SL Power ME10A0603N01 10w-12w single output external power medical sery Datasheet

ME10
10W-12W Single Output External Power
Medical Series
Features


Meets UL/EN/IEC60601-1-2, 4th edition for
EMC*
Approved to EN/IEC/UL60601-1, 3rd edition
applications with and isolation levels which
satisfy the 2 MOPP requirements

Meets DoE Efficiency Level VI Requirements




Up to 12W of AC-DC Power
Universal Input 90-264Vac Input Range






LPS
No load input power
Average Efficiency
Desktop and Wall-Plug versions
Meets EN55011/CISPR11, FCC Part 15.109
Class B Conducted & Radiated Emissions, with
6db margin
E-cap life of >7 years
<1,000,000 Hours MTBF
3 Year Warranty
IP22 Rated Enclosure
Description
A high performance AC to DC external power supply family designed for medical applications. The ME10 Medical Series
external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition, with isolation levels which satisfy the 2
*
MOPP requirements; and designed to UL/EN/IEC60601-1-2, 4th edition for EMC . The ME10 Series models will operate at
universal input range of 90 to 264Vac over the wide temperature range of -20°C to +50°C, delivering full rated output power up
to +40°C and applicable output power derating at 50°C. These models are available in desktop and wall-plug versions, include
an IP22 rating per IEC60529 for the enclosure, and output cable terminated at a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Model
Number
ME10A0503F01
ME10A0603F01
ME10A0703F01
ME10A1203F01
ME10A2403F01
ME10A0503N01
ME10A0603N01
ME10A0703N01
ME10A1203N01
ME10A2403N01
ME10A0503Q01
ME10A0603Q01
ME10A0703Q01
ME10A1203Q01
ME10A2403Q01
ME10A0503B01
ME10A0603B01
ME10A0703B01
ME10A1203B01
ME10A2403B01
www.slpower.com
Volts
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
5.0V
5.9V
7.5V
12.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
2.0A
1.6A
1.3A
1.0A
0.5A
Output
Power
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
10W
10W
10W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
ME10 Rev 2.5
Load
Regulation
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
31-JAN-16
Output
Connector
Input
Configuration
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Interchangeable
Blades (North American
Blade included)²
1
ME10
10W-12W Single Output External Power
Medical Series
Model Selection (continued)
Model
Number
ME10A0503C01
ME10A0603C01
ME10A0703C01
ME10A1203C01
ME10A2403C01
Notes:
Volts
5.0V
5.9V
7.5V
12.0V
24.0V
Output
Current
2.0A
1.6A
1.3A
1.0A
0.5A
Output
Power
10W
10W
10W
12W
12W
Ripple &
Noise¹
75mV pk-pk
75mV pk-pk
75mV pk-pk
120mV pk-pk
240mV pk-pk
Line
Regulation
±1%
±1%
±1%
±1%
±1%
Load
Regulation
±5%
±5%
±5%
±5%
±5%
Output
Connector
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Input
Configuration
Class II Wall-Plug,
Fixed North American
Blades³
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors. For 5V and
6V models, values listed are typical, 100mV pk-pk maximum with 0.1μF ceramic and 47μF low ESR capacitors used at measurement point.
2. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (ME10B0503F01).
5. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
General Specifications
AC Input
Input Current
Inrush Current
Input Fuses
Earth Leakage
Current
Efficiency
100-240Vac, ±10%, 47-63Hz, 1∅
Turn On Time
Less than 700mS @115Vac, full load
115Vac: 0.45A, 230Vac: 0.28A
Hold-up Time
20mS min., at full Load, 100Vac input
264Vac, cold start: will not exceed 40A
F1, F2: 3.15A, 250Vac fuses (line & neutral
lines) provided on all models
Input-GND: <500μA@264Vac, 60Hz, NC
Output-GND: <4mA@264Vac, 60Hz, NC
No Load Input
Power
<0.1W per DoE Efficiency Level VI
Requirements
Transient
Response
Regulation
Drop Test
Vibration
E-Cap Life
www.slpower.com
Short Circuit
Protection
Overvoltage
Protection
10 to 12W continuous – See models chart for
specific voltage model ratings.
Output Voltage
Overload Protection
>87%, typical
Output Power
Ripple and Noise
Overtemperature
Protection
See models chart on pg 1.
See models chart on pg 1.
500µs response time for return to within 0.5%
of final value for any 50% load step over the
range of 5% to 100% of rated load, ∆i/∆t<
0.2A/µs. Max. voltage deviation is +/-3.5%.
Isolation
Safety Standards
Hiccup Mode, auto recovery.
130 to 150% of output voltage, hiccup mode
Input-Output: 2 MOPP
Input-Ground: 1 MOPP
Output-Ground: 1500Vac
EN/IEC/UL60601-1, 3rd edition and
EN/IEC/UL60601-1-11:2010 for Home
Healthcare
-20°C to +50°C
Start Up at -40°C, full load, (warmup period before
Temperature
Derating
Derate output power above 40°C to TBD at
50°C
Storage Temperature
Altitude
Relative Humidity
Operating: 0.003g/Hz, 1.5grms overall, 3
axes, 10 min/axis, 1-500Hz.
Non-Oper.: random waveform, 3 minutes per
axis, 3 axes and Sine waveform, Vib.
frequency/acceleration: 10-500Hz/1g, sweep
rate of 1 octave / minutes, Vibration time of 10
sweeps / axes, 3 axes
>10 year life based on calculations at
115Vac/60Hz & 230Vac/50Hz, ambient 25°C
at 24 hrs per day, 365 days/year, 6 power up
cycles per day.
ME10 Rev 2.5
130 to 180% of rating, Hiccup Mode
Operating
Temperature
See models chart on pg 1.
1.4m from table top to wooden platform, 6
faces.
Will shutdown upon an overtemperature
condition, auto-recovery.
31-JAN-16
all parameters are within published specifications).
-40°C to +85°C
Operating: to 5000m.
Non-operating: -500 to 40,000 ft.
5% to 95%, non-condensing
Shock
Operating: Half-sine, 20gpk, 10mS, 3 axes,
6 shocks total
Non-Operating: Half-sine waveform, impact
acceleration of 100G, Pulse duration of 6
mS, Number of shocks: 3 for each of the
three axis
MTBF
>1,000,000 hours, full load, 110 & 220Vac
input, 25°C amb., per Telcordia 332 Issue 6,
Stress Method.
2
ME10
10W-12W Single Output External Power
Medical Series
EMI/EMC Compliance
Conducted Emissions:
EN55011/CISPR11 Class B, FCC Part 15.107, Class B: 6db margin typ, at 115 and 230Vac
Radiated Emissions:
EN55011/CISPR11 Class B, FCC Part 15.109, Class B: 3db margin typ, at 115 and 230Vac
Common Mode Noise:
High Frequency (100kHz-20MHz): <40mA pk-pk
Electro-Static Discharge (ESD)
Immunity on Power ports:
EN55024/IEC61000-4-2, Level 4: +/- 8kV contact, +/- 15kV air, Criteria A
IEC60601-1-2, 4th Edition, Table 4
Radiated RF EM Fields Susceptibility
EN55022/EN61000-4-3, 10V/m, 80MHz-2.7GHz, 80% AM at 1kHz
IEC60601-1-2, 4th Edition, Table 4
Electrical Fast Transients (EFT) /Bursts:
EN55024/IEC61000-4-4, Level 4, +/- 4.4kV, 100Khz rep rate, 40A, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Surges, Line to Line (Diff Mode) and Line
to GND (CMN Mode)
Conducted Disturbances induced by RF
Fields
Rated Power frequency magnetic fields
Voltage Interruptions, Dips, Sags &
Surges
EN55024/IEC61000-4-5, Level 4, +/-2kV DM, +/-4kV CM, Criteria A
Surpasses IEC60601-1-2, 4th Edition requirements.
EN55022/IEC61000-4-6, 3.6V/m – Level 4, 0.15 to 80Mhz; and 12V/m) in ISM and amateur
radio bands between 0.15Mhz and 80Mhz, 80% AM at 1KHz
IEC60601-1-2, 4th Edition, Table 5
EN55024/IEC1000-4-8, Level 4: 30 A/m, 50/60 Hz
IEC60601-1-2, 4th Edition, Table 4
EN55024/IECEN61000-4-11: --100% dip for 20mS, Criteria A
--100% dip for 5000mS (250/300 cycles), Criteria B
--60% dip for 100mS, Criteria B
--30% dip for 500mS, Criteria A
IEC60601-1-2, 4th Edition, Table 5
Harmonic Current Emissions
EN55011/EN61000-3-2, Class A
Flicker Test
EN61000-3-3
All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
Mechanical Drawing
IEC60320 C14 Receptacle, 2.5 x 5.5 x 9.5mm Barrel Connector
Notes:
www.slpower.com
Interchangeable N.A. Blade, 2.5 x 5.5 x 9.5mm barrel connector
1. All dimensions in mm.
2. Weight = 110g
2. Interchangeable blade models come with North American blade fitted. For other blades (EU, UK, Aust.) order blade kit KT1027K.
3. The unit should not be covered or enclosed to protect against excessive case temperature rise.
ME10 Rev 2.5
31-JAN-16
3
ME10
10W-12W Single Output External Power
Medical Series
Connector Information
Standard models include a 2.5 x 5.5 x 9.5mm straight barrel type connector (Ault #3), center positive. Other standard options are listed below.
The “03” in the standard model number is replaced by the applicable digits below:
Data Sheet © 2016 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for
consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
www.slpower.com
ME10 Rev 2.5
31-JAN-16