Medical Grade Isolation Transformers - Digi-Key

MEDICAL GRADE ISOLATION TRANSFORMERS
Features
● Power Rating 50VA to 3000 VA
● Dual Primary and Dual Secondary windings
● Frequency 50-60Hz
● Isolation 5KVrms Primary-Secondary
● Creepage Distance minimum 8mm
● Leakage Current < 100µA
● Tolerance ±3% on load
● Various Mounting Methods available
● Recognized to UL60601-1 (Agency Reference - XORU2.E251176 & XORU8.E251176)
● Inbuilt Thermal Protection
● Magnetic Shield to minimize Strayfield Emission (Optional)
● Class B Insulation System
● Safety requirements conforms to IEC 601-1 Class I & II equipment
● RoHS Compliant : 2011/65/EU/ (RoHS 2)
● REACH Compliant (EC 1907/2006)
Ratings Table
Transformer
Part Number
Power
0050MD-X-00Y
0100MD-X-00Y
0200MD-X-00Y
0230MD-X-00Y
0300MD-X-00Y
0350MD-X-00Y
0400MD-X-00Y
0500MD-X-00Y
0600MD-X-00Y
0650MD-X-00Y
0760MD-X-00Y
0800MD-X-00Y
(VA)
50
100
200
230
300
350
400
500
600
650
760
800
1000MD-X-00Y
1200MD-X-00Y
1500MD-X-00Y
1800MD-X-00Y
2000MD-X-00Y
2500MD-X-00Y
3000MD-X-00Y
1000
1200
1500
1800
2000
2500
3000
Secondary Ratings
120 VAC
240 VAC
A (rms)
A (rms)
0.42
0.21
0.83
0.42
1.67
0.83
1.92
0.96
2.50
1.25
2.92
1.46
3.33
1.67
4.17
2.08
5.00
2.50
5.42
2.71
6.33
3.17
6.67
3.33
8.33
10.00
12.50
15.00
16.67
20.83
25.00
4.17
5.00
6.25
7.50
8.33
10.42
12.50
No Load
Voltage
(V)
2x135.5
2x131.9
2x129.1
2x128.8
2x127.5
2x126.7
2x126.4
2x125.6
2x125.4
2x125.2
2x124.8
2x124.4
2x124.6
2x123.5
2x123.4
2x123.7
2x122.8
2x123.3
2x121.9
Physical Dimensions **
OD
HT
(mm)
(mm)
81
38
99
41
117
56
117
61
127
62
132
58
132
64
134
70
150
70
157
70
160
71
165
72
168
183
200
206
215
240
265
80
81
90
94
105
115
120
Weight
**
(KG)
0.8
1.1
2.3
2.4
2.8
3.0
3.4
4.3
4.7
5.1
5.9
6.1
8.2
10.0
11.8
13.6
15.5
21.0
26.0
** Physical dimensions and weight given are approximate
Germany: Int.+49 89 841 00 0 ● Ireland: Int.+35 374 954 8666 ● Czech Rep: Int.+420 377 338 351 ● India: Int.+91 427 2433 000 ●
USA: Int. +1 573 303 3675 ext 802
http://www.talema nuvotem.com
Initial Release June 2013
MEDICAL GRADE ISOLATION TRANSFORMERS
Schematic
002
001
003
004
Schematic
Construction
Functional Earth
001
Reinforced insulation
No
No
002
Reinforced insulation
No
Yes (Internal, in-line)
003
Reinforced insulation
Yes
No
004
Reinforced insulation
Yes
Yes (Internal, in-line)
Mechanical Specification
Thermal Protection
Ordering Code
Mounting
● Center potted with through hole
● Metal Disc and Insulating pads (Optional)
Germany: Int.+49 89 841 00 0 ● Ireland: Int.+35 374 954 8666 ● Czech Rep: Int.+420 377 338 351 ● India: Int.+91 427 2433 000 ●
USA: Int. +1 573 303 3675 ext 802
http://www.talema nuvotem.com
Initial Release June 2013