Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

a p p l i c at i o n N o t e
Gas Chromatography/
Mass Spectrometry
Authors
Ruben Garnica
Dawn May
PerkinElmer, Inc.
Shelton, CT USA
Method 8260C by
Purge and Trap Gas
Chromatography
Mass Spectrometry
using the Clarus SQ 8
Introduction
U.S. EPA Method 8260C – Volatile
Organic Compounds (VOCs) by Gas
Chromatography Mass Spectrometry
(GC/MS) is one of the most common
environmental applications for GC/MS.
This method outlines the analysis of
volatile organic compounds in a variety
of solid waste matrices including various air sampling trapping media, ground
and surface water, soils, and sediments
among others. The method requires not
only demonstration of laboratory sample preparation and handling competence
but instrument performance as well. The study presented here demonstrates
the PerkinElmer® Clarus® SQ 8 GC/MS with purge and trap sample introduction
both meets and exceeds the performance criteria set out in method 8260C and
describes the analytical results and instrumental methodology.
Experimental
The PerkinElmer Clarus SQ 8C GC/MS operating in electron ionization mode
with an Atomx purge and trap sample introduction system (Teledyne Tekmar,
Mason, OH) was used to perform these experiments. The purge and trap
conditions are presented in Table 1 and represent standard conditions for
the analysis of method of VOCs by EPA Method 8260C.
Table 1. Purge and Trap Instrument Conditions.
Purge and Trap System:
Tekmar – Atomx
Trap
Tekmar #9 trap
Sample Size
5 mL
Purge Parameters:
Valve Oven Temp
140 °C
Transfer Line Temp
140 °C
Table 2. Clarus SQ 8C GC/MS conditions.
Sample Mount Temp
90 °C
Gas Chromatograph: PerkinElmer Clarus 680
Water Heater Temp
90 °C
Sample Vial Temp
20 °C
Analytical Column:
Sample Equilibrate Time
0.00 min
Injector Temperature: 220 °C
Standby Flow
10 mL/min
Carrier Gas:
Helium @ 1.0 mL/min
Purge Ready Temp
40 °C
Split Flow:
80 mL/min
Condensate Ready Temp
45 °C
Oven Program:
Temperature Hold Time
Presweep Time
0.25 min
Prime Sample Fill Volume
3.0 mL
Sweep Sample Time
Rate
2 min
10 °C/min
100 °C
0 min
30 °C/min
0.25 min
200 °C
4 min
End
Sweep Sample Flow
100 mL/min
Mass Spectrometer: PerkinElmer Clarus SQ 8C
Sparge Vessel Heater
On
Purge Time
11.00 min
GC Transfer Line
Temperature:
Purge Flow
40 mL/min
Purge Temp
40 °C
Condensate Purge Temp
20 °C
Dry Purge Time
2.00 min
Dry Purge Flow
100 mL/min
Dry Purge Temp
20 °C
Water Needle Rinse Volume
7.0 mL
Sweep Needle Time
0.50 min
Desorb Preheat Temp
200 °C
GC Start Signal
Start of Desorb
Desorb Time
2.00 min
Drain Flow
300 mL/min
Desorb Temp
200 °C
Bake Parameters:
2
Elite 624 MS
(30 m x 0.25 mm i.d. x 1.4 μm)
40 °C
Desorb Parameters:
The Clarus SQ 8C GC/MS conditions are presented in Table 2.
The heated transfer line of the concentrator was connected
directly to the injector port of the Clarus 680 GC. A 2 mm
i.d. injector port liner was used inside the capillary injector.
Split flow adjustments will increase or decrease the split
ratio and the resulting sensitivity of the system and allow
the analysts flexibility in optimizing their GC/MS system.
Number of Water Bake Rinses
2
Water Bake Rinse Volume
7.0 mL
Bake Rinse Sweep Time
0.25 min
Bake Rinse Sweep Flow
100 mL/min
Bake Rinse Drain Time
0.40 min
Bake Time
6.00 min
Bake Flow
250 mL/min
Bake Temp
270 °C
Condensate Bake Temp
200 °C
220 °C
Ion Source Temperature:300 °C
Function Type:
Full Scan
Solvent Delay:
0 – 0.50 min
Scan Range:
m/z 35 – 270
Scan Time:
0.20 sec
Interscan Delay:
0.10 sec
Calibration and performance standards were prepared
from commercially available stock standards and diluted in
Class-A volumetric flasks. Calibration standards were made
from a combination of four multi-component standard mixes
including a 76 component 8260 calibration mix, a 6 component 502.2 calibration gas mix, a 3 component VOA surrogate
spike mix, and an 8260 internal standard mix. Stock standards were diluted to an intermediate concentration from
which initial calibration standards were prepared. The calibration levels used in this study are presented in Table 3.
Standards used for the Method Detection Limit (MDL) and
precision experiments were prepared at appropriate concentration levels from intermediate standards. The internal
standard mix was added to all samples such that a resulting
concentration of 20 μg/L was achieved per 5 mL of sample.
The tuning standard, 4-bromofluorobenzene, was prepared
at 50 μg/L. Standards, stock standards, and associated QC/
QA samples were stored in a manner consistent with the
guidelines set out in the method.
Table 3. Calibration points employed in this study.
Calibration Level
Concentration (μg/L)
10.5
21.0
32.0
45.0
510
625
750
8100
9200
Results and Discussion
The GC conditions were optimized to provide adequate
separation while keeping the analytical runtime as short
as possible. Figure 1 demonstrates a mid-level (25 μg/L)
total ion chromatogram (TIC) from this analysis. All of the
expected analyte separation is achieved while allowing for
a maximum of throughput per unit laboratory time. The
injection-to-injection time is further enhanced by the performance of the Clarus 680 GC oven. The oven incorporates
a dual walled design with a unique air flow path to allow
the fastest cooling times of any commercially available GC
oven on the market. Air flow through the Clarus 680 GC
is unidirectional eliminating the mixing of heated exhaust
and cool ambient air allowing for the fastest transport of
heat out of the GC oven. This high speed cooling allows the
user to dramatically decrease the idle time of the instrument
between analytical runs and in this study the injection-toinjection time, measured to be under 30 minutes, was
limited by the purge and trap system.
Table 4 presents the analytical results of these experiments
including the Retention Time (RT), Average Relative Response
Factors (Avg RRF) and Percent Relative Standard Deviation
(% RSD) of the initial calibration, the Method Detection
Limits (MDL), and the method precision as Percent Recovery
(% Rec) and accuracy as % RSD. All of the analytes presented
meet or exceed the minimum method requirements and
boast excellent detection limits, precision and accuracy.
Method Detection Limits were determined by analyzing
seven replicate samples at a concentration of 0.5 μg/L and
1.0 µg/L. Precision and accuracy were measured by analyzing
four replicates at a concentration of 25 μg/L.
The majority of compounds in this report are calibrated
over the entire concentration range as presented in Table 3.
Compounds whose calibration range varied from the bulk of
analytes are highlighted in Table 5. Alternate calibration
ranges are expected for this method especially for compounds
referred to as “bad actors”. In all cases, however, EPA method
criteria are met. Overall improved results including MDL and
precision and accuracy were obtained by operating the ion
source at an elevated temperature than would regularly be
used. Figure 2 shows a time selection of the seven TICs from
the MDL study and illustrates the excellent repeatability of
the system at a low concentration level. The inset presents
the seven extracted ion chromatograms of naphthalene (EIC)
at m/z = 128. The % RSD of the seven naphthalene EICs
was measured to be 3.8%, which includes the sample
handling performed by the purge and trap.
3
Figure 1. TIC of 8260C volatile organic standard mix at 25 μg/L. Numbering references compound as listed in Table 4.
Figure 2. Zoom of the seven MDL samples. TICs overlaid to illustrate the excellent repeatability achieved at low levels. Inset shows the naphthalene EIC at
m/z = 128; the % RSD was measured to be 3.8%.
4
Table 4. Summarized initial calibration results of the analyzed VOCs from 0.5 – 200 μg/L for most compounds.
A hyphen (-) indicates value either not provided by the method or not generated in this work.
A plus sign (+) indicates calibration calculated using linear regression with R2 presented in table.
Avg
# Compound
RT
RRF
1
Dichlorodifluoromethane
+
8260C
Minimum
RRF
% RSD
8260C
% RSD
Criteria
MDL
(µg/L)
Precision Accuracy
(% Rec) (% RSD)
1.38
0.440.1 14 200.1490 2.9
2Chloromethane 1.59 0.72 0.1 0.99920.990 0.15 96
3
Vinyl chloride
1.66
0.790.1 8.2 200.2195 1.8
4Bromomethane
2.00
5
Chloroethane
2.08
0.400.1 4.9 200.2499 1.9
6Trichlorofluoromethane
2.29
7
Diethyl ether
2.58
0.37 - 6.7 200.1198 4.6
0.26
0.57
0.1
0.1
18
5.9
20
20
0.18
0.06
105
109
4.9
4.1
1.2
8
1,1,2-Trichloro-1,2,2-trifluor2.78
0.380.1 13 200.1599 1.1
91,1-Dichloroethene
2.82
10
Carbon disulfide
3.02
1.460.1 5.2 200.0593 1.4
11 Allyl chloride+
3.21 0.17
- 0.99400.990 0.05 76
2.7
12Acetonitrile 3.29 0.20
- 0.99990.990 0.38 96
5.5
13
Methylene chloride
3.37
0.510.1 16 200.2994 2.5
14 trans-1,2 Dichloroethene
3.59
15
Acrylonitrile
3.67
0.44 - 8.8 200.1497 4.6
161,1,-Dichloroethane
4.06
17
Chloroprene
4.11
0.70 - 11 200.2589 1.7
182,2,-Dichloropropane+
4.63 0.15
19cis-1,2,-Dichloroethene
4.66
0.53
0.1
9.6
20
20 Methyl acrylate
4.76
0.92
-
4.2
20
21Propionitrile
4.84
0.22
-
4.4
20
22
Tetrahydrofuran
4.92
0.53 - 13 200.1586 4.7
23Bromochloromethane
4.92
24
Methacrylonitrile
4.96
0.83 - 8.9 200.1290 4.7
25Chloroform
4.99
0.69
0.2
4.2
20
0.09
103
2.1
261,1,1-Trichloroethane
5.14
0.55
0.1
5.5
20
0.20
107
0.9
27 Pentafluorobenzene (ISTD)
5.22
28 Carbon tetrachloride
5.27
0.38
0.1
12
20
0.13
108
29
1,1-Dichloropropene
5.32
0.43 - 5.9 200.2589 1.6
30
Isobutyl alcohol
5.49
0.09 - 11.5 200.2289 5.6
31
Benzene
5.53
1.330.5 6.9 200.2290 1.6
32 1,2-dichloroethane-d4 (Surr)
5.55
33
1,2-Dichloroethane
5.64
0.420.1 9.8 200.0991 2.0
34 1,4-Difluorobenzene (ISTD)
6.00
35Trichloroethene
6.25
0.32
0.2
8.9
20
0.28
103
36
1,2-Dichloropropane
6.56
0.470.1 6 200.1596 1.8
37
Methyl methacrylate
6.67
0.39 - 6.2 200.1791 3.9
38Dibromomethane
6.70
0.19
-
4.9
20
0.10
106
2.7
39Bromodichloromethane
6.87
0.28
0.2
7.4
20
0.17
102
1.9
+
+
0.44
0.46
0.98
0.26
0.1
0.1
0.2
15
9.3
9.2
20
20
20
0.30
0.34
0.16
- 0.99960.990 -
-
4.5
20
102
105
107
1.8
1.3
2.3
56
7.4
0.24
106
1.9
0.14
100
5.3
0.12
100
5.7
0.09
107
3.6
1.2
2.5
402-Nitropropane 7.23 0.06
94
4.7
41cis-1,3-Dichloropropene
7.41
0.47
0.2
8.7
20
0.08
90
1.6
42 Toluene-d8 (Surr)
7.66
- 0.99960.990 -
5
Table 4 continued
Avg
# Compound
RT
RRF
8260C
Minimum
RRF
% RSD
8260C
% RSD
Criteria
MDL
(µg/L)
Precision Accuracy
(% Rec) (% RSD)
43Toluene
7.74
0.67
0.4
6.5
20
0.22
100
1.5
44trans-1,3-Dichloropropene
8.11
0.39
0.1
17
20
0.11
93
2.5
45 Ethyl methacrylate
8.16
0.66
-
11
20
0.07
101
2.6
461,1,2-Trichloroethane
8.31
0.28
0.1
4.3
20
0.06
96
1.9
47Tetrachloroethene
8.35
0.43
0.2
20
20
0.17
123
4.6
481,3-Dichloropropane
8.50
0.65
-
4.4
20
0.06
94
3.1
49Chlorodibromomethane
8.69
0.28
0.1
12
20
0.10
108
3.7
50 Ethylene dibromide
8.81
0.40
0.1
3.6
20
0.07
111
2.8
51 Chlorobenzene-d5 (ISTD)
9.21
52Chlorobenzene
9.24
1.08
0.5
4.7
20
0.23
102
1.0
53Ethylbenzene
9.31
1.34
0.1
2.2
20
0.20
104
1.4
541,1,1,2-Tetrachloroethane
9.31
0.27
-
7.8
20
0.19
105
1.9
55 p- & m-Xylene
9.41
1.09
0.1
4.4
20
0.23
101
1.6
56o-Xylene
9.73
1.15
0.1
2.5
20
0.19
103
2.2
3.8
20
0.18
57Styrene
9.76
1.09
0.3
58Bromoform+
9.91
0.22
0.1
10.01
1.61
0.1
59Isopropylbenzene
0.99960.990 0.10
2.8
20
0.18
106
1.8
104
2.8
106
1.8
+
60cis-1,4-Dichloro-2-butene 10.120.22 - 0.9986
0.9900.09 68
6
61 Bromofluorobenzene (Surr)
10.15
62Bromobenzene
10.25
631,1,2,2-Tetrachloroethane
10.28
64n-Propylbenzene
651,2,3-Trichloropropane
2.2
0.98
-
8.5
20
0.15
98
2.3
1.17
0.3
5.9
20
0.16
93
9.9
10.30
3.21
-
4.8
20
0.20
101
1.1
10.32
1.76
-
5.8
20
0.10
91
3.3
0.9900.05 71
66trans-1,4-Dichloro-2-butene+10.330.25 - 0.9991
2.2
672-Chlorotoluene
10.38
1.88
-
4.4
20
0.19
102
0.9
681,3,5-Trimethylbenzene
10.43
3.00
-
6.3
20
0.21
104
1.1
694-Chlorotoluene
10.47
1.99
-
7.6
20
0.15
100
3.2
70tert-Butylbenzene
10.64
2.57
-
9.2
20
0.18
94
1.3
711,2,4-Trimethylbenzene
10.69
2.94
-
4.8
20
0.19
105
1.2
72sec-Butylbenzene
10.79
3.48
-
5.1
20
0.20
107
1.7
734-Isopropyltoluene
10.88
2.98
-
4.5
20
0.22
97
1.0
741,3-Dichlorobenzene
10.89
1.64
0.6
9
20
0.22
98
0.7
75 1,4-Dichlorobenzene-d4 (ISTD)
10.93
761,4-Dichlorobenzene
10.95
1.65
0.5
9.4
20
0.18
97
1.4
77n-Butylbenzene
11.14
2.10
-
6.5
20
0.17
98
1.2
781,2-Dichlorobenzene
11.19
1.60
0.4
4.5
20
0.12
102
1.5
791,2-Dibromo-3-chloropropane
11.69
0.33
0.05
9.8
20
0.08
80Nitrobenzene+
11.84 0.06
811,2,4-Trichlorobenzene
12.24
1.19
0.2
9.3
20
82Hexachlorobutadiene
12.32
0.50
-
16
20
83Naphthalene
12.46
3.60
-
7.3
841,2,3-Trichlorobenzene
12.65
1.13
0.2
7.2
-
98
3.6
102
5.4
0.16
92
1.3
0.23
91
1.3
20
0.06
102
2.4
20
0.09
94
1.6
0.99990.990
-
Table 5. Alternate calibration range compounds.
Calibration Concentration Range Compound
0.5 – 100 μg/L
Tetrachloroethene
1.0 – 200 μg/L
Bromomethane
Allyl chloride
Acetonitrile
Isobutyl alcohol
2.0 – 200 μg/L
2,2-Dichloropropane
2-Nitropropane
Nitrobenzene
1.0 – 400 µg/L
p- & m-Xylene
Design advancements in the Clarus SQ 8 mass spectrometer
allow such a wide operating range with the Clarifi™ detector
providing much of the advantage. The detector’s enhanced
sensitivity enables full scan mass spectrometry methods to
be performed resulting in library searchable spectra at the
lowest possible concentration levels. With the added sensitivity
of SIFI™ (selected ion full ion) or SIM (selected ion monitoring)
data acquisition at even lower limits are possible for specific
compounds.
The voltage setting on the Clarifi detector was important in
developing the method for such a wide concentration range
and deserves mention here. The UltraTune™ (Standard –
DFTPP/BFB) function built into the TurboMass™ v6.0 software
is designed to set the voltage such that the m/z = 69 peak
of the tune gas FC43 is approximately 80% in the tune window.
This setting is appropriate for general instrument analyses
however users are encouraged to fine tune this setting
to suit their specific analytical needs. In our case this was
achieved by analyzing two samples, the high and low
concentration level of our calibration, at voltage settings
below that set during UltraTune. The optimum voltage
setting produced measurable signal at the lowest level while
not saturating the detector at the highest level. In our case
the initial UltraTune setting was 1570V with test measurements performed at -50V and -100V of this value. The
analysis was performed at 1520V. The stability of the Clarus
SQ 8 system allows this setting to be used for extended
periods of time and need only be reproduced periodically.
The UltraTune (Standard – DFTPP/BFB) function also produces
a satisfactory tune which meets the Tune Evaluation requirements set out in the method. Figure 3 demonstrates the
passing BFB Tune Evaluation Sample. Figure 4 shows the
mass spectrum utilized in the Tune Evaluation Test and all
of the necessary peak intensity comparisons are achieved.
Figure 3. TurboMass 6.0 EPA BFB tune evaluation results.
Figure 4. Mass spectrum of EPA BFB tune evaluation used for test in Figure 3.
Conclusion
The analysis of VOCs by purge and trap GC/MS following
EPA Method 8260C is easily achievable over a wide concentration range using the Clarus SQ 8 GC/MS system.
Satisfactory analytical results were achieved for a concentration range from 0.5 – 200 μg/L for a majority of analyte
compounds using the full scan approach providing library
searchable spectra at all concentration levels. Excellent
detection limits are reported as are accuracy and precision
measurements. A number of technological advances make
the Clarus SQ 8 GC/MS the ideal systems for laboratories
wishing to perform high throughput and sensitivity analyses
with an ease of operation currently unmatched.
PerkinElmer, Inc.
940 Winter Street
Waltham, MA 02451 USA
P: (800) 762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs
Copyright ©2012, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners.
010017B_01