Set BGA Catalog Sheet

© Siemens AG 2015
Analytical Application Sets
Analysis of biogas
Set BGA
■ Overview
■ Application
The efficiency of biogenic production processes and optimum
operation of the plant largely depends on continuous monitoring
of the biogas composition. The basic version of the Set BGA analyzes CH4 and CO2 using the proven ULTRAMAT 23 IR analyzer. The concentrations of O2 and H2S are optionally measured
using electrochemical sensors and also converted into output
signals of 4 to 20 mA. In this manner, the Set BGA contributes to
operational safety and explosion protection in addition to process optimization.
The modular design of the system takes into account the physical conditions of the gas with regard to temperature and moisture in that various gas preparation components for heating or
drying the sample gas can be configured as required.
The Set BGA (biogas analyzer) is a standardized system for stationary, continuous operation for the analysis of landfill gas, sewage gas or biogas.
■ Benefits
4
Standardized complete system
The standardized complete system has a modular configuration
and can thus be used at various measuring locations for different
measuring tasks.
• Simple and fast to configure
• Field-tested and matched Set in rugged industrial design
• Extremely high long-term stability
• The Set BGA is based on the proven ULTRAMAT 23
Field-proven, reliable technologies
• Autocalibration function with ambient air reduces the maintenance requirements
• Detonation protection in accordance with EN 12874
• Modular system design based on long-term tested components
• LEL sensor for cabinet monitoring (optional)
Simple operation
• Intuitive menu guidance
• Configuration on large displays with plain text
• Two freely-configurable limits per measured component
4/38
Siemens AP 01 · 2015
The gas mixture can be explosive at critical concentration ratios
between methane and oxygen. Even if such critical gas compositions occur extremely rarely, the danger of ignition must nevertheless be avoided. For this reason, the Set BGA is designed
with a high safety standard and even the basic version is
equipped with flow monitoring and detonation protection in accordance with EN 12874 in the sample gas path. To increase
safety even further, a gas sensor for monitoring the ambient air
can be connected as an option.
It is also possible to monitoring up to four measuring points using an optional sample switching cabinet. In this case the sample flows are sucked in continuously using a powerful pump in
order to achieve fast measuring times.
© Siemens AG 2015
Analytical Application Sets
Analysis of biogas
Set BGA
■ Design
■ Technical specifications
The Set BGA consists of the following components:
• ULTRAMAT 23 analyzer with four measured components max.
• Analyzer cabinet with modularly configurable gas preparation
components
• Cabinet for measuring point switchover (option)
• Heated line (option)
The ULTRAMAT 23 is selectable with two IR components (CO2
and CH4). Furthermore, the configuration can be equipped with
an electrochemical oxygen sensor and/or an electrochemical
hydrogen sulfide sensor. The corresponding measuring ranges
are listed in the table below.
Technical data of Set BGA
Installation
Ambient temperature
5 … 38 °C, with cabinet heating ± 0 °C
Site
Indoor/outdoor installation (configurable)
Gas inlet conditions
Sample gas pressure
• With pump, depressurized suction mode, selectable with internal or external pump
• Provision must be made for a pressure reduction for pressures greater than
1 200 mbar absolute
Pump performance
Adjustable to 60 … 80 Nl/h
Sample gas temperature
Max. 45 °C, with moisture saturation
Measured component
Smallest measuring
range
Largest measuring
range
CO2
0 … 20 %
0 … 100 %
Supply 1
200 … 240 V AC, 47 … 63 Hz
CH4
0 … 20 %
0 … 100 %
Supply 2
100 … 120 V AC, 47 … 63 Hz
O2
0…5%
0 … 25 %
Power consumption
Approx. 180 VA (without cooler and sample
preparation)
H2S (high)
H2S (low)
0 … 5 000 ppm
0 … 5 ppm
0 … 50 ppm
The ULTRAMAT 23 calibrates the IR components and the electromechanical oxygen sensor automatically with ambient air.
Calibration with calibration gas is recommended once a year or
after oxygen sensor replacement. In order to comply with the
technical specification data, the hydrogen sulfide sensor must
be calibrated every three months. An appropriate calibration
gas is therefore required. It is supplied to the analyzer through a
manually switchable ball valve.
Power supply
Connection systems
Teflon hose
With PVDF screwed glands
Connection systems
Metric (6 mm) or imperial (1/4") selectable
Dimensions
Set BGA measuring system
(W x H x D)
600 x 781 x 600 mm
Sample preparation
(W x H x D)
600 x 600 x 220 mm
4
Weight
Set BGA measuring system
Approx. 50 kg
Sample preparation
Approx. 22 kg
System design
System housing
3-part sheet-steel housing with window
Degree of protection
IP54
Cabinet conditioning
Fan
Cooling system
Peltier cyclone cooler (optional)
Sample preparation
Max. six sample streams can be controlled
using Logo module with fast loop pump in
separate housing
Analog outputs
Per component 0/2/4 … 20 mA; NAMUR,
floating, max. load 750 Ω
Measured components / measuring ranges
Set BGA measuring system
CH4
0 … 100 vol% to 0 … 20 vol% (NDIR)
CO2
0 … 100 vol% to 0 … 20 vol% (NDIR)
O2
0 … 25 vol% to 0 … 5 vol% (electrochemical
or paramagnetic optionally selectable)
H2S
0 … 5 000 ppm (electrochemical); optional
0 … 5 ppm to 0 … 50 ppm (electrochemical);
optional
Safety assemblies
Assembly 1
Detonation protection F501
Assembly 2
Flow measurement with limit monitoring at the
output
Assembly 3
LEL monitoring (optional)
Comment
• The system concept of the Set BGA is based
on the preconfigured ULTRAMAT 23 solutions (7MB2335-…, 7MB2337-…)
• The technical performance data concerning
the measuring response correspond to the
catalog data of the ULTRAMAT 23. The preconfigured version does not contain any
ULTRAMAT 23 add-ons or retrofitting sets.
2-stream sample preparation
Siemens AP 01 · 2015
4/39
© Siemens AG 2015
Analytical Application Sets
Analysis of biogas
Set BGA
■ Selection and ordering data
Set BGA basic configuration, including flame arrestor
Article No.
7MB1955- 7 7 7 7 7
- 7 7 7 7
Cannot be combined
Click on the Article No. for the online configuration in the
PIA Life Cycle Portal.
Gas connections, external
6 mm
0
¼ inch
1
Sample conditioning
Without gas cooling
A
Passive cooling (supplied separately)
B
Peltier cooler, integrated in Set BGA enclosure
C
Enclosure design
Not heated
A
Electrically heated
B
Pump design
Without pump
0
Internal pump in analyzer
1
External pump, fitted in Set BGA enclosure
2
Power supply
4
AC 120 V, 60 Hz
0
AC 230 V, 50 Hz
1
AC 110 V, 50 Hz
2
AC 220 V, 60 Hz
3
Infrared measured components
Version with one measured component
Highly selective ULTRAMAT 23 single-beam infrared gas
analyzer for measuring methane; mounted in 19-inch rack unit
for installation in wall cabinet
Specification:
• ·Measured component CH4
- Smallest measuring range: 0 ... 20 %
- Largest measuring range: 0 … 100 %
0
Version for two measured components
Highly selective ULTRAMAT 23 single-beam infrared gas analyzer for measuring carbon dioxide and methane; mounted in
19-inch rack unit for installation in wall cabinet
Specification:
• 1. Measured component CO2
- Smallest measuring range: 0 ... 20 %
- Largest measuring range: 0 … 100 %
• 2. Measured component CH4
- Smallest measuring range: 0 ... 20 %
- Largest measuring range: 0 … 100 %
1
Version with one measured component
Highly selective ULTRAMAT 23 single-beam infrared gas analyzer for measuring carbon dioxide; mounted in 19-inch rack
unit for installation in wall cabinet.
Specification:
• Measured component CO2
- Smallest measuring range: 0 ... 0.5 %
- Largest measuring range: 0 … 2.5 %
2
Oxygen measurement
Gas analyzer without oxygen sensor
A
Electrochemical oxygen sensor; sensitive to CO2
Specification:
• Smallest measuring range: 0 ... 5 %
• Largest measuring range: 0 ... 25 %
• Repeatability: approx. 0.05 % O2
B
Paramagnetic oxygen measuring cell; no sensor wear
Specification:
• Smallest measuring range: 0 ... 2 %
• Largest measuring range: 0 ... 100 %
• Repeatability: < 1 % of smallest measuring range
C
4/40
Siemens AP 01 · 2015
C
© Siemens AG 2015
Analytical Application Sets
Analysis of biogas
Set BGA
Set BGA basic configuration, including flame arrestor
7MB1955- 7 7 7 7 7
- 7 7 7 7
Cannot be combined
H2S measurement
Without H2S sensor
A
With H2S sensor, 0 … 5 000 ppm
B
B
With H2S sensor, 0 … 5 ppm to 0 … 50 ppm
D
D
Documentation
German, 1 set (paper and CD)
0
English, 1 set (paper and CD)
1
French, 1 set (paper and CD)
2
Further versions (add-ons)
Add "-Z" to Article No. and specify Order code
Heated sample gas line, self-regulating, Ex-proof
Length: 5 m, supplied separately
A01
Length: 10 m, supplied separately
A02
Length: 15 m, supplied separately
A03
Length: 20 m, supplied separately
A04
Length: 25 m, supplied separately
A05
Length: 30 m, supplied separately
A06
Length: 35 m, supplied separately
A07
Communication
PROFIBUS PA interface
A12
PROFIBUS DP interface
A13
4
Fast loop design and sample switching
2-stream sample switching with Logo and external pump
B02
3-stream sample switching with Logo and external pump
B03
4-stream sample switching with Logo and external pump
B04
5-stream sample switching with Logo and external pump
B05
6-stream sample switching with Logo and external pump
B06
Gas sensor for leak monitoring of the Set BGA system
Alarm monitoring: 20 % LEL methane
C01