EXTECH 412355A

User's Guide
Model 412355A Current / Voltage Calibrator
Introduction
Congratulations on your purchase of the Extech Current/Voltage Calibrator. The Model
412355A can measure and source current and voltage. The Oyster Series meters have a
convenient flip up display with neck-strap for hands-free operation. This meter is shipped
fully tested and calibrated and, with proper use, will provide years of reliable service.
Meter Description
1.
LCD display
2.
POWER button
3.
UP and DOWN buttons
4.
MODE button
5.
UNIT button
6.
MEM and ZERO button
7.
Input/Output mini-connector
8.
AC adaptor jack
9.
Battery compartment
10. Neck strap holders
11. Function switch
2
412355A V3.2 7/08
Operation
POWER BUTTON and AUTO POWER OFF FEATURE
1. Use the POWER button to turn the unit ON or OFF. When the unit is powered up, a
short self-test will ensue after which the display will stabilize.
2. This meter can either be powered by one 9V battery or an AC adapter.
3. When the battery symbol appears on the LCD display, replace the battery as soon as
possible. Low battery power may cause inaccurate readings and erratic meter
operation.
4. This instrument is equipped with Auto Power OFF which turns the meter off after 10
minutes of inactivity. To override this feature; press and hold the MODE button until the
display shows
activated).
(auto power off de-activated) or
(auto power off
FUNCTION SLIDE SWITCH
Slide the Function switch on the side of the meter to the desired function (Voltage or
Current)
MODE
Press the MODE button to select either SOURCE (output) or MEASURE (input)
UNIT BUTTON
Press the UNIT button to select:
1.
Voltage mode: mV or V in source.
2.
Current mode: mA or % in source or measure
▲▼ OUTPUT ADJUST BUTTONS
Press the ▲ ▼buttons to increase or decrease the output voltage or current value in the
source mode.
1.
Press the ▲ button once to increase the value in one digit steps.
2.
Press and hold the ▲ button to increase the value in 10 digit steps.
3.
Press and hold the ▲ button > 2 sec and then press the ▼ to increase the value
in 100 digit steps
4.
To decrease the value, use the ▼ button as described above.
ZERO BUTTON
The ZERO button manually zeros the display in the MEASURE mode.
1.
Set the meter to the MEASURE mode
2.
Short the input jack
3.
Press and release the ZERO button.
3
412355A V3.2 7/08
MEM BUTTON
The memory feature provides 5 user settable source values for stepped calibration
outputs. The feature is available for Voltage, mA and % in the SOURCE mode. The
memory values are stored in nonvolatile memory and are not erased when power is turned
off.
Sourcing from the stored memory values:
1.
Select the Source mode
2.
Press the MEM button. The M1 icon (memory location 1) will appear in the display
and the value stored in that location will be displayed and sourced.
3.
Repeated pressing of the MEM button will step through the 5 memory locations.
NOTE: The “SOURCE” icon will blink when the output value has not reached a
stable level. The common cause for the “SOURCE” icon to remain blinking is that the
load impedance is too high in the current mode or too low in the voltage mode.
Storing values into memory:
1.
Select and display a memory location (M1 to M5)
2.
Press the ▲ ▼buttons to adjust the display to the desired source value.
3.
Press and hold the MEM button for > 2 seconds. The displayed value will be stored
into the displayed memory location.
Default memory values.
Five common source values are permanently programmed in memory as the default
values. These values can be replaced by user selected values.
To reset the meter to the default memory values:
1.
Turn the meter on and select the SOURCE mode.
2.
Press and hold the POWER button for >4 seconds.
will briefly appear in the
display and the default values will be stored into memory.
Default Memory Values
M1
M2
M3
M4
M5
mV
0mV
500mV
1000mV
1500mV
2000mV
V
2V
5V
10V
15V
20V
mA
4mA
8mA
12mA
16mA
20mA
%
0%
25%
50%
75%
100%
Over-range / Under-range Indication
Signals above or below the units ranges will be indicated by “HHHH” for above and
“LLLL” for below.
4
412355A V3.2 7/08
Measure and Source
MEASURE (Input)
In this mode, the unit will measure voltage or current.
1.
Select Voltage or Current on the Function switch.
2.
Turn the meter ON
3.
Press the MODE button to select MEASURE
4.
Press the UNIT button to select mA or % if current is selected.
5.
Connect the Calibration Cable to the meter.
6.
Connect the Calibration Cable to the device or circuit under test.
7.
Read the measurement on the LCD display.
SOURCE (Output)
In this mode, the unit will source (output) voltage or current.
1.
Select Voltage or Current on the Function switch.
2.
Turn the meter ON
3.
Press the MODE button to select SOURCE
4.
Press the UNIT button to select V or mV if voltage is selected or to select mA or %
if current is selected.
5.
Connect the Calibration Cable to the meter.
6.
Connect the Calibration Cable to the device or circuit under test.
7.
Use the ▲ ▼buttons to select the desired output value. Use the LCD display to
verify the output level. Alternatively, use the stored calibration values in memory as
described in the MEM button section.
8.
For the -25% to 125% output range the output is 0 to 24mA.
% Display
-25%
0%
25%
50%
75%
100%
125%
mA output
0mA
4mA
8mA
12mA
16mA
20mA
24mA
NOTE: The “SOURCE” icon will blink when the output value has not reached a
stable level. The common cause for the “SOURCE” icon to remain blinking is that the
load impedance is too high in the current mode or too low in the voltage mode.
5
412355A V3.2 7/08
Battery Replacement
When the LOW BAT message appears on the LCD, replace the 9V battery as soon as
possible.
1. Open the calibrator’s lid as far as possible.
2. Open the battery compartment using a coin at the arrow indicator.
3. Replace the battery and close the cover.
Specifications
General Specifications
Display
9999 count LCD
Meter Power
9 volt battery or 9V AC adaptor
Auto Power OFF
Meter automatically powers off after 10 minutes of
inactivity
Current output capability
24mA at 1000 ohms
Operating Temperature
41ºF to 104ºF (5ºC to 40ºC)
Storage Temperature
- 4 F to 140 F (-20 C to 60 C)
Operating Humidity
Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at
104ºF (40ºC)
Storage Humidity
<80%
o
o
o
o
Operating Altitude
7000ft. (2000meters) maximum
Dimensions
3.8 x 4.7 x 1.8” (96 x 118 x 45mm) folded
Weight
12 oz. (340g)
Accessories Supplied
9V battery, AC adaptor and calibration cable with spade
lugs
Range Specifications
Mode
Function
Range
Measure
Voltage
0 to 2000mV
1mV
2.00 to 20.00V
0.01V
Source
Resolution
Current
0.00 to 50.00mA
0.01mA
%
-25.0 to 230.0%
0.1%
Voltage
0 to 2000mV
1mV
2.00 to 20.00V
0.01V
Current
0.00 to 24.00mA
0.01mA
%
-25.0 to 125.0%
0.1%
6
Accuracy (% of reading)
± (0.075% + 1 digit) or ± 3
digits, whichever is greater
412355A V3.2 7/08
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set
forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also provides
NIST certification for most products. Call the Customer Service Department for information
on calibration services available for this product. Extech recommends that annual
calibrations be performed to verify meter performance and accuracy.
Support line (781) 890-7440
Technical Support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Copyright © 2007 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.
7
412355A V3.2 7/08
8
412355A V3.2 7/08