Features
o Available in 0-15 dB or 0-31 dB Configurations.
o Provides a flexible, easy to program, low cost solution
for your bench test/calibration setups and subsystem
applications.
o Relative vs. Nominal attenuation step function.
o DC to 13.0 GHz Operation.
o High Accuracy & Repeatability.
o Power Handling up to 100 Watts average
o Designed to interface with industry standard
communication interfaces:
Description
Aeroflex / Weinschel's design approach uses a highly
adaptable platform that allows configuration of the step
attenuator to the customers requirements. When the controller requests a new attenuation level the input switch
terminates the input signal into a 50 Ohm load. (See Figure
1) This input switch is hot switchable at 100 Watts of input
power. This will remove the high power signal from the main
signal path. With no signal connected to the attenuator path
the controller then commands the series of relays to configure the attenuator for the requested attenuation value. Then
the input switch re-connects the input signal to the attenuator path. The system can be operated with either a remote
controller (IEEE-488 or RS-232) or through front panel
control.
- GPIB/IEEE-488 (HS-488 ready)
- RS-232, RS-422
o Built-in monitoring for switching input power into the
load in case of fan failure.
o Rack Configurable: A Rack Mounting Kit is included
for easily mounting the Model 8312 into any rack or
cabinet that is designed per EIA RS-310 or
MIL-STD-189.
HIGH POWER
HOT SWITCH
RFOUT
RF IN
50 Ω, 100 W
TERMINATION
1 dB
100 W
2 dB
100 W
4 dB
100 W
8 dB
100 W
16 dB
100 W
OPTIONAL
Device
Interface
Bus
SMARTSTEP
CONTROLLER
POWER
SUPPLY
Note: If power failure should occur, the unit will remain in the last selected attenuation state.
Figure 1. Model 8312 Block Diagram
For additional information on the Model 8312, visit our website @ www.aeroflex.com/AW/8312.htm
255
5305 Spectrum Drive, Frederick, MD 21703-7362 s TEL: 301-846-9222, 800-638-2048 s Fax: 301-846-9116
web: www.aeroflex.com/weinschel s email: [email protected]
Revision Date: 9/30/2012
Programmable Attenuators
Specifications
SPECIFICATION
DESCRIPTION
Input Power Requirements
AC
100 to 240 Vac, 50/60 Hz, 50 Watts
Environmental
Operating Temperature
Storage Temperature:
Humidity:
Altitude:
IEEE-488 Bus
Connector:
Protocols:
Indicators:
24-pin per IEEE-488.1
per IEEE-488.2
Remote (RMT), Listen (LSN), Talk (TLK), SRQ (SRQ)
RS-232 Bus
Connector:
Signals:
Baud Rates:
Data Bits:
Handshaking:
Parity:
Indicators:
9-pin male D
TXD, RXD, RTS, CTS, DTR, GND
2400, 9600, 19200, and 38400
8
None, RTS/CTS, XON/XOFF
None, Odd, Even
Tx (Transmit) and Rx (Receive)
RS-422 BUS(3)
Connector:
Signals:
Baud Rates:
Data Bits:
Handshaking:
Parity:
Indicators:
9-pin male D
TXD+, TDX-, RXD+, RTX-, RTS+, RTS-, CTS+, CTS- and signal GND
2400, 9600, 19200, and 38400
8
None, RTS/CTS, XON/XOFF
None, Odd, Even
Tx (Transmit) and Rx (Receive)
RF Characteristics(4)
Connectors:
Frequency Range:
Impedance:
Type N, Female
dc - 13 GHz
50 Ω
SWR:
50 MHz - 5 GHz:
5 GHz - 13 GHz:
Attenuation Range:
15 dB/1 dB steps (8312-15-F)
31 dB/1 dB steps (8312-31-F)
RF Power Rating:
50 MHz - 5 GHz:
5 GHz - 13 GHz:
Attenuation Settings:
100, 000 selections (minimum)
Attenuation Update Rate:
1 second (Typical)
Incremental Accuracy:
Insertion Loss (dB):
0 to +50°C
67° to +167 °F (-55° to +75°C)
96%
40,000' (12,192M)
1.60 (Maximum)
2.30 (Maximun)
100 Watts (Maximum)
50 Watts (Maximum)
Frequency
1-15 dB
16-31 dB
50 MHz - 3 GHz:
3 GHz - 5 GHz:
5 GHz - 13 GHz:
+0.6 dB
+0.6 dB
+2.5 dB
+0.8 dB
+0.8 dB
+3.0 dB
Frequency Range
50 MHz - 3 GHz:
3 GHz - 5 GHz:
5 GHz - 13 GHz:
8312-15-X
3.0
4.0
7.0
8312-31-X
3.5
4.5
8.0
1. GPIB/IEEE-488 model allows user-selectable addresses.
2. RS-232 can be used with standard PC serial port for short and medium distances (up to approximately 50 ft).
3. RS-422, designed for very long distance communications (4000 ft) and & optimized as a single node protocol, typically with one device connected to a
single port.
4. Refer to Individual data sheet for detailed specifications on internal programmables.
256
5305 Spectrum Drive, Frederick, MD 21703-7362 s TEL: 301-846-9222, 800-638-2048 s Fax: 301-846-9116
web: www.aeroflex.com/weinschel s email: [email protected]
Revision Date: 9/30/2012
Programmable Attenuators
Physical Dimensions
NOTE:
All dimensions are given in mm (inches) and are maximum, unless otherwise specified.
MODEL NUMBER DESCRIPTION:
Example:
8312 - XX - X
Basic
Model
Number
Attenuation Connector Location
F = Front
Value (dB)*
R = Rear
* Available in 0-15 dB and 0-31 dB configurations only!
257
5305 Spectrum Drive, Frederick, MD 21703-7362 s TEL: 301-846-9222, 800-638-2048 s Fax: 301-846-9116
web: www.aeroflex.com/weinschel s email: [email protected]
Revision Date: 9/30/2012