APAX-5013 APAX-5017 APAX-5018 8-ch RTD Module 12-ch Analog Input Module 12-ch Thermocouple Input Module NEW NEW RoHS APAX-5013 APAX-5017 COMPLIANT 2002/95/EC Specifications Dimensions (W x H x D) Weight Power Consumption (typical) Analog Input Channels Input Type Protection Isolation between 2,500 VDC Channels and Backplane Environment Operating Temperature -10 ~ 60° C (when mounted vertically) Storage Temperature -40 ~ 70° C Relative Humidity 5 ~ 95% (non-condensing) Ordering Information APAX-5013 APAX-5001 APAX-5002 APAX-5343E 2002/95/EC 8-ch RTD Module 1-slot Backplane Module 2-slot Backplane Module Power Supply for APAX Expansion Module 30 x 139 x 100 mm 170 g 4 W @ 24 VDC (APAX-5017) 3.5 W @ 24 VDC (APAX-5018) Analog Input 8 >10 MW Pt-100, Pt-200, Pt-500, Pt-1000, Balco, Ni 518 RTD (3-wire) Pt-100, Pt-200, Pt-500, Pt-1000: -120 ~ 130° C, -200 ~ 850° C Supports both IEC 60751 ITS90 (0.03851 W/W/° C) and JIS C 1604 (0.03916 W/W/° C) Balco 500: -30 ~ 120° C Ni 518: -80 ~ 100° C, 0 ~ 100° C Yes Configure Different Range for Each Channel Resolution 16-bit Sampling Rate 10 sample/second (total) CMR @ 50/60 Hz 90 dB NMR @ 50/60 Hz 60 dB Noise Rejection Hardware Filter (50/60 Hz) COMPLIANT General Dimensions (W x H x D) 30 x 139 x 100 mm Weight 170 g Power Consumption 2.5 W @ 24 VDC (typical) Temperature Range RoHS Specifications General Channels Input Impedance Input Type APAX-5018 12 (Differential) V, mV, mA (APAX-5017) V, mV, mA, Thermocouple (APAX-5018) Input Impedance Voltage: >10 MΩ (APAX-5017), >1 MΩ (APAX-5018) Current: 120Ω (Current) Voltage/Current Range ±150 mV, ±500 mV, ±1 V, ±5 V, ±10 V, ±20 mA, 0 ~ 20 mA, 4 ~ 20 mA (APAX-5017) ±50 mV, ±100 mV, ±500 mV, ±1 V, ±2.5 V, ±20 mA, 0 ~ 20 mA, 4 ~ 20 mA (APAX-5018) Temperature Range Type J (-210 ~ 1200° C), Type K (-270 ~ 1372° C), (APAX-5018 only) Type T (-270 ~ 400° C), Type E (-270 ~ 1000° C), Type R (0 ~ 1768° C), Type S (0 ~ 1768° C), Type B (300 ~ 1820° C) Configure Different Yes Range for Each Channel Resolution 16-bit with accuracy ±0.1% or better of Full Scale Range (Voltage), ±0.2% or better of Full Scale Range(Current) Note: Refer to manual for Thermocouple accuracy Sampling Rate 12 samples/second (total) Conversion Time 100 ms (per channel) CMR @ 50/60 Hz 92 dB NMR @ 50/60 Hz 60 dB Noise Rejection Hardware Filter (50/60 Hz) Wire Burn-out Detection Yes (APAX-5081 only, for Thermocouple) Protection Isolation between 2,500 VDC Channels and Backplane Supports Over Voltage Protection Environment Operating Temperature -10 ~ 60° C (when mounted vertically) Storage Temperature -40 ~ 70° C Relative Humidity 5 ~ 95% (non-condensing) Ordering Information APAX-5017 APAX-5018 APAX-5001 APAX-5002 APAX-5343E 12-ch Analog Input Module 12-ch Thermocouple Input Module 1-slot Backplane Module 2-slot Backplane Module Power Supply for APAX Expansion Module Online Download www.advantech.com/products