Noxmat 600 gas analyzer brochure

NOxMAT 600
Product Bulletin
Features
• Measures from ppb to 3,000 ppm Full Scale (NO/Nox)
• Four User Programmable Ranges from 0-1 ppm to 0-3,000 ppm
• Fast Response Time
• Auto Ranging
• Auto Calibration
• Output Options: Analog (User Scalable), Digital (RS232) including AK
Protocol & TCP/IP
• Displays & Outputs: NO/Nox/NO2 with Adjustable Time & Hold,
Diagnostics, Alarms & Preventative Maintenance
• Remote Monitoring and Control
• Electronic Sample & Ozone Flow Control
usa.siemens.com/analyticalproducts
Description
The NOxMAT Model 600 CLD NO/NOx digital analyzer is designed
around a state-of-the-art 16-bit microprocessor, with 16 digital
inputs, 16 digital outputs, 16 analog inputs and 4 analog outputs.
The analyzer can be manually operated from the keypad or
remotely via TCP/IP, RS- 232C communications and discrete inputs.
The analyzer display includes screen presentation of all analyzer
alarms. Four levels of password protection are provided. For
precision measurements, the analyzer’s accuracy is increased by
entering calibration curve fit polynomials.
Automatic calibration may be activated locally or remotely and
includes auto cal via preset times. The analyzer may also display
NO, NOx and NO2 via selectable time and hold commands.
Method of Operation
The NOxMAT 600 CLD Analyzer utilizes the principle of
chemiluminescence for analyzing the NO or NOx concentration
within a gaseous sample. In the NO mode, the method is based
upon the chemiluminescent reaction between ozone and nitric
oxide (NO) yielding nitrogen dioxide (NO2) and oxygen. This
reaction produces light which has an intensity proportional to the
mass flow rate of NO into the reaction chamber. The light is
measured by means of a photodiode and associated amplification
electronics. In the NOx mode, NO plus NO2 is determined as
above, however, the sample is first routed through the internal
NO2 in the sample to NO. The resultant reaction is then directly
proportional to the total concentration of NOx. Local operation is
simplified using the 20 button alphanumeric keypad with data
presented on a back lit LCD display. All local operations may be
performed remote via RS-232 and/or TCP/IP.
Applications
• Stack Gases (CEM)
• Scrubber Efficiency
• Turbine/Generator Feedback Control
• Process Chemical Gas Analysis
• Personnel Safety
• Power Plant Stack De-Nitrification
• Vehicle Emissions
Specifications
Detector: Chemiluminescence (CLD) Photodiode (thermally
stabilized with Peltier Cooler)
NO/NOx RANGES: 0-1 * to 3,000 ppm NO or NOx (Four user
programmable ranges) (Higher ranges available upon request)
Response Time: T90 < 2 Seconds to 60 Seconds Adjustable
Resolution: 10 ppb NO/NOx (Displays 5 significant digits)
Repeatability: Better than 0.5% of Full Scale
Linearity: Better than 0.5% of Full Scale
Noise*: Less than 1 % of Full Scale (*1.5% @ 0-1 ppm Full Scale)
Zero & Span Drift: Less than 1 % of Full Scale per 24 Hours
Zero & Span Adjustment: Via front panel, TCP/IP or RS-232
NH3, HCN & SO2 Effect: Not detectable with 100 ppm
CO2 Effect: Less than 1 % with 10% CO2
H2O Effect: Less than 1 % with 1 % H2O
Sample Gas Pressure: 8-25 psig (Standard model)
Sample Flow Rate: 1.5 to 2.0 LPM (Consult factory for other flow
rates)
Converter: Vitreous Carbon Material @ 205.C > 98% efficiency
Ozonator: Ultraviolet Lamp
Air Or O2 Requirements: Less than 0.01 ppm NOx at 350
cc/Min. @ 25 psig (Dew Point < -35°C)
NO/NOx Control: Manual/Remote/Auto Cycle (Remote NOx mode by
dry contact closure)
Outputs Available: TCP/IP, RS232, Four Scalable Analog 0- 10 V/420 mA
Discrete Alarms: (Local & Remote Adjustable) General Fault! TTL
Logic (Ground True) Calibration Failure/ TTL Logic (Ground True)
High Concentration (2 each)/ TTL Logic (Ground True)
Digital Diagnostics: Control Voltages, Temperatures, Pressures,
Flow Parameters
Keypad Displays: Factory Settings,TCP/IP Address, Passwords (4),
Scalable Analog Output Voltages, Full Scale Range Select, Auto Cal
Times
Special Features: Calculated NO2 derived from NOx converter
efficiency, Auto Ranging, Auto Calibration (adjustable through
internal clock) Less than 3 cc Gold Plated Reaction Chamber
Display: 3” x 5” Back lit LCD
Sample Temperature: Up to 50.C Noncondensing
Ambient Temperature: 5 to 40.C
For more information, please contact:
Siemens Industry, Inc.
5980 West Sam Houston Parkway North
Suite 500
Houston, TX 77041
Phone: 713-939-7400
[email protected]
Siemens Industry, Inc.
3333 Old Milton Parkway
Alpharetta, GA 30005
1-800-964-4114
[email protected]
Ambient Humidity: Less than 90% RH Noncondensing
Warm-Up Time: 1 Hour (Typical)
Fittings: 1/4 Inch Tube
Power Requirements: 115/230 (:t100;0) VAC; 50/60 Hz;
200 Watts (350 watts with pump)
Dimensions: 5 1/4 H x 19 W x 17 D (Inches) - short case
Weight: 45 Pounds
Subject to change without prior notice
Order No.: PIAFL-00061-1115
Printed in USA
All rights reserved
© 2015 Siemens Industry, Inc.
usa.siemens.com/analyticalproducts
The information provided in this flyer contains merely general
descriptions or characteristics of performance which in case
of actual use do not always apply as described or which may
change as a result of further development of the products. An
obligation to provide the respective characteristics shall only
exist if expressly agreed in the terms of contract.
All product designations may be trademarks or product names of
Siemens AG or supplier companies whose use by third parties for
their own purposes could violate the rights of the owners.