Viscosity of ULTIMA-Flo Cocktails

Viscosity of ULTIMA-Flo Cocktails
Discussion and Recommendation
To help determine the proper pump for use in a
laboratory automation project, a researcher contacted
our laboratories and requested viscosity data for
ULTIMA-Flo™ M (PerkinElmer part number 6013579)
and ULTIMA-Flo AP (PerkinElmer part number
6013599) for various ratios with water.
Water was added to both ULTIMA-Flo M and
ULTIMA-Flo AP at ratios up to 1:1 and the viscosities
were measured at 25°C. The results are shown below:
N O T E
Problem
Cocktail
No Sample
3:1 With Water
2:1 With Water
1:1 With Water
ULTIMA-Flo AP
14 cSt
22 cSt
22 cSt
19 cSt
ULTIMA-Flo M
17 cSt
28 cSt
23 cSt
23 cSt
In the above table, cSt represents centiStokes and is equivalent to mPa.s (milliPascal seconds).