NI 9244 User Manual and Specifications

USER MANUAL AND SPECIFICATIONS
NI 9244
4-Channel, 400 Vrms L-N, 800 Vrms L-L,
24-Bit Simultaneous Analog Input Module
Français
Deutsch
ni.com/manuals
This document describes how to use the National Instruments 9244
and includes specifications and pin assignments for the NI 9244.
Note The safety guidelines and specifications in this
document are specific to the NI 9244. The other
components in the system might not meet the same safety
ratings and specifications. Refer to the documentation for
each component in the system to determine the safety
ratings and specifications for the entire system.
Related Information
2 |
NI CompactDAQ &
NI CompactRIO Documentation
ni.com/info cseriesdoc
Chassis Compatibility
ni.com/info compatibility
Software Support
ni.com/info softwareversion
Services
ni.com/services
ni.com
|
NI 9244 User Manual and Specifications
Safety Guidelines
Operate the NI 9244 only as described in this user manual.
Hot Surface This icon denotes that the component may
be hot. Touching this component may result in bodily
injury.
Warning This icon denotes a warning advising you to
take precautions to avoid electrical shock.
Do not operate the NI 9244 in a manner not
specified in this user manual. Product misuse can result
in a hazard. You can compromise the safety protection
built into the product if the product is damaged in any
way. If the product is damaged, return it to National
Instruments for repair.
Caution
NI 9244 User Manual and Specifications |
© National Instruments
|
3
Safety Guidelines for Hazardous Voltages
If hazardous voltages are connected to the module, take the
following precautions. A hazardous voltage is a voltage greater
than 42.4 Vpk or 60 VDC to earth ground.
Ensure that hazardous voltage wiring is
performed only by qualified personnel adhering to local
electrical standards.
Caution
Do not mix hazardous voltage circuits and
human-accessible circuits on the same module.
Caution
Make sure that devices and circuits connected
to the module are properly insulated from human contact.
Caution
When module terminals are hazardous voltage
LIVE (>42.4Vpk/60 VDC), you must ensure that devices
and circuits connected to the module are properly
insulated from human contact. You must use the NI 9969
connector backshell kit to ensure that the terminals are
not accessible.
Caution
4 |
ni.com
|
NI 9244 User Manual and Specifications
Figure 1 shows the NI 9969 connector backshell.
Figure 1. NI 9969 Connector Backshell
NI 9244 User Manual and Specifications |
© National Instruments
|
5
Safety Guidelines for Hazardous Locations
The NI 9244 is suitable for use in Class I, Division 2, Groups A, B,
C, D, T4 hazardous locations; Class I, Zone 2, AEx nA IIC T4, and
Ex nA IIC T4 hazardous locations; and nonhazardous locations
only. Follow these guidelines if you are installing the NI 9244 in a
potentially explosive environment. Not following these guidelines
may result in serious injury or death.
Do not disconnect I/O-side wires or connectors
unless power has been switched off or the area is known
to be nonhazardous.
Caution
Do not remove modules unless power has been
switched off or the area is known to be nonhazardous.
Caution
Substitution of components may impair
suitability for Class I, Division 2.
Caution
For Division 2 and Zone 2 applications, install
the system in an enclosure rated to at least IP 54 as
defined by IEC/EN 60529.
Caution
6 |
ni.com
|
NI 9244 User Manual and Specifications
Special Conditions for Hazardous Locations Use in
Europe
This equipment has been evaluated as Ex nA IIC T4 Gc equipment
under DEMKO 12 ATEX 1202658X. Each module is marked
II 3G and is suitable for use in Zone 2 hazardous locations, in
ambient temperatures of -40 °C ≤ Ta ≤ 70 °C. If you are using the
NI 9244 in Gas Group IIC hazardous locations, you must use the
device in an NI chassis that has been evaluated as Ex nC IIC T4,
EEx nC IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment.
Caution The system shall be mounted in an ATEX
certified enclosure with a minimum ingress protection rating
of at least IP54 as defined in IEC/EN 60529 and used in an
environment of not more than Pollution Degree 2.
The enclosure must have a door or cover
accessible only by the use of a tool.
Caution
NI 9244 User Manual and Specifications |
© National Instruments
|
7
Electromagnetic Compatibility Guidelines
This product was tested and complies with the regulatory
requirements and limits for electromagnetic compatibility (EMC)
as stated in the product specifications. These requirements and
limits are designed to provide reasonable protection against
harmful interference when the product is operated in its intended
operational electromagnetic environment.
This product is intended for use in industrial locations. There is no
guarantee that harmful interference will not occur in a particular
installation, when the product is connected to a test object, or if the
product is used in residential areas. To minimize the potential for
the product to cause interference to radio and television reception
or to experience unacceptable performance degradation, install and
use this product in strict accordance with the instructions in the
product documentation.
Furthermore, any changes or modifications to the product not
expressly approved by National Instruments could void your
authority to operate it under your local regulatory rules.
8 |
ni.com
|
NI 9244 User Manual and Specifications
Special Guidelines for Marine Applications
Some products are Lloyd’s Register (LR) Type Approved for
marine (shipboard) applications. To verify Lloyd’s Register
certification for a product, visit ni.com/certification and
search for the LR certificate, or look for the Lloyd’s Register mark
on the product label.
Caution In order to meet the EMC requirements for
marine applications, install the product in a shielded
enclosure with shielded and/or filtered power and
input/output ports. In addition, take precautions when
designing, selecting, and installing measurement probes
and cables to ensure that the desired EMC performance is
attained.
NI 9244 User Manual and Specifications |
© National Instruments
|
9
Connecting the NI 9244
The NI 9244 provides connections for four analog input channels.
Figure 2. NI 9244 Pinout
10
|
ni.com |
AI0
0
AI1
1
AI2
2
Neutral
3
NI 9244 User Manual and Specifications
Connector
The NI 9244 has a 4-terminal, detachable screw-terminal
connector.
Backshell
Caution For safe operation with hazardous voltages,
you must use the NI 9969 Connector Backshell, as shown
in Figure 1, with the 4-terminal connector on the NI 9244.
You can use the backshell with 12 AWG to 24 AWG wires with the
NI 9969. 12 AWG to 14 AWG, 16 AWG, and 18 AWG to 24 AWG
require different strain-relief pieces.
Installing the NI 9969 Using12 AWG to14 AWG Wire
Complete the following steps to install the NI 9969 using 12 AWG
to 14 AWG wires.
NI 9244 User Manual and Specifications
|
© National Instruments |
11
Figure 3. 12 AWG to 14 AWG Installation
2
1
1 Route wires over the strain-relief piece.
2 Secure strain-relief piece and NI 9969 backshell in place using captive
screws.
12
|
ni.com |
NI 9244 User Manual and Specifications
Installing the NI 9969 Using16 AWG Wire
Complete the following steps to install the NI 9969 using 16 AWG
wires.
NI 9244 User Manual and Specifications
|
© National Instruments |
13
Figure 4. 16 AWG Installation
2
1
1 Route wires through the strain-relief pieces.
2 Secure strain-relief piece and NI 9969 backshell in place using captive
screws.
14
|
ni.com |
NI 9244 User Manual and Specifications
Installing the NI 9969 Using18 AWG to 24 AWG Wire
Complete the following steps to install the NI 9969 using 18 AWG
to 24 AWG wires.
NI 9244 User Manual and Specifications
|
© National Instruments |
15
Figure 5. 18 AWG to 24 AWG Installation
2
1
1 Route wires over the strain-relief piece.
2 Secure strain-relief piece and NI 9969 backshell in place using captive
screws.
16
|
ni.com |
NI 9244 User Manual and Specifications
Signals
The NI 9244 has three AI terminals and a Neutral terminal.
You can connect three-phase and single-phase measurement
configurations to the NI 9244. The NI 9244 supports standard
service levels up to 400 Vrms Line-to-Neutral (L-N) and 690 Vrms
Line-to-Line (L-L).
Note You must use 2-wire ferrules to create a secure
connection when connecting more than one wire to a
single terminal on the NI 9244.
Circuitry
Each channel on the NI 9244 provides an independent signal path
and ADC. Each terminal has the same input impedance to ground.
NI 9244 User Manual and Specifications
|
© National Instruments |
17
The NI 9244 returns the voltage between each AI terminal and the
Neutral terminal as well as the voltage between the Neutral
terminal and the chassis ground. Refer to Figure 6 for a diagram of
the equivalent voltages the module returns.
Figure 6. Internal Circuitry for the NI 9244
AI0
AI0
V
AI1
AI1
V
AI2
AI2
Neutral
V
Neutral
V
GND
NI 9244
18
|
ni.com |
NI 9244 User Manual and Specifications
To ensure that measurements to chassis ground are correct, NI
recommends connecting the chassis to earth ground using the
chassis grounding screw. Refer to your chassis manual for
information about connecting the chassis to earth ground.
Connecting Phase Measurements
You can connect three-phase measurement configurations and
single-phase measurement configurations to the NI 9244.
NI recommends using the following phase measurement
configurations for typical power distribution networks. Other valid
connections are possible if the connections do not exceed the safety
rating of the NI 9244.
NI 9244 User Manual and Specifications
|
© National Instruments |
19
Connecting Three-Phase Measurement
Configurations
You can connect WYE or delta measurement configurations to the
NI 9244.
Figure 7. Connecting a 4-Wire WYE Measurement Configuration
Phase A
AI0
Phase B
AI1
AI2
Phase C
Neutral
Neutral
NI 9244
20
|
ni.com |
NI 9244 User Manual and Specifications
Figure 8. Connecting a High Leg Delta Measurement Configuration
Phase A
AI0
Phase B
AI1
Phase C
AI2
Neutral
Neutral
NI 9244
NI 9244 User Manual and Specifications
|
© National Instruments |
21
Figure 9. Connecting a 3-Wire Delta Measurement Configuration
Phase A
AI0
Phase B
AI1
Phase C
AI2
NI 9244
22
|
ni.com |
NI 9244 User Manual and Specifications
Figure 10. Connecting a Corner Grounded 2-Wire Delta
Measurement Configuration
Phase A
AI0
Phase B
AI1
Phase C
AI2
NI 9244
Note The 9244 can not measure the entire tolerance
range or high crest factor signals on 690 Vrms systems in
this configuration. Refer to the Specifications section for
more information about the input range.
NI 9244 User Manual and Specifications
|
© National Instruments |
23
Connecting Single-Phase Measurement
Configurations
You can connect 3-wire or 2-wire single-phase measurement
configurations to the NI 9244.
Figure 11. Connecting a 3-Wire Measurement Configuration
(Split Phase)
AI0
AI1
Neutral
NI 9244
24
|
ni.com |
NI 9244 User Manual and Specifications
Figure 12. Connecting a 2-Wire Measurement
AI0
Neutral
NI 9244
Converting L-N Measurements to L-L
To convert L-N measurements to L-L values, calculate the voltage
difference between the AI channels using your application
software.
Refer to the following equation for an example of converting L-N
measurements to L-L.
Phase A to Phase B Voltage = AIO - AI1
where
AI0 is the reading from Phase A
AI1 is the reading from Phase B
NI 9244 User Manual and Specifications
|
© National Instruments |
25
Converting L-N Measurements to L-Earth
To convert L-N measurements to L-Earth values, add the neutral
channel reading to each AI channel reading.
Refer to the following equation for an example of converting L-N
measurements to L-Earth.
Line to Earth = AIx + Neutral
where
AIx is the analog input channel reading
Neutral is the Neutral channel reading
26
|
ni.com |
NI 9244 User Manual and Specifications
Wiring for High-Vibration Applications
If an application is subject to high vibration, National Instruments
recommends that you either use ferrules to terminate wires to the
detachable screw-terminal connector. Refer to Figure 13 for an
illustration of using ferrules.
Figure 13. 4-Terminal Detachable Screw-Terminal Connector
with Ferrule
NI 9244 User Manual and Specifications
|
© National Instruments |
27
Understanding NI 9244 Filtering
The NI 9244 uses a combination of analog and digital filtering to
provide an accurate representation of in-band signals while
rejecting out-of-band signals. The filters discriminate between
signals based on the frequency range, or bandwidth, of the signal.
The three important bandwidths to consider are the passband, the
stopband, and the alias-free bandwidth.
The NI 9244 represents signals within the passband, as quantified
primarily by passband ripple and phase nonlinearity. All signals
that appear in the alias-free bandwidth are either unaliased signals
or signals that have been filtered by at least the amount of the
stopband rejection.
Passband
The signals within the passband have frequency-dependent gain or
attenuation. The small amount of variation in gain with respect to
frequency is called the passband flatness. The digital filters of the
NI 9244 adjust the frequency range of the passband to match the
data rate. Therefore, the amount of gain or attenuation at a given
frequency depends on the data rate. Figure 14 shows typical
passband flatness for the NI 9244.
28
|
ni.com |
NI 9244 User Manual and Specifications
Figure 14. Typical Passband Response of the NI 9244
0.025
Gain (dB)
0.000
–0.025
–0.050
0
0.1
0.2
0.3
0.4
0.5
Frequency/Data Rate (Hz)
Stopband
The filter significantly attenuates all signals above the stopband
frequency. The primary goal of the filter is to prevent aliasing.
Therefore, the stopband frequency scales precisely with the data rate.
The stopband rejection is the minimum amount of attenuation applied
by the filter to all signals with frequencies within the stopband.
NI 9244 User Manual and Specifications
|
© National Instruments |
29
Alias-Free Bandwidth
Any signal that appears in the alias-free bandwidth of the NI 9244
is not an aliased artifact of signals at a higher frequency. The
alias-free bandwidth is defined by the ability of the filter to reject
frequencies above the stopband frequency, and it is equal to the
data rate minus the stopband frequency.
Understanding NI 9244 Data Rates
The frequency of a master timebase (fM) controls the data rate (fs) of
the NI 9244. The NI 9244 includes an internal master timebase with
a frequency of 12.8 MHz, but the module also can accept an
external master timebase or export its own master timebase. To
synchronize the data rate of an NI 9244 with other modules that use
master timebases to control sampling, all of the modules must share
a single master timebase source. Refer to the software help for
information about configuring the master timebase source for the
NI 9244. Visit ni.com/info and enter cseriesdoc for
information about C Series documentation.
30
|
ni.com |
NI 9244 User Manual and Specifications
The following equation provides the available data rates of the
NI 9244:
f M ÷ 256
fs = -------------------n
where n is any integer from 1 to 31.
However, the data rate must remain within the appropriate data rate
range. Refer to the Specifications section for more information
about the data rate range. When using the internal master timebase
of 12.8 MHz, the result is data rates of 50 kS/s, 25 kS/s,
16.667 kS/s, and so on down to 1.613 kS/s, depending on the value
of n. When using an external timebase with a frequency other than
12.8 MHz, the NI 9244 has a different set of data rates.
The NI 9151 R Series Expansion chassis does not
support sharing timebases between modules.
Note
NI 9244 User Manual and Specifications
|
© National Instruments |
31
Sleep Mode
This module supports a low-power sleep mode. Support for sleep
mode at the system level depends on the chassis that the module is
plugged into. Refer to the chassis manual for information about
support for sleep mode. If the chassis supports sleep mode, refer to
the software help for information about enabling sleep mode.
Typically, when a system is in sleep mode, you cannot
communicate with the modules. In sleep mode, the system
consumes minimal power and may dissipate less heat than it does
in normal mode.
Related Information
Power Requirements
32
|
ni.com |
NI 9244 User Manual and Specifications
Specifications
The following specifications are typical for the range -40 °C to 70 °C
unless otherwise noted.
Input Characteristics
Scaling coefficient ............................ 118,911 nV/LSB
Number of channels.......................... 4 analog input channels
ADC resolution................................. 24 bits
Type of ADC..................................... Delta-Sigma
(with analog prefiltering)
Sampling mode ................................. Simultaneous
Internal master timebase (fM)
Frequency ................................... 12.8 MHz
Accuracy..................................... ±100 ppm max
Data rate range (fs) using internal master timebase
Minimum.................................... 1.613 kS/s
Maximum ................................... 50 kS/s
Data rate range (fs) using external master timebase
Minimum.................................... 390.625 S/s
Maximum ................................... 51.2 kS/s
NI 9244 User Manual and Specifications
|
© National Instruments |
33
f M ÷ 256
Data rates1 (fs) ................................... --------------------- , n = 1, 2, …, 31
n
Input voltage ranges (AIx and Neutral-to-GND, AIx-to-Neutral)
Typical ........................................ 997.5 Vpk
Minimum.................................... 992 Vpk
Overvoltage withstand ...................... 800 Vrms continuous,
1000 Vrms for 1 s
Surge withstand ................................ 5 kV (1.2 µs/ 50 µs)
Input coupling................................... DC
Input impedance, AIx-to-Ground
and Neutral-to-GND ......................... 2 MΩ
1
The data rate must remain within the appropriate data rate range. Refer to the
Understanding NI 9244 Data Rates section for more information.
34
|
ni.com |
NI 9244 User Manual and Specifications
Table 1. DC and AC Accuracy
Measurement Conditions
Percent of
Reading
(Gain Error)
Percent of
Range*
(Offset Error)
Calibrated max (-40 °C to 70 °C)
0.17%
0.14%
Calibrated typ (23 °C ±5 °C)
0.043%
0.012%
Uncalibrated max (-40 °C to 70 °C)
0.42%
0.32%
Uncalibrated typ (23 °C ±5 °C)
0.18%
0.08%
*
Range equals 565.7 V (400 Vrms)
Accuracy specifications are valid for L-L, L-N
and L-Earth measurements.
Note
Input noise at 50 kS/s
L-Earth and N-Earth................... 4.22 mVrms
L-N and L-L1 .............................. 6 mVrms
1
The module returns L-N and N-Earth values only. Refer to Converting L-N
Measurements to L-L and Converting L-N Measurements to L-Earth sections for
details on changing the point of reference of the measurement
NI 9244 User Manual and Specifications
|
© National Instruments |
35
Note When measuring the amplitude of the
fundamental frequency over one or several power cycles
the noise of the measurement reduces significantly
(theoretically with the square root of the number of
samples in the acquisition window).
Nonlinearity (at 25 °C) ..................... 24 ppm
Stability
Gain drift .................................... 5 ppm/°C
Offset drift .................................. 6 mV/°C
Post calibration gain match (channel-to-channel, max)
Up to 20 kHz..................................... 109 mdB
Up to 10 kHz..................................... 39 mdB
Up to 3.8 kHz.................................... 19 mdB
Phase mismatch
channel-to-channel............................ 0.202°/kHz
Phase mismatch
(module-to-module, max) ................. 0.202°/kHz + 360° * fin/ fM
Phase nonlinearity (fs = 50 kS/s)
0 kHz to10 kHz .......................... 0.02° max
0 kHz to 20 kHz ......................... 0.06° max
36
|
ni.com |
NI 9244 User Manual and Specifications
5
Input delay ........................................ 40 --------- /fs + 1.7 μs
512
Passband,
Frequency ......................................... 0.453 * fs
Flatness
0 kHz to 20 kHz ......................... ±100 mdB max
0 kHz to 10 kHz ......................... ±20 mdB max
Negative phase sequence error at 50 Hz and 60 Hz
At 5% unbalance
Maximum............................. 0.13%
Typical ................................. 0.05%
At 1% unbalance
Maximum............................. 0.13%
Typical.................................. 0.05%
Zero phase sequence error at 50 Hz and 60 Hz
At 5% unbalance
Maximum............................. 0.13%
Typical ................................. 0.05%
NI 9244 User Manual and Specifications
|
© National Instruments |
37
At 1% unbalance
Max ...................................... 0.13%
Typ ....................................... 0.05%
Stopband
Frequency ................................... 0.547 * fs
Rejection..................................... -95 dB
Alias-free bandwidth ........................ 0.453 * fs
Anti-alias rejection (fS=50 kS/s) ....... 50 dB
-3 dB bandwidth (fs = 50 kS/s).......... 0.49 * fs
Crosstalk (60 Hz).............................. -100 dB
Crosstalk (1 kHz).............................. -74 dB
CMRR (fin = 60 Hz).......................... -78 dB
SFDR (1 kHz, -60 dBFS).................. -120 dB
Total Harmonic Distortion (THD),
1 kHz................................................. -98 dB
60 kHz............................................... -100 dB
MTBF ............................................... Contact NI for Bellcore
MTBF or MIL-HDBK-217F
specifications.
38
|
ni.com |
NI 9244 User Manual and Specifications
Power Requirements
Power consumption from chassis
Active mode ............................... 332 mW max
Sleep mode ................................. 50 μW max
Thermal dissipation
Active mode ............................... 652 mW max
Sleep mode ................................. 320 μW max
Physical Characteristics
If you need to clean the module, wipe it with a dry towel.
Note For two-dimensional drawings and
three-dimensional models of the C Series module and
connectors, visit ni.com/dimensions and search by
module number.
Screw-terminal wiring ...................... 0.511 mm diameter
(24 AWG) to 2.053 mm
diameter (12 AWG) copper
conductor wire with 7 mm
(0.28 in.) of insulation
stripped from the end
NI 9244 User Manual and Specifications
|
© National Instruments |
39
Torque for screw terminals ............... 0.5 N · m to 0.6 N · m
(4.42 lb · in. to 5.30 lb · in.)
Ferrules ............................................. 0.25 mm2 to 2.5 mm2
Weight............................................... 150 g (5.3 oz)
Safety
Connect only voltages that are within the following limits.
Maximum working voltage, channel-to-earth ground
Up to 2,000 m altitude
Continuous ........................... 400 Vrms,
Measurement Category III
2,001 to 5,000 m altitude
Continuous ........................... 400 Vrms,
Measurement Category II or
300 Vrms,
Measurement Category III
Division 2 and Zone 2 hazardous locations applications
Channel-to-earth ground................... 300 Vrms
Measurement Category III
40
|
ni.com |
NI 9244 User Manual and Specifications
Measurement Category III is for measurements performed in the
building installation at the distribution level. This category refers
to measurements on hard-wired hardware such as hardware in
fixed installations, distribution boards, and circuit breakers. Other
examples are wiring, including cables, bus bars, junction boxes,
switches, socket outlets in the fixed installation, and stationary
motors with permanent connections to fixed installations.
Measurement Category II is for measurements performed on
circuits directly connected to the electrical distribution system.
This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S.
or 230 V for Europe.
Caution Do not connect the NI 9244 to signals or use
for measurements within Measurement Categories or IV.
Hazardous Locations
U.S. (UL) .......................................... Class I, Division 2,
Groups A, B, C, D, T4;
Class I, Zone 2,
AEx nA IIC T4
NI 9244 User Manual and Specifications
|
© National Instruments |
41
Canada (C-UL) ................................. Class I, Division 2,
Groups A, B, C, D, T4;
Class I, Zone 2,
Ex nA IIC T4
Europe (DEMKO) ............................ Ex nA IIC T4 Gc
Safety and Hazardous Locations Standards
This product meets the requirements of the following standards of
safety for electrical equipment for measurement, control, and
laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
•
EN 60079-0:2009, EN 60079-15:2010
•
IEC 60079-0:2007; Ed 5, IEC 60079-15:2010; Ed 4
•
UL 60079-0: Ed 5, UL 60079-15: Ed 3
•
CSA 60075-0:2011, CSA 60079-15:2012
Note For UL and other safety certifications, refer to the
product label or the Online Product Certification section.
42
|
ni.com |
NI 9244 User Manual and Specifications
Electromagnetic Compatibility
This product meets the requirements of the following EMC
standards for electrical equipment for measurement, control, and
laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Industrial
immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
EN 55022 (CISPR 22): Class A emissions
•
EN 55024 (CISPR 24): Immunity
•
AS/NZS CISPR 11: Group 1, Class A emissions
•
AS/NZS CISPR 22: Class A emissions
•
FCC 47 CFR Part 15B: Class A emissions
•
ICES-001: Class A emissions
Note In the United States (per FCC 47 CFR),
Class A equipment is intended for use in commercial,
light-industrial, and heavy-industrial locations.
In Europe, Canada, Australia and New Zealand
(per CISPR 11) Class A equipment is intended for use
only in heavy-industrial locations.
NI 9244 User Manual and Specifications
|
© National Instruments |
43
Note Group 1 equipment (per CISPR 11) is any
industrial, scientific, or medical equipment that does not
intentionally generates radio frequency energy for the
treatment of material or inspection/analysis purposes.
Note For EMC declarations and certifications, refer to
the Online Product Certification section.
CE Compliance
This product meets the essential requirements of applicable
European Directives as follows:
•
2006/95/EC; Low-Voltage Directive (safety)
•
2004/108/EC; Electromagnetic Compatibility Directive
(EMC)
Online Product Certification
To obtain product certifications and the Declaration of Conformity
(DoC) for this product, visit ni.com/certification, search by
module number or product line, and click the appropriate link in
the Certification column.
44
|
ni.com |
NI 9244 User Manual and Specifications
Shock and Vibration
To meet these specifications, you must panel mount the system. If
you are using the NI 9244 with screw terminal, you also must
either affix ferrules to the ends of the terminal wires.
Operating vibration
Random (IEC 60068-2-64)......... 5 grms, 10 Hz to 500 Hz
Sinusoidal (IEC 60068-2-6) ....... 5 g, 10 Hz to 500 Hz
Operating shock
(IEC 60068-2-27).............................. 30 g, 11 ms half sine,
50 g, 3 ms half sine,
18 shocks at 6 orientations
Environmental
Refer to the manual for the chassis you are using for more
information about meeting these specifications.
Operating temperature
(IEC 60068-2-1, IEC 60068-2-2) ..... -40 °C to 70 °C
Storage temperature
(IEC 60068-2-1, IEC 60068-2-2) ..... -40 °C to 85 °C
Ingress protection.............................. IP 40
NI 9244 User Manual and Specifications
|
© National Instruments |
45
Operating humidity
(IEC 60068-2-56).............................. 10% to 90% RH,
noncondensing
Storage humidity
(IEC 60068-2-56).............................. 5% to 95% RH,
noncondensing
Pollution Degree ............................... 2
Maximum altitude1 ........................... 5,000 m
Indoor use only.
Environmental Management
NI is committed to designing and manufacturing products in an
environmentally responsible manner. NI recognizes that
eliminating certain hazardous substances from our products is
beneficial to the environment and to NI customers.
For additional environmental information, refer to the Minimize
Our Environmental Impact Web page at ni.com/environment.
This page contains the environmental regulations and directives
with which NI complies, as well as other environmental
information not included in this document.
1
Refer to the Safety section for more information about altitude.
46
|
ni.com |
NI 9244 User Manual and Specifications
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle,
all products must be sent to a WEEE recycling center.
For more information about WEEE recycling centers,
National Instruments WEEE initiatives, and compliance
with WEEE Directive 2002/96/EC on Waste and
Electronic Equipment, visit ni.com/environment/
weee.
⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё೑ RoHS˅
Ё೑ᅶ᠋ National Instruments ヺড়Ё೑⬉ᄤֵᙃ
ѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ (RoHS)DŽ݇Ѣ
National Instruments Ё೑ RoHS ড়㾘ᗻֵᙃˈ䇋ⱏᔩ
ni.com/environment/rohs_chinaDŽ (For information
about China RoHS compliance, go to ni.com/
environment/rohs_china.)
Calibration
You can obtain the calibration certificate and information about
calibration services for the NI 9244 at ni.com/calibration.
Calibration interval ........................... 1 year
NI 9244 User Manual and Specifications
|
© National Instruments |
47
Worldwide Support and Services
The National Instruments website is your complete resource for
technical support. At ni.com/support you have access to
everything from troubleshooting and application development
self-help resources to email and phone assistance from
NI Application Engineers.
Visit ni.com/services for NI Factory Installation Services,
repairs, extended warranty, and other services.
Visit ni.com/register to register your National Instruments
product. Product registration facilitates technical support and
ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance
with the Council of the European Communities using the
manufacturer’s declaration of conformity. This system affords the
user protection for electromagnetic compatibility (EMC) and
product safety. You can obtain the DoC for your product by visiting
ni.com/certification. If your product supports calibration,
you can obtain the calibration certificate for your product
at ni.com/calibration.
48
|
ni.com |
NI 9244 User Manual and Specifications
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world. For
telephone support in the United States, create your service request
at ni.com/support or dial 1 866 ASK MYNI (275 6964). For
telephone support outside the United States, visit the Worldwide
Offices section of ni.com/niglobal to access the branch office
websites, which provide up-to-date contact information, support
phone numbers, email addresses, and current events.
NI 9244 User Manual and Specifications
|
© National Instruments |
49
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of
their respective companies. For patents covering National Instruments products/technology, refer to the appropriate
location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent
Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and
third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at
ni.com/legal/export-compliance for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data.
© 2014 National Instruments. All rights reserved.
376131B-01
Mar14