PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Unquenched Toluene Standard for Liquid Scintillation Counting Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date 6008500 Lot No. = Assay Date Average Activity Assay 286,200 0.129 112,100 0.050 DPM/Std. + 3 % Serial No. February 23, 2016 µCi DPM/Std. + 2 % µCi February 23, 2016 0 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 0 23 février 2016 0 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 1 February 23, 2016 1 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 1 23 février 2016 1 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 2 February 23, 2016 2 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 2 23 février 2016 2 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 3 February 23, 2016 3 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 3 23 février 2016 3 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 4 February 23, 2016 4 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 4 23 février 2016 4 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 5 February 23, 2016 5 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 5 23 février 2016 5 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 6 February 23, 2016 6 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 6 23 février 2016 6 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 7 February 23, 2016 7 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 7 23 février 2016 7 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 8 February 23, 2016 8 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 8 23 février 2016 8 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 9 February 23, 2016 9 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 9 23 février 2016 9 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 10 February 23, 2016 10 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 10 23 février 2016 10 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 11 February 23, 2016 11 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 11 23 février 2016 11 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 12 February 23, 2016 12 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 12 23 février 2016 12 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 13 February 23, 2016 13 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 13 23 février 2016 13 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 14 February 23, 2016 14 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 14 23 février 2016 14 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 15 February 23, 2016 15 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 15 23 février 2016 15 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 16 February 23, 2016 16 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 16 23 février 2016 16 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 17 February 23, 2016 17 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 17 23 février 2016 17 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 18 February 23, 2016 18 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 18 23 février 2016 18 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 19 February 23, 2016 19 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 19 23 février 2016 19 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 20 February 23, 2016 20 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 20 23 février 2016 20 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 21 February 23, 2016 21 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 21 23 février 2016 21 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 22 February 23, 2016 22 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 22 23 février 2016 22 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 23 February 23, 2016 23 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 23 23 février 2016 23 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 24 February 23, 2016 24 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 24 23 février 2016 24 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 25 February 23, 2016 25 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 25 23 février 2016 25 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 26 February 23, 2016 26 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 26 23 février 2016 26 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 27 February 23, 2016 27 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 27 23 février 2016 27 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 28 February 23, 2016 28 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 28 23 février 2016 28 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificate of Radioactivity / Traceability Product Part Number Radionuclide 3 H (Tritium) Toluene Carbon 14 Toluene Expiration Date Unquenched Toluene Standard for Liquid Scintillation Counting 6008500 Average Activity Assay 286,200 0.129 112,100 0.050 Lot No. = Assay Date DPM/Std. + 3 % µCi DPM/Std. + 2 % µCi Serial No. February 23, 2016 29 February 23, 2016 29 February 23, 2021 REFERENCE STANDARD: National Institute of Standards and Technology SRM 4947C (3H) and SRM 4222C (14C) METHOD OF STANDARDIZATION: The 3H bulk solution is standardized by liquid scintillation counting using SRM 4947C as the reference material. The total uncertainty of the assay is ± 1.6 % The 14C bulk solution is standardized by liquid scintillation counting using SRM 4222C as the reference material. The total uncertainty of the assay is ± 1.3 % The glass vial may have some imperfections due to the vendor's manufacturing process. These imperfections do not affect the performance of the product. I hereby certify that the above information is accurate. Approved by: Alex Wurzel Chemist Date: March 1, 2016 CN 8845270 PerkinElmer Inc. 549 Albany Street, Boston, MA 02118 USA Tel. No. 800-762-4000 Certificat de Radioactivité / Traçabilité Standard Toluène Non-quenché pour Scintillation Liquide PRODUIT: NUMÉRO DE PIÈCE: 6008500 DATE D'ANALYSE / RADIONUCLEIDES 3 H (Tritium) Toluene Carbon 14 Toluene 286200 0,129 112100 0,050 DATE D'EXPIRATION: DOSAGE DE VALEUR: Numéro de lot: NUMÉRO DE SÉRIE DPM/Std. + 3 % 23 février 2016 29 23 février 2016 29 µCi DPM/Std. + 2 % µCi 23 février 2021 Norme de référence: Institut National des Standards et de la technologie SRM 4947C (3H) et 4222C SRM (14C) PROCEDE DE NORMALISATION: La solution en vrac 3H est normalisée par comptage à scintillation liquide à l'aide 4947C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,6% La solution en vrac 14C est normalisée par comptage à scintillation liquide à l'aide 4222C SRM comme matériau de référence. L'incertitude totale de l'essai est de ± 1,3% Le flacon en verre peut avoir quelques imperfections dues au procédé de fabrication du fournisseur. Ces imperfections ne seront pas affecter les performances du produit. Je certifie que les renseignements ci-dessus sont exactes. Approuvé par: Alex Wurzel Date: 1 mars 2016 Chimiste CN 8845270