V2F100-Datasheet3 V1.1

V2F100
100 MHz Voltage to Frequency Converter
Technical Datasheet
The design of this unit by the ESRF is a major leap forward for V2F design,
it has led to two orders of magnitude improvement in resolution as well as
bringing welcome updates to usability and compatibility.
KEY benefits
• 100X the accuracy of other V2F systems
• Low noise
• Easy configuration
The system is a two channel unit, with standard BNC inputs and LEMO outputs. The configuration of
frequency output range (10, 25, 50, 100 MHz), polarity, frequency offset (0 or 12MHz) and input voltage range
(1, 2.5, 5, 10V) are altered with a simple push button interface on the front panel. This configuration is
displayed using LEDs and stored through power cycles with non-volatile memory.
In addition to the standard LEMO outputs, a 2-phase quadrature output and absolute digital SSI output are
available on the rear through a sub-D 15 connector.
Features
V2f Output Frequency
10, 20, 50 or 100 MHz TTL
Input Voltage
0
0
0
0
Input Configuration
Isolated BNC
TTL output frequency offset
0 or 12 MHz
Linearity Error
5ppm
Outputs
Front: LEMO 00, 5V TTL @ 50Ω
10V or 0
-10V
5V or 0
-5V
2.5V or 0
-2.5V
1 V or 0
-1 V
Rear: differential with RS422 electrical levels:
12MHz, 2-phase qaudrature
SSI, 24 bits, clock up to 10 MHz
Noise
1-10 µV / channel
Format
Stand alone unit (100 - 220V IEC/EN 60320-1 input)
Power Consumption
5W
Dimensions
55 x 105 x 109 mm, 0.4 Kg
Web: quantumdetectors.com/v2f100 Email: [email protected] Telephone: +44 1235 44 5795
Performance
V2F100
The following is an example of one of the calibration plots that is sent with each unit. The chart shows both
the linearity (in blue) and the deviation from linearity (in red).
Data taken at the ESRF by the Electronics Support Laboratory.
Availability
The V2F100s are popular at synchrotrons around the globe, with over 40 units now at the ESRF, Spring-8,
Photon Factory, Diamond and elsewhere. The units are produced in batches and kept in stock at Quantum
Detectors. To ensure we have sufficient stock to allocate to your order, please let us know as soon as you can
about your requirement.
User Customisable
The V2F100 is easy to configure using the
CONFIG buttons on the front panel.
• To start configuration press CONFIG for 2s
• To change the setting, tap CONFIG
• To move to the next box, hold CONFIG for 1s
• To close configuration mode hold CONFIG for 2s
Web: quantumdetectors.com/v2f100 Email: [email protected] Telephone: +44 1235 44 5795
QD V2F 1MHz @ 50% of Vin max
QD V2F100 (50MHz mode) @ 50% of Vin max
200
180
160
Measurement Uncertainty (ppm)
V2F100
Comparison of Measurement Uncertainty against Integration Time
for Quantum Detectors V2F and V2F100
140
120
100
80
60
40
20
0
10-4
10-3
10-2
10-1
10-0
101
Integration Time (s)
Analysis by Senior Detector Engineer at Diamond Light Source.
Web: quantumdetectors.com/v2f100 Email: [email protected] Telephone: +44 1235 44 5795