XP Power EME05US12 Ac-dc power supply Datasheet

EME05 Series
AC-DC Power Supplies
5 Watts
• Compact Size
• Medical Approvals
• Single Outputs from 3.3 to 48 V
• PCB Mount
• Encapsulated & Open Frame
• Class II
• Peak Load Capability
• 3 Year Warranty
POWER
GREEN
Dimensions:
The EME05 is a series of open frame and encapsulated AC-DC single output power supplies
designed for medical applications. The series provides two mechanical options including open
frame and encapsulated PCB mount. With approvals to world-wide medical safety standards,
compliance with class B for both conducted and radiated emissions and a 130%, 30s peak load
capability, these class II isolation parts benefit system designers with easy integration into the latest
healthcare products and applications.
EME05:
1.50 x 1.00 x 0.60” (36.1 x 25.4 x 15.2 mm)
EME05-P:
1.40 x 0.94 x 0.69” (35.6 x 23.7 x 17.6 mm)
Models & Ratings
Output Power
Output Voltage
Output Current
Nominal
Peak (1)
Model Number (2)
5W
3.3 VDC
1510 mA
1960 mA
EME05US03
5W
5.0 VDC
1000 mA
1300 mA
EME05US05
5W
9.0 VDC
555 mA
722 mA
EME05US09
5W
12.0 VDC
416 mA
541 mA
EME05US12
5W
15.0 VDC
333 mA
433 mA
EME05US15
5W
24.0 VDC
208 mA
270 mA
EME05US24
5W
36.0 VDC
138 mA
180 mA
EME05US36
5W
48.0 VDC
104 mA
135 mA
EME05US48
Notes
1. Peak load lasting <30 s with a maximum duty cycle of 10%, average output power not to exceed nominal.
2. For Open Frame version add suffix -P to model number, e.g. EME05US12-P.
Summary
Characteristic
Input Voltage Range
Minimum
85
No Load Input Power
1
Typical
Maximum
Units
Notes & Conditions
264
VAC
Derate from 100% at 90 VAC to 90% at 85 VAC
0.3
W
Efficiency
72
86
%
Operating Temperature
-25
+70
°C
Derate linearly from 100% at +50 °C to 50% at +70 °C
EMC
EN55011 Level B Conducted & Radiated, EN601000-3-2, EN61000-3-3, EN60601-1-2
Safety Approvals
EN60601-1, ANSI/AAMI ES60601-1. IEC60601-1
www.xppower.com
EME05 Series
AC-DC Power Supplies
Input
Characteristic
Input Voltage Range
Input Frequency
Input Current - Full Load
No Load Input Power
Inrush Current
Earth Leakage Current
Input Protection
Minimum
85
47
Typical
Units
VAC
Hz
A rms
W
A
Notes & Conditions
Notes & Conditions
±0.5
±1
Units
VDC
%
A
s
ms
ms
%
%
4
%
50
90
120
200
140
200
mV pk-pk
mV pk-pk
mV pk-pk
mV pk-pk
% Vnom
%
0.05
%/˚C
Maximum
Units
%
VAC
kHz
W/in3
kHrs
lb (g)
lb (g
Maximum
264
63
0.10/0.06
0.3
40
At 115/230 VAC
At 230 VAC
Class II construction no earth
Internal T1.0 A/250 VAC fuse fitted in line and neutral
Output
Characteristic
Output Voltage
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
Line Regulation
Load Regulation
Minimum
3.3
Typical
Maximum
48
±1
0
2
14
12
Transient Response
Ripple & Noise
Overvoltage Protection
Overload Protection
Short Circuit Protection
Temperature Coefficient
115
130
No minimum load required
at full load and 115 VAC
Deviation, recovery within 1% in less than 500 µs
for a 25% load change
3.3-5 V versions, 20 MHz bandwidth
9 V versions, 20 MHz bandwidth
12-15 V versions, 20 MHz bandwidth
24-48 V versions, 20 MHz bandwidth
Recycle input to reset
Trip & Restart (hiccup mode)
General
Characteristic
Efficiency
Isolation: Input to Output
Switching Frequency
Power Density
Mean Time Between Failure
Minimum
Typical
80
4000
130
5.5
>400
0.03 (14)
0.053 (24)
Weight
Notes & Conditions
2 MOPP
MIL-HDBK-217F, +25 °C GB
Open frame versions (-P)
Encapsulated version
Efficiency Graphs
EME05US12
EME05US24
90%
90%
80%
80%
y
100%
70%
70%
E
Efficiency
100%
60%
50%
40%
115Vac input
230Vac input
60%
50%
10
20
30
40
50
60
70
80
90
100
40%
10
20
Load (%)
30
40
50
60
70
80
90
100
Load (%)
www.xppower.com
2
EME05 Series
AC-DC Power Supplies
Environmental
Characteristic
Maximum
Units
Operating Temperature
Minimum
-25
Typical
+70
°C
Storage Temperature
-40
+85
°C
95
%RH
3048
m
Notes & Conditions
Derate linearly from 100% at +50 °C to 50% at +70 °C
Cooling
Convection-cooled
Humidity
Operating Altitude
Non-condensing
Shock
IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
Vibration
IEC68-2-6, 2 g, 10 Hz to 500 kHz, 10 mins/cycle, 60 mins each cycle
EMC: Emissions
Phenomenon
Standard
Test Level
Conducted
EN55011
Class B
Radiated
EN55011
Class B
Harmonic Current
EN61000-3-2
Voltage Flicker
EN61000-3-3
Criteria
Notes & Conditions
Class A
EMC: Immunity
Test Level
Criteria
Medical Device EMC
Phenomenon
IEC60601-1-2
Standard
Ed 4.0: 2014
As below
ESD
EN61000-4-2
±8kV contact,
±15kV air discharge
A
Radiated
EN61000-4-3
10 V/m
A
EFT
EN61000-4-4
3
A
Surge
EN61000-4-5
2
A
Conducted
EN61000-4-6
10 Vrms
A
Magnetic Fields
EN61000-4-8
30 A/m
A
70% UT (70 VAC) for 500 ms
A
40% UT (40 VAC) for 100 ms
A
<5% UT (0 VAC) for 10 ms
A
<5% UT (0 VAC) for 5000 ms
B
70% UT (168 VAC) for 500 ms
A
40% UT (96 VAC) for 100 ms
A
EN60601-1-2 (100 VAC)
Dips and Interruptions
EN60601-1-2 (240 VAC)
<5% UT (0 VAC) for 10 ms
A
<5% UT (0 VAC) for 5000 ms
B
Notes & Conditions
80% mod, 80 MHz - 2.7 GHz plus discrete
communcation proximity field frequencies
Load de-rated to 50%
Safety Approvals
3
Safety Agency
Safety Standard
Notes & Conditions
CB Report
IEC60601-1 Including Risk Management
Medical
UL
ANSI/AAMI ES60601-1:2005 & CSA C22.2, No.60601-1:08
Medical
TUV
EN60601-1
Medical
Means of Protection
Category
Primary to Secondary
2 x MOPP (Means of Patient Protection)
IEC60601-1 Ed 3.1
www.xppower.com
EME05 Series
AC-DC Power Supplies
Mechanical Details
Encapsulated
BOTTOM VIEW
1.0 (25.4)
0.20
(5.1)
0.75 (19.05)
0.10 (2.53)
2
1.50
(38.1)
0.60 (15.2)
1
4 Ø0.04
(Ø1.02)
1.30
(33.02)
0.10
(2.54)
3
4
0.10 (2.53)
0.5 (12.7)
Pin Connections
Open Frame (-P)
0.50
(12.7)
1.40 (35.6)
0.120 3.05
0.750 19.05
2
0.05 (1.28)
3
Pin
Single
1
ACN
2
ACL
3
-Vout
4
+Vout
0.94
(23.7)
4
1
0.37
(9.40)
1.30 (33.02)
0.51
(13.0) max
0.69
(17.6)
Ø0.039 (Ø1.00)
Ø0.071 (Ø1.80)
0.08 2.0 max
0.15
(3.8)
0.17
(4.4)
Notes
1. Dimensions in inches (mm).
2. Weight: Open frame versions (-P): 0.03 lbs (14 g)
Encapsulated: 0.053 lbs (24 g))
3. Tolerances:
www.xppower.com
x.xx = ± 0.02 (x.x = ± 0.5)
x.xxx = ± 0.01 (x.xx = ± 0.25)
10 Dec 15
Similar pages