User Guide 400Amp True RMS AC Clamp Meter + NCV Model MA410T Introduction Thank you for selecting the Extech MA410 Clamp Meter. This meter measures AC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Diode Test, Duty Cycle and Continuity. Custom features include Thermocouple Temperature and Non-Contact Voltage detection. The double molded case is designed for heavy duty use. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Extech Instruments is an ISO-9001 Certified company. Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present Double insulation WARNING This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury. CAUTION This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product. PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient over voltages to an appropriate low level. Note – Examples include protected electronic circuits. OVERVOLTAGE CATEGORY II Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Note – Examples include household, office, and laboratory appliances. OVERVOLTAGE CATEGORY III Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations. Note – Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation. OVERVOLTAGE CATEGORY IV Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the installation. Note – Examples include electricity meters and primary over-current protection equipment 2 MA410T-EU-EN-V1.0-7/12 SAFETY NOTES Do not exceed the maximum allowable input range of any function. Do not apply voltage to meter when resistance function is selected. Set the function switch OFF when the meter is not in use. Remove the battery if meter is to be stored for longer than 60 days. WARNINGS Set function switch to the appropriate position before measuring. When measuring volts do not switch to current/resistance modes. Do not measure current on a circuit whose voltage exceeds 600V. When changing ranges always disconnect the test leads from the circuit under test. CAUTIONS Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter. Always remove the test leads before replacing the battery or fuses. Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use. Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard. Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests. Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not "live". If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Function Maximum Input A AC, 400A V DC, V AC 600V DC/AC Resistance, Capacitance, Frequency, Diode Test, Continuity, Temperature 250V DC/AC 3 MA410T-EU-EN-V1.0-7/12 Description Meter Description 1. Current clamp 2. NCV LED indicator 3. Clamp opening trigger 4. MODE button 5. HZ/% button 6. LCD Display 7. Negative input jack 8. Non-Contact Voltage Detector 9. HOLD button 10. Function switch 11. Relative Button 12. Battery compartment (rear) 13. Positive input jack Display icons Description HOLD AUTO DC AC REL V A F Hz % o o F and C n, m, , M, k ))) Data Hold Autoranging Direct Current Alternating Current Low battery Relative Volts (Voltage) Ohms (Resistance) Amperes (Current) Farad (Capacitance) Hertz (Frequency) Duty Ratio Fahrenheit and Celsius units (Temperature) Unit of measure prefixes: nano, milli, micro, mega, and kilo Continuity test Diode test 4 MA410T-EU-EN-V1.0-7/12 Operation NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use. Non-Contact Voltage Detector WARNING: Risk of Electrocution. Before use, always test the Voltage Detector on a known live circuit to verify proper operation. 1. Rotate the Function switch to any measurement position. 2. Place the detector probe tip on the conductor to be tested. 3. If AC voltage is present, the NCV detector light will turn on with a steady red light. NOTE: The conductors in electrical cord sets are often twisted. For best results, move the probe tip along a length of the cord to assure placing the tip in close proximity to the live conductor. NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy may randomly trip the sensor. This is normal operation. AC Current Measurements WARNING: Disconnect the test leads before making clamp measurements. 1. Rotate the Function switch to the 400A position 2. Press the trigger to open jaw. Fully enclose only one conductor. 3. Read the current value in the display. 4. If the value is less than 40A, rotate the function switch to the 40A position to improve resolution. 5 MA410T-EU-EN-V1.0-7/12 AC/DC Voltage, Frequency, Duty Cycle Measurements CAUTION: Do not measure voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter. 1. Rotate the function switch to the V position. 2. Press the MODE button to select AC or DC Voltage. 3. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit. Read the voltage value in the display. Press the Hz % button to display Frequency Press the Hz % button to display Duty Cycle 4. 5. 6. 7. Resistance, Diode, Continuity Measurements Note: Remove power from the device under test before making resistance measurements 1. Set the function switch to the position. 2. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack. 3. Touch the black test probe tip to one side of the device. Touch the red test probe tip to the other side of the device. 4. Read the resistance value in the display. 5. Press the MODE button to select the DIODE mode. The Diode symbol will appear in the display. 6. Press the MODE button to select the Continuity mode. The continuity symbol will appear in the display. If the resistance is <150 ohms the tone will sound. 6 MA410T-EU-EN-V1.0-7/12 Capacitance Measurements WARNING: To avoid electric shock, discharge the capacitor before measuring. 1. Rotate the function switch to the CAP capacitance position. 2. Insert the black test lead banana plug into the negative COM jack. jack. Insert the red test lead banana plug into the positive 3. Touch the black test probe tip to one side of the device. Touch the red test probe tip to the other side of the device. 4. Read the capacitance value in the display. Type K Temperature Measurements 1. Rotate the function switch to the °F or °C temperature position. 2. Insert the Temperature Probe into the input jacks. 3. Place the temperature probe tip(s) where needed. 4. Read the temperature on the display. Note: In case of an open input or a temperature overrange, the meter will display “OL” . Data Hold To freeze the LCD reading, press the HOLD button. The HOLD icon appears on the LCD. Press the HOLD button again to return to normal operation. Relative Press the REL button (REL will appear in the display) to zero the reading and create a reference point. All future reading will be the difference between the actual reading and the stored “REL” reading. In the REL mode, autoranging is disabled and measurements are limited to the range active when the REL button was pressed. Press the REL button to exit the mode. Note: Relative does not function in Frequency or Duty Cycle mode. Automatic Power Off In order to conserve battery life, the meter will automatically turn off after approximately 30 minutes. To turn the meter on again, change the position of the function switch. Low battery indication When the icon appears in the display, the battery should be replaced. Refer to the battery replacement procedure in the maintenance section. 7 MA410T-EU-EN-V1.0-7/12 Maintenance WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case. Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately. Battery Replacement 1. Remove the 2 Phillips head screws that secure the rear battery cover 2. Open the battery compartment 3. Replace the 2 AAA 1.5V batteries. 4. Secure the battery compartment door All EU users are legally bound by the Battery Ordinance to return all used batteries to community collection points or wherever batteries / accumulators are sold! Disposal in household trash or refuse is prohibited! Disposal: Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle. 8 MA410T-EU-EN-V1.0-7/12 Specifications Function Range AC Current 40.00 AAC 0.01A Accuracy (% of reading + digits) ±(2.5% + 12 digits) 400.0 AAC 0.1A ±(2.8% + 8 digits) 50/60Hz Resolution AC Current ranges specified from 5% to 100% of range AC Voltage 50 to 400Hz 4.000 VAC 0.001V 40.00 VAC 0.01V 400.0 VAC 0.1V 600 VAC 1V ±(1.8% + 10 digits) ±(2.5% + 10 digits) AC Voltage ranges specified from 5% to 100% of range DC Voltage 400.0 mVDC 0.1mV 4.000 VDC 0.001V 40.00 VDC 0.01V 400.0 VDC 0.1V 600 VDC Resistance Capacitance Frequency ±(0.8% + 2 digits) ±(1.5% + 2 digits) 1V ±(2.0% + 2 digits) 400.0 0.1 ±(1.0% + 4 digits) 4.000k 0.001k 40.00k 0.01k ±(1.5% + 2 digits) 400.0k 0.1k 4.000M 0.001M ±(2.5% + 3 digits) 40.00M 0.01M ±(3.5% + 5 digits) 40.00nF 0.01nF ±(4.0% + 20 digits) 400.0nF 0.1nF 40.00µF 0.01µF 100.0µF 0.1µF 10 to 10kHz 0.01Hz ±(1.5% + 2 digits) 0.1% ±(1.2% + 2 digits) ±(3.0% + 5 digits) Sensitivity: 15V rms Duty Cycle 0.5% to 99.0% Pulse width: 100s to 100ms, Frequency: 10Hz to 10kHz Temperature Type K -4.0 to 1400°F -20 to 760°C 0.1° <400° 1° >400° ±(3% + 9°F/5°C) Specification does not include probe accuracy. Range of supplied probe is -4 to 482°F (-20 to 250°C). 9 MA410T-EU-EN-V1.0-7/12 General Specifications Clamp jaw opening Display Continuity check Diode test Low Battery indication Over-range indication Measurement rate Thermocouple sensor Input Impedance AC bandwidth AC response Operating Temperature Storage Temperature Operating Humidity Storage Humidity Operating Altitude Battery Auto power OFF Dimensions & Weight Safety Approvals 30mm (1.18") approx. 4,000 count LCD Threshold <150; Test current < 0.5mA Test current of 0.3mA typical; Open circuit voltage 1.5VDC typical Battery symbol is displayed ‘OL’ display 2 readings per second, nominal Type K thermocouple required 10M (VDC and VAC) 50 to 400Hz (VAC) True rms 5C to 40C (41F to 104F) -20C to 60C (-4F to 140F) Max 80% up to 31C (87F) decreasing linearly to 50% at 40C (104F) <80% 2000meters (7000ft) maximum. Two “AAA” 1.5V batteries After approx. 30 minutes 200x66x37mm (7.9x2.6x1.5”); 205g (7.23oz) For indoor use and in accordance with the requirements for double insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage Category III 600V, Pollution Degree 2. CE, Copyright © 2012 Extech Instruments Corporation (a FLIR company) All rights reserved including the right of reproduction in whole or in part in any form ISO-9001 Certified www.extech.com 10 MA410T-EU-EN-V1.0-7/12