XRpad2 3025 Product Note

P ARP O
P LDI CUA TC I TO N NN O
O TTE E
XRpad2 3025 (In Development)
Flat Panel X-ray Detectors
Features and Benefits
•
•
•
•
•
•
•
•
•
•
•
25 cm × 30 cm (10” × 12”) image
High resolution 100 μm pixel pitch (5.0 lp/mm)
Direct deposition Csl, for excellent image quality
Up to 65,536 grey levels (16-bit ADC)
Automatic Exposure Detection (AED)
Wi-Fi interface (Station and Access Point mode)
Docking connector for GigE, power and sync
On-board pixel corrections and storage
Sequence mode for tomosynthesis
Fast preview image
Robust and lightweight design
Applications1
• Digital radiography
Upgrade to Digital
Radiography
Overview
XRpad2 3025 is a wireless, lightweight, cassette-sized
flat panel detector for digital radiography based on a
second-generation XRpad® design from PerkinElmer.
Featuring a 7.4 million pixel image matrix, a best-in-class 100 μm pixel pitch, and a direct
deposited CsI scintillator, the PerkinElmer XRpad2 3025 provides exceptional image quality.
A 25 × 30 cm2 imaging area is provided, along with a removable rechargeable battery. A robust
and lightweight aluminum frame with carbon-fiber entrance window and back cover allow
for easier placement and cleaning. Automatic Exposure Detection, on-board corrections, and
wireless access point mode simplify integration.
We have over 20 years of experience partnering with customers to develop products in a wide
range of X-ray applications. Let our digital imaging experience work for you.
Unless otherwise specified, PerkinElmer Flat Panel X-ray Detectors are components intended to be integrated into products
by X-ray system manufacturers. System manufacturers are responsible for qualifying and validating their products for their
intended uses and meeting all applicable regulatory requirements.
1
Contents in this document are subject to change without notice.
XRpad2 3025 (In Development)
Imaging Performance
Sensor
Panel
Amorphous silicon active TFT-diode array
Limiting Resolution
5 cy/mm
Scintillator
Direct deposition CsI:Tl
Typical MTF
70% (1 cy/mm), 40% (2 cy/mm),15% (4 cy/mm) for RQA5
Pixel Matrix
2508 × 3004
Typical DQE
75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5
Pixel Pitch
100 µm
Advanced Features
Electronics
Sequence Mode
10 fps (2 × 2 binned)
Amplifiers
Low noise ASICs with user selectable gains
On-board Corrections
Offset, gain and defect pixel
ADC
16-bit
On-board Storage
Image storage with tagging
Image Transfer Time
Wired: 300 ms; Wireless: 2000 ms
Fast Preview
4 × 4 binned quick preview image
On-board Memory
1 GB DDR3, 8 GB SDHC card
Environmental
Mechanical
Temperature
10 °C to 35 °C operating
Size
ISO 4090 for 25 cm × 30 cm (10” × 12”) cassette size
Humidity
10% to 85% operating
Active Area
248 mm × 298 mm
Ingress Protection
IPX4 rated (protection against splashing water)
External Dimensions
282 mm (w) × 332 mm (l) × 15.5 mm (h)
Accessories
Weight
1.8 kg (4 lbs)
Battery
Rechargeable battery, 11.1 V
Housing
Aluminum frame with carbon-fiber entrance window
Battery Charger
External two bay charger 100 - 240 V AC 50/60 Hz
Communications
8
7
6 Interface and
Power Unit
Status Display
OLED display with Wi-Fi, LAN, battery, and sensor indicators
Wireless Data I/F
802.11n Wi-Fi standard at 5 GHz
Regulatory
Wired Data I/F
GigE, trigger and power via optional docking connector
Standards
X-ray I/F
Integrated X-ray trigger control
Automatic Exposure Detection
D
Optional
IPU-2 external power supply
5
4 100 – 240 V AC,
GigE and X-ray I/F
ECO
3
REV
EN IEC 60601-1:2006/AC:2010, EN IEC 60601-1-2:2007,
FCC Part 15 subpart B/E, ETSI EN 301 893 V.1.7.1 (2012),
ETSI EN 301 489-1 V1.9.2 (2011), ETSI EN 301 489-17 V2.2.1
(2012), EN ISO 10993-5:2009, EN ISO 10993-10:2013
Mechanical Characteristics (Dimensions in mm)
15.5
282
C
B
332
T
A
MATERIAL AND FINISH REQUIREMENTS:
USA
PerkinElmer, Inc.
2175 Mission College Blvd
Santa Clara, CA 95054 USA
P: +1-844-726-8228
F: +1-408-969-6493
Germany
United Kingdom
PerkinElmer
Dexela Limited
Technologies GmbH & Co. KG
A PerkinElmer Company
PRINTED
FORM
OR TEMPLATE
IS NOT
In
der
Rehbach
22
Wenlock
Business
Centre
37389 VERSION 02
GUARANTEED TO BE THE CURRENT
FRM DRAWING TEMPLATE
65396
Walluf
Germany
50-52WITHIN
WharfTHE
Road
VERSION.
VERIFY
DOCUMENT
C/N 500011504
CONTROL CENTER THE CURRENT VERSION
P: +49-6123-971-300
London N1 7EU United Kingdom
PRIOR TO USE.
F: +49-6123-971-600
P: +44-20-7148-3107
China
PerkinElmer, Inc.
No. 1670 Zhangheng Road
DRAWN
Zhangjiang Hi-Tech Park DATE
Shanghai 201203, PRC
ENGINEER
P: +86-21-60645611 DATE
F: +86-21-60645666
PROPRIETARY AND CONFIDENTIAL
THE INFORMATION CONTAINED IN THIS DRAWING IS THE
SOLE PROPERTY OF PERKINELMER MEDICAL IMAGING.
ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT
THE WRITTEN PERMISSION OF PERKINELMER MEDICAL
IMAGING IS PROHIBITED.
8
7 us through the contact form
6 provided or email [email protected].
5
Visit our website at www.perkinelmer.com/imaging-components/
and contact
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2012-2016, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
011428B_01
PKI
4
UNLESS OTHERWISE SPECIFIED
DIMENSIONS ARE: MM[INCHES]
TOLERANCES ARE:
.X = 0.25[.01]
.XX = 0.13[.005]
ANGLES = 1
DO NOT SCALE DRAWING
3