lbs)
1.2 kg (2.6 lbs)
221 mm (H) x 426 mm (W) x 530 mm (D)
(7 in x 16.76 in x 20.9 in)
With front connectors and rear feet
234 mm (H) x 426 mm (W) x 601 mm (D)
(9.23 in x 16.76 in x 23.67 in)
With front cover and rear feet
234 mm (H) x 426 mm (W) x 612 mm (D)
(9.23 in x 16.76 in x 24.1 in)
80% for temperatures up to 31 °C decreasing linearly to 50% relative humidity at 40 °C
86100D system can operate with supply voltage fluctuations up to ± 10% of the nominal voltage
Mainframe specifications
Horizontal system (time base)
Pattern lock
Scale factor (full scale is ten divisions)
Minimum
2 ps/div (with 86107A or 86100D-PTB: 500 fs/div)
Maximum
1 s/div
250 ns/div
Delays 1
Minimum
24 ns
40.1 ns default, 24 ns min
Maximum
1000 screen diameters or 10 s whichever is smaller
1000 screen diameters or 25.401 μs whichever is smaller
Time interval accuracy 2,3
1 ps + 1.0% of Δ time reading or 8 ps, whichever is smaller
Jitter mode operation 4
Time interval accuracy - jitter mode operation 1 ps
Time interval resolution
≤ (screen diameter)/(record length) or 62.5 fs, whichever is
larger
Display units
Bits or time (TDR mode–meters)
Vertical system (channels)
Number of channels
16 (simultaneous acquisition)
Vertical resolution
54xxx, 83xxx, 86xxx: 14 bit A/D converter (16 or more bits with averaging). N10xx: 16 bit A/D converter
Full resolution channel
Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution from the front panel knob
Adjustments
Scale, offset, activate filter, sampler bandwidth, attenuation factor, transducer conversion factors
Record length
Legacy UI: 16 to 16 K samples
FlexDCA without pattern lock: 16 to 64 K samples
FlexDCA with pattern lock: 16 up to 32 M samples 5
1. Time offset relative to the front panel trigger input on the instrument mainframe.
2. Dual marker measurement performed at a temperature within 5 °C of horizontal calibration temperature.
3. The maximum delay setting is 100 ns and delta time does not span across (28 + Nx4) ns ± 100 ps delay setting, where N=0,1,2….18. If delta time
measurement span exceeds above criteria, time interval accuracy is 8 ps +0.1% of Δ time reading.
4. Characteristic performance. Test configuration: PRBS of length 127 bits, Data and Clock 10 Gb/s.
5. Maximum number of samples depends on pattern, number of active channels and available memory.
05 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications (Continued)
Mainframe specifications (Continued)
86100D with option STR
86100D with option ETR or with options ETR + PTB
Not supported
Not supported
Free run
Free run
DC to 100 MHz
DC to 3.2 GHz
DC to 100 MHz
DC to 3.2 GHz
–1 V to +1 V
Positive or negative
Normal or high sensitivity
Sinusoidal or 200 ps minimum pulse width
200 mVpp
2 Vpp
Tested at 2.5 GHz, 200 mVpp
1.5 ps rms + 50 ppm of delay setting
< 1.0 ps rms + 5 ppm of delay setting (Characteristic)
–1 V to +1 V
Positive or negative
Normal or high sensitivity
Sinusoidal or 200 ps minimum pulse width
200 mVpp
2 Vpp
Tested at 2.5 GHz, 200 mVpp
1.5 ps rms + 50 ppm of delay setting
< 1.0 ps rms + 5 ppm of delay setting (Characteristic)
Module triggering 3
Internal trigger
External direct trigger (DC coupled)
Limited bandwidth
Full bandwidth
Trigger configuration
Triggering level adjustment
Edge select
Hysteresis 2
Trigger sensitivity
Maximum input signal
System jitter
06 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications (Continued)
Mainframe specifications (Continued)
External divided trigger (AC coupled)
Input frequency range
86100D with option STR
Not available with option STR
N/A
Pattern lock frequency range
N/A
Pattern lock pattern length range
N/A
Trigger sensitivity
N/A
Maximum input signal
N/A
System jitter
N/A
86100D with option ETR or with options ETR + PTB
50 MHz - 13 GHz (effective divide by 2)
50 MHz - 15 GHz (characteristic)
Option PTB: 50 MHz - 32 GHz (effective divide by 1)
50 MHz - 13 GHz
50 MHz - 15 GHz (characteristic)
Option PTB: 50 MHz - 32 GHz
1 to 2^23 UI in Scope and Eye/Mask Modes
1 to 2^16 UI in Jitter Mode
Slew rate > 2 V/ns:
200 mVpp
Max Trigger input: 2 Vpp
Option PTB: Max PTB Input: 1.3 Vpp with DC offset up
to + 200 mV
Tested at 10 GHz, 200 mVpp
1.2 ps rms for time delays less than 100 ns, (Maximum)
750 fs rms at minimum delay setting, (characteristic)
Option PTB when PTB is enabled:
Condition 1: 2.4 GHz — < 4.0 GHz trigger, tested at
2.4 GHz, 750 mVpp
≤ 200 fs (characteristic)
< 400 fs, with 54XXX, 8348X, or N1045A (non Option
LOJ) module (characteristic)
Condition 1: 4 GHz — 9.0 GHz trigger, tested at 5 GHz,
750 mVpp
≤ 120 fs (characteristic)
< 400 fs, with 54XXX, 8348X, or N1045A (non Option
LOJ) module (characteristic)
Condition 1: > 9.0 GHz — 44.0 GHz trigger, tested at 10,
20, and 40 GHz, 500 mVpp
≤ 90 fs (characteristic)
< 200 fs, with 54XXX, 8348X, or N1045A
(non Option LOJ) module(characteristic)
Inputs
Trigger input connector
2.92 mm (male)4
Trigger input nominal impedance
50Ω
1.
2.92 mm (male)4
Option PTB: PTB input: 2.92 mm (male)4
50 Ω
Option PTB: PTB input: 50 Ω
The internal precision timebase works with typical digital clock signals, such as a BERT output, as well as sine waves. If the rise time or fall time of the
clock signal is less than 15% of the period of the clock (for example, less than 15 ps for a 10 GHz clock), reduce the edge speed by using an external
low-pass filter. or length of cable. For the lowest jitter, use a signal that is as close as possible to the maximum signal amplitude (1.3 Vpp) and minimize any
sub-harmonics.
2. High Sensitivity Hysteresis mode improves the high frequency trigger sensitivity but is not recommended when using noisy, low frequency signals that may
result in false triggers without normal hysteresis enabled.
3. Module triggering enables precision waveform analyzer modules (86108), clock recovery modules (83496) and TDR modules to be used as a trigger source.
Without support for module triggering, external connections between the module and the mainframe may be required.
4. 86100D mainframes ship with a 2.92mm (f) to (f) adapter for this port.
07 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications (Continued)
Computer system and storage
Computer system and storage
CPU
Mass storage
Operating system
Display 1
Display area
Entire display resolution
Waveform colors
Persistence modes
Waveform overlap
Connect-the-dots
Persistence
Graticule
Grid intensity
Dialog boxes
Front panel inputs and outputs
Cal output
Trigger input
USB 2
Precision timebase input
(Option PTB only)
Rear panel inputs and outputs
Gated trigger input
Video output
GPIB 3
RS-232
LAN
USB 2
Precision timebase delay path
(Option PTB only)
1.
2.
3.
4.
Intel Core i5 3550S CPU at 3.0 GHz
80 GB internal hard drive (default) or 80 GB removable hard drive (Option 090)
Microsoft Windows 7 embedded standard (64 bit)
210.4 mm x 157.8 mm (10.4 inch diagonal color active matrix LCD module incorporating amorphous silicon TFTs)
1024 pixels horizontally x 768 pixels vertically
Select from over 16 colors; user may change color assignment of all traces (channels, waveform memory and
signal processing functions)
Gray scale, color grade, variable, infinite
When two waveforms overlap, a third color distinguishes the overlap area (classic DCA-J interface only)
On/Off selectable
Minimum, variable (100 ms to 40 s), infinite
On/Off
0 to 100%
Opaque or transparent
BNC (female) and test clip, banana plug
2.92 mm 4
Three low-power USB 2.0 ports; Voltage: 5.00 ± 0.25 V; Current: 100 mA each
2.92 mm 4
TTL compatible
VGA, full color, 15 pin D-sub (female) 10
Fully programmable, complies with IEEE 488.2
9 pin D-sub (male)
Four USB 2.0 ports; Voltage: 5.00 ± 0.25 V; Current: 500 mA each
Two female SMA connectors, nominal length jumper cable included
Supports external display. Supports multiple display configurations via Windows XP Pro display utility.
USB Keyboard and mouse included with mainframe.
The GPIB card interface is optional. To include this interface, order 86100D-GPI.
Mainframe ships with 2.92 mm female-female connector saver (P/N 1250-4105).
08 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Specifications (Continued)
Precision time base (86108 configurations)
The 86108 has an integrated precision timebase within the plug-in module. The 86108A/B can be triggered through clock recovery of
the observed signal, through an external reference clock into the precision timebase section, or with the precision timebase operating
on the clock signal recovered from the observed signal. The following specifications indicate the 86100 system timebase specifications
achieved when using the 86108A/B plug-in module. (The 86100 mainframe and the 86108A/B module can also be triggered with
a signal into the mainframe. In this configuration, the basic mainframe specifications are achieved.) If the 86108A/B is placed in a
mainframe that has an integrated precision timebase (86100D-PTB), the integrated mainframe precision timebase is disabled.
Precision time base 86108A/B
Typical jitter (clock recovery (CR) and precision timebase (PTB) configuration)
Maximum jitter (clock recovery and precision timebase configuration) 1
Typical jitter (clock recovery without precision timebase active)
Effective trigger-to-sample delay (CR and PTB configuration, typical)
Typical jitter (trigger signal applied to precision timebase input)
Maximum jitter (trigger signal supplied to precision timebase input) 1
Precision timebase trigger bandwidth
Precision timebase external reference amplitude characteristic
Precision timebase input signal type 2
Precision timebase maximum input level
Precision timebase maximum DC offset level
Precision timebase input impedance
Precision timebase connector type 3
Timebase resolution (with precision timebase active)
Timebase resolution (precision timebase disabled)
86108A
< 60 fs
< 90 fs
< 1.25 ps
< 200 ps
< 60 fs
< 100 fs
2 to 13.5 GHz
(1 to 17 GHz)
1.0 to 1.6 Vpp
Sinusoid
± 2 V (16 dBm)
± 200 mV
50 Ω
3.5 mm male
0.5 ps/div
2 ps/div
86108B-LBW
< 60 fs
< 90 fs
< 1.25 ps
< 200 ps
< 60 fs
< 100 fs
1 to 18 GHz
86108B-HBW
< 50 fs
< 70 fs
< 1.25 ps
< 200 ps
< 60 fs
< 100 fs
1 to 18 GHz
1.0 to 1.6 Vpp
Sinusoid
± 2 V (16 dBm)
± 200 mV
50 Ω
3.5 mm male
0.5 ps/div
2 ps/div
1.0 to 1.6 Vpp
Sinusoid
± 2 V (16 dBm)
± 200 mV
50 Ω
3.5 mm male
0.5 ps/div
2 ps/div
1. Verified with maximum level input signal (~800 mVpp)
2. The precision timebase performs optimally with a sinusoidal input. Non-sinusoidal signals will operate with some degradation in timebase linearity.
3. 86108B-LBW are shipped with a 3.5 mm (female) to 3.5 mm (female) connector saver on the input channels (P/N 5061-5311). The 86108B-HBW is shipped
with 2.4 mm female to 2.4 mm female connector savers on the input channels.
09 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Selection Table
86100 family plug-in module matrix
The 86100 has a family of plug-in modules designed for a broad
range of precision optical, electrical, and TDR/TDT measurements. The 86100D can accommodate up to 4 modules for a total
of 16 measurement channels.
622 Mb/s
1063 Mb/s
1244/1250 Mb/s
2125 Mb/s
2488/2500 Mb/s
2.666 Gb/s
20
62.5
–20
●
●
●
●
●
●
●
750 to 1650
8.5
20
62.5
–16
300 2 1
1
750 to 1650
8.5
20
62.5
–16
1
1
750 to 1650
20
35
62.5
–12
100
1
1
750 to 1650
20
35
62.5
–12
200
1
1
750 to 1650
20
35
62.5
–12
50
86105D
281
1
1
750 to 1650
34
62.5
–7
86115D 3, 6
002
2
0
750 to 1650
20
62.5
–12
102
2
0
750 to 1650
20
62.5
–12
142
2
0
750 to 1650
20
62.5
–12
004 5 4
0
750 to 1650
20
62.5
–11
104
4
0
750 to 1650
20
62.5
–11
144
4
0
750 to 1650
20
62.5
–11
86115D
282
2
0
750 to 1650
34
50
62.5
–7
86116C 3, 6
025
1
1
1300 to 1620
45
80
9
–10
86116C 3, 6
041
1
1
1300 to 1620
65
80
9
–5
1.
2.
3.
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
7
●
7
° °
7
7
° ° ●
7
7
7
° ° °
7
°
7
7
° ° ●
7
7
° ° ●
7
7
7
° ° °
7
7
●
° °
7
7
° ° ●
7
7
7
° ° °
7
°
7
°
°
7
7
°
7
7
°
7
7
°
7
7
°
7
7
°
7
●
7
7
° ° ° ° ° ° °
7
7
7
7
7
7
7
° ° ° ° ° ° °
7
°
7
°
7
°
●
°
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
●
°
°
7
°
7
°
7
°
7
°
7
°
7
°
7
●
●
●
●
●
●
●
●
●
●
●
●
●
7
7
7
7
7
7
7
7
7
●
●
43.018 Gb/s
●
●
°
7
41.250 Gb/s
●
●
●
7
39.813 Gb/s
●
●
●
●
28.05 Gb/s
●
●
7
27.70 Gb/s
°
●
25.80 Gb/s
●
●
14.025 Gb/s
11.317 Gb/s
11.096 Gb/s
10.709 Gb/s
10.664 Gb/s
10.51875 Gb/s
10.3125 Gb/s
8.50 Gb/s
●
°
●
9.953 Gb/s
Filtered data rates
3.125 Gb/s
4.25 Gb/s
5.00 Gb/s
6.25 Gb/s
86105D 3, 6
Fiber input (μm)
155 Mb/s
Unfiltered optical bandwidth (GHz)
Electrical bandwidth (GHz)
8.5
1
Probe power 1
Wavelength range (nm)
750 to 1650
1
TDR/TDT/S-parameters
1
200
No. of electrical channels
100 2 1
No. of optical channels
Mask test sensitivity (dBm) 9
86105C 6
Option
Module
Also consider the N1090 and N1092 DCA-M solutions from
Keysight which support additional data rates. Download the
N1090A, N1092A/B/C/D/E and N1094 DCA-M Optical and
Electrical Sampling Oscilloscopes datasheet from
www.keysight.com/find/N1092A.
10
7
●
●
°
7
●
° ° ° ° ° ° °
10
7
7
7
7
7
7
7
° ° ° ° ° ° ° ●
7
7
7
7
7
7
7
7
° ° ° ° ° ° ° °
7
7
° ° °
7
7
7
° ° °
7
7
7
° ° °
●
7
●
7
●
°
7
°
7
°
7
°
7
°
7
°
°
7
°
7
°
7
°
7
°
7
°
°
7
°
7
°
7
°
7
°
7
°
●
●
●
●
●
°
7
°
7
°
7
°
7
°
7
7
7
°
7
°
7
7
7
° ° °
7
7
7
° ° °
●
●
●
Module has receptacle to supply power for external probe.
Pick any four rates (155 Mb/s to 6.25 Gb/s).
This module is not compatible with the 86100A and 86100B Digital Communication Analyzer (DCA) mainframes. If you would like to upgrade older DCA’s
contact Keysight Technologies and ask for current trade-in deals.
4. The 86108A/B uses all module slots and includes an internal clock recovery and precision timebase capabilities.
5. 4 optical input ports are switched internally to two optical-to-electrical (O/E) converters.
6. All modules with optical channels can use Option IRC to enhance the effective operating range. Reference receivers can be created at any rate within
± 50% of the hardware capability. IRC also corrects non-ideal (but compliant) hardware performance to yield ideal reference receiver responses.
7.
With Option IRC. Requires FlexDCA using Pattern Lock, 86100D mainframe with Option ETR, and signals <= PRBS16-1. For more details, see application
note “Techniques for Higher Accuracy Optical Measurements” (Keysight literature number 5990-8812EN).
8. Requires 86100D mainframe (not compatible with 86100A/B/C).
9. Generally represents the power level where an ideal eye diagram will approach 0% mask margin due to noise of the oscilloscope. Provides a non-specified
figure of merit to compare sensitivities of various optical channels.
10. A 15 Gb/s filter is provided with Option 281 and 282 that closely approximates the response required for 16X Fibre Channel compliance testing. Contact
Keysight for a special option that verifies compliance of the 86105D-281 and 86115D-282 to the 16X Fibre Channel reference receiver specification.
10 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Selection Table (Continued)
86100 family plug-in module matrix (Continued)
54754A
86108A 3, 4
86108B 3, 4
LBW
HBW
86112A
HBW
86118A
N1045A 4
N1045A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1046A 4
N1055A 4
N1055A 4
N1055A 4
N1055A 4
02F/02M
04F/04M
71F
81F
11F
72F
82F
12F
74F
84F
14F
32F/32M
34F/34M
52F/52M
54F/54M
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
2
2
2
2
2
2
2
2
4
1
1
1
2
2
2
4
4
4
2
4
2
4
●
●
●
●
●
●
●
●
●
●
●
Electrical bandwidth (GHz)
Probe power 1
TDR/TDT/S-parameters
No. of electrical channels
No. of optical channels
Option
Module
The 86100 has a family of plug-in modules designed for a broad
range of precision optical, electrical, and TDR/TDT measurements. The 86100D can accommodate up to 4 modules for a total
of 16 measurement channels.
18
32
35
50
20
30
70
60
60
75
85
100
75
85
100
75
85
100
35
35
50
50
1. Module has receptacle to supply power for external probe.
2. This module is not compatible with the 86100A and 86100B Digital Communication Analyzer (DCA)
mainframes. If you would like to upgrade older DCA’s contact Keysight Technologies and ask for
current trade-in deals.
3. The 86108A/B uses all module slots and includes internal clock recovery and precision timebase
capabilities.
4. Requires 86100D mainframe (not compatible with 86100A/B/C).
11 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications
Single-mode and multimode optical/electrical
Multiple and single-mode
optical/electrical modules
Optical channel specifications
Optical channel unfiltered bandwidth
Wavelength range
Calibrated wavelengths
Optical sensitivity 1
Transition time (10 to 90% calculated from TR = 0.48/BW optical)
RMS noise
Characteristic
Maximum
1.
86105C
86105D
86105D
Option 281
8.5 GHz (9 GHz)
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– ≤ 2.666 Gb/s, –20 dBm
–– > 2.666 to ≤ 4.25 Gb/s, –19 dBm
–– > 4.25 to 11.3 Gb/s, –16 dBm
1310 nm/1550 nm:
–– ≤ 2.666 Gb/s, –21 dBm
–– > 2.666 to ≤ 4.25 Gb/s, –20 dBm
–– > 4.25 to 11.3 Gb/s, –17 dBm
20 GHz (22 GHz)
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– 8.5 to 11.3 Gb/s, –9 dBm
–– 14.025 Gb/s, –6 dBm
56 ps
24 ps
34 GHz
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– 15 Gb/s, –9 dBm
–– 25.78 Gb/s, –6 dBm
–– 27.85 to 28.05 Gb/s, –5 dBm
1310 nm/1550 nm:
–– 15 Gb/s, –8 dBm
–– 25.78 Gb/s, –7 dBm
–– 27.95 to 28.05 Gb/s, –6 dBm
1550 nm:
–– 15 Gb/s, –8 dBm
–– 25.78 Gb/s, –8 dBm
–– 27.95 to 28.05 Gb/s, –7 dBm
15 ps
850 nm:
–– ≤ 2.666 Gb/s, 1.3 μW
–– > 2.666 to ≤ 4.25 Gb/s, 1.5 μW
–– > 4.25 to 11.3 Gb/s, 2.5 μW
850 nm:
–– 8.5 to 11.3 Gb/s, 10 μW
–– 14.025 Gb/s, 16 μW
1310 nm/1550 nm:
–– ≤ 2.666 Gb/s, 0.8 μW
–– > 2.666 to ≤ 4.25 Gb/s, 1.0 μW
–– > 4.25 to 11.3 Gb/s, 1.4 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 5 μW
–– 14.025 Gb/s, 8 μW
1310 nm/1550 nm:
–– 8.5 to 11.3 Gb/s, –12 dBm
–– 14.025 Gb/s, –9 dBm
850 nm:
–– ≤ 2.666 Gb/s, 2.0 μW
–– > 4.25 Gb/s to 11.3 Gb/s, 4.0 μW
850 nm:
–– 8.5 to 11.3 Gb/s, 12 μW
–– 14.025 Gb/s, 24 μW
1310 nm/1550 nm:
–– ≤ 2.666 Gb/s, 1.3 μW
–– > 2.666 to ≤ 4.25 Gb/s, 1.5 μW
–– > 4.25 to 11.3 Gb/s, 2.5 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 7 μW
–– 14.025 Gb/s, 12 μW
850 nm:
–– 15 Gb/s, 9 µW
–– 25.78 Gb/s, 17 µW
–– 27.95 to 28.05 Gb/s, 18 µW
–– Unfiltered, 25 µW
1310 nm:
–– 25.78 Gb/s, 13 µW
–– Unfiltered, 18 µW
1550 nm:
–– 15 Gb/s, 8 µW
–– 25.78 Gb/s, 15 µW
–– 27.95 to 28.05 Gb/s, 17 µW
–– Unfiltered, 21 µW
850 nm:
–– 15 Gb/s, 15 µW
–– 25.78 Gb/s, 25 µW
–– 27.95 to 28.05 Gb/s, 25 µW
–– Unfiltered, 35 µW
1310 nm:
–– 15 Gb/s, 12 µW
–– 25.78 Gb/s, 20 µW
–– 27.95 to 28.05 Gb/s, 20 µW
–– Unfiltered, 25 µW
1550 nm:
–– 15 Gb/s, 14 µW
–– 25.78 Gb/s, 23 µW
–– 27.95 to 28.05 Gb/s, 23 µW
–– Unfiltered, 30 µW
Generally represents the power level where an ideal eye diagram will approach 0% mask margin due to noise of the oscilloscope. Provides a non-specified
figure of merit to compare sensitivities of various optical channels.
2. Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.
12 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical (Continued)
Multiple and single-mode
optical/electrical modules
Optical channel specifications
Optical channel unfiltered bandwidth
Wavelength range
Calibrated wavelengths
Optical sensitivity 1
86115D
Option 002, 102, 142 2
86115D
Option 004, 104, 144 2
86115D
Option 282
20 GHz (22 GHz)
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– 8.5 to 11.3 Gb/s, –9 dBm
–– 14.025 Gb/s, –6 dBm
20 GHz (22 GHz)
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– 8.5 to 11.3 Gb/s, –8 dBm
–– 14.025 Gb/s, –5 dBm
1310 nm/1550 nm:
–– 8.5 to 11.3 Gb/s, –12 dBm
–– 14.025 Gb/s, –9 dBm
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, –11 dBm
–– 14.025 Gb/s, –8 dBm
34 GHz
750 to 1650 nm
850 nm/1310 nm/1550 nm
850 nm:
–– 15 Gb/s, –9 dBm
–– 25.78 Gb/s, –6 dBm
–– 27.85 to 28.05 Gb/s, –5 dBm
1310 nm/1550 nm:
–– 15 Gb/s, –8 dBm
–– 25.78 Gb/s, –7 dBm
–– 27.95 to 28.05 Gb/s, –6 dBm
1550 nm:
–– 15 Gb/s, –8 dBm
–– 25.78 Gb/s, –8 dBm
–– 27.95 to 28.05 Gb/s, –7 dBm
15 ps
Transition time (10 to 90% calculated 24 ps
from TR = 0.48/BW optical)
RMS noise
Characteristic
850 nm:
–– 8.5 to 11.3 Gb/s, 10 μW
–– 14.025 Gb/s, 16 μW
Maximum
1.
24 ps
850 nm:
–– 8.5 to 11.3 Gb/s, 12 μW
–– 14.025 Gb/s, 20 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 5 μW
–– 14.025 Gb/s, 8 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 6 μW
–– 14.025 Gb/s, 10 μW
850 nm:
–– 8.5 to 11.3 Gb/s, 12 μW
–– 14.025 Gb/s, 24 μW
850 nm:
–– 8.5 to 11.3 Gb/s, 14 μW
–– 14.025 Gb/s, 30 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 7 μW
–– 14.025 Gb/s, 12 μW
1310/1550 nm:
–– 8.5 to 11.3 Gb/s, 8.5 μW
–– 14.025 Gb/s, 14 μW
850 nm:
–– 15 Gb/s, 9 µW
–– 25.78 Gb/s, 17 µW
–– 27.95 to 28.05 Gb/s, 18 µW
–– Unfiltered, 25 µW
1310 nm:
–– 25.78 Gb/s, 13 µW
–– Unfiltered, 18 µW
1550 nm:
–– 15 Gb/s, 8 µW
–– 25.78 Gb/s, 15 µW
–– 27.95 to 28.05 Gb/s, 17 µW
–– Unfiltered, 21 µW
850 nm:
–– 15 Gb/s, 15 µW
–– 25.78 Gb/s, 25 µW
–– 27.95 to 28.05 Gb/s, 25 µW
–– Unfiltered, 35 µW
1310 nm:
–– 15 Gb/s, 12 µW
–– 25.78 Gb/s, 20 µW
–– 27.95 to 28.05 Gb/s, 20 µW
–– Unfiltered, 25 µW
1550 nm:
–– 15 Gb/s, 14 µW
–– 25.78 Gb/s, 23 µW
–– 27.95 to 28.05 Gb/s, 23 µW
–– Unfiltered, 30 µW
Generally represents the power level where an ideal eye diagram will approach 0% mask margin due to noise of the oscilloscope. Provides a non-specified
figure of merit to compare sensitivities of various optical channels.
2. Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.
13 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical (Continued)
Multiple and single-mode
optical/electrical modules
86105C
Optical channel specifications (Continued)
Scale factor (per division)
Minimum
2 μW
Maximum
100 μW
CW 1 accuracy (single marker,
Single-mode:
referenced to average power
± 25 μW ± 3%
monitor)
Multimode:
± 25 μW ± 10%
CW offset range (referenced two +0.2 to –0.6 μW
divisions from screen bottom)
Average power monitor
–30 to 0 dBm
(specified operating range)
Average power monitor accuracy
Single-mode
± 5% ± 200 nW
± Connector
Uncertainty
Multimode (characteristic)
86105D
86105D
Option 281
86115D
Option 002, 102,
142 2
86115D
Option 004, 104,
144 2
20 μW
20 µW
500 μW
500 µW
Single-mode:
–– ± 25 µW ± 4% (14.025 Gb/s)
–– ± 25 μW ± 2%
–– ± 25 µW ± 4% (15.0 Gb/s)
(8/10 Gb/s)
–– ± 25 µW ± 6%
–– 4% (14 Gb/s)
(25.78125 Gb/s)
–– 6%
–– ± 25 µW ± 6%
(Unfiltered)
(27.7393 Gb/s to 28.05 Gb/s)
–– ± 25 µW ± 6% (Unfiltered)
Multimode:
± 25 μW ± 10%
+1 μW/–3 μW
+1 to –3 mW
20 μW
20 μW
500 μW
500 μW
Single-mode:
Single-mode:
–– ± 25 μW ± 2%
–– ± 25 μW ± 2%
(8/10 Gb/s)
(8/10 Gb/s)
–– 4% (14 Gb/s)
–– 4% (14 Gb/s)
–– 6%
–– 6%
(Unfiltered)
(Unfiltered)
Multimode:
± 25 μW ± 10%
+1 μW/–3 μW
Multimode:
± 25 μW ± 10%
+1 μW/–3 μW
–30 to +3 dBm
–30 to +3 dBm
–30 to +3 dBm
–30 to +3 dBm
± 5% ± 200 nW
± Connector
Uncertainty
± 5% ± 200 nW
± Connector
Uncertainty
± 5% ± 100 nW
± Connector
Uncertainty
(20 to 30 °C)
± 5% ± 200 nW
± Connector
Uncertainty
± 5% ± 200 nW
± Connector
Uncertainty
± 10% ± 200 nW ± 5% ± 200 nW
± Connector
± Connector
Uncertainty
Uncertainty
± 10% ± 200 nW
± Connector
Uncertainty
1. CW refers to an unmodulated optical signal.
2. Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.
± 5% ± 200 nW
± Connector
Uncertainty
14 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical (Continued)
Multiple and single-mode
86105C
optical/electrical modules
Optical channel specifications (Continued)
User calibrated accuracy
Single-mode
± 3% ± 200 nW
± Power meter
Uncertainty
< 5 °C change
Multimode (characteristic)
± 10% ± 200 nW
± Power meter
Uncertainty
< 5 °C change
Maximum input power
Maximum non-destruct average
0.5 mW (–3 dBm)
86105D
86105D
Option 281
86115D
Option 002, 102, 142 1
86115D
Option 004, 104, 144 1
± 2% ± 100 nW
± Power meter
Uncertainty
± 3% ± 200 nW
± Power meter
Uncertainty
± 2% ± 100 nW
± Power meter
Uncertainty
± 10% ± 200 nW
± Power meter
Uncertainty
± 10% ± 200 nW
± Power meter
Uncertainty
± 2% ± 100 nW
± Power meter
Uncertainty
< 5 °C change
± 10% ± 200 nW
± Power meter
Uncertainty
5 mW (7 dBm)
5 mW (7 dBm)
5 mW (7 dBm)
Maximum non-destruct peak
Fiber input
10 mW (10 dBm)
62.5/125 μm
User-selectable
Connector
5 mW (+7 dBm)
1310/1550 nm
Single mode:
5 mW (+8 dBm)
850 nm
Multimode:
3 mW (+5 dBm)
5 mW (+10 dBm)
62.5/125 µm
10 mW (10 dBm)
62.5/125 μm
User-selectable
Connector
10 mW (10 dBm)
62.5/125 μm
User-selectable
Connector
27 dB single-mode
27 dB single-mode
14 dB multimode
14 dB multimode
Polarization dependent loss at
1550 nm
Input return loss (HMS-10 connector
fully filled fiber)
5 mW (+7 dBm)
62.5/125 μm
850 nm > 13 dB
27 dB single-mode
1310 nm/1550 nm
14 dB multimode
> 24 dB
1.
0.2 dB
(characteristic)
> 24 dB (1550 nm
single mode)
> 24 dB (1310 nm
single mode)
> 13 dB (850 nm
multimode)
Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.
± 10% ± 200 nW
± Power meter
Uncertainty
15 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical (Continued)
Multiple and single-mode
optical/electrical modules
Electrical channel specifications
Electrical channel bandwidth
Transition time (10 to 90% calculated
from TR = 0.35/BW)
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
DC offset range (referenced to center
of screen)
Input dynamic range (relative to
channel offset)
Maximum input signal
Nominal impedance
Reflections (for 30 ps rise time)
Electrical input
1.
86105C
86105D
86105D
Option 281
12.4 and 20 GHz
28.2 ps (12.4 GHz)
17.5 ps (20 GHz)
25 and 35 GHz
14 ps (25 GHz)
10 ps (35 GHz)
25 and 35 GHz
12 ps (25 GHz)
7 ps (50 GHz)
0.25 mV (12.4 GHz)
0.5 mV (20 GHz)
0.5 mv (12.4 GHz)
1 mV (20 GHz)
0.25 mV (25 GHz)
0.5 mV (35 GHz)
0.5 mV (25 GHz)
1 mV (35 GHz)
0.25 mV (25 GHz)
0.6 mV (50 GHz)
0.5 mV (25 GHz)
1 mV (50 GHz)
86115D
Option 002, 102,
142 1
1 mV/division
100 mV/division
± 0.4% of full scale ± 2 mV ± 1.5% of (reading-channel offset), 12.4 GHz
± 0.4% of full scale ± 2 mV ± 3% of (reading-channel offset), 20 GHz
± 500 mV
± 400 mV
± 2 V (+16 dBm)
50 Ω
5%
3.5 mm (male)
2.4 mm (male)
Options 102 and 104 do not include filters for 14.025 Gb/s. Options 142 and 144 include filters for 14.025 Gb/s and no other rates.
86115D
Option 004, 104,
144 1
16 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode optical/electrical
High bandwidth single-mode optical/electrical modules
Optical channel specifications
Optical channel unfiltered bandwidth
Wavelength range
Calibrated wavelengths
Optical sensitivity 2
1310 nm
1550 nm
Transition time (10 to 90% calculated
from Tr = 0.48/BW optical)
RMS noise
Characteristic
1310 nm
1550 nm
Maximum
1310 nm
1550 nm
86116C
Option 025
86116C
Option 041
45 GHz
1300 to 1620 nm 3
1300 nm/1550 nm
–9 dBm (17 Gb/s)
–8 dBm (25.8 Gb/s)
–7 dBm (27.7 Gb/s)
–10 dBm (17 Gb/s)
–9 dBm (25.8 Gb/s)
–8 dBm (27.7 Gb/s)
7.4 ps (FWHM) 1
65 GHz
13 μW (17 Gb/s)
17 μW (25.8 Gb/s)
20 μW (27.7 Gb/s)
60 μW (40 GHz)
10 μW (17 Gb/s)
12 μW (25.87 Gb/s)
14 μW (27.7 Gb/s)
40 μW (40 GHz)
18 μW (17 Gb/s)
20 μW (25.8 Gb/s)
30 μW (27.7 Gb/s)
120 μW (40 GHz)
15 μW (17 Gb/s)
18 μW (25.8 Gb/s)
21 μW (27.7 Gb/s)
80 μW (40 GHz)
54 μW (39.8/43.0 Gb/s)
75 μW (55 GHz)
105 μW (60 GHz)
187 μW (65 GHz)
36 μW (39.8/43.0 Gb/s)
50 μW (55 GHz)
70 μW (60 GHz)
125 μW (65 GHz)
102 μW (39.8/43.0 Gb/s)
127 μW (55 GHz)
225 μW (60 GHz)
300 μW (65 GHz)
68 μW (39.8/43.0 Gb/s)
85 μW (55 GHz)
150 μW (60 GHz)
200 μW (65 GHz)
–3 dBm (39.8/43.0 Gb/s)
–5 dBm (39.8/43.0 Gb/s)
1. FWHM (full width half max) as measured from optical pulse with 700 fs FWHM, 5 MHz repetition rate and 10 mW peak power.
2. Generally represents the power level where an ideal eye diagram will approach 0% mask margin due to noise of the oscilloscope. Provides a non-specified
figure of merit to compare sensitivities of various optical channels.
3. Contact Keysight for broader wavelength specifications.
17 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Single-mode and multimode optical/electrical
High bandwidth single-mode optical/electrical modules
Optical channel specifications (Continued)
Scale factor
Minimum
Maximum
CW1 accuracy (single marker, reference to average power
monitor)
CW offset range (referenced two divisions from screen
button)
Average power monitor (specified operating range)
Factory calibrated accuracy
User calibrated accuracy
Maximum input power
Maximum non-destruct average
Maximum non-destruct peak
Fiber input
Input return loss (HMS-10 connector fully filled fiber)
Electrical channel specifications
Electrical channel bandwidth
Transition time (10 to 90% calculated from Tr = 0.35/BW)
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
DC offset range (referenced to center of screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Reflections (for 20 ps rise time)
Electrical input
1.
CW refers to an unmodulated optical signal.
86116C
200 μW/division
5 mW/division
± 150 µW ± 4% (reading-channel offset)
+8 to –12mW
–23 to +9 dBm
± 5% ± 100 nW ± connector uncertainty, 20 to 30 °C
± 2% ± 100 nW ± power meter uncertainty, < 5 °C change
10 mW (+10 dBm)
50 mW (+17 dBm)
9/125 μm, user-selectable connector
20 dB
80 (93), 55 and 30 GHz
6.4 ps (55 GHz)
4.4 ps (80 GHz)
0.5 mV (30 GHz)
0.6 mV (55 GHz)
1.1 mV (80 GHz)
0.8 mV (30 GHz)
1.1 mV (55 GHz)
2.2 mV (80 GHz)
2 mV/division
100 mV/division
± 0.4% of full scale
± 3 mV ± 2% of (reading-channel offset)
± 2% of offset (all bandwidths)
± 500 mV
± 400 mV
± 2 V (+16 dBm)
50 Ω
10% (DC to 70 GHz)
20% (70 to 100 GHz)
1.85 mm (male)
18 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Dual electrical
Dual electrical channel modules
Electrical channel bandwidth
Transition time (10 to 90% calculated from
TR = 0.35/BW)
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
86112A
12.4 and 20 GHz (30 GHz 1)
28.2 ps (12.4 GHz)
17.5 ps (20 GHz)
54754A
12.4 and 18 GHz
28.2 ps (12.4 GHz)
19.4 ps (18 GHz)
0.25 mV (12.4 GHz)
0.5 mV (20 GHz)
0.5 mv (12.4 GHz)
1 mV (20 GHz)
0.25 mV (12.4 GHz)
0.5 mV (18 GHz)
0.5 mv (12.4 GHz)
1 mV (18 GHz)
1 mV/division
100 mV/division
± 0.4% of full scale
± 2 mV ± 1.5% of (reading-channel offset)
(12.4 GHz) ± 0.4% of full scale
± 2 mV ± 3% of (reading-channel offset) (20 GHz)
DC offset range (referenced from center of
screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Reflections (for 30 ps rise time)
Electrical input
Dual electrical channel modules
Electrical channel bandwidth
Transition time (10 to 90% calculated
from TR = 0.35/BW)
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
DC offset range (referenced from center of
screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Reflections (for 30 ps rise time)
Electrical input
1.
± 0.4% of full scale
± 2 mV ± 0.6% of (reading-channel offset)
(12.4 GHz) ± 0.4% of full scale or marker reading
(whichever is greater)
± 2 mV ± 1.2% of (reading-channel offset)
(18 GHz)
± 500 mV
± 400 mV
± 2 V (+16 dBm)
50 Ω
5%
3.5 mm (male)
86118A
50 and 70 GHz
11.7 ps (30 GHz)
7 ps (50 GHz)
0.7 mV (50 GHz)
1.3 mV (70 GHz)
1.8 mv (50 GHz)
2.5 mV (70 GHz)
1 mV/division
100 mV/division
± 0.4% of full scale
± 2 mV ± 2% of (reading-channel offset), (50 GHz)
± 0.4% of full scale
± 2 mV ± 4% of (reading-channel offset), (70 GHz)
± 500 mV
± 400 mV
± 2 V (+16 dBm)
50 Ω
20%
1.85 mm (male)
A special option, 86112-HBW, extends the bandwidth from 20 to 30 GHz. All other specifications remain unchanged.
19 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Dual electrical (Continued)
Dual electrical channel modules
Bandwidth (low/high BW setting)
Transition time (10 to 90% calculated
from Tr = 0.35/BW)
RMS noise
Characteristic (low/high BW setting)
Maximum (low/high BW setting)
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
Low BW setting
High BW setting
CW offset range (referenced from center of screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Reflections (for 30 ps rise time)
Electrical input
CH1 to CH2 skew
86108A
16 and > 32 GHz, (35 GHz)
10 ps
86108B-LBW
20 and 35 GHz
10 ps
86108B-HBW
35 and 50 GHz
7 ps
240 µV/420 µV
350 µV/700 µV
300 μV/500 μV
350 μV/700 μV
600 μV/750 μV
800 μV/980 μV
2 mV/division
100 mV/division
1 mV/division
140 mV/division
1 mV/division
140 mV/division
± 0.7% of full scale, ± 2 mV ± 1.5% of (reading-channel offset) (16 GHz)
± 0.7% of full scale, ± 2 mV ± 3% of (reading-channel offset) (32 GHz)
± 500 mV
± 700 mV
± 700 mV
± 400 mV
± 560 mV
± 560 mV
± 2 V (+16 dBm)
± 2.5 V (+18 dBm)
± 2.5 V (+18 dBm)
50 Ω
50 Ω
50 Ω
5%
5%
5%
3.5 mm
3.5 mm
2.4 mm
< 12 ps
< 10 ps
< 10 ps
20 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Dual electrical (Continued)
Clock recovery
Data rate input range (continuous tuning)
86108A
0.05 to 14.2 Gb/s (requires FW
revision 8.1 or higher)
0.025 to 6.75 GHz
86108B-216
0.05 to 16 Gb/s
86108B-232
0.05 to 32 Gb/s
0.025 to 8 GHz
0.025 to 16 GHz
Minimum input level to aquire lock
175 mVpp
175 mVpp
175 mVpp
Recovered clock random jitter (used as
internal trigger) 1
Internal recovered clock trigger
< 500 fs at 2 Gb/s
< 400 fs at 5 and 10 Gb/s
0.015 to 10 MHz
< 350 fs at data rate < 2 Gb/s
< 300 fs at data rate ≥ 2 Gb/s
0.015 to 20 MHz
< 350 fs at data rate < 2 Gb/s
< 300 fs at data rate ≥ 2 Gb/s
0.015 to 20 MHz
Clock frequency input range (continuous
tuning)
Clock recovery adjustable loop bandwidth
range (user selectable)
Clock recovery loop peaking range
Loop bandwidth accuracy
Tracking range (includes spread spectrum
tracking)
Aquisition range
Maximum consecutive identical digits to lock
Auto relocking
Residual spread spectrum
Front panel recovered clock amplitude
Front panel recovered clock divide ratio
(user selectable)
Recovered clock front panel connector type
Internal frequency counter accuracy
1.
Up to 4 settings (dependent on loop BW)
± 30%
± 2500 ppm ± 0.25%
± 5000 ppm
150
If signal lock is lost, system can automatically attempt to regain phase-lock.
User selectable to enable/disable
–72 ± 3 dB at 33 kHz
–84 ± 3 dB at 33 kHz
–84 ± 3 dB at 33 kHz
0.15 to 1.0 Vpp (0.3 to 1.0 Vpp)
1, 2, 4, 8, 16
1, 2, 4, 8, 16
1, 2, 4, 8, 16
2, 4, 8, 16
2, 4, 8, 16
2, 4, 8, 16
SMA
± 10 ppm
These values indicate performance with the internal precision timebase disabled. With precision timebase enabled, system jitter performance is
dramatically decreased. See page 9.
Phase trimmers
Nominal impedance
Frequency range
Insertion loss (at f max)
Time delay
Connectors
86108B-PT3 (= qty 2 N1027A-PT3)
50 Ω
DC to 26 GHz
0.8 dB
238 to 293 ps
SMA (m-f)
86108B-PT2 (= qty 2 N1027A-PT2)
50 Ω
DC to 50 GHz
0.8 dB
172 to 195 ps
2.4 mm (m-f)
21 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Dual/quad electrical
Dual/Quad electrical channel modules
Channel options
(number of channels; F = female; M = male)
Number of channels 2
Electrical channel bandwidth
Transition time (characteristic)
(10 to 90% calculated from TR = 0.35/BW)
Channel-to-channel skew range
RMS noise
Characteristic
Maximum
Scale factor (per division)
Minimum
Maximum
DC accuracy (single marker)
02F/02M
DC offset range (referenced from center of
screen)
Input dynamic range (relative to channel offset)
Maximum input signal
Nominal impedance
Reflections (for 30 ps rise time)
Electrical input 3
N1045A 1
04F/04M
32F/32M
N1055A 1
34F/34M
52F/52M
54F/54M
4
35 GHz 4, 5
4
35/50 GHz 5
2
4
20/35/45/60 GHz
17.5/10/7.8/5.8 ps
2
35 GHz 4, 5
10 ps
2
35/50 GHz 5
10 /7 ps
± 100 ps
± 150 ps
275/425/500/750 µV
950 µV (60 GHz)
600 µV
730 µV
1 mV/division
100 mV/division
± 2 mV ± 4% of (reading-channel
offset), (60 GHz)
± 500 mV
1 mV/division
100 mV/division
± 2 mV ± 4% of (reading-channel offset), (50 GHz)
± 400 mV
± 2 V (+16 dBm)
50 Ω
20%
1.85 mm (female or male option)
± 400 mV
+2 V/–1 V
50 Ω
20%
2.92 mm (female or male option)
600/750 µV
950 µV
± 500 mV
1.85 mm (female or male option)
TDR system
TDR system
(Mainframe with 54754A module)
Rise time
TDR step flatness
Low level
High level
TDR system
(Mainframe with N1055A module)
Step rise/fall time (transmitted 6)
1.
2.
3.
4.
5.
6.
Oscilloscope/TDR performance
Normalized characteristics
40 ps nominal
< 25 ps normalized
≤ ± 1% after 1 ns from edge
≤ + 5%, –3% < 1 ns from edge
0.00 V ± 2 mV
± 200 mV +2 mV
N1055A - 3xx
(35 GHz bandwidth)
Without TDR calibration
With TDR calibration
< 18 ps
Adjustable from 15 ps,
characteristic
Adjustable from larger of 17 ps or 0.08 x time/div
Maximum: 5 x time/div
≤ 0.1%
N1055A - 5xx
(50 GHz bandwidth)
Without TDR calibration
< 7 ps
With TDR calibration
Adjustable from 6 ps,
characteristic
Module is supported by 86100D DCA-X mainframe and later.
Upgradable from two channel to four channel after purchase (return to Keysight).
Connector style is the same on all channels and is selected at time of order.
Upgradable from 35 to 50 GHz after purchase (return to Keysight).
Tuned to be –3 dB (± measurement uncertainty) at stated bandwidth(s) using NIST traceable swept-sine test system.
Defined as the transition time at the output of the remote head. It is calculated by deconvolving the receiver transition time from the measured transition
time when the remote head is terminated with a short. Measured on a negative TDR step.
22 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Specifications describe warranted performance while characteristics provide useful, nonwarranted information and performance.
Characteristics are listed below in italics.
Multiple electrical channel modules
Channel options
Number of channels
Electrical channel bandwidth
(maximum) 1
Electrical channel bandwidth settings
(user-selectable)
Transmision time (10 to 90% calculated
from TR = 0.35/bandwidth)
Channel-to-channel skew range
RMS noise
Characteristic
71F/72F/74F
1/2/4
75 GHz
N1046A
81F/82F/84F
1/2/4
85 GHz
45, 60, 75 GHz
45, 60, 75, 85 GHz
7.8, 5.9, 4.7 ps
7.8, 5.9, 4.7, 4.2 ps
45, 50, 75, 85, 100, 122 GHz
(characteristic)
7.8, 5.9, 4.7, 4.2, 3.5, < 3.2 ps
± 100 ps
± 100 ps
± 100 ps
400, 580, 740 μV
400, 580, 740, 860 μV
Maximum
550, 700, 950 μV
550, 700, 950, 1150 μV
400, 580, 740, 860, 1005,
2000 μV
550, 700, 950, 1150, 1350, 2000
(characteristic) μV
Scale factor (per division)
–– Minimum
–– Maximum
DC accuracy (VAVG measurement)
1 mV/division
100 mV/division
1 mV/division
100 mV/division
–– Specified at calibration temperature ± 0.5 °C. (Perform a new
module calibration if hardware
skew has been applied.)
–– Specified at calibration temperature ± 5 °C
DC offset range (referenced from
center of screen)
Input dynamic range (relative to
channel offset)
Maximum input signal
Nominal impedance (characteristic)
11F/12F/14F
1/2/4
100 GHz
1 mV/division
100 mV/division
± 2 mV (characteristic)
± 2 mV ± 4% of (reading – channel offset)
± 500 mV
± 500 mV
± 500 mV
± 400 mV
± 400 mV
± 400 mV
± 2V (+16 dBm)
50 Ω
± 2V (+16 dBm)
50 Ω
± 2V (+16 dBm)
50 Ω
How does Keysight test and setup your module?
The N1046A bandwidth is tested using a very fast electrical impulse that is generated by a photodiode driven by a femtosecond optical
impulse source. The high bandwidth photodiode is calibrated and NIST traceable through 110 GHz. For frequencies beyond 110 GHz,
photodiode performance is based on a linear extrapolation of the photodiode response through 110 GHz but test data beyond 110 GHz
is not provided.
User note:
For customers needing the maximum possible bandwidth, the 122 GHz setting is recommended. For customers needing high bandwidth with “matched” performance across channels, the 100 GHz setting is recommended.
1.
Tuned to be —3 dB (± measurement uncertainty) at stated bandwidth(s) EXCEPT for the 122 GHz setting which is tuned for maximum bandwidth while
keeping channel noise ≤ 2.5 mV RMS.
23 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Modules Specifications (Continued)
Clock recovery
The DCA platform offers clock recovery capability in the form of standalone clock recovery solutions such as the N1076/N1077A. For
more information on standalone clock recovery solutions, please download the Electrical and Optical Clock Data Recovery Solutions
datasheet which covers the N1076, N1077 and N1078 solutions. It is available from www.keysight.com/find/N1077A
Typical System Configurations
86100D Infiniium DCA-X Mainframe
86100D hardware options
86100D software options (select any):
Trigger Options (select one only):
STR – Standard – for basic eye measurements only
ETR – Enhanced – for pattern waveforms, jitter analysis
and some TDR applications
PTB – Integrated precision timebase for low residual jitter
061/062 – Add matlab analysis package
200 – Enhanced jitter analysis
201 – Advanced waveform analysis
202 – Enhanced Impedance and S-Parameters
300 – Advanced amplitude analysis/RIN/Q-scale
500 – Productivity package
SIM – InfiniiSim-DCA de-embedding/embedding
86100DU – 400 PLL and jitter spectrum analysis
86100D – 401 advanced eye analysis (Jitter on PRBS31)
Remote Connection Options (select one only):
GPI – GPIB card interface installed
GPN – No GPIB card
DCA plug-in modules (for typical 1 applications)
Electrical/PLL
1 to 12 Gb/s
86112A
Dual channels
BW > 20 GHz each
83496B
Electrical clock recovery
(#100, 200, 300)
Electrical/PLL
1 to 16/32 Gb/s
(high performance)
86108B
Dual channels
BW > 35/50 GHz
with integrated precision
timebase & clock recovery
Electrical
Optical
20, 40 Gb/s
(high performance)
1 to 12Gb/s
86118A
9 GHz optical channel
20 GHz electrical channel
Dual remote heads
BW > 70 GHz each
86107A or
86100D-PTB
precision timebase
(#40)
86105C
Optical
20, 40 Gb/s
86105D-281
34 GHz optical channel
(780 to 1630 nm)
86116C
65 GHz optical channel
80 GHz electrical channel
86107A (#40) or
86100D-PTB
TDR/TDT
Single-ended, differential
54754A
Dual 18 GHz channels
N1055A
2 or 4 channels
35 or 50 GHz channels
Note - TDR/TDT modules
also perform as
electrical-only receivers
precision timebase
1.
Contact your local Keysight sales representative to help configure a system for your specific application.
24 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Software Applications
86100D-200 enhanced jitter analysis
86100D-SIM InfiniiSim-DCA
Option 200 provides extensive and accurate jitter decomposition,
which is increasingly important as data rates increase and margins
reduce. Quickly customize your view of many parameters and take
advantage of advanced features such as jitter spectrum analysis.
At higher data rates, the effect of measurement fixtures and
channels is prominent and can be accounted for through the
de-embedding capabilities within Option SIM. You may also view
measurements on live and simulated signals, allowing you to
compare design intent to actual performance.
86100D-201 advanced waveform analysis
Take advantage of powerful features in Option 201 to generate
much deeper waveform files, integrate MATLAB analysis, and use
equalizers (CTLE, LFE, DFE), User Operators and User Defined
Measurements.
86100D-202 enhanced impedance and S-parameter
Option 202 works in TDR mode and provides live S-parameter
responses, including magnitude (return loss, insertion loss, etc.),
phase and group delay. Single-ended, differential and mixedmode responses are available simultaneously. Also enables quick
and easy saving of measurement results in Touchstone files (.s2p,
.s4p, etc.).
86100D-300 amplitude analysis/RIN/Q-factor
Extend the many capabilities from jitter mode into the amplitude
domain and see the decomposition of the amplitude into several
factors. Option 300 also reports relative intensity noise (RIN) and
Q-factor.
86100D-400 PLL and jitter spectrum
Option 400 quickly characterizes key parameters of phase-locked
loops and provides the jitter spectrum of your signal including
spread spectrum clocks. Control of the jitter source and receiver
is integrated to ensure fast results.
86100D-401 advanced EYE analysis
For device testing with long patterns and to obtain BER-contour
mask testing, Option 401 integrates with the classic or FlexDCA
interfaces to decompose the jitter and amplitude interference
measurements into the key parameters. When using the
embedded capability within FlexDCA or the included automation
application, you may characterize jitter on simultaneous multiple
lanes and obtain concise and visual results.
86100D-9FP PAM-N analysis software
Accurately and quickly analyze Pulse Amplitude Modulation
(PAM) signals as described in IEEE 802.3bj Clause 94 and
developing standards such as OIF-CEI-56G and IEEE 400G.
86100D-TFP IEEE TDECQ Analysis software
A lower cost subset of option 9FP capabilities that targets optical
Transmitter production test applications
N1012A OIF CEI compliance and debug application
The N1012A application lets you fully characterize the ~120 transmitter test parameters (including 28G-VSR), reducing your test time
from hours to minutes. Included are return loss tests and several
utilities to improve your test productivity. Debug mode enables you
to characterize your devices well beyond the parameters prescribed
in the implementation agreements.
N1014A SFF-8431 compliance and debug
application
The N1014A application lets you fully characterize ~70 test
parameters including all transmitter tests for host, module and host
copper and all test signals for receiver testing. Return loss tests and
utilities are also included.
N1019A user-defined application
You may create your own test application or suite of tests using the
DCA-X and other instruments. Quickly and intuitively create groups
of tests, test descriptions, user prompts and test limits. Use concise
HTML reports to share your multiple test results with your users.
N1081-4A IEEE802.3 Ethernet applications
The extensive requirements for IEEE802.3-2012, 802.3bj and
802.3bm are covered in these four applications, which comprise
> 400 tests. Characterize your device for one, four or ten lanes, and
analyze trends over time/temperature and between devices.
N1085A PAM4 Ethernet and OIF-CEI
Measurement Application
The N1085A software application covers PAM-4 transmitter
measurements outlined in IEEE P802.3bs and four OIF-CEI-4.0
(56G) clauses.
86100D-BFP Automatic Fixture Removal (AFR)
When measuring physical layer devices with non-coaxial interfaces,
test fixtures or probes are often used to connect the device under
test (DUT) to the measurement equipment. For accurate measurements of the DUT, the fixtures or probes need to be characterized
and their effects removed from the composite measurement of
the DUT plus test fixture or probe combination. This option adds
the Automatic Fixture Removal (AFR) capability from Keysight’s
powerful N1930B Physical Layer Test System (PLTS) software to
help you characterize a fixture or probe with non-coaxial interfaces.
Fixtures and probes to be characterized can be single-ended or
differential.
25 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Measurements
The 86100D DCA-X features two user interfaces for optimum
ease-of-use. It includes the classic DCA interface for complete
backwards compatibility with earlier DCA mainframes. It also
includes the new FlexDCA interface that provides new measurements and powerful analysis capability in a fully customizable
application.
The following measurements are available from the tool bar, as
well as the pull down menus. The available measurements depend
on the DCA-X operating mode.
Oscilloscope mode
–– Time
Rise time, fall time, jitter RMS, jitter p-p, period, frequency,
+ pulse width, – pulse wWidth, duty cycle, delta time, [Tmax,
Tmin, Tedge—remote commands only]
–– Amplitude
Overshoot, average power, V amptd, V p-p, V rms, V top,
V base, V max, V min, V avg, OMA (optical modulation
amplitude)
Eye/mask mode
–– NRZ eye measurements
Extinction ratio, jitter RMS, jitter p-p, average power,
crossing percentage, rise time, fall time, one level, zero Level,
eye height, eye width, signal to noise, duty cycle distortion,
bit rate, eye amplitude
–– RZ eye measurements
Extinction ratio, jitter RMS, Jitter p-p, average power,
rise time, fall time, one level, zero level, eye height, eye
amplitude, opening factor, eye width, pulse width, signal to
noise, duty cycle, bit rate, contrast ratio
Mask test
–– Open mask, start mask test, exit mask test, filter, mask test
margins, mask margin to a hit ratio, mask test scaling, create
NRZ mask
Option 201 advanced waveform analysis
–– Measurements
Deep memory pattern waveform, user-defined
measurements through MATLAB interface
–– Data displays
Equalized waveform
Option 202 enhanced impedance and
S-parameters
–– Markers and delta markers on live frequency domain
responses such as S-parameter magnitude, phase and group
delay
Option 300 amplitude analysis/RIN/Q-factor
(requires Option 200)
–– Measurements
Total interference (TI), deterministic interference (dual-dirac
model, DI), random noise (RN), periodic interference (PI), and
Inter-symbol interference (ISI), RIN (dBm or dB/Hz), Q-factor
–– Data displays
TI histogram, RN/PI histogram, ISI histogram
Option 400 PLL and jitter spectrum measurement
software
–– Jitter spectrum/phase noise measurements integrated jitter:
Total jitter (TJ), random jitter (RJ), deterministic jitter (DJ);
DJ amplitude/frequency, jitter spectrum graph, jitter versus
time graph, frequency versus time graph, jitter histogram,
post processed jitter measurements, phase noise graph
dBc/Hz versus frequency
–– Phase locked loop (PLL) measurements PLL bandwidth,
PLL peaking, data rate, jitter transfer function (JTF) graph,
observed jitter transfer (OJTF) graph, JTF model.
Option 401 advanced eye analysis
–– * Jitter analysis on long patterns
Option 200 enhanced jitter analysis software
–– Measurements
Total jitter (TJ), random jitter (RJ), deterministic jitter (DJ),
periodic jitter (PJ), data dependent jitter (DDJ), duty cycle
distortion (DCD), intersymbol interference (ISI), sub-rate jitter
(SRJ), asynchronous periodic jitter frequencies, subrate jitter
components
–– FlexDCA adds the following measurements:
Data dependent pulse width shrinkage (DDPWS),
uncorrelated jitter (UJ), J2, J9
–– Data displays
TJ histogram, RJ/PJ histogram, DDJ histogram, composite
histogram, DDJ versus bit position, bathtub curve (log or Q
scale)
FlexDCA version: Total jitter (TJ), deterministic jitter (DJ), random
jitter (RJ), J2, J5, J9. Jitter can be measured on long patterns
such as PRBS23, PRBS31, live traffic.
Microsoft Excel version: All of the above, plus BER contour mask
testing.
26 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Measurements (Continued)
Option 500 productivity package including TDEC
Option 500 enables Rapid Eye acquisition and TDEC (transmitter
dispersion and eye closure) measurements. Rapid Eye achieves
two significant benefits. First, unlike conventional sampling
and data display, when an eye mask test is performed, every
acquired sample will be compared to the mask, as the central
eye is composed of all acquired samples. Effective throughput is
improved at least 60%. Second, incomplete eye diagram displays
that can occur when triggering at sub-rates are eliminated. TDEC
is a complex eye diagram measurement required for new high
data rate standards deployed over multimode fiber. It is easily
and quickly achieved with the automated TDEC measurement.
(IEEE 802.3bm and T11 FC-PI-6P require a specific TDEC channel
bandwidth, achieved with 86105D/86115D-Option 168.)
–– Time at max
–– Time at min
–– Time at amplitude
–– Limit lines
–– Time at edge
Additional capabilities
Standard functions
–– Standard functions are available through pull down menus
and soft keys, and some functions are also accessible
through the front panel knobs.
Markers
–– Two vertical and two horizontal (user-selectable)
Option 401 advanced EYE analysis
–– Jitter measurements
Total jitter (TJ), random jitter (RJ), deterministic jitter (DJ), J2
jitter (J2), J5 jitter (J5), J9 jitter (J9)
–– Amplitude measurements
Total interference (TI), random noise (RN), deterministic
interference (DI), eye opening
–– Mask test
Pass/fail status, BER limit
Option SIM InfiniiSim-DCA
2-port de-embedding and embedding; 4-port de-embedding and
embedding; add simulated random jitter and noise
Option 9FP PAM-N analysis software
–– Electrical and optical signals
–– Eye width, eye height, eye skew
–– Level amplitude, level noise, level skew
–– Linearity measurements
–– TDECQ, Outer OMA, Outer ER
Option TFP IEEE TDEFCQ Analysis software
–– TDECQ, Outer OMA, Outer ER
TDR/TDT mode (requires TDR module)
Some also usable within measurement regions:
Prior to average add:
–– Time domain responses available in volts, percent reflection
and ohms (where applicable)
–– X-axis in time or distance
–– Average
–– Min
–– Max
–– Amplitude at time
–– Excess capacitance
–– Excess inductance
–– Rise time
–– Fall time
–– Delta time
Limit tests
–– Acquisition limits
–– Limit test “Run Until” conditions — Off, # of Waveforms, # of
Samples
–– Report action on completion — Save waveform to memory,
save screen image
–– Measurement limit test
–– Specify number of failures to stop limit test
–– When to fail selected measurement — Inside limits, outside
limits, always fail, never fail
–– Report action on failure — Save waveform to memory, save
screen image, save summary
–– Mask limit test
–– Specify number of failed mask test samples
–– Report action on failure — Save waveform to memory, save
screen image, save summary
Limit lines
–– With the limit line editor, you can modify an existing limit line
or create a new limit line from a current waveform or manual
data entry.
–– Limit lines are displayed boundaries used for pass/fail testing
in oscilloscope and TDR/TDT instrument modes, and they
are similar to masks in Eye/Mask mode.
–– Any data point that occurs above an upper limit line or below
a lower limit line causes a failed condition. A failed portion of
a waveform that is beyond the limit line boundary is displayed
in red.
Waveform autoscaling
Autoscaling provides quick horizontal and vertical scaling of both
pulse and eye-diagram (RZ and NRZ) waveforms.
Gated triggering
Trigger gating port allows easy external control of data
acquisition for circulating loop or burst-data experiments. Use
TTL-compatible signals to control when the instrument does and
does not acquire data.
27 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Measurements (Continued)
Configure measurements
–– Thresholds
–– 10%, 50%, 90% or 20%, 50%, 80% or custom
–– Eye boundaries
–– Define boundaries for eye measurements
–– Define boundaries for alignment
–– Format units for
–– Duty cycle distortion — Time or percentage
–– Extinction/Contrast ratio — Ratio, decibel or percentage
–– Eye height — Amplitude or decibel (dB)
–– Eye width — Time or ratio
–– Average power — Watts or decibels (dBm)
–– Top base definition
Automatic or custom
–– Δ time definition
–– First edge number, edge direction, threshold
–– Second edge number, edge direction, threshold
–– Jitter mode
–– Units (time or unit interval, watts, volts, or unit amplitude)
–– Signal type (data or clock)
–– Measure based on edges (all, rising only, falling only)
–– Graph layout (single, split, quad)
Quick measure configuration
When using the classic DCA interface, “Quick Measure” measurements are initiated by pressing the <Multi-Purpose> button on
the front panel.
–– Four user-selectable measurements for each mode, eyemask, TDR, etc.
–– Default settings (Eye/Mask mode) extinction ratio, jitter RMS,
average power, crossing percentage
–– Default settings (Oscilloscope mode) rise time, fall time,
period, V amptd
Histograms
Configure
–– Histogram scale (1 to 8 divisions)
–– Histogram axis (vertical or horizontal)
–– Histogram window (adjustable window via marker knob)
Math measurements - classic DCA user interface
–– Four user-definable functions operator — Magnify, invert,
subtract, versus, min, max
–– Source — Channel, function, memory, constant, response
(TDR)
Signal processing measurements - FlexDCA
–– Math — Add, subtract, multiply, average, invert, maximum,
minimum, median
–– Signal processing — Difference (differentiate), summation
(integrate), interpolation (linear, Sin(x)/x), filters: 4th order
bessel, butterworth, gaussian
–– Transforms — FFT, versus
–– Equalizer (Option 201) — Linear feed-forward
–– Equalizer (LFE, up to 64 taps)
–– Simulation (Option SIM) — De-embedding, embedding,
random jitter, random noise
Calibrate - classic DCA user interface
–– All calibrations
–– Module (amplitude)
–– Horizontal (time base)
–– Extinction ratio
–– Probe
–– Optical channel
–– Front panel calibration output level
–– User selectable –2 to 2 V
Utilities
–– Set time and date
–– Remote interface
–– Set GPIB interface
–– Touch screen configuration/calibration
–– Calibration
–– Disable/enable touch screen
–– Upgrade software
–– Upgrade mainframe
–– Upgrade module
28 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Measurements (Continued)
Easier calibrations
IVI-COM capability
Calibrating your instrument has been simplified by placing all
the performance level indicators and calibration procedures in a
single high-level location. This provides greater confidence in the
measurements made and saves time in maintaining equipment.
Interchangeable Virtual Instruments (IVI) is a group of new
instrument device software specifications created by the IVI
Foundation to simplify interchangeability, increase application
performance, and reduce the cost of test program development
and maintenance through design code reuse. The 86100D
IVI-COM drivers are available for download from the Keysight
website.
Stimulus response testing using the Keysight
Technologies, Inc. N490X, M8000 Series BERTs
Error performance analysis represents an essential part of digital
transmission test products. Together, the Keysight 86100D and
BERT products create a powerful test solution. If stimulus only
is needed, the 81133A and 81134A pattern generators work
seamlessly with the 86100D.
Transitioning from the Keysight 83480A and
86100A/B/C to the 86100D
While the 86100D has powerful new functionality that its
predecessors don’t have, it has been designed to maintain
compatibility with the Keysight 86100A, 86100B, 86100C and
Keysight 83480A digital communications analyzers and Keysight
54750A wide-bandwidth oscilloscope. All modules used in the
Keysight 86100A/B/C, 83480A and 54750A can also be used in
the 86100D. Since the 86100D includes the classic DCA interface,
the remote programming command set for the 86100D designed
for the 86100A/B/C will work directly. Some code modifications
are required when transitioning from the 83480A and 54750A,
but the command set is designed to minimize the level of effort
required.
VXII.2 and VXII.3 instrument control
The 86100D DCA-X provides LAN based instrument control.
29 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Ordering Information
86100D
Infiniium DCA-X mainframe
86100D
Hardware options
86100D-STR
Standard trigger
86100D-ETR
Enhanced trigger
86100DU-ETR Enhanced trigger upgrade kit
86100D-PTB
Precision timebase integrated in the mainframe
86100DU-PTB Upgrade mainframe with integrated precision timebase
86100D-GPI
GPIB card interface installed (default)
86100D-GPN No GPIB card interface
86100D-090
Removable hard drive
86100D-092
Internal hard drive (default)
86100D software options
86100D-xxx
Are permanent fixed-node licenses installed by the
factory
86100DU-xxx Are permanent fixed-node licenses installed by the
customer (software upgrades)
86100DT-xxx
Are permanent transportable licenses installed by the
customer
The “-xxx” option code provides the same functionality independent of the
prefix (i.e., 86100D or 86100DU or 86100DT)
86100D-BFP
Automatic fixture removal
86100D-061
MATLAB - Basic Oscilloscope package
86100D-062
MATLAB - Standard Oscilloscope package
86100D-200
Enhanced jitter analysis software
86100D-201
Advanced waveform analysis software
86100D-202
Enhanced impedance and S-parameter software
86100D-300
Amplitude analysis/RIN/Q-factor
86100DU-400 PLL and jitter spectrum software 1
86100D-401
Advanced EYE analysis software 1
86100D-SIM
InfiniiSim-DCA software
86100D-500
Productivity package software
86100D-9FP
PAM-N analysis software, fixed perpetual license
N1012A
OIF CEI compliance and debug application
N1014A
SFF-8431 compliance and debug application
N1019A
User-defined application
N1081A
IEEE 802.3 KR/KR4 application
N1082A
IEEE 802.3bj XLAUI, CAUI-10 and nPPI; IEEE 802.3bm
CAUI-4 (Option -4TP)
N1083A
IEEE 802.3 40GBASE-CR4 and 100GBASE-CR10
N1084A
IEEE802.3 KR4/CR4 application
N1085A
PAM4 Measurement application for Ethernet and OIF-CEI
IEEE Measurements option 1TP
OIF-CEI Measurements option 4TP
Misc options
86100D-AFP
Module slot filler panel
86100D-AX4
Rack mount flange kit
86100D-AXE
Rack mount flange kit with handles
86100D-UK6
Commercial cal certificate with test data
86100DU-WN7 Upgrade to Windows 7 operating system
Note:
–– Option BFP requires Option 202.
–– For complete PAM4 analysis capability, Option 9FP requires Options
200/201/300/401.
–– Options 200, 201, and SIM require Option ETR (enhanced trigger).
–– Option 300 requires Options 200 and ETR.
–– Option 400 requires Microsoft Office Excel 2007/2010.
Optical/electrical modules
86105C
9 GHz optical channel; single-mode and multimode,
amplified (750 to 1650 nm) 20 GHz electrical channel
86105C-100
155 Mb/s through 8.5 Gb/s (choose 4 filter rates from
Options 86105C-110 through 86105C-195)
86105C-110
155 Mb/s
86105C-120
622 Mb/s (also covers 614 Mb/s)
86105C-130
1.063 Gb/s
86105C-140
1.244/1.250 Gb/s (also covers 1.229 Mb/s)
86105C-150
2.125 Gb/s
86105C-160
2.488/2.500 Gb/s (also covers 2.458 Gb/s)
86105C-170
2.666 Gb/s
86105C-180
3.125 Gb/s (also covers 3.072 Gb/s)
86105C-190
4.250 Gb/s
86105C-193
5.0 Gb/s
86105C-195
6.250 Gb/s (also covers 6.144 Gb/s)
86105C-200
8.5, 9.953, 10.3125, 10.519, 10.664, 10.709, 11.096,
11.317 Gb/s
86105C-300
Combination of rates available in 86105C-100 and
86105C-200
86105D2
20 GHz optical channel; single-mode and multimode,
(750 to 1650 nm); filters for 8.5, 9.953, 10.3125, 10.519,
10.664, 10.709, 11.096, 11.317, 14.025 Gb/s; 35 GHz
electrical channel
86105D-100
Identical capability as 86105D, 14.025 Gb/s filter not
included
86105D-200
Identical capability as 86105D, only filter provided is
14.025 Gb/s
86105D-IRC 3 System impulse response correction calibration
86105D-281
34 GHz optical channel, filters for 15, 25.78 , 27.95,
28.05 Gb/s (contact Keysight for additional 14.025 Gb/s
filter) 4. 50 GHz electrical channel
1.
Fixed node customer installed or transportable customer installed
(86100DU or 86100DT). Not available as a factory-installed license.
2. This module is not compatible with the 86100A and 86100B DCA
mainframes. If you want to upgrade older DCAs, contact Keysight
Technologies to discuss current trade-in deals.
3. System impulse response correction calibration provides a unique
calibration file for the optical channel(s). The FlexDCA user interface uses
this calibration file to create ideal reference receiver responses for more
accurate and consistent transceiver compliance testing. IRC also allows
reference receivers to be defined at any data rate within ± 50% of the
hardware response. This allows optical receivers to significantly extend
their operating range.
4. A 15 Gb/s filter is provided with Option 281 and 282 that closely
approximates the response required for 16X Fibre Channel compliance
testing. Contact Keysight for a special option that verifies compliance
of the 86105D-281 and 86115D-282 to the 16X Fibre Channel reference
receiver specification.
30 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Ordering Information (Continued)
Optical/electrical modules (Continued)
86115D 2
20 GHz multi-optical port plug-in module; single-mode
and multimode (750 to 1650 nm); filters for 8.5, 9.953,
10.3125, 10.519, 10.664, 10.709, 11.096, 11.317, 14.025
Gb/s
86115D-002
Two optical channels with filters for all rates listed (8.5 to
14.025 Gb/s)
86115D-102
Identical capability as 86115D-002, 14.025 Gb/s filters
not included
86115D-142
Identical capability as 86115D-002, only filters provided
are 14.025 Gb/s
86115D-282
Two optical channels with filters for 15, 25.78, 27.95,
28.05 Gb/s (contact Keysight for 14.025 Gb/s filter) 4
86115D-004
Four optical ports with filters for all rates listed (8.5
to 14.025 Gb/s) mulitplexed to two optical channels
through 2 integral 1X2 optical switches
86115D-104
Identical capability as 86115D-004, 14.025 Gb/s filters
not included
86115D-144
Identical capability as 86115D-004, only filters provided
are 14.025 Gb/s
86115D-IRC
System impulse response correction calibration
86116C 1, 2
40 to 65 GHz optical/80 GHz electrical sampling module,
1300 to 1620 nm
86116C-IRC 3
System impulse response correction calibration
This module is not compatible with the 86100A and 86100B DCA
mainframes. If you want to upgrade older DCAs, contact Keysight
Technologies to discuss current trade-in deals. All optical modules
have FC/PC connectors installed on each optical port. Other connector
adapters available as options are: Diamond HMS-10, DIN, ST and SC.
Select exactly one reference receiver option:
86116C-025
40 GHz optical/80 GHz electrical channels,
17.0/25.8/27.7 Gb/s reference receiver
86116C-041
65 GHz optical/80 GHz electrical channels,
39.81/41.25/43.02 Gb/s reference receiver
Dual/Quad electrical channel modules
86112A
Dual 20 GHz electrical channels
86112A-HBW Dual 30 GHz electrical channels
86118A
Dual 70 GHz electrical remote sampling channels
86118A-H01
Differential de-skew
N1045A 2
2/4 port 60 GHz electrical remote head
N1045A-02F
2 channel remote head, 1.85 mm, female
N1045A-02M 2 channel remote head, 1.85 mm, male
N1045A-04F
4 channel remote head, 1.85 mm, female
N1045A-04M 4 channel remote head, 1.85 mm, male
N1046A
100 GHz, 1/2/4 port electrical remote sampling head
N1046A-71F
1 channel, 75 GHz remote head, 1 mm, female
N1046A-81F
1 channel, 85 GHz remote head, 1 mm, female
N1046A-11F
1 channel, 100 GHz remote head, 1 mm, female
N1046A-72F
2 channel, 75 GHz remote head, 1 mm, female
N1046A-82F
2 channel, 85 GHz remote head, 1 mm, female
N1046A-12F
2 channel, 100 GHz remote head, 1 mm, female
N1046A-74F
4 channel, 75 GHz remote head, 1 mm, female
N1046A-84F
4 channel, 85 GHz remote head, 1 mm, female
N1046A-14F
4 channel, 100 GHz remote head, 1 mm, female
TDR/TDT modules
54754A 1, 5
Differential TDR module with dual 18 GHz TDR/electrical
channels
N1055A 1, 2
35/50 GHz, 2/4 port, TDR/TDT remote head
N1055A-FS1
Fast sampling, mandatory option
N1055A-32F
35 GHz, 2 channel remote head, 2.92 mm, female
N1055A-32M
35 GHz, 2 channel remote head, 2.92 mm, male
N1055A-34F
35 GHz, 4 channel remote head, 2.92 mm, female
N1055A-34M 35 GHz, 4 channel remote head, 2.92 mm, male
N1055A-52F
50 GHz, 2 channel remote head, 1.85 mm, female
N1055A-52M
50 GHz, 2 channel remote head, 1.85 mm, male
N1055A-54F
50 GHz, 4 channel remote head, 1.85 mm, female
N1055A-54M 50 GHz, 4 channel remote head, 1.85 mm, male
N1055A-FS1
Fast sampling
1.
2.
3.
4.
5.
86100D-ETR is recommended when;
a. More than 1 TDR module is connected to the same DUT.
b. Using a DCA module equipped with a rear-panel trigger circuit.
Examples include 54754A, 83496x, and 86108A/B modules. If
operating these modules in an 86100D with Option STR, an external
cable (such as P/N 5062-6690) must be connected from the module’s
front panel trigger/clock output to the 86100D’s trigger input.
Requires 86100D mainframe (not compatible with 86100A/B/C).
This module is not compatible with the 86100A and 86100B DCA
mainframes. If you want to upgrade older DCAs, contact Keysight
Technologies to discuss current trade-in deals.
Supports both DCA-X and DCA-M platforms. For detailed
N1076A/N1077A specifications, refer to the data sheet located at
www.keysight.com/find/N1077A. Standalone clock recovery instruments
are controlled via a USB connection to an 86100D DCA-X mainframe or a
user-supplied PC running N1010A FlexDCA software.
Included with each of these TDR modules is a TDR trigger cable, two 50 Ω
SMA terminations and one SMA short.
31 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Ordering Information (Continued)
External clock recovery solutions 4
N1076A electrical clock recovery
N1076A-216
Clock recovery range: 50 MBd to 16 GBd
N1076A-232
Clock recovery range: 50 MBd to 32 GBd
N1076A-JSA
Jitter spectrum analysis
N1076B electrical clock recovery
N1076B-216
Clock recovery range: 125 MBd to 16 GBd
N1076B-232
Clock recovery range: 125 MBd to 32 GBd
N1076B-264
Clock recovery range: 125 MBd to 64 GBd (56 GBd for
PAM4 signals)
N1076B-EVA
Integrated variable equalizers (mandatory option)
N1076B-JSA
Jitter Spectrum Analysis
N1077A optical/electrical clock recovery
N1077A-216
Clock recovery range: 50 MBd to 16 GBd
N1077A-232
Clock recovery range: 50 MBd to 32 GBd
N1077A-SMS
Internal SM and MM splitters
N1077A-SXT
No splitter (supplied by user)
N1077A-JSA
Jitter spectrum analysis
N1078A optical/electrical clock recovery
N1078A-216
Clock recovery range: 125 MBd to 16 GBd
N1078A-232
Clock recovery range: 125 MBd to 32 GBd
N1078A-264
Clock recovery range: 125 MBd to 64 GBd
N1078A-S50
Internal 50-50 SM optical splitter
N1078A-SXT
No splitter (splitter supplied by user)
N1078A-JSA
Jitter spectrum analysis
N1078A-EVA
Integrated variable equalizers in electrical input path
(mandatory option)
Precision waveform analyzer module 1, 3
Dual electrical channel module with integrated clock recovery and
precision timebase.
86108A-100
Dual 32 GHz electrical channels, integrated clock
recovery (50 Mb/s to 14.2 Gb/s) with integrated
precision timebase
86108A-001
Two 3.5 mm phase trimmers for skew adjustment
86108A-002
Two precision 3.5 mm 18 inch cables
86108A-400
Auxiliary clock recovery Input
86108B options
86108B-LBW
86108B-HBW
86108B-216
86108B-232
86108B-300
86108B-400
86108B-PTB
86108B-JSA
86108B-A23
86108B-CA2
86108B-CA3
86108B-DC2
86108B-DC3
86108B-PT2
86108B-PT3
1.
2.
3.
4.
5.
Dual 35 GHz electrical channels
Dual 50 GHz electrical channels
Clock recovery 50 Mb/s to 16 Gb/s
Clock recovery 50 Mb/s to 32 Gb/s
Adjustable loop bandwidth/peaking
Auxiliary clock recovery input
Integrated precision timebase
Jitter spectrum analysis and software clock recovery
emulation
Two adapters, 2.4 mm (f) to 3.5 mm (f)
Matched cable pair, 2.4 to 2.4 mm, 24 inch
Matched cable pair, 3.5 to 3.5 mm, 18 inch
Two DC blocks, 2.4 mm, 16 V, 50 kHz to 50 GHz
Two DC blocks, 3.5 mm, 16 V, 50 kHz to 26.5 GHz
Two 2.4 mm phase trimmers for ext. skew adjustment
Phase trimmers
Nominal impedance
Frequency range
Insertion loss (at f max)
Time delay
Connectors
Two 3.5 mm phase trimmers for ext. skew adjustment
Phase trimmers
Nominal impedance
Frequency range
Insertion loss (at f max)
Time delay
Connectors
86100D-ETR is recommended when;
a. More than 1 TDR module is connected to the same DUT.
b. Using a DCA module equipped with a rear-panel trigger circuit.
Examples include 54754A, 83496x, and 86108A/B modules. If
operating these modules in an 86100D with Option STR, an external
cable (such as P/N 5062-6690) must be connected from the module’s
front panel trigger/clock output to the 86100D’s trigger input.
Requires 86100D mainframe (not compatible with 86100A/B/C).
This module is not compatible with the 86100A and 86100B DCA
mainframes. If you want to upgrade older DCAs, contact Keysight
Technologies to discuss current trade-in deals.
Supports both DCA-X and DCA-M platforms. For detailed
N1076A/N1077A specifications, refer to the data sheet located at
www.keysight.com/find/N1077A. Standalone clock recovery instruments
are controlled via a USB connection to an 86100D DCA-X mainframe or a
user-supplied PC running N1010A FlexDCA software.
Included with each of these TDR modules is a TDR trigger cable, two 50 Ω
SMA terminations and one SMA short.
32 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Ordering Information (Continued)
Warranty options (for all products)
R1280A
Customer return repair service
R1282A
Customer return calibration service
Accessories
See the "DCA Accessories" Guide for available accessories
Connectivity solutions
For a wide range of test adapters to connect to one or more lanes for
SFP+, QSFP+, fibre channel, PCIe and many others, please see adapters
information from Wilder Technologies at: http://www.wilder-tech.com/
Call Keysight for connectivity and probing solutions not listed above.
Firmware and software
Firmware and software upgrades are available through the Web or your
local sales office. www.keysight.com/find/dcax
N1010A FlexDCA Remote Access Software FlexDCA Pro runs on the
86100D as well as on a PC or laptop. Software licenses allow connectivity to an 86100C or D mainframe, advanced analysis or productivity
packages, and more. http://www.keysight.com/find/flexdca_download
FlexDCA express is the same as FlexDCA Pro except that it does not
included any functionality requiring a license. It is available free of charge
from http://www.keysight.com/find/flexdca_express
86100D DCA-X brochure
For more information on the features and benefits of the 86100D DCA-X
and DCA modules, download the 86100D DCA-X brochure. Go to:
www.keysight.com and search on 5989-5822EN.
33 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Keysight Oscilloscopes
Multiple form factors from 20 MHz to > 90 GHz | Industry leading specs | Powerful applications
34 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. The business that became Keysight was a founding
member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo
are registered US trademarks of the PCI Industrial Computer Manufacturers
Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web
inside your test systems. The business that became Keysight was a founding
member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Download your next insight
Keysight software is downloadable
expertise. From first simulation through
first customer shipment, we deliver the
tools your team needs to accelerate from
data to information to actionable insight.
–– Electronic design automation (EDA)
software
–– Application software
–– Programming environments
–– Productivity software
Learn more at
www.keysight.com/find/software
Start with a 30-day free trial.
www.keysight.com/find/free_trials
35 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
http://www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and
find warranty information.
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—onestop calibration, repair, asset management, technology refresh,
consulting, training and more—helps you improve product quality
and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure
your instruments are operating to specification, so you can rely on
accurate measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and
product breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
Brazil
Mexico
United States
(877) 894 4414
55 11 3351 7010
001 800 254 2440
(800) 829 4444
Asia Pacific
Australia
China
Hong Kong
India
Japan
Korea
Malaysia
Singapore
Taiwan
Other AP Countries
1 800 629 485
800 810 0189
800 938 693
1 800 11 2626
0120 (421) 345
080 769 0800
1 800 888 848
1 800 375 8100
0800 047 866
(65) 6375 8100
Europe & Middle East
Austria
Belgium
Finland
France
Germany
Ireland
Israel
Italy
Luxembourg
Netherlands
Russia
Spain
Sweden
Switzerland
United Kingdom
0800 001122
0800 58580
0800 523252
0805 980333
0800 6270999
1800 832700
1 809 343051
800 599100
+32 800 58580
0800 0233200
8800 5009286
800 000154
0200 882255
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
www.keysight.com/find/86100D
This information is subject to change without notice.
© Keysight Technologies, 2018
Published in USA, February 28, 2018
5990-5824EN
www.keysight.com
Similar pages