Fuji EDSX5-95H Small flange remote seal type pressure transmitter Datasheet

SMALL FLANGE REMOTE SEAL TYPE
PRESSURE TRANSMITTER
FKW...5
DATA SHEET
The FCX – AIII small flange remote seal type pressure transmitter accurately measures gauge pressure and transmits
a proportional 4 to 20mA signal.
The transmitter utilizes a unique micromachined capacitance silicon sensor with state-of-the-art microprocessor
technology to provide exceptional performance and functionality.
Totally welded construction of the seals assures excellent
reliability in high temperature and highly corrosive process
conditions.
FEATURES
1. Directly connectable to 1-1/2in and 2in flanges
The transmitter is connectable to 1-1/2in and 2in pipes
without a reducer.
2. Connectable to 1/2in and 3/4in pipes
Use of direct mounting adapter allows the transmitter to
be connected to the following process.
1/2in and 3/4in flanges
Screw connection 1/2-14NPT, 3/4-14NPT, Rc1/2, Rc3/4
3. Minimum environmental influence
The “Floating Cell” design which protects the pressure
sensor against changes in temperature, and overpressure
substantially reduces total measurement error in actual
field applications.
4. Fuji/HART® bilingual communications protocol
FCX – AIII series transmitter offers bilingual communications to speak both Fuji proprietary protocol and HART®.
Any HART® compatible devices can communicate with
FCX – AIII.
5. Application flexibility
Various options that render the FCX – AIII suitable for
almost any process applications include:
– Full range of hazardous area approvals
– Built-in RFI filter and lightning arrester
– 5-digit LCD meter with engineering unit
– Stainless steel electronics housing
– Wide selection of materials
– High temperature, vacuum seals
6. Burnout current flexibility (Under Scale: 3.2 to 4.0mA,
Over Scale: 20.0 to 22.5mA)
Burnout signal level is adjustable using Model FXW Hand
Held Communicator (HHC) to comply with NAMUR NE43.
7. Dry calibration without reference pressure
Thanks to the best combination of unique construction
of mechanical parts (Sensor unit) and high performance
electronics circuit (Electronics unit), reliability of dry
calibration without reference pressure is at equal level
as wet calibration.
SPECIFICATIONS
Functional specifications
Service:
Liquid, gas, or vapour
Span, range, and overrange limit:
Span limit [kPa]{bar}
Type
Min.
F KW
2
F KW
3
F KW
4
50
{0.5}
300
{3}
1000
{10}
Max.
500
{5}
3000
{30}
10000
{100}
Range limit
[kPa]{bar}
−100 to +500
{−1 to +5}
−100 to +3000
{−1 to +30}
−100 to 10000
{−1 to 100}
Overrange
limit
[MPa] {bar}
1.5
{15}
4.5
{45}
15
{150}
–
Lower range limit (vacuum limit) ;
Silicone fill sensor: See Fig. 1
Fluorinated fill sensor: Atmospheric pressure
– Conversion factors to different units;
1MPa=103kPa=10bar=10.19716kgf/cm2=145.0377psi
1kPa=10mbar=101.9716mmH2O=4.01463inH2O
Output signal:
4 to 20mA DC with digital signal superimposed on the 4 to 20mA signal.
Power supply:
Transmitter operates on 10.5V to 45V DC
at transmitter terminals.
10.5V to 32V DC for the units with optional arrester.
EDSX5-95h
Date
Aug. 10, 2011
FKW...5
Load limitations: see figure below
Temperature limit:
Ambient: −15 to +65°C
(−15 to +60°C for arrester option)
(−10 to +60°C for fluorinated oil fill transmitter)
(−10 to +60°C for silicone oil “H”, “S”)
(−15 to + 45°C for capillary length more than 7m)
For explosionproof units (flameproof or intrinsic
safety), ambient temperature must be within the
limits specified by each standard.
Process:
[Ω]
Load resistance
2000
1500
1474Ω
R=
E-10.5
(Imax+0.9)x10-3
1000
577Ω
Operating
area
500
24V
250
0
45V
Fill fluid
10
16.35
10.5V
Power voltage (DC)
50
E[ V ]
Fluorinated oil W, A and D
Hazardous locations: SEE TABLE2
Zero/span adjustment:
Zero and span are adjustable from the
HHC(1). Zero and span are also adjustable
externally from the adjustment screw.
Damping:
Adjustable from HHC or local configurator
unit with LCD display.
The time constant is adjustable between
0.06 to 32 seconds.
Zero elevation/suppression:
Zero can be elevated or suppressed within
the specified range limit of each sensor
model.
Normal/reverse action:
Selectable from HHC(1)
Indication:
Analog indicator or 5-digit LCD meter, as
specified.
Burnout direction: Selectable from HHC(1)
If self-diagnostic detect transmitter failure, the analog signal will be driven to either “Output Hold”, “Output Overscale”
or “Output Underscale” modes.
“Output Hold”: Output signal is hold as the value just
before failure happens.
“Output Overscale”:
Adjustable within the range 20.0mA to
22.5mA from HHC(1)
“Output Underscale”:
Adjustable within the range 3.2mA to
4.0mA from HHC(1)
4
20
22.5 [mA]
Over scale
Burnout
Normal operating range
Under scale
Probable under range
Burnout
Probable over range
Output limits conforming to NAMUR NE43 by order.
Loop-check output:
Transmitter can be configured to provide
constant signal 3.2mA through 22.5mA by
HHC(1).
Silicone oil
−20 to 120°C
0 to 250°C
Y and G
−40 to 180°C
S
0 to 250°C
2
(1) HHC: Hand Held Communicator
Atmospheric
pressure
2.7kPa abs
{20mmHg
abs}
Storage: −40 to +70°C
Humidity limit:
0 to 100% RH
Communication: With HHC(1) (Model FXW, consult Data
Sheet No. EDS8-47), following items can
be remotely displayed or configured.
Note: HHC’s version must be higher than
7.0 (or FXW
1– 4), for FCX AIII.
Local configurator with LCD display (option):
Local confi gurator with 3 push button
and LCD display can support following
items.
Items
By communication By local configurator
with FXW
(with 3 push button)
Display
Set
Display
Set
Tag No.
v
v
v
v
Model No.
v
v
v
v
Serial No. & Software Version
v
—
v
—
Engineering unit
v
v
v
v
Range limit
v
—
v
—
Measuring range
v
v
v
v
Damping
v
v
v
v
Output mode
v
—
v
—
Burnout direction
v
v
v
v
Calibration
v
v
v
v
Output adjust
—
v
—
v
Data
v
—
v
—
Self diagnoses
v
—
v
—
Printer (In case of FXW with
printer option)
v
—
—
—
External switch lock
v
v
v
v
Transmitter display
v
v
v
v
Linearize
v
v
—
—
Rerange
v
v
v
v
Saturate current
v
v
v
v
Write protect
v
v
v
v
History
– Calibration history
– Ambient temperature history
v
v
v
—
v
v
v
—
EMC Conformity: EN61326-1: 2006
(Note)
Lower limit
of
static press.
Process
temperature
H
Note: For communication with HHC (Model: FXW), min. of 250 is required.
3.2
13th digit of
“Code symbols”
Performance specifications
Physical specifications
Reference conditions, silicone oil fill, SS316 isolating diaphragms, 4
to 20mA analog output in linear mode and capillary length 3m.
Accuracy rating: (including linearity, hysteresis, and repeatability)
(Standard)
For spans greater than 1/10 of URL: ±0.25% of span
For spans below 1/10 of URL:
x
± 0.17+0.08 0.1 URL % of span
Span
(Option) (Code; 21th digit H)
For spans greater than 1/10 of URL: ±0.1% of span
For spans below 1/10 of URL:
x
± 0.05+0.05 0.1 URL % of span
Span
(
)
(
)
Stability:
±0.2% of upper range limit (URL) for 10
years.
Temperature effect:
Effect per 28°C change between the limits of −15°C and +65°C
Zero shift: ±0.5%/28°C
(x equal to 1/6.5 URL or more)
URL
Zero shift; ± (0.5
)%/28°C
6.5 x x
(x less than 1/6.5 URL)
Total shift; ±0.75%/28°C
(x less than 1/6.5 URL or more)
URL
Total shift; ±(0.25 + 0.5
)%/28°C
6.5 x x
(x less than 1/6.5 URL)
Where, x:
Calibrated span
URL:
Maximum span (Upper Range
Limit)
Note : Above specifications are based on the
conditions that flange and sensor unit are
at the same temperature and in the same
level. If temperature is different at flange,
capillary or sensor unit, output variation
may increase.
Overrange effect: Zero shift; 0.2% of URL/(1.5 x URL)
Supply voltage effect:
Less than 0.005% of calibrated span per
1V
Update rate:
60 msec
Step response: Time constant: 0.3s (at 23°C)
Dead time: 0.12s
(without electrical damping)
Dielectric strength:
500V AC, 50/60Hz 1 min., between circuit
and earth.
Insulation resistance:
More than 100M/500V DC.
Internal resistance for external field indicator:
12 or less
Electrical connections:
G1/2, 1/2-14 NPT, Pg13.5, or M20 x 1.5
conduit, as specified.
Process connections:
JIS;
10K, 20K, 30K, 63K −40, 50A
10K, 20K, 30K, 63K −15, 20A (with Adapter)
ANSI/JPI;
150LB, 300LB, 600LB, −1 1/2", 2"
150LB, 300LB, 600LB, −1/2", 3/4" (with
Adapter)
Screw connection (with Adapter) ;
Rc1/2, Rc3/4, 1/2-14NPT, 3/4-14NPT
Diaphragm extension:
0, 50, 100, 150, or 200mm as specified.
(See model code. Extended diaphragm
is available only with 316L stainless steel
diaphragm)
Process-wetted parts material:
Diaphragm: 316L stainless steel, Hastelloy-C
Monel or Tantalum
Flange face: 316 stainless steel, Hastelloy-C
lining, Monel lining or Tantalum lining
Extension: 316 stainless steel
(Refer to “Code symbols”)
Non-wetted parts material:
Electronics housing: Low copper die-cast
aluminum alloy finished with polyester
coating (standard), or 316 stainless
steel (ASTM CF8M), as specified.
Capillary: In case of 11th code “D. E. L”,
PE armored stainless steel.
In case of 11th code “Q. R. S”, stainless steel armored stainless steel.
Mounting flange: 304 stainless steel or
carbon steel, as specified.
Fill fluid: Silicone oil (standard) or fluorinated oil
Mounting bracket: 304 stainless steel.
Environmental protection:
IEC IP67 and NEMA 6/6P
Mounting:
On 60.5mm (JIS 50A) pipe using mounting bracket, direct wall mounting
Mass {weight}:
Transmitter approximately 8.2 to 11.2kg
without options.
Add; 0.5kg for mounting bracket
4.5kg for stainless steel housing
option
1.5kg per 50mm extension of diaphragm
3
FKW...5
Optional features
ACCESSORIES
Indicator:
A plug-in analog indicator (2.5% accuracy)
An optional 5-digit LCD meter with engineering unit is also available.
Local configurator with LCD display:
An optional 5 digits LCD meter with 3
push buttons can support items as using
communication with FXW.
Arrester:
A built-in arrester protects the electronics
from lightning surges.
Lightning surge immunity: 4kV (1.2 ×
50µs).
Oxygen service: Special cleaning procedures are followed
throughout the process to maintain all
process wetted parts oil-free.
The fill fluid is fluorinated oil.
Chlorine service: Oil-free procedures as above. Includes
fluorinated oil for fill.
Degreasing:
Process-wetted parts are cleaned, but the
fill fluid is standard silicone oil. Not for use
on oxygen or chlorine measurement.
Vacuum and high temperature service:
Special silicone oil and filling procedure
are applied.
See Fig.1.
Optional tag plate:
An extra stainless steel tag for customer
tag data is wired to the transmitter.
Coating of cell: Cell's surface is finished with epoxy/polyurethane double coating. Specify if environment is extremely corrosive.
Filled liquid:
Silicone (Code Y,G)
[kPa abs]
{m bar abs}
Silicone (Code S)
Operating pressure
101
{1010}
53.3
{533}
Operating area
13.3
{133}
Non-operating
area
2.7
{27}
−40 0
60 85 120
180
Process temperature
230 250
[°C ]
Fig. 1 Relation between process temperature and
operating pressure
4
Hand-held communicator:
(Model FXW, refer to Data Sheet No.
EDS8-47)
CODE SYMBOLS
1 2 3 4 5 6 7 8
Digit
4
Description
Note
<Conduit connection>
Conduit connection Case type
G1/2
T type
1/2-14NPT
T type
Pg13.5
T type
M20×1.5
T type
G1/2
L type
1/2-14NPT
L type
Pg13.5
L type
M20×1.5
L type
5
<Flange>
Material
Size and rating
304 stainless
JIS 10K 40A
steel
JIS 10K 50A
JIS 20K 40A
JIS 20K 50A
JIS 30K 40A
JIS 30K 50A
JIS 63K 40A
JIS 63K 50A
ANSI/JPI 150LB 1 1/2"
ANSI/JPI 150LB 2"
ANSI/JPI 300LB 1 1/2"
ANSI/JPI 300LB 2"
ANSI/JPI 600LB 1 1/2"
ANSI/JPI 600LB 2"
Carbon steel
JIS 10K 40A
JIS 10K 50A
JIS 20K 40A
JIS 20K 50A
JIS 30K 40A
JIS 30K 50A
JIS 63K 40A
JIS 63K 50A
ANSI/JPI 150LB 1 1/2"
ANSI/JPI 150LB 2"
ANSI/JPI 300LB 1 1/2"
ANSI/JPI 300LB 2"
ANSI/JPI 600LB 1 1/2"
ANSI/JPI 600LB 2"
None
40A, 1 1/2B
(Wafer type)
50A, 2B
Note 1
Direct mounting adapter connection (* 1)
6
<Span limit [kPa] {bar} >
50.......500 ( 0.5...5)
300.....3000 ( 3...30)
1000...10000 (10...100)
7
<Material/diapharagm extension>
Diaphragm
Diaphragm extension (mm)
Flange face
316L stainless steel 316 stainless steel
0
Note 2
316L stainless steel 316 stainless steel
50
316L stainless steel 316 stainless steel
100
(* 2)
316L stainless steel 316 stainless steel
150
316L stainless steel 316 stainless steel
200
316L stainless steel 316 stainless steel
0
+Au coating
Hastelloy-C
0
Hastelloy-C
Hastelloy-C
0
316 stainless steel
Monel
0
Monel
Tantalum
0
Tantalum
Note1: (*1) Direct mounting adapter type is specified at 16th to 20th digit.
Direct mounting adapter is available only for 7th digit code "V".
Note2: (*2) Diaphragm extension is available only for 2" (50A) flanges.
F KW
5
9 10 11 12 13
14 15 16 17 18 19 20
0
21
Digit No.
of code
5
6
7
8
S
T
V
W
0
1
2
3
4
5
6
7
A
B
C
D
E
F
G
H
J
K
L
M
N
P
Q
R
S
T
U
V
W
X
Y
2
3
4
V
A
B
C
D
J
H
W
M
T
5
FKW...5
1 2 3 4 5 6 7 8
Digit
9
10
11
12
13
14
15
Description
<Indicator and arrester>
Indicator
Arrester
None
None
Analog, 0 to 100% linear scale
None
Analog, custom scale
None
None
Yes
Analog, 0 to 100% linear scale
Yes
Analog, custom scale
Yes
Digital, 0 to 100% linear scale
None
Digital, custom scale
None
Digital, 0 to 100% linear scale
Yes
Digital, custom scale
Yes
Digital, 0 to 100% linear scale
(Local configurator unit with LCD display)
None
Digital, custom scale
(Local configurator unit with LCD display)
None
Digital, 0 to 100% linear scale
(Local configurator unit with LCD display)
Yes
Digital, custom scale
(Local configurator unit with LCD display)
Yes
<Approvals for hazardous locations>
None (for ordinary locations)
TIIS, Flameproof (Cable gland seal) (*7)
TIIS, Intrinsic safety
FM, Flameproof (or explosionproof) (*10)
FM, Intrinsic safety and nonincentive
FM Combined of flameproof and intrinsic safety (*10)
ATEX Flameproof (*9)
ATEX Intrinsic safety
ATEX Type n
ATEX Combined of flameproof and intrinsic safety (*9)
IECEx Scheme, Flameproof (*9)
IECEx Scheme, Intrinsic safety
CSA, Flameproof (or explosionproof) (*10)
CSA, Intrinsic safety and nonincentive
NEPSI, Flameproof (or exprosionproof)
NEPSI, Intrinsic safety
NEPSI, Combined of flameproof and intrinsic safety
<Capillary and mounting bracket>
Mounting bracket
Capillary
Armor of Capillary
Stainless steel
1.5m
PE
Stainless steel
3m
PE
Stainless steel
5m
PE
(*5)
Stainless steel
6m
PE
Stainless steel
7m
PE
Stainless steel
8m
PE
Stainless steel
10m
PE
Stainless steel
1.5m
Stainless steel
Stainless steel
3m
Stainless steel
Stainless steel
5m
Stainless steel
Stainless steel
6m
Stainless steel
Stainless steel
7m
Stainless steel
Stainless steel
8m
Stainless steel
Stainless steel
10m
Stainless steel
<Options>
Stainless steel elec. housing Coating of cell
Extra SS tag plate
None
None
None
Yes
None
None
None (*3)
None
Yes
Yes
None
Yes
None
Yes
Yes
(*6)
Yes
Yes
Yes
<Special applications and fill fluid>
Treatment
Fill fluid
Standard
Silicone oil
Standard
Fluorinated oil
Degreasing
Silicone oil
Fluorinated oil (7th digit code "V", "W", "A", "B", "C" and "D")
Oxygen service
Chlorine service
Fluorinated oil (7th digit code "H" and "T")
High temp. 250°C
Silicone oil (7th digit code "V", "W", "A", "B", "C", "D" and "H")
High temp. and vacuum (250°C) Silicone oil (7th digit code "V", "W", "A", "B", "C" and "D")
<Teflon membrane>
None
Yes (Available for 7th digit code "V", "J", "W.
Not available for 5th digit code "Y" and 13th digit code "H", "S".)
<Bolt/nut> (*4)
None
6th digit code "2", "3"
Cr-Mo alloy hexagon socket head cap bolt/
carbon steel nut
6th digit code "4"
Cr-Mo alloy hexagon bolt/carbon steel nut
304 stainless steel bolt/304 stainless steel nut
316 stainless steel bolt/316 stainless steel nut
Note3: (*3)
Note4: (*4)
Note5: (*5)
Note6: (*6)
6
Note
F KW
9 10 11 12 13
14 15 16 17 18 19 20
5
21
Digit No.
of code
A
B
D
E
F
H
L
P
Q
S
1
2
4
5
A
C
G
D
H
V
X
K
P
M
R
T
E
J
F
S
U
Note 7
Note 10
Note 10
Note 9
Note 9
Note 9
Note 10
D
E
L
F
M
N
P
Q
R
S
T
V
W
X
Note 5
Y
B
M
N
P
Q
Note 3
Note 6
Note 6
Y
W
G
A
D
H
S
Note 11
Note 11
Y
C
Note 4
Customer tag number can be engraved on standard stainless steel
name plate. If extra tag plate is required, select "Yes".
In case of tropical use, select stainless bolts and nuts.
Available for 13th digit code "Y, W, G, A, D".
Not available for 10th digit code "C".
Y
A
B
E
U
Note7:
Note9:
Note10:
Note11:
(*7)
(*9)
(*10)
(*11)
Available for 4th digit code "S".
Available for 4th digit code "6", "8", "T", "W".
Available for 4th digit code "6", "T".
Treatment: Standard.
Specifications of connecting pipeless adapter {for 15, 20A (1/2, 3/4") connection} and others
Note 1. When odering the instrument with connecting pipeless adapter, specify “Y” in the 5th digit of Code Symbol, and
specify 16th digit to 20th digit.
When odering the instrument without connecting pipeless adapter, nothing should be filled in the 16th to 20th digit.
2. Unless otherwise described in the specifications, leave the 21st digit blank.
1 2 3 4 5 6 7 8
Digit
Description
Note
16, 17 <Process connection (connecting pipeless adapter)>
JIS 10K 15A
JIS 10K 20A
JIS 20K 15A
JIS 20K 20A
JIS 30K 15A
JIS 30K 20A
JIS 63K 15A
JIS 63K 20A
ANSI/JPI 150LB 1/2"
ANSI/JPI 150LB 3/4"
ANSI/JPI 300LB 1/2"
ANSI/JPI 300LB 3/4"
ANSI/JPI 600LB 1/2"
ANSI/JPI 600LB 3/4"
Screw connection Rc1/2
Screw connection Rc3/4
Screw connection Rc1/2 - 14NPT
Screw connection Rc3/4 - 14NPT
18
<Material (connecting pipeless adapter)>
Note 12
Adapter
Bolts/nuts
(* 12)
316 Stainless Steel
Cr-Mo steel/carbon steel
19
<Vent/drain (for connecting pipeless adapter)>
Standard
Long type
20
<Gasket (for connecting pipeless adapter)>
Standard (Teflon)(Only Y, W, G, A and D can be specified on 13th digit).
For high temperature (spiral gasket) (Only H and S can be specified on
13th digit).
Note 13
21
<Other options> (* 13)
High accuracy type
Instruction manual attached
Instruction manual unattached
High accuracy type
Instruction manual unattached
F KW
9 10 11 12 13
14 15 16 17 18 19 20
21
5
Digit No.
of code
1 1
1 2
2 1
2 2
3 1
3 2
6 1
6 2
1H
1T
2H
2T
4H
4T
SR
S2
SN
ST
W
A
N
1
2
H
L
T
Note12: (* 12) For connection of transmitter receiving pressure unit and connecting pipeless adapter
Note13: (* 13) In other option is not necessary, 21st digit code is blank.
In case of 21st digit code is blank, instruction manual attached.
7
FKW...5
ORDERING INFOMATION
When ordering this instrument, specify.
1. CODE SYMBOLS
2. Measuring range.
3. Output orientation (burnout direction) when abnormality is occurred in the transmitter.
Hold / Overscale / Underscale
Unless otherwise specified, output hold function is supplied.
4. Indication method (indicated value and unit) in case of the actual scale (code D, H, P, S 2,5 on 9th digit).
5. Tag No. (up to 14 alphanumerical characters), if required.
8
OUTLINE DIAGRAM (Unit:mm)
<AMP. case: L type>
IN CASE OF 4TH DIGIT CODE “S, V, W”
ZERO ADJ. SCREW
φ81
8
IN CASE OF 4TH DIGIT CODE “T”
103
φ81
103
10
MTG. PIPE JIS 50A (2B)
38
57
“B” CONDUIT CONN.
37
U-BOLT M8
172
(21)
WITH ANALOG IND.
(11)
4 HOLES 9 DIA.
WITH DIGITAL IND.
N HOLES h DIA
φ
E
P
φ48±1
φF
φ49
φD
INNER OF GASKET
FLANGE
20
L
t
163
DETAIL “B”
(CONDUIT. CONN.)
126
41
φ15
J
S (m)
1.5
3
5
6
7
8
10
EARTH TERMINAL
198
K
M
11th digit of the
code symbols
D, Q
E, R
L, S
F, T
M, V
N, W
P, X
43
SEE TABLE1
conduit conn.
J
K
M
7th digit of the
code symbols
L
V, W, J, H, M, T
A
B
C
D
0
50
100
150
200
S
G1/2
18
2
T
1/2-14NPT
16
4
V
Pg13.5
W
M20×1.5
10.5 4.5
2-M8×12
S
4
TABLE 1
φD
φE
0, G
1, H
2, J
3, K
4, L
5, M
6, N
7, P
A, Q
B, R
C, S
D, T
E, U
F, V
140
155
140
155
160
165
175
185
127
152
156
165
156
165
105
120
105
120
120
130
130
145
95.4
120.6
114.3
127
114.3
127
CorrosionStandard
resistant
materials
specification materials
specification
Note 1
Note 2
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
t
P
N-φh
FLANGE
36
36
38
38
42
42
52
54
37.5
39.5
41
42.5
42.5
45.5
60
71
60
71
60
71
60
71
60
71
60
71
60
71
4-19
4-19
4-19
8-19
4-23
8-19
4-25
8-23
4-16
4-20
4-23
8-20
4-23
8-20
JIS-10K-40A
JIS-10K-50A
JIS-20K-40A
JIS-20K-50A
JIS-30K-40A
JIS-30K-50A
JIS-63K-40A
JIS-63K-50A
ANSI/JPI-150LB-1 1/2B
ANSI/JPI-150LB-2B
ANSI/JPI-300LB-1 1/2B
ANSI/JPI-300LB-2B
ANSI/JPI-600LB-1 1/2B
ANSI/JPI-600LB-2B
Note1) In case of the 7th digit of type code is V, A, B, C, D, J or W.
Note2) In case of the 7th digit of type code is H, M, T.
CONNECTION DIAGRAM
<SS TAG PLATE>
2-M6 SCREW DEPTH 9
INNER OF
GASKET
φF
5th digit of the
code symbols
H.P.
62
70
120
φF
Q
16
MASS. APPROX.
(kg)
8.2
9.2
10.2
10.7
11.2
φ49
4th digit of the
code symbols
20
<WAFER TYPE>
40A, 1 1/2B
50A, 2B
OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN)
10th digit code C (Cable gland type)
NOTE1)
G1/2
DETAIL “C”
φ33.5
φ13
−/CK− CK+
75
19
106
20
17
+
Q
60
71
"C" CONDUIT CONN.
G1/2
CABLE GLAND
NOTE1) IN CASE OF 10TH CODE C , φ11 CABLE IS SUITBLE.
9
FKW...5
<AMP. case: T type>
IN CASE OF 4TH DIGIT CODE “5, 7, 8”
IN CASE OF 4TH DIGIT CODE “6”
EARTH TERMINAL
8
10
MTG. PIPE JIS 50A (2B)
φ81
100
100
38
57
φ81
ZERO ADJ. SCREW
(21)
37
WITH ANALOG IND.
(11)
U-BOLT M8
185
4 HOLES 9 DIA.
WITH DIGITAL IND.
“B” CONDUIT CONN.
P
φ48±1
φF
φ49
φD
φ
E
N HOLES h DIA
INNER OF GASKET
FLANGE
20
L
185
t
127
20
DETAIL “B”
(CONDUIT. CONN.)
φ15
K
J
S (m)
226
M
11th digit of the
code symbols
D, Q
E, R
L, S
F, T
M, V
N, W
P, X
1.5
3
5
6
7
8
10
SEE TABLE1
7th digit of the
code symbols
L
V, W, J, H, M, T
A
B
C
D
0
50
100
150
200
5
G1/2
18
2
6
1/2-14NPT
16
4
7
Pg13.5
8
M20×1.5
10.5 4.5
16
MASS. APPROX.
(kg)
8.2
9.2
10.2
10.7
11.2
2-M8×12
S
H.P.
62
70
120
4
TABLE 1
φD
φE
0, G
1, H
2, J
3, K
4, L
5, M
6, N
7, P
A, Q
B, R
C, S
D, T
E, U
F, V
140
155
140
155
160
165
175
185
127
152
156
165
156
165
105
120
105
120
120
130
130
145
95.4
120.6
114.3
127
114.3
127
φF
CorrosionStandard
resistant
materials
specification materials
specification
Note 1
Note 2
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
73
84
100
84
t
P
N-φh
FLANGE
36
36
38
38
42
42
52
54
37.5
39.5
41
42.5
42.5
45.5
60
71
60
71
60
71
60
71
60
71
60
71
60
71
4-19
4-19
4-19
8-19
4-23
8-19
4-25
8-23
4-16
4-20
4-23
8-20
4-23
8-20
JIS-10K-40A
JIS-10K-50A
JIS-20K-40A
JIS-20K-50A
JIS-30K-40A
JIS-30K-50A
JIS-63K-40A
JIS-63K-50A
ANSI/JPI-150LB-1 1/2B
ANSI/JPI-150LB-2B
ANSI/JPI-300LB-1 1/2B
ANSI/JPI-300LB-2B
ANSI/JPI-600LB-1 1/2B
ANSI/JPI-600LB-2B
Note1) In case of the 7th digit of type code is V, A, B, C, D, J or W.
Note2) In case of the 7th digit of type code is H, M, T.
CONNECTION DIAGRAM
−/CK− CK+
75
+
<SS TAG PLATE>
20
10
2-M6 SCREW DEPTH 9
INNER OF
GASKET
φF
5th digit of the
code symbols
Q
conduit conn.
J
K
M
φ49
4th digit of the
code symbols
20
<WAFER TYPE>
40A, 1 1/2B
50A, 2B
Q
60
71
<AMP. case: L type> (with direct mount adaptor) screw connection type
10
IN CASE OF 4TH DIGIT CODE “S, V, W”
103
φ81
103
8
IN CASE OF 4TH DIGIT CODE “T”
φ81
ZERO ADJ. SCREW
MTG. PIPE JIS 50A (2B)
38
U-BOLT M8
57
37
“B” CONDUIT CONN.
172
(21)
(11)
4 HOLES 9 DIA.
WITH ANALOG IND.
S
φ84
WITH DIGITAL IND.
25
HIGH PRESS. CONN. D
VENT/DRAIN PLUG
20
23
163
126
DETAIL “B”
(CONDUIT. CONN.)
41
43
M
φ15
K
EARTH TERMINAL
198
J
SEE TABLE1
4th digit of the
code symbols
conduit conn.
J
K
M
G1/2
18
2
11th digit of the
S (m)
code symbols
4
D, Q
E, R
L, S
V
Pg13.5
10.5
4.5
M, V
7
W
M20×1.5
16
4
N, W
8
P, X
10
S
1/2-14NPT 16
T
1.5
3
5
F, T
6
2-M8×12
62
70
120
TABLE 1
16th digit of the 17th digit of the
code symbols
code symbols
Press. conn. “D”
S
R
Rc 1/2
S
N
1/2-14NPT
S
2
Rc 3/4
S
T
3/4-14NPT
OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN)
CONNECTION DIAGRAM
<SS TAG PLATE>
10th digit code C (Cable gland type)
NOTE1)
G1/2
φ13
φ33.5
−/CK− CK+
75
19
106
20
17
+
DETAIL “C”
"C" CONDUIT CONN.
G1/2
CABLE GLAND
NOTE1) IN CASE OF 10TH CODE C , φ11 CABLE IS SUITBLE.
11
FKW...5
<AMP. case: T type> (with direct mount adaptor) screw connection type
10
MTG. PIPE JIS 50A (2B)
EARTH TERMINAL
8
IN CASE OF 4TH DIGIT CODE “5, 7, 8”
IN CASE OF 4TH DIGIT CODE “6”
38
U-BOLT M8
57
(21)
“B” CONDUIT CONN.
37
φ81
φ81
100
100
ZERO ADJ. SCREW
WITH ANALOG IND.
(11)
185
4 HOLES 9 DIA.
WITH DIGITAL IND.
S
25
HIGH PRESS. CONN. D
VENT/DRAIN PLUG
20
23
185
DETAIL “B”
(CONDUIT. CONN.)
127
20
226
K
M
φ15
J
SEE TABLE1
4th digit of the
code symbols
5
6
conduit conn.
J
K
M
G1/2
18
2
1/2-14NPT 16
4
11th digit of the
S (m)
code symbols
7
Pg13.5
10.5
4.5
8
M20×1.5
16
4
D, Q
E, R
L, S
1.5
3
5
F, T
6
M, V
7
N, W
8
P, X
10
TABLE 1
16th digit of the 17th digit of the
code symbols
code symbols
Press. conn. “D”
S
R
Rc 1/2
S
N
1/2-14NPT
S
2
Rc 3/4
S
T
3/4-14NPT
CONNECTION DIAGRAM
−/CK− CK+
75
+
<SS TAG PLATE>
20
12
2-M8×12
62
70
120
<AMP. case: L type> (with direct mount adaptor)
IN CASE OF 4TH DIGIT CODE “S, V, W”
38
57
U-BOLT M8
37
172
(21)
(11)
φ81
16.4011
IN CASE OF 4TH DIGIT CODE “T”
103
φ81
103
MTG. PIPE JIS 50A (2B)
8
10
ZERO ADJ. SCREW
“B” CONDUIT CONN.
WITH ANALOG IND.
4 HOLES 9 DIA.
S
WITH DIGITAL IND.
N HOLES h DIA.
φ84
t
25
φF φE φD
80
20
163
126
VENT/DRAIN PLUG
41
43
DETAIL “B”
(CONDUIT. CONN.)
EARTH TERMINAL
K
198
M
φ15
J
SEE TABLE1
2-M8×12
16th digit of the
code symbols
17th digit of the
code symbols
φD
φE
φF
t
N-φh
1
1
95
70
51
12
4-15
JIS-10K-15A
1
2
100
75
56
14
4-15
JIS-10K-20A
2
1
95
70
51
14
4-15
JIS-20K-15A
2
2
100
75
56
16
4-15
JIS-20K-20A
3
1
115
80
55
18
4-19
JIS-30K-15A
3
2
120
85
60
18
4-19
JIS-30K-20A
6
1
120
85
55
23
4-19
JIS-63K-15A
6
2
135
95
60
25
4-23
JIS-63K-20A
1
H
89
60.3
34.9
11.5
4-16
ANSI/JPI-150LB 1/2B
1
T
98
69.9
42.9
13
4-16
ANSI/JPI-150LB 3/4B
2
H
95
66.7
34.9
14.5
4-16
ANSI/JPI-300LB 1/2B
2
T
117
82.5
42.9
16
4-20
ANSI/JPI-300LB 3/4B
4
H
95
66.7
34.9
14.5
4-16
ANSI/JPI-600LB 1/2B
4
T
117
82.5
42.9
16
4-20
ANSI/JPI-600LB 3/4B
FLANGE
62
70
120
4th digit of the
code symbols
conduit conn.
J
K
G1/2
S
18
M
11th digit of the
code symbols
2
T
1/2-14NPT
16
4
V
Pg13.5
10.5
4.5
W
M20×1.5
16
4
D, Q
S (m)
1.5
E, R
3
L, S
5
F, T
6
M, V
7
N, W
8
P, X
10
TABLE 1
OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN)
CONNECTION DIAGRAM
<SS TAG PLATE>
10th digit code C (Cable gland type)
NOTE1)
G1/2
φ13
φ33.5
−/CK− CK+
75
19
106
20
17
+
DETAIL “C”
"C" CONDUIT CONN.
G1/2
CABLE GLAND
NOTE1) IN CASE OF 10TH CODE C , φ11 CABLE IS SUITBLE.
13
FKW...5
<AMP. case: T type> (with direct mount adaptor)
10
MTG. PIPE JIS 50A (2B)
EARTH TERMINAL
8
IN CASE OF 4TH DIGIT CODE “5, 7, 8”
IN CASE OF 4TH DIGIT CODE “6”
(21)
(11)
38
U-BOLT M8
57
φ81
φ81
100
100
ZERO ADJ. SCREW
37
WITH ANALOG IND.
185
"B" CONDUIT CONN.
WITH DIGITAL IND.
4 HOLES 9 DIA.
S
N HOLES h DIA.
φ84
t
25
φF φE φD
20
185
23
80
127
20
VENT/DRAIN PLUG
DETAIL “B”
(CONDUIT. CONN.)
K
M
226
φ15
J
SEE TABLE1
2-M8×12
16th digit of the
code symbols
17th digit of the
code symbols
φD
φE
φF
t
N-φh
1
1
95
70
51
12
4-15
JIS-10K-15A
1
2
100
75
56
14
4-15
JIS-10K-20A
FLANGE
2
1
95
70
51
14
4-15
JIS-20K-15A
2
2
100
75
56
16
4-15
JIS-20K-20A
3
1
115
80
55
18
4-19
JIS-30K-15A
3
2
120
85
60
18
4-19
JIS-30K-20A
6
1
120
85
55
23
4-19
JIS-63K-15A
6
2
135
95
60
25
4-23
JIS-63K-20A
1
H
89
60.3
34.9
11.5
4-16
ANSI/JPI-150LB 1/2B
1
T
98
69.9
42.9
13
4-16
ANSI/JPI-150LB 3/4B
2
H
95
66.7
34.9
14.5
4-16
ANSI/JPI-300LB 1/2B
2
T
117
82.5
42.9
16
4-20
ANSI/JPI-300LB 3/4B
4
H
95
66.7
34.9
14.5
4-16
ANSI/JPI-600LB 1/2B
4
T
117
82.5
42.9
16
4-20
ANSI/JPI-600LB 3/4B
62
70
120
4th digit of the
code symbols
5
conduit conn.
J
G1/2
−/CK− CK+
75
+
<SS TAG PLATE>
20
14
18
M
2
6
1/2-14NPT
16
4
7
Pg13.5
10.5
4.5
8
M20×1.5
16
4
TABLE 1
CONNECTION DIAGRAM
K
11th digit of the
code symbols
D, Q
S (m)
1.5
E, R
3
L, S
5
F, T
6
M, V
7
N, W
8
P, X
10
TABLE 2
Authorities
ATEX
Authorities
Intrinsic safety
Ex II 1 G
Ex ia IIC T5 Tamb = –40°C to +50°C
Ex ia IIC T4 Tamb = –40°C to +70°C
Entity Parameters:
Ui=28V, Ii=94.3mA, Pi=0.66W,
Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator),
Ci=36nF (With Arrester), Li=0.7mH (With analog indicator)
Factory
Mutual
13th digit
Y,G,H,S
Y,G,H,S
Y,G,H,S
Y,G,H,S
W,A,D
Tamb
Factory
Mutual
Class I
Div.1 Groups B, C, D
T6 Type 4X
Class II III
Div.1 Groups E, F, G
T6 Type 4X
Tamb max = +60°C
CSA
Class I
Div.1 Groups C, D
Class II
Div.1 Groups E, F, G
Class III
Div.1
Note) “Seal Not Required” enclosure is allowed.
Class I
Div.1 Groups A, B, C, D
Class II
Div.1 Groups E, F, G
Class III
Div.1
Temp Code T5
Tamb max = +50°C
Temp Code T4
Tamb max = +70°C
Entity Parameters:
Vmax=28V, Imax=94.3mA, Ci=25nF (Without Arrester),
Ci=36nF (With Arrester),Li=0.6mH (Without analog meter),
Li=0.7mH (With analog meter)
TIIS
Ex ia IIC T4
Tamb max = +60°C
Entity Parameters:
Ui=28V, Ii=94.3mA, Pi=0.66W,
Ci=38.4nF, Li=0.694mH
IECEx
Scheme
Ex ia IIC T4
Tamb = –40°C to +70°C
Ex ia IIC T5
Tamb = –40°C to +50°C
Entity Parameters:
Ui=28V, Ii=94.3mA, Pi=0.66W,
Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator),
Ci=36nF (With Arrester), Li=0.7mH (With analog indicator)
NEPSI
Ex II 2 GD
Ex d IIC T6 IP66/67 T85°C
Tamb = −40°C to +65°C
Ex d IIC T5 IP66/67 T100°C
Tamb = −40°C to +85°C
−40°C to +85°C
−20°C to +80°C
−20°C to +60°C
−40°C to +60°C
−10°C to +60°C
Entity Parameters:
Vmax=78V, Imax=94.3mA, Pi=0.66W,
Ci=35.98nF, Li=0.694mH
CSA
ATEX
Class I II III
Div.1 Groups A, B, C, D, E, F, G
T4 Entity Type 4X
Model code
9th digit
A,B,D
L,P,1,2
Q,S,4,5
E,F,H
−
TIIS
Ex do IIB+H2 T4
Tamb max = +55°C
Maximum process temp. = +120°C
IECEx
Scheme
Ex d IIC T5 IP66/67
Tamb = −40°C to +85°C
Ex d IIC T6 IP66/67
Tamb = −40°C to +65°C
NEPSI
Ex d IIB+H2 T6
Tamb = −40°C to +60°C
13th digit
Y,G,H,S
Y,G,H,S
Y,G,H,S
Y,G,H,S
A,W,D
ATEX
Ex II 3 GD
EEx nL IIC T5 Tamb = −40°C to +50°C
EEx nL IIC T4 Tamb = −40°C to +70°C
Specific Parameters:
Model without arrester:
Ui=42.4V, Ii=113mA, Pi=1W,
Ci=25.18nF, Li=0.694mH
Model with arrester:
Ui=32V, Ii=113mA, Pi=1W,
Ci=35.98nF, Li=0.694mH
EEx nAL IIC T5 Tamb = −40°C to +50°C
EEx nAL IIC T4 Tamb = −40°C to +70°C
Specific Parameters:
Model without arrester:
Umax=42.4V, Imax=113mA, Pmax=1W,
Model with arrester:
Umax=32V, Imax=113mA, Pmax=1W
Tamb
−40°C to +85°C
−20°C to +80°C
−20°C to +60°C
−40°C to +60°C
−10°C to +60°C
Type n
Nonincendive
Authorities
Ex ia IIC T4
Ex d IIB+H2 T6 / Ex ia IIC T4
Model code
9th digit
A,B,D
L,P,1,2
Q,S,4,5
E,F,H
−
Flameproof
Factory
Mutual
(pending)
Class I II III
Div.2 Groups A, B, C, D, F, G
T4 Entity Type 4X
Model code
13th digit
9th digit
Entity Parameters:
Ui=42.4V, Ii=113mA, Pi=1W,
Ci=35.98nF, Li=0.694mH
A,B,D
Y,G,H,S
−40°C to +85°C
L,P,1,2
Y,G,H,S
−20°C to +80°C
Q,S,4,5
Y,G,H,S
−20°C to +60°C
E,F,H
Y,G,H,S
−40°C to +60°C
W,A,D
−10°C to +60°C
−
CSA
Tamb
Class I
Div.2 Groups A, B, C, D
Class II
Div.2 Groups E, F, G
Class III
Div.2
Temp Code T5
Tamb max = +50°C
Temp Code T4
Tamb max = +70°C
Entity Parameters:
Vmax=28V, Ci=25.18nF (Without Arrester),
Ci=35.98nF (With Arrester), Li=0.694mH
15
FKW...5
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan
Similar pages