Downloads for 7500 Voltage Ampere (V A) Power VPM Series Toroidal Medical Transformer (VPM240-31200)

POWER TRANSFORMER
CHASSIS MOUNT :
TOROIDAL MEDICAL SERIES
VPM240-31200
Description:
The toroidal construction inherently reduces stray fields, increases efficiency and minimizes size
compared to traditional EI transformers. The addition of a Flux Band further reduces the
remaining stray fields. The shield between Primary and Secondary improves safety, reduces
common mode signals and minimizes leakage current. Built with a Class F (155°) insulation
system. A 140°C self-resetting thermal switch is included in each primary.
Photo for illustration only
RoHS
2011/65/EU
Electrical Specifications (@25C)
1. Maximum Power: 7500VA
2. Input Voltages: 100, 120, 220, 240VAC, 50/60Hz
3. Output Voltages: 120VAC @62.40A or 240VAC CT @ 31.20A
4. Voltage Regulation: 0.9% TYP from full load to no load
5. Temperature Rise: 45°C TYP
6. Hipot: 4000VAC, Primary to Secondary, Primary & Secondary to Shield & mounting surface
7. Efficiency: 95% TYP. @ full load
Agency File:
UL: File E122529, UL 60601-1/(R) 2012 Medical Electrical Equipment – Part 1
CE: ES 60601-1 (IEC 60601-1:2005, MOD)
cUL: C22.2 No. 60601-1:14, Medical Electrical Equipment – Part 1
CB Certified.
OD
Dimensions: Inches (mm)
O.D.
I.D.
HT.*
13(330)
3.8(96)
5.5(140)
*Add 0.188 (3) to the height for mounting hardware
Weight: 62Kg
BOLT
MOUNTING
METAL
PLATE
Mounting:
FLAT
WASHER
RUBBER
PAD
Transformer is provided with one metal mounting plate, two rubber pads, M12 x 140mm bolt, nut,
spring and flat washer.
Connections:
H
Transformer is provided with 12” (305mm) long, 0.5” (12mm) stripped and tinned, stranded
UL 1015 lead wire. Primaries are 10AWG, Secondaries are 10AWG, and Shield is 18AWG.
The GRN/YEL shield lead is typically grounded. Do not lift transformer by leads!
Input Options:
100VAC: Input to Gray & Blue, jumper White & Brown, jumper Blue & Violet.
120VAC: Input to White & Blue, jumper White & Brown, jumper Blue & Violet.
220VAC: Input to Gray & Violet, jumper Blue & Brown
240VAC: Input to White and Violet, jumper Blue & Brown
Output Options:
120VAC: Output from Black & Red, jumper Black & Orange, jumper Red to Yellow
240VAC: Output from Black & Yellow, jumper Red & Orange
PRI
Primary and secondary windings are designed to be connected in series or parallel. Windings are not
intended to be used independently.
RoHS Compliance: Meets the requirements of 2011/65/EU, known as the RoHS 2 initiative.
460 Harley Knox Blvd.
Perris, California 92571
NUT
RUBBER
PAD
WHT
GRY
BLK
BLU
RED
BRN
ORG
VIO
YEL
GRN/YEL
* At printing, this document is considered “uncontrolled”. Contact Triad Magnetics’ website for current version
Web: www.TriadMagnetics.com
Phone 951-277-0757
Fax 951-277-2757
SPRING
WASHER
SCHEMATIC
Publish Date: March 2, 2016
SEC