MA-COM 3MV1

Catalog: 1654001
Issue Date: 06.2011
Corcom Product Guide
Multi-purpose Medical Filter for Power Line Noise Protection
MV Series
UL Recognized
CSA Certified
VDE Approved
MV Series
Specifications
• Multi-purpose medical filter
• Improved Line to Ground performance
• A good solution to emission or immunity problems
• Meets leakage current requirements of UL2601 for
health care equipment
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
.07 mA
@250 VAC 50 Hz:
.13 mA
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
Rated Voltage (max):
250 VAC
Operating Frequency:
50/60 Hz
Rated Current:
3 to 20A
Operating Ambient Temperature Range
(at rated current Ir):
-10°C to +40°C
In an ambient temperature (Ta) higher than +40°C
the maximum operating current (Io) is calculated as
follows: Io = Ir (85-Ta)/45
Ordering Information
Electrical Schematic
6 MV 1
Input / Output Style
1 – .250 [6.3] spade terminals
LINE
LOAD
MV Series
Current Rating
3, 6, 10 or 20A
Available Part Numbers
3MV1
10MV1
6MV1
20MV1
54
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com
Catalog: 1654001
Issue Date: 06.2011
Corcom Product Guide
Multi-purpose Medical Filter for Power Line Noise Protection (continued)
MV Series
Case Styles
MV1 (3, 6, 10A)
Performance Data
Typical Insertion Loss
Measured in closed 50 Ohm system
B
C




6MV
db
100
db
100
90
90
80
80
70
70
60
60
50
50
40
40
30
30
20
20
10
0



20MV1
10
0.1
1
0
10
30
Frequency in MHz
20MV
db
100
db
100
90
90
80
80
70
70
60
60
50
50
40
40
30
30
20
20
10
E
0.1
10MV
0
1
RFI Power Line Filters
D
E
A
3MV
1
10
30
Frequency in MHz
1
10
30
Frequency in MHz
10
0.1
1
0
10
30
Frequency in MHz
0.1
Common Mode / Asymmetrical (L-G)
Differential Mode / Symmetrical (L-L)
D B
Minimum Insertion Loss
Measured in closed 50 Ohm system
2.00
Common Mode / Asymmetrical (Line to Ground)
50.8




A
C



Case Dimensions
Part No.
3MV1
6MV1
10MV1
20MV1
A
B
C



























D










E
























































Differential Mode / Symmetrical (Line to Line)















































55
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
For email, phone or live chat, please go to
te.com/help
corcom.com