EVB90614 User Manual old 407 DownloadLink 5196

USER MANUAL
MLX90614 Evaluation board
MLX90614 Single and Dual zone Infra Red
thermometer Evaluation Board
EVB90614
User Manual
390129061401
Rev 003
Page 1 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
Table of Contents
1 Introduction ..................................................................................................................... 3
2 Host computer requirements ......................................................................................... 3
3 Installing the software .................................................................................................... 3
4 Getting started with MLX90614 evaluation board ........................................................ 4
4.1 MLX90614 evaluation board overview......................................................................................................................................... 4
4.2 Connecting the EVB90614 evaluation board ............................................................................................................................... 5
5 Getting started with MLX90614 Configurator ............................................................... 5
5.1 MLX90614 Configurator overview................................................................................................................................................ 5
5.2 Working with MLX90614 Configurator ......................................................................................................................................... 5
5.3 Configuration of MLX90614 modules........................................................................................................................................... 8
5.4 Manual sending commands....................................................................................................................................................... 11
5.5 Demonstration of SMBus and PWM temperature measuring ..................................................................................................... 12
5.6 Quick customization of MLX90614 thermometers...................................................................................................................... 15
5.7 Updating EVB90614 firmware ................................................................................................................................................... 17
6 Troubleshooting............................................................................................................ 19
7 Disclaimer...................................................................................................................... 20
♦APPENDIX A – EVB90614 commands set ...................................................................... 21
♦APPENDIX B – EVB90614 schematics............................................................................ 22
390129061401
Rev 003
Page 2 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
1 Introduction
The EVB90614 is designed to support MLX90614 infrared thermometer modules.
The communication between PC and the evaluation board is accomplished by USB.
The Demonstration Kit contains the following items:
1. FS USB demonstration board, pre-programmed with USB
bootloader and demonstration firmware.
2. A standard USB cable for use in communicating with the board.
3. 1 pc. MLX90614AAA (single zone 5V)
4. 1 pc. MLX90614BAA (single zone 3V)
5. CD-ROM, containing the USB HID driver and EVB90614 software
The EVB90614 is designed to allow OEMs to configure the MLX 90614 IR thermometer
for virtually any application quickly. OEMs can quickly experiment with temperature
ranges, optics, etc. to find the best IR configuration to meet the application needs. Once
the best configuration is established, low volume OEMs can easily configure IR modules
for their own use. For high volume OEMs, Melexis can supply special configured
modules, ready to install into the customer’s application. Please contact IR Sales at
Melexis for quotations.
The EVB90614 allows engineers to customize the MLX90614 for high accuracy IR
measurements in an R&D or Laboratory setting.
2 Host computer requirements
To communicate with and program the EVB90614 evaluation board, the following
hardware and software requirements must be met:
• PC-compatible system
• An available USB port
• CD-ROM drive (for use with the accompanying CD)
• Microsoft Windows 98, Second Edition (98SE), Windows 2000 Desktop or Windows XP
Note1: EVB90614 is HID USB device and will require no drivers on PCs that support HID class USB
devices (except Microsoft Windows 98, Second Edition (98SE)).
Check www.melexis.com for most recent release of the software.
3 Installing the software
As an USB device, the demonstration board requires very little effort to install.
Most of the work is done by the operating system. The software is installed simply by
running the file “setup.exe”.
The software will be installed in a separate Melexis directory and can be accessed from
Start/Programs/Melexis/ Mlx90614 Configurator.
Note2: If your version of Windows does not support the HID USB devices you might need to install USB
driver (check www.melexis.com for such driver).
390129061401
Rev 003
Page 3 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
4 Getting started with MLX90614 evaluation board
4.1 MLX90614 evaluation board overview
1
2
3
4
5
6
7
Fig.4.1
Top view of evaluation board is shown on Fig.4.1.The main elements include:
1. Extension connector (70553-0003 from www.molex.com) allowing additional
MLX90614 modules to be connected to SMBus.
Connector pin description:
pin1 - SCL/Vz (pin 1 is marked on the plastic body of the connector)
pin2 - PWM/SDA
pin3 - Vdd
pin4 - Vss
2. ZIF socket for MLX90614
3. Button “next”
4. Button “start”
5. Button “reset”
6. RGB status LED
7. USB "B" Receptacles
The EVB90614 board receives its power supply only from the USB cable (Bus-Powered
Device). External power supply source is not needed.
390129061401
Rev 003
Page 4 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
4.2 Connecting the EVB90614 evaluation board
To connect the evaluation board:
1. Unbox and unwrap the board, and set it on a non-conductive surface near the
host system.
2. Connect the USB cable (supplied in the kit) to an open USB port on the host
system or a USB hub connected to the host system, and to the USB connector on the
board. The LED will shine in BLUE.
3. EVB90614 evaluation board is HID compliant device so a special USB device
driver is not needed (except when using Window 98 & 98SE).
4. Check the board connection. This can be done from Start/Settings/Control Panel/
System/Hardware/Device Manager (Fig. 4.2).
Fig. 4.2
5 Getting started with MLX90614 Configurator
5.1 MLX90614 Configurator overview
The MLX90614 Configurator is a Windows-based software application designed to
be used with the EVB90614 evaluation board for evaluating Melexis MLX90614 infrared
thermometers. Using this software, features and performance of MLX90614 infrared
thermometers can be evaluated. All basic configurations of MLX90614 are also supported
in user-friendly format. Configuration is written in EEPROM of each MLX90614 and will be
the POR default afterwards. This way the MLX90614 can be further used without EVB in
target applications.
5.2 Working with MLX90614 Configurator
Insert a MLX90614 thermometer in the ZIF socket (see Fig. 4.1). Make sure that
module benchmark coincides with the mark on the PCB.
The software can be launched from Start/Programs/Melexis/ MLX90614 Configurator.
390129061401
Rev 003
Page 5 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
The Configurator automatically recognizes the module type, address and voltage supply
(Fig. 5.1). The status LED is shinning GREEN.
Fig. 5.1
If there is no thermometer module inserted on the EVB or the module is not a MLX90614,
the user will be notified (Fig. 5.2). Click the button OK. The main panel looks as shown in
Fig. 5.3.
Fig. 5.2
Fig. 5.3
IF a module is put incorrectly the user will be notified (Fig. 5.4).
Fig. 5.4
Note that the buttons Power off and Refresh are alternating.
The MLX90614 Configurator gives the opportunity to connect a few MLX90614 modules in
a network (Fig.5.5). Note that if in a network both 3V (MLX90614Bxx) and 5V
(MLX90614Axx) modules are present, currently only the 3V modules will be available.
390129061401
Rev 003
Page 6 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
Fig. 5.5
Note3: When more than one module is connected on the bus an error message (see Fig. 5.4) will be shown.
Note that this doesn’t mean mandatorily that there is incorrectly put module but the user is encourages
carefully to check the modules before click button Yes.
Note4: If in a SMBus network there is a module with address zero it will be not reported. A module with
SMBus address zero will be reported if it is the only one connected to board. Also modules with the same
SMBus addresses connected to the network will cause conflict on the bus when operation reading is done.
Note5: It is strongly recommended to save factory default EEPROM of every new MLX90614 before
doing anything else with it (See below how to do that).
Note6: schematics for PCBs to attach multiple modules using connector 1 can be found on page 20.
Main panel description
Turn off module power supply.
(Re-)start system to refresh information about the modules.
Opens “Console Utility” window for manual sending of commands.
See Appendix A for commands set.
Opens “Configuration uploader” window for quick replication of
MLX90614 infrared modules’ configuration.
Opens “Advanced” window from which user can upgrade EVB firmware,
change measuring unit of the temperature scale and relocate the log file.
Opens current EVB90614 User Manual. It is recommended to check
www.melexis.com for updates and related documents.
390129061401
Rev 003
Page 7 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
Opens “Configuration utility” window used for configuring MLX90614
infrared thermometers.
Opens “Measure utility” window for temperature measurements.
Dual zone module identification picture.
Single zone module identification picture.
Module type and SMBus slave address identification fields.
Show which EVB is active (more than one EVB can be connected to
the one PC).
5.3 Configuration of MLX90614 modules
From the main panel push the button Config. ”Configuration utility” window is
opened (Fig. 5.6).
Fig. 5.6
Fig. 5.7
From this window the user can configure the MLX90614 module. A short description
of every setting field follows bellow. For a detailed description of the settings please refer to
the MLX90614 datasheet. These settings are stored in MLX90614 EEPROM and will be
the power-up default configuration.
●To max, To min - defines the temperature ranges for object temperature in PWM mode.
390129061401
Rev 003
Page 8 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
●Ta max, Ta min - defines temperature ranges for ambient temperature in PWM mode.
●PWM/SMBus/Relay - chooses between SMBus, PWM and Thermo Relay mode.
●PWM mode - defines single or extended PWM mode
●PWM struct - selects what data will be transmitted via the PWM.
●PWM period - defines the PWM period. A list of available settings as drop-down menu.
●PWM period repetitions - defines how many times each PWM period will be repeated.
Any even number of repetitions from 0 to 62 is valid.
●PWM pin configuration - chooses between NMOS “open drain” and “push pull” PWM
output.
●Sensor zone - defines the number IR sensors. This number is factory programmed and is
recommended not to be altered.
●SMBus address - set SMBus slave address. All values between 0 and 127 are valid. All
MLX90614 will accept communication with slave address 0. Therefore in SMBus network a
slave address 0 will be useless.
●F.I.R filter- drop-down list of settling time available via embedded in MLX90614 digital
filter. Slower settling comes with less noise.
●I.I.R filter [%] – drop-down list of available values for impulse limit. 100% passes all
impulse, 50% cuts impulse magnitude in half via embedded in MLX90614 IIR filter. Note
that the settling time depends on both FIR and IIR filters settings.
Some options are available from the drop-down menu while other must be typed by the
user. For example to change the object temperature range double click on To min or To
max field and type the new range.
To change between PWM and SMBus mode double click on PWM/SMBus/Relay field and
select from the drop-down menu the desired option (Fig.5.8).
Fig. 5.8
Note that when the Thermo relay option is selected the fields To max and To min change
in Hysteresis and Threshold respectively (Fig.5.7). Hysteresis adds to threshold on each
side. Trip points will be Threshold+Hysteresis for rising temperature and
Threshold-Hysteresis for decreasing temperature.
Tick
to view the full EEPROM memory in hexadecimal view (Fig.5.9).
Note that only highlighted cells are accessible by the user (see MLX90614 datasheet for
more information). All other EEPROM cells contain factory calibration values and are
protected against erase and writes.
390129061401
Rev 003
Page 9 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
Fig. 5.9
Note 7: Do not write SMBus address 0x00 in any module when you make a sensor network because reading
this module will cause all devices on the network to reply. This will result in confusion on the SMBus.
Click button Read to read the complete EEPROM memory. Starting the SW or clicking
“Refresh” after “Power Off” will automatically read the EEPROM, too.
Click button Write to write changed EEPROM cells. Factory calibration values will not be
altered. Writing EEPROM cell is automatically preceded by erasure.
Click button Save file to save EEPROM in a bin file. Refer to Fig.5.10.
Click button Load file to load a bin file. If the loaded bin file is to be written into MLX90614
EEPROM a “Write” must be clicked afterwards. Refer to Fig.5.11.
Fig. 5.10
390129061401
Rev 003
Fig. 5.11
Page 10 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
5.4 Manual sending commands
From the main panel push the button Console. ”Console utility” window is opened
(Fig.5.12).
Information field
Command field
Fig. 5.12
Type a valid command in the Command field, push Execute (or button ‘Enter’ on the
keyboard). The result will be reflected in the Information field. The supported commands
are described in APPENDIX A. The console utility is provided for extended functionality. All
read/configure/write as well as measure/log data operations can be done without this utility.
390129061401
Rev 003
Page 11 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
5.5 Demonstration of SMBus and PWM temperature measuring
From the main panel push the button Measure. The ”Measure utility” window is
opened (Fig.5.13).
Fig. 5.13
Note8: If the used module is single zone only two thermometers will be present on the screen (Fig.5.14).
Fig. 5.14
Push the button Start to start temperature measurement. With the MLX90614 configured in
PWM mode the PWM output depends on the ranges written in EEPROM. These ranges
390129061401
Rev 003
Page 12 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
are automatically read and filled in the Min/Max boxes on this screen. An example is
shown on Fig.5.15.
Fig. 5.15
In the field “Measure period” the user can make the measuring faster or slower.
Max and Min fields permit user to change graphic scales.
When the measurement is started and Do log field is ticked data will be acquired in a
log.csv file. If the SMBus mode is selected the following RAM registers will be logged in the
log.csv file (See MLX90614 datasheet for more information):
● Ambient sensor data - SMBus RAM address=0x03h
● IR sensor 1 data - SMBus RAM address= 0x04h
● IR sensor 2 data - SMBus RAM address= 0x05h
● Linearized ambient temperature Ta - SMBus RAM address =0x06h
●Linearized object temperature (IR1) TOBJ1 - SMBus RAM address= 0x07h
●Linearized object temperature (IR2) TOBJ2 - SMBus RAM address = 0x08h
If PWM mode is selected logged data will depend on the chosen PWM structure.
Every time a measurement is started the log file is continued, with the entire EEPROM
saved as new heading row.
Note9: In Regional and Language Options (available from Start/Settings/Control panel) list separator need
to be “,” and decimal symbol – “.” in order to get the CSV log file opened in an easy to use form in EXCEL.
Note10: Note that the CSV log file is the same for all measurements, there is not a separate file for every
module that is measured.
To change the temperature measuring unit (°C, °F or K) and to relocate the log.csv file:
390129061401
Rev 003
Page 13 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
1. Click the button Advanced from main panel
2. Click the button Options available in “Advanced” window.
3. From “Software settings” window choose the desired settings (Fig. 5.16).
Fig. 5.16
Thermo Relay option can be simulated with MLX90614 Configurator. For example let
choose Thermo Relay option in the “Config” window and set the following parameters:
Threshold=27ºC, Hysteresis=1ºC. Let open “Measure utility” window (Fig. 5.17) and press
the button Start. The Thermo Relay output should be in logical “0” if the object temperature
is bellow 26ºC how it is shown on the Fig. 5.17.When the object temperature become
above 28ºC the Thermo Relay output is switching over in logical “1” (Fig. 5.18). Min and
Max fields show min and max temperature threshold by which the Thermo Relay output will
switch over in logical “0” and “1” respectively. Note that the module is indeed in SMBus
mode by SMBus Request command and the shown below pictures show how the real
Thermo Relay output should keep itself.
Fig. 5.17
Fig. 5.18
To use EVB90614 with modules with fixed PWM output (no SMBus access is available) the
following option can be used:
1. Click the button Advanced from main panel
4. Click the button PWM-M available in “Advanced” window.
390129061401
Rev 003
Page 14 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
The window “PWM detached measurement module” is opened (Fig. 5.19).
Fig. 5.19
All settings must be entered manually by the user. In Slowdown field user can write value
between 0 and 100. Bigger value in this field will make the thermometer readings slower
and easier.
5.6 Quick customization of MLX90614 thermometers
Push the button Uploader on the main panel. The “Configuration uploader” window
is opened (Fig. 5.20).
This window is very similar to “Configuration utility” window but here the user cannot
change the EEPROM.
Click the button “Select EEPROM” to open the window “Open EEPROM file” (See Fig.
5.11). Browse to select the desired EEPROM configuration file and click the button “Open”.
The chosen EEPROM configuration is loaded (Fig. 5.21). Tick
the full EEPROM memory in hexadecimal view if desired.
Fig. 5.20
390129061401
Rev 003
to view
Fig. 5.21
Page 15 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
The user has two choices to load an EEPROM configuration in a MLX90614 module:
● Upload
● Quick upload
When the Upload button is pushed the software automatically checks the module
power supply before writing the EEPROM memory of the module. This process takes some
time. This option is useful when the user wants to configure MLX90614 modules of
different versions. It is recommended to use this option as it provides additional safety (a
3V module would not be subjected to overvoltage stress unintentionally).
When long series of MLX90614 modules need to be configured with the same
configuration the Quick upload option might be preferred. In this case the user must initially
choose the module voltage supply manually in the field below the button Quick upload
(See Fig. 5.18 or Fig. 5.19) and push the button Quick upload. In this case the software
does not make module power supply verification so the time of module configuration is
reduced. Tick SMBus increment field if you want the SMBus address field to be
incremented automatically.
Note that in both cases the users cannot enter the SMBus device address. And
indeed this is not necessary because the software automatically uses address 0x00 which
is universal address for all MLX90614 devices. The users must realize also that when
using the “Configuration uploader” window only one MLX90614 device should be present
or all devices in the network are changed. If you work in a network and want to change the
configuration of a specific module in the network use ”Configuration utility” window ( see
5.3 Configuration of MLX90614 modules).
390129061401
Rev 003
Page 16 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
5.7 Updating EVB90614 firmware
For possible future upgrades of the EVB FW a bootloader is programmed into the
EVB MCU. This makes possible for the user to load updates into the EVB using the same
software.
This can be done by two ways:
Updating EVB90614 firmware – method 1
1. Start MLX90614 Configurator
2. Push the button Advanced from the main panel. “Advanced” window is opened
5.22).
(Fig.
Fig. 5.22
3. Push the button Upload HEX. A warning message appears (Fig. 5.23).
Fig. 5.23
4. Click Yes and in a new browse window to find the firmware file will appear, select the
correct file and click button Open (Fig. 5.24).
Fig. 5.24
390129061401
Rev 003
Page 17 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
Wait until a message box appears (Fig. 5.25).
Fig. 5.25
Close this window. The EVB90614 board is updated and ready for work.
If an incompatible file is loaded warning messages will appear.
Follow these messages to load a compatible file.
Updating EVB90614 firmware – method 2
1. Close MLX90614 Configurator
2. Press and hold the button start on the board
3. Press and release the button reset on the board
4. Release the button start
After these steps the board is manually entered in bootloader mode. The LED is turned off.
5. Start MLX90614 Configurator. The following message appears (Fig. 5.25).
Fig. 5.25
6. Click OK and follow step 4 in method 1
390129061401
Rev 003
Page 18 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
6 Troubleshooting
If the USB communication with EVB90614 fails or the MLX90614 Configurator
hangs, close the program, push the button reset on the board (the LED must shines BLUE)
and launch the program again.
390129061401
Rev 003
Page 19 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
7 Disclaimer
Devices sold by Melexis are covered by the warranty and patent indemnification
provisions appearing in its Term of Sale. Melexis makes no warranty, express, statutory,
implied, or by description regarding the information set forth herein or regarding the
freedom of the described devices from patent infringement. Melexis reserves the right to
change specifications and prices at any time and without notice. Therefore, prior to
designing this product into a system, it is necessary to check with Melexis for current
information. This product is intended for use in normal commercial applications.
Applications requiring extended temperature range, unusual environmental requirements,
or high reliability applications, such as military, medical life-support or life-sustaining
equipment are specifically not recommended without additional processing by Melexis for
each application.
The information furnished by Melexis is believed to be correct and accurate. However,
Melexis shall not be liable to recipient or any third party for any damages, including but not
limited to personal injury, property damage, loss of profits, loss of use, interrupt of business
or indirect, special incidental or consequential damages, of any kind, in connection with or
arising out of the furnishing, performance or use of the technical data herein. No obligation
or liability to recipient or any third party shall arise or flow out of Melexis’ rendering of
technical or other services.
© 2006 Melexis NV. All rights reserved.
For the latest version of this document, go to our website at
www.melexis.com
Or for additional information contact Melexis Direct:
Europe, Africa, Asia:
Phone: +32 1367 0495
E-mail: [email protected]
America:
Phone: +1 603 223 2362
E-mail: [email protected]
ISO/TS 16949 and ISO14001 Certified
390129061401
Rev 003
Page 20 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
♦APPENDIX A – EVB90614 commands set
1. Read MLX90614 RAM address
Syntax:
Operands:
rr
address
0(0x0) ≤ address ≤31(0x1F)
2. Read MLX90614 EEPROM address
Syntax:
Operands:
re
address
0(0x0) ≤ address ≤31(0x1F)
3. Write MLX90614 EEPROM address
Syntax:
Operands:
we address value
0(0x0) ≤ address ≤31(0x1F)
0(0x0) ≤
value ≤65535(0xFFFF)
4. Restart MLX90614 (turn off-turn on module power supply)
Syntax:
Operands:
rt
no operands
5. Send Request (switch to SMBus mode)
Syntax:
Operands:
rq
no operands
6. Capture PWM
Syntax:
Operands:
390129061401
Rev 003
cp
no operands
Page 21 of 22
User Manual
Aug/2007
USER MANUAL
MLX90614 Evaluation board
♦ APPENDIX B – EVB90614 schematics
Note that the components marked “Optional” are not populated. They are assumed to give
the user the option to develop his own applications with the board.
ICSP
J1
R33
(Vss)
PGD
Vpp
PGC
6
5
3
4
2
4
2
2
4
3
1
1
3
4
RB4
optional
1ks R10
RA2
RA1
6
12
ICCK"
4
1,8ks R19
3
2
2ks
R21
1ks
optional
R22
C13
1
820s R23
optional
U8
+
22/10TBs
TP1
Tx
22ns
optional
4k7s R25
.
optional
SN74LVC2G14DBVR
RC5/D+
RC4/D-
RC6/Tx
44
43
42
1ks
R24
optional
LED1
LM1-TPP1-01
13
ICDT"
RB4/KBI0
14
15
RB5/KBI1/PGM
16
17
RB7/KBI3/PGD
-MCLR/Vpp/RE3
18
20
19
RA0/AN0
RB6/KBI2/PGC
RD2
RD3
1ks R32 41
R26
+ C9
2,2ks R18
optional
RD1
1
2
3
5
RC7/Rx
1ks R31 40
ICPORTS"
22/10TBs
C8
22ns
6
RD4
CRST"
34
6
RA1/AN1
RC0/T1OSO
TP2
Rx
U SB_D -
65
U SB_D +
5
1
-CS
2
Vdd
optional
0,1s
C28
8
optional
0s
R41 optional
RD4
6
5
RD3
4
RD2
3
RD1
RD0
2
1
Vss
Optional
3
M3mm
SO
M4
M3mm
-Hold
M3
M3mm
7
4
M2
M3mm
Vss
22/10TBs
optional
-WP
M1
+ C20
SCK
22/10TBs
optional
SI
+ C19
6
22/10TBs
optional
U13
+ C18
25LC640
.
1
2
3
4
5
7
.
5V
DD+
GND
miniUSB
5
8
0s
7
.
CN6
.
9
USB-B
0s
RA2/AN2/Vref-
RD5/P1B
+ C16
.
5V
DD+
GND
1
2
3
4
RA3
Vdd
RD6/P1C
OSC2/CLKO/RA6
FB1
600Z
.
CN1
21
OSC1/CLKI
4V8
C15
1,0s
8
RD7/P1D
RD0
33
9
Vss
Vss
1ks R30 39
47*s
R20 32
U6
PIC18F4550
Vdd
Out
GND
In
U7
LM2931
C14
3
1
Y1
20MHz31
22ps
C12
Vdd
1ks R29 38
30
RE2/AN7
Vusb
C11
22ps
RB0/INT0/SDA
R
29
RE1/AN6
37
28
RB1/INT1/SCL
B G
C10
22ns
RE0/AN5
.
330ks R17
optional
10
RB2/INT2
0,22s C17
R16
11
RB3/CCP2'
RA5/AN4/-SS/HLVDIN
RC2/CCP1/P1A
26
27
1,0s
3
RA3/AN3/Vref+
25
R14
RA4/T0CKI
RC1/T1OSI/CCP2'
12ks
R13
10ks R28 36
33ks
optional
1ks R84
optional
1ks R9 5
Vsup
24
62ks
2
1
optional
1ks R7
23
1ks
R15
22
4
SMB2
U3
optional
NC
GND
2
C7
0,1s
M LX9 06 14
R11
1ks
3
5
SMB1
1
Vcc
(LTC1694IS5)s
U4
6
1
PESD5V0L5UY
.
SCL
.
Vss
.
10ns
C6
U2
10ks R27 35
R12
Vdd
.
4
9
33ks
optional
R6
1,5ks
5
4
On/OFF ADJ
R3
0s
SCL
7
2
3
8
C4
NFM21P C105
GND
3
.
PWM
+
2
5
Out
Q2
NDS332P
18ks R4
Q1
NDS332P
18ks R2
SCL/Vz
PWM/SDA
1
2
Vdd
Vss
3
4
R5
1,5ks
R1
J2
PWM
C3
22/10TBs
U5
In
Optional
Vdd
+
LP2980-ADJ
1
P 12302SCT2
S1
5
R43
33ks
10ns
WM4902-ND J3
C5
1,0s
S1JP 1 SMD
5
R42
33ks
C2
Vdd
2
33ks
PWM
C21
22/10TBs
optional
P 12302SCT1
P 12294SCT1
S1
5
1
CN4
SCL
22ps
C1
TP3
Vss
1
2
3
4
5
1
TP4
SCL/Vz
1
1
TP5
PWM/SDA
SCL/Vz
PWM/SDA
Vdd
Vss
Vss
Alternative
TP6
Vdd
1
7
6
UX40_MB_5P
.
.
.
.
9
8
Vss
Vdd
optional
J4
Note: If user decides to add own firmware using EVB90614 he must use program memory above 0x4000
(see PIC18F4550 datasheet). Also EEPROM memory addresses between 0x40 and 0x43 must be
maintained 0x00 or 0xFF otherwise the EVB90614 will not be recognized from the host when it is in
bootloader mode. For the latest information about this opportunity check www.melexis.com .
390129061401
Rev 003
Page 22 of 22
User Manual
Aug/2007